UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Similar documents
Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Methanol Extraction of high level soil samples by USEPA Method 8260C

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Analytical Trap Comparison for USEPA Method 8260C

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Volatile Organic Compounds in Every Day Food

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Maximizing Sample Throughput In Purge And Trap Analysis

Solid Phase Micro Extraction of Flavor Compounds in Beer

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Ed George and Anaïs Viven Varian, Inc.

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Optimizing. Abstract: is standardd. procedures. altered to

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Determination of Volatile Organic Compounds in Air

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

STANDARD OPERATING PROCEDURES

2017 EPA Method Update Rule and EPA Method 624.1

Solid Phase Micro Extraction of Flavor Compounds in Beer

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Optimizing Standard Preparation

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Maximizing Production While Minimizing Costs

Study of Residual Solvents in Various Matrices by Static Headspace

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

U.S. EPA Method 8270 for multicomponent analyte determination

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Author. Abstract. Introduction

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Reactions of Vinyl Chloride and Methanol in a Quadrupole Ion Trap Mass Spectrometer during VOC Analysis

Application Note 032

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

STANDARD OPERATING PROCEDURES

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Thermal Desorption Technical Support

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

If you have any questions concerning this report, please feel free to contact me. REPORT OF LABORATORY ANALYSIS

EPA TO-17 Volatile Organic Compounds

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

STANDARD OPERATING PROCEDURES

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Fast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Transcription:

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring Application Note Abstract The United States Environmental Agency (USEPA) has established the Unregulated Contaminant Monitoring Program to collect data for contaminants suspected to be present in drinking water. 1 The USEPA is currently proposing Unregulated Contaminant Monitoring Rule 3 (UCMR 3) which will collect data from January 2013 through December 2015. Because of the impact on public health and safety, it is vital to have a system that can detect and analyze for these compounds at low levels. This study will validate the analysis of the Volatile Organic Compounds (VOCs) present on the UCMR 3 analyte list. USEPA Method 524.3, used to monitor the VOCs, allows for the use of Single Ion Monitoring (SIM) to reach the part-pertrillion (ppt) levels required by UCMR 3. 2 This study uses a Teledyne Tekmar Stratum Purge and Trap Concentrator (PTC) and AQUATek 100 Autosampler coupled to a Thermo ISQ Single Quadrupole GC-MS. A calibration curve and detection limits will be determined for the VOCs on the UCMR 3 analyte list. Introduction The USEPA Contaminant Monitoring Program reviews the list of analytes every five years based on the Contaminant Candidate List; a list of compounds not regulated by the national primary drinking water regulations. These contaminants are known or anticipated to occur in public water systems. The compounds being monitored in UCMR 3 are 28 List 1 chemicals for Assessment Monitoring and two List 3 viruses for Pre-screen Testing. Assessment monitoring is required for those compounds with existing analytical methods, while the pre-screening methods are new analytical techniques for monitoring the viruses. The thirty contaminants that will be monitored by UCMR 3 are divided into several groups and can be found in Table 1. This study focuses on the analysis of the nine VOCs, which can be analyzed using USEPA Method 524.3, Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry according to UCMR 3. USEPA Method 524.3 allows for the use of SIM scan for low-level analysis. This ensures that contaminants present in water sources can be analyzed at the part-per-trillion level and monitored accordingly. This study will use a Stratum PTC and AQUATek 100 Autosampler with a Thermo ISQ GC- MS to establish a calibration and detection limit for the nine VOCs listed as contaminants in UCMR 3.

Unregulated Contaminant Monitoring Rule 3 Analyte List 1,2,3-Trichloropropane 17-β-estradiol 1,3-Butadiene 17-α-ethynylestadiol Chloromethane 16-α-hydroxyestradiol 1,1-Dichloroethane Hormones Equilin Volatile Organic Compounds n-propylbenzene Estrone Bromomethane Testosterone sec-butylbenzene 4-androstene-3,17-dione Chlorodifluoromethane Perfluorooctane sulfonate Bromochloromethane Perfluorooctanoic acid Metals Vanadium Molybdenum Cobalt Perfluorinated Compounds Perfluorononanoic acid Perfluorohaxane sulfonic acid Perfluoroheptanoic acid Strontium Perfluorobutane sulfonic acid Synthetic Organic Compounds 1,4-Dioxane Enteroviruses Viruses Oxyhalide Anion Chlorate Noroviruses Table 1: UCMR 3 Proposed Contaminants List (VOC Compounds in Yellow) Experimental-Instrument Conditions The Stratum PTC and AQUATek 100 autosampler were coupled to the Thermo ISQ GC/MS system for analysis. The Thermo ISQ was configured with a Restek Rtx-624 20m x 0.18mm x 1.0µm column. The GC-MS parameters are outlined in Tables 2 and 3. The SIM ions used for quantification and conformation for the analytes in this study can be found in Table 4. A Tekmar proprietary #9 trap was the analytical trap of choice. Table 5 outlines the PTC and autosampler conditions. GC Parameters MS Parameters GC: Thermo Trace GC Ultra MSD: Thermo ISQ Column: Restek Rtx -624 20m x 0.18mm x 1.0µm Source: 275 C Oven Program: 45 C for 4.5 min, to 100 C at 15 C/min, for 0 min, to 240 C at 25 C/min for 0.33 min Transfer Line Temp: 230 C Inlet: 200 C Solvent Delay: 0.5 min Column Flow: 1.0 ml/min SIM ions: See Table 4 Gas: Helium Ionization Mode: EI+ Pressure: 21.76 psi Emission Current 25 ua Split Ratio: 25:1 Tables 2 & 3: Thermo ISQ GC/MS Parameters

Compound Retention Time Q-Ion Confirmation Ion Chlorodifluoromethane 1.14 min 51 67 Chloromethane 1.26 min 50 52 1,3-Butadiene 1.33 min 54 53 Bromomethane 1.53 m in 94 96 1,1-Dichloroethane 3.48 min 63 65 Bromochloromethane 4.45 min 128 130 n-propylbenzene 9.80 min 91 120 1,2,3-Trichloropropane 9.94 min 110 112,75 sec-butylbenzene 10.30 min 105 134 Table 4: Retention Times, Quantification Ions, and Confirmation Ions for SIM scan Stratum PTC and AQUATek 100 Parameters Variable Value Variable Value Pressurize Time 0.25 min Purge Time 6.00 Sample Transfer Time 0.50 min Purge Temp 20 C Rinse Loop Time 0.50 min Purge Flow 60 ml/min Sweep Needle Time 0.30 min Dry Purge Time 2.00 min Bake Rinse On Dry Purge Temp 30 C Bake Rinse Cycles 1 Dry Purge Flow 100 ml/min Bake Rinse Drain Time 0.50 min GC Start Start of Desorb Presweep Time 0.35 min Desorb Preheat Temp 245 C Water Temp 90 C Desorb Drain On Valve Oven Temp 150 C Desorb Time 1.00 min Transfer Line Temp 150 C Desorb Temp 250 C Sample Mount Temp 90 C Desorb Flow 300 ml/min Purge ready Temp 35 C Bake Time 2.00 min Condenser Ready Temp 40 C Bake Temp 280 C Condenser Purge Temp 20 C Bake Flow 400 ml/min Standby Flow 10 ml/min Condenser Bake Temp 200 C Pre-Purge Time Pre-Purge Flow Sample Heater Sample Preheat Time 0.5 min 40.0 ml/min Off 1.00 min Sample Temp 40 C Table 5: Stratum PTC and AQUATek 100 Parameters (Stratum PTC Parameters are in Blue)

Calibration Data A 1ppm stock solution containing the analytes of interest was prepared in methanol. From this stock solution, an eight-point calibration curve from 2.5-1000ppt was prepared in de-ionized water. Thermo Scientific EnviroLab Forms software was used to process the calibration data. Figure 1 illustrates calibration curve results for sec-butylbenzene, one of the contaminants to be monitored according to UCMR 3. Curve data was evaluated using quadratic regression, with equal weighting. The results of these calibrations can be found in Table 6. Internal standards and surrogates were added to each sample in 10uL aliquots to the sample providing a constant 50ppt final concentration. The internal standards and surrogates were evaluated for reproducibility; the results can be found in Table 7. Figure 1: Calibration Curve for sec-butylbenzene.

Method Detection Limits Contaminant Quadratic Regression Factor (r 2 ) Chlorodifluoromethane 0.9998 Chloromethane 0.9996 1,3-Butadiene 1.0000 Bromomethane 0.9998 1,1-Dichloromethane 0.9998 Bromochloromethane 0.9996 n-propylbenzene 0.9997 1,2,3-Trichloropropane 0.9994 sec-butylbenzene 1.0000 Table 6: Calibration Curve data for UCMR 3 VOC Contaminants Internal Standard/Surrogate %RSD 1,4-Diflurorobenzene 4.52 Chlorobenzene-d5 4.27 4-Bromofluorobenzene 3.02 1,2-Dichlorobenzene-d4 2.77 1,4-Dichlorobenzene-d4 5.57 Table 7: Internal Standard/Surrogate Standard Data In order to complete this ppt analysis, SIM scanning was used. As recommended by the USEPA, the ions utilized for the analysis of each contaminant can be found in Table 3. Figures 2 and 3 are images from Thermo Scientific EnviroLab Forms software illustrating the quantification and confirming ions for 1,2,3- trichloropropane (TCP), individually and overlaid to show there are no interferences. Figure 2: Quantification Ion (110m/z) and Confirming Ions (112m/z and 75 m/z) for TCP

Figure 3: Ion Overlay of the Quantification Ion (110m/z) and Confirming Ions (112, 75 m/z) for TCP Method detection limits (MDLs) were established by analyzing seven replicate samples at a concentration of 10 ppt for each contaminant. The detection limit is a statistical determination which identifies the minimum concentration of an analyte that can be identified, measured, and reported with 99% confidence that the concentration is greater than zero. 2 The detection limit is determined with the following equation, where S is the standard deviation and t is the t-value for the 99% confidence level. For seven replicates, the t value is 3.143. 2 DL = S t (n 1,1 α=0.99) Table 8 illustrates the replicate data and MDL results for the VOC contaminants in UCMR 3. Repetition Analyte 1 2 3 4 5 6 7 Average Stdev %RSD MDL Chlorodifluoromethane 7.86 7.69 7.31 7.22 8.18 7.20 8.01 7.64 0.40 5.23 1.26 Chloromethane 9.68 10.31 9.96 10.41 10.36 9.82 10.28 10.12 0.29 2.89 0.92 1,3-Butadiene 7.22 7.00 8.27 8.25 6.81 7.27 6.49 7.33 0.69 9.37 2.16 Bromomethane 8.52 8.45 8.37 9.01 8.89 9.29 8.50 8.72 0.35 3.98 1.09 1,1-Dichloromethane 9.99 10.22 10.51 10.07 9.68 10.20 9.49 10.02 0.34 3.44 1.08 Bromochloromethane 9.37 9.34 9.85 9.72 9.72 10.86 9.32 9.74 0.54 5.53 1.69 n-propylbenzene 8.31 8.61 8.30 8.86 8.57 8.32 8.01 8.43 0.28 3.27 0.87 1,2,3-Trichloropropane 10.33 10.36 9.90 10.42 10.13 9.74 10.44 10.19 0.28 2.70 0.86 sec-butylbenzene 8.94 9.04 9.27 9.05 8.92 9.18 8.70 9.01 0.19 2.07 0.59 Table 8: MDL data for VOC Contaminants Listed in UCMR 3

Conclusions The USEPA has established the Unregulated Contaminant Monitoring Program to collect data for contaminants suspected to be present in drinking water. The current proposal, UCMR 3, will collect data on 30 contaminants from January 2013 through December 2015. This study validates the capability of Teledyne Tekmar Stratum PTC and AQUATek 100 Autosampler coupled with a Thermo ISQ for low-level detection of the nine VOC contaminants listed in UCMR 3. Using a SIM scan, as permitted in USEPA method 524.3, MDLs were established for all VOC contaminants to be monitored according to UCMR 3. With concerns for public health and safety, notification levels of harmful compounds are being pushed lower and lower, driving the demand for more sensitive instrumentation. All contaminants in this study had MDL levels under 2.5 ppt, proving purge and trap to be a superb analytical tool for this analysis. References 1. Unregulated Contaminant Monitoring Program http://water.epa.gov/lawsregs/rulesregs/sdwa/ucmr/ 2. USEPA Method 524.3, Measurement of Purgeable Organic Compounds by Capillary Column Gas Chromatography/Mass Spectrometry (GC/MS), Revision 1, 2009.