Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Similar documents
PAR Method Method to 2.0 mg/l Pb TNTplus 850

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Oxygen Demand, Chemical

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Oxygen Demand, Chemical

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Nitrogen, Total Inorganic

Chloramine (Mono) and Nitrogen, Free Ammonia

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Chlorine, Free and Total, High Range

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Chlorine, Free and Total, High Range

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Chlorine, Free and Total, Low Range

DR/4000 PROCEDURE. IRON, Total

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Method to 0.50 mg/l NH 3 N Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows

Trihalomethane Formation Potential (THMFP)

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. 2. The display will show:

TPH (Total Petroleum Hydrocarbons)

Quaternary Ammonium Compounds

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

CHLORINE, TOTAL (0 to 4.00 mg/l)

DR/4000 PROCEDURE. HARDNESS, Total

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Mercaptoacetic Acid Method*

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

Polychlorinated Biphenyls (PCB) in Soil

NITROGEN, AMMONIA, High Range, Test N Tube

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

NITRITE, Low Range (0 to mg/l NO 2- -N)

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

DPD Test N Tube Method *

NITROGEN, TOTAL, HR, Test N Tube

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

NITROGEN, TOTAL, Test N Tube * Method 10071

SILICA, Low Range (0 to 1.60 mg/l)

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

p-dimethylaminobenzaldehyde Method*

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Manganese III Digestion Method * (with chloride removal)

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

Steam-Assisted Gravity Drainage Water Analysis Handbook

ANALYSIS CERTIFICATE OF

SILICA, Low Range SiLR

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

ALLOWAY METHOD OUTLINE

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

Phosphate total HR TT mg/l P b) Phosphomolybdenum Blue

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Ground Water & UST Samples: Containers, Preservation and Hold Times Table

Steam-Assisted Gravity Drainage Water Analysis Handook

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

MERCURY (0.1 to 2.5 µg/l) Method 10065

Instruction Manual. Waterproof Portable Colorimeter C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105. Part of Thermo Fisher Scientific

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Nitrogen total LR VARIO TT mg/l N b) Persulphate Digestion

Sample Preparation LCW 909. Total Kjeldahl Nitrogen, TKN NB: NEW! LCW 909 GB

Schools Analyst Competition

#18 Analysis of Water (Chlorine, Phosphorus, Fluoride, and Color)

Transcription:

Aluminum DOC316.53.01004 Chromazurol S Method Method 10215 0.02 to 0.50 mg/l Al TNTplus 848 Scope and application: For drinking water, surface water, swimming pool water, wastewater and process analysis. Test preparation Instrument-specific information Table 1 shows all of the instruments that have the program for this test. The table also shows the adapter and light shield requirements for the applicable instruments that can use TNTplus vials. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information for TNTplus vials Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR 1900 9609900 or 9609800 (A) Before starting DR 3900, DR 3800, DR 2800: Install the light shield in Cell Compartment #2 before this test is started. Review the safety information and the expiration date on the package. The recommended sample ph is 2.5 3.5. The sample temperature must be 20 23 C (68 73 F) for accurate results. The recommended temperature for reagent storage is 15 25 C (59 77 F). DR 1900: Go to All Programs>LCK or TNTplus Methods>Options to select the TNTplus number for the test. Other instruments automatically select the method from the barcode on the vial. Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. Items to collect Description Quantity Aluminum TNT848 Reagent Set 1 Pipet, adjustable volume, 1.0 5.0 ml 1 Pipet tips, for 1.0 5.0 ml pipet 1 Refer to Consumables and replacement items on page 4 for order information. 1

Sample collection and storage Test procedure Collect samples in clean glass or plastic bottles that have been cleaned with 6 N (1:1) hydrochloric acid and rinsed with deionized water. To preserve samples for later analysis, adjust the sample ph to less than 2 with concentrated nitric acid (approximately 2 ml per liter). No acid addition is necessary if the sample is tested immediately. Keep the preserved samples at room temperature for a maximum of 6 months. Before analysis, adjust the ph to 2.5 3.5 with 5 N sodium hydroxide solution. Correct the test result for the dilution caused by the volume additions. 1. Use a pipet to add 2.0 ml of Solution A to the test vial. 2. Use a pipet to add 3.0 ml of sample to the test vial. 3. Add one level spoonful of Reagent B to the vial. 4. Tighten the cap on the vial and invert until completely mixed. Start 5. Start the reaction time of 25 minutes. 6. DR 1900 only: Select program 848. Refer to Before starting on page 1. 7. Insert the Zero vial into the cell holder. DR 1900 only: Push ZERO. The instrument zero is set. 8. When the timer expires, clean the vial. 9. Insert the vial into the cell holder. DR 1900 only: Push READ. Results show in mg/l Al. 2 Aluminum, Chromazurol S TNTplus Method (0.50 mg/l)

Reagent blank correction For the best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results automatically with the reagent blank adjust option. Measure the reagent blank value when a new lot of reagent is used. 1. Use deionized water as the sample in the test procedure to measure the reagent blank value. 2. Set the reagent blank function to on. The measured reagent blank value is shown. 3. Accept the blank value. The reagent blank value is then subtracted from all results until the reagent blank function is set to off or a different method is selected. Note: As an alternative, record or enter the reagent blank value at a different time. Push the highlighted reagent blank box and use the keypad to enter the value. Sample blanks Interferences A sample blank is the portion of the test result that is from color or turbidity in the sample and not from the analyte. Samples that do not have color or turbidity have no sample blank value. If the sample blank value is within the permitted range, the sample blank value is automatically subtracted from the test result. Table 2 shows that the ions were individually examined to the given concentrations and do not cause interference. No cumulative effects or influences of other ions were found. Higher concentrations of heavy metals other than those given, as well as fluoride, phosphate and relatively rare elements (e.g., beryllium, thorium, titanium, zirconium and vanadium) interfere with the determination. Aluminum oxide hydrates and hydroxide are only partially found. Verify the measurement results with sample dilutions or standard additions. Table 2 Interfering substances Interfering substance Mg 2+, K +, Na +, NH + 4, Cl -, NO - 3, SO 2-4, Ca 2+ Ag+, Mn 2+ Cd 2+, Co 2+, Ni 2+, Sn 2+, Pb 2+, PO 3-4 Cu 2+, Hg 2+ Fe 2+, Fe 3+, Zn 2+, Si 4+ Cr 3+, Cr 6+ F - Interference level 500 mg/l 100 mg/l 50 mg/l 10 mg/l 5 mg/l 0.5 mg/l 0.1 mg/l Accuracy check Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: 100-mg/L Aluminum Standard Solution 250-mL volumetric flask, Class A 1.0-mL volumetric pipet, Class A and pipet filler safety bulb Deionized water Aluminum, Chromazurol S TNTplus Method (0.50 mg/l) 3

1. Prepare a 0.4-mg/L aluminum standard solution as follows: a. Use a pipet to add 1.0 ml of a 100-mg/L aluminum standard solution into the volumetric flask. b. Dilute to the mark with deionized water. Mix well. Prepare this solution daily. 2. Use the test procedure to measure the concentration of the prepared standard solution. 3. Compare the expected result to the actual result. Note: The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are small variations in the reagents or instruments. Summary of Method Chromazurol S forms a green lake with aluminum in weak acidic acetate-buffered solutions. The amount of color formed is directly proportional to the amount of aluminum in the sample. The measurement wavelength is 620 nm. Consumables and replacement items Required reagents Description Quantity/Test Unit Item no. Aluminum TNT848 Reagent Set 1 24/pkg TNT848 Required apparatus Description Quantity/test Unit Item no. Pipet, adjustable volume, 1.0 5.0 ml 1 each BBP065 Pipet tips, for 1.0 5.0 ml pipet 1 75/pkg BBP068 Light shield, DR 3800, DR 2800, DR 2700 1 each LZV646 Light shield, DR 3900 1 each LZV849 Recommended standards Description Unit Item no. Aluminum Standard Solution, 10-mL Voluette Ampule, 50 mg/l as Al 16/pkg 1479210 Aluminum Standard Solution, 100-mg/L as Al 3+ 100 ml 1417442 Optional reagents and apparatus Description Unit Item no. Flask, volumetric, Class A, 250 ml each 1457446 Nitric Acid, concentrated 500 ml 15249 Sample blank vials for TNTplus methods 5/pkg TNT919 Sampling bottle with cap, low density polyethylene, 500-mL 12/pkg 2087079 Sodium Hydroxide Standard Solution, 5.0 N 100 ml MDB 245032 Syringe, 10-cc, Luer-Lock tip each 2202400 Test tube rack, stainless steel each 1864100 Water, deionized 4 L 27256 4 Aluminum, Chromazurol S TNTplus Method (0.50 mg/l)

Aluminum, Chromazurol S TNTplus Method (0.50 mg/l) 5

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company/Hach Lange GmbH, 1989 2017. All rights reserved. 01/2017, Edition 9