DETERMINING SIZE OF SOURCE FOR HANDHELD INFRARED THERMOMETERS THEORY AND PRACTICE

Similar documents
A Study on the Effects of Bandwidth of IR Thermometry Measurements

Infrared Calibration Development at Fluke Corporation Hart Scientific Division

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

Infrared Thermometer Calibration A Complete Guide

Traceability and Quality Control in a Radiation Thermometry Laboratory

Traceability and Quality Control in a Radiation Thermometry Laboratory

Traceability and Quality Control in a Radiation Thermometry Laboratory

Standards for Radiation Thermometry

RADIOMETRIC COMPARISON BETWEEN A NATIONAL LABORATORY AND AN INDUSTRIAL LABORATORY

DIRECT RADIOMETRIC TECHNIQUES

ASSET INTEGRITY INTELLIGENCE. Featured Article. ACHIEVING A COMPREHENSIVE FIRED HEATER HEALTH MONITORING PROGRAM By Tim Hill, Quest Integrity Group

Infrared Temperature Calibration 101 Using the right tool means better work and more productivity

Influence of sun radiation on results of non-contact temperature measurements in far infrared range

Sensors - April A Certified-Emissivity Blackbody for Calibrating Infrared Thermometers

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

Thermal Image Resolution on Angular Emissivity Measurements using Infrared Thermography

Characterization of the VIIRS Blackbody Emittance

D501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect

Inside Wall Temperature Measurements of DSTs Using an Infrared Temperature Sensor

Validation of NIST s Low Temperature Infrared Spectral Radiance Scale

IR spectral characterization of customer blackbody sources: first calibration results

Standard Guide for Determination of the Thermal Resistance of Low-Density Blanket-Type Mineral Fiber Insulation 1

Emissivity: Understanding the difference between apparent and actual infrared temperatures

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

Pre-lab Quiz/PHYS 224. Your name Lab section

The Quantum Model of the Hydrogen Atom

The Thermal Sieve: a diffractive baffle that provides thermal isolation of a cryogenic optical system from an ambient temperature collimator

EXPERIMENT 18 THE PHOTOELECTRIC EFFECT

High Accuracy High Speed in Your Process

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

On the effects of temperature dependence of spectral emissivity in industrial radiation thermometry

Any first year text, sections on atomic structure, spectral lines and spectrometers

ABB temperature measurement Radiation thermometry. Measurement made easy. Process temperature measurement practice--non-contacting

TIR100-2 Measurement of Thermal Emissivity

Pin Fin Lab Report Example. Names. ME331 Lab

Surrounding Effects on Temperature and Emissivity Measurement of Equipment in Electrical Distribution System

Operator's Manual. optris BR 400. Calibration source

Lab 6. RC Circuits. Switch R 5 V. ower upply. Voltmete. Capacitor. Goals. Introduction

Uncertainty of temperature measurements by infrared thermography for metal cutting applications

Method to Remove the Effect of Atmosphere and Ambient Radiation on Colorimetric Temperature Measurement

Switch + R. ower upply. Voltmete. Capacitor. Goals. Introduction

UNCERTAINTY IN THE TEMPERATURE OF SILICON WAFERS MEASURED BY RADIATION THERMOMETRY BASED UPON A POLARIZATION TECHNIQUE

Blackbody Radiation. George M. Shalhoub

InfraRed Thermometer with Laser Pointer

Laser Optics-II. ME 677: Laser Material Processing Instructor: Ramesh Singh 1

User Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration

Novel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007

Calibrating the Thermal Camera

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

INFRARED THERMOMETER TM0819

Model 556X User s Manual. Optical Isolator

Hot solutions for temperature measurement. Contact and non-contact thermometers from Fluke

SOME ASPECTS OF PYROMETRY OF "WHITE" OBJECTS V.E.

Mathematical determination of emissivity and surface temperature of aluminum alloys using multispectral radiation thermometry

Switch. R 5 V Capacitor. ower upply. Voltmete. Goals. Introduction

Introduction. Procedure. In this experiment, you'll use the interferometer to EQUIPMENT NEEDED: Lens 18mm FL. Component holder.

Photoelectric Effect

Chamber Development Plan and Chamber Simulation Experiments

PROCESS CONTROL BASIS FOR A COST-EFFECTIVE SELECTIVE SOLDERING PROCESS

LASER TRAPPING MICRO-PROBE FOR NANO-CMM

PASSIVE MICROWAVE IMAGING. Dr. A. Bhattacharya

Lecture 11: Introduction to diffraction of light

Advanced Heat and Mass Transfer by Amir Faghri, Yuwen Zhang, and John R. Howell

Black Body Radiation

Measurement method for the proficiency testing program

TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS

Order No. G Date: September 8, 2010 REPORT NO CRT-001 TEST OF SMF/PAPI SYSTEM RENDERED TO

A Novel and Cost Effective Radiant Heatflux Gauge. Samim Safaei. M.S. Candidate Worcester Polytechnic Institute, Worcester, MA

Coherence and width of spectral lines with Michelson interferometer

Switch. R 5 V Capacitor. ower upply. Voltmete. Goals. Introduction

MEASUREMENTS ACCELERATION OF GRAVITY

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Lab 2: Mach Zender Interferometer Overview

OIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Chapter 6. Fiber Optic Thermometer. Ho Suk Ryou

Switch. R 5 V Capacitor. ower upply. Voltmete. Goals. Introduction

Infrared thermography

Stick-Style Water-Resistant Infrared Thermometer for Foodservice Applications with 4:1 Distance-to-Sight Ratio Model

PHYS 160 Astronomy Test #2 Fall 2017 Version A

Introduction to Infrared Thermometry

Measuring Planck s Constant By Martin Hackworth

ONYX -MCE MULTI-CHANNEL OPTICAL FIBER PYROMETERS WITH ACTIVE EMISSIVITY COMPENSATION PRECISION TEMPERATURE MEASUREMENT FOR DEMANDING INDUSTRIAL

Lecture 3a: Surface Energy Balance

Thermal Resistance Measurement

INFRAMET. 2.1 Basic laws

GS110. IR Thermal Imaging System for Low-Emissivity Glass with Automatic Emissivity Correction. Manual Addendum

Lab Manual: Determination of Planck s constant with x-rays

Edward S. Rogers Sr. Department of Electrical and Computer Engineering. ECE318S Fundamentals of Optics. Final Exam. April 16, 2007.

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

Thermionic emission and Frank-Hertz oscillations

ENVR 416 Aerosol Technology - Laboratory Session Fall 2007

Characterization of high temperature solar thermal selective absorber coatings at operation temperature

Atomic Spectra HISTORY AND THEORY

HL-800K Infrared Laser Thermometer. -50 C to +800 C (13:1 ratio) User Manual

Experimental Uncertainty (Error) and Data Analysis

Lecture 9: Introduction to Diffraction of Light

Lednium Series Optimal X OVTL09LG3x Series

Canal Velocity Indexing at Colorado River Indian Tribes (CRIT) Irrigation Project in Parker, Arizona using the SonTek Argonaut SL

b( ) ) ( ) PROBLEM = W / m = 1991 W / m. = W / m 4 m = 3164 W.

Transcription:

DETERMINING SIZE OF SOURCE FOR HANDHELD INFRARED THERMOMETERS THEORY AND PRACTICE Frank Liebmann Fluke Corporation Hart Scientific Division 799 Utah Valley Dr. American Fork, Utah 84003 801-763-1600 Frank.Liebmann@hartscientific.com Abstract - Infrared (IR) thermometry is a very useful form of temperature measurement. It has advantages over contact thermometry. These advantages typically include a quicker response time and not having to interfere with the system being measured. However, it is typically not as accurate as contact thermometry. Part of the difference in accuracy is due to the fact that IR thermometers themselves are not as accurate as contact thermometers. The other part of the difference can be due to the user not controlling factors which can cause greater uncertainty of measurement. Major uncertainties in IR thermometry can be a result of emissivity, repeatability and size of source effect (SSE). Size of source (sometimes referred to as spot size) can be especially difficult because information on size of source is difficult to find. This paper discusses size of source as it applies to IR thermometers. It covers experimentation done on handheld IR thermometers to determine size of source using accepted practices. It also compares this data to manufacture s specifications given with the tested instruments. It demonstrates how to apply this knowledge to account for the uncertainty caused by not having a source of proper size and shows how to apply this data to an uncertainty budget. SIZE OF SOURCE EFFECT AND SPOT SIZE The purpose of the experimentation described in this paper is to put forth a method for an IR calibrator user to easily evaluate the uncertainty caused by surface non-uniformity combined with size of source effect issues. As a basis for discussion, a few terms and concepts must first be discussed. Size of Source Effect Size of source effect is the phenomenon caused by lens aberrations and by light scattered from within the optical system [1]. The same reference goes on to describe the effect as a change in temperature indication when an automatic optical pyrometer views a radiant target being changed in diameter [1]. In this paper, size of source effect (SSE) will refer to the ratio of the amount of energy contained within a given circle compared to the amount of energy contained within a circle of infinite diameter received by an IR thermometer. This ratio is specified by β throughout this paper. This ratio is used since it is more convenient for calculation of uncertainties at multiple temperatures. The ramification of SSE is that the user of an IR thermometer must be concerned with how large of an area is needed to obtain an accurate measurement. When using an IR calibrator, the user must also be concerned with the uniformity of the calibrating surface. Spot Size Diagrams Many IR thermometers come with spot size diagrams. These tools are a good guideline to determine how much target size is needed for a measurement. However, they can be misleading for two reasons. First, their area only includes a percentage of the entire energy received by the IR thermometer. Second, the percentage of energy within the spot diagram is not necessarily specified by a standard. Thus 80%, 90%, 95% or 99% of the energy or power could be within the specified spot size. This can cause a great amount of confusion and uncertainty when measuring surfaces.

IR CALIBRATOR UNIFORMITY AND SIZE When determining uncertainty, it is important to know the uniformity of the IR calibrator target as well as knowing the IR thermometer s SSE. This becomes more of a factor when the size of the surface being measured is less than what the IR thermometer is actually measuring. In other words, this must be considered when calibrating IR thermometers where 100% of SSE is larger than the calibration source or target size. Hart s Definition of IR Calibrator Uniformity 418X uniformity for a diameter (d) is defined as the difference in average temperature in any circle with a diameter (D) less than or equal to a diameter (d). This average is shown mathematically in (1). T D / π AVG (1) 1 4 ( D) = TdA = T ( r, θ ) rdθdr A πd 0 0 This can be calculated for a given diameter on a flat plate by applying a Riemann sum to samples within the given diameter D. To define the total uniformity for a diameter d, all average temperature values must be computed between 0 and d. The difference of the maximum and minimum average temperature (T AVG-MAX - T AVG-MIN ) between D=0 and d is the uniformity for diameter d as shown in (). TUNIF ( d) = TAVG ( DMAX ) TAVG ( DMIN ) where DMIN < d D < d MAX D MAX is the diameter of largest average temperature D MIN is the diameter of smallest average temperature () To measure this distribution (T(r,θ)), an IR thermometer measures a number of points on the target s surface. These measurements demonstrate the surface s temperature gradient. Uniformity Testing at Hart Uniformity testing done at Hart Scientific has found that uniformity as tested on the 418X IR Calibrators has been well within the published specification limits. These specifications are shown in Table 1. This testing means that a user of these instruments can feel comfortable using the specifications to calculate uncertainty due to SSE combined with target uniformity. Table 1: 418X IR Calibrator Uniformity Specifications Model Temperature Uniformity at 50mm Uniformity at 15mm 4180-15 C ±0.10 C ±0.15 C 4180 0 C ±0.10 C ±0.10 C 4180 10 C ±0.0 C ±0.5 C 4181 35 C ±0.10 C ±0.10 C 4181 50 C ±0.5 C ±0.50 C 4181 500 C ±0.50 C ±1.00 C

SIZE OF SOURCE EFFECT TESTING The purpose of this SSE testing was two-fold. First, knowledge of how manufacture s spot size diagrams relate to SSE within a circle was desired. Second, a practical method was needed to test SSE for handheld IR thermometers. This method needed to be inexpensive and non-time-consuming, but still have good repeatability. Practical Size of Source Testing at Hart To test SSE for handheld devices, a simple method was devised based on a Target Size Test Method [] outlined by ASTM. A set of apertures was constructed at increasing diameters from 0mm to 150mm at 1.5mm increments. They were constructed out of an inexpensive material. The aperture holes were placed so that they would be centered on the center of the source. A fixture to hold the apertures was constructed. The source used for this test was a Hart Scientific 4181 IR Calibrator. For most of the tests, the target was controlled at 50 C. The one exception was IR Thermometer which has a measurement limit of 00 C. The ASTM standard states the temperature source is stabilized at a value near the top of the calibration range of the radiation thermometer []. For this test, 50 C was chosen due to safety concerns with the use of the apertures at higher temperatures. The apertures were placed in front of the target at a distance of 100mm. The handheld thermometer was mounted on a tripod. It was centered on the aperture at a specified distance from the aperture. For smaller apertures, the IR thermometer was moved in the horizontal and vertical position until the signal was maximized. For the larger apertures, the center was found by finding where a 1% drop off in energy was located on either side, then going to the center of these two points. This was done in both the horizontal and vertical directions. The temperature reading of the measurements through the apertures was recorded. The test was done three times to ensure repeatability. The temperatures were used to compute percent of signal at each temperature. ASTM suggests an equation for a narrowband instrument based on Planck s Law []. However, since all the handhelds tested were wideband instruments, Planck s Law was numerically evaluated for the instruments bandwidth. This equation is shown in (3). The emissivity setting of the IR thermometer was taken into account. Some of the instruments do not have an adjustable emissivity setting. In all cases except for IT Thermometer, the emissivity setting is 0.95. For IR Thermometer, the emissivity setting is 0.97. S β ( D) = ε λ λ S = S c1 c1 dλ + ( 1 ε ) 5 c λ1 5 c λ exp 1 λ exp λtmeas λt 1 1 ( D) S S MAX ( 0) ( 0) λ BG dλ (3) In (3), the measured temperature for each aperture diameter is T MEAS. T BG is what the IR thermometer uses for the background temperature in its temperature calculation. This number is measured and calculated internal to IR thermometers. For this test, 3 C was used for calculation. The values λ 1 and λ are the spectral band limits of the IR thermometer. For instance, if the IR thermometer is an 8 14 μm instrument, λ 1 and λ will be 8 and 14 μm respectively.

S (0) is the signal of the measurement made with the 0mm diameter aperture. The S -MAX is the maximum signal measured in a set of measurements. S (D) is the signal for a given diameter aperture. β(d) is a calculation of the ratio of the energy received from the target at a given aperture diameter compared to the maximum signal received. Using (3) accounts for the energy from background radiation and the emissivity setting of the IR thermometer. This is predicted by Kirchoff s Law [1]. It also accounts for signal from the aperture, accounted for by S (0). The results of these tests are shown in Table. A curve fit was performed on the SSE data for the purpose of analysis. The target equation for the curve fit is a variant of the Weibull Cumulative Distribution Function [3] shown in (4). Distance Table. Comparison of SSE Testing Results Spot Size Ratio of Maximum Power (#) (mm) (mm) 1xSS xss 3xSS 4xSS 1 380 38 0.884 0.970 0.997 1.000 100 40 0.88 0.996 1.000 1.000 3 300 38 0.940 0.993 0.999 1.000 4 300 5 0.816 0.97 0.99 0.997 5 300 4 0.971 0.998 1.000 1.000 6a 300 19 0.98 0.998 1.000 1.000 6b 600 19 0.980 0.993 0.997 0.999 7 1150 19 0.898 0.955 0.963 0.975 a β = 1 c D exp b 1 a e D exp d (4) Table 3 gives the typical repeatability of the tests. The repeatability at smaller aperture sizes tended to be worse. This is most likely due to lack of signal coupled with alignment difficulty using smaller apertures. Table 3. Generalized Repeatability of Measurements Typical Ratio of Max Power (β) Repeatability 0.000 to 0.800 0.008 0.800 to 0.900 0.008 0.900 to 0.950 0.003 0.950 to 0.980 0.00 0.980 to 0.990 0.00 0.990 to 0.999 0.001 1.000 0.000

Comparison to Manufactures Spot Size Diagrams As shown in Table, the SSE data does not always correlate to the spot size shown in the manufacture s documentation. This is because scatter is not always considered as part of the spot size in these devices. For most of the IR thermometers tested, 99.9% of the energy was with in 4 times the spot size. At diameters less than 4 times the spot size, the results varied. For most of the IR thermometers tested, their specifications state that typically 90% of the energy is within the specified spot size. The two exceptions to this are IR Thermometer 1 (specified at 80%) and IR Thermometer 7 (not specified). In the case of IR Thermometer 7, the test showed 90% of the energy being within 1 spot size. For the rest of this paper, examples will be based on IR Thermometer 3 s test results. Figure 1 shows the Cumulative Distribution of Power versus diameter with a curve fit for IR Thermometer 3. The curve fit is based on a Weibull Distribution as shown in (4). The vertical line and arrow represent the spot size specification in the IR thermometer s manual. Note the decaying effects of scatter beyond the arrow. Figure 1. IR Thermometer 3: Spot Size and Cumulative Distribution of Power Signal SSE Testing CDF #3 1.000 0.900 0.800 0.700 0.600 0.500 0.400 0.300 0.00 0.100 0.000 0 50 100 150 Diameter (mm) Measured Curve Fit Simplified Mathematics Since (3) may be difficult to evaluate, consideration was given to using Planck s Law for a single wavelength shown in (5). IR Thermometer 3 is an 8 14 μm instrument. For λ, a value of 10 μm was used. This is because at 50 C, the radiated energy tends to be closer to the lower wavelengths. S c1 = ε 5 c λ exp λt MEAS Δλ + 1 1 ( 1 ε ) Δλ c 5 c λ exp λt BG 1 (5) The difference between calculation using the simplified method shown in (5) and the integrated method from (3) is shown in Figure. Most of the error occurs at the smaller aperture diameters.

Figure : Error Caused by Simplified Equation Error Caused by Simplified Equation Error (ratio of maximum signal) 0.004 0.003 0.00 0.001 0.000-0.001-0.00-0.003-0.004 0 50 100 150 Aperture Diameter (mm) A second equation that was considered is included in the ASTM standard []. This equation is modified to calculate β and is shown in (6). It is meant to calculate the SSE at levels of 99% and above and may not be suited for calculating lesser SSE. The ASTM standard is for radiation thermometers with emissivity settings of 1.00 used to measure blackbodies. This type of measurement is not always possible with handheld instruments. ΔT β = 1 c 1 exp λt c λt (6) The difference between calculation using the ASTM equation shown in (6) and the integrated method from (3) is shown in Figure 3. Most of the error occurs at the smaller aperture diameters. Figure 3. Error Caused by ASTM Equation Error Caused by ASTM Equation 0.080 Error (ratio of maximum signal) 0.060 0.040 0.00 0.000-0.00-0.040 0 50 100 150 Aperture Diameter (mm)

ACCOUNTING FOR UNCERTAINTY Once the SSE is characterized, this data can be used to calculate uncertainty caused by SSE and IR calibrator uniformity issues. Using Test Data and Uniformity Data If 100% of the signal is contained within the target size, the target s uniformity specification can be used completely for evaluation. For this example, the 4181 s uniformity specification is used. At 50 C, the uniformity is specified as ±0.5 C at a 50mm diameter and as ±0.50 C at a 15mm diameter. This means that 100% of the signal will have an uncertainty of ±0.5 C. Referring to the graph in Figure 1, 3% of the signal comes between the diameters of 50mm and 15mm. This will cause an additional uncertainty of 3% of ±0.50 C or ±0.0 C. Since these uncertainties are correlated, they should be summed giving an uncertainty of ±0.7 C. Having Less Source then SSE A different problem occurs if the source is smaller than the SSE needed to get an accurate measurement from an IR thermometer. To illustrate this problem, a 6.5mm diameter target at 50 C will be considered. For simplicity, it is assumed the target has complete temperature uniformity on its surface. From the SSE testing illustrated in Figure 1, it has been demonstrated IR Thermometer 3 has 98.3% of its detected signal within this diameter. This means that when the IR thermometer is properly centered on the target, 1.7% of its signal will come from areas not on the target. By applying the inverse of (3), it can be calculated that this can cause up to.7 C of error in the calibration. Inclusion in an Uncertainty Budget Using the testing method described in Section 3 with the analysis from this section, type B uncertainties can be calculated as shown above. These uncertainties can be quite large and should be considered for any IR thermometer calibration. CONCLUSION Knowledge of size of source is important for the calibration and use of IR thermometers. To determine size of source, a simple test can be done as outlined in this paper. This helps the user of these instruments to determine uncertainty caused by the uncertainty of surface uniformity of an IR calibrator. Even more important, uncertainty can be calculated for the case when the IR calibrator source is smaller than 100% of the IR thermometer s SSE. This uncertainty calculation will lead to better knowledge of the accuracy of IR thermometer calibrations and measurements. ACKNOWLEDGEMENTS The author would like to express thanks to the many individuals that helped in this research and research into uncertainties in general at Fluke Hart Scientific. Special thanks goes out to Tom Wiandt of Fluke Hart Scientific for his guidance in developing IR testing and uncertainty analysis at Hart, Stephen King of Raytek for his help in understanding the nature of IR optics and scatter, and the NIST Optical Technology Division for their advice and education in radiation thermometry. REFERENCES [1] D.P. DeWitt G.D. Nutter, Theory and Practice of Radiation Thermometry. Wiley Interscience, New York, 1988, pp. 70-7, 409-411.

[] E 156-95 in Annual Book of ASTM Standards Vol. 14.03. ASTM International, West Conshohocken, PA, 005, pp. 490-491. [3] NIST/SEMATECH e-handbook of Statistical Methods. 7/18/006. National Institute of Standards and Technology. <http://www.itl.nist.gov/div898/handbook/eda/section3/eda3668.htm>.