Application Note CBRNE Detecting Chemical Warfare Agents using the µraid

Similar documents
MM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE

NIGAS. i o n. N B C D e t e c t. The Neutron Induced Gamma Spectrometer

Compendium of Chemical Warfare Agents

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

DETECTING VX METABOLITES IN A HUMAN VICTIM KEL DANIEL MARSHALL UNIVERSITY FORENSIC SCIENCE PROGRAM APRIL 11, 2014

Chemistry Module Quiz for Science and Technology for NPTS Spring 2017

OPTIMIZING THE PERFORMANCE OF AUTOMATED MONITORING SYSTEMS FOR AIRBORNE TOXIC CHEMICALS. Dr. Gary D. Sides August 5, 2014

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Chemical Warfare Agents. Hanna Hunziker, Céline Prange Analy:cal Strategy, 13. October 2015

Organisation for the Prohibition of Chemical Weapons. Inspectorate

Improvised Chemical Devices, Chemical Security and the CWC

The RAID-M had the highly sensitive TDI library in Library A of the instrument as shown in Table 1. Table 1: Highly sensitive TDI library in ppb.

(U) Indicators for Terrorist Use of Toxic Industrial Chemicals

Chemical Munitions in the Baltic Sea the Lithuanian experience

ENGINEERING AND CONSTRUCTION BULLETIN

SU photoresist and SU-8 Developer (1-methoxy-2-propanol acetate) were

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority:

TITLE: Hazmat Recognition and Identification

Hazard Communication for Dangerous and Harmful Materials

Laboratory Safety and Fundamental Equipment Post-Assessment Exam Page 1 of 9

RADIATION SAFETY GUIDELINES FOR NON-USERS

HAZCOM - Training

7/8/2013. What is GHS?

The OPCW Verification Regime

(FM 8-285, NAVMED P AFM

LABORATORY CHEMICAL HYGIENE PLAN

Technical Inquiry Sarin/GB Overview

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Detection of Chemical Warfare Agents by Transportable GC/MS

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

Series 2. Continuous monitoring of trace chemical vapours

PUBLIC EMPLOYEE HAZARDOUS CHEMICAL PROTECTION AND RIGHT TO KNOW ACT O.C.G.A

The Globally Harmonized System of Classification and Labelling of Chemicals (GHS)

Chemical Health and Safety General Program

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry

The Globally Harmonized System (GHS) for Hazard Classification and Labelling. Development of a Worldwide System for Hazard Communication

Hazard Communication Standard: Safety Data Sheets

DETECTING AND MEASURING CHEMICAL WARFARE AGENTS IN REAL TIME USING THE TRACE ATMOSPHERIC GAS ANALYZER (TAGA)

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Mobile Mass Spectrometer in daily use Experiences made by non-scientific operators

Detection and identification of chemical warfare agents in environmental samples. Martin Köhler, Mario Francesco Mirabelli

Real Time On-Site Odor and VOC Emission Measurements Using a znose

Hazard Communication Right-to-Know Program

Science and Technology. Chemical Weapons Detection

"GHS CONTAINER LABELING"

Storing, using and disposing of unsealed radioactive substances in a Type C Laboratory: Extract of regulatory requirements

3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures. A. Laboratory Specific Information and Signatures

C h a p t e r 5 : W o r k p l a c e H a z a r d o u s M a t e r i a l s I n f o r m a t i o n S y s t e m ( W H M I S )

TargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

Guidelines for the Labelling of Chemical Products

Industrial Hygiene Report

Chemical Safety. Peter Yeung D.G. Manager. Safety Office The University of Hong Kong

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

HAZARD COMMUNICATION PROGRAM

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR

SAMPLE PAGES. Hazard Communication Program. [Company name]

HALF LIFE. NJSP HMRU June 10, Student Handout CBRNE AWARENESS Module 4 1. Objectives. Student will

Needs and Challenges in the Field Detection and Identification of Military Chemical Threats

Controlled Substances: TSCA the Toxic Substances Control Act

Radioactive Waste Policies

Product Stewardship Summary

Product Description and Specification

WHAT S WRONG WITH THIS PICTURE?

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

APPLICATION FOR AUTHORIZATION

A. Identify the highly penetrating radioactive emission that exposed the photographic plates.

REPORT OF THE SCIENTIFIC ADVISORY BOARD ON DEVELOPMENTS IN SCIENCE AND TECHNOLOGY FOR THE THIRD REVIEW CONFERENCE REMARKS BY THE CHAIR

HAZARD COMMUNICATION PROGRAM 2015 UPDATE GHS AWARENESS TRAINING CSULB COE

Proven robustness for large-scale metabolomics studies using the Phenomics Workhorse

Facilities Management

Radiation Emergencies -Preparing for Response-

Ion Mobility Spectrometer (IMS) Standalone IMS with integrated micro-computer and color display

Technical know-how in thermal analysis measurement

Introduction to IH Analytical Chemistry

Introduction and Initial Field Data of a Novel, Portable Nanoparticle Sizing Instrument

Mercury Vapour Analysers: Finding the Right Fit for Your Needs

Safety in the Chemistry Laboratory

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

SOUTH AFRICAN NATIONAL STANDARD

Global Harmonization and Hazard Communication

The Claviature of Gas Analysis

Arizona Division of Occupational Safety and Health Administration. 800 W. Washington Street, Phoenix, AZ Consultation:

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

SECTION 1. Identification of the substance/mixture and of the company/undertaking. Low Density Polyethylene LF2207M

UNIVERSITY OF WISCONSIN-RIVER FALLS HAZARD COMMUNICATION PROGRAM Table of Contents

TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL

Nitrogen, Total Inorganic

The Globally Harmonized System (GHS) for Hazard Classification and Labeling. Development of a Worldwide System for Hazard Communication

Collective Protection 2005

Multi-residue analysis of pesticides by GC-HRMS

Evaluating Leaks in a Soil Gas Sample Train

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

DRAIN TRACING DYE RED E123

HazCom and Global Harmonization Are You In Compliance? BISC June, 2018

Method for the determination of 1,3-butadiene

( V o l a t i l i t y )

Ionizing Radiation Awareness (Non-User)

MATERIAL SAFETY DATA SHEET

Transcription:

Application Note CBRNE 1825475 Detecting Chemical Warfare Agents using the µraid Abstract The personal protection of military personnel against chemical warfare agents (CWAs) and toxic industrial chemicals (TICs) is essential and can only be achieved using reliable, rapid and sensitive detection equipment. The micro Rapid Alarm and Identification Device (μraid) is a personal detector that has been designed to detect low concentrations (ppb range) of CWAs and TICs rapidly. It provides direct identification of an agent and quantifies the concentration range. Authors Franziska Lange, Wolfgang Heller, Thomas Elßner Bruker Daltonik GmbH, Leipzig, Germany Keywords Instrumentation and Software Introduction CWAs are classed as weapons of mass destruction and use the toxic properties of the chemicals to disable, injure or kill personnel. The chemicals are classified by their effects on human health into discrete groups; nerve, blister, choking and blood agents. Typical examples and their toxicological characteristics are listed in Table 1. A period of disarmament started by the end of the cold war; around 1990. The Chemical Weapons Convention (CWC) is the most recent arms control agreement. It outlaws the production, stockpiling and use of chemical weapons with the force of international law. The compliance of the CWC is supervised by the Organization for the Prohibition of Chemical Weapons (OPCW), who were a Nobel Peace Prize winner in 2013. 189 nations signed the convention, covering Chemical Warfare Agents Personal detection Ion Mobility Spectrometry µraid XIMS plus

approximately 98 % of the global population [1]. Despite this convention there have still been proven cases of CWAs being used in recent years; for example, in the Iraq War in Abu Sayda in 2007 [2] and in the Syrian civil war around Damascus in 2013 [3]. These examples serve to prove the continued need for military and emergency response personnel to be protected against such threats. Ion mobility spectrometry (IMS) is well suited for the gas phase detection and identification of hazardous chemicals. Due to the documented advantages of speed and sensitivity, the technique is highly appropriate in personal detection devices where a rapid response is essential. In this Application Note we describe the use of a novel instrument applicable for personal chemical detection in the field. Experimental Material CWA standards were kindly provided by the VPO CZ s.p., division VTÚO Brno. The purity of the material was 89.2 to 99.3 % according to the provider. Laboratory test For the determination of detection limits, response and recovery times, analyses were performed on the compact µraid personal detector in laboratory tests. For this purpose two standard instruments were used in parallel and connected to the GCWA_2011, a lab type gas device that produces constant and defined agent concentrations at adjustable humidity. The air inlet used bespoke gas tubes connected to the GCWA_2011. The experimental setup is shown in Figure 1. All of the tubing that was used for this experiment was made of Teflon as it is chemically inert and the risk of vapour adsorption on to the tube walls is extremely low. Experimental setup internal drawing pump µraid to fume µraid internal drawing pump open ending regulation of humidity inert teflon tube Tenax tube for quantitative reference analysis closed ending injection by a nanoliter valve storage container for the agent in gaseous state Figure 1: Schematic view of the setup of a gaseous agent supply at defined concentrations (GCWA_2011 by VPO CZ) and detection by two devices in parallel. filtered air The µraid is equipped with its own integrated sampling pump that is rated at approximately 20 L h -1. Analyses were carried out at ambient temperature (24 25 C; 33 C for VX) and ambient pressure (994 1007 mbar). The humidity in the gas flow was between 12 and 20 % for G agents, 22 and 80 % for HD, respectively. Quantitative references for CWAs The quantitative reference for the agent concentration was taken by sampling the agent containing air from the generator to a Tenax sampling tube (see Figure 1) followed by an analysis by the Ellman enzymatic method [4] and/or gas chromatography pulsed flame photometric detector (GC-PFPD; for organophosphorus compounds), by thymolphthalein (T-135) method [5] and/or GC-PFPD (sulphur mustard) or by gas chromatography mass spectrometry (GC-MS, nitrogen mustard). Tenax tubes were extracted with 2-propanol. Software The Bruker IMS software XIMS plus was used for detailed analyses of the experiments. Field test To show the usability and robustness of the instrument under realistic conditions supplementary outdoor tests were performed in addition to the laboratory test. Field tests were done at the Nevada Test Site in the United States in an open air test region. Chlorine (Cl 2 ) was used as CWA test agent and released at IDLH level (Immediately Dangerous to Life or Health, 10 ppm) from a storage cylinder. The purity was stated to be > 99 %. To simulate possible interferences a reference type smoke was released in addition to the test sample from a storage cylinder as well. Results Detectable agents The detection capability of the µraid includes a clear identification of an agent and not only an agent group (e.g. nerve, blister). Identification is achieved by the detection of corresponding agent peaks in the spectra measured in the positive and negative mode. An example IMS spectrum of Soman (GD), exhibiting a monomer dimer system recorded in the positive mode, is shown in Figure 2. The ratio of the resulting peak areas changes dependent on the agent concentration. At a two peak system (monomer and dimer) the peak area of the monomer decreases and the peak area of the dimer increases with a rising agent input. In general, identification based on a dimer system leads to more reliable detection results.

Table 1: Detectable nerve, blister and blood agents and associated toxicological properties and determined alarm limits Selected detectable agents Agent class Abbr. Agent name 1 LCt 50 [mg min m -3 ] [6, 7, 8] AEGL 3, 2 10 min [ppb] Alarm limit [mg m - ³] Alarm limit [ppb] Nerve GA Tabun 135 110 GB Sarin 70 64 GD Soman 70 25 GF Cyclosarin 50 53 VX VX 30 3 VXR Russian VX n.d. 3 n. d. Blister and Blood HD Sulphur Mustard 1,500 590 HN Nitrogen Mustard 1,500 311 L Lewisite 1,200 453 AC Hydrogen Cyanide 5,000 27,000 0.05 7 0.05 9 0.05 7 0.05 7 0.05 5 0.05 5 0.06 9 0.03 4 0.25 30 1 900 1 LCt 50 lethal concentration time (The concentration at which a gas kills 50 % of the exposed population) 2 AEGL 3 Acute Exposure Guideline Levels (The airborne concentration of a substance above which it is predicted that the general population, including susceptible individuals, could experience life-threatening health effects or death) 3 n. d.... no data IMS spectrum of Soman + Reactant Ion Peak GD Monomer Figure 2: IMS spectrum of GD (80 µg m -3 ) detected by the µraid and depicted using the software XIMS plus. At K 0 = 2.39 cm² Vs -1 the reactant ion peak (RIP) is shown. The characteristic agent peaks are present at 1.36 cm² Vs -1 (monomer) and 0.99 cm² Vs -1 (dimer) in the positive mode. GD Dimer A selection of the CWAs that the μraid can detect is shown in Table 1. This list is not exhaustive; further substances can be programmed into the integral libraries. In general, detection results are displayed by means of symbols and bar graphs (see Figure 3, left). Up to two detected agents can be displayed simultaneously by bars related to a specific agent concentration range. Additionally, the abbreviated names of the detected agents are given. In the right hand part of the result field a GHS hazard pictogram (Globally Harmonized System of Classification and Labelling of Chemicals) for toxic agents is displayed. Depending on the substance class the concentration bars are colour coded (red = nerve agent, yellow = blister agent, cyan = TICs, magenta = pulmonary and blood agents). If the response signal does not correspond to any substance in the library so-called pollution level indicators display a substance unspecific contamination level. If the pollution level or the increase rate of the pollution level indicator exceeds a predefined threshold level, a pollution alarm is given (cf. Figure 3). Additionally, a warning symbol appears. The purpose of this pollution indicator is to warn the operator that other substances are present that are not contained in the detection library. If a chemical agent is detected the alarm for this agent will be displayed as a priority; despite the presence of a pollution alarm.

Display readings of the µraid Concentration bar(s) GHS symbol of detected agent(s) Warning symbol Abbreviation(s) of detected agent(s) Pollution indicators in positive and negative mode Figure 3: Left: display reading at an alarm of the agents GD and L; right: display reading at a pollution alarm. Response and recovery times Response and recovery times are important parameters that show the true capability of the detection device. Response time is defined as the time elapsed between presence of the substance to be analyzed and the response of the detection system. Recovery time is the amount of time the device needs to get back to a clean status after the sample input has ceased. Low values for both response times and recovery times are essential for effective detection in the field. The aim of a personal detector is to provide an immediate alert and then be ready for the next detection as rapidly as possible. In Figure 4 examples of the response (t a ) and the recovery time (t b ) for GD are depicted. In Table 2 response times for several agents are summarized. For all agents the response time is within seconds. The response to the monomer is faster than to the dimer. Exemplarily a monomer of GD at 80 µg m -3 can be detected at 23 seconds, the dimer will be detected at 39 seconds. Recovery times are in the range of seconds to a few minutes, depending on the agent and concentration applied; this is substance dependent because different chemical agents exhibit different physical properties. Response and recovery time of Soman 0.80 0.70 0.60 51 µg m - ³ 80 µg m - ³ 280 µg m - ³ 0.50 0.40 0.30 0.20 0.10 0.00 Figure 4: Relative dimer peak area of GD (K 0 = 0.99 cm² Vs -1 ) during the agent challenge at three selected concentrations. Response time (t a ) and recovery time (t b ) are shown for an agent concentration of 80 µg m -3.

Table 2: Determined response times for CWAs using the µraid. Test site Response times Agent Test concentration [µg m - ³] Response time [s] GA 57 26 GB 97 21 GD 80 39 GF 64 50 VX 84 8 HD 104 5 HN 29 23 L 380 39 Figure 5: Outdoor testing at the Nevada National Security Site, United States. Alarm limits Table 1 details the alarm limits determined for the CWAs based on a dimer system obtained with the experimental setup described in Figure 1. All alarm limits were in the ppb range and are significantly below the AEGL 3, 10 min values (see also Table 1) except for VX. Due to the very low response time of only 8 seconds for VX the detection of this agent in the relevant range can be ensured nevertheless. Due to the fact that the determination of the alarm limits relies on the detectability of a monomer and a dimer, the real detection limits of the µraid, based on only monomer identification, are significantly below the alarm limits depicted in Table 1. When very low agent concentrations are expected, a special library based on a monomer identification can be used, albeit that the reliability of the identification result is lowered. Considering the LCt 50 and AEGL 3 values (Table 1) the alarm limits and response times determined demonstrate the extreme suitability of the µraid as an early warning tool for CWAs in the field. Field Test A field test was conducted to show the capabilities of the µraid in real conditions. Four instruments were used in parallel arranged to cover almost all areas in a test field. Because of the high toxicity of the aforementioned CWAs the less toxic but quite dangerous TIC Chlorine was released at a level of 10 ppm (AEGL 3, 10 min... 50 ppm). A picture of the test site is shown in Figure 5. Figure 6 illustrates the test setup and the obtained results. In interference tests (smoke) three out of four instruments gave very good detection results (75 to 100 %). Nevertheless, instrument #3803 gave lower detection alarms than the other instruments (67 and 38 %) and it was concluded that this was due to the position of the detector in the test area. Instrument #3803 was placed somewhat off to the side of the generated cloud. It was concluded that for instrument #3803 not all gas clouds reached its inlet. Conclusion The µraid is an extremely effective personal chemical detector and can provide essential threat detection in the field. Detection limits are in the ppb level and response times are in the region of a few seconds. Ease of use coupled with a low maintenance requirement make the µraid highly suitable for all users in combating the threat from CWAs and TICs. Acknowledgements Measurements of CWAs were done with the kind support of VPO CZ, division VTÚO Brno. Measurements of TICs in field tests at the Nevada National Security Site were done with the kind support of the U.S. Department of Homeland Security, Science and Technology Directorate, Chemical and Biological Defense Devision.

Bruker Daltonics is continually improving its products and reserves the right to change specifications without notice. Bruker Daltonics 12-2013, CBRNE, 1825475 Field test setup and obtained results Release and detection of Cl 2 without interferences Release and detection of Cl 2 with smoke interferences 20 m µraid #3810 9 of 9 detections 20 m µraid #3810 7 of 8 detections µraid #3802 8 of 9 detections µraid #3802 8 of 8 detections 10 m 10 m µraid #3798 9 of 9 detections Smoke Generator µraid #3798 6 of 8 detections Trailer Chemical Transfer Line Release Point (1 meter AGL) 10 m 20 m µraid #3803 6 of 9 detections Trailer Chemical Transfer Line Release Point (1 meter AGL) 10 m 20 m µraid #3803 3 of 8 detections Figure 6: Field test using four instruments, Cl 2 release, left: intake without interferences, right: intake with smoke interferences (AGL above ground level). References [1] Federal Office of Economics and Export Control, Germany, www.bafa.de [2] A. Richardt, F. Sabath; CBRN Protection: Managing the Threat of Chemical, Biological, Radioactive and Nuclear Weapons; Andre Richardt, Birgit Hülseweh, Bernd Niemeyer, Frank Sabath (Eds.); Wiley, 2012; page 11 [3] United Nations Mission to Investigate Allegations of the Use of Chemical Weapons in the Syrian Arab Republic, Report on the Alleged Use of Chemical Weapons in the Ghouta Area of Damascus on 21 August 2013 [4] G. L. Ellman (1959); Tissue sulfhydryl groups; Arch. Biochem. Biophys. 82 (1): 70 7 [5] F. M. Issa, H. Z. Youssef & R. M. Issa (1975); Spectrometric determination of bis(2-chloroethyl) sulfide (sulk mustard HD) and tris(chloroethyl) amine (nitrogen mustard HN) using thymolphthalein; Egypt J Chem. 18 (2): 257-64 [6] Federation of American Scientists, www.fas.org [7] Federal Department of Defence, Civil Protection and Sport, Switzerland, www.labor-spiez.ch [8] www.armystudyguide.com [9] United States Environmental Protection Agency, www.epa.gov Bruker Detection Division of Bruker Daltonik GmbH Leipzig Germany Phone +49 (341) 2431-30 Fax +49 (341) 2431-404 detection@bruker.com www.bruker.com Bruker Detection Corp. Billerica, MA USA Phone +1 (978) 663-3660 Fax +1 (978) 667-5993 detection@bruker.com