Suppression in anion chromatography. More sensitive analysis of anions and organic acids

Similar documents
881 Compact IC pro 882 Compact IC plus. Compact ion chromatography for intelligent, reliable and precise analysis

883 Basic IC plus. The smallest Metrohm ion chromatograph

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise.

861 Advanced Compact IC

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Air Monitoring. Semi-continuous determination of ambient air quality

Ion chromatography. Ion chromatography

Theory of Ion Chromatography

Analysis of various clays with Metrohm Combustion IC

Intelligent Ion Chromatography. 850 Professional IC 872 Extension Module 858 Professional Sample Processor

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur

Dionex IonPac AS28-Fast-4µm column

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0

ION CHROMATOGRAPHY SYSTEM S 150

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur using pyrohydrolysis

Metrohm Inline Sample Preparation. Ion chromatography analyses could not be easier

columns IonPac AS17-C Anion Exchange Column

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY

Determination of trace anions in concentrated hydrofluoric acid

Metrohm Inline Sample Preparation. Making your ion chromatography more efficient

Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample...

Thermo Scientific Dionex eluent suppressors for ion chromatography

Ion Analysis of Hydraulic Fluids in Fracking Process

columns IonPac AS18 Anion-Exchange Columns

Ion Chromatography. Introduction

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Ion Analysis of Hydraulic Fluids in Fracking Process

columns IonPac AS22 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Automation in ion chromatography. Save time and money through automated sample preparation and analysis

Thermo Scientific. Anion-Exchange Column

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Automation in ion chromatography. Save time and money through automated sample preparation and analysis

suppressors Eluent Suppressors for Ion Chromatography

Sulfur speciation in mining leachates by ion exchange with direct UV detection, using a perchlorate mobile phase.

IC5000 Ion Chromatography System

Ion Chromatography (IC)

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Ion Chromatography. Anion Exchange. Chromatography Ion Exchange Theory. Dr. Shulamit Levin

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Expanded capabilities in ammonium and amine detection

Overcoming Challenging Matrices in Ion Chromatography

TP Analytical Chemistry II 7 th, December 2012

Data Pack. Ion Chromatography Methods. June 2009

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

4. Ion chromatography (IC) 4.1. Educational aims and objectives

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Waters. Ion Chromatography

Determination of Trace Cations in Power Plant Waters Containing Morpholine

suppressors SRS -ULTRA Self-Regenerating Suppressor

Cation Exchange HPLC Columns

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

anion-exchange column

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Novel Electrodeionization Devices: Applications in Inorganic Analysis

The column program. The whole world of ion chromatography

Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography 1

The column program. The whole world of ion chromatography

Inline sample preparation for the determination of anions in sodium hydroxide

IonPac Trace Cation Concentrator

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Anion Analysis of Ophthalmic Irrigation Solutions with the Shimadzu Prominence IC System Craig Young, PhD, Sr. Tech Support Specialist

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

Mobile coulometer for water determination anywhere you want

The determination of trace anions in concentrated phosphoric acid

ProPac WCX-10 Columns

HPLC Praktikum Skript

groups I and II cations, amines, divalent transition metals

The Determination of Trace Anions in Concentrated Phosphoric Acid

PRODUCT MANUAL OMNIPAC PCX-500 GUARD COLUMN (PCX-500 GUARD, P/N ) OMNIPAC PCX-500 ANALYTICAL COLUMN (PCX-500 ANALYTICAL, P/N )

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

It s Time to Improve Component Standards and Measurement (Abstract: Component Cleanliness Should Industry Care?)

Analysis of pulping liquors by suppressed ion chromatography

Chromatographic Methods of Analysis Section - 4 : Ion Exchange Chrom. Prof. Tarek A. Fayed

The power to increase productivity. Thermo Scientific Dionex Reagent-Free Ion Chromatography (RFIC) System Capabilities

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Determination of Inorganic Anions in Wastewater by Ion Chromatography

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Anion Exchange HPLC Columns

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Biochemistry. Biochemical Techniques HPLC

Subject: Ionic Analysis of PrintedCircuit Boards, Printed board Assemblies and Materials, Ion Chromatography Method

columns IonPac SCS 1 Silica Cation Separator

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Transcription:

Suppression in anion chromatography More sensitive analysis of anions and organic acids

What is suppression? 02 Suppression plays a key role in the analysis of anions and organic acids using ion-exchange chromatography and conductivity detection. Suppression in anion chromatography decreases the background conductivity of the eluent minimizes baseline noise transforms analytes in free anions with protons as counterions (which involves a remarkable increase in the conductivity signal) optimizes the signal-to-noise ratio increases the detection sensitivity of the measurement system The Metrohm Suppressor Module The patented Metrohm Suppressor Module (MSM) is a suppression solution that is just as smart as it is reliable. The Metrohm anion suppressor consists of a small rotor, containing three cartridges filled with cation exchanger resin. While one cart ridge is used for suppression, a re gene ration step is performed on the second one. The third cartridge is rinsed automatically during this time. Thus, a freshly regenerated suppressor cartridge is always available for every new sample. And now: STREAM the green way of suppression Suppression with the MSM is optimized and simplified even further thanks to STREAM (Suppressor Treatment Reusing Eluent After Measurement). STREAM refers to a process whereby, rather than simply disposing of the sup pressed eluent after the detection, the eluate is used to rinse the regenerated suppressor unit. STREAM also leads to a significant reduction in regene rant consumption. The advantages of STREAM at a glance No need for ultrapure water as a rinsing solution Less waste Regenerant consumption reduced by a factor of 3 Significant reduction in demand for consumables As a result, the system can be operated for much longer without requiring any intervention by the user. Connection to 1 st cartridge Connection to 2 nd cartridge Connection to 3 rd cartridge

03 Highlights of the Metrohm Suppressor Module for anion chromatography 10 year warranty (anion suppressor) 100% pressure stability 100% resistance to organic solvents Short conditioning time Minimum noise of < 0.2 ns/cm STREAM the green way of suppression Cost-effective and robust Application can be changed at any time with no negative impact on system performance Excellent signal-to-noise ratio enables anions and organic acids to be analyzed down to the ultratrace range

04 Chemical suppression for anions The MSM consists of three cartridges filled with cation exchanger material. The first cartridge is used for suppres sion. Simultaneously, the second cartridge is regenerated with diluted acid (e.g., sulfuric acid) and the third cartridge is rinsed with the eluate or water. Before each analysis, the suppressor is rotated 120 so that a freshly regenerated and rinsed cartridge is always available for suppression. Reactions that take place during chemical suppression Analyte: Na + Cl + RSO 3 H + g H + Cl + RSO 3 Na + R = substrate This example of an anion analysis includes a sodium analyte counterion. This ion is replaced with a proton with an equi valent conductivity that is five times higher. This signifi cantly increases the conductivity of the sample solution and therefore also the signal strength. Salts from weakly dissociated acids (e.g., sodium carbonate/sodium hydrogen carbonate) are used as eluent. Eluent: Na + H + RSO 3 H + g H 2 + RSO 3 Na + Sequential suppression for anions Sequential suppression combines chemical suppression and CO 2 suppression. For this, a Metrohm CO 2 Suppressor (MCS) is used in addition to the MSM. Sequential suppression reduces the background conductivity further (< 1.5 µs/cm), minimizes the impact of carbonate in the sample matrix (> 98% of carbonate re moved) and improves the separation of the injection peak from early eluting anions, such as fluoride. This also further increases the sensitivity of the analysis of anions and organic acids. The MCS consists of a gas-permeable membrane, which is surrounded by a chamber and contains a vacu um. The carbonic acid produced during chemical sup pres sion breaks down into carbon dioxide and water. The carbon dioxide diffuses through the gas-permeable mem brane and is removed. This allows the chemical equi librium to shift more and more towards carbon dio xide and water. As less and less carbonic acid and there fore less disso cia ted carbonic acid is present, background conductivity is reduced even further. This in turn improves the signal-tonoise ratio and allows lower detection limits than those that would be possible with chemical sup pression alone. The reaction that takes place in the Metrohm CO 2 Suppressor (MCS) H + + H H 2 H 2 O + CO 2 (h) R = substrate The eluent counterions are also replaced with protons. The carbonic acid that is produced in this way is unstable and only weakly dissociated, meaning that lower background conductivity is measured. Depending on the eluent composition, background conductivity values of 10 to 20 µs/cm are typical for chemical suppression.

05 Precision with every analysis A fresh suppressor cartridge is used for each measurement with the MSM. This means that users can be sure they are always working with a cation an exchanger that is delivering its full performance: today, tomorrow, and for many years to come. The chromatograms below demonstrate the outstanding reproducibility of measurements provided by the MSM here with sequential suppression for anion analysis: 1 2 3 4 5 7 6 30 injections, separation of the standard anions on Metrosep A Supp 5-100/4.0, eluent 3.2 mmol/l Na 2 / 1.0 mmol/l NaH, flow rate 0.7 ml/min, loop 20 μl, 45 C, sequential suppression Peak 1 2 3 4 5 6 7 Anions Fluoride Chloride Nitrite Bromide Nitrate Phosphate Sulfate Concentration in mg/l 0.25 0.25 0.25 0.25 0.25 0.25 0.25 Relative standard deviation in % (n = 30) 0.26 0.06 0.42 0.17 0.30 0.46 0.25

Eluate MCS Conductivity detector Peristaltic pump MSM 06 Flow diagram for STREAM with peristaltic pump: the eluate is suppressed in the MSM and MCS before it is detected in the conductivity detector. The regenerant is supplied by a peristaltic pump. Regeneration two versions Two options are available when it comes to regenerating the suppressor: the new STREAM process and the conventional version. STREAM is the standard version that comes with every Metrohm IC. As described, STREAM reduces the consumption of chemicals and increases convenience for the user. Conventional suppression is used if the eluent is coupled to another detector following conductivity detection (e.g., PCR-UV/VIS or amperometric detection but also IC-MS or IC-ICP/MS couplings). Thus, a second peristaltic channel is necessary to transfer the rinsing water through the suppressor. Free choice of suppressor The suppressor rotor can be freely selected with ion chromatographs from the 940 Professional IC Vario and 930 Compact IC Flex series. Metrohm offers three che mical anion suppressors with different capacities (MSM, MSM-HC, MSM-LC). The MSM is the suppressor rotor for universal use in routine applications The MSM-HC is the high-capacity suppressor rotor for gradient applications or isocratic applications with very strong eluents The MSM-LC is the low-capacity suppressor rotor with a minimum dead volume for applications with 2-mm columns All three suppressor rotors MSM-HC, MSM, and MSM-LC have the same design and vary solely in terms of their capacity.

Eluate MCS Conductivity detector MSM 800 Dosino 07 Flow diagram for STREAM with Dosino Regeneration: the eluate is suppressed in the MSM and MCS before it is detected in the conductivity detector. An 800 Dosino supplies the regenerant in this system. Regenerant supply For STREAM, the regenerant can be supplied either by a peristaltic pump or an 800 Dosino. The latter of the two versions is known as «Dosino Regeneration» and offers the following advantages: The 800 Dosino requires less maintenance than the peristaltic pump MagIC Net can monitor the volume of regenerant consumed and informs the user if the regenerant needs to be refilled As the peristaltic pump is not required, it can be used for sample preparation techniques such as Inline Ultrafiltration and Inline Dialysis Suppression from a different perspective Until now, suppression has been considered in the context of determining anions and organic acids using ionexchange chromatography and conductivity detection. However, the use of an anion suppressor is recommended if anions are to be determined using UV/VIS detection (in particular between 190 230 nm) too. The sup - pression of the background signal decreases the noise in the UV/VIS detector. In addition, any interferences caused by car bonate in the sample matrix can be avoided; for example, when determining iodide in mi neral water. Lower de tec tion limits can be attained when the suppressor is com bined with the 944 Pro fes sional UV/ VIS Detector Vario. Absorbance [mau] 4.0 3.6 3.2 2.8 2.4 2.0 1.6 1.2 0.8 0.4 with MSM without MSM 0 1 2 3 4 5 6 7 8 9 10 11 12 Time [min] An additional application for the suppressor is inverse suppression. This is used in the determination of or ga nic acids with ion-exclusion chromatography and con ductivity detection, for example. Producing lithium salts from the organic acids in the anion suppressor can signi ficantly increase sensitivity (by a factor of 2-25 de pending on the pk a value of the acid). As a result, this type of ion chromatography can even be used to determine low concentrations. UV/VIS detection at 218 nm with and without suppression: determination of nitrite, bromide, and nitrate

Technical information on the anion suppressor rotors MSM MSM-HC MSM-LC Design Robust and chemically resistant micro packed-bed suppressor Capacity Approx. 90 min* Approx. 350 min* Approx. 45 min** (0.45 meq) (1.8 meq) (0.13 meq) Dead volume < 50 µl < 250 µl < 10 µl Regeneration Chemical regeneration with STREAM Maximum flow rate No irreversible damage possible Backpressure limitations No irreversible damage possible, no specific backpressure required for operation Solvent stability 100% solvent stability Temperature range No limit High-capacity suppressor Low-capacity suppressor Application Universal use for gradient applications or with minimum dead isocratic applications with volume for applications very strong eluents with 2-mm columns Factory warranty 10 years *... Standard conditions for Metrosep A Supp 5-100/4.0: eluent 3.2 mmol/l Na 2 / 1.0 mmol/l NaH, flow rate 0.7 ml/min **... Standard conditions for Metrosep A Supp 10-100/2.0: eluent 5.0 mmol/l Na 2 / 5.0 mmol/l NaH, flow rate 0.2 ml/min Ordering information Suppressor rotors 6.2832.000 MSM Rotor A 6.2842.000 MSM-HC Rotor A 6.2844.000 MSM-LC Rotor A Accessories 6.2835.010 Connecting piece for MSM-HC and SPM 6.2842.020 Adapter for Suppressor Vario (always required if 6.2832.000 or 6.2844.000 is used in combination with 940 Professional IC Vario or 930 Compact IC Flex) Subject to modifications Layout by Ecknauer+Schoch ASW, printed in Switzerland at Metrohm AG, CH-9100 Herisau 8.000.5134EN 2015-08 Dosino Regeneration 2.800.0010 800 Dosino 6.5330.190 IC equipment: Dosino Regeneration www.metrohm.com