HI Propylene Glycol Refractometer

Similar documents
HI Propylene Glycol Refractometer INSTRUCTION MANUAL

Instruction Manual HI Refractometer for Sodium Chloride Measurements

Instruction Manual HI Refractometer for Invert Sugar Measurements

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

Instruction Manual HI Refractometer for Sucrose in Wine and Grape Products Measurements

HI Seawater Refractometer

INSTRUCTION MANUAL Milwaukee Refractometer MA888 Refractometer for Ethylene Glycol Measurements

Instruction Manual. HI Refractometer for Sucrose in Wine and Grape Products Measurements

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

INSTRUCTION MANUAL Milwaukee Refractometer MA886 Refractometer for Sodium Chloride Measurements

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

INSTRUCTION MANUAL Milwaukee Refractometer MA872 Refractometer for Fructose Measurements

HI 96811, HI HI 96813, HI 96814

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

MA882, MA883, MA884, MA885

Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer

HI REDUCING SUGARS ISM for wine analysis

Instruction Manual HI 96738C Chlorine dioxide ISM

Instruction Manual HI 96728C Nitrate ISM

HI Hardness ISM

Instruction Manual HI Potassium LR and MR ISM

Instruction Manual HI 96720C Calcium Hardness ISM

Instruction Manual HI 96710C ph & Chlorine ISM

IR/Thermistor Pen. Instruction Manual

SUGAR/BRIX REFRACTOMETER INSTRUCTION MANUAL

Today s Weather 24 Hour Forecaster

HI HI 8633 HI HI 8734 HI

HI Free & Total Chlorine ISM

PHOTOMETER FOR BOAR SEMEN ANALYSIS

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

INSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7

OPERATING INSTRUCTIONS

Bante210 Benchtop ph Meter Instruction Manual

Non - Contact Forehead Infrared Thermometer. Care Q7 USER MANUAL - 1 -

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

DHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual

Rapid RH Portable ph Meter. User's Manual

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

Eclipse Refractometer General Models

WS-7047TWC Wireless 433 MHz Weather Station With Rainfall and Temperature. Instruction Manual

Fischer Instruments Chrome and Black Wood Base Weather Station with Barometer, Hygrometer, Thermometer and Quartz Clock User Manual

Owner's Manual. for AUTOMATIC LEVEL. NBL Series. FOR CUSTOMER SERVICE, PARTS & REPAIR, CALL Toll Free:

Owner's Manual NSL100B BUILDERS LEVEL NSL500B TRANSIT LEVEL

WS-7211U Wireless 433 MHz Weather Station. Instruction Manual

IR F ( C) ENGLISH INSTRUCTION MANUAL. Thermometer

Fischer 1508BTH-45 5" Brass Barometer with Temperature & Humidity User Manual

Ambient Weather WS-152B 6" Contemporary Barometer User Manual

ph 5 Meter Manual The LaMotte 5 Series

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual

Fischer Instruments Stainless Steel Barometer User Manual

WeatherHub2 Quick Start Guide

DIGITAL ABBE REFRACTOMETER INE-WYA-2S OPERATING INSTRUCTION PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

INSTRUCTION MANUAL DTH Fax: (503)

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL

Portable ph/orp/ion Meter D-73 Instruction Manual CODE:GZ A

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual

ph / Temperature Meter for Cheese with Application Specific Probe

Operating Manual. Turb 355 IR/T. Portable turbidimeter. ba75454e03 09/

Ambient Weather WS-228TBH 9" Brushed Aluminum Traditional Barometer with Temperature and Humidity, Radiant Blue User Manual

Fischer Banjo Weather Station with Thermometer, Hygrometer, Barometer User Manual

One-hand ph/temperature measuring instrument

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:

Ambient Weather DHR70B-BRASS Handheld Fishing Barometer User Manual

POTASSIUM ION SELECTIVE ELECTRODE

KD2. User s Manual. Thermal Properties Analyzer. version 6. Decagon Devices, Inc.

SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL

CAUTION: BEFORE YOU BEGIN, ALWAYS MAKE SURE THE WEAPON IS UNLOADED AND SAFE TO HANDLE.

TH2821B Portable LCR Meter

Model ST-616CT. Instruction Manual. Infrared Thermo-Hygrometer. reedinstruments. www. com

Field Pro Operation and Installation Manual

LEAPERS, INC. RED/GREEN DOTS

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Density of an Unknown

HL-800K Infrared Laser Thermometer. -50 C to +800 C (13:1 ratio) User Manual

1450 First Ave. Page 1 of 10 Chippewa Falls, WI (800) or (715) Fax: (715)

Model: WT-3181PL Instruction Manual DC: Indoor/Outdoor Atomic Wall Clock. Get Started

Eppendorf UVette. Instructions for use

ICE MANTLE MAKER MODEL 452 User Maintenance Manual/Handbook

HS515G-M. PARALOW Circle Dot Sight. User's Manual. Holosun Technologies Inc. Phone: (909) Fax: (909)

Mini Environmental Quality Meter

Instruction Manual C401/301/201/102/101. Portable Colorimeter. 99 Washington Street Melrose, MA Phone Toll Free

Chemistry MultiMeasure Sensor PS-2170

SkyScan EWS-PRO - Manual -

InfraRed Thermometer with Laser Pointer

Vet Photometer DP 700 Operating Manual. Version V5.9

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Anemometer Thermometer Hygrometer

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.

Drive electronics pq12

Guide to ph Analysis.

product manual H-3220A Benkelman Beam

WS-9018U Wireless Weather Station. Instruction Manual

Coating Thickness Gauge TW-8. art. nr Thoma Tel./Fax: Operation Manual

ICE MANTLE MAKER MODEL 452 User Maintenance Manual/Handbook

CONTENTS. La Crosse Technology, Ltd. Page 1

VIBRATION METER. Instruction Manual.

Instruction Manual SEFRAM Thermal Camera

Spectrum 22ED Spectrophotometer #FOT Absorbency 0A to 1.99A Photometric accuracy 2.0% T Power requirement 115/230 VAC; Hz automatic Size

Transcription:

Instruction Manual HI 96832 Propylene Glycol Refractometer www.hannainst.com 1

Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct use of this instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. TABLE OF CONTENTS PRELIMINARY EXAMINATION... 2 GENERAL DESCRIPTION... 3 SPECIFICATIONS... 3 PRINCIPLE OF OPERATION... 4 MEASUREMENT UNITS... 5 FUNCTIONAL DESCRIPTION... 5 DISPLAY ELEMENTS... 6 MEASUREMENT GUIDELINES... 7 CALIBRATION PROCEDURE... 7 MEASUREMENT PROCEDURE... 8 TO CHANGE MEASUREMENT UNIT... 9 TO CHANGE TEMPERATURE UNIT... 9 MAKING A STANDARD SOLUTION...10 BATTERY REPLACEMENT...10 WARRANTY...10 ERROR MESSAGES...11 PRELIMINARY EXAMINATION Remove the instrument from the packing materials and examine carefully to ensure no damage has occurred during shipping. If any damage has occurred, notify your Dealer or closest Hanna Customer Service Center. Each instrument is supplied with: 9 V battery Instruction manual Note:Save all packing material until you are sure that the instrument functions correctly. A defective instrument must be returned in its original packing. All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA. 2

GENERAL DESCRIPTION The HI 96832 Digital Refractometer is a rugged portable, water resistant device that utilizes the measurement of refractive index to determine the % volume and freezing point of propylene glycol based solutions. Propylene glycol is used in many applications including: pharmaceutical solvent humectant food additive emulsification agent moisturizer in cosmetics, toothpaste, tabacco products, etc. hand sanitizer and lotion antifreeze The HI 96832 device benefits from Hanna s years of experience as a manufacturer of analytical instruments. The digital refractometer eliminates the uncertainty associated with mechanical refractometers and is easily portable for use in the field. The HI 96832 refractometer is an optical device that is simple and quick to use. Samples are measured after a simple user calibration with distilled or deionized water. Within seconds, the refractive index and temperature are measured and converted into one of two measurement units; % Volume or Freezing Point. The instrument utilizes internationally recognized references for unit conversion and temperature compensation for propylene glycol solutions (e.g. CRC Handbook of Chemistry and Physics, 87 th Edition). The temperature (in C or F) is also displayed on the large dual level display along with helpful message codes. Key features include: IP65 waterproof protection Automatic Temperature Compensation (ATC) Battery operation with Low Power indicator (BEPS) Automatically turns off after 3 minutes of non-use SPECIFICATIONS % Volume F reezing Point C ( F) R ange: 0 to 100 % 0 to -51 C (32 to -59.8 F) 0 to 80 C (32 to 176 F) R esolution: 0.1 % 0.1 C (0.1 F) 0.1 C (0.1 F) A ccuracy: ± 0.3 % ± 0.5 C (±1.0 F) ±0.3 C (±0.5 F) Temperature Compensation: Automatic between 0 and 40 C (32-104 F) Measurement Time: Approximately 1.5 seconds Minimum Sample Volume: 100 μl (cover prism totally) Light Source: Yellow LED Sample Cell: Stainless Steel ring and flint glass prism Case Material: ABS Enclosure Rating: IP 65 3

Battery Type / Life: Auto-Off: Dimensions: Mass: PRINCIPLE OF OPERATION 1 X 9V / 5000 readings After 3 minutes of non-use 19.2(W) x 10.2(D) x 6.7(H) cm 420 g Propylene Glycol determinations are made by measuring the refractive index of a solution. Refractive Index is an optical characteristic of a substance and the number of dissolved particles in it. Refractive Index is defined as the ratio of the speed of light in empty space to the speed of light in the substance. A result of this property is that light will bend, or change direction, when it travels through a substance of different refractive index. This is called refraction. When passing from a material with a higher to lower refractive index, there is a critical angle at which an incoming beam of light can no longer refract, but will instead be reflected off the interface. The critical angle can be used to easily calculate the refractive index according to the equation: sin (θ critical ) = n 2 / n 1 Where n 2 is the refractive index of the lower-density medium; n 1 is the refractive index of the higher-density medium. In the HI 96832 refractometer, light from an LED passes through a prism in contact with the sample. An image sensor determines the critical angle at which the light is no longer refracted through the sample. Specialized algorithms then apply temperature compensation to the measurement and convert the refractive index to: % Volume or Freezing Point. 4

MEASUREMENT UNITS Freezing point is displayed as a temperature from 0.0 to -51.0 C corresponding to 0-60 % by volume. When the display shows LO, the freezing point is at a minimum (below -51 C). FUNCTIONAL DESCRIPTION TOP VIEW 1. Liquid Crystal Display (LCD) 2. READ Key (User Measurement) 3. ZERO Key (User Calibration) 4. RANGE Key (User Measurement Unit) 5. ON/OFF 6. Stainless Steel Sample Well and Prism 7. Secondary Display 8. Primary Display 5

BOTTOM VIEW DISPLAY ELEMENTS 9. Battery Cover 10.Battery Compartment 1. Battery (blinks when low battery condition detected) 2. Measurement in Progress Tag 3. SETUP: Factory Calibration Tag 4. CAL: Calibration Tag 5. Automatic Temperature Compensation (blinks when temperature exceeds 0-40 C / 32-104 F range) 6. Primary Display (displays measurement and error messages) 7. Temperature Units 8. Secondary Display (displays temperature measurements; when blinking, temperature has exceeded operation range: 0-80 C / 32-176 F) 9. Range indicator 6

MEASUREMENT GUIDELINES Handle instrument carefully. Do not drop. Do not immerse instrument under water. Do not spray water to any part of instrument except the sample well located over the prism. The instrument is intended to measure propylene glycol solutions. Do not expose instrument or prism to solvents that will damage it. This includes most organic solvents and extremely hot or cold solutions. Particulate matter in a sample may scratch the prism. Absorb sample with a soft tissue and rinse sample well with deionized or distilled water between samples. Use plastic pipettes to transfer all solutions. Do not use metallic tools such as needles, spoons or tweezers as these will scratch the prism. Cover sample well with hand if measuring in direct sun. To reduce the effects of evaporation or absorption of water when taking readings over a period of time, the prism and sample well can be covered with plastic wrap. CALIBRATION PROCEDURE Calibration should be performed daily, before measurements are made, when the battery has been replaced, between a long series of measurements, or if environmental changes have occurred since the last calibration. 1. Press the ON/OFF key, then release. Two instrument test screens will be displayed briefly; all LCD segments followed by the percentage of remaining battery life. The meter will then briefly display an indication of the measurement unit set. When the LCD displays dashes, the instrument is ready. 2. Using a plastic pipette, fill the sample well with distilled or deionized water. Make sure the prism is completely covered. Note: If the ZERO sample is subject to intense light such as sunlight or another strong source, cover the sample well with your hand or other shade during the calibration. 7

3. Press the ZERO key. If no error messages appear, your unit is calibrated. (For a description of ERROR MESSAGES see page 11). Note: The 0.0 screen will remain until a sample is measured or the instrument is turned off. 4. Gently absorb the ZERO water standard with a soft tissue. Use care not to scratch the prism surface. Dry the surface completely. The instrument is ready for sample measurement. Note: If the instrument is turned off the calibration will not be lost. MEASUREMENT PROCEDURE Verify the instrument has been calibrated before taking measurements. 1. Wipe off prism surface located at the bottom of the sample well. Make sure the prism and sample well are completely dry. 2. Using a plastic pipette, drip sample onto the prism surface. Fill the well completely. Note: If the temperature of the sample differs significantly from the temperature of the instrument, wait approximately 1 minute to allow thermal equilibration. 3. Press the READ key. The results are displayed as % Volume or Freezing Point. Note: The last measurement value will be displayed until the next sample is measured or the instrument is turned off. Temperature will be continuously updated. Note: The ATC tag blinks and automatic temperature compensation is disabled if the temperature exceeds the 0-40 C / 32-104 F range. 8

4. Remove sample from the sample well by absorbing with a soft tissue. 5. Using a plastic pipette, rinse prism and sample well with distilled or deionized water. Wipe dry. The instrument is ready for the next sample. TO CHANGE MEASUREMENT UNIT Press the RANGE key to select the measurement units. The instrument toggles between the two measurement scales each time the key is pressed and the primary display indicates Vol % Volume or FrPt for Freezing Point. When the instrument displays the screen with 4 dashes, the instrument is ready for measurement. A number on the display indicates the selected unit: 1 denotes % Volume and 2 denotes Freezing Point (as indicated on the instrument cover). Note: The temperature measurement is not displayed in Freezing Point mode. TO CHANGE TEMPERATURE UNIT To change the temperature measurement unit from Celsius to Fahrenheit (or vice versa), follow this procedure. 1. Press and hold the ON/OFF key continuously for approximately 8 seconds. The LCD will display the all segment screen followed by a screen with the model number on the primary display and the version number on the secondary display. Continue pressing the ON/OFF key. 8 seconds 2. While continuing to hold the ON/OFF key, press the ZERO key. The temperature unit will change from C to F or vice versa. Note: The selected unit will be used in Freezing Point mode. 9

MAKING A STANDARD SOLUTION To make a Propylene Glycol Solution, follow the procedure below: To make an X % propylene glycol solution volumetrically add X ml of high purity propylene glycol (CAS #: 57-55-2; MW 76.09) to a 100 ml Grade A volumetric flask. Use distilled or deionized water to bring the flask close to the total volume, mix, allow the solution to return to room temperature. Once the solution has returned to room temperature use distilled or deionized water to bring the total volume to 100 ml. Mix solution well before use. Propylene Glycol Total Volum e Expected Freezing Point Value 10 % V 10.00 ml 100.00 ml -3.4 C (25.9 F) 40 % V 40.00 ml 100.00 ml -21.3 C (-6.3 F) BATTERY REPLACEMENT To replace the instrument s battery, follow these steps: Make sure the instrument is off. Turn instrument upside down and remove the battery cover by turning it counterclockwise. Extract the battery from its location. Replace with a new 9V battery making sure to observe polarity. Insert the back battery cover and fasten it by turning clockwise to engage. WARRANTY The HI 96832 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your dealer. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase. 10

ERROR MESSAGES Err Error Code LO Primary display LCD Description General failure. Cycle power to instrument. If error persists, contact Hanna. S a m p l e e x c e e d s measurement range. m i n i m u m HI Primary display LO Primary display CAL segment ON HI Primary display CAL segment ON t LO Primary display CAL segment ON S a m p l e e x c e e d s measurement range. m a x i m u m W ron g sol u ti on u sed to ze ro i n stru m en t. Use d ei on i zed or distilled water. Press Z ERO. W ron g sol u ti on u sed to ze ro i n stru m en t. Use d ei on i zed or distilled water. Press Z ERO. Temperature exceeds ATC low limit (0 C) during calibration. t HI Primary display CAL segment ON Temperature exceeds ATC high limit (40 C) during calibration. Air Prism surface insufficiently covered. ELt nlt Too m u c h e xt e r n a l l i g h t f or measurement. Cover sample well with hand. LED light is not detected. Contact Hanna. B attery segment blinking <5 % of battery life is remaining. Temperature values are blinking 0.0 C or 80.0 C Temperature measurement out of range (0 to 80 C). A TC segment blinking Outside temperature compensation range (0 to 40 C). S ETUP segment blinking Factory calibration lost. Contact Hanna. 11

Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.com Local Sales and Customer Service Office Printed in EUROPE (ROMANIA) MAN96832 01/10 12