Chapter 11 Conventional Gas Chromatography

Similar documents
Instrumentation. Components of a gas chromatograph

Gas Chromatography (GC)

2401 Gas (liquid) Chromatography

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

CHAPTER 6 GAS CHROMATOGRAPHY

Chapter 31 Gas Chromatography. Carrier Gas System

Chapter 27: Gas Chromatography

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

GC Instruments. GC Instruments - Sample Introduction

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Chromatography. Gas Chromatography

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

10/27/10. Chapter 27. Injector typically 50 C hotter than oven

Gas Chromatography. Introduction

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Introduction to Gas Chromatography

Gas chromatography. Advantages of GC. Disadvantages of GC

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

Chemistry Gas Chromatography: Separation of Volatile Organics

Understanding Gas Chromatography

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

Chapter 1. Chromatography. Abdul Muttaleb Jaber

Chemistry Instrumental Analysis Lecture 28. Chem 4631

Ch24. Gas Chromatography (GC)

GC Instruments. GC Instruments - Columns

Principles of Gas- Chromatography (GC)

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Biochemistry. Biochemical Techniques. 12 Gas Liquid Chromatography

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Gas Chromatography CHEM Dr. Reem M. Alghanmi st term

7 INSTRUMENTAL CHROMATOGRAPHY

Introduction to Capillary GC

GAS CHROMATOGRAPHY (GC)

Luminescence transitions. Fluorescence spectroscopy

Principles of Instrumental Analysis

Gas Chromatography. A schematic diagram of a gas chromatograph

Volatile organic compounds (VOCs):

Chapter content. Reference

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. CHAPTER 26

High Pressure/Performance Liquid Chromatography (HPLC)

Activity in the FID Detection Port: A Big Problem if Underestimated

Experiment 6 Simple and Fractional Distillation

Instrumental Chemical Analysis

Lab 3 Guide: Gas Chromatography (GC) (Sept 8-14)

Determination of Volatile Substances Proof of Food Adulteration

Gas Chromatography notes

Rapid Catalyst Screening Reactors

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid

Gas chromatography. Flow measurement. Flow control. Injection methods. Flow measurement


Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Trajan SGE GC Columns

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Gas Chromatography (GC)

Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid.

1. Carrier gas supply. - Flow rate measurement

/Chapter 27.ppt

Selection of a Capillary

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

CEE 772: Instrumental Methods in Environmental Analysis

High Performance Liquid Chromatography

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Selection of a Capillary GC Column

Introduction and Principles of Gas Chromatography

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

Experiment 3: Analysis of Unknown Hydrocarbons by GC, NMR, and Chemical Tests for Unsaturation and Aromaticity

G a s C h r o m a t o g r a p h y Determination of Ethanol in Wine by Head-Space Gas Chromatography

Method for the determination of 1,3-butadiene

Thermal Soil Desorber Instructions September 2006

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography

Fundamentals of GC: Introduction: Sample introduction:

Introduction to Chromatographic Separations

Analytical Chemistry

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method

Chromatography and Functional Group Analysis

Methods of pollution control and waste management - laboratory. Adsorptive removal of volatile organic compounds from gases streams

Introduction to Capillary GC

High Performance Liquid Chromatography

High Performance Liquid Chromatography

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

CEE 772: Instrumental Methods in Environmental Analysis

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

Experiment 1: Thin Layer Chromatography

Muliple Gas#3 plus Sulfur GC Configuration

CEE 772: Instrumental Methods in Environmental Analysis

DET REPORT. 1.) CATALYTIC COMBUSTION IONIZATION DETECTION (CCID) - MORE CHARACTERISTICS OF THE SELECTIVITY FOR n PARAFFINS. NO.

Transcription:

Chapter 11 Conventional Gas Chromatography

Gas Chromatography GC is the first instrumental chromatographic method developed commercially It is relatively easy to introduce a stable flow and pressure for the mobile phase, that is the carrier gas The equipment is simple. All that is really needed is cylinder of compressed air, pressure regulator and a valve

Introduction Principles Partition of molecules between mobile phase and stationary phase Separation technique Gas is the mobile phase and liquid (GLC) or solid (GSC) is the stationary phase GLC: liquid is coated on an inert solid; separation is the result of variable solubilities in the liquid phase GSC: Particulate solid like molecular sieve is the stationary phase; separation is the result of variable adsorption on the solid surface

Abundanc e Gas Chromatography E A D Gas Chromatograph C Sample: mixture of volatile liquids (~1 L) B Gas Chromatogram A B C E D 0 5 10 15 20 Time (minutes)

Uses of GC Separation and analysis of organic compounds Testing purity of compounds Determine relative amounts of components in mixtures Compound identification Isolation of pure compounds (micro scale work)

Advantages of the GC Speed: minutes or seconds Resolution: complex samples Sensitivity: 10-9 or even 10-12 g/s Versatility: Gases, liquids or solids (qualitative & quantitative) GLC is more common than GSC in terms of flexibility and resolution Salts and other ionic compounds + high molecular weight may not be determined by the GC. They can be determined by liquid chromatography or pyrolysis chromatography

GC Process 1. Column is selected, packed with liquid phase, and installed 2. Sample injected with microliter syringe into the injection port where it is vaporized and mixed into the carrier gas stream (helium, nitrogen, argon). 3. Sample becomes partitioned between the moving gas phase and the stationary liquid phase. 4. The time spent by different compounds of the sample in vapor phase is a function of their vapor pressure 5. The more volatile compounds arrive at the end of the column first and pass into the detector

Factors Affecting Separation Boiling Points of Components in Sample Low boiling compounds have higher vapor pressures. Boiling point increases with increasing molecular weight Flow Rate of Carrier Gas Choice of Liquid Phase (Solubility in the liquid stationary phase determines the retention time in the stationary phase) Molecular weights, functional groups, and polarities of component molecules are factors in selecting liquid phase. Length of Column Similar compounds require longer columns than dissimilar compounds. Isomeric mixtures often require quite long columns

Mechanism of Separation of Components of a Mixture by Chromatography Stationary Phase Start Mobile Phase Affinity of molecule For the stationary Phase determines Residence time In this phase. high affinity low affinity

GC Instrumentation Carrier gas Sample Injector Column Detector

Schematic Diagram of a Gas Chromatograph

RESET Gas Chromatography System Filters/Traps Data system H Regulators Syringe/Sampler Inlets Air Hydrogen Gas Carrier Column Detectors gas system Inlet (Injector) column detector Readout system

Gas Flow

Carrier Gas Carrier gases, must be chemically inert, Include helium, argon, nitrogen, carbon dioxide, and hydrogen. The choice of gases is often dictated by the detector used. Associated with the gas supply are pressure regulators, gauges, and flow meters. The carrier gas system often contains a molecular sieve to remove water or other impurities.

Purity of Carrier Gases Impurities (particularly O 2 & H 2 O) can chemically change the liquid phase and thus the t R and reduce the lifetime of the column. False peaks may appear Liquid stationary phases (polyesters, polysiloxanes, polyamides) degrade by O 2 & H 2 O Contaminant from column may desorb in H 2 O causing a high detector background (baseline drift and noise) Traces of hydrocarbons cause a high background in the FID Purity should be >99.995%

Carrier gas: mainly He and N 2 Two stage pressure regulator 1. It indicates the pressure left in the cylinder (min. 40 psi) 2. It indicates the increasing pressure delivered to the GC (min. = 20 psi)

Function Introduce sample Vaporize sample Split sample Injection systems Main Components It is a metal block containing: Heaters Temperature sensors Septum holder on the front Connection for the column on the rear

Packed column injection Injector is a simple portion of the GC system when packed columns are used There are two basic approaches for these injectors Injection ports Sampling loops/ valves

Various Types of Injection ports Split - only a small portion (10:1 and 100:1) of the injected sample goes on column Splitless - all material injected goes on column Usually split/splitless modes are provided to the same instrument. Cool on-column injection (Direct on column cold injection. It is used for sensitive materials) Programmed temperature - sensitive materials (more durable method than OC)

http://elchem.kaist.ac.kr/vt/chem-ed/sep/gc/graphics/packdinj.gif Rubber septum serves for about 30 injections in ordinary care 5-10 injections in case of large syringes

Injection Port/ Temperature Effects The purpose of the port is to flash vaporize the sample and introduce it into the column T injector should be > 50 o C above T column Septum of the injector must be: Stable at the T inj Replaced regularly to maintain seal

Syringe Injection Methods Major source of precision error results from poor injection technique Both automatic and manual injection methods are available If automatic equipment is present, use it. If not, several approaches can be tried to help reduce injection errors

Injection of liquid sample 10 µl syringe is the most popular device Load the desired volume of liquid, then draw the plunger back to pull the liquid out of the needle Insert the needle quickly through the septum as far as it will go Depress the plunger and immediately remove the needle from the injection port

Gas Sampling Loops (Sampling Valves) Introducing a constant amount of a gas can be difficult with a syringe Gas sampling loops and valves offer a high precision (+/- 1%) means of introducing gases Equipment is relatively inexpensive and only requires a constant temperature for easy use

Gas Sampling Loops/Valves Valves give better reproducibility Require less skill Can be easily automated

Injection valve

Injection Port Temperature The temp. should be 20-30 o C hotter than the boiling point of the least volatile component But low enough to prevent sample decomposition and septum bleed Temp. may be checked by raising it and watching : Position, or area, or shape of the peaks. Drastic changes mean the temp. setting is high It should be 10% above that of the column to ensure rapid volatilization of the sample. The efficiency of the column is almost constant under this condition Components may be vaporized at a temp. ~100 o C below its atmospheric boiling point

Very high boiling point or temp. sensitive material can be handled by dilution with volatile solvent that permits lowering the injection temp. This will lower the sensitivity! Try various temperatures until peak broadening becomes apparent With temp. programming techniques low injection temps become very practical. No rush to vaporize the high boiling components

Effect of injection port temperature on resolution a: Methanol; b: Ethanol; c: ispropanol Boling points: 65 82 o C

Gas Chromatographic Columns 1. Packed Columns for GLC 2. Packed Columns for GSC 3. Capillary Columns

Types of columns Packed columns 1/8 1/4 OD, stainless steel or glass tube 6-20 feet in length Capillary (open tubular) columns 0.1-0.5 mm ID 10-100 meters in length

Capillary (Open tubular columns)

Packed Columns 1. Column material 2. packing material (solid or inert support) 3. Stationary phase

Column Material Stainless Steel : most common. adsorbs some compounds particularly polar ones & especially water. Copper tubing. It is good for trace water analysis. Plastics: are limited due to permeability & temperature limit (used for reactive or highly corrosive chemicals H 2 S, HF Teflon, polypropylene and nylon tubing are available. Glass : If glass were not difficult to form into columns & relatively fragile it would be the very best choice for tubing ( used for pesticides & steroid).

Solid Support (Stationary Phase Support) Thermally stable Diatomaceous earth is the most common (nearly, all silica) solid support. Diatomaceous earth: Skeletons of microscopic unicellular algae (diatoms) consisting of microamorphous hydrous silica marketed under Chromosorb Polar analyte species such as, Alcohols or Aromatic hydrocarbons are adsorbed physically on the silicate surfaces of diatomaceous earth. Adsorption results in distorted peaks broadened with tail This catalytic activity may lead to sample decomposition

Reasons for adsorption activity Silicates + Water Silanol groups on the silicate surface OH OH OH OH Si O Si O Si O Si Si-OH groups have strong affinity for polar organic molecules

Treatment of Solid Supports Non-acid washed (NAW) an untreated form Acid washed (AW) use HCl Removes metals, impurities, Reduces surface activity and absorption Acid washed Dimethyldichlorosilane treated (AW-DMCS) Cl OH Si + (CH 3 ) 2 SiCl 2 HCl CH 3 O Si CH 3 Si CH 3 CH 3 O Si CH 3 Si HCl CH 3 OH

Stationary Liquid Phase

Characteristics of Stationary Liquid Phase The stationary phase should provide separation of the sample with a reasonable column life Suitable phase is chosen on the basis of : Experience or Experiment. It is desirable to have maximum information about the sample composition : bp.range, components expected & their structure Stationary phases should have similar chemical structure to the sample components

( is the separation factor)

Commonly Used Liquid Phases PHASE TEM. LIMITS Good for 1. SQUALANE 0/125oC Nonpolar 2. OV-1, SE-30 100/350 oc 3. DEXSIL-300 50/350 Oc (Most thermally stable) 4. OV-17; SP-2250 0-350 Moderately polar 5. QF-1; OV-210; 0-275 SP-2401 6. CARBOWAX-20M 60/225 Strongle polar 7. DEGS 20/200 8. OV-275 20/250

Squalane Saturated, highly branches, C-30 hydrocarbon Non-polar Limited temperature range: 0-125 oc Separate hydrocarbons Standard reference for Rohrschneider and McReynolds constants

Non-polar Temperature range: 100-350oC Most widely used liquid phase Separate all sample types OV-1 ( SE-30 )AND SP-2100

OV-17 (SP-2250) 50% methyl, 50% phenylpolysiloxane Semi-polar Temperature range : 0-350oC Widely used to separate drugs, steroids, carbohydrates

Carbowax 20-M HO ( - CH 2 CH 2 O-) n H Polymeric polyethylene glycol Polar Temperature range : 60-225oC Widely used for polar samples

Column Temperature 1. As the Column temp. increases a sample component spends more time in the mobile phase. * This will cause a decrease in the t R (Faster separation) * t R doubles for every 30 o C decrease. 2. Increasing the temp. decreases the band broadening since it leads to a decrease in the available time for diffusion in the column. 3. The lower the temperature the better is the separation * The column temp. should not be less than about 10 o C below the bp of the highest-boiling sample component (other wise distorted peaks may result). * Roughly, a temp. equal or slightly above the average b.p of a sample results in a reasonable elution time (20 to 30 min.). 4. Optimum column temp. depends upon: b.p. of the sample Degree of separation required.

Isothermal and temperature programming analysis The column is placed in the oven If the temperature is held constant during the entire analysis then it is called analysis under an isothermal mode. If the temperature is varied during the analysis, then the process is called temperature programming

Column Temperature Effect (Isothermal Analysis) Isothermal chromatographic analysis is one which is performed at a constant column temperature.

Conclusion Higher temp. enables rapid analysis but loss in resolution. Lower temp. achieves better resolution but longer analysis time

Narrow Boiling Range Samples Isothermal column temperature should be used. Select temperature 20-50 o C lower than boiling range of sample when thin films are handled. Use highest temperature that still allows adequate resolution and stability to shorten analysis time.

PERCENT LOADING OF STATIONARY PHASE How much liquid phase should be coated on the support? For analytical columns: 5 10% Higher loading ( up to ~ 30% ) can be applied for light gases (C 1 to C 4 hydrocarbons) Low Loading (1-5% ) with very high boiling compounds The liquid phase should cover the support almost completely. If the support is inert, this condition is not a must

LIQUID PHASE AND TEMPERATURE LIMIT To avoid decomposition, column bleeding and high retention Practical Temperature Limits depend upon the following: * Nature of liquid phase * Percent loading * Sensitivity of detector Temp. limits reported in most cases for TCD Limits; for FID may be about 100 o C less than that for the TCD.

Conditioning the column Why? Remove: impurities in liquid phase ; residual solvent from coating How? Install the column but do not connect to the detector Set the carrier flow 30 ml/min for 1/8 inch Heat for 1 hr at 100 o C Raise the temperature to slightly below the b.p. of liqid phase (about 30 o C higher than expected operating column operation temp.(. Continue heating over night Cap the column when it is removed from the instrument

Deatectors Features of Detectors A device that measures physical properties (preferred), not chemical properties The detector generates an electrical signal proportional to the sample concentration Detector and connections must be hot enough (20 to 30 o C above the column temp. or the boiling point of the highest boiling component) so that condensation of the sample or liquid phase does not occur

Peak broadening or disappearance is characteristic for condensation in the connections Ionization type detectors must be maintained at temp. high enough to avoid not only condensation of sample but also the water or by-products formed in thr ionization process

Most Common GC Detectors Most common detectors roughly in order from most common Thermal Conductivity Detector (TCD or hot wire detector), Flame Ionization Detector (FID), Electron Capture Detector (ECD), Photo Ionization Detector (PID), Flame Photometric Detector (FPD), Thermionic Detector VERY expensive choices: Atomic Emission Detector (AED) Ozone- or Fluorine-Induced Chemiluminescence Detectors. All of these (except the AED) produce an electrical signal that varies with the amount of analyte exiting the chromatographic column. Fourier Transform Infrared Detector (FTIR) Mass Spectrometer (MS) Other: UV, FT-NMR

Schematic of a thermal conductivity detector, TCD

Two pairs of TCDs are used in gas chromatographs. One pair is placed in the column effluent to detect the separated components as they leave the column. Another pair is placed before the injector or in a separate reference column. The resistances of the two sets of pairs are then arranged in a bridge circuit. The heated element may be a fine platinum, gold, or tungsten wire or, alternatively, a semi conducting thermistor. The resistance of the wire or thermistor gives a measure of the thermal conductivity of the gas. Elution heat loss increased resistance needed to balance bridge = recorded

Flame Ionization Detector Basic Principle The effluent from the column is mixed with hydrogen and air, and ignited. Organic compounds burning in the flame produce ions and electrons which can conduct electricity through the flame. A large electrical potential is applied at the burner tip, and a collector electrode is located above the flame. The current resulting from the pyrolysis of any organic compound is measured which is proportional to the carbon content of the molecule entering.

Ring electrode: stainless Steel gauze (+ve electrode) Flame jet serves As Ve Electrode (FID) sample burned in H 2 /air flame sample must be combustible must use electrometer ppm sensitivity destructive

Qualitative analysis by Chromatographic methods Qualitative analysis is based on retention data Retention time t R is characteristic of a substance, compared to a standard. Reproducibility of retention depends upon several experimental conditions: column length and diameter, stationary and mobile phases, column packing, column temperature, mobile phase flow rate and others

Retention time, t R t R : It is the time elapsed from the point of injection to the peak maximum Adjusted t R : It is the time from the maximum of unretained peak (the peak of the mobile phase or the air peak) to the peak maximum of a certain component t M (hold up time): is the time required for the mobile phase to be eluted completely from the column

Same column under Same conditions Has been used Unknown alcohol components Standard sample

Component 1 is used as the reference; it should be present or added to the sample and compatible with the sample Peak of component 1 must be close (but resolved) to the sample peak

When component 3 is suspected, add more of this component to the sample and watch any change in its peak

Basis for Quantitative Analysis The peaks in the chromatogram are the basis for quantitative analysis Peaks of interest should fulfill the following requirements: must be undistorted must be well separated Must have a large S/N ratio must have a flat baseline Peak shape: The ideal chromatographic peak is symmetric and narrow Peak integration The peak height or, better, the area must be determined and this is done by the computer Calculation External standard method Internal standard method Internal normalization

External standard method This method is common to most quantitative analysis techniques. It allows the measurement of the concentration of one or more components that elute in a chromatogram containing, perhaps, many peaks. This method, employing the absolute response factor, K, is used in the following way (Single point calibration method)

Multilevel calibration In a multilevel calibration several different amounts of the standard are prepared and analyzed. A regression method is used (e.g. linear least-square) and this leads to a more precise value for C unk. This quantitative method is the only one adapted to gaseous samples. This simple method is used in industry for repetitive analyses. For such analyses, chromatograph must be equipped with an autosampler,

Multilevel External Calibration of Fatty Acids Detector Response C 18 Peak Area (cm 2 ) 10 C 16 8 6 C 14 4 2 0.5 1.0 1.5 2.0 2.5 3.0 Retention Time The content % of C fatty acids = 14 C C + C + C Sample Concentration (mg/ml) = the content % of C fatty acids 14