Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Similar documents
Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University. and

ALLOWAY METHOD OUTLINE

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

STANDARD OPERATING PROCEDURES

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

QAM-I-111 Operation and Calibration of the Conductivity Meter

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

SECTION D.2 AMMONIA NITROGEN

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Oxygen Demand, Chemical

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Glossary of Common Laboratory Terms

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

SOP-C-120 Determination of ph

CHEMISTRY DEPARTMENT

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

Oxygen Demand, Chemical

DR/4000 PROCEDURE. HARDNESS, Total

DR/4000 PROCEDURE. IRON, Total

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

LOW-RANGE TOTAL CHLORINE

GENERAL INSTRUCTIONS GENERAL PREPARATION

Experiment #7. Determination of an Equilibrium Constant

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Method to 0.50 mg/l NH 3 N Powder Pillows

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

EXPERIMENT 6. Properties of Buffers INTRODUCTION

Method to 0.50 mg/l NH 3 N Powder Pillows

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

GENERAL INSTRUCTIONS GENERAL PREPARATION

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Test Method: CPSC-CH-E

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

DPD Test N Tube Method *

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Nitrogen, Total Inorganic

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

STANDARD OPERATING PROCEDURE No. 52 STATIC ACID GENERATION (NAG) TEST

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

NITROGEN, TOTAL, Test N Tube * Method 10071

Laboratory Certification Workshop

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Chlorine, Free and Total, High Range

Determining the Concentration of a Solution: Beer s Law. Evaluation copy. Figure 1

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

California Environmental Protection Agency Air Resources Board

25. Qualitative Analysis 2

Manganese III Digestion Method * (with chloride removal)

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Approximate Volatile Acids by Titration

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline)

P E R O X Y G E N S A sector group of Cefic

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURES

Transcription:

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: OEWRI Laboratory Manager Date: Approved by: OEWRI Director Date: September 30, 2015 OEWRI-SOP-004 1

Table of Contents 1 Identification of the test method... 3 2 Applicable matrix or matrices... 3 3 Detection Limit... 3 4 Scope of the test method... 3 5 Summary of test method... 3 6 Definitions... 3 7 Interferences... 4 8 Health and safety... 4 9 Personnel qualifications... 4 10 Equipment and supplies... 4 11 Reagents and standards... 4 12 Sample collection, preservation, shipment and storage... 4 13 Quality control... 5 14 Calibration and standardization... 5 15 Procedure... 5 16 Data acquisition, calculations, and reporting... 6 17 Computer hardware and software... 6 18 Method performance... 6 19 Pollution prevention... 6 20 Data assessment and acceptable criteria for quality control measures... 6 21 Corrective actions for out-of-control or unacceptable data... 7 22 Waste management... 7 23 References... 7 24 Tables, diagrams, flowcharts and validation data... 7 25 Calibration Log Sheet... 8 26 ph/specific Conductivity Bench Sheet... 8 27 Bench Procedure Sheet... 8 2

1 Identification of the test method Laboratory analysis of water samples for conductivity using the Cole-Parmer Traceable Portable Conductivity Meter. 2 Applicable matrix or matrices This method is suitable for the determination of conductivity in potable and surface waters and wastewaters with a conductivity of up to 199.9 ms/cm. 3 Detection Limit The manufacturer claims a range of detection from 0.00 to19.99 µs, 0.0 to 199.9 µs, 0 to 1999 µs, 0 to 19.99 ms, and 0 to 199.9 ms, with a resolution of 0.05% of range. 4 Scope of the test method This standard operating procedure provides the Missouri State University (MSU) laboratory personnel with guidance on the procedure for determining conductivity in water samples. This method is limited to the determination of conductivity in water samples collected from natural bodies of water with conductivity values of <20 ms/cm. 5 Summary of test method Conductivity is the ability of a solution to pass an electric current. Current is carried by inorganic dissolved solids, such as chloride, nitrate, sulfate and phosphate ions in the solution, as well as such cations as sodium, calcium, magnesium, iron and aluminum. Organic materials like oils, phenol, alcohols and sugars do not carry electric current well (Hach 1989). Conductivity is performed by measuring the voltage drop, or resistance, between two electrodes immersed in a test solution. In the Cole-Parmer Traceable Portable conductivity meter, temperature compensation is automatic. The temperature compensation circuit measures the temperature of the sample and from the deviation from 25ºC calculates and correct the reading 2% for each degree of deviation (Hach 1989)(Cole-Parmer 2015). A 150 ml aliquot of sample is decanted into a clean 250 ml beaker. The conductivity probe is inserted into the sample. Allow time for the reading to stabilize and record reading. Rinse probe thoroughly with deionized water before analyzing the next sample. 6 Definitions 6.1 Analytical batch: The set of samples processed at the same time 6.2 Field duplicates (FD): Two samples taken at the same time and place under identical circumstances and that are treated identically throughout field and laboratory procedures. Analysis of field duplicates indicates the precision associated with sample collection, preservation, and storage as well as laboratory procedures. Collect one set of field duplicates for every ten samples collected. 6.3 Laboratory reagent blank (LRB): An aliquot of deionized water treated as a sample in all aspects, except that it is not taken to the sampling site. The purpose is to determine if the analytes or interferences are present in the laboratory environment, the reagents, or the apparatus. Analyze one LRB for every twenty samples analyzed. 3

6.4 Laboratory duplicates (LD): Two aliquots of the same environmental sample treated identically throughout a laboratory analytical procedure. Analysis of laboratory duplicates indicates precision associated with laboratory procedures but not with sample collection, preservation or storage procedures. Analyze one set of LDs for every ten samples analyzed. 6.5 Method detection limit (MDL) -- The lowest level at which an analyte can be detected with 99 percent confidence that the analyte concentration is greater than zero. 7 Interferences When measuring very low conductivity levels (approximately 2 µs/cm or less), the sample must be protected from atmospheric gases such as carbon dioxide or ammonia. These gases dissolve readily in water and would cause a rapid change in conductivity readings. 8 Health and safety This analysis involves handling freshwater samples that may contain live microorganisms and therefore pose some threat of infection. Laboratory personnel who are routinely exposed to such water samples are encouraged to protect themselves from water borne illnesses by wearing clean disposable gloves and washing their hands frequently. 9 Personnel qualifications Laboratory and field personnel shall have a working knowledge of this analytical procedure and will have received training from an MSU employee knowledgeable of the proper sample analysis procedures. Prior to the first batch of sample analyses, the analyst will complete a demonstration of capability exercise as described below in the Quality control section. 10 Equipment and supplies 10.1 Cole-Parmer Traceable Portable Conductivity Meter, EW-19601-03, with Cole- Parmer 4362 Conductivity Probe, EW-19601-07. 10.2 Laboratory glassware: 250 ml beakers 11 Reagents and standards 11.1 Deionized water (DI) 11.2 0.001 M Potassium chloride solution: Weigh 74.55 mg of KCl. Dissolve with deionized water having a conductivity value of close to 0 µs/cm. Dilute to 1L with DI. The conductivity of this solution should be 146.9 µs/cm. 11.3 Purchased conductivity standard: 1000µS/cm purchased from Fisher Scientific. Catalog number 2243-1. This is 495 mg/l as NaCl. 12 Sample collection, preservation, shipment and storage 12.1 See the SOP for water sample collection procedures. 12.2 Samples should be refrigerated as soon as possible after collection. 4

12.3 The holding time for conductivity is 28 days (Standard Methods, 2005). 13 Quality control 13.1 Laboratory duplicates (LD) and Field duplicates (FD) reproducibility: a. Carry out two replicates on the same sample. b. The relative percent difference (RPD) between the two conductivity values should be 20% of their average value. c. Use equation 1 to calculate RPD: Equation 1: RPD (%) = (A B) x 100% (A + B)/2 Where: A = conductivity of first aliquot (µs/cm), and B = conductivity of duplicate aliquot (µs/cm) d. Analyze one set of duplicates for every 10 samples analyzed. e. Record the RPD on the bench sheet. 13.2 Blank (LRB): At least one blank should be measured with each set of samples. a. For the blank measurement analyze 150mL of deionized water. b. The conductivity value for the blank should be less than 2 µs/cm. If it is not, evaluate the procedure and correct sources of error. c. Analyze one blank for every 10 samples analyzed. 14 Calibration and standardization 14.1 Clean the probe using deionized water. 14.2 Soak the probe in DI for 30 minutes. 14.3 Remove the probe from the water and fling out drops clinging inside the probe. 14.4 Immerse the probe to or beyond the vent holes in a beaker containing the calibration solution. Agitate vertically to remove entrapped air. 14.5 Repeat steps 14.3 and 14.4 one more time. 14.6 Press the POWER I key and CND key. Verify that the LO BAT indication does not appear. 14.7 Record the final reading in the instrument log book. 15 Procedure 15.1 Sample Analysis a. Pour approximately 150mL of sample into a 250mL beaker. b. Press the POWER I key and CND key. Verify that the LO BAT indication does not appear. c. Insert the probe into the sample. Immerse the tip to or beyond the vent holes and agitate vertically to be sure air bubbles are not entrapped in the probe. 5

d. Allow time for the reading to stabilize (about 30 seconds). e. Rinse the probe thoroughly with DI after each measurement. f. Record the conductivity value on the bench sheet. 15.2 Measuring Dilute Samples a. If the conductivity of the sample exceeds the range of the instrument, the sample can be diluted and the conductivity calculated. b. The affects of the dilution water must be considered c. Determine the conductivity of the dilution water. d. Calculate using equation 2. Equation 2: C S = (V T x C M ) [C D x (V T - V S )] / V S Where: C S = Conductivity of sample V T = Total Volume (sample + dilution water) C M = Conductivity measured (sample + dilution water) C D = Conductivity of dilution water V S = Volume of sample (total dilution water) 16 Data acquisition, calculations, and reporting 16.1 Record the conductivity values for each sample on the bench sheet (see below). 16.2 Record any other calculation results (e.g., dilution, duplicate RPD) in the comments section of the bench sheet. 17 Computer hardware and software 17.1 Word: This document and attached bench sheet are prepared using Microsoft Word. The Word document file name for this SOP is: OEWRI-Conductivity-SOP- 003 17.2 Excel: Quality control charts are created using Excel. 18 Method performance 18.1 The desired performance criteria for this measurement are: a. Detection limit: 2 µs/cm b. Precision: ± 20% RPD a. Minimum Quantification Interval: 0.1 µs/cm 19 Pollution prevention All wastes from these procedures shall be collected and disposed of according to existing waste policies within the MSU Geography, Geology, and Planning Department. Volumes of reagents made should mirror the number of samples being analyzed. These adjustments should be made to reduce waste. Diluted potassium or sodium chloride solutions are not hazardous and can be disposed of in the laboratory sinks. 20 Data assessment and acceptable criteria for quality control measures 20.1 The analyst should review all data for correctness (e.g., calculations). 20.2 Precision values are calculated for pairs of duplicate analyses. 20.3 Record the precision values as a percent on the bench sheet. 6

20.4 The desired precision is ± 20% RPD. 20.5 The desired detection limit is 2 µs/cm 20.6 The completed bench sheet is reviewed by the analyst s supervisor or the OEWRI Laboratory Manager. 21 Corrective actions for out-of-control or unacceptable data 21.1 Quality control charts will be created for charting precision and blank values. 21.2 The results for precision and blank data are compared to the acceptable values for this analysis; ± 20% and 2 µs/cm, respectively. 21.3 If a precision value exceeds 20% RPD then the analyst should write in the comments section of the bench sheet: These data are associated with an out-ofcontrol duplicate analysis. The UCL = 20%. Note: UCL is the Upper Control Limit (i.e., 20%). 21.4 If a blank value exceeds 2 µs/cm then the analyst should write in the comments section of the bench sheet: These data are associated with a blank value that exceeds the detection limit of 2 µs/cm. 21.5 The samples can be reanalyzed because the sample volume will not have been depleted after the initial analysis. 21.6 If data are unacceptable for any reason, the analyst should review their analytical technique prior to conducting this analysis again. 21.7 Any maintenance steps taken to correct the operation of the instrument should be recorded in the instrument log book. 22 Waste management The wastes generated in this method are not hazardous. They can be discarded in the following manner: the water can be discarded in the laboratory. 23 References 23.1 Cole-Parmer Traceable Portable Conductivity Meter Specifications. Cole- Parmer. Web. 16 Sept. 2015. 23.2 Hach Model 44600 Conductivity/TDS Meter manual. Hach Company, Loveland, CO. 1989. 23.3 Standard Methods for the Examination of Water and Waste Water. Method 2510. APHA, 21st Edition, 2005. 23.4 Water Sample Collection. 2006. OEWRI SOP. Document name: OEWRI-Water Sample Collection-SOP-001. 24 Tables, diagrams, flowcharts and validation data 24.1 There are no tables, diagrams, flowcharts or validation data for this method. 7

24.2 See page 9 for the ph/conductivity calibration sheet. The analyst should make a copy of this form for each batch of samples analyzed. 24.3 See page 10 for the ph/conductivity analysis bench sheet. The analyst should make a copy of this form for each batch of samples analyzed. 24.4 See page 11 for the Conductivity bench procedure sheet. 8

Ozarks Environmental and Water Resources Institute (OEWRI) at Missouri State University ph/sc Calibration Log Date of Calibration: Analyst: Oakton ph Meter ph 4 Pre Post Lot # ph 7 Pre Post Lot # Cole-Parmer Conductivity Meter Conductivity (µs/cm) Pre Post Lot # Comments Date of Calibration: Analyst: Oakton ph Meter ph 4 Pre Post Lot # ph 7 Pre Post Lot # Cole-Parmer Conductivity Meter Conductivity (µs/cm) Pre Post Lot # Comments Date of Calibration: Analyst: Oakton ph Meter ph 4 Pre Post Lot # ph 7 Pre Post Lot # Cole-Parmer Conductivity Meter Conductivity (µs/cm) Pre Post Lot # Comments 9

Ozarks Environmental and Water Resources Instititute (OEWRI) at Missouri State University ph, Conductivity, and Turbidity Bench Sheet Project: Date: Analyst: Site ID Temperature ( C) Sp. Cond (µs/cm) ph DO (ppm) Turbidity (NTU) Stage (cm) 10

Conductivity Bench Procedures Conductivity Calibration 1. Insert probe in DI water for ~30 minutes to rinse probe. 2. Select the conductivity standard for calibration. 3. Pour ~4cm of the solution into two separate containers. 4. Power on the meter 5. Rinse the probe into one of the two containers and then gently stir the probe. 6. Dip rinsed probe into the other container (Tap probe on the bottom of container to remove any bubbles). Let probe stabilize to the solution temperature. 7. Press CAL key more than 2 seconds to enter the calibration (probe will automatically detect the conductivity value of the solution and blinks the value on the LCD). 8. Press the UP/DOWN keys to change the value on the primary display to match the value to the standard. (can be adjusted up to ±20%, if difference is greater it means cleaning or replacing probe is needed). Press enter once correct reading is selected. Analysis Procedure 1. Turn on the meter. 2. Rinse the probe with DI water. 3. Dip probe into sample making sure there are no air bubbles trapped on the slot of the probe. 4. Stir the probe gently in the sample to create a homogenous sample, allowing a few seconds for the temperature reading to approach the solution temperature 5. Take reading (when the reading is stable READY will display on the left-middle of the LCD. 11