Automating Gradient Method Development in Flash Chromatography For Greater Productivity and Minimizing Solvent Use

Similar documents
Scott Anderson, Kathy Lawrence Dennis McCreary, Melissa Wilcox. Grace Davison Discovery Sciences

Optimized Flash Chromatography Purification: From TLC to large scale in three steps

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Basic Principles for Purification Using Supercritical Fluid Chromatography

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience.

Biotage 2012 Webinar series

Minimizing Solvent Impact on Purification of Nitrogencontaining

Flash chromatography purification of high pka organic compounds with Teledyne Isco s specialty RediSep columns

Chromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Alltech Alltima HP Introduction

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Fast methods for the determination of ibuprofen in drug products

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

Introduction to Capillary GC

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

ACD/AutoChrom Assisted Method Development for Challenging Separations. Vera Leshchinskaya February 7, 2018

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

LC Technical Information

RediSep Rf C18 Flash Column Loading Techniques

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Materials for Pharmaceutical Manufacturing

Strategies for the Flash Purification of Highly Polar Compounds

Do you need trouble-free speed? The NEW Chromolith HR

Peptide Isolation Using the Prep 150 LC System

Lab scale isolation of compounds by Flash Chromatography

Determination of trace anions in concentrated hydrofluoric acid

CHROMATOGRAPHY, AND MASS SPECTRAL. FRACTIONS OF Lasianthus

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Varian Galaxie Chromatography Data System for Preparative HPLC

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis

Gemini-NX Performance paired with Axia Technology

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application

Do you need trouble-free speed? The NEW Chromolith HR

Experiment 1: Thin Layer Chromatography

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Polymer analysis by GPC-SEC. Technical Note. Introduction

Chromatography: Thin-Layer Chromatography (TLC) & Column Chromatography

TLC Introduction. tlc Choose Stationary Phase. Select Visualization Technique. Choose a Mobile Phase

Expediting Achiral SFC Method Development Using a Multi-Channel SFC System with MS Detection

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

HPLC Background Chem 250 F 2008 Page 1 of 24

Open Column Chromatography, GC, TLC, and HPLC

Optical Isomer Separation Columns and Packing Materials

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Laboratory 1 CSI Meadville: Preparative purification of a mixture

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Lab 10 Guide: Column Chromatography (Nov 3 9)

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Comparison of different aqueous mobile phase HPLC techniques

LUMEFANTRINUM LUMEFANTRINE

LC and LC/MS Column Selection Flow Chart

[ a ppl ic at ion no t e ]

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Abstract. Introduction

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

for Acclaim Mixed-Mode HILIC-1 Column

Chapter 23 Introduction to Analytical Separations

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

ADVANTAME. Not less than 97.0% and not more than 102.0% on the anhydrous basis. Sweetener, flavour enhancer

Chiral Separation Using SFC and HPLC

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

BRIEFING. (EXC: K. Moore.) RTS C Propylparaben C 10 H 12 O Benzoic acid, 4 hydroxy, propyl ester; Propyl p hydroxybenzoate [ ].

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

Automatization for development of HPLC methods

Chromatography Extraction and purification of Chlorophyll CHM 220

Prelab Reading Assignment: Laboratory Techniques in Organic Chemistry, 4 th Ed. Chapter 19

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

EXPERIMENT #1 SEPARATION AND RECOVERY OF ORGANIC COMPOUNDS, THIN LAYER CHROMATOGRAPHY, COLUMN CHROMATOGRAPHY, CRYSTALLIZATION AND MELTING POINTS

The Nanostream (Pasadena, CA) Veloce system,

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Protein Sequencer PPSQ-31A / 33A C297-E055B

Hypersil BDS Columns TG 01-05

Chromatography. Gas Chromatography

Page 1. Electronic Supplementary Material (ESI) for Green Chemistry This journal is The Royal Society of Chemistry 2012

THIN LAYER CHROMATOGRAPHY

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

This method describes the identification of the following prohibited colorants in cosmetic products:

SPE Introduction. general chromatography. solid phase extraction. How to Choose an SPE Product. Solid-Phase Processing Methods

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

Experiment Nine Thin Layer Chromatography

High Performance Liquid Chromatography

Automated LC Method Development: Fundamentals and Columns, Hardware, and Software

Liquid Chromatography

Dionex IonPac AS28-Fast-4µm column

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

ACD/Labs Software Impurity Resolution Management. Presented by Peter Russell

Transcription:

Automating Gradient in Flash For Greater Productivity and Minimizing Solvent Use Melissa Wilcox, Rakesh Bose, Scott Anderson, Kathy Lawrence Grace Davison Discovery Sciences 251 Waukegan Rd. Deerfield IL 615 U.S.A. Phone: 1-8-255-8324 Website: www.discoverysciences.com PP248 Grace Davison Discovery Sciences

Abstract Productivity demands in today s laboratories require the chemist to minimize time spent purifying compounds. Faster, greener flash chromatography methods increase throughput, improve productivity, save solvent, and reduce operating costs. Meeting these goals requires gradient methods that deliver the required resolution in the fastest possible time. The RevealX Operating System of the Reveleris flash chromatography system automatically generates gradient profiles using only two chromatographic separations (TLC or HPLC) as inputs and provides methods for both normal phase and reversed phase systems. The chemist can choose a gradient profile based on either highest purity or fastest speed. In this work both of these gradient profiles were compared with traditional gradient development routes for normal and reversed phase chromatography examples. This work shows a productivity gain, when using this operating system for gradient method development, by increasing chromatographic resolution, reducing time spent optimizing the separation and reducing solvent used compared to traditional methods. Introduction This paper compares the productivity of the RevealX Operating System on a Reveleris flash chromatography system with traditional flash method optimization. Flash purifications are improved for quality, speed, and cost easily by the control capability of the Navigator feature of the operating system. Productivity of the final flash purification method was determined by measuring resolution for the chromatographic compound pairs, time savings, and solvent used to run the method. Experiments were conducted to evaluate the following separations and mixture types: a two component normal phase (NP), three component normal phase (NP) and two component reversed-phase (RP) separation. All chromatography comparisons were made using the same instrument and flash cartridge or HPLC column. An overview of two development pathways and overall process is shown below. Sample Optimize eluant solvent system by TLC RevealX Operating System Optimization Only 2 TLC/HPLC isocratic runs off line Multiple TLC/HPLC runs to meet acceptance criteria. Chemist performs optimization off line Reveleris Navigator software automatically generates the desired method TLC or HPLC separation data is converted into a flash method Final Purity Run and Result/Fraction collection Speed Run and Result/Fraction collection Run and Result/Fraction collection

2-Component Normal Phase Experimental: Equipment: TLC tank Cartridge: Reveleris Silica 12g Part No 5146131 1 x 1cm Reveleris Si plates Mobile Phases: A: Hexane, B: Ethyl acetate Competitor s software system Sample Mix: alpha tocopherol, delta tocopherol in hexane Reveleris flash instrument Part No 5148513 Optimization Procedure: development is conducted by repeatedly spotting TLC plates with sample mixture and developing the plate in a solvent chamber containing enough solvent to raise to a 1cm height from the bottom with a mix of a solvent pair. The plate is allowed to sit in the chamber and wick up an isocratic mix of solvent until it nears the top of the plate. Optimization takes place by the chemist selecting TLC plate conditions that best separate the components of interest (visible by UV light irradiation or chemical staining) from 1-3 near eluting components or impurities. The resulting component Rfs (measured as height of elution) are used as software input values for gradient predictor software required to generate flash gradient method conditions. Gradient: Flow Rate: 36mL/min Time: 4.2 15.2 18.2 18.2 19.2 %B: 1 1 The RevealX Operating System Procedure: Two isocratic TLC runs are made with the sample mixture using the same solvent pair. The Rf values are entered into the Reveleris Navigator method predictor feature, which automatically generates an ideally optimized method of choice. One of two method options is selected: either a Purity method designed for maximizing resolution or a Speed method for minimizing run time/solvent use. optimization is automatically performed by the software system. Gradients: Purity : Flow Rate: 36mL/min Time:.6 5.3 8.1 %B: 3 3 9 9 Speed : Flow Rate: 36mL/min Time:.6 2. 4.5 %B: 3 3 9 9 Results: 2-Component Normal Phase Performance Purity or Flash System Speed Time (min) 52 52 78 Solvent Used (ml) 5 5 75 Resolution 2.94 2.4 1.5 Time (min) 8.1 4.5 19.2 Solvent Used (ml) 292 162 691 Total Time (min) 6.1 56.5 Total Solvent Used (ml) 342 212 1.2 Reveleris Flash System Purity 8 6 4 2. 2 4 6 8 1 12 14 16 18 Min.16.12.8.4 1.2 Reveleris System Speed 8 6 4 2. 2 4 6 8 1 12 14 16 18 Min.16.12.8.4 Competitor s 1 8 6 4 2 2. 4 6 8 1 12 14 16 18 Min.2.16.12.8.4 2-Component NP Resolution Results 2-Component NP TIme Used 2-Component NP Solvent Used 3.5 3.5 3. 3. 2.94 2.5 2.5 2. 2. 1.5 1% increase in Rs for Purity 2.4 1.5 Time Used (Min) 12. 1 2. 1. 1. 8. 8. 6. 6. Time Used (min) 4. 4. 2. 2. 4% drop in time used for Reveleris System Speed 6.1 58.5 56.5 Solvent (ml) 9 9 8 8 7 7 6 6 5 5 4 4 Solvent (ml) 3 3 2 2 1 1 7% decrease in solvent use for Reveleris System Speed 342 342 212 212 1. Purity Reveleris Speed System Competitors Competitor s Purity Speed.. Reveleris Purity Reveleris Speed Competitors Purity Speed Competitor s Purity Speed Competitor s Reveleris Reveleris Optimizer So we see that the competitor method has the lowest resolution, longest time, and uses the most solvent. By using the RevealX operating system and choosing speed as the rate limiting factor, the baseline resolution between the peaks is still achieved and the minimum amount of solvent is used with the shortest time.

3-Component Normal Phase Experimental: Equipment: TLC tank Cartridge: Reveleris Silica 12g Part No 5146131 1 x 1cm Reveleris Si plates Mobile Phases: A: Hexane, B: Ethyl acetate Competitor s software system Sample Mix: alpha tocopherol, delta tocopherol, Reveleris flash instrument Part No 5148513 methyl paraben in hexane/ethyl acetate Optimization Procedure: predictor software for three components is not available. The method predictor software previously used in 2-component NP was employed to generate a flash method gradient based on the two Rf values having the least difference. Gradient : Flow Rate: 36 ml/min Time: 3.2 3.2 4.2 15.2 18.2 18.2 19.2 %B: 1 1 The RevealX Operating System Procedure: Two isocratic TLC plate tests were conducted, instead of the usual numerous tests, with the three component sample mix. Three Rf values are entered into the Reveleris Navigator method predictor feature including the component of interest and two of it s most closely eluting impurities. An optimized method was automatically generated and parameters loaded in preparation in advance of the flash purification run. The method optimizes the separation for all three components for maximum resolution, least time and minimal solvent use. Results: 3-Component Normal Phase Performance Gradient : Flow Rate: 36mL/min Time:.6 2. 3.9 6.4 %B: 3 3 9 27 27 Time (min) 52 78 Solvent Used (ml) 5 75 Resolution peak 1 & 2 2.8 1.46 Resolution peak 2 & 3 3.5 3.4 Time (min) 6.4 19.2 Solvent Used (ml) 23 691 Total Time (min) 58.4 Total Solvent Used (ml) 28 1 8 6 4 2 Peak 1 Peak 2 Peak 3. 2 4 6 8 1 12 14 16 18 Min.1.8.6.4.2 1 8 6 4 2. 2 4 6 8 1 12 14 16 18 Min.3.24.18.12.6 3.5 3. 2.5 2.8 2. 1.5 3-Component NP Resolution Results Peaks 1 & 2 Peaks 2 & 3 Peaks 2 & 3 3.5 3.4 4% increase in Rs for critical pair separation Peaks 1 & 2 1.46 Time Used (Min) 12. 1 2 1 1. 8 8. 6 6. Time Used (min) 4 4. 3-Component NP TIme Used 4% drop in time used for Reveleris System 58.4 58.4 1. Reveleris System 2 2.. Reveleris R e v ele System l 3-Component NP Solvent Used 9 9 8 8 Solvent (ml) 7 7 6 6 5 5 4 4 3 3 Solvent Used (ml) 28 6% decrease in solvent use for 2 2 1 1 Reveleris R e v ele System l Similarly with this example, the competitor method has the lowest resolution, particularly between peaks 1 and 2, as well as using the most solvent and taking the longest time.

2-Component Reversed-Phase Experimental: Equipment: HPLC Instrument: Agilent 11 HPLC Column: Reveleris C18 4.6 x 1mm Reveleris flash instrument Part No 5148513 Cartridge: Reveleris C18 12g Part No 515213 Mobile Phase: A: DI water, B: Methanol Sample Mix: 1% benzoic acid.5% caffeine in 3% MeOH 7% DI water Optimization Procedure: Gradient method development is performed by the chemist on an HPLC system using an analytical column. Multiple runs are made to optimize the gradient through a manual iterative approach resulting in several analysis runs. This HPLC method is then converted to a flash gradient method manually, by calculating volumes and scaling up accordingly. Gradient: Flow Rate: 3mL/min Time:.5 5.8 %B: 4 4 65 The RevealX Operating System Procedure: Two scouting runs were performed, versus the usual unknown iterative number of runs, at different isocratic mobile phase conditions using a Reveleris HPLC analytical column using the same solvent pair. Retention times of the component of interest and its nearest impurity were entered into the Reveleris Navigator method predictor software to automatically generate two optimized methods. A Purity method to maximize resolution/target compound recovery and a Speed method to minimize time/solvent are used. Gradients: Purity : Flow Rate: 3mL/min Time:.9 6.9 1.6 %B: 38 38 55 55 Speed : Flow Rate: 3mL/min Time:.9 2.7 6. %B: 38 38 55 55 Results: 2-Component Reversed-Phase Performance Purity Speed Time (min) 67.6 62.5 13 Solvent Used (ml) 5 5 125 Resolution 1.95 1.83 1.58 Time (min) 1.6 6. 5.8 Solvent Used (ml) 318 18 174 Total Time (min) 78.2 68.5 135.8 Total Solvent Used (ml) 368 23 299 2-Component RP Resolution Results 2-Component RP TIme Used 2. 2. 1. 9 1.9 1. 8 1.8 1. 7 1.7 1. 6 1.6 1. 5 1.5 1. 4 1.4 1. 3 1.3 1. 2 1.2 1. 1 1.1 1. 1. 1.95 1.95 Purity 1.83 1.83 Speed 2% increase in Rs for Reveleris System Purity 1.58 Reveleris Purity Reveleris Speed 1.58 Time Used (Min) 1 6. 16. 1 4. 14. 1 2. 12. 1. 1. 8. 8. 6. 6. Time Used (min) 4. 4. 2. 2... 5% drop in time used for Speed 78.2 78.2 Purity 68.5 68.5 Speed 135.8 Reveleris Purity Reveleris Speed Competitors 2-Component RP Solvent Used 4 4 3 5 35 3 3 368 368 4% decrease in solvent use for Speed 299 Solvent (ml) 25 2 5 2 2 15 1 5 Solvent Used (ml) 1 1 23 55 Purity Speed Reveleris Purity Reveleris Speed Competitors The RevealX operating system is equally valuable for time and solvent savings with reversed-phase separations. The data again shows increased resolution with lower solvent consumption and reduced time.

Conclusion The RevealX Operating System Navigator feature was successfully employed to guide flash purification method optimization for quality, speed, or reduced costs. It is now possible to easily improve methods using Speed or Purity options within the method predictor software using Reveleris Navigator. 2-Component NP separation was optimized for Purity resulting in twice the resolution to the traditional approach. The Speed method uses 7% less solvent and 4% less total time. The 3-Component NP separation employing the Reveleris Navigator method resulted in 4% higher resolution in peaks 1 & 2, uses 6% less solvent than the traditional method and an improvement in total time saved by 4%. The 3 Component NP Reveleris Navigator is the only method to solve 3 component flash separations automatically, which can take multiple calculations in competitive software. The 2-Component RP Reveleris Navigator Purity and Speed optimization resulted in 4% solvent and 5% total time saved with 2% increased resolution compared to a manual iteratively development method. This is the only known RP optimizer software to automatically generate flash methods from LC data. The traditional methods require repeated experimentation until the method is optimized or the Rf values meet some set criteria; some chromatographic knowledge to perform this is required. The Reveleris Navigator automatically performs this optimization in all 3 examples. An additional feature of the Reveleris Navigator is to automatically transfer HPLC reversed or normal phase method details into a Flash method. The RevealX Operating System allows a relatively inexperienced chromatographer to develop methods that separate target compounds with increased purity (resolution), with less solvent (time) and to do this with less development steps compared to traditional methods. The information presented herein is derived from our testing and experience. It is offered for your consideration and verification. Since operating conditions vary significantly, and are not under our control, we disclaim all warranties on the results that may be obtained from the use of our products. W. R. Grace & Co.-Conn. and its subsidiaries can not be held responsible for any damage or injury occurring as a result of improper installation or use of its products. Grace reserves the right to change prices and/or specifications without prior notification. NAVIGATOR and REVEALX are trademarks of Alltech Associates, Inc. ALLTECH and REVELERIS are trademarks, registered in the United States and/or other countries, of Alltech Associates, Inc. GRACE DAVISON DISCOVERY SCIENCES is a trademark of W. R. Grace & Co.-Conn. GRACE and GRACE DAVISON are trademarks, registered in the United States and/ or other countries, of W. R. Grace & Co.-Conn. This trademark list has been compiled using available published information as of the publication date of this brochure and may not accurately reflect current trademark ownership or status. Grace Davison Discovery Sciences is a product group of W. R. Grace & Co.-Conn. Alltech Associates, Inc. is a wholly owned subsidiary of W. R. Grace & Co.-Conn. Copyright 211 Alltech Associates, Inc. All rights reserved. Printed in the USA.