J. K. Wang+ and C. S. Tsai

Similar documents
EFFECTS OF ACOUSTIC SCATTERING AT ROUGH SURFACES ON THE

Vector diffraction theory of refraction of light by a spherical surface

6th NDT in Progress Lamb waves in an anisotropic plate of a single crystal silicon wafer

ACOUSTIC TRANSMISSION WITH MODE CONVERSION PHENOMENON

MEASUREMENT OF REFLECTANCE FUNCTION FOR LAYERED STRUCTURES USING FOCUSED ACOUSTIC WAVES INTRODUCTION

Analysis of second-harmonic generation microscopy under refractive index mismatch

Modeling microlenses by use of vectorial field rays and diffraction integrals

THE ULTRASONIC FIELD OF FOCUSED TRANSDUCERS THROUGH A LIQUID- M. EI Amrani Intercontrole 13 rue du Capricome, Rungis Cedex, France

High-Resolution. Transmission. Electron Microscopy

Nondestructive Determination of Elastic Constants of Thin Plates Based on PVDF Focusing Ultrasound Transducers and Lamb Wave Measurements

Spatial Frequency and Transfer Function. columns of atoms, where the electrostatic potential is higher than in vacuum

gives rise to multitude of four-wave-mixing phenomena which are of great

DETERMINATION OF ELASTIC CONSTANTS OF ANISOTROPIC MATERIALS FROM OBLIQUE

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Lamb Wave Behavior in Bridge Girder Geometries

AP5301/ Name the major parts of an optical microscope and state their functions.

SUBSURFACE WAVES IN SOLIDS WITH CURVED SURFACE AND ACOUSTICAL IMPEDANCE ON IT

ESTIMATION OF LAYER THICKNESS BY THE COST FUNCTION OPTIMIZATION: PHANTOM STUDY

ULTRASONIC INSPECTION, MATERIAL NOISE AND. Mehmet Bilgen and James H. Center for NDE Iowa State University Ames, IA 50011

Laser Optics-II. ME 677: Laser Material Processing Instructor: Ramesh Singh 1

5. LIGHT MICROSCOPY Abbe s theory of imaging

Engineering Physics 1 Prof. G.D. Vermaa Department of Physics Indian Institute of Technology-Roorkee

C-) ISUMMARY REPORT LINEAR AND NONLINEAR ULTRASONIC INTERACTIONS ON LIQUID-SOLID BOUNDARIES

Canalization of Sub-wavelength Images by Electromagnetic Crystals

C 44 (SAM) (AMS) 3,520±170 m/s 3,240±180 m/s g/cm 3 S 2. 18±2 GPa

Ultrasonic reflection and material evaluation

Nature of Light Part 2

An alternative method to specify the degree of resonator stability

of polymer-composite-, 8) metal- 9) and multilayeredplates 10) were also achieved by the normal incidence of 1. Introduction

Transmission Electron Microscopy

Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown and Twiss Photon Antibunching Setup

Lamb Waves in Plate Girder Geometries

Weak-Beam Dark-Field Technique

Digital Holographic Measurement of Nanometric Optical Excitation on Soft Matter by Optical Pressure and Photothermal Interactions

Two-Dimensional simulation of thermal blooming effects in ring pattern laser beam propagating into absorbing CO2 gas

1532 J. Acoust. Soc. Am. 102 (3), September /97/102(3)/1532/8/$ Acoustical Society of America 1532

A Single-Beam, Ponderomotive-Optical Trap for Energetic Free Electrons

Single Emitter Detection with Fluorescence and Extinction Spectroscopy

OPTI 511L Fall A. Demonstrate frequency doubling of a YAG laser (1064 nm -> 532 nm).

Phys102 Lecture Diffraction of Light

On the Interpretation of Phase-contrast Images

Astronomy 203 practice final examination

Shear waves in solid-state materials

v. K. Kinra, Y. Wang and C. Zhu Center for Mechanics of Composites Department of Aerospace Engineering Texas A&M University College Station, TX 77843

Version 087 EX4 ditmire (58335) 1

Phased Array Inspection at Elevated Temperatures

2.71. Final examination. 3 hours (9am 12 noon) Total pages: 7 (seven) PLEASE DO NOT TURN OVER UNTIL EXAM STARTS PLEASE RETURN THIS BOOKLET

A small object is placed a distance 2.0 cm from a thin convex lens. The focal length of the lens is 5.0 cm.

6th NDT in Progress Evaluation of metal weld-zone by using Poisson s ratio distribution

APPLICATION-DIRECTED MODELING OF RADIATION AND PROPAGATION OF ELASTIC WAVES IN ANISOTROPIC MEDIA: GPSS AND OPOSSM

SIMULATION OF THE INSPECTION OF PLANAR NON MAGNETIC MATERIALS WITH ELECTRO MAGNETIC ACOUSTIC TRANSDUCERS

Remarkable properties of Lamb modes in plates

A model for the ultrasonic field radiated by an Electro-Magnetic Acoustic Transducer in a ferromagnetic solid

Transducer design simulation using finite element method

Signal Loss. A1 A L[Neper] = ln or L[dB] = 20log 1. Proportional loss of signal amplitude with increasing propagation distance: = α d

SIMULATION OF THE INSPECTION OF PLANAR NON MAGNETIC MATERIALS WITH ELECTRO MAGNETIC ACOUSTIC TRANSDUCERS

Lab 2: Mach Zender Interferometer Overview

Addition of a Lens Before the Second Focus of a Prolate-Spheroidal IRA

On the study of elastic wave scattering and Rayleigh wave velocity measurement of concrete with steel bar

G. Chang, R.B. Givens, J.W.M. Spicer, and J. C. Murphy

Physics 3312 Lecture 7 February 6, 2019

MEMS Metrology. Prof. Tianhong Cui ME 8254

Phase Retrieval for the Hubble Space Telescope and other Applications Abstract: Introduction: Theory:

Polarization control and sensing with two-dimensional coupled photonic crystal microcavity arrays. Hatice Altug * and Jelena Vučković

Scattering of light from quasi-homogeneous sources by quasi-homogeneous media

Paper V. Acoustic Radiation Losses in Busbars. J. Meltaus, S. S. Hong, and V. P. Plessky J. Meltaus, S. S. Hong, V. P. Plessky.

Einstein Classes, Unit No. 102, 103, Vardhman Ring Road Plaza, Vikas Puri Extn., Outer Ring Road New Delhi , Ph. : ,

Time reversal reconstruction of finite sized sources in elastic media

Citation. J. Mod. Opt. 60(3), (2013). 1. M.-S. Kim, A. C. Assafrao, T. Scharf, C. Rockstuhl, S. F. Pereira, H. P. Urbach, H. P.

CHEM 681 Seminar Mingqi Zhao April 20, 1998 Room 2104, 4:00 p.m. High Resolution Transmission Electron Microscopy: theories and applications

IMAGE CONTRAST FORMED BY SCATTERED X-RAYS

( PART : B DESCRIPTIVE )

Modeling Focused Beam Propagation in a Scattering Medium. Janaka Ranasinghesagara

IR SPECTROSCOPY FOR BONDING SURFACE CONTAMINATION CHARACTERIZATION

A COMPARISON OF DIFFERENT METHODS FOR THE DETECTION OF A WEAK ADHESIVE/ADHEREND INTERFACE IN BONDED JOINTS

PROBLEM OF X-RAY SYNCHROTRON BEAM COLLIMATION BY ZONE PLATE

Ambiguity of optical coherence tomography measurements due to rough surface scattering

Laboratory 3: Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown, and Twiss Setup for Photon Antibunching

Graduate School of Engineering, Kyoto University, Kyoto daigaku-katsura, Nishikyo-ku, Kyoto, Japan.

n The visual examination of the image of a point source is one of the most basic and important tests that can be performed.

EFIT SIMULATIONS FOR ULTRASONIC NDE

ECE 185 ELECTRO-OPTIC MODULATION OF LIGHT

Optics.

Optical Tweezers. BGGN 266, Biophysics Lab. June, Trygve Bakken & Adam Koerner

EFFICIENT MODELING OF FINITE ACOUSTIC BEAM EXCITATION AND DETECTION OF INTERFACE AND BULK WAVES ON PLANAR AND CYLINDRICAL

Optimisation using measured Green s function for improving spatial coherence in acoustic measurements

ELASTIC MODULI OF SILICON CARBIDE PARTICULATE REINFORCED ALUMINUM METAL MATRIX COMPOSITES

Determination of absorption coefficient by digital holographic measurement of optical excitation

PRINCIPLES OF PHYSICAL OPTICS

ANALYSIS OF FREQUENCY CHARACTERISTICS ON NON-INVASIVE ULTRASONIC-DOPPLER FLOW MEASUREMENT FOR METAL PIPES

1. Consider the biconvex thick lens shown in the figure below, made from transparent material with index n and thickness L.

APPLICATION OF SCANNING ACOUSTIC MICROSCOPY TO RESIDUAL STRESS

Topic 4: Waves 4.3 Wave characteristics

PHY410 Optics Exam #3

Electromagnetic Theory for Microwaves and Optoelectronics

ULTRASONIC INSPECTION OF TITANIUM ALLOYS WITH A TIME REVERSAL

Efficiency analysis of diffractive lenses

EXEMPLARY PROBLEMS APPENDIX B CHAPTER 1

SURFACE-DISPLACEMENT IMAGING USING OPTICAL BEAM DEFLECTION. S.E. McBride* and G.C. Wetsel, Jr.* Southern Methodist University Dallas, Texas 75275

Phase-Shifting Interferometry and Combined Methods

Transcription:

TRANSMISSION SCANNING ACOUSTIC MICROSCOPY FOR TILTED PLATE SPECIMENS* J. K. Wang+ and C. S. Tsai Department of Electrical Engineering University of California. Irvine. CA 92717 I. INTRODUCTION In transmission scanning acoustic microscopy (SAM) [1.2.3] the plate-specimen under examination is usually aligned with its surface normal parallel to the lens axis in order to optimize the spatial resolution. However. we have found that in many instances a higher acoustic transmission and additional image contents can be facilitated by tilting the specimen such that its surface normal is at a small angle with respect to the lens axis (Fig. 1). For a small tilt angle. the image degradation due to reduced spatial resolution is insignificant with relatively small numerical aperture lenses. It is thus transmitting transducer receiving transducer Fig. 1 Geometry of a 150 MHz Transmission SAM Used to Measure Angular Power Spectra of Plate Specimens *This work was supported in part by the MICRO Project of the University of California and the Industrial Matching Fund from Rockwell In terna t ional. +Presently with Industrial Electronics Group Technology Center. Hughes Aircraft Company. 253

254 J. K. WANG AND C. S. TSAI desirable to study plate specimens under this particular mode of operation. It is to be noted that reflection scanning acoustic microscopy that utilizes oblique incidence of the acoustic wave has been reported [4]. In this paper, the basic mechanisms of these ang1edependent acoustic transmission and image contrasts are studied through an angular-spectrum technique [4-9]. The analysis to be presented shows that acoustic resonance, mode conversion and the numerical aperture of the confocal lenses are the dominant factors in determining the transmission contrast of a plate specimen. Application of this ti1tspecimen mode of operation to acoustic imaging of isotropic and anisotropic materials is also addressed. Acoustic resonance at normal incidence was recently utilized by us for characterization of defects in thick specimens using a 150 MHz transmission SAM [10]. For a collimated acoustic beam resonance occurs at several incidence angles that depend upon the specimen thickness [11]. For a thin specimen this resonance transmission may occur at an incidence angle which is greater than the shear-wave critical angle [11, 12]. For a focused acoustic beam as in SAM, the angle-dependent acoustic transmission as a function of the numerical aperture of the acoustic lens is analyzed. Another dominant factor that affects the transmission angular spectrum concern mode conversion which refers to transfer of the incident 10gitudinal waves into the shear waves in the specimen. The calculated transmission angular spectrum in the presence of mode conversion is compared with that obtained without mode conversion. The relative contribution of each kind of wave to the image contast is thus determined as a function of the incidence angle and the numerical aperm ture of the acoustic lenses. II. THEORETICAL FORMULATION This section presents an analysis for calculating the induced voltage at the receiving transducer as a function of the tilt angle of the specimen and other relevant parameters in a confocal transmission scanning acoustic microscope. For a confocal transmission scanning acoustic microscope, with unti1ted specimen, Wickramasinghes' theory [8] can be applied to show that: V ~ ~ P O ( r ) P 5 ( r ) u O ( r ) U 5 ( r ) T ( K r / K ) d r folio rpo (r)p 5 (r) U o (r) U 5 (r)t(sin1/l)dr (1) V: Transducer output voltage r: radial distance from the lens axis pupil functions of the transmitting and receiving lenses, respectively

TRANSMISSION SCANNING ACOUSTIC MICROSCOPY 255 acoustic fields on the planes 0 and 5, respectively, in the absence of the specimen K: wave vector of a Fourier (plane-wave) component of the focused acoustic beam K r : the component of K on the focal plane K /K = r/f = sinp [7] r P: the angle between the lens axis and the propagation direction of a plane-wave component f: focal length of the two identical confocal lenses and T: the acoustic transmission coefficient of the water-specimenwater system. It is to be noted that measurement of T for an aluminum plate in water using transmission probes was previously reported. [13]. Now for a specimen tilted at an angle 8 as shown in Fig. 1, Eq. (1) is modified to yield: where V(sin8) = t {fooorpo(r)ps(r)uo(r)us(r)[t(sin l) + T(sin 2)]1 dr (2) l = 8 + sin-l(r/f) and 2 = 8 - sin-l(r/f). The reason for separating the transmission coefficient T into two parts is that one half of the cross section of the focused acoustic beam is incident on the specimen at angles l where 8 ~ l 8 ~ TI/2 (radian), while the other half is at angles 2 where 8 - T I / 2 ~ 28. ~ For a collimated acoustic beam incident on a lossy plate specimen which is immersed in a liquid, the transmission coefficient is given by [11]: where T(8) = 2N/[2M + j(n 2 - M2 + 1)] (3) M= (Z2/Zl)cOS 2 28 S cot P + (Z2S/Zl)sin 2 28 S cot Q N = (Z2/Zl)cOS 2 28 S /sin P + (Z2S/ Zl)sin 2 28s/sin Q Z 1 = Plcl/cos8, Z2 = P2cL/cos8L, Z2S = P2cS/cos0S (4) P - ~ d Q - KSd and

256 J. K. WANG AND C. S. TSAI PI' P2: Mass densities of water and the specimen, respectively c l : longitudinal wave velocity in water c L, c S : longitudinal and shear wave velocities in specimen 8: incidence angle (from water) 0 L, 8. S refraction angles in specimen for the longitudinal and shear waves d: specimen thickness, and complex wave vectors in thickness direction of the specimen for longitudinal and shear waves (imaginary part accounts for the acoustic attenuation coefficient). By solving Eq. (3) as a function of the incidence angle 8, we can determine the angles at which resonance transmission occurs. Eq. (3) shows implicitly that the larger the specimen thickness, the higher the number of resonance transmission. However, as the specimen thickness increases the peaks of the resonance transmissions become weaker and are eventually reduced to a small value due to increased acoustic attenuations in the specimen. Such resonance transmission affects the tranducer voltage not only through enhanc8ment of acoustic transmission, but also through sharp phase changes (180 ) in the transmitted planewave components such as the calculated example shown in Figure 2. As a result, when the specimen is tilted to one of the resonance angles in a transmission SAM, some of the incident acoustic Fourier components will incur sharp phase changes while others will not. As a result, destructive interference occurs among these two portions of the spherical wavefront. The degree of destructive interference is a sensitive function of local material composition (or elastic parameters) and irregularity in the front surface of the specimen. In other words, the 320 O r - - - - - - - ~ - - - - - - - - ~ - - - - _ ~ ~ g> 240 :g, Q) UI IU 160 ~ a.. aluminum 80 Fig. 2 O ~ - - ~ ~ - L ~ - - - - ~ ~ - - ~ ~ o 3 6 9 12 Angle of Incidence (degree) Calculated Phase Changes in the Plane Acoustic Waves Transmitted Through an Aluminum Plate o! 0.156 rom in Thickness (Frequency Thickness Product = 2.34XIO meter/second)

TRANSMISSION SCANNING ACOUSTIC MICROSCOPY 257 image contrast of the specimen should be enhanced accordingly. Experimental verification of this contrast enhancement is given in Section III. It should be noted that in reflection SAM such sharp phase changes also occur in the leaky Rayleigh wave radiated when an acoustic wave impinges upon a specimen at the Rayleigh wave critical angle [4]. Equations (2)-(4) form the basis for generation of the numerical solutions for the output voltage of the transmission SAM with a tilted specimen. The merit of the above formulation is that the product P O (r)p 5 (r)u O (r)u 5 (r) in the integrand can be viewed as a generalized pupil function to include the effects of lens geometry, nonuniform acoustic field in the beam cross section, acoustic attentuation in water, and spherical aberration, etc. [9]. These effects are included in generating the numerical results to be given in Section III. Finally, it should be noted that we have implicitly neglected the effect of the specimen thickness on the acoustic fields in Eq. (1). In general, the acoustic fields on both sides of a plate specimen cannot be assumed to be the same without introducing some error. However, for the case involving confocal acoustic lenses of relatively small numerical aperture and long focal length such as the ones employed in this study, this error does not drastically affect the results. This observation is in agreement with the conclusion that for a specimen thickness less than the depth of field of the beam (or range resolution) plane-wave approximation results in only a small phase error in the transmitted acoustic beam [14]. III. NUMERICAL AND EXPERIMENTAL RESULTS We have calculated the angular power spectra of plate specimens in a confocal transmission SAM at 150 MHz using Eq. (4)-(7). For example, Fig. 3(a) and 3(b) show the calculated results for an aluminum plate of 254 ~ m in thickness with the numerical aperture of the acoustic lenses as a parameter. It is seen that both the number and the intensity of the enhanced transmission angles decrease as the numerical aperture of the acoustic lenses is increased. This is accounted for by the fact that destructive interference among the Fourier components of varying propagation directions increases with the numerical aperture of the incident acoustic beam. To verify the aforementioned theoretical predictions an experiment was carried out using an aluminum plate of the same thickness as that used in obtaining Fig. 3. The experimental setup as shown in Fig. 1 is the same as that employed in the conventional transmission SAM [1-3] except that a precision rotation jig with an accuracy of one-fourth of a degree is used to rotate the specimen plane relative to the focal plane. The numerical aperture of the acoustic lenses used for the experimentis 0.156. Figure 4 shows that angular spectra measured. By comparing Fig. 3(b) with Fig. 4, we see that the agreement between theory and experiment is quite satisfactory and the theory correctly predicts the incidence angle for maximum acoustic transmission, namely, 6.3 0 IV. CONTRAST ANALYSIS It is also desirable to tilt the specimen by a small angle such that the angular contrast, which is defined as the slope of the angular

258 J. K. WANG AND C. S. TSAI 0.7 Fig. 3, ' -o.s -0.4,! 0.4 O.S sine Calculated Transmission Angular Spectra of an Aluminum Plate of 254 ~ m Thick for 150 MHz Converging Acoustic Beams of Different Numerical Aperture: (a) N.A. = 0.031, (b) N.A. = 0.156 power spectrum (dv/d8), is a maximum. In other words, a small incident angle is searched to maximize the following expression: dv dv @ = cos8 dsin6 = t c o s ~ ~ P O ( r ) P 5 ( r [ ) c Uo O s ( ~ lr T ) ' U+ ( 5 sc( ior n s ) ~ l 2( ) T s' i n ~ dr 2 ) ] 1 where T' ( s i n refers ~ i ) to differentiating T(sinc/li) lvith respect to its argument. Note that dv/d0 is approximately equal to dv/dsin0 for a small tilt angle 0 for the specimen. Near the angle of maximum angular contrast any small variation in the incidence angle due to local r o u g h ~ ness or curvature in the surface of the specimen would result in a large variation in the transmitted acoustic power. In other words, the angular contrast consitutes an important contrast mechanism in the transmission SAM. From Eq. (5), we see that it is a function of the lens pupil function, and the orientation plus the acoustic parameters of the specimen. An appropriate incident angle for the aluminum specimen studied was found to be 5 0 ~ (sing 6 ~ 0.1) as can be seen from Fig. 4. Acoustic transmission images of the same aluminum plate as described in Section III, but with lossy adhesive layer residing on a great portion of the surface of the plate, were used to verify this o b s e ~ a t i o n. The adhesive layer was of the commercial product Loctite 420 and was approximately 0.5 mil thick. Figures 5 (a)-(c) show the image obtained. It should be mentioned that the boundary of the adhesive layer (partly shown) coincides with that of the dark area in Fig. 5. By first comparing the bright areas in these images, we see that the acoustic transmission is considerably higher at the incidence angle of 5 0 ~ 6. Furthermore, additional image contents are seen in the portion of the image that corresponds to the adhesive area. Thus, it confirms the prediction that by examining the specimen at both normal incidence and oblique incidence of suitable angle a higher degree of image detail can be achieved. (5)

TRANSMISSION SCANNING ACOUSTIC MICROSCOPY 259 0.3 Transducer Voltage 0.2 0.5 sine Fig. 4 Measured Angular Spectrum of an Aluminum Plate of 254 ~ m Thick Using Acoustic Lenses With N.A. = 0.156 6 Fig. 5 Transmission Acoustic Images of Surface Adhesive Area (Dark) on an Aluminum Plate of 254 ~ Thick m as a Function of Specimen Tilt Angle 0: (a) 0=0 ; (b) 0=3 ; and (c) 0=6 The theoretical study presented thus far can be extended to treat an anisotropic specimen. Since the acoustic transmission coefficient for anisotropic materials depend on not only the tilt angle but also the orientation of the specimen surface with respect to the crystal axis. the enhanced image contrast discussed above should be even more readily observed. We have seen in Figs. 3(a) and 3(b) that acoustic resonance greatly affects the transmission angular spectra and image contrast. Let us now examine the case in which acoustic resonance is unimportant in the angular range of concern such as an aluminum membrane of 3 microns in thickness. The calculated angular spectra based on Eq. (2) show that the normalized transducer v o 1 t is ~ g0.37. e independent of the incident angle for both collimated beam and focused beam. with the latter being formed by the acoustic lens with a numerical aperture equal to 0.156. This suggests that there exists little angular image contrast when acoustic resonance is insignificant because dv/d8 is near zero. In a related study on image contrast with thin aluminum specimens (one-quarter to one acoustic wavelength in thickness) [15]. we have however experimentally observed that the transmission image contrast was drastically altered if the numerical aperture of the acoustic lens is large enough

260 J. K. WANG AND C. S. TSAI to excite a leaky lamb wave. It should be noted that in this situation the model that includes only longitudinal and shear waves would not be adequate. We shall now point out a major difference between transmission and reflection SAM's in terms of contrast mechanisms. First, we recall that in the conventional reflection SAM the transducer voltage [7] is given as follows: f co 2 2 2 V(Z) = 0 r[uo(r)] PO(r)P'O(r)R(r/f) exp[-j(koz/f )r ]dr (6) where Z refers to the location of the front surface of the specimen along the lens axis with the acoustic focus located at the origin of the coordinate system. Any nonuniformity or roughness on the specimen surface will correspond to variation in the Z value, and therefore V(Z). The image contrast as a result of such variation in V(Z) can be defined as the axial contrast or Z-contrast and mathematically given as follows: 2 () (co 2 J 2 2 1 KOr)d (7) ~ ~ Z = J O r[uo(r)] PO(r)PO (r)r(r/f) I exp[-j(koz/f )r ] J (-j -;z- r Secondly, due to excitation of leaky Rayleigh waves, contrast reversal may also occur as the speciman is translated along the lens axis [3,16]. However, in the transmission SAM the Z-contrast as defined above is often insignificant because there is no dependence on Z in Eq. (1). Rather, the primary variation in the image as the specimen is translated along the lens axis is due to the defocusing effect. Our measurements with acoustically thick specimens (thickness of many acoustic wavelengths) also support this observation. Consequently, we may conclude that the angular contrast (dv/d8) due to acoustic resonance constitutes the dominant mechanism for material signature in a transmission SAM for acoustically thick specimens. We have also found experimentally that contrary to thick specimens both the angular contrast and the axial contrast are important in transmission SAM for thin specimens with a thickness in the order of one acoustic wavelength or smaller [15]. In this case it is believed that the axial contrast is caused by the interference between the transmitted longitudinal wave and the leaky Lamb wave [11]. The strength of the longitudinal wave is represented by the factor T ( s i in n ~ Eq. ) (1). There is one interesting problem concerning the role that shear wave plays in acoustic imaging using the transmission SAM [17]. The shear wave in the specimen is generated from the obliquely incident longitudinal wave through mode conversion, and the strength of the shear wave generated increases with the incident angle. The transmission angular spectra for the same aluminum specimen as described in Section III with and without mode conversion were calculated using Eq. (2), and shown in Fig. 6(a) and Fig. 6(b). respectively. The latter case was obtained simply by letting cs in Eq. (4) equal zero. The resultant acoustic transmission coefficient is as follows: T (0) = o '1 + ZL) sinp 2 (8)

TRANSMISSION SCANNING ACOUSTIC MICROSCOPY 261 0.2 0.1 T r a n s d u c e r V o l t a g e ~ 9, 9t -0:5-0.4-0.3-0.2-0.1 0 0.1 0.2 0.3 0.4 ( 0.2 ~ sin9 a ) Fig. 6, t, 0 0.1 9, 9t -0.5-0.4-0.3-0.2-0.1 0 0.1 0.2 0.3 0.4 (b) sample tilt angle sin9 Effect of Mode Conversion on the Transmission Angular Spectrum of an Aluminum Plate of 254 ~ m Thick: (a) With Mode Conversion (b) Without Mode Conversion t From these plots we first observe that for a small tilt angle (sing < 0.2) mode conversion only slightly affects the transmitted acoustic power-or the angular contrast of the image. This is so because only a small fraction of the incident longitudinal wave is converted into the shear wave. Next, as the tilt angle is increased beyond the longitudinalwave critical angle ( s i n0.24), 0 ~ ~ the mode conversion greatly affects both the transmitted acoustic power and the angular contrast of the image. Fig. 6(b) indicates that the transducer voltage rapidly diminishes as sing is increased beyond 0.37. The difference between sin-l 0.37 and 8 ~ is approximately equal to the half angle of the acoustic lens. All Fourier components of the focused acoustic beam propagate at angles larger than 8 ~ when the acoustic beam is incident upon the specimen at the angle sin- l 0.37. Finally, at a tilt angle far beyond the shear wave critical angle 8 t,namely, sing >0.45 in Fig. 6 the shear wave due to mode conversion becomes the only source which contributes to formation of the acoustic image. It is clear that the transmission angular spectrum can be used to quantitatively determine the differences that mode conversion brings to the transmitted acoustic power and the angular image contrast in a transmission SAM. V. CONCLUSION We have studied both theoretically and experimentally the transmission angular spectra and the image contrast of a tilted plate specimen using transmission scanning acoustic microscope. The transmitted acoustic power and the intrinsic image contrast as a function of the tilt angle of the specimen are studied in detail. Detailed study on an aluminum plate specimen was carried out using a 150 MHz transmission SAM that employs confocal lenses of relatively small numerical aperture and long focal length. A good agreement between

262 J. K. WANG AND C. S. TSAI the theoretical predictions and the experimental results has been obtained. It is concluded that the angular contrast which originates from acoustic resonance and mode conversion constitutes the dominant contrast mechanism in a transmission SAM for acoustically thick specimens. Specifically, by orienting the specimen to a small tilt angle significant enhancement in transmitted acoustic power and/or angular image contrast due to acoustic resonance can be achieved with only slight degradation in spatial resolution. It is also found that, for a small tilt angle, the effect of mode conversion on the acoustic transmission and the image contrast is negligible for a SAM that utilizes relatively small numerical aperture lenses. The analytical results presented should be applicable to a transmission SAM operating at frequencies higher than 150 MHz. REFERENCES 1. R. A. Lemons and C. F. Quate, App1. Phys. Lett., Vol. 24, pp. 163-166, February 1974. 2. C. S. Tsai, S. K. Wang, and C. C. Lee, App1. Phys. Lett., Vol. 31, pp. 317-320, September 1977. 3. C. F. Quate, A. Ata1ar, and H. K. Wickramasinghe, Proc. IEEE, Vol. 67, pp. 1092-1114, 1979. 4. C. E. Yeack and M. Chodorow, J. App1. Phys., Vol. 51, pp. 4637-4644, September 1980. 5. D. C. Champeney, Academic Press, New York, 1973. 6. M. McGuirk and C. K. Carnig1ia, J. Opt. Soc. Am., Vol. ~ pp., 103-107, January 1977. 7. 8. A. Ata1ar, J. App1. Phys., Vol. 49, pp. 5130-5139, October, 1978. H. K. Wickramasinghe, J. App1. PfiYs., Vol. 50, pp. 664-668, February, 1979. 9. J. K. Wang and C. S. Tsai, Journal of Applied Physics, Vol. 55, pp. 80-88 (Jan. 1, 1984). 10. J. K. Wang, C. S. Tsai, and C. C. Lee, Proc. Rank Prize Fund Conference on Scanned Image Microscopy, Ed. by E. A. Ash, Academic Press, New York, pp. 137-147, 1980. 11. L. M. Brekhovskikh, "Waves in Layered Media," First Ed., Academic 12. A. 13. I. 14. S. 15. J. 16. R. Press, New York, 1960, p. 68. Schoch, Ergeb. exakt. Naturw., Vol. ~ p., 127, 1950. R. Smith, D. A. Sinclair, and H. K. Wickramasinghe, 1981 IEEE Ultrasonics Sym. Proc., pp. 591-596, IEEE Cat. No. 81CH1689-9. D. Bennett, IEEE Trans. Sonics and Ultrasonics, Vol. SU-29. pp. 316-320, November, 1982. -- K. Wang, Ph.D. Thesis, Carnegie-Mellon University, January, 1983. D. Weg1ein, App1. Phys. Lett., Vol. 34, pp. 179-181, February, 1979. -- 17. J. B. Jipson, App1. Phys. Lett., Vol. 35(5), pp. 385-387, September, 1979.