Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Similar documents
System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Solid Phase Micro Extraction of Flavor Compounds in Beer

Maximizing Sample Throughput In Purge And Trap Analysis

Optimizing. Abstract: is standardd. procedures. altered to

Maximizing Production While Minimizing Costs

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Solid Phase Micro Extraction of Flavor Compounds in Beer

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Optimizing Standard Preparation

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Methanol Extraction of high level soil samples by USEPA Method 8260C

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Determination of Volatile Organic Compounds in Air

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Ed George and Anaïs Viven Varian, Inc.

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

EPA TO-17 Volatile Organic Compounds

2017 EPA Method Update Rule and EPA Method 624.1

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Analytical Trap Comparison for USEPA Method 8260C

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Target Compound Results Summary

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

CALA Directory of Laboratories

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

STANDARD OPERATING PROCEDURES

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Thermal Desorption Technical Support

Associates. Project No.: October 25, 2017

Copies: Dave Favero, RACER File

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

In-Tube Extraction Sample Preparation for Gas Chromatography

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Table 1 Soil and Groundwater Sampling Summary Table

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Volatile Organic Compounds in Water PBM

January 19, Dear Mr. Nightingale:

317 Elm Street Milford, NH (603) Fax (603)

Readiness thru Health

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

Headspace Technology for GC and GC/MS: Features, benefits & applications

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Study of Residual Solvents in Various Matrices by Static Headspace

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Date Issued: April 01, 2013 Expiration Date: June 30, 2013

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

C146-E209. Headspace Samplers. HS-20 Series

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

Transcription:

Optimal Conditions for USEPA Method 8260 Analysis Anne Jurek Introduction: Over the past decade, the need for environmental laboratoriess using purge and trap systems to report at or below the Method Detection Limit (MDL) has created manyy new challenges. As a result of the required lower detection levels, water management, analyte migration and carryover reduction have become a critical concern to meet linear calibration criteria. This paper will present the optimum purge and trap system parameters used to generate USEPA Method 8260B data. The conditions utilized will provide the necessary sensitivity, linearity and accuracy compliant to the method. In particular, the revolutionary advancements used to virtually eliminate carryover and manage the moisture will be highlighted. Analytical results including calibrations factors, method detection limits and reproducibility data will be presented. Discussion: As part of the process of purging the volatile target compounds from a water matrix, water vapor can also be carried along with the purge gas. This moisture can trigger interferencee in the GC/MS total ion chromatograms causing elevated detection levels, poor vacuum readings, and inconsistent analytical results. The management of this moisture prior to GC introduction is crucial in obtaining consistent and reliable data. The EST Encon Evolution Purge and Trap Concentrator employs a unique Moisture Reduction Trap (MoRT) to decrease the amount of water vapor introduced to the GC. Unlike other concentrators the Evolution positions the moisture management device before the analytical trap to remove the water from the purge gas as illustrated in Figure 1. Throughh the use of an 8-port valve, the Evolution s MoRT is excluded from the desorb pathway during the transfer of analytes to the GC as displayed in Figure 2. Figure 1: Purge Flow Path Figure 2: Desorb Flow Path The Evolution s desorb pathway minimizes deadd volume between the analytical trap and GC injection port to deliver superb peak resolutions and sensitivity over thee entire chromatogram as shown in Figure 3.

Another feature incorporated in the Encon Evolution which enables superior peak resolution is the Desorb Pressure Control (DPC) parameter. This featuree allows the trap to be pressurized to near to the GC inlet pressure prior to desorb. The balance of pressure between the trap and GC inlet assures the analytes are transferred in a tight band eliminating the surge in gas flow commonly associated with systems which do not balance pressure and use water management procedures during the critical desorb step. As detection limits are pushed lower, carryover within the purge and trap system has become of greater concern. In the past, efforts have been made to reduce carryover with new tubing materials andd programmable flow rates during the bake cycle, but little has been done to deal with primary cause of the carryover the spargee vessel itself. The Evolution employs a patented mode which heats the sparge vessel during the bake process as shown in Figure 4. Since this heat takes place during the normal Evolution bake cycle, no additional time is added to the overall cycle time. The result is lower carryover which aids in superior analytical results. A further problem associated with producing quality analytical data is internal standard reproducibility. The EST Centurion WS Autosampler eliminates this problem by utilizing a unique delivery mechanism consisting of no moving components. The volume programmable valve injects the internal standard directly into the sample stream rather than passing the sample through a fixed volume loop or valve. Experimental Results: Figure 3: Total ion chromatogram of thee 50ppb 8260B standard The EST Analytical Centurion WS Autosampler and Encon Evolution Purge and Trap Concentrator were interfaced to a GC/MSD in order to determine the optimal conditions required to achievee the desired chromatographic resolutions and sensitivity over the entire compound list in compliance with all USEPA Method 8260B criteria. The system conditions used to obtain the results are shown in Tables 1 and 2. The linear range of the system was first established by analyzing a nine point calibration curve (0.5-200ppb) according to the outlined procedures in the method. The internal standard concentration was held constant at 50ppb. The percent relative standard deviation () for each compound were below the method criteria of <15%. Thesee results are listed in Tablee 4.

Purge and Trap Concentrator EST Encon Evolution Trap Type Vocarb 3000 Valve Oven Temp. 130ºC Transfer Line Temp. 130ºC Trap Temp. 35ºC Moisture Reduction Trap (MoRT) Temp. 39ºC Purge Time 11 min. Purge Flow Dry Purge Temp. ambient Dry Purge Flow Dry Purge Time 1.0 min. Desorb Pressure Control On Desorb Pressure 12psi Desorb Time 1.0 min. Desorb Temp. 260ºC Moisture Reduction Trap (MoRT) Bake Temp. 230ºC Bake Temp 260ºC Sparge Vessel Bake Temp. 110ºC Bake Time 8 Bake Flow Purge and Trap Auto-Sampler EST Centurion WS Sample Size 5mL Internal Standard Volume 5µL Table 1: Purge and Trap Parameters GC/MS Inlet Inlet Temp. Inlet Head Pressure Mode Split Ratio Column Oven Temp. Program Column Flow Rate Gas Total Flow Source Temp. Quad Temp. MS Transfer Line Temp. Scan Range Scans Solvent Delay Agilent 6890A/5973 inert XL Split/Splitless 200ºC 17.311 psi Split 40:1 Rxi-624Sil MS 20m x 0.18mm I.D. 1µm film thickness 45º C hold for 1 min., ramp 18ºC/min to 220ºC, hold for 0.3 min. 0.8mL/min Helium 38.8mL/min 230ºC 150ºC 180ºC m/z 35-265 3.12 scans/sec 0.7 min Table 2: GC/MS Parameters

The precision and accuracy of the method was determined byy analyzing seven replicatee standards spiked at 50ppb. Method detection limits were established by running seven replicate standards at 1ppb. Both the MDLs and the precision and accuracy resultss are listed in Table 4. To demonstrate the Encon Evolution s ability to virtually eliminate carryover; two blank samples weree analyzed immediately following a 200ppb standard. The carryover results of the last four heavy compounds in the first blank after the 200ppb standard are displayed in Table 3. Compound % Carryover 1,2,4-Trichlorobenzene 0.20 Naphthalene 0.24 Hexacholorbutadiene 0.19 1,2,3-Trichlorobenzene 0.21 Table 3: Percent Carryover Results In order to the present the precision of the internal standard delivery system, the internal standard compound response was analyzed over the 10 replicate samples at a concentration of 1ppb. The precision is expressed as and the results are displayedd in Figure 5. 500000 Compound Area Count 400000 300000 200000 100000 0 1 2 3 4 5 6 7 8 9 10 Pentafluorobenzene, 1.84 1,4 Difluorobenzene, 1.81 Chlorobenzene d5, 1.69 1,4 Dichlorobenzene d4, 2.08% Run Number Figure 5: The Internal Standard Delivery Precision

Compound Ave. RF MDL Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethanee 1,1-Dichloroethene Acetone Iodomethane Carbon Disulfide Methylene Chloride MTBE trans-1,2-dichloroethene Vinyl acetate 1,1-Dichloroethane cis-1,2-dichloroethene 2-Butanone 2,2-Dichloropropane Bromochloromethane Chloroform 1,1,1-Trichloroethane 2-Chloroethylvinyletherr Carbon Tetrachloride 1,1-Dichloropropene Benzene 1,2-Dichloroethane Trichloroethene 1,2-Dichloropropane 10.97 5.01 5.11 6.70 4.67 5.77 6.80 7.22 12.65 10.42 3.17 5.43 5.05 7.01 6.04 4.59 6.83 2.13 8.24 3.16 4.04 7.15 8.21 3.91 1.18 6.42 3.89 2.66 0.309 0.614 0.529 0.291 0.387 0.685 0.400 0.189 0.592 1.270 0.514 1.501 0.576 1.372 1.015 0.496 0.367 0.721 0.348 0.910 0.812 0.431 0.677 0.762 2.301 0.775 0.384 0.364 0.18 0.22 0.11 0.19 0.25 0.12 0.07 0.06 0.21 0.17 0.06 0.10 0.14 0.18 0.11 0.07 Dibromomethane Bromodichloromethanee cis-1,3-dichloropropene Toluene trans-1,3-dichloropropane 4.30 3.49 3.31 2.26 3.17 0.210 0.436 0.540 0.905 0.462 0.14 0.14 50ppb 50ppb %Recovery Compound 90.76 4.82 Tetrachloroethene 91.19 4.744 1,3-Dichloropropane 96.94 5.055 Dibromochloromethane 98.74 4.70 2-Hexanone 97.93 4.744 1,2-Dibromoethane 98.28 5.18 Chlorobenzene 100.10 5.055 1,1,1,2-Tetrachloroethane 100.01 3.444 Ethylbenzene 102.45 4.01 Xylene (m+p) 90.28 5.57 Styrene 97.12 3.633 Xylene (o) 97.60 2.722 Bromoform 98.13 3.70 Isopropylbenzene 80.03 9.52 Bromobenzene 100.33 4.09 1,2,3-Trichloropropane 98.90 4.90 1,1,2,2-Tetrachloroethane 92.73 2.65 n-propylbenzene 84.26 5.47 2-Chlorotoluene 100.38 2.95 4-Chlorotoluene 98.38 3.78 1,3,5-Trimethylbenzene 98.22 4.87 tert-butylbenzene 99.80 2.47 sec-butylbenzene 102.37 5.20 1,2,4-Trimethylbenzene 98.15 5.02 1,3-Dichlorobenzene 96.51 4.27 1,4-Dichlorobenzene 99.60 2.69 Isopropyltoluene 101.51 3.32 1,2,-Dichlorobenzene 101.72 3.34 n-butylbenzene 1,2-Dibromo-3-102.60 3.14 chloropropane 101.43 2.95 1,2,4-Trichlorobenzene 99.47 2.96 Naphthalene 98.56 4.16 Hexachlorobutadiene 99.16 2.79 1,2,3-Trichlorobenzene Ave. 50ppb 50ppb MDL %Recovery RF 4.35 0.278 104.80 4.79 1.46 0.544 0.14 100.64 2.66 3.51 0.343 0.10 102.65 2.47 5.53 0.332 105.84 2.95 2.44 0.310 0.12 101.88 2.28 2.79 1.043 0.07 96.25 3.40 4.78 0.282 0.18 98.76 4.91 1.94 1.852 98.61 4.51 1.81 1.430 0.19 96.73 3.96 1.14 1.185 96.64 3.35 1.40 1.431 0.07 97.50 3.71 4.95 0.257 0.07 102.04 2.95 2.57 1.611 0.10 100.18 4.39 2.37 1.240 0.06 94.16 2.84 3.30 0.704 0.10 96.26 2.45 1.61 0.818 91.43 2.37 2.91 3.800 0.12 96.94 3.92 1.73 0.787 95.78 3.87 2.08 0.796 95.00 3.73 2.20 2.730 0.07 96.46 3.66 1.86 2.394 96.92 4.32 2.27 0.727 0.10 96.98 4.45 2.02 2.818 0.07 95.72 3.56 1.82 1.474 95.06 3.28 5.02 1.537 0.04 92.16 3.06 1.46 2.965 0.11 96.15 3.94 2.52 1.405 94.72 2.77 1.50 2.651 0.10 97.11 4.11 6.99 6 0.19 98.12 2.87 2.19 0.932 0.10 95.18 2.70 3.25 2.934 97.86 2.42 2.74 0.387 0.21 94.51 3.96 1.92 0.891 96.67 2.80 Table 4:, MDL and Precision and Accuracy Data Conclusion: The instrument configuration and operating conditions described above produced outstanding performance for USEPA Method 8260B. All quality control criteria were easily met or exceeded. In addition to its superior analytical performance, the EST Enconn Evolution and Centurion WS Autosampler also offer a number of other unique features and benefits. These include a patented mode which heats the sparge vessel during bake, the fiber optic foam sensor, and the electronic pressure control system with built in diagnosticss including an automatic leak check for guaranteed system integrity assurance.