Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Similar documents
US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Methanol Extraction of high level soil samples by USEPA Method 8260C

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Optimizing. Abstract: is standardd. procedures. altered to

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Solid Phase Micro Extraction of Flavor Compounds in Beer

Maximizing Sample Throughput In Purge And Trap Analysis

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Maximizing Production While Minimizing Costs

Optimizing Standard Preparation

Analytical Trap Comparison for USEPA Method 8260C

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Ed George and Anaïs Viven Varian, Inc.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

2017 EPA Method Update Rule and EPA Method 624.1

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

EPA TO-17 Volatile Organic Compounds

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Determination of Volatile Organic Compounds in Air

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

CALA Directory of Laboratories

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Target Compound Results Summary

Volatile Organic Compounds in Every Day Food

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

STANDARD OPERATING PROCEDURES

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Copies: Dave Favero, RACER File

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Associates. Project No.: October 25, 2017

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Volatile Organic Compounds in Water PBM

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Thermal Desorption Technical Support

Table 1 Soil and Groundwater Sampling Summary Table

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Study of Residual Solvents in Various Matrices by Static Headspace

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

In-Tube Extraction Sample Preparation for Gas Chromatography

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

317 Elm Street Milford, NH (603) Fax (603)

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

METHOD MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN WATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY. Revision 4.

Non-Potable Water A2LA Certificate

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

Laboratory Report Number(s) Lone Tree Road (Deep) 10/5/15, 10/19/15, 6/13/16. Address Sample Date(s) Analyses Requested

/J~ GOLDER. Golder Associates Inc. golder.com. Project No April 25, 2018

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Date Issued: April 01, 2013 Expiration Date: June 30, 2013

Transcription:

Validation of USEPA Method 524.2 Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS Application Note By: Nathan Valentine Abstract The US EPA developed Method 524.2¹, Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry, for identifying and measuring purgeable volatile organic compounds (VOC) in surface water, groundwater, and drinking water. Due to the required sensitivity of this method, the use of purge and trap gas chromatography is vital. This study will utilize a Teledyne Tekmar Stratum Purge and Trap Concentrator (PTC) and AQUATek 100 Autosampler in conjunction with a Perkin-Elmer Clarus 600 GC/MS. A linear calibration and Method Detection Limits (MDLs) will be established for Method 524.2¹ compounds. Introduction Clean drinking water is something that many of us take for granted. This is due, in part, to the stringent water quality criteria established by the US EPA. Methods such as 524.2¹ allow for detection of a widerange of VOCs at very low levels. VOCs are known drinking water contaminants with potentially harmful long- and short-term health effects. The reason methods can be created with such precise, low-level performance criteria is through the use of Purge and Trap technology. For this study, a Stratum PTC was used in conjunction with an AQUATek 100 Autosampler. This set-up allows for complete automation of sample preparation for the analysis of liquid samples for purge and trap. Through the features the AQUATek 100 provides, such as the 100-position sample tray and standard addition vessels, efficiency and throughput can be greatly increased, leading to cost and time savings. Utilizing a Perkin-Elmer Clarus 600 GC/MS, a linear calibration was performed and percent Relative Standard Deviation (%RSD) and Method Detection Limits (MDLs) were determined for the full list of compounds. A 25mL purge volume was used and all performance criteria of EPA Method 524.2¹ were met. Experimental-Instrument Conditions The Stratum PTC and AQUATek 100 Autosampler were coupled to a Perkin-Elmer Clarus 600 GC/MS for analysis. Teledyne Tekmar s proprietary #9 trap was the analytical trap used. The GC was configured with a Perkin-Elmer Elite-624 20m x 0.18mm x 1.0µm column. The GC/MS parameters are outlined in Tables 1 and 2. Table 3 outlines the P&T and autosampler conditions.

GC: Column: Oven Program: GC Parameters Perkin-Elmer Clarus 600 Gas Chromatograph Perkin-Elmer Elite-624 20m x 0.18mm x 1.0µm 35 C for 2 min, to 200 C at 10 C/min, for 0 min, to 240 C at 50 C/min MSD: Source: MS Parameters Clarus 600T Quadrupole Mass Spectrometer 200 C Transfer Line Temp: 200 C Inlet: 220 C Solvent Delay: 0.5 min Column Flow: 0.8mL/min Scan Range: 35-300 m/z Gas: Helium Scan Time: 0.2 sec Pressure: 16.1 psi Inter-scan Delay: 0.1 sec Split Ratio: 80:1 Ionization Mode: EI+ Tables 1 & 2: GC and MSD Parameters Stratum PTC and AQUATek 100 Parameters Variable Value Variable Value Pressurize Time 0.95 min Purge Time 11.00 Sample Transfer Time 1.25 min Purge Temp 20 C Rinse Loop Time 0.85 min Purge Flow 40mL/min Sweep Needle Time 0.35 min Dry Purge Time 0.0 min Bake Rinse On Dry Purge Temp 20 C Bake Rinse Cycles 1 Dry Purge Flow 100mL/min Bake Rinse Drain Time 0.60 min GC Start Start of Desorb Presweep Time 0.35 min Desorb Preheat Temp 245 C Water Temp 90 C Desorb Drain On Valve Oven Temp 150 C Desorb Time 2.00 min Transfer Line Temp 150 C Desorb Temp 250 C Sample Mount Temp 90 C Desorb Flow 300mL/min Purge ready Temp 35 C Bake Time 4.00 min Condenser Ready Temp 40 C Bake Temp 280 C Condenser Purge Temp 20 C Bake Flow 200mL/min Standby Flow 5mL/min Condenser Bake Temp 200 C Pre-Purge Time 0.5 min Pre-Purge Flow 40.0mL/min Sample Heater Off Sample Preheat Time 1.00 min Sample Temp 40 C Table 3: Stratum PTC and AQUATek 100 Parameters (Stratum PTC Parameters are in Blue) Calibration Data A 50ppm working calibration standard was prepared in methanol. Calibration standards were prepared in 50mL volumetric flasks filled to volume with de-ionized water over a range of 0.2ppb to 50ppb. Samples were transferred to headspace free 40mL vials for analysis. The Internal Standard (IS) and Surrogate Standards (SS) were prepared in methanol at a 25ppm concentration. After transferring to the standard

vessel on the AQUATek 100, the IS/SS was added in 5µL volumes to each sample, bringing the final concentration of 5ppb. Perkin-Elmer TuroboMass software was used to process the calibration data. Relative response factors were evaluated for linearity and %RSD with results for all compounds listed in Table 4. Method detection limits were also established for all compounds by analyzing seven replicates at a concentration of 0.2ppb. MDL results for all compounds were below 0.2ppb. Compound Name Average RRF % RSD Minimum Detection Limit Dichlorodifluoromethane 0.286 8.6 0.14 Chloromethane 0.559 9.4 0.12 Vinyl Chloride 0.371 6.3 0.13 Bromomethane 0.153 11.2 0.20 Chloroethane 0.228 8.7 0.15 Trichlorofluoromethane 0.283 5.4 0.17 Diethyl Ether 0.143 9.0 0.09 1,1-Dichloroethene 0.223 7.3 0.17 Acetone 0.049 10.4 0.13 Iodomethane 0.137 0.9983* 0.04 Carbon Disulfide 0.681 7.4 0.16 Allyl Chloride 0.180 7.2 0.17 Methylene Chloride 0.220 7.0 0.15 trans-1,2-dichloroethene 0.241 8.3 0.15 MTBE 0.368 8.1 0.09 Acryl Nitrile 0.041 11.5 0.11 1,1-Dichloroethane 0.476 5.1 0.14 2-Butanone 0.066 14.6 0.06 2,2-Dichloropropane 0.382 7.7 0.13 cis-1,2-dichloroethene 0.273 6.5 0.16 Methyl Acrylate 0.092 9.6 0.10 Propionitrile 0.015 10.8 0.11 Methacrylonitrile 0.045 11.0 0.07 Bromochloromethane 0.078 5.4 0.11 Tetrahydrofuran 0.012 13.3 0.08 Chloroform 0.405 4.6 0.14 1,1,1-Trichloroethane 0.322 5.4 0.15 Carbon Tetrachloride 0.257 4.9 0.13 1-Chlorobutane 0.699 6.4 0.12 1,1-Dichloropropene 0.398 5.1 0.15 Benzene 1.241 4.3 0.15 1,2-Dichloroethane 0.183 6.5 0.11 Trichloroethene 0.262 7.0 0.15 Methyl Methacrylate 0.079 9.3 0.08 1,2-Dichloropropane 0.233 2.3 0.17 Dibromomethane 0.067 6.6 0.15 Ethyl Methacrylate 0.162 8.7 0.04 Bromodichloromethane 0.240 5.0 0.10 2-Nitropropane 0.001 16.5 0.19 Chloroacetonitrile 0.010 13.9 0.12 cis-1,3-dichloropropene 0.332 3.5 0.09 4-Methyl-2-Pentanone 0.035 7.8 0.08 Toluene 0.685 5.3 0.16

trans-1,3-dichloropropene 0.227 5.9 0.10 1,1,2-Trichloroethane 0.115 8.3 0.08 Tetrachloroethene 0.433 9.7 0.19 2-Hexanone 0.077 9.7 0.08 1,3-Dichloropropane 0.249 6.4 0.11 Dibromochloromethane 0.109 5.7 0.11 1,2-Dibromoethane 0.098 8.7 0.09 Chlorobenzene 0.648 4.8 0.14 1,1,1,2-Tetrachloroethane 0.172 4.9 0.14 Ethylbenzene 1.263 4.9 0.16 m,p-xylene 1.004 3.6 0.31 o-xylene 0.988 4.4 0.14 Styrene 0.673 2.2 0.15 Nitrobenzene 0.004 9.6 0.15 Bromoform 0.045 15.9 0.06 Isopropylbenzene 1.255 4.1 0.14 Bromobenzene 0.192 3.6 0.15 1,1,2,2-Tetrachloroethane 0.089 8.8 0.07 trans-1,4-dichloro-2-butene 0.028 6.0 0.05 1,2,3-Trichloropropane 0.084 5.9 0.10 n-propylbenzene 1.440 3.5 0.16 2-Chlorotoluene 0.871 3.4 0.14 4-Chlorotoluene 0.977 3.9 0.14 1,3,5-Trimethylbenzene 1.002 3.5 0.13 tert-butylbenzene 0.837 3.5 0.15 1,2,4-Trimethylbenzene 0.950 4.0 0.12 sec-butylbenzene 1.331 3.6 0.14 1,3-Dichlorobenzene 0.436 2.8 0.14 p-isopropyltoluene 0.953 6.9 0.14 1,4-Dichlorobenzene 0.420 2.8 0.14 1,2-Dichlorobenzene 0.360 4.7 0.13 n-butylbenzene 0.896 6.3 0.14 Hexachloroethane 0.113 0.9999* 0.14 1,2-Dibromo-3-chloropropane 0.012 0.9955* 0.10 1,2,4-Trichlorobenzene 0.237 6.1 0.13 Hexachlorobutadiene 0.109 8.6 0.14 Naphthalene 0.348 10.5 0.07 1,2,3-Trichlorobenzene 0.180 8.1 0.12 Average 0.376 7.0 0.13 * Indicates compounds using correlation coefficients from linear regression (r²). Table 4: Calibration and MDL Data for EPA Method 524.2

Figure 1: Total Ion Chromatogram of a 10ppb Calibration Standard for Method 524.2 Conclusions Accuracy and precision are critical when dealing with drinking water analyses because of the impact on public health and safety. This study demonstrates the capabilities of the Teledyne Tekmar Stratum PTC and AQUATek 100 Autosampler coupled with a Perkin-Elmer Clarus 600 system, which met all performance criteria outlined in US EPA Method 524.2¹. By completely automating the sample preparation, without compromising sensitivity, efficiency and throughput can be greatly increased while saving time and money. References 1. USEPA Method 524.2, Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry (GC/MS), Revision 4.1, 1995.