DPD Test N Tube Method *

Similar documents
CHLORINE, TOTAL (0 to 4.00 mg/l)

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

Chlorine, Free and Total, High Range

Chlorine, Free and Total, High Range

Chlorine, Free and Total, Low Range

NITROGEN, AMMONIA, High Range, Test N Tube

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Chloramine (Mono) and Nitrogen, Free Ammonia

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Manganese III Digestion Method * (with chloride removal)

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Mercaptoacetic Acid Method*

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

NITRITE, Low Range (0 to mg/l NO 2- -N)

Method to 0.50 mg/l NH 3 N Powder Pillows

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Method to 0.50 mg/l NH 3 N Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

NITROGEN, TOTAL, HR, Test N Tube

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Nitrogen, Total Inorganic

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

DR/4000 PROCEDURE. IRON, Total

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

SILICA, Low Range (0 to 1.60 mg/l)

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

NITROGEN, TOTAL, Test N Tube * Method 10071

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

PAR Method Method to 2.0 mg/l Pb TNTplus 850

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Oxygen Demand, Chemical

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

p-dimethylaminobenzaldehyde Method*

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Trihalomethane Formation Potential (THMFP)

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Instruction Manual. Waterproof Portable Colorimeter C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105. Part of Thermo Fisher Scientific

Understanding Chlorine Measurement. Rebecca Luedee Environmental Sales

Oxygen Demand, Chemical

Quaternary Ammonium Compounds

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

ALLOWAY METHOD OUTLINE

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

RAPID LIQUID TESTING. Hach Company, 1996 All rights reserved. Printed in the U.S.A dld/dp ed

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

TPH (Total Petroleum Hydrocarbons)

DR/4000 PROCEDURE. HARDNESS, Total

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Parameter Method Range # of Tests Code Page. Acidity (as % Oleic acid) titration % acidity 6 HI

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Bromine 10 T mg/l Br 2 DPD

LOW-RANGE TOTAL CHLORINE

Ammonia Gas Sensing Combination Electrode

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

RAPID KJELDAHL BENCHNOTES

Reagents for Chemical Test Kits

SMART Spectro. L Mott TEST INSTRUCTIONS. v MN. Spectro pho to m eter

Dissolved Oxygen. Water Quality Test Kit Instruction Manual Code 7414/5860

Nitrogen total LR VARIO TT mg/l N b) Persulphate Digestion

#18 Analysis of Water (Chlorine, Phosphorus, Fluoride, and Color)

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

The content of this operation manual is protected by copyright laws. No part of this operation manual may be reproduced in any form.

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Thermo Scientific Orion AQUAfast AQ3070 Colorimetry Meter. User Guide

Name Period Date. Lab 9: Analysis of Commercial Bleach

Ammonia Gas Sensing Combination Electrode

Polychlorinated Biphenyls (PCB) in Soil

Transcription:

CHLORINE, FREE (0 to 5.00 mg/l) DPD Test N Tube Method * Method 10102 For water, wastewater, and seawater 1. Enter the stored program number for Test N Tube free chlorine (Cl 2 ). Press: PRGM The display will show: PRGM? Note: For most accurate results, perform a Reagent Blank Correction using deionized water (see Section 1). 2. Press: 10 ENTER The display will show mg/l, Cl2 and the ZERO icon. 3. Insert the COD/TNT Adapter into the cell holder by rotating the adapter until it drops into place. Then push down fully to insert it. Note: For increased performance, a diffuser band covers the light path holes on the adapter. Do not remove the diffuser band. 4. Fill an empty Test N Tube vial with sample (the blank). Note: Fill to the top of the Hach logo oval mark. Note: Samples must be analyzed immediately and cannot be preserved for later analysis. * Adapted from Standard Methods for the Examination of Water and Wastewater 149

CHLORINE, FREE, continued 5. Wipe the outside of the blank vial with a towel. Note: Wiping with a damp cloth followed by a dry one removes fingerprints and other marks. 6. Place the blank in the adapter. Push straight down on the top of the vial until it seats solidly into the adapter. Note: Do not move the vial from side to side as this can cause errors. 7. Cover the vial tightly with the instrument cap. 8. Press: ZERO The cursor will move to the right, then the display will show: 0.00 mg/l Cl2 Note: If Reagent Blank Correction is on, the display may show limit. See Section 1. 9. Remove the cap from a Free Chlorine DPD- TNT tube. Add 10 ml of sample. Note: Fill to the top of the Hach logo oval mark. Note: A pink color will develop if chlorine is present. 10. Cap and invert at least 10 times to dissolve the powder. This is the prepared sample. Note: Use slow, deliberate inversion for complete recovery. Ten inversions should take at least 30 seconds. One inversion equals turning the vial upside down, then returning it to an upright position. 11. Within 30 seconds after mixing, wipe the prepared sample vial with a towel, then place it in the adapter. Push straight down on the top of the vial until it seats solidly into the adapter. Note: Do not move the vial from side to side as this can cause errors. 12. Cover the vial tightly with the instrument cap. Press: READ The cursor will move to the right, then the result in mg/l free chlorine will be displayed. 150

Sampling and Storage Analyze samples for chlorine immediately after collection. Free chlorine is a strong oxidizing agent and is unstable in natural waters. It reacts rapidly with various inorganic compounds and more slowly oxidizes organic compounds. Many factors, including reactant concentrations, sunlight, ph, temperature, and salinity influence decomposition of free chlorine in water. Avoid plastic containers since these may have a large chlorine demand. Pretreat glass sample containers to remove any chlorine demand by soaking in a dilute bleach solution (1 ml commercial bleach to l liter of deionized water) for at least 1 hour. Rinse thoroughly with deionized or distilled water. If sample containers are rinsed thoroughly with deionized or distilled water after use, only occasional pretreatment is necessary. A common error in testing for chlorine is obtaining an unrepresentative sample. If sampling from a tap, let the water flow for at least 5 minutes to ensure a representative sample. Let the container overflow with the sample several times, then cap the sample containers so there is no headspace (air) above the sample. Perform the analysis immediately. Accuracy Check Standard Additions Method a) Snap the top off a HR Chlorine PourRite Ampule Standard Solution. b) Use a TenSette Pipet to add 0.1 ml of the standard to the reacted sample (this is the spiked sample). Swirl to mix. c) Analyze the spiked sample, beginning at Step 8 of the procedure. d) Calculate the concentration of mg/l chlorine added to the sample: 0.1 vol. standard added mg/l chlorine added = ( ) Label value ( mg L Cl 2) ---------------------------------------------------------------------------------------------------------------------------------------- 10.1( sample + standard volume ) e) The spiked sample result (step c) should reflect the analyzed sample result + the calculated mg/l Cl 2 added (step d). f) If this increase does not occur, see Standard Additions, Section 1 for more information. 151

Method Performance Precision In a single laboratory using a standard solution of 2.53 mg/l chlorine and two representative lots of reagents with the instrument, a single operator obtained a standard deviation of ±0.14 mg/l chlorine. Estimated Detection Limit (EDL) The estimated detection limit for program 10 is 0.03 mg/l Cl 2. For more information on derivation and use of Hach s estimated detection limit, see Section 1. Interferences Acidity Alkalinity Bromine Interfering Substance Chlorine Dioxide Chloramines, organic Interference Level and Treatment Greater than 150 mg/l CaCO 3. May not develop full color or color may fade instantly. Neutralize to ph 6-7 with 1 N sodium hydroxide. Determine amount to be added on separate sample aliquot, then add the same amount to the sample being tested. Correct for volume addition (See Section 1, Correcting for Volume Additions in the DR/800 Series Procedures Manual). Greater than 250 mg/l CaCO 3. May not develop full color or color may fade instantly. Neutralize to ph 6-7 with 1 N sulfuric acid. Determine amount to be added on separate sample aliquot, then add the same amount to the sample being tested. Correct for volume addition (See Section 1 Correcting for Volume Additions). May interfere Hardness No effect at less than 1,000 mg/l as CaCO 3 Iodine Manganese, oxidized (Mn 4+, Mn 7+ ) or Chromium, oxidized (Cr 6+ ) 1. Adjust sample ph to 6-7. 2. Add 3 drops potassium iodide (30 g/l) to a 25-mL sample. 3. Mix and wait one minute. 4. Add 3 drops sodium arsenite (5 g/l) and mix. 5. Analyze 10 ml of the treated sample as described in the procedure. 6. Subtract the result from this test from the original analysis to obtain the correct chlorine concentration. 152

Monochloramine Ozone Peroxides Interfering Substance Extreme sample ph and highly buffered samples For conventional free chlorine disinfection (beyond the breakpoint), typical monochloramine concentrations are very low. If monochloramine is present in the sample, its interference in the free chlorine test depends on the sample temperature, relative amount of monochloramine to free chlorine, and the time required to do the analysis. Typical interference level of monochloramine in the free chlorine test are listed below (as mg/l Cl 2 ). May interfere Interference Level and Treatment NH 2 Cl Sample Temp. C ( F) as Cl 2 5 (40) 10 (50) 20 (68) 30 (83) 1.2 mg/l +0.15 +0.19 +0.30 +0.29 2.5 mg/l 0.35 0.38 0.55 0.61 3.5 mg/l 0.38 0.56 0.69 0.73 Adjust to ph 6-7. See Interferences, Section 1. Pollution Prevention and Waste Management Samples treated with sodium arsenite for manganese or chromium interferences will be hazardous wastes as regulated by Federal RCRA for arsenic (D004). See Section 3 for more information on proper disposal of these materials. Summary of Method Chlorine in the sample as hypochlorous acid or hypochlorite ion (free chlorine or free available chlorine) immediately reacts with DPD (N,N-diethyl-p-phenylenediamine) indicator to form a magenta color which is proportional to the chlorine concentration. 153

REQUIRED REAGENTS Quantity Required Description Per Test Unit Cat. No. Test N Tube DPD Free Chlorine Reagent...1 vial... 50/pkg... 21055-45 Test N Tube Vials...1 vial... 6/pkg... 22758-06 REQUIRED APPARATUS Caps, white...1 cap... 6/pkg... 22411-06 COD/TNT Adapter... 1... each... 48464-00 OPTIONAL REAGENTS Chlorine Standard Solution, PourRite ampule, 50-75 mg/l, 2 ml... 20/pkg... 14268-20 Potassium Iodide Solution, 30 g/l... 100 ml * MDB... 343-32 Sodium Arsenite, 5 g/l...100 ml* MDB... 1047-32 Sodium Hydroxide Standard Solution, 1.000 N...100 ml* MDB... 1045-32 Sulfuric Acid Standard Solution, 1.000 N...100 ml* MDB... 1270-32 OPTIONAL APPARATUS Beaker, 50 ml... each... 500-41H ph Meter, sension 1, portable, with electrode... each... 51700-10 ph Paper, ph 1 to 11 ph... 5 rolls/pkg... 391-33 Pipet, TenSette, 0.1 to 1.0 ml... each... 19700-01 Pipet Tips, for 19700-01 TenSette Pipet... 50/pkg... 21856-96 Pipet Tips, for 19700-01 TenSette Pipet... 1000/pkg... 21856-28 PourRite Ampule Breaker... each... 24846-00 Test Tube Rack... each... 18641-00 * Marked Dropper Bottle - contact Hach for larger sizes. 154