Rapid, sensitive, and easy UHPLC-MS/MS analysis of fungicides in fruit juices with QuEChERS

Similar documents
LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Fast methods for the determination of ibuprofen in drug products

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Determination of urea in ultrapure water by IC-MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Analysis of Plasticizer Contaminants in Beverages and Milk using an Automated System Based on Turbulent-flow Chromatography Coupled to LC-MS/MS

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

LC-MS/MS Analysis of Phytocannabinoids and their

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Validation Report 18

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Fully automated QuEChERS clean-up and LC/MS-QQQ analysis of pesticides in fruits and vegetables.

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Definitive EtG/EtS LC-MS/MS Analysis:

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Automated and accurate component detection using reference mass spectra

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

Determination of trace anions in concentrated hydrofluoric acid

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

A New Polymer-based SPE Sorbent to Reduce Matrix Effects in Bioanalytical LC-MS/MS

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Liquid Chromatography Mass Spectrometry. Shimadzu Corporation, UK; 2 Phytocontrol, France; 3 Shimadzu France, France. Trimesium S + CH 3 ETU

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

Appendix II- Bioanalytical Method Development and Validation

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Determination of Benzo(a)pyrene in Sausage and Preserved Ham

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Highly Sensitive and Rugged GC/MS/MS Tool

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

Multi-residue analysis of pesticides by GC-HRMS

Transcription:

APPLICATIO OTE 21691 Rapid, sensitive, and easy UHPLC-MS/MS analysis of fungicides in fruit juices with QuEChERS Author Aaron Lamb and Brian King, Thermo Fisher Scientific, Runcorn, UK Keywords Food safety, Quantitation, Fungicides, Small molecule, Conazole, Carbendazim, Thiabendazole, Imazalil, Fenbuconazole, Dfenoconazole, Conazole, Fruit juice, Apple juice, Orange juice, UHPLC, QuEChERS, Vanquish GOLD UHPLC column, Vanquish Flex Binary UHPLC system, TSQ Endura triple quadrupole mass spectrometer Application benefits Quantitation of relevant fungicides in fruit juices down to below 1 ng/ml level without SPE workup Simplified extraction procedure with QUECHERS sample treatment only UHPLC-MS/MS analysis with sub-four minute gradients Goal To develop a rapid and sensitive assay for the quantitation of fungicides in fruit juices (orange and apple) with a separation gradient time of less than four minutes. To achieve a lower limit of quantification of 1 ng/ml for fungicides in fruit juice matrices while reducing cost per sample with a simple QuEChERS extraction. of all fungicides is required to be greater than 70 120 with an RSD of less than 20, in line with SATE guidelines. 1 Introduction Methyl 2-benzimidazole carbamate, most commonly known as carbendazim, is a widely used broad-spectrum benzimidazole fungicide and a decomposition product of benomyl. Carbendazim is used to control plant diseases in cereals and fruit, including citrus fruits, bananas, strawberries, pineapples, and pome fruits. Although not permitted for use to treat citrus

fruit in the USA and Australia, it is permitted in the EU, and European Regulation 559/2011 sets a limit for carbendazim and benomyl (sum of carbendazim and benomyl expressed as carbendazim) at 0.2 mg/kg in oranges. Incidences of MRL exceedance have been common in the EU, with 23 Rapid Alert otifications in 2011 for levels of carbendazim as high at 4 mg/kg in fruit, vegetables, and herbs from Africa, S. America, and Asia. 2 Orange juice from Brazil imported into the USA has been found to contain carbendazim, and an action limit of 0.01 mg/kg (10 ng/ml) has been applied by the FDA. 3 Many methods in widespread use for monitoring carbendazim have been developed for multi-residue determination of fungicides and employ a variety of sample preparation and cleanup techniques. In recent years the QuEChERS (Quick, Easy, Cheap, Effective, Rugged and Safe) method has become widely adopted for handling fruit such as oranges. These methods usually use this process as a sample pretreatment prior to the utilization of an instrumental technique, such as column switching, to give further cleanup or a precursor to a lengthy SPE extraction. For higher analytical throughput, recent methods have utilized sub-10 minute separation gradients. Legacy methods in continued to use longer run times. Existing methods can quantify these fungicides to >10 ng/ml with a multi-step sample preparation but none are able to use a simple extraction with a UHPLC-MS/MS separation gradient of less than 4 minutes to quantify at 1 ng/ml levels. Other related fungicides, the conazoles, are also used for treatment of fruit and they are more stable than carbendazim so have a longer-term impact. All these fungicides can be found in fruit that has been sprayed during cultivation and is then used to produce commercial juices and fruit drinks. These can vary widely in source or type (Figure 1). In these cases, the LLOQ is limited to 10 ng/ml for the conazole fungicides. To decrease this level further, the use of SPE and sample dry down is required. The retention of these less polar compounds is longer and the methods require acceleration to give a cycle time of 5 minutes. The method developed was designed to determine the levels of these fungicides in orange juice with a minimum amount of sample cleanup. Using QuEChERS with a H H O carbendazim OCH ³ fenbuconazole Cl H CH ³ Figure 1. Structures of carbendazim and conazole fungicides. thiabendazole range of commercial prepared juices, it was possible to develop a fast sample preparation protocol that allowed the analysis of these fungicides by UHPLC-MS/MS down to 1 ng/ml levels within four minutes without the need for further sample cleanup or post-extraction concentration. The Thermo Scientific Hypersep Dispersive SPE (QuEChERS) products are available in a range of formats to meet different application requirements. Both extraction tubes and dispersive SPE were used in this analysis to provide a straightforward cleanup of difficult matrices. By using the Thermo Scientific Vanquish Flex UHPLC platform, the detection of the specific fungicides was quantitated with excellent retention time reproducibility with RSD of 0.05 compared to typical values of 0.1 for competitors. The Thermo Scientific Hypersil GOLD VAQUISH column has an endcapped ultra-pure porous silica material giving exceptional peak shape and resolution for HPLC and LC-MS. The Vanquish GOLD column and the capabilities of the Vanquish Flex UHPLC platform allow a combination of high separation efficiency and fast analysis to give increased sample throughput compared to methods with multi-step liquid extraction and SPE steps. The Thermo Scientific TSQ Endura triple quadrupole mass spectrometer has ultrafast selected-reaction monitoring (SRM) of 500 SRM/s, with up to 30,000 definable SRMs. This enables quantification of more compounds in less time. The ion optics, RF-lens, ion beam guide with neutral blocker, and quadrupole mass filter combine to reduce noise and increase sensitivity for enhanced quantitative performance. S O H ² C O Cl O imazalil difenoconazole O Cl Cl Cl 2

All chromatography and MS data was processed using Thermo Scientific Chromeleon CDS software, which provides chromatography labs with compliance-ready data management, unified instrument control, and simplified analysis and data reporting for chromatography and mass spectrometry. The method developed was able to exceed the sensitivity of a competitive QuEChERS method by giving 1 ng/ml sensitivity for all the fungicides investigated. This was achieved without a separate SPE step and any dry down concentration. The runtime of < 4 minutes compared to 10 minutes for an online method and 7.2 minutes for a comparable UHPLC method. Experimental Recommended consumables Deionized water, 18.2 MΩ cm resistivity from Thermo Scientific Smart2Pure system (P/ 5012984) Fisher Scientific Optima UHPLC-MS grade methanol (P/ A458-1) Fisher Scientific Optima UHPLC-MS grade acetonitrile (P/ A956-1) Fisher Scientific Analytical grade formic acid (P/ F/1900/PB08) Fisher Scientific Analytical grade ammonia (P/ A/3295/PB05) Hypersil GOLD VAQUISH C18 UHPLC column 50 mm 2.1 mm, 1.9 µm (P/ 25002-052130-V) Thermo Scientific Virtuoso Vial Identification System (P/ 60180-VT100) Virtuoso vial, clear 2 ml kit with septa and cap (P/ 60180-VT402) HyperSep Dispersive SPE (QuEChERS) 50 ml Tube 25-pk (P/ 60105-316) HyperSep Dispersive SPE Clean Up 15 ml Tube 25-pk (P/ 60105-327) Thermo Scientific 22 ml Storage Vials Kit 200-pk (P/ 22-CV-CP) 50 ml Centrifuge tubes (P/ 05-539-13) 15 ml Centrifuge tubes (P/ 11849650) Standards The reference grade standards used were purchased from a reputable supplier: Carbendazim Thiabendazole Imazalil Fenbuconazole Difenoconazole (internal standard) Samples All juices were purchased from a local supermarket. Sample handling equipment Benchtop centrifuge with 50 ml and 15 ml dual tube rotor Vortex mixer Stock solution preparation Standard stock solutions Separate stock solutions, fungicide standards, and the internal standard were prepared at 10 mg/ml in 100 glacial acetic acid. Calibration stock solutions A mixed fungicide solution was then prepared at 10 µg/ml from the fungicide 10 mg/ml stock solution in a 5:95 methanol/water 0.1 formic acid solution. From this mixed standard, spiking solutions were prepared at 1.0, 2.5, 5, 10, 25, 50, and 100.0 ng/ml. Quality control (QC) stock solutions A separate preparation of the mixed standard was used to make quality control spiking solutions (3.0, 40.0, and 80.0 ng/ml) in in a 5:95 methanol/water 0.1 formic acid solution. Internal standard stock solution An internal standard solution spiking solution was prepared at 1,000 ng/ml Standard and QC preparation Calibration curve preparation The appropriate calibration standard (2.5 ml) was added to the juice and made up to volume in a 50 ml volumetric flask. Three 15 ml sub-aliquots were taken of each QC and transferred to separate labeled vials and to these solutions 750 µl of the internal standard was added and the vials were mixed well. Fisherbrand Adapt-a-Rack (blue) (P/ 15340370) 3

Extraction procedure The following QuEChERS procedure was followed: Transfer 15 ml of sample to 50 ml QuEChERS extraction tubes + 15 ml acetonitrile. Vortex until homogeneous and leave to stand for 10 minutes and vortex once more. Centrifuge at 7,400 rpm for 2 minutes. Transfer 10 ml of top layer to dispersive SPE QuEChERS tubes, vortex until homogeneous, and then centrifuge at 7,400 rpm for 2 minutes. Dilute the supernatant 250 µl + 750 µl of 5 methanol/95 0.1 ammonia solution. Samples/standards and QCs The appropriate QC standard (1.25 ml) was added to the juice, 750 µl of the internal standard was added, and the vials were mixed well. Matrix blank preparation The dilution solution (1.25 ml) was added to 15 ml of juice matrices and made up to volume in a 25 ml volumetric flask. The dilution solution was 95:5:0.1 v/v/v water/methanol/ammonia. A 15 ml sub-aliquot was taken and transferred to separate labeled vials and mixed well. Mobile phase blanks To 950 µl of acetonitrile, 50 µl of 95:5:0.1 v/v/v water/ methanol/ammonia was added. Then, 250 µl of this solution was diluted with 750 µl of 95:5:0.1 v/v/v water/ methanol/ ammonia to produce the blank solution. Matrix post-spiked standard preparation Matrix blanks (900 µl) were transferred to separate autosampler vials. These were spiked with 50 µl of the appropriate QC standard as well as 50 µl of the internal standard. Then, 250 µl of this solution was diluted with 750 µl of 95:5:0.1 v/v/v water/methanol/ammonia to produce the matrix-matched post-spiked solutions. Separation conditions Instrumentation Analyses were performed using a Vanquish Flex Binary UHPLC system consisting of: System Base Vanquish Flex (P/ VF-S01-A) Binary Pump F (P/ VF-P10-A-01) Split Sampler FT (P/ VF-A10-A) Column Compartment H (P/ VH-C10-A) Active Pre-heater (P/ 6732.0110) Column: Hypersil GOLD VAQUISH C18 UHPLC column 50 mm 2.1 mm 1.9 µm (P 25002-052130-V) Mobile phase A: Water/0.5 formic acid Mobile phase B: Methanol/0.5 formic acid Gradient: Table 1 Table 1. LC gradient conditions. Time (min) A B 0 90 10 4 5 95 6 5 95 6.1 90 10 10 90 10 Flow rate: 0.4 ml/min Column temperature: 40 C Column thermostatting mode: Still Air Pre-heater: On 32 C Injection details: 20 µl Injection wash solvent: 20:80 methanol/water (v/v) MS conditions Instrumentation Mass analysis was performed with the TSQ Endura MS. Instrumental conditions are listed in Table 2 and the compound transition details are listed in Table 3. 4

Table 2. MS source and analyzer conditions. Ionization conditions Polarity HESI Positive Spray voltage 3500 V Vaporiser temperature 400 C Sheath gas pressure 50 Arb Aux gas pressure 15 Arb Capillary temp 350 C Collision pressure 1.5 mtorr Scan time 0.25 s Q1 (FWHM) 0.7 Q3 (FWHM) 0.7 Data processing The Thermo Scientific Chromeleon 7.2 SR4 Chromatography Data System was used for data acquisition and analysis. Table 3. Compound transition details. Compound Carbendazim Thiabendazole Imazalil Fenbuconazole Difenoconazole Precursor (m/z) 192.1 202.1 297.1 337.1 406.1 Product (m/z) (1) 160.1 175.1 159.0 125.0 251.0 Confirmation (m/z) (2) 132.1 131.2 176.1 194.0 188.1 Collision energy 1 (V) 33 36 22 29 28 Collision energy 2 (V) 20 28 26 19 46 Results and discussion Extraction of juices Methods for the analysis of fungicides in fruit juices sometimes utilize a three-step process of non-optimized QuEChERS matrix cleanup, dry down, and reconstitution followed by SPE. This is time-consuming and expensive in terms of labor and material expenditure. In this modified QuEChERS method, we demonstrate a simple optimized QuEChERS matrix cleanup followed by UHPLC-MS/MS analysis. The use of the salting out step with a mix of magnesium sulfate, sodium chloride, and sodium citrate was shown to provide an initial cleanup that was refined by the use of a selective dispersive SPE step to remove matrix contaminants prior to dilution and injection into the UHPLC system. This eliminated the dry down/reconstitution and further SPE steps, saving time and money. give the most reproducible peak shape while maintaining method sensitivity. This was maintained for all batches. Dilution of the acetonitrile QuEChERS extracts in 95:5:0.1 v/v/v water/methanol/ammonia rather than mobile phase starting conditions was found to improve peak shape for the earlier eluting peaks. The basic solution gave better chromatography and consistent signal response. Injection of the calibration and QC matrix standards showed excellent retention time reproducibility with of between 0.05 and 0.16 for the four fungicides. All the standards retentions showed excellent stability on the programmed gradient for 22 samples (Table 4). Method optimization During method development, the injection volume of QuEChERS extract was initially set at 50 µl but due to the high acetonitrile concentration the early eluting components showed peak tailing and broadening even after dilution. The injection volume was evaluated at levels between 50 µl and 5 µl. A volume of 20 µl was found to 5

Table 4. Peak summary data for 22 replicates of a fungicide standard mix at 40 C and a flow rate of 0.4 ml/min. Parameter Carbendazim Thiabendazole Imazalil Fenbuconazole Difenoconazole (IS) Maximum (min) 1.477 1.632 2.655 3.223 3.425 Average (min) 1.474 1.630 2.653 3.219 3.421 Minimum (min) 1.469 1.628 2.651 3.215 3.420 SD 0.002 0.002 0.002 0.002 0.002 RSD 0.16 0.13 0.08 0.06 0.05 QC solutions were 3 ng/ml, 40 ng/ml, and 80 ng/ml concentrations and representative chromatograms are shown below in Figures 2 and 3, which show separation of the four compounds from the internal standard. Figure 2. Representative chromatogram for Low QC batch with IS and extracted ion chromatograms. Figure 3. Representative chromatogram for Mid QC batch with IS. 6

UHPLC Calibration Linearity The calibration linearity was investigated by carrying out injections of matrix-matched calibration standards for 1.0 to 100 ng/ml. The results from the calibration line accuracy versus the true value were within 100 ± 20 for all compounds across the range (Table 5). The matrix-matched calibrations were prepared in duplicate. When plotting the calibration curve, the coefficient of determination was >0.997 for all fungicides (Figure 4). Table 5. Calibration line accuracy of true concentration. Accuracy relative to true amount Compound 1 2.5 5 10 25 50 100 Carbendazim 82.2 95.3 95.2 100.7 94.7 98.6 100.9 Thiabendazole 97.5 90.3 91.6 92.1 94.8 97.2 103.6 Imazalil 117.0 90.8 89.8 104.6 102.4 93.9 103.1 Fenbuconazole 113.0 98.7 89.8 98.0 94.4 88.2 107.4 Component Peak ame: Carbendazim Ret. Time: 1.477 min Coefficient of determination: 0.99993 Calibration Line Area 550 500 400 300 200 100 0 0 10 20 30 40 50 60 70 80 90 100 110 120 Amount Peak ame: Thiabendazole Ret. Time: 1.632 min Coefficient of determination: 0.99971 Area 180 160 140 120 100 80 60 40 20 0 0 10 20 30 40 50 60 70 80 90 100 110 120 Amount Peak ame: Imazalil Ret. Time: 2.651 min Coefficient of determination: 0.99953 Area 80.0 70.0 60.0 50.0 40.0 30.0 20.0 10.0 0.0 0 10 20 30 40 50 60 70 80 90 100 110 120 Amount Peak ame: Fenbuconazole Ret. Time: 3.219 Coefficient of determination: 0.99707 Area 80.0 70.0 60.0 50.0 40.0 30.0 20.0 10.0 0.0 0 10 20 30 40 50 60 70 80 90 100 110 120 Amount Figure 4. Calibration lines for four fungicides. 7

Quality control samples accuracy and precision The QC level accuracy for six independent preparations at three different levels was found to be in the range of 86.9 to 108.7, within the guidelines of ±15 of true values for all samples with a range of 1.2 to 9.4 RSD showing good precision (Table 6). Table 6. QC accuracy of true concentration n=6 preparations in orange juice matrix. Compound QCL at 3 ng/ml (n=6) at QCL QCM at 40 ng/ml (n=6) at QCM QCH at 80 ng/ml (n=6) at QCH Carbendazim 108.7 5.7 100.6 3.9 99.1 3.9 Thiabendazole 94.4 9.4 90.3 4.1 99.7 3.0 Imazalil 86.9 5.5 100.9 6.4 101.7 2.4 Fenbuconazole 92.8 1.2 97.3 4.1 101.9 1.5 Six aliquots of orange juice were prepared at the three QC levels and processed in parallel. The potential for using the procedure for other juice matrices was explored by processing an equivalent apple juice. The recovery for both was measured against six over-spiked samples in the original blank matrix and determined to be within the 70 120 w/v values and with repeatability of RSD <20, in line with SATE/11945/2015 guidelines. from the orange juice was between 81.1 and 91.6, and the recovery from the apple juice was similar at between 80.0 and 90.1 (Tables 7 and 8). Thiobendazole gave slightly lower recovery values compared to the other fungicides in both matrices but was within the guidance criteria. The calculated recovery for the fungicides in both matrices was in the range of 2.1 to 12, with the means between 4 and 6. This gave very reproducible recovery of all fungicides from either matrix (Figures 5 and 6). It was considered that the method could be applied to those juices which were most commonly available. Table 7. Fungicide recovery from orange juice. Compound at QCL at QCL at QCM at QCM at QCH at QCH Average Carbendazim 90.7 3.5 77.2 3.1 84.4 6.4 84.1 Thiabendazole 87.3 7.8 76.0 2.5 80.0 7.3 81.1 Imazalil 87.0 7.1 85.4 4.5 90.8 2.8 87.7 Fenbuconazole 94.3 12.0 90.2 2.1 90.2 3.5 91.6 Table 8. Fungicide recovery from apple juice. Compound at QCL at QCL at QCM at QCM at QCH at QCH Average Carbendazim 85.6 7.2 82.1 3.2 84.3 3.4 90.1 Thiabendazole 74.2 4.3 80.0 2.2 80.9 3.7 82.8 Imazalil 84.4 7.1 85.4 3.6 84.8 4.5 84.4 Fenbuconazole 82.1 11.7 81.8 3.7 84.4 2.9 87.0 8

100 Table 9. Suppression of signal from orange and apple juice. 95 Orange Signal Suppression (Matrix Effects) 90 85 Carbendazim 16.9 Thiabendazole 32.6 Imazalil 3.4 Fenbuconazole 1.5 80 75 Apple Signal Suppression (Matrix Effects) Carbendazim 3.2 70 Carbendazim Thiabendazole Imazalil Fenbuconazole Fungicide Figure 5. and RSD comparisons for four fungicides in orange juice matched-matrix QC samples. Thiabendazole 12.2 Imazalil 1.3 Fenbuconazole -0.6 50.0 100 45.0 40.0 95 35.0 90 85 Supression 30.0 25.0 20.0 15.0 80 10.0 75 5.0 70 Carbendazim Thiabendazole Imazalil Fenbuconazole Fungicide Figure 6. and RSD comparisons for four fungicides in an apple juice matrix-matched QC samples. 0.0 Carbendazim Thiabendazole Imazalil Fenbuconazole Fungicide Figure 7. Orange juice matrix suppression at 40 ng/ml level calculated 1-(Matrix OS)/(Solution standard). 50.0 Matrix effects The matrix effect was determined comparing the MS component signals at MIDQC against a solution MIDQC. The orange juice matrix was found to have a suppression of up to 32 on signal for thiabendazole, but imazalil and fenbuconalzole showed little suppression at these levels. Overall there was less suppression shown by the apple juice matrix (Table 9 and Figures 7 and 8). This was most apparent with carbendazim, which was reduced to less than 5 in apple juice compared to 17 in the orange juice. The levels for samples introduced in the matrix from other applications were over 35 for thiabendazole and could only be reduced by further SPE cleanup. 5 Supression 45.0 40.0 35.0 30.0 25.0 20.0 15.0 10.0 5.0 0.0 Carbendazim Thiabendazole Imazalil Fenbuconazole Fungicide Figure 8. Apple matrix suppression at 40 ng/ml level calculated 1-(Matrix OS)/(Solution standard). 9

The matrix effects were less with apple juice compared to orange juice, which had some effects on the thiabendazole and carbendazim. As these are early eluting peaks, it would be more likely that polar matrix components will be present as part of the injection solution. Detection Limits The calibration curve for all four fungicides was confirmed for a linear range of 1 ng/ml to 100 ng/ml. Below 1 ng/ml, replicates of 0.5 ng/ml standard in orange juice matrix were run and the presence of all five fungicides confirmed at this level. Table 10. MDL for seven replicate injections at 1 ng/ml level. To determine the method detection limit for the components this was calculated by the equation 5 : Detection limit = S t (n 1, 1 - α = 0.99) The symbol S represents the standard deviation of replicate analyses, n represents the number of replicates, t (n - 1, 1 - α = 0.99) represents the Student s t value for the 99 confidence level with n 1 degrees of freedom. Using this value and the response from seven replicates of the 1 ng/ml standard calibration solution, it was determined that the method detection limit for each fungicide could be quantified at levels below 0.5 ng/ml (Table 10). Carbendazim ng/ml Thiabendazole ng/ml Imazalil ng/ml 0.325 0.231 0.399 0.381 Fenbuconazole ng/ml The extracted ion chromatograms showed sufficient response for quantification and confirming peaks. The fenbuconazole confirmation peak showed a small doublet at this low 1 ng/ml level but retained 100 confirmation and calibration (Figure 9). Figure 9. Extracted ion chromatograms at 1 ng/ml. 10

Comparative results For comparison, application alternatives to manual SPE have been suggested by direct injection onto turbulent flow columns 2 or by means of automation. 4 These methods have been applied to single fungicides and require expensive additional hardware. The analysis of three fungicides with SPE and LC-MS/ MS was shown to give an LLOQ limit of 2 ng/ml for carbendazim and two other conazole fungicides but required extensive SPE followed by dry down and concentration steps to meet the detection limits. 6 The application of QuEChERS alone has been proposed before 5 but was able to reach lower limits of 1 ng/ml only with carbendazim and showed 10 ng/ml with other conazole fungicides in a similar matrix. This modified QuEChERS method required only a bench centrifuge for processing, and consumable use was limited to the solvents used for the extractions from the 50 ml tubes. Tube racking was found to make the workflow much easier to control for the batch process. The method used here meets the major requirements of the EC guidelines on analytical quality control for pesticides residues analysis in food and drink 1 and has shown applicability within related matrices. Conclusions A lower limit of quantification of 1 ng/ml for carbendazim, imazalil, and fenbuconazole from a fruit juice matrix with only a QuEChERS sample preparation procedure was determined. Rapid separation of the target fungicides of this Application ote in less than four minutes was achieved by using 50 mm columns packed with Vanquish GOLD 1.9 µm C18 materials. The QuEChERS method was used without further sample preparation, such as SPE, evaporation, and resuspension, to give lower costs and faster workup. of all fungicides tested in this Application ote and using the modified QuEChERS method was greater than 75 with a RSD of <5 for the QC samples. References 1. Guidance document on analytical quality control and method validation and method validation procedures for pesticides residue analysis in food and feed. European Commission SATE/11945/2015 2015 rev 0. 2. Hollosi, L.; Bousova, K.; Ates, E.; Mittendorf, K. Thermo Fisher Scientific Application ote A52252: Determination of Carbendazim and Benomyl Residues in Oranges and Orange Juice by Automated Online Sample Preparation Using TLX-LC-MS/MS, 2012. 3. GB 14870-1994: Maximum Residue Limits of Carbendazim in Foods; Standardization Administration of China (SAC), General Administration of Quality Supervision, Inspection and Quarantine of the People s Republic of China: Beijing, 1994. 4. Fernandes V.C.; Domingues V.F.; Mateus,. Determination of Pesticides in Fruit and Fruit Juices by Chromatographic Methods. J. Chromatogr. Sci., 2011, 49, 715-730. 5. Xiongfeng, H.; Qun, X.; Rohrer, J. Thermo Fisher Scientific Application ote A 1067: Determination of Carbendazim in Orange Juice, 2016. 6. Dioumaeva, I. Agilent Application ote: 5991-0051E: LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120, 2012. 7. Young, M.S.; van Tran, K.; Shia, J.C.; Burgess, J.A.; Mullen, L.; Fountain, K.J. Waters Corporation Application ote 720004457E: QuEChERS Sample Preparation for LC/ MS and LC/UV Determination of Carbendazim and Other Conazole Fungicides in Orange Juice, 2012. Find out more at thermoscientific.com/quechers For Research Use Only. ot for use in diagnostic procedures. 2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. ot all products are available in all countries. Please consult your local sales representatives for details. A21691-E 0917S