Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Similar documents
Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

The Polyarc System Product Brochure. A Breakthrough Innovation Revolutionizing GC/FID Technology

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

1.0 Reactive Gas Chromatography Design and Setup Procedure for Catalyst Pelletization & Inlet Packing Software and Method.

Introduction to Capillary GC

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Eliminating the Need for Standards and Calibration in GC/FID Analysis of Hydrocarbons and Oxygenates

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Analytical Trap Comparison for USEPA Method 8260C

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Study of Residual Solvents in Various Matrices by Static Headspace

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Application. Gas Chromatography February Introduction

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Author. Abstract. Introduction

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Analysis of BTEX in Natural Water with SPME

GAS CHROMATOGRAPHY (GC)

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

U.S. EPA Method 8270 for multicomponent analyte determination

Understanding Gas Chromatography

Selection of a Capillary GC Column

Monomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

STANDARD OPERATING PROCEDURES

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Trajan SGE GC Columns

Introduction to Capillary GC

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Secrets of GC Column Dimensions

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

METHOD 8015C NONHALOGENATED ORGANICS BY GAS CHROMATOGRAPHY

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Selection of a Capillary

Luminescence transitions. Fluorescence spectroscopy

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Gas Chromatography (GC)

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Water Injections in GC - Does Water Cause Bleed?

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

METHOD 8015D NONHALOGENATED ORGANICS USING GC/FID

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Headspace Technology for GC and GC/MS: Features, benefits & applications

Chemistry Gas Chromatography: Separation of Volatile Organics

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Ed George and Anaïs Viven Varian, Inc.

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Original. Organic Standards Section, Organic Analytical Chemistry Division, National Metrology Institute of Japan (NMIJ),

Optimizing. Abstract: is standardd. procedures. altered to

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2),

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

BUTYL ALCOHOL in urine by GC-FID Code GC05510

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Methanol Extraction of high level soil samples by USEPA Method 8260C

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Transcription:

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID Application Note Volatile Organic Compounds (VOCs) Author Andrew Jones Activated Research Company 7561 Corporate Way Eden Prairie, MN 55344 Andrew.Jones@activatedresearch.com Abstract The concentrations of oxygenates commonly used as fuel additives were determined with an average error of 1.7%, and within gravimetric uncertainties, using an ARC Polyarc reactor in series with a flame ionization detector (FID) on a gas chromatograph (GC). The analysis demonstrates a greater than 5-fold speed-up of analysis with less cost and less introduction of human error. This method is applicable to a wider range of volatile organic compounds (VOCs) that have importance in environmental and industrial testing. Introduction Oxygenated compounds are added to fuels to increase octane number and allow for more complete combustion and less pollution. These have included C1-C4 alcohols (e.g., methanol and tert-butanol) and C5-C7 ethers (e.g., diisopropyl ether). The water solubility of these compounds has led to their presence in groundwater and soils around the world in areas where fuel leaks have occurred. Concerns about the presence of these compounds in aquafers and drinking water has led to the need for accurate testing and monitoring of their levels. The analysis of oxygenates presents a unique challenge to the analytical chemist because of their varied response in GC detectors and the difficulty in preparing and handling volatile standards. Compounds are typically identified and quantified using GC with mass spectrometry and 5-point calibration curves, such as those specified in EPA method 8260B for volatile organic compounds (VOCs), defined as those compounds with boiling points less than 200 C. Accurate analysis requires expensive instruments and pumps, time-consuming 5-point calibrations, regular (sometimes weekly) purchases of standards, and the careful preparation, handling and storage of those standards. These requirements limit the feasibility and sometimes the accuracy of oxygenate analysis. Here, we show that the accurate analysis of oxygenates and VOCs is possible without standards, using inexpensive GC/FID equipment and the Polyarc reactor from Activated Research Company (ARC). The concentrations of six volatile ethers and alcohols are determined with an average error of 1.7% from a single injection and an arbitrary internal standard. These results provide a new avenue for environmental sampling that is quick and inexpensive while providing high accuracy. Experimental An Agilent 7890A GC equipped with a capillaryoptimized FID and an ARC Polyarc reactor (PA-RRC- A02) were used for the analysis. Helium (99.999%, Praxair) was used for carrier and FID makeup. Air (zero grade, Praxair) and H 2 (99.999%, Praxair) were supplied to the ARC electronic flow control module (PA-MFC-A09) and to the FID. The effluent of the GC column was sent directly to the inlet of the Polyarc reactor via a zero-dead volume union (PA-CPM-R46). The reactor effluent was connected directly to the FID. Figure 1 illustrates the configuration. A mixture of oxygenates (Restek, 30626, Lot A0115237) was transferred to a chilled (0 C) sample vial for analysis. The sample was injected using an automated sample handler and a 10 μl syringe. Activated Research Company, LLC 2015-2016 1

The sample was also analyzed without a Polyarc reactor on an identical setup, but with the FID H 2 flow rate at 35 sccm. In addition, the sample was analyzed using an Agilent 5973 mass spectrometer connected to the column effluent. Here, the end of the capillary column was connected to a tee (Agilent, G3184-60065), which was connected to the inlet of the Polyarc/FID and the mass spectrometer via a retention gap column (Agilent, 160-2635-5, 2 ft., 0.1 mm ID). Figure 1. Diagram of GC setup with Polyarc/FID, FID and MS options. GC conditions Front inlet Split/splitless Inlet liner Agilent P/N 5190-2295 Inlet Temperature 250 C Inlet Mode 5:1 Split Inlet Pressure 11.27 psi Septum purge flow 3 sccm Oven 40 C (2 min), 10 C/min to 80 C Column HP-5 (30 m 0.32 mm 0.25 µm) Syringe 10 µl Injection 0.2 µl FID conditions Temperature 315 C H2 1.5 sccm Air 350 sccm Makeup 20 sccm (He) Sampling 100 Hz Polyarc reactor conditions Setpoint 293 C H2 35 sccm Air 2.5 sccm Results and Discussion Table 1 describes the concentrations and elution order of alcohols and ethers in the mixture studied. The concentrations are reported as determined gravimetrically from the supplier. The unstressed (vial maintained at moderate temperatures) concentration uncertainties were 2.1 % for each component. The elution order was determined from the literature and confirmed with GC/MS. No other major components were observed. The concentration of methanol solvent was estimated from its density and the ideal volume of other species. Table 1. Oxygenate sample information including elution order and concentration. Elution Short Name Name Concentration (µg/) 1 MeOH Methanol 764225 2 TBA Tert-butanol 9912 3 MTBE methyl-tert-butyl ether 1986 4 DIPE diisopropyl ether 1990 5 ETBE ethyl-tert-butyl ether 1988 6 TAME tert-amyl methyl ether 1988 7 TAEE tert-amyl ethyl ether 1992 Activated Research Company, LLC 2015-2016 2

Figure 2. GC-Polyarc/FID chromatogram of oxygenates mixture. Figure 2 shows the FID signal after Polyarc reactor conversion to methane of MeOH, TBA, MTBE, DIPE, ETBE, TAME, and TAEE. Baseline separation was obtained for all species. A small peak on the MeOH peak tail is most likely due to dimethyl ether formed from the reaction of MeOH on acid sites before chromatographic separation. The Polyarc reactor converts compounds to methane in a two-step reaction scheme (Figure 1). First, compounds are oxidized until each carbon atom is converted to CO 2. Second, the resulting CO 2 is reduced to CH 4 in the presence of H 2. The resulting CH 4 is measured by the FID, and because non-carbon species are invisible to the FID, the resulting detector response and sensitivity is equivalent on a per carbon basis regardless of the original compound composition and type. The universal and uniform carbon response has been demonstrated on a variety of different molecules. The complete conversion of carbon to methane is the result of the special catalyst blend, reactor flow-path design, and the temperatures and conditions of the Polyarc reactor, and has been verified by mass spectrometry for a number of molecules and concentrations. Thermodynamic calculations ensure that >99.9% of carbon will become methane at equilibrium under the conditions of Polyarc operation [1]. The uniform response of the Polyarc/FID detector to carbon, allows for the quantification of species using, RF = 1 = area concentration, (1) area standard /concentration standard where RF is the relative response factor of the species to the internal standard (IS), area is the integrated detector response, and concentration is the molar carbon concentration in the mixture. The RF is unity for all organic molecules when the Polyarc reactor is used. ETBE is used as the IS for this study, however, any component could be used (or added to the mixture) because of the uniform response of the Polyarc/FID to carbon. The concentration of alcohols and ethers in the sample are thus calculated from the carbon concentration of the IS and the relative peak areas of the IS and the analyte, area concentration = concentration area standard. (2) standard Sample calculation The carbon concentration of ETBE in the undiluted sample can be calculated from its gravimetric concentration, 1988 μg/, molecular weight, 102.18 g/mol, and number of carbons per molecule, 6, 1987.92 10 6 g ETBE 1 sample 1 mol ETBE 102.18 g 6 mol C 1 mol ETBE = 116.73 10 6 mol C. Activated Research Company, LLC 2015-2016 3

The concentration of any other species can now be calculated from its area ratio to ETBE. For example, the integrated detector response (i.e., area under the curve) of MTBE is 740.2, whereas the area of ETBE is 770.7. From Eq. (2), we find that the carbon concentration of MTBE is, 740.2 pa s 770.7 pa s 116.73 10 6 mol C = 112.1 10 6 mol C. Figure 4 compares the actual and measured concentrations of all compounds. The concentrations span nearly three orders of magnitude from ~10 3 to ~10 6 μg/. This demonstrates the effective analysis range of the GC-Polyarc/FID from 1000 ppm to ~100% (pure solvent). Previous studies have shown detection sensitivity below 100 ppb with a similar setup indicating a linear dynamic range of about 7 orders of magnitude. This equates to a measured concentration of 1977 μg/. The gravimetric concentration was 1986 μg/, yielding a total measurement vs. gravimetric error of, 1977 1986 %Error = 100 = 0.5% 1986 Figure 3 shows the quantification errors using the GC- Polyarc/FID setup with Eq. (2) for the determination of compound concentrations from their FID peak areas. Absolute errors range from 0.5% for MTBE to 3.4% for tert-butanol. Five out of the six components (internal standard is not included) have quantification errors below the gravimetric unstressed uncertainty of sample preparation (<2.1%) from the manufacturer. The choice of internal standard has a negligible influence on average error values indicating the choice of internal standard is arbitrary because all components are converted to methane before detection. Figure 3. GC-Polyarc/FID quantification errors of compounds in Table 1. ETBE (elution number 5) is used as the internal standard. The dotted line represents the gravimetric unstressed uncertainty of the compounds and the dashed line represents the average of the measurement errors. Figure 4. Parity plot of GC-Polyarc/FID measured and gravimetric (actual) concentrations of compounds in Table 1. The concentrations of compounds in Table 1 could alternatively be determined with a similar accuracy using GC/FID or GC/MS (MeOH would saturate the MS) and five point calibration curves. The purchase and preparation of standards, and the subsequent calibrations, however, can represent a large investment in time and money. In this example a five point calibration curve for each of the 6 components relative to the internal standard would require 30 samples, or a 31-fold increase in the number of injections (97% reduction in samples with the Polyarc/FID). This could be reduced to 5 samples by using 7-component mixtures as standards, but the effort in producing these mixtures and the cost of the standards would be similar. In addition, there is a significant chance for the introduction of error during standard preparation and analysis. The use of GC- Polyarc/FID, even in tandem with GC/MS, represents a significant reduction in preparation and analysis cost and time. The response factors of an FID have been tabulated for some compounds in the literature [2]. The accuracy of these relative response factors on various instruments with different operating conditions is unsubstantiated. The response factors of MeOH and TBA (other compounds are not available) have been reported as 0.23 and 0.74, respectively. Using these tabulated response factors with injections on the Activated Research Company, LLC 2015-2016 4

GC/FID system leads to quantification errors of 28% for TBA. The effective carbon number (ECN) concept was introduced to account for deviations in the response factors of the FID due to certain functional groups present in molecules. The ECN values published by Scanlon et al. [3] are commonly used to calculate relative response factors. Figure 5 shows the analysis error using the ECN method and FID peak areas from the injection of the mixture in Table 1 into a GC/FID (ETBE as the internal standard). Quantification errors range from 0.6% for MTBE to 44.2% for MeOH, with an average error of 10%. The ethers have an average analysis error of 1.6%, which is not surprising considering the similar relative responses of these species in the FID due to their similar ether functionality. TBA and MeOH have much higher errors of 9.5% and 44.2% suggesting the ECN model does not capture the different response of alcohols and ethers on this FID. These results strongly discourage the use of the ECN model when comparing alcohols and ethers. References 1. Beach, C., Krumm, C., Spanjers, C., Maduskar, S., Jones, A., and Dauenhauer, P., Analyst, 141 (2016) 1627-1632. 2. Dietz, W. A., J. Gas Chrom., 5 (1962) 68-71. 3. Scanlon, J. T., and Willis, D. E., J. Chrom. Sci., 23 (1985) 333-340. Contact Us For more information or to purchase a Polyarc reactor, please contact us at 612-787-2721 or contact@activatedresearch.com. Please visit our website for details and additional technical literature, www.activatedresearch.com. 2016 Activated Research Company, LLC Printed in the USA July 06, 2016 PA-APP-1625 Figure 5. GC-FID quantification errors using the ECN model [3]. Conclusions A 7 component mixture of oxygenates including ethers and alcohols commonly added to fuels was quantified using the GC-Polyarc/FID with an average error of 1.7% and concentrations that range over three orders of magnitude. This high accuracy is obtained without the use of calibrations, potentially reducing the analysts workload by more than 31-fold while maintaining exceptional accuracy. It is found that literature tabulated response factors and effective carbon numbers do not accurately reflect the different sensitivity of alcohols and ethers in the FID and should be avoided. Activated Research Company, LLC 2015-2016 5