PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Similar documents
Application Note. Abstract. Authors. Environmental

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Rely on Rxi -PAH Columns

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Determination of 24 PAHs in Drinking Water

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood Using GC/MS

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Selection of a Capillary

Analysis. Priority LC GC. Zebron PAH Accurately quantitate EU and EPA PAHs in less than 28 minutes

Selection of a Capillary GC Column

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

This project has been funded with support from the European Commission. This publication reflects the views only of the authors, and the Commission

ENVIRONMENTAL analysis

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Challenging Pesticide Analysis Using an Agilent J&W DB-35ms Ultra Inert GC Column

Application Note. Abstract. Authors. Introduction. Food Safety

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Ultra-Inert chemistry for Trace Level Analysis

Katherine K. Stenerson, Michael Ye, Michael Halpenny, Olga Shimelis, and Leonard M. Sidisky. Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA

Author. Abstract. Introduction

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

The Suite for Environmental GC Analysis

CALIBRATION OF GC/MS METHOD AND VALIDATION OF THE MODIFIED SAMPLE PREPARATION FOR DETERMINATION OF POLYCYCLIC AROMATIC HYDROCARBONS

Choosing the Correct GC Column Dimensions and Stationary Phase

Report on the third inter-laboratory comparison test organised by the Community Reference Laboratory for Polycyclic Aromatic Hydrocarbons

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

Sample Pre-Treatment : Introduction to Commonly Used Techniques for Gas

Understanding Gas Chromatography

CERTIFICATION REPORT. Certification of the Mass Fraction of Polycyclic Aromatic Hydrocarbons (PAHs) in Toluene

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Facility: Mansfield Revision 3. Title: PAHs by SPME Page 1 of 11

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

Highly Sensitive and Rugged GC/MS/MS Tool

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Application. Gas Chromatography March 1998

Application Note. Abstract. Authors. Introduction

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Reviewing Material Safety Data Sheets to Verify Significant Drinking Water Threats

Reliably measure trace-level contaminants and meet regulatory requirements with confidence

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Andrei Medvedovici, Florina Micăle, Florentin Tache

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Appendix 1: Polycyclic Aromatic Compounds: Nomenclature and Analysis

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Application. Gas Chromatography February Introduction

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Fast and Accurate Analysis of PBDEs in a Single Run, Including BDE-209

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Fast and Accurate Analysis of PBDEs in a Single Run, Including

Environmental Forensic Principals for Sources Allocation of Polycyclic Aromatic Hydrocarbons

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Volatile organic compounds (VOCs):

APPENDIX A TO PART 136 METHODS FOR ORGANIC CHEMICAL ANALYSIS OF MUNICIPAL AND INDUSTRIAL WASTEWATER METHOD 610 POLYNUCLEAR AROMATIC HYDROCARBONS

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Determination of Volatile Substances Proof of Food Adulteration

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Determination of Polycyclic Aromatic Hydrocarbons (PAH) Adsorbed on Soot Formed in Pyrolysis of Acetylene at Different Temperatures

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Single-Laboratory Validation of a GC-MS Method for the Determination of 27 Polycyclic Aromatic Hydrocarbons (PAHs) in Oils and Fats

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

ILLA SpA VIA GHISOLFI GUARESCHI, NOCETO (PR)

Introduction to Capillary GC

Test Report. Report No. ECL01H Page 1 of 5

Introduction to Capillary GC

Transcription:

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices Food Quality and Environmental Author Ken Lynam Agilent Technologies, Inc. 280 Centerville Road Wilmington, DE 19808 Abstract The European Union (EU) regulates a series of PAHs found primarily in food matrices. This list is referred to as the EU 1 +1 list. The US Environmental Protection Agency (EPA) regulates a series of 16 PAHs historically addressed as environmental pollutants. Both lists contain unique analytes that present different separation challenges. However, there are eight analytes that are common to both lists. In this study, resolution of all 24 combined regulated PAHs is achieved, in under 28 minutes using an Agilent J&W DB-EUPAH 20 m x 0.18 mm, 0.14 µm High Efficiency GC column. Resolution of 23 of the 24 combined regulated PAHs is shown using an Agilent J&W DB-ms 20 m x 0.18 mm, 0.18 µm High Efficiency GC column in under 22 minutes. Both the Agilent J&W DB-EUPAH and DB-ms columns are excellent column choices for analysis of the regulated PAHs. The Agilent J&W DB-EUPAH is recommended when separation of benzo[b,j,k]fluoranthene isomers is required.

Instruction Table 1. EU and US-EPA Regulated PAH Compounds Polycyclic aromatic hydrocarbons (PAHs) are a large class of organic compounds containing two or more fused aromatic rings. PAHs often result from the incomplete combustion of organic substances such as wood, coal, and oil. The European Union PAH regulation has focused on these substances as potential contaminants in the food supply. A main source of potential human exposure to PAHs is through heat processing of meat and dairy products, such as grilling and smoking [1]. There are serious health concerns regarding PAHs since many are classified as carcinogenic or mutagenic [2]. In 200, the European Commission recommended the monitoring of fifteen EU priority PAHs along with an additional PAH highlighted by the Joint FAO/WHO Expert Committee on Food Additives (JECFA) [3]. The EU 1+1 priority PAHs along with the US-EPA-regulated PAHs are provided in Table 1 [4]. There are eight PAHs that are common to both the EU-1 +1 list and the US-EPA list. A % phenyl methylpolysiloxane stationary phase column is the most commonly used GC column for PAH analysis. This nonpolar column yields good resolution for the 16 US-EPA PAHs [,6], however, three critical pairs of the 1+1 EU PAHs co-elute and are difficult to resolve by mass spectrometry [7,8]. These challenging pairs are benz[a]anthracenecyclopenta[c,d]pyrene- chrysene,benzo[b]fluoranthenebenzo[k]fluoranthene-benzo[j]fluoranthene, and indeno[1,2,3- cd] pyrene-dibenz[a,h]anthracene. Agilent J&W DB-EUPAH, a midpolar GC column, improves the resolution of these critical pairs allowing for more accurate detection and quantitation of the 1 +1 EU priority PAHs. Another set of challenging analytes is the four dibenzopyrene isomers. Due to their high molecular weight (MW 302), these isomers are prone to discrimination and poor peak shape. Broad, tailing peaks make reliable quantitation difficult and decrease the signal-to-noise ratio, resulting in an increase in the limits of detection. Limiting analyte dwell time on the column and in the GC/MS interface can offset these deleterious chromatographic effects. Shorter columns with thinner film thickness and high operating temperatures are all factors that collectively can improve peak shapes for these analytes. Peak # Component CAS # MW EU 1+1 EPA 1 Naphthalene 91-20-3 128 x 2 Acenaphthylene 208-96-8 12 x 3 Acenaphthene 83-32-9 14 x 4 Fluorene 86-73-7 166 x Phenanthrene 8-01-8 178 x 6 Anthracene 120-12-7 178 x 7 Fluoranthene 206-44-0 202 x 8 Pyrene 129-00-0 202 x 9 Benzo[c]fluorene 20-12-9 216 x Benz[a]anthracene 6--3 228 x x 11 Cyclopenta[c,d]pyrene 27208-37-3 226 x 12 Chrysene 218-01-9 228 x x 13 -Methylchrysene 3697-24-3 242 x 14 Benzo[b]fluoranthene 20-99-2 22 x x 1 Benzo[k]fluoranthene 207-08-9 22 x x 16 Benzo[j]fluoranthene 20-82-3 22 x 17 Benzo[a]pyrene 0-32-8 22 x x 18 Indeno[1,2,3-cd]pyrene 193-39- 276 x x 19 Dibenz[a,h]anthracene 3-70-3 278 x x 20 Benzo[g,h,i]perylene 191-24-2 276 x x 21 Dibenzo[a,l]pyrene 191-30-0 302 x 22 Dibenzo[a,e]pyrene 192-6-4 302 x 23 Dibenzo[a,i]pyrene 189--9 302 x 24 Dibenzo[a,h]pyrene 189-64-0 302 x Regulated PAH compounds shown in plain text are included only in the US- EPA set, compounds in Italic are included only in the EU 1+1 list, and the compounds in bold are included in both the US-EPA and EU 1+1 lists. 2

Experimental Table 4. Flow Path Supplies GC EU PAH standard (Agilent p/n 190-0487) and US-EPA mixture (Agilent p/n 800-603) were diluted separately to a concentration of 2 µg/ml using class A glassware and pipets. These solutions were then mixed 1:1 to for a final concentration of 1-2 µg/ml of all 24 regulated PAHs. Table 2. Chromatographic Conditions DB-EUPAH Column Sample: 0. µl 1-2 µg/ml EU + EPA PAH combined standards (EU PAH standard Agilent p/n 190-0487 and EPA PAH standard Agilent p/n 800-603) GC/MS: Agilent 7890A GC System with an Agilent 97C Series GC/MSD, TAD, and an Agilent 7873B automatic liquid sampler Column: Agilent J&W DB-EUPAH 20 m 0.18 mm, 0.14 µm (Agilent p/n 121-9627) Carrier: Helium 60 cm/sec 1.8 ml/min constant flow Oven: 70 C (0.8 min), 70 C/min to 180 C, 7 C/min to 230 C (6 min), 40 C/min to 280 C ( min) 2 C/min to 33 C ( min) Inlet: 300 C splitless, purge 0 ml/min at 0.2 min Inlet liner: Helix double taper deactivated (Agilent p/n 188-398) MSD: Sim/Scan mode 0-400 AMU, transfer line 340 C, source 340 C, quad 10 C Vials: Amber screw top glass vials (Agilent p/n 183-2072) Vial Caps: Screw caps (Agilent p/n 182-0723) Vial inserts: 0 µl glass/polymer feet (Agilent p/n 181-8872) Syringe: µl (Agilent p/n 183-4729) Septum: Advanced green (Agilent p/n 183-479) Inlet Seal: Gold plated inlet seal (Agilent p/n 188-367) Inlet liners: Helix double taper deactivated (Agilent p/n 188-398) Ferrules: 0.4 mm ID short; 8/1 vespel/graphite (Agilent p/n 181-3323) 20 magnifier : 20 Magnifier loop (Agilent p/n 430-20) Table 3. Chromatographic Conditions DB-ms Column Sample: 0. µl 1-2 µg/ml EU + EPA PAH combined standards (EU PAH standard Agilent part # 190-0487 and EPA PAH standard Agilent part # 800-603) GC/MS: Agilent 7890A GC System with an Agilent 97C Series GC/MSD, TAD, and an Agilent 7873B automatic liquid sampler Column: Agilent J&W DB- ms 20 m 0.18 mm, 0.18 µm (Agilent p/n 121-22) Carrier: Helium 60 cm/sec 1.8 ml/min constant flow Oven: C (0.4 min), 2 C/min to 200 C, 8 C/min to 280 C, C/min to 320 C (2 min) 2 C/min to 33 C ( min) Inlet: 300 C splitless, purge 0 ml/min at 0.2 min Inlet liner: Helix double taper deactivated (Agilent part # 188-398) MSD: Sim/Scan mode 0-400 AMU, transfer line 340 C, source 340 C, quad 10 C 3

Discussion of Results Figure 1 shows the separation of all 24 analytes included in the EU 1 + 1 and US-EPA PAH lists on a DB-EUPAH column. This separation was accomplished in under 28 minutes on an Agilent J&W DB-EUPAH 20 m x 0.18 mm, 0.14 µm column (Agilent p/n 121-9627). The selectivity of this midpolar column is necessary to resolve the benzo(b,j,k) fluoranthene isomers. High temperature stability is also required for elution of the high boiling point dibenzopyrenes. The injection volume was reduced to 0. µl in order to scale the separation to the 0.18 or high efficiency GC (HEGC) format. This is often a necessary step when working with high efficiency columns because the higher efficiency of narrow bore columns is at the expense of sample loading capacity. Here the separation focused on achieving faster analysis. An example of a separation maximizing resolution with longer retention on this column was described previously [9]. Agilent J&W DB-EUPAH Separation of Combined EU 1 + 1 and US-EPA PAHs 2 3 4 4 14 1 20 6 17 19 12 16 18 8 7 9 13 21 22 1 11 23 24 0 2 4 6 8 12 14 16 18 20 22 24 26 Component 1 Naphthalene 2 Acenaphthylene 3 Acenaphthene 4 Fluorene Phenanthrene 6 Anthracene 7 Fluoranthene 8 Pyrene 9 Benzo[c]fluorene Benz[a]anthracene 11 Cyclopenta[c,d]pyrene 12 Chrysene 13 -Methylchrysene 14 Benzo[b]fluoranthene 1 Benzo[k]fluoranthene 16 Benzo[j]fluoranthene 17 Benzo[a]pyrene 18 Indeno[1,2,3-cd]pyrene 19 Dibenz[a,h]anthracene 20 Benzo[g,h,i]perylene 21 Dibenzo[a,l]pyrene 22 Dibenzo[a,e]pyrene 23 Dibenzo[a,i]pyrene 24 Dibenzo[a,h]pyrene Figure 1. EU and US-EPA regulated PAH separation on an Agilent J&W DB-EUPAH 20 m 0.18 mm, 0.14 µm column (Agilent p/n 121-9627). Chromatographic conditions as in Table 2 and flow path supplies as in Table 4. 4

Figure 2 shows the resolution of 23 of the 24 analytes included in the EU 1 + 1 and US-EPA PAH lists on an Agilent J&W DB-ms column. Benzo[j] fluoranthene is not resolved from benzo[k]fluoranthene using this column. However, when it is sufficient to report the sum of the benzoflouranthene isomers, the Agilent J&W DB-ms column is an excellent choice for the 24 regulated PAHs. This separation was accomplished in under 22 minutes on an Agilent J&W DB-ms 20 m 0.18 mm, 0.18 µm column (Agilent p/n 121-22). The Agilent J&W DB-ms separation offers a 27 % faster analysis time when compared to the DB-EUPAH separation shown in Figure 1. There are a number of best practices to consider when optimizing a GC/MS system for PAH analysis. The use of retention gaps and/or inlet backflushing can reduce maintenance and cycle times. Close examination of injection parameters such as injection volume, inlet temperature, purge time activation, solvent focusing and holding the oven temperature stable during injection can all contribute to better results. Minimizing inlet and system dwell time by operating at high linear velocities can also improve results. Another best practice for PAHs is to keep heated zones well insulated and hot to reduce the potential for system cold spots and the resultant signal loss. Conclusions All 24 (EU 1 + 1 and US-EPA) regulated PAHs are resolved using an Agilent J&W DB-EUPAH 20 m 0.18 mm, 0.14 µm column. The benzo[b,j,k]fluoranthene isomers were adequately separated for individual quantitation. This is the column of choice when resolution of benzo[j]fluoranthene is required. This separation was accomplished in under 28 minutes. 23 of 24 (EU 1 + 1 and US-EPA) regulated PAHs resolve using an Agilent J&W DB-ms 20 m 0.18 mm, 0.18 µm column. This column is an excellent choice when benzoflouranthene isomers are reported as a sum of the isomers and speed of analysis is critical. This separation was accomplished in under 22 minutes. Agilent J&W DB-ms Separation of Combined EU 1 + 1 and US-EPA PAHs 1 3 1 12 4 1-16 11 14 20 8 7 2 6 17 19 18 9 13 21 22 23 24 4 6 8 12 14 16 18 20 Component 1 Naphthalene 2 Acenaphthylene 3 Acenaphthene 4 Fluorene Phenanthrene 6 Anthracene 7 Fluoranthene 8 Pyrene 9 Benzo[c]fluorene Cyclopenta[c,d]pyrene 11 Benz[a]anthracene 12 Chrysene 13 -Methylchrysene 14 Benzo[b]fluoranthene 1 Benzo[k]fluoranthene 16 Benzo[j]fluoranthene 17 Benzo[a]pyrene 18 Indeno[1,2,3-cd]pyrene 19 Dibenz[a,h]anthracene 20 Benzo[g,h,i]perylene 21 Dibenzo[a,l]pyrene 22 Dibenzo[a,e]pyrene 23 Dibenzo[a,i]pyrene 24 Dibenzo[a,h]pyrene Figure 2. EU and US-EPA regulated PAH separation on an Agilent J&W DB- ms 20 0.18 mm, 0.18 µm (Agilent p/n 121-22) column. Chromatographic conditions as in Table 3 and flow path supplies as in Table 4.

Factors to consider in optimizing EU 1 +1 and US-EPA PAH analyses Choose an Agilent J&W DB-EUPAH when the resolution of benzo[b,j,k]fluoranthene isomers (24 of 24 peaks in combined set) is required. Choose an Agilent J&W DB-ms when benzo[b,j,k]fluoranthene isomers can be reported as a sum of the isomers. The Agilent J&W DB-ms resolves 23 of 24 regulated PAHs in 27 % faster cycle time than the Agilent J&W DB-EUPAH column. Consider the use of retention gaps and inlet backflushing to reduce cycle time and maintenance. Achieve faster analysis times with no loss of resolution using 0.18 mm id high efficiency GC columns. Optimize injection volume, temperature, purge time, and solvent focusing for best results on your instrument. Minimize inlet and system dwell time with high linear velocities. Keep heated zones hot to avoid cold spots and signal loss. 6. Kenneth Lynam and Doris Smith, "Polycyclic Aromatic Hydrocarbon (PAH) Analysis Using an Agilent J&W DB- ms Ultra Inert Capillary GC Column," Agilent Technologies publication 989-9181EN, July 2008. 7. Jose Angel, Gomez-Ruiz, Thomas Wenzl Evaluation of Gas Chromatography Columns for the Analysis of the 1 +1 EU-priority Polycyclic Aromatic Hydrocarbons (PAHs) Anal Bioanal Chem, (2009) 393: p. 1697-1707 8. L. R. Bordajandi, M. Dabrio, F. Ulberth, H. Emons, Optimization of the GC-MS Conditions for the Determination of the 1 EU Foodstuff Priority Polycyclic Aromatic Hydrocarbons J. Sep.Sci. 2008, 31, 1769-1778. 9. Doris Smith and Ken Lynam, GC/MS Analysis of European Union (EU) Priorty Polycyclic Aromatic Hydrocarbons (PAHs) using an Agilent J&W Db-EUPAH GC Column with a Column Performance Comparison, Agilent Technologies publication 990-4883EN, October 29, 2009 References 1. P. Mottier, V. Parisod, R. J. Turesky, J. Agric. Food Chem. 48 (2000), p. 1160. 2. Y. V. Pashin and L. M. Bakhitova, "Mutagenic and Carcinogenic Properties of Polycyclic Aromatic Hydrocarbons," Environ Health Prospect, 30 (1979),p.18-9. 3. European Union, Commission Recommendation 200/8/EC, Off. J. Eur. Comm. L34 (200) 43. 4. Environmental Protection Agency (EPA) of the United States of America, Compendium Method TO-13A, EPA, Cincinnati, OH, USA, 1999.. Frank David and Matthew S. Klee, "Fast Analysis of Polynuclear Aromatic Hydrocarbons using Agilent Low Thermal Mass (LTM) GC/MS and Capillary Flow Technology Quickswap for Backflush," Agilent Technologies publication 990-341EN, April 2009. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 20 Printed in the USA May 2, 20 990-872EN