Spectroscopy: Introduction. Required reading Chapter 18 (pages ) Chapter 20 (pages )

Similar documents
Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

25 Instruments for Optical Spectrometry

R O Y G B V. Spin States. Outer Shell Electrons. Molecular Rotations. Inner Shell Electrons. Molecular Vibrations. Nuclear Transitions

Instrumental Analysis: Spectrophotometric Methods

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Analytical Spectroscopy Review

Lecture 0. NC State University

MOLECULAR AND ATOMIC SPECTROSCOPY

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I

Reflection = EM strikes a boundary between two media differing in η and bounces back

Ultraviolet-Visible and Infrared Spectrophotometry

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Outline of Recombinant DNA technology. Application of UV spectroscopy in recombinant DNA technology

Skoog Chapter 6 Introduction to Spectrometric Methods

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Ultraviolet-Visible and Infrared Spectrophotometry

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out.

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Electronic Excitation by UV/Vis Spectroscopy :

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Overview of Spectroscopy

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

Chem 310 rd. 3 Homework Set Answers

UNIT I COLORIMETER AND SPECTROPHOTOMETERS PART A

JABLONSKI DIAGRAM INTERACTIONS BETWEEN LIGHT AND MATTER LIGHT AS A WAVE LIGHT AS A PARTICLE 2/1/16. Photoelectric effect Absorp<on Emission ScaDering

Introduction to Spectroscopic methods

II. Spectrophotometry (Chapters 17, 19, 20)

Spectrophotometry. Introduction

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

Classification of spectroscopic methods

Chem 434 -Instrumental Analysis Hour Exam 1

Lecture 3: Light absorbance

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Chem 155 Midterm Exam Page 1 of 10 Spring 2010 Terrill

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision

Chem Homework Set Answers

Chapter 10. Spectroscopic Methods. An early example of a colorimetric analysis is Nessler s method for ammonia, which was.

World Journal of Pharmaceutical Research SJIF Impact Factor 8.074

Absorption spectrometry summary

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

Instrumental Chemical Analysis

Spectrochemical methods

Ch 313 FINAL EXAM OUTLINE Spring 2010

levels. The signal is either absorbance vibrational and rotational energy levels or percent transmittance of the analyte

Chapter 10. Spectroscopic Methods. An early example of a colorimetric analysis is Nessler s method for ammonia, which was.

Questions on Instrumental Methods of Analysis

Instrumental Chemical Analysis

CEE 772: Instrumental Methods in Environmental Analysis

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology

Course: M.Sc (Chemistry) Analytical Chemistry Unit: III

Photochemical principles

Chapter 4 Ultraviolet and visible spectroscopy Molecular Spectrophotometry

Chapter 5 Electrons In Atoms

Introduction to FT-IR Spectroscopy

Physics 30: Chapter 5 Exam Wave Nature of Light

Lab 10: Spectroscopy & the Hydrogen Atom Phy208 Fall 2008

EXPERIMENT 17: Atomic Emission

EMISSION SPECTROSCOPY

Chapter 17: Fundamentals of Spectrophotometry

Absorption photometry

Chapter 5 Light and Matter

Reference. What is spectroscopy? What is Light? / EMR 11/15/2015. Principles of Spectroscopy. Processes in Spectroscopy

Properties of Electromagnetic Radiation Chapter 5. What is light? What is a wave? Radiation carries information

高等食品分析 (Advanced Food Analysis) I. SPECTROSCOPIC METHODS *Instrumental methods: 1. Spectroscopic methods (spectroscopy): a) Electromagnetic radiation

JABLONSKI DIAGRAM 2/15/16

PAPER No. 12: ORGANIC SPECTROSCOPY MODULE No. 7: Instrumentation for IR spectroscopy

Lab 5: Spectroscopy & the Hydrogen Atom Phy248 Spring 2009

2101 Atomic Spectroscopy

The Fundamentals of Spectroscopy: Theory BUILDING BETTER SCIENCE AGILENT AND YOU

Analytical Chemistry II

Spectroscopy. a laboratory method of analyzing matter using electromagnetic radiation.

Ultraviolet-Visible Spectroscopy

Spectroscopy. a laboratory method of analyzing matter using electromagnetic radiation

n ( λ ) is observed. Further, the bandgap of the ZnTe semiconductor is

SPECTROPHOTOMETERS. Visible and UV-Visible

Spectroscopy Problem Set February 22, 2018

Atomization. In Flame Emission

Chemistry Instrumental Analysis Lecture 15. Chem 4631

APAS Laboratory { PAGE } Spectroscopy SPECTROSCOPY

The Spectrophotometer and Atomic Spectra of Hydrogen Physics 246

C101-E111. Talk Letter. Vol.2 February 2009

Infrared Spectroscopy: Identification of Unknown Substances

1 WHAT IS SPECTROSCOPY?

Chapter 17: Fundamentals of Spectrophotometry

INTERNATIONAL JOURNAL OF INSTITUTIONAL PHARMACY AND LIFE SCIENCES

The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer s Law.

Skoog Chapter 7 Components of Optical Instruments

UNIT 2 UV-VISIBLE SPECTROMETRY

Modern Instrumental Methods of Analysis Prof. J. R. Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore

MOLEBIO LAB #4: Using a Spectrophotometer

= (fundamental constants c 0, h, k ). (1) k

Sunlight. 1 radiation.

Experiment 3 Electromagnetic Radiation and Atom Interaction

Chapter 5 Electrons In Atoms

Transcription:

Spectroscopy: Introduction Required reading Chapter 18 (pages 378-397) Chapter 20 (pages 424-449)

Spectrophotometry is any procedure that uses light to measure chemical concentrations Properties of Light

electromagnetic radiation energy that moves by means of oscillating, perpendicular electric and magnetic fields. "Visible Light" is one small part of electromagnetic radiation. Wavelength (λ) distance between successive crests of a wave Frequency (ν) number of wave crests passing a point in space per unit time (second) one hertz is one cycle per second (1/s) The wavelength and frequency of light are related: λ x ν = c (c = speed of light in vacuum = 2.998 x 10 8 m/s)

Regions of the Electromagnetic Spectrum Frequency Wavelength Energy Name Uses 10 20 to 10 21 10-12 Nuclear Gamma Rays Cancer treatment 10 17 to 10 19 10-10 Electronic X rays imaging 10 15 to 10 16 10-7 Electronic Ultraviolet Black Lights 10 13 to 10 14 10-6 Electronic Visible Illumination 10 12 to 10 13 10-4 Vibrational Infrared Heating 10 9 to 10 11 10-2 Rotational Microwave Cooking 10 5 to 10 8 10 2 Radio Signal Transmission

photon individual bundles of energy; each photon has an energy that is directly proportional to the frequency: E photon = hn photon (h = 6.626 x 10-34 J s) Often, we like to express the frequency in terms of inverse wavelength, E = h n = h c/l = h c s where σ is called the wavenumber. We will use this in infrared spectrometry.

Light of various wavelengths or energies does different things with matter: The absorption of light by a molecule causes an increase in the energy of the molecule. This may be in terms of an increase in vibrational, rotational, electronic or other types of molecular energy states. Typically, in the UV-Vis region, we will disturb electronic states by the absorption of light in the range of 200-800 nm.

Excited State Ground State Absorption at l If a molecule absorbs energy from a photon, then an electron can be promoted to a higher energy level in that molecule. That electron originally "resided" in a molecular orbital of the molecule which is at a relatively low energy, we call that the GROUND STATE. The new place of residence for the excited electron is a higher energy molecular orbital. The excited energy level of the molecule has not only an electronic energy level associated with it, but also vibrational and rotational energy levels.

Basically, we have to dissipate the quantized energy in the different vibrational, rotational and electronic states of the molecule. Absorption at l } Vibrational States Excited Electronic State Relaxation Vibrational States Vibrational States } Ground Electronic State Time 1 Time 2 Vibrational States

So, the absorbed energy can be dissipated in a lot of different ways, as shown in the JABLONSKI DIAGRAM below:

Fluorescence occurs on a timescale of about 10-6 sec. Phosphorescence is much slower, about 1 to 10-6 sec. This is because the electron has undergone spin inversion. Remember the Pauli Exclusion Principle.

What if we have a sample holder known as a cuvette in the path of Electromagnetic radiation of some energy: l l l l l Lamp l Wavelength Selector Detector

If the incident light has an incident power, P o, and the exiting light has intensity, P, then the transmittance through the path length b (cm) of the sample is: T = P/P o (Range of 0 to 1) The absorbance at that wavelength is: A(λ) = -log(p/po) = -log(t)

If we do some calculus, we come up with the Beer-Lambert Law: A(l) = e l b C (dimensionless) Beer's Law will deviate at high concentrations of analyte. Solvent is chemically affected by high concentrations of analyte, thus impacting e The distribution of analyte components as a result of equilibria events is drastically affected by high analyte concentrations An absorption spectrum is a graph showing how A varies with λ. λ max Absorption Spectrum

Note: Any substance that absorbs visible light will appear colored when white light is transmitted through it or reflected from it. The substance absorbs certain wavelengths of the white light, and our eyes detect the wavelengths that are not absorbed. The observed color is called the complement of the absorbed color.

The Absorption Spectrophotometer Light from a continuous source is passed through a monochromator. This monochromatic light travels through a sample of pathlength b, and the radiant power of the emergent light is measured. Visible light generated from a quartz halogen lamp. UV light generated from a deuterium arc lamp. cuvette sample cell that has flat quartz faces. Quartz transmits both UV and visible light plastic or glass only transmits visible light.

Single-beam Spectrophotometer First measure Po for a reference cuvette containing pure solvent. Then replace reference cuvette (containing the blank) with cuvette containing sample to measure P. Block Diagram of Single-beam Spectrophotometer P 0 P Light source Monochromator Sample cell Detector Read out Inconvenient, somewhat inaccurate, poorly suited to continuous measurements

Double-beam Spectrophotometer Light passes alternately through the reference and sample cuvettes. A chopper is a mirror that rotates in and out of the light path diverting the light between the reference and sample cuvettes. Routine procedure is to first record a baseline spectrum with two reference cuvettes. The absorbance of the reference is then subtracted from the absorbance of the sample to obtain the "true" absorbance at each wavelength.

Double-beam Spectrophotometer

Precautions Most spectrophotometers operate best at values for A 0.4 0.9 If absorbance too high, intensity is hard to measure. If absorbance too low, hard to distinguish between reference and sample Slight mismatches between reference and sample cuvettes leads to systematic error Samples must be dust free. Cuvettes must be wiped clean before use.

Light Sources Any object that is heated emits radiation. Emission from real objects such as a tungsten filament light bulb emulate blackbody radiation (the emission is a continuous spectral distribution). Visible and infrared lamps as light sources approach blackbody radiators. The radiation from an object's surface expressed as power per unit area is the exitance (emittance), M. M = s T 4 Where s is the Stefan-Boltzmann constant ~ 5.7 X 10-8 W/(m 2 K 4 ) T = Temperature (K)

Spectral Distribution of Blackbody Radiation

Lamps for Absorption Spectrometers Lamps are broad-band, polychromatic light sources. Typically they are inexpensive and stable. i) Visible and Near Infrared: Tungsten Lamp ii) Ultraviolet and Visible: Quartz Halogen Lamp iii) Ultraviolet: Deuterium Arc Lamp iv) Infrared: nichrome wire

Spectral Distribution of Different Light Sources

Lasers Lasers provide ~ single l Very bright sources for spectroscopy Properties of Lasers: Monochromatic (only one wavelength) Collimated (emit in one direction) Polarized (only one electric field vector) Coherent (electric/magnetic fields in phase) expensive high maintenance

Laser Operation Principles

Operation of a Laser Population Inversion: A higher energy state has a greater population of electrons than a lower energy state Lifetimes of electrons in various energy levels: E2 >> E3 or E1 Two popular lasers: helium-neon laser ~ red light Laser diode ~ solid state, p-n junction ~ red, or near IR

Wavelength Selection Devices Prisms were used in older instruments, Quartz or salt crystals. Monochromators separate wavelengths of light; they consist of both entrance & exit slits, mirrors, and diffraction grating or refraction lens/prisms, and filters.

Grating Monochromators Polychromatic light is collimated (focused) into a beam of parallel rays by a concave mirror (monochromatic-one wavelength; polychromaticmany wavelengths). Rays strike the reflection grating (see next figure) and different wavelengths are diffracted (separated) at different angles. Diffracted light is focussed by a second concave mirror so that only one wavelength passes through the exit slit at a time.

Grating Equation: nλ= d(sinθ-sinφ) n = diffraction order (1 n) d = groove spacing = angle of reflection = angle of incidence

Components of a Grating 1) diffraction grating Spectrophotometer

A diffraction grating is ruled with a series of closely spaced parallel grooves separated by distance d. These are often constructed from aluminum metal and coated with a non oxidative coating applied. When light is reflected from the grating, each groove behaves as a source of light. When adjacent rays are in phase, they reinforce each other. When adjacent rays are out of phase, the partially or completely cancel each other. Thus can be aligned to allow only certain wavelengths to pass through.

2) Slits Slits are constructed by machining a sharp edge onto two metal pieces. These lie in a plane and the spacing between them, the slit width, can be adjusted. The smaller the slit width, the better the spectral resolution. 3) Filters: Filters are used to pass on only desired wavelengths of light. A filter could be colored glass. Most likely they are also based on constructive or destructive interference of light waves.

4) Interferometers - alternative wavelength selection process

Interferometers allows all wavelengths to simultaneously reach the detector Radiation from source reaches beam splitter, where half of the radiation hits the moving mirror and half hits the fixed mirror. The beams reflect and re-combine, the emerging radiation for a wavelength exhibits constructive or destructive interference.

Interferometers With constant mirror velocity, the wavelength modulates in a regular sinusoidal manner. Both the sampling rate of radiation reaching the detector and the mirror velocity is modulated by a helium-neon laser. The resulting detector signal typically is stored as a time domain spectrum (interferogram). Converted to a spectrum in the frequency domain using the mathematical process of Fourier Transform.

Fourier Transform: 1) allows for signal averaging 2) allows all wavelengths to be monitored simultaneously 3) mathematical process that converts data obtained in the time domain to be converted into the frequency domain.

Fourier Transform FT is used quite extensively in IR (infrared), mass spectroscopy and in NMR (nuclear magnetic resonance) applications. In addition to IR spectroscopy, the use of Fourier transform has become quite common in certain spectroscopic methods due to the affordability of PCs.

Types of Detectors A detector produces an electric signal when struck by photons. Phototube detector

Phototube Detector The phototube is used frequently as a detector in UV-Vis spectrometers. The cathode consists of a photo-emissive surface. Electrons are ejected from the cathode proportional to the radiant power (photons) striking its surface. The emitted electrons are attracted to the anode. The accompanying voltage is fed to an amplifier and converted to a signal.

Photomultiplier Tubes The Photo Multiplier Tube, (PMT) is similar to the photo tube, but is a vast improvement.

Photomultiplier Tubes In addition to the cathode and anode, the PMT has dynodes, which produce a cascade effect on the electron emission production. Each photon causes a ~ 10 7 additional electrons to be produced. The PMT possesses high sensitivity, good S/N ratio, and excellent dynamic range. PMTs are highly sensitive to visible and UV excitations at extremely low power conditions, (very low concentrations of analyte). Intense light sources (such as daylight or stray light) can destroy and damage PMTs.

PHOTO DIODE ARRAY: silicon diode a) Schematic cross-sectional view of photodiode array b) Picture of array with 1024 elements, each 25 µm wide and 2.5 mm high. The central black rectangle is the photosensitive area. The entire chip is 5 cm in length.

PHOTO DIODE ARRAY Silicon diode detectors, such as those used in a photo diode arrays (PDA), are composed of reverse-biased p-n junctions formed on a silicon chip. The doping of different elements in the silicon can produce electron rich and electron poor domains. As seen in (a) holes and electrons are randomly distributed. The addition of an outside power source (b) causes a depletion layer where conductance is zero. When a photon of light strikes the junction (c), holes and electrons form which provide the completion of the circuit. The current is proportional to the radiant power.

Photodiode Arrays PDAs are a series of silicon photo diodes, with each having a storage capacitor, and a switch that are combined in a integrated circuit on a silicon chip. The number of sensors (silicon photodiodes) in a PDA range from 64 to 4096.

Photodiode Arrays The slit width of the instrument allows the radiation to be dispersed over the entire array, allowing the spectral information to be accumulated simultaneously. PDAs are not as sensitive nor have the same S/N ratio as the PMT, but one gains the advantage of gathering multichannel information (all of spectrum collected simultaneously). Advantage of the PDA is recording the entire spectrum in a fraction of the time required for a conventional scanning spectrometer to scan one wavelength at a time. An example of PDA use is in atomic emission spectroscopy (AES), UV-vis spectrophotometry, fluorescence spectrometry, Raman spectrometry.

Errors in Spectrophotometry Choice of the Wavelength and Bandwidth

For Quantitative Analysis Absorbance should not be too large or small(~0.4-0.9 units) Absorbance should be measured at the lambda max providing no interference from other species at this lambda. Monochromator bandwidth should be as large as possible but small compared to the spectral bandwidth. Slit width to small allows less light to reach the analyte, smaller S/N ratio and precision. Too wide a monochromator band width (slit width) distorts the peak shapes of the spectrum.

As shown in the previous figure; the instrumental setting with the band width at 0.1 nm and the choice of 609 nm for l max would be desirable for the spectra shown. At the 2.0 nm setting there is no fine structure shown in this spectrum.

Stray Light, Electrical Noise, Cell Positioning Stray light arises from two major sources: 1) Misdirected rays coming from the monochromator. 2) Light coming from outside the instrument such as the sample compartment lid not closed properly.

Other concerns; the correct choice of the sample cell (does it require glass or quartz?) the alignment of the sample cell (and/or the sample cell holder) dust & fingerprints on the cell

Obtaining Quantitative Information from UVvis Absorption Spectrophotometry Beer's Law (aka Beer-Lambert Law) A = e X b X C where c is the concentration (M), b is the path length (cm), and e is the molar absorptivity (M 1 cm 1). e is wavelength dependent. Beer's Law works for dilute (~ 0.01 M) solutions in which the absorbing species is not participating in a concentration-dependent equilibrium.