Spectrophotometry Materials

Similar documents
Introduction to Spectroscopy: Analysis of Copper Ore

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Experiment 6: Determination of the Equilibrium Constant for Iron Thiocyanate Complex

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore

EXPERIMENT #3 A Beer's Law Study

Photosynthesis. Introduction

Determination of the Equilibrium Constant. HIn - + H 2 O = H 3 O + + In 2-, [H 3 O + ][In 2- ] K ai = [HIn - ]

Introduction to Spectroscopy: Analysis of Copper Ore

MORE LIGHTS, COLOR, ABSORPTION!

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

MOLEBIO LAB #4: Using a Spectrophotometer

EXPERIMENT 5: PHOTOMETRIC DETERMINATION OF AN EQUILIBRIUM CONSTANT

Investigating Food Dyes in Sports Beverages. Sample

Experimental Procedure Lab 402

Introduction to Spectroscopy: Analysis of Copper Ore

Beer's- Lambert Law and Standard Curves. BCH 312 [Practical]

Determining the Concentration of a Solution: Beer s Law. Evaluation copy. Figure 1

For simplicity, we ll represent BTB s ionization in a solution by the equilibrium: HBTB = H + + BTB -

Experiment 7A ANALYSIS OF BRASS

Chemical Equilibrium: Finding a Constant, Kc

The Determination of an Equilibrium Constant

Determining the Concentration of a Solution: Beer s Law

aa + bb cc + dd Equation 1

MEASUREMENT: PART II

Spectrophotometric Study of Biomolecules: Absorption Spectrum of Hemoglobin and Quantitative Assay of Serum Protein Concentration

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Chemical Equilibrium: Finding a Constant, Kc

Lab #12: Determination of a Chemical Equilibrium Constant

The Determination of an Equilibrium Constant

Bio 120 Lab 5: Quantitative Analysis

Determining the Concentration of a Solution: Beer s Law

Experiment 11 Beer s Law

Chemical Equilibrium: Finding a Constant, Kc

AP Biology Lab 4 PLANT PIGMENTS AND PHOTOSYNTHESIS

Experiment 13I THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Experiment 11 Beer s Law

Lab 13.3 Determining K c via Colorimetry

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample

MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT. Lab Report ROOM NO: FE E309

Preparation of Standard Curves. Principle

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Photosynthesis. Photosynthesis is the conversion of light energy to chemical energy and its subsequent use in the synthesis of organic molecules.

Enzymes. Lab Exercise 7. Introduction. Contents. Objectives

The Determination of an Equilibrium Constant

PRELIMINARY ACTIVITY FOR

Lab 5: Calculating an equilibrium constant

Chemical Equilibrium: Finding a Constant, Kc

REVIEW OF LAB TECHNIQUES

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

PLANT PIGMENTS AND PHOTOSYNTHESIS LAB

VISIBLE SPECTROSCOPY

Atomic Theory: Spectroscopy and Flame Tests

CH 112 Special Assignment #4 Chemistry to Dye for: Part C

Chemistry 213. A KINETIC STUDY: REACTION OF CRYSTAL VIOLET WITH NaOH LEARNING OBJECTIVES

1iI1E. The Determination of 0 an Equilibrium Constant [LU. Computer

Chemistry 112 SPECTROPHOTOMETRIC DETERMINATION OF AN EQUILIBRIUM CONSTANT

CHM 152 updated May 2011 Lab 6: Experimentally Determining an Equilibrium Constant using Spectrophotometry

CHM112 Lab Determination of an Equilibrium Constant Grading Rubric

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Finding the Constant K c 4/21/15 Maya Parks Partners: Ben Seufert, Caleb Shumpert. Abstract:

Virtual Lab 5 Photosynthesis

EXPERIMENT. Beer s Law

EXPERIMENT 6: Photometric Determination of an Equilibrium Constant

Experimental Procedure Overview

Instructor s Advance Preparation

CHEM Lab 7: Determination of an Equilibrium Constant using Spectroscopy

Laboratory Measurements and Procedures

The Effect of Alcohol on Biological Membranes

Exp 03 - Reaction Rate

Lab Investigation 4 - How could you make more of this dye?

Experiment 12H, Parts A and B

DETERMINATION OF K c FOR AN EQUILIBRIUM SYSTEM

EXPERIMENT 14. ACID DISSOCIATION CONSTANT OF METHYL RED 1

2 (aq) [FeSCN [Fe 3JSCN] Figure 1

Atomic Theory: Spectroscopy and Flame Tests

Atomic Theory: Spectroscopy and Flame Tests

Experiment 12: SPECTROSCOPY: EMISSION & ABSORPTION

Lab #3 ph and Buffers

Determination of an Equilibrium Constant

Basic Instrumentation. Learning Objectives:

Spectrophotometric Determination of Iron

Introduction. Concepts Kinetics Order of reaction Reaction rate Colorimetry. Background

Modified Adams Assay for Phenolics in Wine

LAB 05 Enzyme Action

C.S.Rye: Mascot Murder Mystery

Spectrophotometric Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant

Determination of an Equilibrium Constant

Beer's Law and Data Analysis *

Spectroscopy II Introduction: Spectrophotometry and Fluorometry

Quantitative Proteins Estimation by Lowry method

Atomic Theory: Spectroscopy and Flame Tests

Plop Plop, Fizz Fizz, Oh What A Relief It Is (Which Pain Reliever Works Fastest)

Experiment #7. Determination of an Equilibrium Constant

we might also expect the reaction rate to be influenced by ph. In fact, the rate has been reported to follow the rate law:

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Transcription:

Spectrophotometry Materials Item per Class per Bench Genesys 10UV Spectrophotometer 6 1 13 ml test tubes box 7 Test tube racks 6 1 1% Albumin solution 25 ml/one flask 2 ml 0.7% Albumin solution (unknown conc.) 25 ml/one flask 1 ml Sharpies 6 1 Kim wipe boxes 6 1 1 ml pipettes 36 6 DI Water Bottles 6 1 Pipette Pumps 6+ (1 ml pipettes) at least 1 Biuret reagents bottles 6 1 Disposable transfer pipettes (droppers) 6 1 0

Spectrophotometry Abstract: When you look at a green leaf, it appears green to you because of the leaf pigments ability to absorb red and blue/violet and transmit green visible light. In this lab you will learn about the spectrophotometer, an instrument used to detect light energy that is transmitted by various substances. The spectrophotometer is useful because the absorption and transmission of light energy, the basis for one of the most widely used procedures for determining the concentration of substances in a solution. You will learn to use the spectrophotometer, become familiar with its parts, and use this instrument to determine the concentration of an unknown substance through the construction of a standard curve. You will use this instrument several times during this laboratory course, so make sure that you become comfortable in using this instrument! This lab also includes the principles of the Scientific Method, a series of steps commonly used to make observations, formulate hypotheses, carry out experiments and interpret the data generated from these experiments. Genesys 10UV Spectrophotometer 1. Locate the following on the spectrophotometer. As you complete the lab, come back to this section and record its use: 1

2. Identify the following on the spectrophotometer at your bench. a. On/Off Switch b. Measure Blank button c. Set nm button d. Sample Holder Determine the concentration of an unknown substance by constructing a Standard Curve When light of specific wavelength (color) is passed through a solution, some of the light energy will be absorbed. The remaining light energy is detected and expressed on a meter as either percent transmittance (%T) or absorbance (A). The spectrophotometer can be utilized to determine the concentration of a solution using a set of standards of known concentration. A higher concentration means a larger amount of light will be absorbed. By graphing a standard curve using the absorbance you determine for each standard, the unknown concentration can be extrapolated. Follow these directions carefully. Do not rush, or you will end up having to do the entire experiment again! Based upon my understanding of the effects of concentration on absorbance, I hypothesize that tube with the _highest concentration will have the highest absorbance because a higher concentration will have more color and in turn absorbs more light. Furthermore, I hypothesize that the tube with the lowest concentration will have the lowest absorbance because a lower concentration will have less color and in turn absorb less light. SM: The independent variable in this experiment is concentration SM: The dependent variable in this experiment is absorbance 2

Turn on the spectrophotometer. 1. Turn on the spectrophotometer, allow it to initialize. Prepare your samples: 2. Obtain 7 test tubes. a. Label one of them U for unknown. b. Label one B for blank. c. Label the remaining tubes #1- #5. (Steps 3-9 pertain ONLY to tubes #1- #5) 3. Add 1 ml of distilled water to tubes #2- #5. 4. To tube #1, add 1 ml of standard protein solution (1% albumin). Record the concentration of the standard protein solution here: 0.01 mg/ml Check your understanding At this point, what is the difference between tube #1, and tubes #2- #5? Tube #1 has protein while tubes #2---#5 only have water. 5. To tube #2, add 1 ml of standard protein solution. Shake gently. 6. With a fresh pipette, remove 1 ml from tube #2 and add it to tube #3. Shake gently. 7. With a fresh pipette, remove 1 ml from tube #3 and add it to tube #4. Shake gently. 8. With a fresh pipette, remove 1 ml from tube #4 and add it to tube #5. Shake gently. 9. With a fresh pipette, remove 1 ml of fluid from tube #5 and discard the fluid. At this point, you should have approximately the same amount of fluid in each of tubes #1- #5. Check your understanding What happened to the concentration of protein in each subsequent tube as you proceeded from step 5 through step 8? (Check with your instructor if you are not sure.) The concentration is reducing in half each step. 10. Add 1 ml of unknown protein to tube U. 11. Add 1 ml of distilled water to tube B. Check your understanding What is the difference between tube B and all of the rest of the tubes? Why will you use this tube as your blank? Tube B is the only tube with no protein in it. 12. To ALL tubes that you have prepared, add 2 mls of Biuret reagent and allow the color to develop for 5 minutes. The blank as well. 3

Determine Peak Wavelength: Each molecule has a wavelength fingerprint, or peak wavelength. The peak wavelength (λmax) is the best wavelength to make measurements since it will yield the highest absorbance values and thus the highest sensitivity; high sensitivity means that lower concentrations of protein can be detected and measured accurately. This can be determined by measuring the absorbance of a substance over a range of wavelengths to create an absorbance spectrum. 1. Place the blank (tube 0) in the sample holder of the spectrophotometer. 2. Enter the lowest wavelength (420 nm). 3. The distilled water blank has no color and should not absorb any visible light. If the absorbance is not zero, re-click set blank to calibrate to zero on the absorbance scale. 4. Remove the blank and replace it with tube 1 (1% albumin). This is the sample you will use to determine the absorption spectrum of albumin. 5. After the meter has stabilized (5-10 sec) read the absorbance value and record the wavelength and absorbance value in Figure 1 below. 6. Remove tube 1 and adjust the filter to the next wavelength in Figure 1. 7. Put the blank back into the spectrophotometer and readjust for zero absorbance at the new wavelength. The spectrophotometer should be recalibrated with the blank often, especially when you change the wavelength. 8. Insert tube 1 and measure its absorbance at the new wavelength. Record the absorbance in Figure 1. 9. Complete Figure 1 for the other wavelengths by repeating steps 1-7 and measuring the absorbance of the contents of tube 1. Results: Wavelength 420 nm 460 nm 490 nm 530 nm 570 nm 610 nm 660 nm Figure 1. Absorption Spectrum for Albumin Absorbance Record the wavelength with the peak absorbance here: 530 nm 4

Use the spectrophotometer set at the peak wavelength to find the absorbance (A) for each sample (cuvette). 1. Place the blank sample in the diamond shaped sample holder again. 2. Once the blank is oriented correctly, close the lid of the sample holder and press the Zero button to set the blank to 0.00 Absorbance. Check your understanding If the blank is set to zero absorbance, what does this mean in terms of the percent of light that will be transmitted through the cuvette (%T)?100% T Is there anything in your blank cuvette that should attribute an absorbance value? No 3. Remove the blank test tube from the sample holder and determine the absorbance of each of your other test tube (one at a time) by: (there is no need to the blank again) a. removing fingerprints from the cuvette with a soft wipe b. inserting each test tube into the sample holder c. closing the lid and d. recording the reading from the digital display in Figure 2. Results: Serial Dilution Tube # mg/ml protein Absorbance 1 2 3 4 5 B (blank) U (Unknown) Figure 2. Comparative concentrations and absorbances in known and unknown samples. 0? 5

Note: It is important to note that you must construct a fresh standard curve every time you are determining the concentration of unknowns. You should prepare the standards and your unknowns together. You cannot use a friend s standard curve and expect accurate results for your unknown! To construct your standard curve, you will plot the absorbance of each dilution on the Y- axis against the concentration of each dilution on the X- axis, as shown in Figure 3. You will draw a line of best fit using the data points you plot. By drawing a straight line from (a) to the curve at (b) and dropping from the curve to (C), the concentration of an unknown solution can be determined. Figure 3. Using a standard curve to determine the concentration of an unknown protein sample. 6

Absorbance Use the blank graph provided in Figure 4 to create your standard curve by plotting each of the absorbance values that you determined from tubes #1- #5. Make sure that you label both the X axis and the Y Figure 4. Standard curve correlating sample concentration to absorbance. Protein Concentration (mg/ml) 0.005 0.01 Figure 2. Standard curve correlating sample concentration to absorbance. 7

Conclusions: What is the concentration of your unknown (include units!)? Points for Discussion: 1. Why are you not able to use a standard curve generated in one lab to assess the concentration of an unknown in another lab? List at least 2 reasons. The spectrophotometer needs to be calibrated every time it is used. Lab conditions may be variable. 2. How does the way you connect the data points affect your conclusions about the concentration of your unknown? If all of the data points lie in a straight line except for one, should you ignore the one point or shift the line to connect the points? The outlier may be ignored or retested to verify an error. 3. Would a hand generated line of best fit or a computer generated line yield the most accurate results? Why? Computer generated line may be more accurate because it can eliminate human error. 8