LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Similar documents
TANDEM MASS SPECTROSCOPY

Analysis of Polar Metabolites using Mass Spectrometry

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

LC-MS Based Metabolomics

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

Mass Spectrometry in MCAL

CEE 772: Instrumental Methods in Environmental Analysis

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

M M e M M H M M H. Ion Sources

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry

Molecular Mass Spectrometry

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

MS Goals and Applications. MS Goals and Applications

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Choosing the metabolomics platform

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Mass Spectrometry (MS)

Molecular Mass Spectrometry

CHROMATOGRAPHY AND MASS SPECTROMETER

1. The range of frequencies that a measurement is sensitive to is called the frequency

Introduction to LC-MS

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass spectrometry.

Translational Biomarker Core

Quattro Micro - How does it work?

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

MS Goals and Applications. MS Goals and Applications

Introduction to the Q Trap LC/MS/MS System

International Journal of Pharmaceutical Research & Analysis

Selecting Detectors for Compounds with No Optical Absorbance

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry Instrumental Analysis Lecture 34. Chem 4631

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Ch. 22 Mass Spectrometry (MS)

Types of Analyzers: Quadrupole: mass filter -part1

Liquid Chromatography in conjunction with Mass Spectrometry (LC-MS)

Introduction to Mass Spectrometry (MS)

High Pressure/Performance Liquid Chromatography (HPLC)

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Lecture 15: Introduction to mass spectrometry-i

Mass Spectrometry. Electron Ionization and Chemical Ionization

Mass Spectrometry for Chemists and Biochemists

Chromatography & instrumentation in Organic Chemistry

LC MS analysis of metabolites. Basis of Chromatography

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Mass Spectrometry. What is Mass Spectrometry?

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Harris: Quantitative Chemical Analysis, Eight Edition

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

ABI 3200 Q TRAP LC/MS/MS System

Spectroscopy and Chromatography

Technical Note. Introduction

1) In what pressure range are mass spectrometers normally operated?

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008

Particle Position Relative Mass Relative Charge Proton Nucleus 1 +1 Neutron Nucleus 1 0 Electron Orbitals 1/ Atomic Symbol

BST 226 Statistical Methods for Bioinformatics David M. Rocke. January 22, 2014 BST 226 Statistical Methods for Bioinformatics 1

Basics of Mass Spectrometry

Presentation Basic Introduction to Instrumentation Matrix Effects Challenges

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

2101 Atomic Spectroscopy

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Mass spectrometry: forming ions, to identifying proteins and their modifications Stephen Barnes, PhD

This is the total charge on an ion divided by the elementary charge (e).

Webinar: How to be More Successful with your ACQUITY QDa Detector Eric van Beelen

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

Introduction to GC/MS

Principles of Gas- Chromatography (GC)

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Chromatography. Mrs. D. MEENA MPharm PA & QA KTPC

Mass spectrometry: forming ions, to identifying proteins and their modifications Stephen Barnes, PhD

Mini review for final

Chapter 5. Complexation of Tholins by 18-crown-6:

Introduction to Pharmaceutical Chemical Analysis

Electrospray Ion Trap Mass Spectrometry. Introduction

Mass Spectrometry - part 1


Hplc Lc Ms And Gc Method Development And Validation Guideline For Academic And Industrial Scientists Involved In Method Development And Validation

Research Equipment in the UCA Chemistry Department

Secondary Ion Mass Spectroscopy (SIMS)

Chapter content. Reference

770-9P. Sensitivity and Selectivity - A Case Study of LC/MS Enantioselective Resolution of Bupivacaine Using Vancomycin as a Chiral Stationary Phase

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

BIOINF 4399B Computational Proteomics and Metabolomics

Introduction. Chapter 1. Learning Objectives

CHIRAL SEPARATION USING THIN LAYER CHROMATOGRAPHY

Transcription:

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid chromatography with mass analysis capability of Mass spectrometry. It is a method that combines separation power of HPLC with detection power of Mass spectrometry. In LC-MS we remove the detector from the column of LC and fit the column to interface of MS. In the most of the cases the interface used in LC-MS are ionization source.

Theory of LC/MS HPLC is a method for separating a complex mixture in to its components. High sensitivity of mass spectroscopy provides the information for identification of compounds or structural elucidation of compounds. Combination of these two techniques is LC-MS As the metabolites appear from the end of the column they enter the mass detector, where the solvent is removed and the metabolites are ionized.

LC-MS System Components Mass spectrometers work by ionizing molecules and then sorting and identifying the ions according to their mass-to-charge (m/z) ratios.

PROBLEMS IN COMBINING HPLC AND MS HPLC Liquid phase operation 25-50 deg. C No mass range limitations Inorganic buffers 1 ml/min eluent flow is equivalent to 500 ml/min of gas MS Vacuum operation 200-300 deg. C Up to 4000 Da for quadrupole MS Requires volatile buffers Accepts 10 ml/min gas flow

MOBILE PHASE The mobile phase is the solvent that moves the solute through out column. General requirements:- (1) Low cost, UV transparency, high purity. (2) Low viscosity, low toxicity, non flammability. (3) Non corrosive to LC system component. Solvent strength and selectivity:- It is the ability of solvent to elute solutes from a column.

COLUMN The use of di-functional or tri-functional silanes to create bonded groups with two or three attachment points leading to phases with higher stability in low or higher ph and lower bleed for LCMS Most widely used columns for LC-MSare:- (1) fast LC column. the use of short column. (15-50mm) (2) Micro LC column. theuse of large column. ( 20-150mm)

Sample preparation Sample preparation generally consists of concentrating the analyte and removing compounds that can cause background ion or suppress ionization. Example of sample preparation include:- 1. On Column concentration -to increase analyte concentration. 2. Desalting - to reduce the sodium and potassium adduct formation that commonly occurs in electro spray. 3. Filtration- to separate a low molecular-weight drug from proteins in plasma, milk, or tissue.

INTERFACES LC-MS systems include a device for introducing samples (such as an HPLC )an interface for connecting such device, an ion source that ionizes samples, an electrostatic lens that efficiently introduces the generated ions, a mass analyzer unit that separates ions based on their mass-to-charge (m/z) ratio, and a detector unit that detects the separated ions. In an LC-MS system, however, if the LC unit is simply connected directly to the MS unit, the liquid mobile phase would vaporize, resulting in large amounts of gas being introduced into the MS unit. This would decrease the vacuum level and prevent the target ions from reaching the detector. So interfaces are to be used.

TYPES OF INTERFACES It is difficult to interface a liquid chromatography to a massspectrometer cause of the necessity to remove the solvent. The commonly used interfaces are:- (1) Electrospray ionization (ESI) (2) Thermospray ionization (TSI) (3) Atmospheric pressure chemical ionization (APCI) (4) Atmospheric pressure photoionization(appi)

Electro Spray Ionization (ESI) ESI draws sample solutions to the tip of a capillary tube, where it applies a high voltage of about 3 to 5 kv. A nebulizer gas flows from outside the capillary to spray the sample. This creates a fine mist of charged droplets with the same polarity as the applied voltage. As these charged particles move, the solvent continues to evaporate, thereby increasing the electric field on the droplet surface. When the mutual repulsive force of the charges exceeds the liquid surface tension, then fission occurs. As this evaporation and fission cycle is repeated, the droplets eventually become small enough that the sample ions are liberated into the gas phase

Electro Spray Ionization (ESI) ESI provides the softest ionization method available, which means it can be used for highly polar, least volatile, or thermally unstable compounds.

Atmospheric pressure chemical ionization (APCI) APCI vaporizes solvent and sample molecules by spraying the sample solution into a heater (heated to about 400 C) using a gas, such as N2. Solvent molecules are ionized by corona discharge to generate stable reaction ions.

Thermospray ionization (TSI) They are of 2 type: a) Real-TSP ionization b) Discharge electrode for external ionization and repeller electrode

Atmospheric pressure photoionization(appi) The LC eluent is vaporized using a heater at atmospheric pressure. The resulting gas is made to pass through a beam of photons generated by a discharge lamp (UV lamp) which ionizes the gas molecules.

Mass Analyser They deflect ions down a curved tubes in a magnetic fields based on their kinetic energy determined by the mass, charge and velocity. The magnetic field is scanned to measure different ions. Types of mass analyzer:- (1) Quadrapole mass filter. (2) Time of flight (3) Ion trap (4) Fourier transform ion cyclotron resonance (FT-ICR)

Quadrupole Mass Analyzer A Quadrupole mass filter consists of four parallel metal rods with different charges Two opposite rods have an applied + potential and the other two rods have a - potential The applied voltages affect the trajectory of ions traveling down the flight path For given DC and AC voltages, only ions of a certain mass-to-charge ratio pass through the quadrupole filter and all other ions are thrown out of their original path

TOF (Time of Flight) Mass Analyzer TOF Analyzers separate ions by time without the use of an electric or magnetic field. In a crude sense, TOF is similar to chromatography, except there is no stationary/ mobile phase, instead the separation is based on the kinetic energy and velocity of the ions.

Ion Trap Mass Analyzer It uses an electric field for the separation of the ions by mass to charge ratios. The electric field in the cavity due to the electrodes causes the ions of certain m/z values to orbit in the space.

Fourier transform ion cyclotron resonance (FT-ICR) Uses a magnetic field in order to trap ions into an orbit inside of it. In this analyzer there is no separation that occurs rather all the ions of a particular range are trapped inside, and an applied external electric field helps to generate a signal.

Applications of LC-MS Pharmaceutical Applications: Rapid chromatography of benzodiazepines Identification of bile acid metabolite Biochemical Applications: Rapid protein identification using capillary LC/MS/MS and database searching. Clinical Applications: High-sensitivity detection of trimipramine and thioridazine

Applications of LC-MS Food Applications: Identification of aflatoxins in food Determination of vitamin D3 in poultry feed supplements Environmental Applications: Detection of phenylurea herbicides Detection of low levels of carbaryl in food Forensic Applications: illegal substances, toxic agents Explosives

References Instrumental Methods Of Chemical Analysis By Willard Instrumental Methods of Analysis By Gurdeep Chatwal Textbook of Pharmaceutical Analysis - By Ravi Shankar Instrumental Methods of Analysis - By Skoog. www.impactanalytical.com/lc-ms-analysis.asp www.shimadzu.com/an/lcms/index.html www.enotes.com/topic/liquid_chromatography-mass_spectrometry www.spectroscopynow.com www.sciencedirect.com Raffaelli A, Saba A. Atmospheric pressure photoionization spectrometry. Mass Spectrum Rev. 2003;22:318 31 Google Images mass