Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Similar documents
Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Abstract. Experimental Sample Preparation

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Detection of 9-tetrahydrocannabinol ( 9-THC) in human urine by Solid Phase Extraction and HPLC.

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

LC-MS/MS Analysis of Phytocannabinoids and their

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Application Note. Edgar Naegele. Abstract

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Application Note. Abstract. Authors. Environmental

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Psychoactive Drugs Urine LC-MS/MS Analysis Kit User Manual

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

Determination of Cannabinoids (THC) in Biological Samples

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Authors. Introduction. Abstract. Forensics

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Appendix II- Bioanalytical Method Development and Validation

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

ENVIRONMENTAL analysis

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Confirmation and Quantitation of THC in Oral Fluid Using an MRM Method on a GC-Triple Quadrupole MS

Modernizing the USP Monograph for Acetaminophen

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

AppNote 7/2014. Automating Liquid-Liquid Extractions using a Bench-top Workstation KEYWORDS ABSTRACT. Sample Preparation, Lab Automation, LC/MS/MS

A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Opiates in Urine by SAMHSA GC/MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Supporting Information

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Detection of Phencyclidine in Human Oral Fluid Using Solid Phase Extraction and Liquid Chromatography with Tandem Mass Spectrometric Detection

Tecan AC Extraction Plate

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

770-9P. Sensitivity and Selectivity - A Case Study of LC/MS Enantioselective Resolution of Bupivacaine Using Vancomycin as a Chiral Stationary Phase

Transcription:

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source Application Toxicology Authors Jürgen Wendt Agilent Technologies Sales and Support GmbH Waldbronn, Germany Jörg Röhrich Institute of Legal Medicine, University Mainz Mainz, Germany Abstract LC/MS/MS is a useful analytical technique for the analysis of amphetamines and cannabinoids in biological matrices. Amphetamines ionize well in electrospray ionization (ESI), whereas cannabinoids exhibit better sensitivity with atmospheric pressure chemical ionization (APCI). Using a 1.8-µm particle size RRHT column for the LC separation, the Agilent G1978B multimode ion (MMI) source was utilized in order to achieve a balanced response for both compound classes in a single analysis. The presented method exhibits good within-day and dayto-day reproducibility. The coefficients of variation ranged from 3 to 15%; most of the coefficients were in the 5 to 10% range. Introduction Cannabis is the most widely used illicit drug in Europe and in the United States of America. Furthermore, the abuse of sympathomimetic drugs such as amphetamine derivatives has increased considerably during the last years. Consequently, driving under the influence of drugs of abuse has become a serious problem for traffic safety. Due to this reason, driving after consumption of cannabis and amphetamines, including their methylenedioxy-derivatives methylenedioxymethamphetamine (MDMA) and methylenedioxyethylamphetamine (MDE), was sanctioned by the German Road Traffic Act in 1998. Since then, the number of toxicological analyses of serum for Δ9-tetrahydrocannabinol (THC) or amphetamine derivatives has increased enormously. Therefore laboratories need analytical methods that can handle a large number of samples in a relative short time. An appropriate technique to meet these needs is LC/MS/MS. Amphetamines are basic polar compounds and ionize well in electrospray ionization (ESI), whereas the relatively nonpolar cannabinoids exhibit better sensitivity with atmospheric pressure chemical ionization (APCI) (Figure 1). To use the optimum ionization technique for each class of drug in a single run, the Agilent G1978B multimode ion (MMI) source (Figure 2) was evaluated in order to achieve a balanced response for both compound classes. The MMI source can operate in either ESI or APCI modes or in mixed mode, which is simultaneous ESI and APCI. The choice and parameters of ionization mode can be timeprogrammed during the run.

Figure 1. Comparison of the MDA and THC response in ESI and APCI modes. each of methanolic solution of D 11 -amphetamine, D 11 -methamphetamine, D 5 -MDA, D 5 -MDMA, D 6 -MDE, and D 9 -THC-COOH and 0.1 ng/µl each of D 3 -THC and D 3 -THC-OH). The sample was mixed for 3 minutes and the mixture was centrifuged at 3,000 rpm for 10 minutes. Solid-phase extraction was either automated, using the Caliper Rapid- Trace SPE workstation, or was done manually, using a vacuum manifold. The supernatant was applied to a solid-phase extraction column (Bakerbond SPE C18, 500 mg), which had been conditioned by flushing with 2 3 ml of methanol and 2 ml of water. The column was rinsed with 2 2 ml water, 2 2 ml water/ methanol (80:20; v/v), and 1 ml of 0.1 M acetic acid. The column was dried for 10 minutes. Figure 2. Experimental Reagents 2 Design of the multimode source. All solvents and reagents were analytical grade. Methanol, acetone, acetic acid, dichloromethane, 2-propanol, and ammonia were purchased from E. Merck (Darmstadt, Germany) or from Sigma- Aldrich (Deisenhofen, Germany). Solid-phase extraction columns were purchased from Mallinckrodt Baker (Griesheim, Germany), and all drug standard solutions and deuterated compounds were purchased from Cerilliant (Austin, TX). Sample Preparation A 1-mL sample of serum was diluted with 6 ml of phosphate buffer (0.1 M, ph 6). Then 50 µl of the internal standard mixture was added (1 ng/µl The elution was carried out in two steps. First the cannabinoids were eluted with 3 ml of dichloromethane/acetone (50:50; v/v), followed by elution of amphetamines, opiates, and cocaine/ metabolites with 3 ml of dichloromethane/ 2-propanol/ammonia (40:10:2; v/v/v). Both extracts were evaporated under a slight stream of nitrogen at 30 C, reconstituded in 0.1 ml methanol, and added together. This combined SPE fraction was diluted with water (ratio 1:4) to improve the chromatographic peak shape. LC/MS/MS Method The LC/MS/MS consisted of an Agilent 1200 Rapid Resolution liquid chromatograph and an Agilent G6410A Triple Quadrupole mass spectrometer. Different ZORBAX columns were evaluated in combination with different solvents, flow rates, and column parameters to optimize the speed of the analysis while maintaining a good chromatographic resolution.

The best results were obtained with a 1.8-µm particle size ZORBAX SB-C18 column (2.1 100 mm) using a water/acetonitrile gradient (both containing 0.1% formic acid). The detailed LC conditions are listed in Table 1. Table 1. LC Method Column Zorbax RRHT SB-C18 (2.1 mm id 100 mm, 1.8 µm) p/n 828700-902 Column temperature 70 C Mobile phase A: 0.1% formic acid in water B: 0.1% formic acid in acetonitrile Flow rate Gradient Stop time Post time Injection volume 0.6 ml/min 10% B at 0 min 10% B at 2 min 95% B at 8 min 95% B at 11min 10% B at 11.5min 15 min None 10 µl (sample diluted with water 1:4 to improve peak shape) In addition to the standard ESI and APCI sources, an Agilent G1978B Multimode source was coupled to the mass spectrometer. The MMI source was operated in mixed mode (ESI and APCI simultaneously in positive polarity) or alternatively in pure ESI and APCI modes, switching between these ionization techniques based on a chromatographic time scale. The optimized source parameters are shown in Table 2. Determination of the optimal MRM transitions for both analytes and internal standards was carried out by flow injection analysis of the single components at concentration levels around 1 µg/ml. See Table 3. Results and Discussion Four possible MMI modes were investigated (Figure 3). Using an MMI method that begins in ESI mode and switches to APCI mode after five minutes resulted in the best overall compound responses. The LC method was not fully optimized for speed (Table 1) because the vaporizer temperature is changed from 175 to 250 C after the switch of the ionization mode, and that change requires some short time before the cannabinoids elute. The total run time, including the re-equilibration time of the column at starting gradient conditions, was 15 minutes. Table 2. Optimized MMI-Parameters Neb. press Drying gas Drying gas Charging Capillary Vaporizer Corona MMI mode (psi) flow (L/min) temp ( C) voltage (V) voltage (V) temp ( C) current (µa) Mixed 40 5 300 2000 2000 200 2 ESI 60 5 300 2000 2000 175 0 APCI 20 5 300 2000 2000 250 4 Table 3. Data Acquisition Parameters for the MRM Transitions RT Precursor Frag CE Product ion CE Product ion 2 Compound (min) (M H) + (V) (V) (m/z) (V) (m/z) Amphetamine 1.3 136 100 15 91 10 119 D 11-Amphetamine 1.3 147 100 15 127 15 97 MDA 1.4 180 100 15 105 15 135 D 5-MDA 1.4 185 100 10 168 15 138 Methylamphetamine 1.5 150 100 15 91 10 119 D 11-Methylamphetamine 1.5 161 100 15 127 15 97 MDMA 1.9 194 100 10 163 15 135 D 5-MDMA 1.9 199 100 15 165 10 135 MDE 2.6 208 100 15 135 15 147 D 6-MDE 2.6 214 100 15 166 15 136 THC-OH 7.9 331 110 30 193 30 201 D 3-THC-OH 7.9 334 110 20 316 25 196 THC-COOH 8.1 345 110 30 193 30 299 D 9-THC-COOH 8.1 354 120 22 308 25 196 THC 8.6 315 110 30 193 30 259 D 3-THC 8.6 318 100 30 196 30 105 3

Figure 3. Comparison of the different MMI modes. Method Validation The LC/MS/MS method was validated for the detection and quantification of THC, THC-OH, THC-COOH, amphetamine, methamphetamine, MDA, MDMA, and MDE in serum. The validation of the method was carried out according to Peters et al [1] and the German Society of Toxicology and Forensic Chemistry (GTFCh). The method validation was performed by using a Microsoft Excel-based validation program (VALISTAT [2]). Drug-free serum was used as a blank matrix for the validation measurements. Seven calibration standards were prepared. The different calibration levels were obtained by spiking the blank serum with 50 µl of methanolic solutions containing appropriate amounts of the analytes. The calibration levels are shown in Table 4. Table 4. Calibration Range for Amphetamines and Cannabinoids in Serum Samples Compound Cal 1 Cal 2 Cal 3 Cal 4 Cal 5 Cal 6 Cal 7 Cal 8 Amphetamine 0 10 20 40 60 80 100 500 Methamphetamine 0 10 20 40 60 80 100 500 MDA 0 10 20 40 60 80 100 500 MDMA 0 10 20 40 60 80 100 500 MDE 0 10 20 40 60 80 100 500 THC 0 0.5 1 2 3 4 5 25 THC-OH 0 0.5 1 2 3 4 5 25 THC-COOH 0 5 10 20 30 40 50 250 4

A seven-point calibration curve for each compound was obtained by measuring of the calibration standards in six replicate injections. The calibrations were linear in the range tested and the correlation coefficients were > 0.98 for all compounds. The S/N calculations for calibration standard Cal 3, which represents the limit of quantitation, were based on peak-to-peak noise definition and no smoothing was applied. All quantifier and qualifier ions of the amphetamines and cannabinoids can be easily detected, even when diluting the methanol-reconstituted SPE fractions with water (ratio 1:4) to improve the chromatographic peak shape (Figure 4). Intra-assay and inter-assay precision data were obtained from two analyses in a series performed on eight different days at two concentration levels (low, high). The intra-assay precision (within-day reproducibility) is defined as the mean value of the eight coefficients of variation (CV) from the two measurements carried out on one day. Inter-assay precision (day-to-day reproducibility) is the coefficient of variation from the average of the eight mean values of the two measurements carried out on one day. The intra-assay coefficients of variation ranged from 2.9 to 15.3 % (Table 5). The day-to-day coefficients of variation ranged from 3.4 to 15.3 % Figure 4. S/N calculation for standard Cal 3. 5

www.agilent.com/chem Table 5. Inter-Assay and Intra-Assay Precision at Two Concentration Levels (Cal 3 and Cal 7) Intra-assay precision in % Inter-assay precision in % Compound CalStandard 3 CalStandard 7 CalStandard 3 CalStandard 7 Amphetamine 4.3 2.9 4.9 5.6 Methamphetamine 4.7 4.7 6.1 5.5 MDA 7.6 4.2 8.4 4.6 MDMA 5.9 3.2 5.9 3.4 MDE 8.5 5.2 8.5 5.6 THC 9.5 5.8 10.0 6.3 THC-OH 8.4 8.7 11.6 8.7 THC-COOH 15.3 5.4 15.3 5.5 Conclusions The use of a 1.8-µm particle size RRHT column for the LC separation provides a faster analysis (cycle time 12 min) than GC/MS (cycle time 45 min). Due to the polarity differences of the two compound classes, the use of the multimode ion source allows the detection of the eight compounds with an optimal response for each compound (switching the ionization mode on a time-based scale leads to the best results). In comparison to the established GC/MS method, the RRLC/QQQ method shows a higher sensitivity and selectivity (considering an injection volume of 1 µl in GC/MS and 10 µl in LC/MS/MS with a dilution factor of 4). The presented method exhibits good within-day and dayto-day reproducibility. The coefficents of variation ranged from 3 to 15%; most of the coefficients were in the 5 to 10% range. Acknowledgements The authors are very grateful to John Hughes, PhD, (Agilent Technologies Inc., Pleasanton, CA) for reviewing the manuscript and making helpful comments. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. In the future, other drugs of abuse (opiates like morphine, 6-acetylmorphine, and codeine as well as cocaine and its metabolites) will be included in this RRLC/QQQ method. Also, the use of online SPE will be evaluated. References 1. F. T. Peters and H. H. Maurer, Bioanalytical Method Validation and Its Implications for Forensic and Clinical Toxicology A review, Accred. Qual. Assur. 7, 441 449 (2002) 2. G. Schmitt, M. Herbold, and F. Peters, Methodenvalidierung Im Forensisch-Toxikologischen Labor, Arvecon Walldorf (2003) Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2008 Printed in the USA June 10, 2008 5989-8368EN