Determination of Polybrominated Diphenylethers (PBDE) in Sediment and Sewage Sludge Application

Similar documents
GC Analysis of Polybrominated Flame Retardants Application

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

ISO INTERNATIONAL STANDARD

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Fast and Accurate Analysis of PBDEs in a Single Run, Including

Fast and Accurate Analysis of PBDEs in a Single Run, Including BDE-209

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

PosterReprint INTRODUCTION

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Author. Abstract. Introduction

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

ENVIRONMENTAL analysis

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

APPLICATION NOTE Authors Jane Cooper, 1 Jordi Parera, 2 Esteban Abad, 2 Richard Law,¹ and Cristian Cojocariu 1

Bromierte Flammschutzmittel die erfolgreiche kritische Trennung von BDE 49/71 und BDE209

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

U.S. EPA Method 8270 for multicomponent analyte determination

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Modernizing the USP Monograph for Acetaminophen

Methods of Analysis of Polychlorinated Biphenyl Congeners Using an Application-Specific Capillary GC Column

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

HRMS PBDE Analysis on a Thermo DFS. Dave Hope, Pacific Rim Laboratories Inc

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

Technical Report. Abstract: Pu Wang 1, Huizhong Sun 1, Qinghua Zhang 1, Feifei Tian 2, Lei Cao 2

Application Note. Abstract. Authors. Introduction

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Food analysis. Solutions for Your Analytical Business Markets and Applications Programs

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

GC/MS/MS Dioxin and Dioxin-like PCBs in Feed and Food. Jessica Westland, LSAG Applications Chemist

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Evaluation of Capillary Columns for General Performance Parameters

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Learning centre Management Chemicals Issues of Importance for Sustainable Development. CSD-19 May 3, Beijing, April 2010

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

A New Approach to the Analysis of Chlorinated Paraffins by Gas Chromatography Quadrupole Time of-flight Mass Spectrometry

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Sulfotepp impurities in Chlorpyrifos EC formulations

Soil and Sludge. Determination of selected polybrominated diphenylethers (PBDEs) by gas chromatography-mass spectrometry (GC-MS).

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

SUPELCO. The Analysis of all 209 PCB Congeners on the SPB -Octyl and MDN -5S Capillary Columns. Katherine K. Stenerson and Leonard M.

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Determination of EU-Banned Disperse Dyes by LC/MSD TOF Application

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

BACKGROUND What are PBDEs? Regulatory Status of PBDEs in Canada

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Benzoylecgonine in Urine by SAMHSA GC/MS

The Analysis of Organophosphate Pesticides by LC/MS Application

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

The Analysis of Fluoroquinolones in Beef Kidney Using HPLC Electrospray Mass Spectrometry Application

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Application Note. Authors. Abstract. Introduction. Environmental

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Chromatography. Gas Chromatography

Water Injections in GC - Does Water Cause Bleed?

Transcription:

Deteration of Polybroated Diphenylethers (PBDE) in Sediment and Sewage Sludge Application Environmental Authors Sonia Wanner Centre d'analyse et de Recherche (CAR) 76 route du Rhin F-67441 Illkirch France Isabelle Chanel Agilent Technologies France Parc Club du Moulin à Vent 33 rue du Docteur Levy F-69693 Venissieux France Abstract Generally used as flame retardants, polybroated diphenylethers (PBDE) have become chemicals of significant environmental concern. While little toxicological information is available, PBDEs have been detered to be persistent and bio-accumulative substances, similar to well-known environmental contaants such as polychlorinated biphenyls (PCBs). Therefore, environmental laboratories are asked to analyze polybroated diphenylethers (flame retardants) in sediment and sewage sludge. This application note describes the successful separation of all PBDEs, including the most difficult, decabroated diphenylether. Examples include standards as well as real samples of sewage sludge with quantitative data. Introduction With increasing frequency, environmental laboratories are asked to analyze PBDEs (flame retardants) in sediment and sewage sludge. See Figure 1. Figure 1. O Structure of PBDEs. Br 5-1 Broated flame retardants (BFRs) are a group of chemicals added to many products, including computers, TVs, and household textiles, in order to reduce fire risk. Two substances, decabromodiphenyl ether (DecaBDE) and tetrabromobisphenol A (TBBP-A), account for about 5% of world use of broated flame retardants. Two other polybroated diphenyl ethers (PolyBDE) - octabromodiphenyl ether (OctaBDE) and pentabromodiphenyl ether (PentaBDE) - are used commercially, but in much smaller quantities than DecaBDE. Heating (for example, during manufacture of plastics) and burning of materials containing PBDEs and other BFRs can produce polybroated

dibenzo-p-dioxins and dibenzofurans, which have similar toxicological effects to chlorinated dioxins. Research has shown that low-level exposure of young mice to PBDEs causes permanent disturbances in behavior, memory, and learning (Eriksson et al., 1998) [1]. PBDEs have also been shown to disrupt the thyroid hormone system in rats and mice; these systems are a crucial part of the development of the brain and body (Darnerud and Thuvander, 1998 [2]; Hallgren and Darnerud, 1998) [3]. The release of these organic pollutants can be revealed by analyses of sewage sludge produced by municipal waste-water treatment plants. Therefore, the European community has given a directive (2/6/CE) [4] for water to analyze four PBDEs (BDE-99, BDE-1, BDE-25, BDE-29) and is now working on an ISO norm ISO/CD 2232 to analyze eight PBDEs (BDE-47, BDE-99, BDE-1, BDE-153, BDE-154, BDE-183, BDE-25, BDE-29). This analysis starts with an extraction of broated diphenyl ethers (BDEs) from the dried sample of sediment or sewage sludge by a solvent (for example, hexane or other solvents suitable to get high extraction rates). The extract is cleaned, with silica, for example, if necessary. After concentration, the BDEs are separated by capillary gas chromatography (GC) and detected with a suitable system. A calibration over the total procedure using an internal standard (ISTD) mix is used to calculate the concentration in the sample. When analyzing PBDE with GC, a number of problems arise: [5] Adsorption to glass surfaces Discriation of high molecular weight compounds Degradation of the heavier congeners Irreproducible results Disappearing peaks This application note gives analysts the necessary tools to attempt low-level detection of PBDE by gas chromatography/mass spectrometry (GC/MS). Materials and Methods Samples All sewage sludge and sediment samples were provided by municipal waste-water treatment plants. Ten grams of sediment or 1 g of sewage sludge is liquid extracted. The extract is cleaned on silica and the clean extract is concentrated in 1-mL hexane prior to GC analysis. Standards and ISTDs The project for the European norm 2232 (2/6/CE) is requesting analysis of four PDBE (BDE-99, BDE-1, BDE-25 and BDE-29) and recommends TetraBDE (BDE-77) as ISTD. (See Table 1.) These standards were purchased commercially and were of the highest grade available. A test mixture of pentabde (BDE-99, BDE-1), octabde (BDE-25), and decabde (BDE-29) was used for the evaluation in order to obtain a GC analysis with little or no discriation. BDE-77 was used as ISTD. Standard solutions containing.1;.5;.1;.2;.25;.5 ng/µl of pentabde,.5; 1; 2; 3; 4; ng/µl of decabde and octabde, and.2 ng/µl of ISTD were prepared in hexane. Table 1. Selected BDEs Name Formula Abbreviation Molar mass g/mol 3,3',4,4'-tetraBDE C 12H 6Br 4O BDE-77 481.715 2,2',4,4',5-pentaBDE C 12H 5Br 5O BDE-99 564.6911 2,2',4,4',6-pentaBDE C 12H 5Br 5O BDE-1 564.6911 2,3,3',4,4',5,5',6-octaBDE C 12H 2Br 8O BDE-25 81.384 DecaBDE C 12Br 1O BDE-29 959.1714 2

GC Conditions The selection of column and injection parameters is of great importance for the GC analysis of PBDE, especially for the high molecular weight congeners. See Table 2. The temperature of the GC is of great importance since some congeners decompose at temperatures just above 3 C. Thermal degradation is a function of temperature and time; thus, by choosing a column with as little retention for the BDE congeners as possible and shortening the column to the imum length required for the separation, thermal degradation can be imized. In addition, pulsed injection allows shorter injection time and also helps to imize risk of thermal degradation. A pulsed splitless injection and a DB-1 3 m,.32 mm, thin film,.1 µm, really imizes the time each PDBE stays in both the injector and in the column and avoids degradation. Table 2. Optimized Run Conditions Column: DB-1 Part number: 123-131 Length: 3 m Diameter:.32 mm Film thickness:.1 µm Carrier: Helium at 58 cm/s Flow rate 2.5 ml/ Injector: 2 µl Pulsed splitless at 25 C Oven: 6 C for 2 utes 6 C 2 C at 1 C/ 2 C for 2 utes 2 C 3 C at 2 C/ 3 C for 25 utes Detector MS Agilent 5973 inert MSD SIM mode Group 1 / 3 / m/z 486; 484; 326 Group 2 / 2 / m/z 46; 564; 566 Group 3 / 24 / m/z 642; 644; 562 Group 4 / 28 / m/z 799; 797 Quad temperature 15 C Source temperature 23 C Transfer line temperature 3 C 3

Results and Discussion The chromatograms (Figures 2 and 3) show very good peak shapes for each PBDE and a high response for the most critical decabde (BDE-29) (see Figure 4) using the optimized run conditions listed in Table 2. 5 18 16 14 2 4 6 12 1 8 6 4 2 7 8 1 3 9 1 5. 1. 15. 2. 25. 3. 35. 4. 45. Figure 2. Total Ion Chromatogram (TIC) of a standard mixture at 2 2 ng/µl 1 2: pentabde 3 9: octabde 1: decabde 1 6 2 PentaDBE Ion 46 5. 1. 15. 2. 25. 3. 35. 4. 45. 1 6 2 OctaBDE Ion 643 5. 1. 15. 2. 25. 3. 35. 4. 45. Figure 3. Selected Extracted Ion Chromatograms (EICs) of a standard mixture at.5 5 ng/µl. 4

7 6 Ion 799. (798.7 to 799.7) 5 BDE-29 4 3 2 1 27.5 28. 28.5 29. 29.5 3. 3.5 31. 31.5 32. Figure 4. EIC of BDE-29 in a standard mixture at 5 ng/µl. In order to have a precise quantitation, five point calibration curves from.1 to.25 ng/µl for pentabde and from.5 to 4 ng/µl of octa and decabde were achieved with ISTD BDE-77 at.2 ng/µl. (See Figure 5.) For all components, the R 2 values range from.996 to 1, meeting the AFNOR requirements for valid quantitation (See Table 3). Table 3. Calibration Curve Summary Using Optimized Analysis Conditions with GC/MS Compound Calibration range (ng/µl) Target ion m/z Qualifier ion m/z R 2 value BDE-77 ISTD.2 486. 326. ISTD BDE-99.1.25 45.8 563.6 1 BDE-1.1.25 45.8 563.6 1 BDE-25.5 4 641.6 643.6.99 BDE-29.5 4 799.4 797.4.996 Figure 5. Calibration curve for decabde (BDE-29) by GC/MS. 5

Concerning the limit of detection (LOD), the lower level at 1 µg/kg (1 µg/kg for sewage sludge\) for pentabde and 5 µg/kg (5 µg/kg for sewage sludge) for octabde and decabde in sediment, which is 1 pg/µl of pentabde or.5 ng/µl of octa and decabde in solution, is easily achieved. This is the case even for decabde because a very good signal-to-noise ratio was achieved, as shown in Figure 6. 28 Ion 799. 24 2 16 BDE-29 12 8 26. 28. 3. 32. 34. 36. Figure 6. EIC of BDE-29 in a standard mixture at.5 ng/µl, which is the required LOD 5 µg/kg of sediment. Real sewage sludge samples were analyzed using the run conditions listed in Table 2. Figure 7 shows one example. The EICs of the different PDBE show that only one pentabde and the decabde are present in this sample, and they were quantified in a quantitation report showed on Table 4. 28 BDE-77 24 2 Ion 486. Ion 565.6 Ion 799. 16 12 8 4 BDE-99 BDE-29 15. 16. 17. 18. 19. 2. 21. 22. 23. 24. 25. 26. 27. 28. 29. Figure 7. Overlaid EICs for BDE-77, 99, and 29 from sewage sludge. 6

Table 4. Quantitation Report of Real Sewage Sludge ISTDs RT QIon Response Conc units Dev() BDE 77 19.43 486 82327.2 ng/µl.2 Target compounds Qvalue BDE 1 2.14 46 22687.72 ng/µl 11 BDE 99 2.48 46 17372.45 ng/µl 87 BDE 25 23.91 642 17336.3417 ng/µl 57 BDE 29 28.31 799 64526 5.253 ng/µl 9 Summary By combining the highly inert thin film DB-1 with the Agilent 689 gas chromatograph and the Agilent 5973 inert MSD, laboratories can achieve accurate quantitation of PBDE in sediments and sewage sludge. References 1. P. Erikkson, E. Jakobsson, and A. Fredriksson, Developmental neurotoxicity of broated flame retardants, polybroated diphenyl ethers and tetrabromo-bis-phenol A. (1998) Organohalogen Compounds, 35: 375-377. Acknowledgements Doique Eisenblaetter (CAR, Illkirch) and Bruno Schnepp (CARSO, Lyon) are acknowledged for supporting the partnership with Agilent Technologies France. Jean-Paul Chauve (Agilent Technologies France) and Eberhardt Kuhn (Agilent Technologies USA) are acknowledged for their technical support. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. 2. S. Hallgren and P.O. Darnerud, Effects of polybroated diphenyl ethers (PBDEs), polychlorinated biphenyls, (PCBs) and chlorinated paraffins (CPs) on thyroid hormone levels and enzyme activities in rats. (1998) Organohalogen Compounds, 35: 391-394. 3. P.O. Darnerud and A. Thuvander, Studies on immunological effects of polybroated diphenyl ether (PBDE) and polychlorinated biphenyl (PCB) exposure in rats and mice. (1998) Organohalogen Compounds, 35: 415-418. 4. European framework directive on water policy N o (2/6/CE) 5. E. Kuhn, J. Ellis, and S. Wilbur, GC Analysis of Polybroated Flame Retardants, Agilent Technologies publication 5989-94EN www.agilent.com.chem 7

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 24 Printed in the USA June 4, 24 5989-111EN