Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Similar documents
Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

Research of Polychlorinated Biphenyls (PCBs) in Vegetables by GC-MS/MS

Quantitative Analysis of Opioids Using a Triple-Quadrupole GC/MS/MS

Quantitative Determination of Dioxins in Drinking Water by Isotope Dilution using Triple Quadrupole GC-MS/MS

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

Pittcon P

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Chiral Separation Using SFC and HPLC

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Fast and Accurate Analysis of Fluorotelomer Alcohols and Acrylates using Triple Quadrupole GC-MS/MS

Simultaneous determination of 74 kinds of pesticides in Chinese herbal medicine listed in Chinese pharmacopoeia (2015) by GPC-GC-MS/MS

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer)

Fast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

Fast and Accurate Quantitation of Perfluorinated Sources from Textiles using Gas Chromatography-Triple Quadrupole Mass Spectrometry

Determination of 59 potential Allergens in Perfumes by twin-line fast GCMSMS

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

WADA Technical Document TD2003IDCR

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Technical Report. Abstract: Pu Wang 1, Huizhong Sun 1, Qinghua Zhang 1, Feifei Tian 2, Lei Cao 2

Highly Sensitive and Rugged GC/MS/MS Tool

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

C146-E209. Headspace Samplers. HS-20 Series

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Gas Chromatograph Mass Spectrometer GCMS-TQ8040 C146-E251

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Gas Chromatograph Mass Spectrometer GCMS-TQ8050 C146-E326

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

U.S. EPA Method 8270 for multicomponent analyte determination

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Multi-residue analysis of pesticides by GC-HRMS

Protein Sequencer PPSQ-31A / 33A C297-E055B

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Screening of pesticides in water using SPE on line

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

Benzodiazepines and Barbiturates

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Hints and Tips for APGC Coupled with Xevo TQ-S for Food and Environmental Analysis

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Solid Phase Micro Extraction of Flavor Compounds in Beer

Highly sensitive quantitative analysis of Felodipine and Hydrochlorothiazide from plasma using LC/MS/MS

Evaluation of a novel glass vial overcoming adsorption effect for pharmaceutical drugs

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Methods of Analysis of Polychlorinated Biphenyl Congeners Using an Application-Specific Capillary GC Column

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Maximizing Sample Throughput In Purge And Trap Analysis

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

ApplicationNOTE Abstract

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Consumables Guide for the Cannabis Testing Lab

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

GC-CI-MS analysis of TMS derivatives

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

EPA Office of Water Method Update Rule Final August 28, 2017 William Lipps October 2017

Liquid Chromatograph Mass Spectrometer LCMS-8040 C146-E188

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Liquid Chromatograph Mass Spectrometer LCMS-8040 C146-E188A

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Sensitive Detection of Pyrethroids in Surface Water and Sediment

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Fast Analysis of Chlorinated Phenoxy Herbicides in Environmental Samples by Negative ESI LC-MS/MS 14337

Transcription:

PO-CON111E Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry Pittcon 201 1120-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole Lock and Laura Chambers Shimadzu Scientific Instruments, Columbia, MD, USA

Introduction This poster describes development of a Multiple Reaction Monitoring (MRM) instrument method using a commercially available triple quadrupole GC/MS/MS for detection and quantitation of N-nitrosamines after extraction as described by USEPA Method 521 (2004) 1. GC/MS/MS in MRM mode produces significant improvements in selectivity and specificity, as well as dramatically lower detection limits than single quadrupole GC/MS, especially in complex matrices producing background interferences. In addition, this poster presents final instrument configuration and operating conditions, as well as instrument validation results, including estimated MDLs, and precision & accuracy as evaluated using standards at various concentration levels. Method 521 tandem MS with CI results are compared to the electron impact ionization triple quadrupole results. The intent was to create a determination step simpler than the techniques described in the method; therefore, this method uses 70 Electron Volt EI ionization, and simple syringe injection of 2 micro-liters of sample. Eliminating CI and LVI simplifies the method. Experimental This study was conducted using a Shimadzu GCMS-TQ8040 (Figure 1) configured with a Restek capillary column designed specifically for the analysis of N-Nitrosamines. The GC was operated in constant linear velocity mode, providing the best chromatographic resolution, symmetrical peak shapes, and enhanced sensitivity for complete separation of all target analytes from each other and from the solvent peak. A commercial The objective of this poster was to develop a simpler determination step than that described in the existing EPA 521 method and the various other N-nitrosamine experiments reported elsewhere 2,3. Because Method 521 is a drinking water method, there is very little modification allowed. Even though Section.11.3 of the method states: The tandem mass spectrometer may be either a triple quadrupole or an ion trap, the same section also states during the method development, only ion trap spectrometers were used. Changing a detector is not allowed unless there is sufficient data to support the change. Unlike the flexibility allowed with 40 CFR Part 13. for wastewater, modifications of drinking water methods must go through a full ATP evaluation; since drinking water regulations are national standards, single laboratory validations are not permitted 3. This poster demonstrates a proof of concept and demonstrates that the triple quadrupole MS/MS detection technique provides data that are as equally effective as tandem ion trap MS/MS. mixture of seven N-nitrosamine compounds, one internal standard, and one surrogate was used to prepare calibration curves ranging from 0.5 50 ng/l. The standards were prepared in methylene chloride. Chromatographic conditions were established and the MRM method was optimized for each component. The instrument operating conditions are shown in Table 1. Figure 1: Shimadzu GCMS-TQ8040 2

Table 1: GCMS-TQ8040 Operating Conditions Instrument : GCMS-TQ8040 Column : Stabilwax-MS (Restek PN 1073), 30 m x 0.25 mm x 0.25 um df Oven Program : 50 C, hold 2.0 minutes, 15 C/minute to 130 C, 20 C/minute to 220 C, hold 4.0 minutes Injector : Pulse splitless (300kPa for 1.0 mimute) 200 C Single taper w/wool, 3.5 mm ID x 5.0 x 5 (Restek PN 2333.5) Injection volume, 2.0 ul Carrier Gas Column Carrier Gas : Helium Constant linear velocity mode, 40.0cm/sec Total Flow 50.0 ml/min, Column Flow = 1.22 ml/min Purge Flow 3.0 ml/min Interface Temperature : 220 C Mass Spectrometer Ion Source Temperature : 200 C MS Operating Mode : Acquisition Mode, MRM CID gas, Argon (200 kpa) Solvent cut time,.5 minutes Detector voltage set relative to tune + 0.2 kv Threshold = 0.0 Ionization type, EI Electron Voltage, 70 ev Event time, 0.3 sec Analysis Time GC Cycle Time : 15.83 minutes : 20.0 minutes Calibration A 7-point calibration curve of 0.5 to 50 ng/l was analyzed using the conditions described in Table 1. The curves of all nine components were evaluated using linear regression and of the calculated response factors. Table 2 lists calibration data and the abbreviations used for the remainder of this text. Linear calibration curves with internal standards (IS) were established for seven N-nitrosamines as shown in Figure 2. The linearity with correlation coefficient (R2) greater than 0. across the calibration range of 0.5 ppb 50.0 ppb was obtained. A MRM chromatogram from a mid-point standard is shown in Figure 3. 3

Table 1: GCMS-TQ8040 Operating Conditions Abbreviation Retention Time (min) % RSD or r 2 N-Nitrosodi-n-propylamine (IS) -IS 8.21 IS N-Nitrosodimethylamine d- (SURR) (SURR).808 Surr N-Nitrosodimethylamine.818 0.7 N-Nitrosomethylethylamine 7.350 0.8 N-Nitrosodiethylamine 7.0 0.8 N-Nitrosodi-n-propylamine 8.8 0. N-Nitrosodi-n-butylamine 10.438 0. N-Nitrosopiperidine 10.707 0.8 N-Nitrosopyrrolidine 10.2 0. Figure 2: Calibration Curves Figure 3: Mid-Point Standard MRM Chromatogram for 20 ng/l Calibration Standard 4

MRM Method Development MRM transitions were monitored for each component. Quantitative and qualitative transitions were selected to provide maximum sensitivity and as independent confirmation of the compounds identity. The Ion Shield High Efficiency EI source minimized fragmentation at 70 ev, providing an optimum abundance and transmission of ions into the quadrupoles. Method settings were made to provide enough sensitivity to easily detect and quantify the target analytes at concentrations equal to or better than Method 521. MRM transitions and collision energies (CE) for each compound are shown in Table 3. Table 3: GCMS-TQ8040 MRM Transitions and Collision Energies Compared to Product Ions Given in Method 521 Precursor Quantitative Product CE (V) Precursor Qualitative Product CE (V) Method 521 Product Ion -IS 78.00 50.10 144.00 50.10 15 7 (7) - (SURR) 80.00 50.10 80.00 4.10 18 4(5) 74.00 44.10 74.00 42.10 15 43 (5) 88.00 71.10 88.00 73.10 1 (1) 102.00 85.10 102.00 5.10 15 75 (75) 130.00 113.20 130.00 88.10 8 (8) 11.00.10 158.15.1 57 (103) 114.00 84.10 114.00 7.10 55 (55) 100.00 55.10 100.00 8.10 () Method 521 product ions are based on methanol as the ionization gas. Values in parentheses are for an acetonitrile ionization gas. The proposed method monitors two transitions per analyte. The most sensitive transition was chosen for quantitative analysis. The other transition is used as qualitative verification of the identity of each peak. Instrument Detection Limit An instrument detection limit (IDL) study was made using eight replicate injections at 0.5, 1.0 and 1.25 ng/l standards. These estimated Method Detection Limit (MDL) results were compared to Method 521 detection limits and are shown in Table 4. The IDL study for the triple quadrupole method was made using un-extracted standard solutions. Detection limits for extracted samples will be slightly higher. The data indicates that detection limits in extracts should be essentially equivalent to Method 521 MDLs. Table 4: Method Detection Limits 0.5 ng/l standard IDL ng/l 1.0 ng/l standard 1.25 ng/l standard Method 521 MDL (ng/l) 0.14 0.22 0.1 0.28 0.0 0.11 0.0 0.28 0.08 0.05 0.10 0.2 0.07 0.07 0.11 0.32 0.10 0.08 0.1 0.3 0.07 0.0 0.05 0. 0.12 0.14 0.22 0.35 5

Precision and Accuracy Eight replicates of 10 ppb and 2.0 ppb were made to determine precision and accuracy. Table 5 and Table list the results of the precision and accuracy studies compared to EPA 521. Table 5: Accuracy and Precision at 10 ng/l Precision and Accuracy (10 ng/l) Triple Quad Method EPA 521 10 2.1 88.7 3.8 101 1.1 8.5 4.5 8. 1.0 87.5.1 8.7 0. 7.0 10 5.4 1.3 8.4.4.4 1.4 1.8 3.7 4.7 1. 101 5.0 Table : Accuracy and Precision at 2.0 ng/l Precision and Accuracy (2.0 ng/l) Triple Quad Method EPA 521 103 2. 4.7 12 7.8 1.3 81.8.. 2.2 84..0 5.5 1.5 81.7 8.0.8 3.0 85.2 1 7.5 2.0 8.3 20 8.4 4.4 2. 12 The triple quadrupole method results were on standards, not extracts, so it is expected that the recoveries and precision would be better than what would be obtained on extracted samples. However, the data indicates no problem with the instrumental method and that recovery and precision in extracted samples should be essentially equivalent to results obtained on an ion trap detector. Surrogate Standard Stability Surrogate recovery was monitored throughout the entire sequence of calibration and analysis of check standards. These analyses were performed in separate batches over a period of two days. These data are shown in Figure 4. Method 521 requires that surrogate recovery be within 70-130 %. The excellent recovery (within 80-120% ) of the surrogate standard throughout the run indicates the triple quad method is rugged and suitable for routine laboratory use.

Conclusion Triple quadrupole analysis with 70 EV EI ionization and 2uL direct injections simplifies detection and quantitation of the N-Nitrosamine compound. Detection limits, precision, and accuracy appear equal to or better than Method 521. Using a triple quadrupole GCMS, such as the Shimadzu References Figure 4: Surrogate Standard Recovery GCMS-TQ8040, for N-nitrosamines in drinking water is a viable alternative to EPA Method 521. This poster evaluated standards only. Our evaluation indicates that triple quadrupole GCMS is a suitable alternative to the ion trap detector described in Method 521. 1. J.W. Munch and M.V. Basset; Method 521 Determination of Nitrosamines in Drinking Water by Solid Phase Extraction and Capillary Column Gas Chromatography with Large Volume Injection and Chemical Ionization Tandem Mass Spectrometry (MS/MS) Version 1, September 2004; NERL; ORD; U.S.EPA; Cincinnati Ohio 2. Cardona, Steiniger, Silcock, and Semyonov; Use of GC-MS Triple Quadrupole Instrumentation for Nitrosamine Analysis; Poster Presentation, Thermo Scientific, NEMC2015 3. Honnold, Curtis, and Walker; Analysis of N-nitrosamines in water by Gas Chromatography, Electron Impact Ionization Tandem Mass Spectrometry; Poster Presentation, Agilent Technologies, NEMC2015 4. EPA 815-R-15-007; Protocol for the Evaluation of Alternate Test Procedures for Organic and Inorganic Analytes in Drinking Water, February 2015 First Edition: March, 201 www.shimadzu.com/an/ For Research Use Only. Not for use in diagnostic procedure. This publication may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country. The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu. Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its affiliates, whether or not they are used with trademark symbol TM or. Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own. The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject to change without notice. Shimadzu Corporation, 201