MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application

Similar documents
METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

enable measurement. This method separates these isotopes effectively.

URANIUM IN SOIL. Analytical Procedure (2 GRAM SAMPLE) 1. SCOPE

TECHNETIUM-99 IN WATER

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Contact Person(s) : Anna Berne APPLICATION

TECHNETIUM-99 IN SOIL

TECHNETIUM-99 IN WATER

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM)

Determination of trace anions in concentrated hydrofluoric acid

THORIUM, PLUTONIUM, AND URANIUM IN WATER

DISCLAIMER: This method:

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Automation of the radiochemical procedures for the sequential separation of radionuclides

Samples compliant with the WATER

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Test Method: CPSC-CH-E

ELECTRODEPOSITION OF ACTINIDES

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

EXPERIMENT: LIMITING REAGENT. NOTE: Students should have moles of reactants in DATASHEET converted into masses in grams prior to the lab period.

Eichrom Technologies, Inc. Analytical Procedures Rev. 1.5 February 10, 2005 Page 1 of 9

1.1. This is a method for the separation and measurement of 228 Ra in water via its beta emitting 228 Ac daughter.

Uranium from water sample

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Rapid Separations. Activity Radioactive Solutions. Lawrence Jassin Eichrom Technologies LLC March 3, 2008 Pittcon 2008

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

Method NumberUSTUR 150

SAFETY NOTE: Before beginning this procedure, read all of the Material Safety Data Sheets for the chemicals listed in Section 5 of this procedure.

RARE EARTH FLUORIDE MICROPRECIPITATION

BATTERY INDUSTRY STANDARD ANALYTICAL METHOD

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Analysis of High Fired Plutonium Oxide and Other Actinides in MAPEP Soil Samples

NICKEL-63/59 IN WATER

Sample Analysis Design PART III

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Automating Sample Purification MC-ICPMS. Elemental Scientific ICP ICPMS AA

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Pu and Np-237 in seawater samples Version /03/14. Summary

&OAE- 9b I 0 IS& - - I Determination of Plutonium in Urine: Evaluation of Electrothermal Vaporization Inductively Coupled Plasma Mass Spectroscopy.

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Revised material in Section 4, Analytical Chemistry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Fergus Keenan Shona McSheehy Julian Wills

Method Number USTUR 510

Determination of urea in ultrapure water by IC-MS/MS

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

On the use of ICP-MS for measuring plutonium in urine

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

Rapid Analytical Methods for Determination of Actinides

ANALYTICAL TASK EXPERIMENTAL PROCEDURE

Rapid separation of uranium and plutonium by extraction chromatography for determination by thermal ionisation mass spectrometry

columns IonPac AS17-C Anion Exchange Column

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

QAM-I-116 Preparation of Labware

Rapid Methods for the Determination of Sr-90 in Steel and Concrete Samples

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Supporting Information

The problems as we saw them were;

Development of a Sampling and Analysis Method for the Measurement of Hexavalent Chromium in Ambient Air

Sample Analysis Design PART II

ION CHROMATOGRAPHY SYSTEM S 150

ALLOWAY METHOD OUTLINE

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+

DR/4000 PROCEDURE. IRON, Total

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

CHEMISTRY DEPARTMENT

THE URANIUM DETERMINATION IN COMMERCIAL IODINATED SALT

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction

California Environmental Protection Agency Air Resources Board

Determination of Selenium-79 in Aqueous Samples

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

Transcription:

Analytical/Inorganic MM800 (a) Ion Exchange and ICP/MS of Uranium in Water 1.0 Scope and Application This procedure can be used to determine U concentration or isotopic-ratio composition in groundwater and drinking water, using ion exchange (IE) and inductively coupled plasma-mass spectrometry (ICP-MS) techniques, at concentrations > 10 pg/ml. The information provided by this procedure will allow for evaluation and determination of 235 U enrichment or depletion. 2.0 Summary of Method The IE and ICP-MS determination of U concentration and isotopic abundance was obtained by adding a 233 U spike solution with known isotopic and elemental concentration to the sample before sample preparation. A 10-mL, 1-bed volume (BV) column of Dowex-1X8 anion exchange resin in the chloride form and 2 M HCl as the mobile phase provides for a U chlorocomplex having a distribution coefficient (Kd) of 10 2 (Kraus and Nelson 1956). This allows for 100 ml of sample solution to be processed. Standard operating procedures for ICP-MS are outlined in section 8.3. 3.0 Interferences Isobaric interferences for U isotopes may be observed for U.S. Department of Energy (DOE) wastes due to the presence of Pu, Pa, and Np. However, these elements may be considered as analogs to the lanthanide elements, which do not produce anion chloro-complexes in HCl solutions. This should be confirmed using a Pu, Pa, and Np spike. Other interferences, due to high mass loading, may also be observed in the presence of elements such as Zn and Fe. However these elements do not produce isobaric interferences for the determination of U using ICP-MS. Table 1 lists possible isobaric interferences for uranium isotopic determination. For more details on interpreting interferences, refer to U.S. Environmental Protection Agency (EPA) Method 6020 CLP-M v. 8.0. (a) This method was supplied by O. T. Farmer, G. M. Mong, A. Sharma, and C. J. Cemetson (Pacific Northwest Laboratory, Richland, Washington). MM800-1

DOE Methods Table 1. Isobaric Interferences for Uranium Plutonium Protactinium Neptunium Element Nuclides: Element Nuclides: Element Nuclides: Nuclide Weight Half-Life Nuclide Weight Half-Life Nuclide Weight Half-Life 236 Pu 236.046 2.87 y 234 Pa 234 6.69 h 235 Np 235 1.085 y 237 Pu 237.048 45.2 d 234m Pa 234 1.17 m 236 Np 236 1.6E05 y 238 Pu 238.049 87.7 y 236m Np 236.05 22.5 h 238 Np 238 2.117 d 4.0 Safety Ordinary laboratory safety procedures are assumed. Waste from DOE sites, either as leachates or as specific chemical separation process material, may contain significant amounts of fission products or actinides. Shielding and radiochemical safe-handling practices are assumed for those instances where radioactivity is suspected. 5.0 Apparatus and Materials 5.1 Sample Preparation Nalgene beaker (250 ml) Nalgene cover glass adjustable hot plate Plastic pipettes (0.1, 1.0, and 10.0 ml) Four place balance (0.001 g) Milli-pore vacuum Filtration system (0.45-µm) 5.2 Ion Exchange Column with zero head space (10 ml) Dowex 1 X 8 anion resin Advanced Gradient Pump (Dionex Corp.) DQP Pump (Dionex Corp.) Eluent Delivery Module (Dionex Corp.) AS40 Autosampler (Dionex Corp.) Foxy 200 Fraction Collector (Dionex Corp.) MM800-2

Analytical/Inorganic 5.3 Inductively Coupled Plasma-Mass Spectrometry 6.0 Reagents ICP-MS Fission PQ2 U5000 Ultrasonic Nebulizer (USN) American Society for Testing and Materials (ASTM) Type II water (ASTM D1193) Concentrated HCl (Seastar) Thirty percent H 2 O 2 (Reagent Grade) Concentrated HNO 3 (Seastar) Natural isotopic U standard {National Institute for Standards and Technology (NIST)} Uranium-233 standard 7.0 Sample Collection, Preservation, and Handling This procedure implements processes specified in Method 6020 CLP-M Version 8.0 whenever applicable. Special care should be taken to avoid contamination while collecting samples. All samples should be placed in a sealed container. Proper shielding should be provided to handle the sample safely. 8.0 Procedure 8.1 Relative Uranium Concentration Procedure 8.1.1 The sample is shaken, and 10 ml of the sample are transferred to a 20-mL plastic sample bottle. Then 0.1 ml of (1:1) HNO 3 :H 2 O are added to the sample. 8.1.2 The ICP-MS is optimized using a 10-ng/mL solution of natural U for a maximum 238 U count rate. Each sample solution is compared to this standard for relative concentration. This procedure will determine the amount of 233 U spike that will be added in the actual sample preparation. The ratio of 233 U/ 238 U in the final sample solution is between 0.5 to 1.5. The amount of 233 U spike is determined by the cognizant scientist and should be near the value expected for the sample U concentration or at a minimum at a detectable amount (5 to 10 times the instrument detection limit). If the 238 U is above 0.1 µg/ml in this solution, the sample should be further diluted. 8.2 Acid Digestion Procedure for IC and ICP-MS Analysis 8.2.1 One hundred milliliters of a well-mixed sample are transferred to a 250-mL Nalgene beaker. MM800-3

DOE Methods 8.2.2 One milliliter of (1:1) HNO 3 :H 2 O 2 and 2 ml of 30% H 2 O 2 are added to the sample to oxidize U(IV) to U(VI) and break down any organic material. 8.2.3 An appropriate amount of a spike solution of 233 U (as defined by the above procedure) is added to the samples being prepared. The samples are labeled with the sample identification. 8.2.4 The sample is covered with a watch glass or similar cover and heated on a steam bath or hotplate for 2 h at 95 C or until the sample volume is reduced to between 25 and 35 ml. 8.2.5 The sample is cooled and filtered with a 0.45-µm filter to remove insoluble material. 8.2.6 Seventeen milliliters of 12 M HCl are added to the digestion vessel, and the sample volume is adjusted to 100 ml, which will provide an ~2 M HCl sample solution. The sample is now ready for ion chromatography (IC) separation. 8.3 Ion Exchange 8.3.1 Column Characteristics The column characteristics are established before any column separation of U. This is initiated by conditioning the resin to the chloride form using 2 to 4 BV of 2 M HCl. A solution of 100 ng/ml of natural U in 2 M HCl is processed through the column and directly into the ICP-MS to determine the number of BV required for U breakthrough (BT). The determination of U BT will establish the volume of sample in 2 M HCl that can be processed with this column. However, this procedure requires that 15 BV be obtained before U BT. The volume of H 2 O necessary to elute a 100-mL (10 BV) U solution off the column should be determined. Initially, the column is conditioned with 2 to 4 BV of 2 M HCl. This is followed by loading 100 ml (10 BV) of a 2 M HCl U solution on the column. In the next step, 2 BV of 2 M HCl are processed; these fractions are diverted to waste. The U is then eluted using H 2 O directly into the ICP-MS, and the BV of H 2 O required to remove U is determined. These characteristics are used to separate and collect U in groundwater and drinkingwater samples. (Note: If the 233 U spike solutions are contaminated with Pu, Np, or Pa or the 2 M HCl eluent is contaminated with U, then an IE separation or clean-up can be performed using this procedure. If this is done, then an analysis of the solution is required to determine the U concentration and isotopic abundance of this solution using ICP-MS. The need for any spike cleanup will MM800-4

Analytical/Inorganic 8.4 Sample Processing be displayed in the blank solution if any alteration in the U isotopic ratios after IE separation occurs.) (Kunin 1956; Marcus 1968). The system configuration for separating and eluting U is defined in Figures 1 through 4. In Figure 1, the anion column is equilibrated with 20 ml of 2 M HCl to convert the resin to the Cl - form, with this solution being diverted to waste. In Figure 2, the gradient pump is turned off, and valve 5 is switched to the off position; the sample pump is put inline to the column, and 100 ml of sample are loaded. The output from the column is also diverted to waste. This solution will contain lanthanide elements, actinides, alkali, alkaline earth elements, and most transition elements (Kraus and Nelson 1956; Hudgen 1956). In Figure 3, the gradient pump is turned on, valve 5 is switched to the on position, and the column is washed with 20 ml of 2 M HCl. This is to remove all elements that have little tendency to form chloro-complexes in 2 M HCl. This solution is also diverted to waste. In Figure 4, the eluent is changed to H 2 O, and U is eluted with 20 ml of H 2 O. This solution is collected by the fraction collector and is ready for analysis by ICP-MS. 8.5 ICP-MS 8.5.1 Instrument Detection Limit The instrument detection limit (IDL) is three times the average of the standard deviations obtained from the analysis of a standard solution (each analyte in reagent water) at a concentration of 3 to 5 times the expected IDL, with seven consecutive measurements. For general purposes, once an IDL has been established, it is valid until either 1) a significant change in instrument response occurs, or 2) there is a major change or repair to the instrument (e.g., a new sample interface or quadrupole); certain projects may have more stringent requirements regarding the frequency and/or protocol of determining the IDL (Marcus 1968). MM800-5

Gradient Pump E1 = H O 2 E1 = 2 M HCl Sample Pump Column Waste 1 Waste 2 Fraction Collector Gradient Fraction Sample Pump Time Flow % E1 % E2 Collector Flow 0.0 5.0 0.0 100 On Off Waste 2 0.0 4.0 5.0 0.0 100 On Off Waste 2 0.0 Figure 1. Column Conditioning with 2 M HCl

Gradient Pump E1 = H O 2 E1 = 2 M HCl Sample Pump Column Waste 1 Waste 2 Fraction Collector Gradient Fraction Sample Pump Time Flow % E1 % E2 Collector Flow 4.0 0.0 0.0 0.0 Off Off Off 5.0 24.0 0.0 0.0 0.0 Off Off Off 5.0 Figure 2. 100-mL Sample Loading on Column

Figure 3. Wash Column with 2 M HCl Gradient Pump E1 = H O 2 E1 = 2 M HCl Sample Pump Column Waste 1 Waste 2 Fraction Collector Gradient Fraction Sample Pump Time Flow % E1 % E2 Collector Flow 24.0 5.0 0.0 100 On Off Waste 2 0.0 28.0 5.0 0.0 100 On Off Waste 2 0.0

Gradient Pump E1 = H O 2 E1 = 2 M HCl Sample Pump Column Waste 1 Waste 2 Fraction Collector Gradient Fraction Sample Pump Time Flow % E1 % E2 Collector Flow 28.0 5.0 100 0.0 On Off On 0.0 32.0 5.0 100 0.0 On Off On 0.0 Figure 4. Elution of Uranium with H 2 O

DOE Methods 9.0 Instrument Operation and Analysis The following procedures are to be used in conjunction with the VG Plasmaquad operating and user s manuals. The procedures below assume that the instrument is operating and functioning normally. 9.1 Normal Operating Conditions Operating conditions typically change on a daily basis. The following parameters are guidelines for proper instrument performance: Plasma Gas Auxiliary Gas Nebulizer Gas Forward Power Reflected Power Sampling Height (mm above loadcoil) ~13 L/min (Lpm) ~0.8 Lpm ~0.9 Lpm 1250 W < 5 W 12 mm 9.2 Tuning the Instrument At least 30 min should be allowed for the instrument to equilibrate before analyzing any samples. The ICP-MS is then tuned by aspirating the tune solution and adjusting the lens voltages for optimal sensitivity. The element in the tune solution having the median mass (e.g., In at 115 amu) is typically tuned first, with the low and high masses being tuned subsequently only if their resulting sensitivities are determined to be insufficient. 9.2.1 Instrument Response and Precision Check The instrument response is determined using the tune solution and shall be at least 1 x 10 6 counts per sec per ppm at 115.0 amu by pneumatic nebulization (assuming the tune solution contains In) to be considered acceptable for analysis. Tuning an instrument using the ultrasonic nebulizer (USN) (Cetac Model U- 5000) enhances instrument response by 10 to 50 times. Therefore a tune solution with element concentrations approximately 1 to 10% that of the pneumaticnebulizer tune solution is typically used. Both the instrument and the nebulizer must be optimized. The level and tuning settings on the USN are adjusted to maximize output (indicated on the output meter); output may be slightly compromised, however, to optimize the aerosol density and/or instrument response. The ICP-MS is tuned in the same manner with the USN as with the pneumatic nebulizer. Instrument response using the USN must be at least 5 x 10 7 MM800-10

Analytical/Inorganic cps/ppm at middle mass to be acceptable for analysis. The USN chiller must be running for at least 30 min before data acquisition using the USN and preferably before USN operation. 10.0 Calculations A mass bias correction (MBC) for any isotope of U, relative to 233 U, is determined using a solution having a known U molar concentration. The following equation (1) represents the calculation for the MBC for 235 U: M B C = CPS CPS Moles Moles 235 233 235 233 U U U U (1) where CPS stands for counts per second for the desired U isotope, represented by XXX U, obtained from the instrument. These values are corrected for the background signal before insertion in the above equation (1). 10.1 Isotopic Mass The following equation (2) represents U isotopic mass (IM) (in g) calculation in a given sample. It can be used for any available isotope of U that is represented by XXX U: IM = XXX 233 ( CPS U ) ( Moles U) 233 CPS U MBC - Z At. Wt. (2) where XXX U = desired U isotope At. Wt. = atomic weight of U isotope under consideration Z = moles of XXX U/ mole of 233 U (moles of XXX U associated with spike) MM800-11

DOE Methods 10.2 Isotopic Concentration The isotopic concentration can be determined by introducing a known amount of 233 U as the spike. The isotopic concentration of any U isotope can be determined relative to 233 U by the following equation (3): IC = ( ) IM g Sample Volume ml ( ) (3) 10.3 Total Uranium Isotopic Mass The total U IM is the sum of all the observed U isotopes. Since 234 U and 236 U are relatively low in natural abundance, these isotopes can be neglected in total U IM determinations. The following equation (4) is used: n 235 238 Total U = IM = U( g ) + U( g) (4) where n is the number of U isotopes. 10.4 Isotopic Abundance Once the total U is determined, the isotopic abundance (IA) of any U isotope can be obtained by taking their concentration respective to the total U. This is shown in the following equation (5): 11.0 Quality Control IA = IM n IM (5) Once IA is known, depletion or enrichment can be determined by comparing the ratio with 0.715, which is the naturally occurring ratio. If the ratio is less then 0.715, U is depleted, and if it is greater then 0.715, it is enriched. 11.1 Sample Preparation Check Sample 11.1.1 A matrix sample solution is used having a known concentration and IA of U spiked with 233 U and processed through all steps of the sample-preparation procedure. This solution validates the sample preparation. MM800-12

Analytical/Inorganic 11.2 Instrument Check Sample 11.2.1 To evaluate the instrument response, a known natural U standard is spiked with a known amount of 233 U. This is followed by performing a preparation check. The data obtained by instrument check (Table 2) and prep-check (Table 3) are listed as follows: Table 2. Instrument Check Run No. 235 U (ppm) 238 U (ppm) 235 U (Abundance) Total (ppm) 1 0.00733 1.014 0.00718 1.021 2 0.00716 0.999 0.00712 1.006 3 0.00725 0.994 0.00714 1.001 4 0.00720 0.995 0.00718 1.003 5 0.00717 1.000 0.00712 1.008 6 0.00738 1.025 0.00714 1.033 7 0.00718 0.994 0.00717 1.001 Table 3. Preparation Check Run No. 235 U (ppm) 238 U (ppm) 235 U (Abundance) Total (ppm) 1 0.0193 2.682 0.00715 2.701 2 0.0188 2.612 0.00715 2.631 3 0.0189 2.612 0.00717 2.630 4 0.0190 2.644 0.00712 2.663 5 0.0188 2.620 0.00713 2.639 6 0.0195 2.697 0.00719 2.716 7 0.0191 2.639 0.00718 2.658 MM800-13

DOE Methods 12.0 Method Performance The average, standard deviation, and % RSD for instrument check samples and prep-check samples were calculated for establishing the method performance. These values are listed in Table 4 as follows: Table 4. Instrument and Preparation Checks Instrument Check Variable 235 U (ppm) 238 U (ppm) 235 U Abundance Total (ppm) Average 0.00722 1.003 0.00715 1.010 Standard Deviation 0.0001 0.0110 0.00002 0.0111 % rsd 1.16 1.10 0.35 1.10 Preparation Check Variable 235 U (ppm) 238 U (ppm) 235 U Abundance Total (ppm) Average 0.01905 2.643 0.00716 2.663 Standard Deviation 0.003 0.0314 0.00003 0.0317 % rsd 1.35 1.19 0.35 1.19 13.0 References Hudgen, J. E. 1956. Ion Exchange in Analytical and Radiochemistry. Ion Exchange and Chromatography in Analytical Chemistry. Fifty-Ninth Annual Meeting Papers. ASTM Special Technical Publication No. 195. American Society for Testing and Materials, Philadelphia, Pennsylvania, pp. 11-26. Kraus, K. A., and F. Nelson. 1956. Metal Separations by Anion Exchange. Ion Exchange and Chromatography in Analytical Chemistry. Fifty-Ninth Annual Meeting Papers. ASTM Special Technical Publication No. 195. American Society for Testing and Materials, Philadelphia, Pennsylvania, pp. 27-57. Kunin, R. 1956. Ion Exchange in Analytical Chemistry. Ion Exchange and Chromatography in Analytical Chemistry. Fifty-Ninth Annual Meeting Papers. ASTM Special Technical Publication No. 195. American Society for Testing and Materials, Philadelphia, Pennsylvania, pp. 3-10. MM800-14

Analytical/Inorganic Marcus, Y., and A. S. Kertes. 1968. Ion Exchange and Solvent Extraction of Metal Complexes, Wiley Interscience, New York, p. 943. MM800-15