GPC/SEC Column Selection & Method Development

Similar documents
Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE

Clearing the Confusion: GPC, SEC, GFC What, When, Why, and How?

Gel Permeation Chromatography Basics and Beyond eseminar March 13, Jean Lane Technical and Applications Support LSCA, Columns and Supplies

Practical Steps in GPC Method Development

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

GPC/SEC An essential tool for polymer analysis

Comparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation

GFC separation of water-soluble polymers with TSKgel PW-type columns (2) using the PWXL series. Contents

Optimizing GPC Separations

Maximizing Performance Through GPC Column Selection

Polymer analysis by GPC-SEC. Technical Note. Introduction

GPC/SEC Troubleshooting Volume 1

Gel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC)

Aqueous and polar GPC/SEC columns. Product guide

No.106. Aqueous SEC Columns for Analysis of Cationic Polymers: TSKgel PWXL-CP Series. Contents. Page

Lecture 4 : Gel Permeation or Size Exclusion Chromatography

GPC - Gel Permeation Chromatography. aka Size Exclusion Chromatography- SEC

GPC/SEC standards. Product guide

Dr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1

GPC/SEC standards. Product guide

Tips & Tricks GPC/SEC: Inter-Detector Delay

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 25: CHROMATOGRAPHIC METHODS AND CAPILLARY ELECTROPHORESIS

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

An Introductions to Advanced GPC Solutions

GPC/SEC Start-up Kit

FAQ's. 1 - Which column would be the most appropriate for my application?

Tips & Tricks GPC/SEC: Protein Analysis with Size-Exclusion Chromatography

GFC Analysis of Water-Soluble Polymers with TSKgel PW-type Columns. Contents

High Performance Liquid Chromatography

High Performance Liquid Chromatography

AGILENT AQUEOUS AND POLAR GPC/SEC COLUMNS

Tips & Tricks GPC/SEC: From a Chromatogram to the Molar Mass Distribution

GPC-/SEC-Columns. MZ-Gel SD. MZ-Gel Super FG SEC with fluorinated Eluents

Latest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments

Chapter content. Reference

Determination of Polymer Modifier in Asphalt

Open Column Chromatography, GC, TLC, and HPLC

Macromolecular Chemistry

Gel Permeation Chromatography - GPC

High Performance Liquid Chromatography

Cation Exchange HPLC Columns

Waters GPC User Guide and Tutorial for Using the GPC in the Reynolds Research Group 2 nd Edition: April 2012

LC Technical Information

Packings for HPLC. Packings for HPLC

Analysis of Star Polymers Using the Agilent 1260 Infinity Multi-Detector GPC/SEC System

High Performance Liquid Chromatography

Phenogel. GPC/SEC Columns. Sample Elution. Technical Specifications 10 3 Å 10 6 Å

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Polymer Analysis by Gel Permeation Chromatography

Biochemistry. Biochemical Techniques HPLC

Sem /2007. Fisika Polimer Ariadne L. Juwono

PHENOGEL GPC/SEC COLUMNS

Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques

Shodex TM ODP2 HP series columns

Comprehensive Polymer Analysis Strategies

Protein separation and characterization

Organic GPC/SEC columns. Product guide

Particle Characterization Laboratories, Inc.

[ CARE AND USE MANUAL ]

LC and LC/MS Column Selection Flow Chart

Liquid Chromatography

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

HPLC Background Chem 250 F 2008 Page 1 of 24

penta-hilic UHPLC COLUMNS

for Acclaim Mixed-Mode HILIC-1 Column

Instrumental Chemical Analysis

C18 Column. Care & Use Sheet

LC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i

Sepax SRT -C, Zenix -C SEC Phases

How to use GPC/SEC for compositional analysis

Characterization of polyphenylene sulphide using the Agilent PL-GPC 220 High Temperature GPC System with triple detection

GPC Streamliner. » Continues on Page 2. Focus: accurate GPC/SEC. Sieve Curves. In This Issue

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

Liquid Chromatography

Benefit of light scattering technologies (RALS/LALS/MALS) and multidetection characterization in life science research?

Chromatographic Separation

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

Dilution(*) Chromatography

Setting the Standard for GPC. Complete Guide for GPC / SEC / GFC Instrumentation and Detection Technologies. The Right Instrument for Your Application

Solvent Flexibility for Size-Based Polymer Analysis Using the Advanced Polymer Chromatography (APC) System

ACQUITY UPLC HSS Columns

TSKgel Size Exclusion Chromatography Columns

Optical Isomer Separation Columns and Packing Materials

Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis

ProPac WCX-10 Columns

Polymer Separations by Chemical Composition Using SEC-Gradients

HPLC COLUMNS WILEY-VCH. Theory, Technology, and Practice. Uwe D. Neue with a contribution from M. Zoubair El Fallah

TSKgel Size Exclusion Chromatography Columns

Prelab Reading Assignment: Laboratory Techniques in Organic Chemistry, 4 th Ed. Chapter 19

Chemistry Instrumental Analysis Lecture 31. Chem 4631

EcoSEC - semi-micro GPC/SEC system for high throughput polymer analysis

Polymers Reactions and Polymers Production (3 rd cycle)

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

Guide to Choosing and Using Polymer Reversed Phase Columns!

Size Exclusion Chromatography: Method Development

The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction

Instrumental Analysis II Course Code: CH3109. Chromatographic &Thermal Methods of Analysis Part 1: General Introduction. Prof. Tarek A.

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Transcription:

GPC/SEC Column Selection & Method Development LIVE WEBCAST: Tuesday April 27, 2010 Topics Introduction to GPC/SEC Column Characteristics & Column Selection Additional Method Development Troubleshooting Page 1

Introduction to GPC/SEC GPC: Gel Permeation Chromatography SEC: Size Exclusion Chromatography GFC: Gel Filtration Chromatography GPC/SEC/GFC is pore-size limited, diffusion-controlled liquid chromatography (LC) method. Page 2

Introduction to GPC/SEC Separation mechanism GPC/SEC Column Filled with porous particles; diffusion into the pores when size fits Sample Mixture size : hydrodynamic radii color: different chemistry GPC/SEC Detector Elugram Signal Intensity vs. Elution Volume; 1) solvent elutes 2) large sizes 3) small sizes Page 3

Introduction to GPC/SEC Required Items Columns are the heart of the system! Without sufficient separation only limited information (even when intelligent detection is used). Isocratic pump Injection system GPC/SEC data acquisition and evaluation software 1 or more detectors (depending on application) Calibration Standards, Reference Materials, Validation Standards Page 4

Introduction to GPC/SEC Applications a) Polymer/Biopolymer characterization using various detectors b) Separation of mixtures into discrete fractions Typical samples Synthetic polymers, molar mass range 100 to 30 000 000 Da Synthetic copolymers Water soluble (charged) polymers, molar mass range 100 to 30 000 000 Da Proteins Polysaccharids Polypeptides. Page 5

Introduction to GPC/SEC Results GPC/SEC Results Complete Molar mass distribution (MMD) M n, M w, M z, PDI % material above/below MM ranges [h], branching index g, structure Radius of gyration, branching, structure Composition (Copolymers) Concentration residual monomer, solvent, additive content Substance Idenfification Comment One of a few methods! + other averages On-line viscometer required On-line MALLS required 2 concentration detectors required Detector response calibration required FTIR/NMR/ESI-MS/MS required Page 6

Introduction to GPC/SEC From Detector Elugram to Molar Mass Distribution Elugram Y: Signal Intensity X: Elution Volume Left: high molar masses Right: low molar masses Sample information AND system parameters contribute to V e required for MMD: (log) M = f(v e ) Page 7

Introduction to GPC/SEC From Detector Elugram to Molar Mass Distribution How to get the molar mass M at a elution volume V e : Calibration with narrow or broad standards, universal calibration with Mark-Houwink coefficients Universal calibration curve with on-line viscometer Measure M online with on-line light scattering detector (LS, dn/dc required) Page 8

Introduction to GPC/SEC Calibration Curve (depends on accessible pore volume) Page 9

Introdcution to GPC/SEC From Detector Elugram to Molar Mass Distribution Molar Mass Distribution Y: w(logm) (transformed!) X: M (log scale, to compress) Left: low molar masses Right: high molar masses System parameters eliminated when using the calibration curve only sample information present Inter-laboratory sample comparison Page 10

Introduction to GPC/SEC Calibration Curve Has all information about the column material s pore size distribution! (Used in inverse SEC to characterize networks!) Users can learn about Molar mass separation range Resolution in a molar mass range Page 11

Introduction to GPC/SEC Mixture Polystyrene injected onto 2 different column sets (overlaid with corresponding calibration curve) Slope of the calibration curve: Steep Shallow/Flat low resolution (in this molar mass area) high resolution (in this molar mass area) Page 12

Column Characteristics Ideal GPC/SEC column Minimal/No interactions between sample and column material (stationary phase) High resolution (large pore volume) Minimal shear degradation (especially for high molar masses) Low column backpressure Good mechanical, chemical and thermal stability Page 13

Column Characteristics Other important analytical tasks Information content High throughput Time savings Save eluent High sample loading for fractionation Precision Analytical Challenge Time Environmental issues Page 14

Column Characteristics Optimize 4 Parameters to find the best column(s) for your application. Column Type/Dimensions Chemistry Column Material Particle Size Column Material Porosity Column Material Page 15

Column Characteristics Column Type/Dimensions Task: Find the column that meets your application needs Typical Dimensions [mm] Preparative HighSpeed Analytical Semi-micro ID: 20 Length: 300/600 ID: 20 Length: 50 ID: 8 mm Length: 300 ID: 4.6 mm Length: 250 Flow-rate [ml/min] 3-10 3-10 0.5 1 0.2-0.5 Time consumption per column [min] 12.5 2 12.5 10 Eluent consumption per column [ml] 78.1 (6.25 ml/min) 12.5 (6.25 ml/min) 12.5 (1 ml/min) 3.5 (0.3 ml/min) Maximum loading per column* 10 mg/ml, Up to 1000 µl 2 mg/ml, 20 µl 2 mg/ml, 20 µl 2 mg/ml, 5 µl * Depends largely on molar mass and polydispersity, loading maximized (not separation) Page 16

Column Characteristics Chemistry: Column Material Column materials: - mostly designed either for organic solvents or aqueous solvents - either polymer beads (copolymer networks, usually manufactured via suspension polymerization) or silica based Page 17

Column Characteristics Chemistry: Column Material Task: GPC/SEC must be interaction free Apply different solvent interaction dominates Page 18

Column Characteristics Chemistry: Column Material Polymer Packing Inorganic packing Polarity Non Medium Polar Non Medium Polar Chemistry St-DVB Acrylic, Polyester OH-acrylic SiO 2 -C 18 SiO 2 -diol SiO 2 Solvents THF, TCM, Toluol DMF, NMP, DMAc, DMSO, (all) Water (ph 1.5 to 13) Aqueous (ph < 7) Examples SDV, POLEFIN GRAM SUPREMA, NOVEMA PFG PROTEEMA Samples PS, PMMA PVC, PC, Resins, PE, PP, etc. PU, Starch, Cellulose, Polyimide etc. PEG/PEO, Pullulan, Dextran, PAA, etc. Proteins, Petides, PLA, Polyester, POM, etc. Page 19

Column Characteristics Chemistry: Column Material Task: GPC/SEC must be interaction free Balance Polarity: Sample Solvent Stationary Phase To avoid doing HPLC on a GPC/SEC column! Page 20

Molecular weight [g/mol] Molecular weight [g/mol] Intensity [V] Intensity [V] Column Characteristics -0,035 Solvent: DMF -0,004-0,006-0,040-0,008-0,010-0,045-0,012-0,014-0,050-0,016 7,0 7,5 8,0 8,5 9,0 9,5 10,0 10,5 11,0 11,5 12,0 12,5 13,0 Elution volume [ml] -0,018 7 8 9 10 11 12 13 Elution volume [ml] Polystyrene 1620 Da on divinylbenzene gel Polystyrene 1620 Da on polyester gel 10 5 10 4 PS PMMA PEO 10 5 PS PMMA PEO 10 4 10 3 10 3 6 8 10 12 Elution volume [ml] 5 6 7 8 9 10 11 12 Elution Volume (ml] Calibration curves on divinylbenzene gel Calibration curves on polyester gel Page 21

Column Characteristics Particle Size Optimum packing and resolution: sperical particles with narrow particle size distribution lower particle size results in: higher plate count higher resolution higher column backpressure higher shear forces lower mechanical stability (large porosity) Page 22

Column Characteristics Particle Size High viscous solvents (e.g. water, DMAc) use higher particle sizes to reduce back pressure and shear forces. High molar masses use higher particle sizes to avoid shear degradation. Typical particle sizes Linear polymers > 10 6 /10 7 Da Oligomers, proteins 5 µm (low viscous solvents) 10 µm (high viscous solvents) 10-20 µm (to avoid flow-induced degradation) 3 µm (reduce flow-rate to reduce column back pressure) Page 23

Column Characteristics Particle Size 660 000 Da 17 000 Da Influence particle size on resolution: Same protein mixture injected using the same chromatographic conditions but columns with different particle size: PSS SUPREMA 10µm (8x300 mm) PSS SUPREMA 5µm (8x300 mm) Page 24

Column Characteristics Porosity Linear/Mixed Bed Column Single Porosity Column Many different pore sizes in the same column (by syntesis or mixture of single porosity materials) Constant resolution over a wide molar mass range, wide linear molar mass range 100Å 100 000Å Narrow pore size distribution Highest resolution in a smaller molar mass range Page 25

Molar mass Column Characteristics Porosity Pore type Advantage Disadvantage Single porosity column Efficient, optimized, low cost Viscous fingering Linear column Fast, universal, inject band dilution Low efficiency Page 26

Column Characteristics Porosity Use column combination/bank to increase: Resolution: Add column of the same porosity (increase accessible pore volume) Doubling column length increases resolution by a factor of 1.4 ( Length ). Separation range: Add column of different porosity Avoid: column mismatch Disadvantages of column combinations: increased analysis time, eluent consumption, back pressure Page 27

Column Characteristics Porosity Use multiple columns of the same porosity to increase resolution Dimer: 36 000 Da Monomer: 18 000 Da Dimer: 36 000 Da Monomer: 18 000 Da Protein (Monomer/Dimer): PSS SUPREMA 5µm, 100Å (8x300 mm) PSS SUPREMA 5µm, 2x 100Å (8x300 mm each) Page 28

Column Characteristics Porosity Increase separation range: add column of different porosity but be aware of column mismatch, do not confuse mismatch with higher resolution. Mismatch Mismatch If the pore size distributions of the columns do NOT match: bimodal peaks, peak shoulders may occur Exclusion Example data: Dextrans on a column bank with porosity mismatch Page 29

Column Characteristics Porosity Mismatch-free combination/bank Example data: Dextrans on a column bank SUPREMA 1000 Å and 30 Å Page 30

Column Characteristics Porosity Calibration curve can show mismatch (2 slopes), but this must not be the case. Potentially problematic are the following combinations: Linear column with low porosity single porosity column Example: Linear + 100 Å to increase resolution for oligomers Single porosity column combinations with a wide gap in porosity Example: 10 000 000 Å + 1000 Å Page 31

PSS WinGPC scientific V 6.20, Instanz #1 Detektor Signal [V] Column Characteristics Porosity Calibration curve can show mismatch (2 slopes), but this must not be the case. -- linear and 100 Å -- 10 7 Å and 1000 Å -- mismatch-free 100 -- linear and 100 Å -- 10 7 Å and 1000 Å -- mismatch-free 90 80 70 60 50 40 30 20 10 0 10 11 12 13 14 15 16 17 18 Elution Volume [-] Polystyrene: 48 000 g/mol Page 32

Column Characteristics Porosity How to avoid column mismatch? Use column combinations/banks recommended by the manufacturer. Never combine linear columns with single porosity column. If in doubt (e.g. because of unexpected shoulders): - have a look at the calibration curve - overlay different samples and see if the shoulder appears always at the same elution volume - use broad standards to detect column mismatch Page 33

Additional Method Development For macromolecules with or without functional groups neutral polymers negatively charged polymers positively charged polymers Application in organic solvents aqueous solvents Page 34

Additional Method Development Why additional method development and solvent additives? To overcome non-size exclusion effects. Potential problems: ion exchange ion inclusion/exclusion (intramolecular) electrostatic interactions adsorption sample aggregation Column material is made by suspension polymerization, there are potential residuals/charges on column and imperfections in surface chemistry. Adding electrolyte eliminates ion exchange/ion exclusion, aggregate formation and shields electrostatic interactions. Page 35

Additional Method Development Salt effect DMF 0.05m LiBr Norbonenderivative GRAM 10µm 100 Å DMF Method development includes variation of ionic strength. 3 4 5 6 7 8 9 10 V e /ml Page 36

Additional Method Development Organic Solvents funct. Group -OH -NH 2 -COOH suitable additives Alcohol (Methanol, Ethanol) Diethylamine Acetic acid, Trifluoro acetic acid (TFA) Solvent: + 5g/L LiBr (DMF, DMAc, DMSO) + Additive: 0,05...2% (lowest possible concentration) Columns: (SDV), GRAM, GRAL Page 37

Additional Method Development Aqueous Solvents Page 38

Additional Method Development Aqueous Solvents, Neutral Polymers: - Polymer with functional groups: -CO-NH2, -OH, Polysacharides Solvent (buffer): 0,05M...0,1M NaNO 3, NaCl, Na 2 SO 4 + Additive highly unpolar parts: + 20% Acetonitrile or + 20% Methanol Columns: SUPREMA Page 39

Additional Method Development Aqueous Solvents, Polyanions: - Polymer with functional groups: -COO - Na +, -SO 4- Na + Solvent (buffer): 0,05M...0,1M NaNO 3, NaCl, Na 2 SO 4 or 0,05M...0,1M Na 2 HPO 4 (ph = 9) (buffer) + Additive highly unpolar parts: + 20% Acetonitrile or + 20% Methanol Columns: SUPREMA, MCX Page 40

Additional Method Development Aqueous Solvents, Polycations: - Polymer with functional groups: -NH4+, Cl- Solvent (buffer): 0.1 Vol% - 1.5 Vol% Formic acid to 0.1 Vol% - 1.5 Vol% Trifluoro acetic acid (TFA) + Additive (+ additional salt) Columns: NOVEMA, SUPREMA-MAX Page 41

Troubleshooting Column Performance Tests: Plate count (half height method): Measured using monodisperse low molar mass sample. Depends on: N Vp 2 V 5.54 w p 0.5 sample, detector, loading, tubing length, column dimensions, packing, solvent, flow rate 2 Theoretical plate counts/m 3µm 167 000 5µm 100 000 10µm 50 000 20µm 25 000 40µm 12 000 Determine plate count for your system and verify regularly. Test single columns if column bank fails. Page 42

Troubleshooting Column Performance Tests: Resolution: R s V2 V1 log( M1 M 2) 2 ( 1 2) 2D ( 1 2) D V slope calibration curve peak volume peak variance Value and system dependent, using different standards yield different results. Specific resolution (sample independent): R sp log R M s 1 M 2 Determine the resolution and verify regularly. Page 43

Troubleshooting Installing new columns: 1) Installation of a column from a different manufacturer: The distance from ferrule to the end of the column can be very different. Change all fittings and ferrules connected to the column, otherwise you may damage your column. 2) Be aware of column flow direction. Operating the flow rate in the reverse direction is only part of troubleshooting or operating after a long storage time. 3) Thread the column fittings finger tight into the system. Do not overtighten the fitting. Overtightening may damage the column and the column head. Page 44

Troubleshooting Storing columns: To store columns: 1) Remove all salt solutions with pure solvents and plug tight with the original end plugs. 2) Keep columns with volatile mobile phases in a refrigerator (4 C) after use to prevent solvent evaporation. Never let the column temperature fall below the freezing point of the storage solvent. This can destroy the stationary phase. 3) Occasionally solvent is lost during long-term storage. When the expected pressure does not build up, it is an indication of a partially evaporated column solvent. Page 45

Troubleshooting Recovering partially dry columns: To re-wet a partially dry column: 1) Install the column in the reverse direction. 2) Fill the column with solvent, using a flow rate of 0.1 ml/min, until no more bubbles appear at the column outlet. 3) Change the column to the correct flow direction and use 0.1 ml/min for 2h. 4) Increase the flow rate slowly to the operating flow rate with an increment of 0.2 ml/min every 5 minutes. Page 46

Troubleshooting Cleaning Columns: When columns lose efficiency or you suspect the presence of foreign material adsorbed on the column: 1) Remove the column from the detector. 2) Install the column in reverse direction for clean-up. 3) Flush the column at 0.1 ml/min flow rate with a solvent that dissolves the suspect impurities and is fully compatible with your system. Aqueous columns: Organic columns: use variation of ph, buffer concentration, or mixtures of organic and aqueous eluents. use variation of eluent polarity. Use mixtures of appropriate solvents (e.g. THF in Toluene resp. chloroform) or solvent additives (TFAc in THF). Page 47