Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1

Similar documents
The design and operational theory of ph electrodes is a very complex subject, explored only briefly here. What is important to understand is that thes

The calibration of Hygrometer (LectureandTraining )

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth

#13 Electrochemical Cells

ELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY

Measuring State Parameters of the Atmosphere

CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Ordinary Level. PHYSICS 5054/02 Paper 2 Theory October/November 2003

Design and Development of Semi Cylindrical Capacitive Type Humidity Sensor


D501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect

Measuring State Parameters of the Atmosphere

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015

5mm Silicon PIN Photodiode, T-1 3/4 PD333-3B/H0/L2

SPECIFICATIONS INFRARED LAMP LED

3mm Photodiode PD204-6B

Fig. Electrochemical Cell/ Potentiometric Titration

Analysis of cations and anions by Ion- Selective Electrodes (ISEs)

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

KITUI WEST SUB-COUNTIES KCSE REVISION MOCK EXAMS 2015 PHYSICS PAPER 2

Q1. Why does the conductivity of a solution decrease with dilution?

MEASUREMENT OF ph: INTRODUCTION TO BUFFER AND BASICS OF ph Meter

and another with a peak frequency ω 2

: INSTRUMENTATION AND PROCESS CONTROL COURSE CODE : 6071 COURSE CATEGORY : A PERIODS/ WEEK : 5 PERIODS/ SEMESTER : 75 CREDIT : 5 TIME SCHEDULE

Reference Electrode Salt Bridge Kit

Elizabethtown College Department of Physics and Engineering PHY104

Electrochemical Cell

Lab #14: Electrochemical Cells

1. Mark the correct statement(s)

3URGXFWLQIRUPDWLRQ 5HVHUYDWLRQRISURSHUW\ULJKWV

Prof. Katherine Candler! E80 - Spring 2013!!!!! (Notes adapted from Prof. Qimin Yang s lecture, Spring 2011)!

Chapter 18 Electrochemistry. Electrochemical Cells

Chapter 24 Photonics Question 1 Question 2 Question 3 Question 4 Question 5

Technology Brief 9: Capacitive Sensors

Buffer Preparation. Learning Objectives:

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

Carbonized Electrospun Nanofiber Sheets for Thermophones

ORP Probe. Mini ORP. Reads. +/-2000mV. Range SMA. Connector. 95% in 1s. Response time 100 PSI. Max pressure. 60m (197 ft) Max depth 1 80 C

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

ORP Probe ORP. Reads. +/-2000mV. Range. 95% in 1s. Response time 100 PSI. Max pressure. 60m (197 ft) Max depth 1 80 C. Temperature range C.

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background

Technology Document Version

Technical Data Sheet. Pb Free. Specification GR101 SSC. Customer. Rev. 02 January 서식번호 : SSC- QP (Rev.0.

Unit - 3 ELECTROCHEMISTRY VSA QUESTIONS (1 - MARK QUESTIONS) 3. Mention the purpose of salt-bridge placed between two half-cells of a galvanic cell?

Stage 4 G-03: CONDUCTIVITY OF SOLUTIONS CP: USP. May 2015 INTRODUCTION

Electrochemistry. Michael Faraday s law of electromagnetic induction says that whenever a conductor is

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Advanced Level

2. (a) Differentiate between rare metal thermocouples and base metal thermocouples.

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

SUPPLEMENTARY INFORMATION

Contents. 2. Fluids. 1. Introduction

General Medicine 2016/17

Electro Analytical Methods

Experimentation. Third Edition. Wheeler. Ahmad R. Ganji. School of Engineering. With Third Edition contributions by. V. V. Krishnan. Brian S.

The Design and Development of a Low Cost Wood Moisture Sensor

Designing Information Devices and Systems I Discussion 8B

PRACTICAL 3 ph AND BUFFERS

Specification LR521. Drawn Approval Approval. Rev. 05 December 서식번호 : SSC-QP (Rev.00)

Chapter 3 Electrochemical methods of Analysis-Potentiometry

Titration of a strong acid with a strong base with Cobra4

possesses negative potential & undergoes oxidation preferably act as ANODE

FOCUS: Water Pollkon Prevention and Control

Humidity Calibration Solutions

Technical Data Sheet 3ψ Infrared Piranha

SAMPLE PAPER-05 (solved) PHYSICS (Theory) Class XII. Time allowed: 3 hours Marks: 70

The original definition of ph, is the negative base 10 logarithm of the hydrogen ion concentration, or: ph = - log 10 [H + ]

VALLIAMMAI ENGINEERING COLLEGE

Capacitance Level Measurement Electronic Inserts EC 37 Z, EC 47 Z

ph Electrode basics ph Electrodes Model 19 B /1 =1 m 24 B /1 =1 m

SPECIFICATIONS BLUE OVAL LAMP LED

(Refer Slide Time: 02:27)

CBSE_2014_SET_3 Physics

b. Which bulb is brightest? Justify your answer.

Chapter 17 Electrochemistry

LAMP UTW/S400-X9/T10(WS)

Dr. Anand Gupta

ph Probes Lab vs. Process Care, Calibrations and Comparison Presented by: Michael Blackburn Technical Sales Specialist at Hach Company

5mm Infrared LED HIR383C/L289

Lecture 18 Capacitance and Conductance

FIRST PUBLIC EXAMINATION SUBJECT 3: PHYSICAL CHEMISTRY

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Advanced Level

Humidity Calibration Solutions

Understanding ph Troubleshooting and Diagnostic information

CHAPTER 4 THERMAL CONDUCTIVITY AND VISCOSITY MEASUREMENTS

AP Physics C Electricity & Magnetism Mid Term Review

Part 2. Sensor and Transducer Instrument Selection Criteria (3 Hour)

Specification LR530 SSC CUSTOMER. Rev. 07 January acriche.com. Document No. : SSC-QP (Rev.00)

Chapter 6. Fiber Optic Thermometer. Ho Suk Ryou

electrons out of, or ionize, material in their paths as they pass. Such radiation is known as

15. Compare the result with the value you have taken above Compare the calculated pressure value with the actual pressure value that you have

THM80X Series Feature Application. Temperature & Humidity / Flow Measuring Specialist

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

Coimisiún na Scrúduithe Stáit State Examinations Commission

Specification LR530. Drawn Approval Approval. Rev. 08 December Document No. : SSC-QP (Rev.00)

Sensors and Transducers. mywbut.com

GB/T / IEC 60093:1980

INSTRUMENTATION ECE Fourth Semester. Presented By:- Sumit Grover Lect., Deptt. of ECE

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Amount of Aluminium required to get 56 Kg of Iron = 27 Kg Amount of Aluminium required to get 1120 Kg of Iron = = 30 o C

Soil Water Content & Soil Water Potential

Transcription:

Module 2 Measurement Systems Version 2 EE IIT, Kharagpur 1

Lesson 8 Measurement of Level, Humidity and ph Version 2 EE IIT, Kharagpur 2

Instructional Objectives At the end of this lesson, the student will be able to: 1. Name different methods for level and moisture measurements 2. Explain the basic techniques of level and humidity measurement 3. Explain the principle of ph measurement 4. Explain the necessity of using special measuring circuit for ph measurement 1. Introduction Level, humidity and ph are three important process parameters and their measurement find wide application in chemical and manufacturing industries. In this chapter we would provide a brief overview of the different techniques adopted for measurement of liquid level and humidity. The basic principle of ph measurement and the construction of ph electrodes are explained in section 4. 2. Level Measurement There are several instances where we need to monitor the liquid level in vessels. In some cases the problem is simple, we need to monitor the water level of a tank; a simple float type mechanism will suffice. But in some cases, the vessel may be sealed and the liquid a combustible one; as a result, the monitoring process becomes more complex. Depending upon the complexity of the situation, there are different methods for measuring the liquid level, as can be summarized as follows: (a) Float type (b) Hydrostatic differential pressure gage type (c) Capacitance type (d) Ultrasonic type (e) Radiation technique. Some of the techniques are elaborated in this section. Hydrostatic Differential Pressure type The hydrostatic pressure developed at the bottom of a tank is given by: p = h ρ g where h is the height of the liquid level and ρ is the density of the liquid. So by putting two pressure tapings, one at the bottom and the other at the top of the tank, we can measure the differential pressure, which can be calibrated in terms of the liquid level. Such a schematic arrangement is shown in Fig. 1. The drum level of a boiler is normally measured using this basic principle. However proper care should be taken in the measurement compensate for variation of density of water with temperature and pressure. Version 2 EE IIT, Kharagpur 3

Capacitance type This type of sensors are widely used for chemical and petrochemical industries; and can be used for a wide range of temperature (-40 to 200 o C) and pressure variation (25 to 60 kg/cm 2 ). It uses a coaxial type cylinder, and the capacitance is measured between the inner rod and the outer cylinder, as shown in Fig. 2. The total capacitance between the two terminals is the sum of (i) capacitance of the insulating bushing, (ii) capacitance due to air and liquid vapour and (iii) capacitance due to the liquid. If the total capacitance measured when the tank is empty is expressed as C 1, then the capacitance or the liquid level of h can be expressed as: 2πε 0 ( ε1 ε 2 ) h C t = C1 + ln ( r2 / r1 ) where, ε 1 is the relative permittivity of the liquid and ε 2 is the relative permittivity of the air and liquid vapour ( 1). Hence a linear relationship can be obtained with the liquid level. The advantage of capacitance type sensor is that permittivity of the liquid is less sensitive to variation of temperature and can be easily compensated. Version 2 EE IIT, Kharagpur 4

Ultrasonic type Ultrasonic method can be effectively used for measurement of liquid level in a sealed tank. An ultrasonic transmitter/receiver pair is mounted at the bottom of the tank. Ultrasonic wave can pass through the liquid, but gets reflected at the liquid-air interface, as shown in Fig.3. The time taken to receive the pulse is measured, that can be related with the liquid level. For accurate measurement, variation of speed of sound with the liquid density (and temperature) should be properly compensated. Radiation technique Radioactive technique also finds applications in measurement of level in sealed containers. Radioactive ray gets attenuated as it passes through a medium. The intensity of the radiation as it passes a distance x through a medium is given by: I( x) = I e α x where I 0 0 is the incidental intensity and α is the absorption co-efficient of the medium. Thus if we I 0 measure the intensity of the radiation, knowing, and α, x can be determined. There are several techniques which are in use. In one method, a float with a radioactive source inside is allowed to move along a vertical path with the liquid level. A Geiger Muller Counter is placed at the bottom of the tank along the vertical path and the intensity is measured. The basic scheme is shown in Fig. 4. The method used in a batch filling process of bottles, uses a source-detector assembly that can slide along the two sides of the bottle, as shown in Fig. 5. As soon as the source-detector assembly passes through the liquid-air interface, there would be a large change in the signal received by the detector. Radioactive methods, though simple in principle, find limited applications, because of possible radiation hazards. However radioactive methods are routinely used for level measurement of grains and granular solids. Version 2 EE IIT, Kharagpur 5

3. Humidity Measurement Humidity measurement finds wide applications in different process industries. Moisture in the atmosphere must be controlled below a certain level in many manufacturing processes, e.g., semiconductor devices, optical fibres etc. Humidity inside an incubator must be controlled at a very precision level. Textiles, papers and cereals must be dried to a standard storage condition in order to prevent the quality deterioration. The humidity can be expressed in different ways: (a) absolute humidity, (b) relative humidity and (c) dew point. Humidity can be measured in different ways. Some of the techniques are explained below. Hygrometer Many hygroscopic materials, such as wood, hair, paper, etc. are sensitive to humidity. Their dimensions change with humidity. The change in dimension can be measured and calibrated in terms of humidity. Psychrometer Psychrometric method for measurement of relative humidity is a popular method. Two bulbs are used- dry bulb and wet bulb. The wet bulb is soaked in saturated water vapour and the dry bulb is kept in the ambient condition. The temperature difference between the dry bulb and wet bulb is used to obtain the relative humidity through a psychrometric chart. The whole process can also be automated. Dew point measurement If a gas is cooled at constant pressure to the dew point, condensation of vapour will start. The dew point can be measured by placing a clean glass mirror in the atmosphere. The temperature of the mirror surface is controlled and reduced slowly; vapour starts condensation over the mirror. Optical method is used to detect the condensation phenomena, and the temperature of the mirror surface is measured. Conductance/Capacitance method of measurement Many solids absorb moisture and their values of the conductance or capacitance change with the degree of moisture absorption. Moisture content in granules changes the capacitance between two electrodes placed inside. By measuring the capacitance variation, the moisture content in the Version 2 EE IIT, Kharagpur 6

granules can be measured. Similarly, moisture content in paper and textiles change their resistance. A schematic arrangement for measurement of moisture content in paper or textiles using Resistance Bridge is shown in Fig. 6. Infrared Technique Water molecule present in any material absorb infrared wave at wavelengths 1.94µm, 2.95 µm and 6.2µm. The degree of absorption of infrared light at any of these wavelengths may provide a measure of moisture content in the material. 4. Measurement of ph ph is a measure of hydrogen ion concentration in aqueous solution. It is an important parameter to determine the quality of water. The ph value is expressed as: 1 ph = log10 C Where C is the concentration of H + ions in a solution. In pure water, the concentration of H + ions is 10-7 gm/ltr at 25 o C. So the ph value is 1 ph = = 7. 7 log10 10 The advantage of using ph scale is that the activities of all strong acids and bases can be brought down to the scale of 0-14. The ph value of acidic solutions is in the range 0-7 and alkaline solutions in the range 7-14. The ph value of a solution is measured by using ph electrode. It essentially consists of a pair of electrodes: measuring and reference electrode, both dipped in the solution of unknown ph. These two electrodes essentially form two half-cells; the total potential developed is the difference between the individual electric potential developed in each half cell. While the potential developed in the reference cell is constant, the measuring cell potential is dependent on the hydrogen ion concentration of the solution and is governed by Nernst s equation: RT E = E0 + ln( ac) nf Where: Version 2 EE IIT, Kharagpur 7

E= e.m.f of the half cell E 0 = e.m.f of the half cell under saturated condition R= Gas constant (8.314 J/ 0 C) T= Absolute temperature (K) N= valance of the ion F= Faraday Constant = 96493 C a= Activity co-efficient (0 a 1) ; for a very dilute solution, a 1 C= molar concentration of ions. Measuring Electrode The measuring electrode is made of thin sodium ion selective glass. A potential is developed across the two surfaces of this glass bulb, when dipped in aqueous solution. This potential is sensitive to the H + ion concentration, having a sensitivity of 59.2 mv/ph at 25 0 C. Fig. 7 shows the basic schematic of a measuring probe. The buffer solution inside the glass bulb has a constant H + ion concentration and provides electrical connection to the lead wire. Reference Electrode The basic purpose of a reference electrode is to provide continuity to the electrical circuit, since the potential across a single half cell cannot be measured. With both the measuring and reference cells dipped in the same solution, the potential is measured across the two lead wires. A reference electrode should satisfy the following basic requirements: (i) The potential developed should be independent of H + ion concentration. (ii) The potential developed should be independent of temperature (iii) The potential developed should not change with time. Considering all these requirements, two types of reference electrodes are commonly used: (i) Calomel (Mercury-Mercurous Chloride) and (ii) Silver-Silver Chloride. The construction of a Calomel reference electrode is shown in Fig. 8. The electrical connection is maintained through the salt bridge. Version 2 EE IIT, Kharagpur 8

Sometimes the reference and measuring electrodes are housed together, as shown in Fig. 9. This type of electrode is known as Combination Electrode. The reference electrode used in this case is Silver-Silver Chloride. The combination is dipped in the solution whose ph is to be measured and the output voltage is the difference between the e.m.f.s generated by the measuring glass electrode and the reference electrode. Version 2 EE IIT, Kharagpur 9

Measuring scheme The sensitivity of ph probe is around 59.2mv/pH at 25 0 C. This sensitivity should be sufficient for measurement of voltage using ordinary electronic voltmeters. But, that is not the case; special measuring circuits are required for measurement of ph voltage. This is because of the fact that the internal resistance of the ph probe as a voltage source is very high, in the order of 10 8-10 9 Ω. This is because of the fact; the electrical path between the two lead wires is completed through the glass membrane. As a result, the input resistance for of the measuring device must be at least ten times electrode resistance of the electrode. FET-input amplifier circuits are normally used for amplifying the voltage from the ph probe. Not only that, the insulation resistance between the leads must also be very high. They are normally provided with moisture resistance insulation coating. The voltage in the ph probe is temperature dependent, as evident from Nernst equation. As a result suitable temperature compensation scheme should also be provided in the measuring scheme. Review Questions 1. How would you measure level of a liquid inside a sealed tank? Explain with a schematic arrangement any one of the methods. 2. Name different techniques used for level measurement of a liquid. Explain the principle of operation of hydrostatic differential pressure level gage. 3. Name few instances where measurement of humidity/ moisture finds important applications in industry. 4. How the moisture content in solids can be measured? Give an example and show the schematic arrangement. Version 2 EE IIT, Kharagpur 10

5. Define ph of a solution. What is the hydrogen ion concentration of a solution if the ph of the solution is 5.0? 6. Explain with simple sketches the construction of measuring electrode and reference electrode. Why two electrodes are required for ph measurement? 7. Why temperature compensation scheme should be provided in ph measurement? 8. What special arrangements are to be provided for amplifying the voltage generated in a ph electrode? Justify. Version 2 EE IIT, Kharagpur 11