The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

Similar documents
PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Rely on Rxi -PAH Columns

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Application Note. Abstract. Authors. Environmental

Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood Using GC/MS

Application Note GCMS-03 Reliable high-throughput quantification of melamine and related analogs below regulatory limits in food using the EVOQ GC-TQ

CALIBRATION OF GC/MS METHOD AND VALIDATION OF THE MODIFIED SAMPLE PREPARATION FOR DETERMINATION OF POLYCYCLIC AROMATIC HYDROCARBONS

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

Determination of 24 PAHs in Drinking Water

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

This project has been funded with support from the European Commission. This publication reflects the views only of the authors, and the Commission

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

ENVIRONMENTAL analysis

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

Report on the third inter-laboratory comparison test organised by the Community Reference Laboratory for Polycyclic Aromatic Hydrocarbons

Highly Sensitive and Rugged GC/MS/MS Tool

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

ILLA SpA VIA GHISOLFI GUARESCHI, NOCETO (PR)

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Katherine K. Stenerson, Michael Ye, Michael Halpenny, Olga Shimelis, and Leonard M. Sidisky. Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Estimation of LOQ for the Analysis of Persistent Organic Pollutants, in particular PCDD/Fs and PCBs

Hints and Tips for APGC Coupled with Xevo TQ-S for Food and Environmental Analysis

Environmental Forensic Principals for Sources Allocation of Polycyclic Aromatic Hydrocarbons

Technical Report. Abstract: Pu Wang 1, Huizhong Sun 1, Qinghua Zhang 1, Feifei Tian 2, Lei Cao 2

WADA Technical Document TD2003IDCR

APPENDIX G. Data Management Rules. Dioxin Data Report Appendix G. Lower Duwamish Waterway Superfund Site: T-117 Early Action Area

Application Note GCMS-09 Evaluation of Rapid Extraction and Analysis Techniques for Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood by GC/MS/MS

Analysis. Priority LC GC. Zebron PAH Accurately quantitate EU and EPA PAHs in less than 28 minutes

Application Note # CA

CERTIFICATION REPORT. Certification of the Mass Fraction of Polycyclic Aromatic Hydrocarbons (PAHs) in Toluene

Selection of a Capillary

Signal, Noise, and Detection Limits in Mass Spectrometry

PAHs in Parking Lot Sealcoats. Performance Study. Asphalt Based. Coal Tar Sealcoat Sealcoat. Sealcoat. Scrapings. Asphalt Based.

ApplicationNOTE Abstract

Using the Agilent 7000B Triple Quadrupole GC/MS for Parts per Trillion Detection of PAH Metabolites in Human Urine

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Single-Laboratory Validation of a GC-MS Method for the Determination of 27 Polycyclic Aromatic Hydrocarbons (PAHs) in Oils and Fats

Selection of a Capillary GC Column

The direct effect of the harbour. micropollutant concentration. Workshop Characterization of atmospheric pollution in harbour areas 26 June 2013

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

SPE AND GC MS INVESTIGATION OF ORGANIC CONTAMINANTS IN ATMOSPHERIC PRECIPITATION

ApplicationNOTE THE ANALYSIS OF POLYCHLORINATED BIPHENYLS (PCBS) BY GC-HIGH RESOLUTION MASS SPECTROMETRY USING THE MICROMASS AUTOSPEC ULTIMA NT

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Dissemination patterns and risks impersonated of PAHs contaminated in the surface and subsurface soil

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

MEASUREMENT OF DRY DEPOSITION AMOUNT OF PAHS IN ZONGULDAK REGION

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

Occupational exposure to polycyclic aromatic hydrocarbons in various technological processes i INTRODUCTION

GC/MS/MS Dioxin and Dioxin-like PCBs in Feed and Food. Jessica Westland, LSAG Applications Chemist

Determination of PAH profiles by GC-MS/MS in salmon muscle meat processed with four cold smoking techniques

Investigating Source, Age, Maturity, and Alteration Characteristics of Oil Reservoirs Using APGC/MS/MS Analysis of Petroleum Biomarkers

T H E ANA LYSIS O F DIOX INS AND F U R A NS USING H RG C- H IG H R E SO LU T IO N MS W IT H T H E AU T OS P EC- ULTIMA NT

Analytische Qualitätssicherung Baden-Württemberg

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Mass Spectrometry. Ionizer Mass Analyzer Detector

This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Validation of GC-MS/MS for Detection and Confirmation of Low-Level Dioxins

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

BenchTOF-dx. High-performance time-of-flight mass spectrometer for gas chromatography

Polycyclic Aromatic Hydrocarbons in Water by GC/MS PBM

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Multi-residue analysis of pesticides by GC-HRMS

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Method 8270C PAH. FD- Field Duplicate Samples

Facility: Mansfield Revision 3. Title: PAHs by SPME Page 1 of 11

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

Determination of 59 potential Allergens in Perfumes by twin-line fast GCMSMS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method)

Andrei Medvedovici, Florina Micăle, Florentin Tache

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Reviewing Material Safety Data Sheets to Verify Significant Drinking Water Threats

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Gas Chromatography Mass Spectrometry (GCMS)

VGA-100 Instrument Detection Limits and Linearity

Food analysis. Solutions for Your Analytical Business Markets and Applications Programs

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

ApplicationNOTE ULTRA TRACE ANALYSIS OF DIOXINS AND FURANS IN HUMAN ADIPOSE TISSUE USING SFE-LC EXTRACTION/CLEANUP AND THE WATERS AUTOSPEC ULTIMA NT

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Quantitative Determination of Dioxins in Drinking Water by Isotope Dilution using Triple Quadrupole GC-MS/MS

FUTURE CONFIRMATORY CRITERIA

Transcription:

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products Bruno Veyrand, Aline Brosseaud, Philippe Marchand, Fabrice Monteau, François Andre, Bruno Le Bizec, Timothy Jenkins 2, Jean-Marc Joumier 3 LABERCA, ENV, Atlanpôle La Chantrerie, BP 577, NANTES Cedex 3, France. 2 Waters Corporation, Manchester, UK. 3 Waters EHQ, St Quentin en Yvelines FR BACKGROUND Polycyclic aromatic hydrocarbons Polycyclic aromatic hydrocarbons (PAHs) are a group of chemicals that are formed during the incomplete burning of coal, oil, gas, wood, garbage or other organic substances, such as tobacco and charbroiled meat. There are more than different compounds. PAHs generally occur as complex mixtures (for example, as part of combustion products such as soot), not as single compounds. A few PAHs are used in medicines and to make dyes, plastics, and pesticides. Others are contained in asphalt used in road construction. They can also be found in substances such as crude oil, coal, coal tar pitch, creosote, and roofing tar. Food can be contaminated by an accumulation of PAHs in the food chain, due to their lipophilic properties and propensity to deposit in adipose tissue. Moreover, PAHs are also present due to heat processes such as smoking, grilling and smoke-drying. Exposure to PAHs is a major concern for human health, as PAHs have been classified as carcinogenic. In the European Union, a new legislation was adopted in 25 [] and provided a list of 5 PAHs to monitor: Cyclopenta[c;d]pyrene, Benz[a]anthracene, Chrysene, 5-Methylchrysene, Benzo[b]fluoranthene, Benzo[j]fluoranthene, Benzo[k]fluoranthene, Benzo[a]pyrene, Indeno[,2,3-c,d]pyrene, Dibenz[a,h]anthracene, Benzo[g,h,i]perylene, Dibenzo[a,l]pyrene, Dibenzo[a,e]pyrene, Dibenzo[a,i]pyrene, Dibenzo[a,h]pyrene. PAHs in food The concentration of PAHs in food is regulated by the European Union (28/25 regulation), which determines maximum allowable limits. The objective of this application note is to demonstrate the relevance of GC/MS/MS (gas chromatography with tandem quadrupole mass spectrometry) in the detection and quantification of the broad range of PAH compounds in food products. The GC/MS/MS technology will be compared to other analytical approaches. INSTRUMENTATION For GC/MS experiments, a single quadrupole instrument was used. For GC/MS/MS experiments, a Waters Quattro Micro GC tandem quadrupole mass spectrometer was used. Injector and transfer line temperatures were set at 28 C and 33 C respectively. Source temperature was set to 23 C. GC column was a 5 phenyl substitued polydimethylsiloxane based phase with the following characteristics: 3 m x.25 mm id., film thickness.25 μm. The GC temperature parameters were set as follows C ( min), 2 C.min - until 24 C ( min), 5 C.min - until 32 C ( min). Helium was used as carrier gas at ml.min -. Electron ionization (EI) was operated at 7 ev in any case. Waters Quattro micro GC

RESULTS Comparison of GC/MS and GC/MS/MS Fragmentation of PAHs after electron ionization (EI) leads exclusively to the observation of molecular ion (M + ). The mass spectrum is generally poor in fragment ions so that identification of each compound on single MS is limited to the M + signal. Typical example of observed ion chromatograms for 3 C 3 -pyrene (internal standard), 3 C 6 -fluoranthene (internal standard), fluoranthene and pyrene in smoked tuna sample is shown on Figure. For some traces such as the one shown for 3 C 6 -fluoranthene (m/z 28) and fluoranthene/pyrene (m/z ), the signal is affected by co-extracted compounds. Moreover the signal to noise ratio of 3 C 3 -pyrene is weak even at.5 μg.kg - level. The main advantage of the MS/MS technique is the possible fragmentation of PAHs in the collision cell using high collision energies (3 to 5 V) to produce specific product ions. Ion 25. (24.7 to 25.7): 7267.D 4 35 3 25 2 5 m/z = 25 5 8.2 8.4 8.6 8.8 9. 9.2 9.4 9.6 9.8 Ion 28. (27.7 to 28.7): 7267.D 25 2 m/z = 28 5 8.2 8.4 8.6 8.8 9. 9.2 9.4 9.6 9.8 Ion 22. (2.7 to 22.7): 7267.D 4 2 8 6 4 2 m/z = 22 8.2 8.4 8.6 8.8 9. 9.2 9.4 9.6 9.8 Ion. (.7 to.7): 7267.D 3 25 2 5 m/z = 5 8.2 8.4 8.6 8.8 9. 9.2 9.4 9.6 9.8 Figure. Ion chromatograms (GC/MS, EI ionization, SIM acquisition) corresponding to 3 C 3 -pyrene (m/z: 25; internal standard), 3 C 6 -fluoranthene (m/z 28; internal standard), fluoranthene and pyrene (both m/z 22 and ) in a smoked tuna sample.

The monitored transitions are shown for the same sample on Figure 2. The results show very significant S/N improvement. The diagnostic absence/presence of target compounds is greatly facilitated, and the quantification does not suffer from the presence of any interference. The given example is transposable to all compounds; moreover, the internal standard signals ( 3 C-analyte) do not disturb the corresponding 2 C-analyte. TF 62623 2: MRM of 6 Channels EI+.7 25 > 23 27 4.8e5 25>23 62623 2: MRM of 6 Channels EI+ 9.72 28 > 26 887 3.2e5 28>26 62623 2: MRM of 6 Channels EI+ 9.72 22 > 52 42.7.52e5 328 22>52 2 62623 2: MRM of 6 Channels EI+.7 22 > 2 7772 2.87e6 22>2 9.72 28868 9.25 9.5 9.75..25.5.75 Figure 2. Ion chromatograms (GC/MS/MS, EI ionization, SRM acquisition) corresponding to 3 C 3 -pyrene (m/z: 25>23; internal standard), 3 C 6 -fluoranthene (28>26; internal standard), fluoranthene and pyrene (both 22>52 and 22>2) in a smoked tuna sample.

Limit of detection achievable by GC/MS/MS Figure 3 shows an example of a smoked salmon sample in which concentration of benzo[a]pyrene has been determined at.4 μg/kg of wet matter. Even at this concentration level, the S/N to noise ratio is still substantial, so that the detection is made possible down to low ppt level. Fulfillment of identification criteria The high stability of PAHs makes them difficult to fragment even under strong EI conditions so that the observation of diagnostic product ions is rather limited. A direct consequence of this is the difficulty to unambiguously identify target analytes, especially with regard to the criteria specified in the 22/657/EC decision. MS/MS is a detection technique which enables at least 2 specific transitions for each compound to be obtained, guaranteeing reliable analyte identification. SF2 6262 5: MRM of 5 Channels EI+ 7.67 256 > 254 8 4.9e5 Internal Standard Benzo[a]pyrene 3 C 4 6262 5: MRM of 5 Channels EI+ 7.68 252 > 25 5743.4e5 Benzo[a]pyrene 252>25 S/N = 4 6262 5: MRM of 5 Channels EI+ 7.68 252 > 226 82 2.22e4 Benzo[a]pyrene 252>226 S/N = 3 3 7. 7.2 7.4 7.6 7.8 8. 8.2 8.4 Figure 3. Ion chromatograms (GC/MS/MS, EI ionization, MRM acquisition) corresponding to 3 C 4 -benzo[a]pyrene (256>254; internal standard), benzo[a]pyrene (252>25 and 252>226) in a smoked salmon sample at.4 μg/kg wet matter.

Figure 4 shows the monitoring of 4 transitions for benzo[ghi]perylene in cockle sample at 9.9 μg.kg -. It was found that the typical number of identification points that could be obtained for each PAH was at least 7. Application to Certified Reference Materials Figure 5 shows the calibration curve obtained by GC/MS/MS for the quantification of Benzo [a] pyrene. The calibration curve is obtained with a good correlation coefficient (r²=.9999), which allows the determination of concentration with high accuracy. An example of the quantification is made for a.5 g sample of mussel tissue (Standard Reference Material 2977, National Institute of Standards & Technology). Certified concentration for phenanthrene is 35. ± 3.8 μg/kg (dry matter basis). The isotope dilution approach (using 3 C 6 -Phenanthrene as internal standard) combined with GC/MS/MS analysis allows for high precision. In our example, the concentration was calculated at 34.8 μg/kg of dry matter (within of the assigned value). 5.66.53 62625 6: MRM of 9 Channels EI+ 2.3 282 > 28 3962 6.96e5 62625 6: MRM of 9 Channels EI+ 276 > 274 2.3 27663 7.4e6 62625 6: MRM of 9 Channels EI+ 2.3 276 > 272 93599 2.44e6 62625 6: MRM of 9 Channels EI+ 2.3 276 > 25 2233 2.67e5 62625 6: MRM of 9 Channels EI+ 2.3 276 > 226 29 2.4e5 3 C 2 Internal standard 276>274 276>272 276>25 276>226 2. 2.5 2. 2.5 22. 22.5 Figure 4. Ion chromatograms (GC/MS/MS, EI ionization, MRM acquisition) corresponding to 3 C 2 -benzo[g;h;i]perylene (m/z: 282>28; internal standard) and benzo[g;h;i]perylene (m/z: 276>274, 276>272, 276>25, 276>226) in a cockle sample at 9.9 μg.kg -.

Compound name: Benzo[a]pyrene Correlation coefficient: r =.999976, r^2 =.999952 Calibration curve:.74859 * x + -.2527 Response type: Internal Std ( Ref 3 ), * ( IS Conc. / IS ) Curve type: Linear, Origin: Exclude, Weighting: Null, Axis trans: None 699 Compliance with EU legislation Figure 6 shows a positive result from the analysis of a smoked fish sample. The concentration of Benzo[a]pyrene was determined as 6.94 (±.73) μg.kg - of wet matter. The maximum level in smoked fish permitted by European legislation (Regulation EC No. 28/25) is 5 μg.kg -. Response - pg 5 5 2 25 3 35 4 Figure 5. Calibration curve (GC-MS/MS, EI ionization, MRM acquisition) for Benzo [a] pyrene (252>25) versus 3 C 4 -Benzo [a] pyrene (256>254; internal standard). 6.24-23562 5: MRM of 5 Channels EI+ 256 > 254.79e5 6.44 6852 23562 5: MRM of 5 Channels EI+ 252 > 25 4.4e6 6.44 4478 23562 5: MRM of 5 Channels EI+ 252 > 226 6.44 5.85e5 23832 5.25 5.5 5.75 6. 6.25 6.5 6.75 7. 7.25 7.5 Figure 6. Ion chromatograms corresponding to 3 C 4 -benzo[a]pyrene (256>254; internal standard), benzo[a]pyrene (252>25 and 252>226) in a positive sample of smoked fish.

CONCLUSION REFERENCES This application note demonstrates that GC/MS/MS offers significant advantages for PAHs analysis in food products compared to classical GC/MS approaches. Several benefits can be achieved including;. Commission Recommendation of 4 February 25 on the further investigation into the levels of polycyclic aromatic hydrocarbons in certain foods. Off. J. Eur. Union L34 (25) 43. () Good sensitivity (down to ppt levels) in complex matrices using MRM. (2) Confident confirmatory analysis using multiple MRMs (up to 7 identification points per congener). (3) Excellent quantitative accuracy for analysis of a Certified Reference Material. With the greater sensitivity, accuracy and confidence in results afforded by the Quattro micro GC, the food safety analytical laboratory has the ability to generate more meaningful data and satisfy strict regulatory requirements. 2. Commission Regulation (EC) No 28/25 of 4 th February 25 amending Regulation (EC) No. 466/2 as regards polycyclic aromatic hydrocarbons. Off. J. Eur. Union L34 (25) 3. 3. 22/657/EC. Commission Decision of 4 th August 22 implementing Council Directive 96/23/EC concerning the performance of analytical methods and the interpretation of results. 4. Commission Directive 25//EC of 4 th February 25 laying down the sampling methods and the methods of analysis for the official control of the levels of benzo [a] pyrene in foodstuffs. Waters and Quattro micro GC are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 26 Waters Corporation Produced in the U.S.A. Sept 26 72899EN SE-PDF