Electrochemical Techniques: Anodic Stripping Voltammetry

Similar documents
Electrochemical Techniques: Cyclic Voltammetry

Standard Operating Procedure. edaq Potentionstat. Miramar College Potentiostat. -Report by Marianne Samonte, Dec 2009

# 12 ph-titration of Strong Acids with Strong Bases

EXPERIMENT 16 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

Operating Instructions for Ammonium ISE Specifications

Air Quality 2 - Determination of Cd and Pb by Anodic Stripping Voltammetry

#13 ph-titration of Weak Acids with Strong Bases

Potentiometry: Ion-Selective Electrodes

Conductometric Titration & Gravimetric Determination of a Precipitate

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

Related concepts Electrolyte, electrical conductance, specific conductance, ion mobility, ion conductivity, conductometry, volumetry.

Acid-Base Titration. Evaluation copy

Heavy metal analysis of river water using the ItalSens HM1 sensor

Titration of a strong acid with a strong base with Cobra4

Acid-Base Titration. Sample

Chemical Reactions: The Copper Cycle

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Final Concentration 0 excess 0.1 M 0.1 M

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Unit 2 B Voltammetry and Polarography

STOICHIOMETRY AND THE CHEMICAL REACTION

Using Conductivity to Find an Equivalence Point

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

Ka Acid Dissociation Constant Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.1.16

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

GENERAL INSTRUCTIONS GENERAL PREPARATION

Determination of an Equilibrium Constant

NaOH (aq) + HCl (aq) NaCl (aq) + H 2 O (l)

#14 Determination of Equivalent Mass by Electrolysis

Chemistry 1B Experiment 14 65

Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask

#09 Investigating the Relationship between the Mass of a Liquid and its Volume Ken Lyle, St. John s School, Houston, TX

Conductimetric Titration and Gravimetric Determination of a Precipitate

Sodium Chloride - Analytical Standard

Electrochemistry. To use principles of electrochemistry to understand the properties of electrochemical cells and electrolysis.

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION

ph Titration Curves You will need about 10 ml of ~6M NaOH solution and about 40 ml of each acid solution and about 1.5 grams of KHP.

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Shown below is a sample titration curve for a diprotic acid. Note the two equivalence points.

Electrochemistry. Part I: Electrochemical Activity from Chemical Reactions. Part II. Electrochemical activity from cell potentials.

Measuring Enthalpy Changes and Gas Laws

symphony Multisensor Probes

Voltaic Cells. 100 ml graduated cylinder Emery cloth 150 ml beakers, 3 Salt bridge Voltmeter Wires with alligator clips, 2

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

Experimental Procedure

Chemistry 212 Lab, Spring Design of the Experiment: Standard and Non-Standard Reduction Potentials For Metal/Metal Ion Half-Cells

Exercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES

Experiment 8 Introduction to Volumetric Techniques I. Objectives

EXPERIMENT 6. Properties of Buffers INTRODUCTION

Determination of Zinc, Cadmium, Lead and Copper in Effluent Sample

Introduction to Spectroscopy: Analysis of Copper Ore

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Electrochemistry. Conductivity of strong and weak electrolytes LEC 06. What you need: What you can learn about. Principle and tasks

Electrochemical Cells

Acid-Base Titration Curves Using a ph Meter

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

ANALYSIS OF LEAD IN SEAWATER

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006

GENERAL INSTRUCTIONS GENERAL PREPARATION

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Synthesis of Benzoic Acid

Determination of Orthophosphate Ion

DIFFUSION THROUGH MEMBRANES STANDARDS B C.4 INTRODUCTION

Exercise 4-4. Titration of a Buffer Solution EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Buffer solutions

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

1) A ph/mv meter or an ion meter, either line operated or portable.

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Introduction to Spectroscopy: Analysis of Copper Ore

Cyclic Voltammetry. Objective: To learn the basics of cyclic voltammetry with a well-behaved echem system

Electrochemistry and Concentration Effects on Electrode Potentials Prelab

To determine relative oxidizing and reducing strengths of a series of metals and ions.

05 Reduction potentials: Micro-voltaic cells. An electrochemical cell uses an oxidation reduction (redox) reaction to produce electrical energy.

EXPERIMENT 9 SALTWATER CONDUCTANCE: The Effect of Concentration

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background

Determination of Chloride using Potentiometry

Experiment C-10 Titration of a Strong Acid and a Strong Base

Acid Rain. Evaluation copy

Acid-Base Titration Curves Using a ph Meter

Chemistry MultiMeasure Sensor PS-2170

CHEM 334 Quantitative Analysis Laboratory

ELECTROCHEMICAL CELLS

Apply the ideal gas law (PV = nrt) to experimentally determine the number of moles of carbon dioxide gas generated

Experiment 20-Acid-Base Titration: Standardization of KOH and Determination of the Molarity and/or Percent Composition of an Acid Solution

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

6 Acid Base Titration

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Determination of the K a of a Weak Acid and the K b of a Weak Base from ph Measurements

Canine brain natriuretic peptide,bnp ELISA Kit

The Determination of an Equilibrium Constant

Reaction Stoichiometry

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

ELECTROCHEMICAL CELLS

POGIL LAB EXERCISE 15 HOW DO YOU STANDARDIZE AN ACID AND BASE?

GRIGNARD REACTION Synthesis of Benzoic Acid

QUALITATIVE ANALYSIS OF GROUP I CATIONS

Thermodynamics and the Solubility of Sodium Tetraborate Decahydrate

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.

Exercise 4-3. Titration of Weak Acids EXERCISE OBJECTIVE DISCUSSION OUTLINE. The 5% rule DISCUSSION

Transcription:

Electrochemical Techniques: Anodic Stripping Voltammetry Quantitative Determination of Copper, Lead and Cadmium in Tap Water at a Thin Mercury Film Electrode. 1. Aim To use a standard additions method to determine the levels of Copper, Lead and Cadmium in a typical tap water sample by Anodic Stripping Voltammetry (ASV) at a thin mercury film electrode. 2. Background Anodic Stripping Voltammetry (ASV) is a very sensitive electroanalytical techniques where a reducing potential is initially applied to a working electrode which can immobilise the reduced analyte species onto its surface. Subsequently an oxidising potential sweep is applied to the electrode and the analyte is re-oxidised at its characteristic oxidation potential and the cathodic current generated is proportional to the amount of analyte initially deposited onto the electrode. In this experiment copper, lead and cadmium ions are reduced at the surface of a thin mercury film electrode plated onto a glassy carbon disk electrode. The metals produced are immobilised by forming an amalgam with the mercury. The sensitivity of this technique arises due to the pre-concentration of the analyte species prior to analysis. Figure 2.1 illustrates the processes occurring at the electrode surface during ASV. Prior to Hg deposition Hg deposition at 1300 mv Analyte pre-concentration at 900 mv Re-oxidation Figure 2.1 Formation of the thin mercury film electrode, pre-concentration and re-oxidation of analyte. Anodic Stripping Voltammetry Page 1 of 14 TEXP01_1509

3. Equipment Required Per Group: Computer with Echem software installed ER466 Integrated Potentiostat System, or an e-corder unit with EA161 Potentiostat From the ET014 EChem Electrode Kit: Magnetic stirrer Glassy Carbon voltammetric disk electrode Leakless Ag/AgCl Miniature Reference electrode Pt/Ti Auxiliary electrode 3 ml mini cells with stirrer bar Retort stand with 3 bosheads & clamps 1 ml auto-pipette and tips 500 ml beaker 3 x 20 ml beakers 4 x 10 ml beakers Nitrogen gas cylinder with a needle valve and hose terminating in Teflon tubing drawn down to a 1 mm OD at the tip 250 ml filter flask Tap fitted with a vacuum adapter 1 m of Teflon tubing with one end drawn down to 1 mm OD at the tip, and the other end inserted through the rubber bung that fits into the neck of the filter flask 100 ml volumetric flask 50 ml volumetric flask A wash bottle filled with milli-q water Per class: Heavy metal waste container Printer Parafilm to cover reagents in beakers Anodic Stripping Voltammetry Page 2 of 14 TEXP01_1509

3. Reagents Required Per Group: Deionised water (milli-q grade) 50 ml Chloride-Acetate Buffer Electrolyte (CLAC) 1 ml 1000 ppm Mercury Plating Solution (MPS) in CLAC 250 ml of Tap Water 1 ml each of 1000 ppm Copper, Lead and Cadmium standards 4. Safety Notes & Instrument Care Dispose of all solutions in the heavy metal waste container Wear gloves and protective eyewear throughout the experiment Avoid spilling liquids on the hardware by mixing reagents away from the units Do not touch the tip of the glassy carbon electrode. This will coat the electrode with an greasy film which prevents mercury from being deposited. 5. Procedure 5.1 Equipment Setup 5.1.1 Set up the potentiostat using the instructions in the manual on the edaq Software Installer CD. 5.1.2 Place the magnetic stirrer onto the base of the retort stand and attach the bossheads and clamps in the configuration shown in Figure 5.1. 5.1.3 Attach the filter flask to the clamp facing the rear of the setup and attach a length of rubber hose between the suction nozzle of the vacuum adapter and the nozzle of the flask. 5.1.4 Place the bung with protruding tubing into the neck of the filter flask. The tubing will be used to suck out solutions from the mini-cell when the tap fitted with the vacuum adapter is turned on. Only apply suction when solutions need to be removed from the cell. 5.1.5 Inspect the tip of the glassy carbon electrode (NB: avoid touching the tip with your fingers). The surface should be black and shiny. If it is dull and grey (from previous use), gently wipe the surface with a small piece of lint free tissue. Dispose of tissue in heavy metal waste container. 5.1.6 Place the stirrer bar into the mini cell and pipette 2 ml of CLAC and 40 µl of MPS into the cell. 5.1.7 Insert the glassy carbon working electrode into the largest hole in the lid of the mini cell. Adjust the rubber o-ring on the electrode so the electrode tip is immersed and Anodic Stripping Voltammetry Page 3 of 14 TEXP01_1509

there is a clearance of at least 3 mm between the tip and the stirrer bar. Insert the reference electrode into the medium sized hole and the auxiliary electrode into one of the smaller holes. Align the tips of the two electrodes with that of the working electrode. Tape the pin ends (not the pins though) of the electrode together to keep everything in place. 5.1.8 Place the mini cell onto the magnetic stirrer and secure with the lower clamp. Turn on the magnetic stirrer and set to a medium rate. Check that the electrode tips are still immersed whilst the magnetic stirrer is on. NB, DO NOT ADJUST THE RATE OF STIRRING DURING THE EXPERIMENT. TURN THE MAGNETIC STIRRER ON AND OFF USING A SWITCH OR AT THE POWER POINT. 5.1.9 Make sure that needle valve of the nitrogen supply is closed and insert the fine Teflon tube end of the gas line through the remaining hole in the lid. Adjust the cylinder regulator to give a pressure reading of not more than 0.5 bar. (0.1 bar ideal). Slowly open the needle valve, adjusting the flow to give a slow steady stream of gas bubbles. 5.1.10 Gently bubble nitrogen gas through the solution for approximately 1 minute, then adjust the nitrogen tube so nitrogen gas is still being pumped into the cell, but does not affect solution convection in the cell. 5.1.11 Connect the Red alligator clip from the potentiostat cable to the Auxiliary Electrode, the Green to the Working Electrode and the Yellow to the Reference Electrode. 5.1.12 Open EChem software. Anodic Stripping Voltammetry Page 4 of 14 TEXP01_1509

Vacuum Adapter Filter Flask Suction Tube Mini-Cell with Electrodes Nitrogen gas line Retort stand with clamps Magnetic Stirrer Figure 5.1 ASV Equipment Setup 5.2 Plating the Thin Mercury Film onto the Glassy Carbon Electrode 5.2.1 In EChem, set the current range of the potentiostat to 20 µa in the potentiostat dialog box. 5.2.2 Select Linear Stripping Voltammetry from the Technique menu and use the following parameters to plate the mercury film: Range Speed Deposition Potential Deposition Time Ramp Rate Start Potential Finish Potential Sample Width Rest Time ±2 V 10 khz -1300 mv 300 s 1000 mv/s -1300 mv 100 mv 0.5 ms 10 s Anodic Stripping Voltammetry Page 5 of 14 TEXP01_1509

5.2.3 Run the procedure SWITCHING THE MAGNETIC STIRRER OFF DURING THE REST AND STRIPPING STEPS. Label the resulting voltammogram mercury plating in the page comments. 5.3 Sample Collection and Preparation of Standard Additions 5.3.1 Collect approximately 250 ml of tap water. Each group should select different taps from around campus and make an estimate of the frequency of the taps use. 5.3.2 Using the tap water sample as the diluent, prepare a combined standard with an addition concentration of 5 ppm Cu, 1 ppm Pb and 1 ppm Cd in the 100 ml volumetric flask. For now assume that the concentration of these elements in the tap water is zero. 5.3.3 Make a second serial dilution with tap water to 250 ppb Cu, 50 ppb Pb and 50 ppb Cd, using the 50 ml volumetric flask (i.e. dilute 2.5 ml of the standard produced in 5.3.2 to 50 ml). This is the top standard to prepare other standards for analysis. 5.3.4 Using the top standard, prepare the following 5 standards in the 10 ml beakers, again using the tap water sample as the diluent. Addition Concentration [Cu] [Pb] [Cd] (ppb) (ppb) (ppb) Addition 1 50 10 10 Addition 2 100 20 20 Addition 3 150 30 30 Addition 4 200 40 40 Addition 5 250 50 50 5.4 Removing Solution from Mini-Cell and Cleaning 5.4.1 Make sure that the sink is not blocked and turn on the tap connected to the vacuum adapter, to apply suction to the filter flask. 5.4.2 Switch off the magnetic stirrer. 5.4.3 Leaving the electrodes in situ, remove the nitrogen gas line, and replace with the suction tube. Suck the entire contents of the cell into the filter flask. 5.4.4 Remove the suction tube and fill the cell with milli-q water using the wash bottle. 5.4.5 Repeat steps 5.4.3 to 5.4.4 a total of three times. 5.4.6 Pipette 2.5 ml of fresh CLAC into the empty mini-cell for the next run. 5.4.7 Switch magnetic stirrer back on. Anodic Stripping Voltammetry Page 6 of 14 TEXP01_1509

5.5 Running Tap Water Sample 5.5.1 Pipette 0.5 ml of the Tap Water sample into the cell with the 2.5 ml of CLAC. 5.5.2 Rinse the nitrogen gas line and repeat the de-gassing step outlined in 5.1.10. 5.5.3 In the Linear Stripping Voltammetry dialog box, select the following parameters: Range ±1 V Speed 10 khz Current range 20 µa Deposition Potential -900 mv Deposition Time 60 s Ramp Rate 1000 mv/s Start Potential -900 mv Finish Potential 50 mv Step Height 2 mv Step Width 2 ms Sample Period 0.5 ms Rest Time 10 s Cleaning Potential 50 mv Cleaning Time 2 s 5.5.4 Collect a total of three voltammograms for the Tap Water sample, labelling each page accordingly. If the measured current exceeds 20 µa during a run, increase the current range to a suitable value and repeat the run. 5.5.5 The voltammograms should look similar to Figure 5.2. The largest hump should be around 100 mv, which corresponds to copper. There may also be smaller humps at around 400 mv and 600 mv, which corresponds to lead and cadmium respectively. 5.5.6 When you are happy with your voltammograms for the Tap Water sample, clean the mini-cell using the procedure outlined in 5.4, and pipette 2.5 ml of fresh CLAC into the mini-cell. 5.6 Running Standard Additions 5.6.1 Repeat steps 5.5.2 to 5.5.6 substituting the tap water sample with 0.5 ml of Addition 1. 5.6.2 Repeat step 5.6.1 with the remaining Additions in turn. Anodic Stripping Voltammetry Page 7 of 14 TEXP01_1509

Figure 5.2 Typical Voltammogram Obtained 6. Analysis 6.1 Peak Areas 6.1.1 For integrating peak areas, change the display from I vs E to I vs t in the display pop-up menu on the bottom left hand corner of the main EChem window. See Figure 6.1. 6.1.2 Open the Data Pad window under the Windows menu. In the first column select Page Number from the Selection Information options. In the second column select Integral from the General Statistics options. 6.1.3 Drag the marker from the bottom left corner of the main EChem window, to where you think the copper peak begins. Select a second marker by selecting Ctrl before clicking in the empty marker box. Drag the second marker to the other side of the copper peak. The tie line between the two markers represents the baseline of the voltammogram. 6.1.4 Adjust the markers until you feel that the tie line is a good approximation of the baseline of the peak. Using the mouse select the region between the markers, See Figure 6.1. Anodic Stripping Voltammetry Page 8 of 14 TEXP01_1509

Figure 6.1 Selection of a Peak for Integrating 6.1.5 Select Add to Data Pad from the Windows menu, or press Ctrl+D to record the page number and the area between the baseline and the curve for the selection. 6.1.6 Annotate the data pad to record which element the peak belongs to and the repeat that was being performed. 6.1.7 Repeat 6.1.3 to 6.1.6 for all pages and for all peaks and record peak areas in Table 7.1. It may be difficult to identify peaks for lead and cadmium in the tap water sample, because these elements are usually less abundant in drinking water than copper. 6.1.8 For the tap water sample and each addition, work out the average peak area of each element for the three repeats performed and record in Table 7.1. 6.4 Standard Additions Calibration 6.4.1 Call the unknown concentration of Cu, Pb and Cd in the 0.5 ml tap water sample [CuTW], [PbTW], [CdTW]. Assuming the concentrations of these elements is zero in tap water, the concentrations of the additions would be equivalent to those listed in 5.3.4. For each element X: [Xactual] = [XTW] + [Xaddition] and Areameasured = AreaTW + Areaaddition Anodic Stripping Voltammetry Page 9 of 14 TEXP01_1509

6.4.3 In a spreadsheet application, plot Xaddition on the x axis and Areameasured on the y axis, and use least squares regression to fit a straight line through the points. Do this for each of the elements and report the R 2 value for each plot. 6.4.4 Determine the equations of the fitted lines in the form y = mx + b; where m = slope b = y-intercept 6.4.5 If the concentration of an analyte element is zero ([XTW]=0), the fitted line should pass through the origin. If the analyte element is present in the tap water sample, the fitted line in the corresponding plot is shifted up. 6.4.6 To calculate the concentration of the unknown, either; a. extrapolate the line fitted in 6.4.3 back to the x-axis. [XTW] is equal to the absolute value of the x-intercept. b. substitute y=0 into the equation determined in 6.4.4. [XTW] is equal to the absolute of x value when y=0. Anodic Stripping Voltammetry Page 10 of 14 TEXP01_1509

7. Results 7.1 Peak Areas Sample Cd ( µa.ms) Pb ( µa.ms) Cu ( µa.ms) Tap Water Run 1 Tap Water Run 2 Tap Water Run 3 Average= Addition 1 Run 1 Addition 1 Run 2 Addition 1 Run 3 Average= Addition 2 Run 1 Addition 2 Run 2 Addition 2 Run 3 Average= Addition 3 Run 1 Addition 3 Run 2 Addition 3 Run 3 Average= Addition 4 Run 1 Addition 4 Run 2 Addition 4 Run 3 Average= Addition 5 Run 1 Addition 5 Run 2 Addition 5 Run 3 Average= 7.2 Tap Water Concentrations [CuTW] [PbTW] [CdTW] Anodic Stripping Voltammetry Page 11 of 14 TEXP01_1509

8. Questions 8.1 What is the purpose of the reducing the analyte at the mercury film electrode? To pre-concentrate the analyte, which increases the sensitivity of the technique. 8.2 Why should the rate of stirring in the mini-cell not be altered throughout the experiment? The amount of analyte ions reduced at the mercury film electrode is dependant on both solution concentration, the period the reducing potential is applied and solution convection. Both the reduction period and the rate of convection need to remain constant so the signal due to changes in solution concentration is monitored. 8.3 What is the purpose of bubbling nitrogen gas through the solution? To remove oxygen, which can interfere with the redox processes occurring at the surface of the electrode. 8.4 List some of the advantages and disadvantages of using the standards additions method of calibration? The matrix of the standards are matched to the matrix of the sample and the magnitude of matrix interference is similar in both solutions. Matrix differences between samples and standards can often cause errors when normal calibration methods are used. The main disadvantage of using standard additions is that only one sample can be analyses per calibration run. 8.5 List some possible sources of error in the experiment? Contamination in the reagents, variability in defining the peaks between runs, degradation of the mercury film, cross contamination between solutions, measurement errors, etc. 8.6 Why is copper expected to be the largest peak in tap water? Most tap water runs through copper piping and copper is a more abundant element than lead and cadmium. 8.7 Do you think the method used in this experiment would be a suitable for the routine analysis of cadmium in drinking water? No. On most occasions, the amount of cadmium in drinking water would be below the detection limits of the techniques. Anodic Stripping Voltammetry Page 12 of 14 TEXP01_1509

9. Instructors Notes 9.1 Computer System and Software Required One computer is required per group. EChem software must be installed for this experiment. 9.2 Reagents Make sufficient Chloride-Acetate Buffer Electrolyte (CLAC) to supply at least 50 ml per group. An additional 100 ml is required to make the Mercury Plating solution. Batch Volume 500 ml 7.3 g of Sodium Chloride 1.35 g of Sodium Acetate 0.6 ml of Glacial Acetic Acid 500 ml milli-q water 1000 ppm Mercury Plating Solution (MPS) in CLAC Batch Volume 100 ml 135 mg of Mercuric Chloride 100 ml of CLAC It is best to supply each group with small vials of each of the 1000 ppm Cu, Pb and Cd standards to prevent contaminating expensive commercial standards. Only 1 ml of each is required per group. 9.3 Preparation of Glassware Prior to practical class, all glassware should be soaked overnight in 0.2% HCl and then rinsed with milli-q water. Air dry the glassware, as wiping may introduce contaminants. If beakers are to be left standing for some time, seal with parafilm. 9.4 Other Water Samples The experimental technique outline works well with sea water and other natural water samples, however the solutions may need to be filtered to remove sediment and treated with UV light to destroy organic material. These treatments may need to be applied to tap water samples depending on the quality of the supply. Anodic Stripping Voltammetry Page 13 of 14 TEXP01_1509

9.5 Hints and Tips It is not uncommon for newcomers to analytical electrochemistry to have teething difficulties with ASV experiments. The most common problems are: 1. The working electrode should be scrupulously clean before the experiment. Especially no oily deposits (eg fingerprints) should be left on the tip of the electrode as these stop the formation of an even mercury film. 2. The mercury film itself is actually a collection of small mercury droplets with an overall gray appearance. Inexperienced users often want to deposit more mercury to get a mirror like finish but this would require the deposition of more mercury than is required. Also excess mercury can fall from the electrode mid-experiment. 3. Control of solution ph is critical for most ASV experiments. Most transition metal ions are hydrolysed to an appreciable extent at ph 7. That is hexa-aquametal ions, [M(H2O)6] 2+, will convert to hydroxyl species eg [M(H2O)5(OH)] + as the ph is raised, eventually resulting in the formation of insoluble hydroxides and oxides. These processes interfere with the simple chemistry of the conversion of a metal ion in aqueous solution to a metal atom in the mercury phase in the ASV experiment. One of the first signs of trouble is a shift in the ASV peak position (and sometimes an unsymmetrical peak). In general ASV of simple metal dications should be done at around ph 4 or less. Even ph 5 would be too high for copper. Use of a narrow range indicator paper can be useful to determine if the ph has drifted out of range far. However the ph should not be so low as to cause dissolution of the substrate metal out of the mercury amalgam. For example while ph 2 would probably be OK for copper determination it would be too low for ASV of zinc. A ph of about 4 is usually a good compromise. 4. Another problem can be the temptation to use too high concentration of substrate ions. ASV is normally used to detect metal ion in the ppb ranges (or even ppt). If using ppm solutions, you normally would have use very short deposition times to avoid putting too much substrate into the mercury film - which leads to a mercury amalgam that no longer behaves ideally (eg changes in viscosity can occur). Again the first sign of trouble can be changes in peak position, misshapen and peaks even double-humped peaks. 5. Changes in peak position (especially drift in the one direction over several days) can indicate that the reference electrode is in need of regeneration or replacement. Reference electrodes do age! If you are using a reference electrode without a knowledge of its history or age then do not be surprised by strange peak positions. Anodic Stripping Voltammetry Page 14 of 14 TEXP01_1509