Prima PRO Process Mass Spectrometer

Similar documents
Thermo Scientific Prima PRO Process Mass Spectrometer

Prima PRO Process Mass Spectrometer

THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION

Defining quality standards for the analysis of solid samples

POLY-OLEFIN PROCESS GAS ANALYSIS BY MASS SPECTROMETRY

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

2401 Gas (liquid) Chromatography

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

SmartNotes. What is meant by the term interference?

HYPHENATED TECHNOLOGY GUIDE

of mass spectrometry

HYPHENATED TECHNOLOGY GUIDE

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

SIFT-MS. technology overview

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

The Claviature of Gas Analysis

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

Author. Abstract. Introduction

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Methods of pollution control and waste management - laboratory. Adsorptive removal of volatile organic compounds from gases streams

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Introduction to GC/MS

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Technical know-how in thermal analysis measurement

Extrel Application Note

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

TANDEM MASS SPECTROSCOPY

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Gas Chromatography (GC)

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

Secondary Ion Mass Spectroscopy (SIMS)

Chapter 5. Stoichiometry

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

air protection technology

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Application Note S/SL

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

EA-IRMS: Fast and Precise Isotope Analysis of Liquids on a Delta V Isotope Ratio MS with a High Temperature Conversion Elemental Analyzer

Mass Spectrometry. General Principles

CHEM 429 / 529 Chemical Separation Techniques

HELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Determining Chemical Formulas

IRMS. perspective. Isotope Ratio Mass Spectrometry.

Rapid Catalyst Screening Reactors

Harris: Quantitative Chemical Analysis, Eight Edition

SERVOTOUGH SpectraScan

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Monomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

1) In what pressure range are mass spectrometers normally operated?

Low level Os isotopic measurements using multiple ion counting

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Introduction to Fourier Transform Infrared Spectroscopy

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

MAX300-BIO Bioreactor and Fermentation Gas Analyzer. Providing Solutions for Gas Analysis

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

An ion source performs the following two functions:

Chemistry 311: Topic 3 - Mass Spectrometry

Accurate and Precise Sulfur Analysis using the Thermo Scientific FlashSmart Elemental Analyzer

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

2. Separate the ions based on their mass to charge (m/e) ratio. 3. Measure the relative abundance of the ions that are produced

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

White Paper. Overview: NDIR Definition:

Mass Spectrometry. Introduction EI-MS and CI-MS Molecular mass & formulas Principles of fragmentation Fragmentation patterns Isotopic effects

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

Headspace Technology for GC and GC/MS: Features, benefits & applications

Offgas from Radiolysis Difficult analysis needing High Resolution Mass Spectrometry

The AccuTOF -DART 4G: The Ambient Ionization Toolbox

Reactions of Vinyl Chloride and Methanol in a Quadrupole Ion Trap Mass Spectrometer during VOC Analysis

Quadrupole Mass Spectrometry Concepts. Mass spectrometers for residual gas analysis: Intermediate Level Users Guide

Welcome to Organic Chemistry II

Ionization Techniques Part IV

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

Transcription:

APPLICATION NOTE Prima PRO Process Mass Spectrometer No. xxxx Controlling emissions by multi-component analysis of Volatile Organic Compounds (VOC) Authors: Peter Traynor, Thermo Fisher Scientific, Franklin, Mass, USA Robert Wright, Thermo Fisher Scientific, Winsford, Cheshire, United Kingdom Keywords Volatile Organic Compounds (VOC) Total Organic Carbon (TOC) Activated carbon adsorption Emissions control Magnetic Sector Secondary Electron Multiplier (SEM) Flame Ionization Detector (FID) Introduction To comply with local regulations, air containing VOCs generated by a large pharmaceutical site is fed to an activated carbon adsorption plant at a rate of 60,000 cubic meters/hour. The plant comprises three activated carbon beds; at any one time just one bed is being used, while the other two are being regenerated with steam. Activated carbon has small, low-volume pores that increase surface area; the surface area of one gram of activated carbon is more than 500 m 2, making it ideal for adsorbing VOCs. The plant is shown in schematic form in Figure 1. Figure 1 Schematic of activated carbon adsorption plant with three carbon beds

Historically, the activated carbon adsorption plant operated by switching the plant air between beds approximately every 2.5 hours. This was the maximum rate, since it took around 5 hours to regenerate a bed for use again. The adsorption plant inlet and outlet streams were monitored with Flame Ionization Detector (FID) analyzers; these only provide a reading of Total Organic Carbon (TOC), with no information on individual VOC concentrations. This online analysis was backed up by periodic laboratory analysis of average component concentrations with a Gas Chromatograph-Mass Spectrometer (GC-MS), using sample collection by adsorption trapping. Improving process control by implementing Process Mass Spectrometry Plant management were concerned that switching beds on a simple time basis was potentially inefficient, with beds being switched prematurely, resulting in excess energy costs as steam was being used unnecessarily. They decided they needed a new gas analysis system with improved sensitivity and accuracy, providing the composition of each individual VOC. Real-time monitoring of individual component concentrations would enable the plant to predict when bed changeover was required, allowing more efficient steam use while still preventing over-limit emissions. They also needed the new gas analysis system to provide a continuous and complete log of all emissions at the site, providing evidence to regulating authorities that compliance with emission standards was being maintained. Examples of VOC results Example MS data from the time when the operation of the activated carbon plant was based on switching beds at fixed time intervals are shown in Figures 2 to 5. 4000 3500 3000 2500 2000 1500 1000 500 0 Total Organic Carbon: Inlet 0 2 4 6 8 10 12 14 16 18 20 22 24 Time Time (hours) (hours) Figure 2 Inlet stream Total Organic Carbon 140 Total Organic Carbon: Outlet 120 100 80 60 40 20 0 0 2 4 6 8 10 12 14 16 18 20 22 24 Time Time (hours) (hours) Figure 3 Outlet stream Total Organic Carbon

Figure 2 shows the variable and sometimes high TOC level in the carbon adsorption plant inlet stream, while Figure 3 shows a much lower TOC level in the outlet stream, typically reduced by a factor of 20, where there is a regular accumulation effect between bed switches every 2.5 hours. VOCs: Inlet 800 700 600 500 400 300 200 100 0 Methyl Chloride Dichloromethane Chloroform 0 2 4 6 8 10 12 14 16 18 20 22 24 Time Time (hours) (hours) Figure 4 Three individual VOCs in the inlet stream VOCs: Outlet 25 Methyl Chloride Dichloromethane Chloroform 20 15 10 5 0 0 2 4 6 8 10 12 14 16 18 20 22 24 Time Time (hours) Figure 5 Three individual VOCs in the outlet stream, showing unnecessary bed switching Figure 4 shows three of the major inlet VOCs, methyl chloride, dichloromethane and chloroform. Figure 5 shows these same three VOCs in the outlet. Of all the VOCs, the chlorides and methyl bromide have the lowest acceptable emission standards. It can be seen from Figures 3 and 5 that on five occasions the bed was switched unnecessarily, when the outlet VOC levels were relatively low. These five unnecessary bed switches out of a total of ten bed switches over the 24 hour period equates to consuming 70% more steam for bed regeneration than is necessary. This was the primary reason for installing the process mass spectrometer: to control the bed switching process to conserve energy and reduce the plant running costs. TOC and individual VOC results for inlet and outlet are summarized in Table 1, showing the mean inlet TOC level of 692 reduced to 36 in the outlet.

Table 1 24 hours inlet and outlet stream data from VOC adsorption plant Component Inlet mean Inlet max. Inlet min. Outlet mean Outlet max. Outlet min. Methanol 5.1 17.5 0.0 6.0 10.5 0.0 Ethanol 5.3 36.0 0.0 0.1 2.4 0.0 Methyl Chloride 20.7 463.9 0.0 3.5 17.9 0.0 Acetone 77.6 407.6 0.0 0.0 2.7 0.0 2-Propanol 11.9 30.4 5.2 0.8 1.6 0.0 Methyl Ethyl Ketone 27.4 167.1 3.1 1.1 4.8 0.0 Tetrahydrofuran 0.0 0.6 0.0 0.0 0.0 0.0 Dichloromethane 65.0 700.1 4.0 5.4 20.4 0.0 Toluene 3.9 51.3 1.0 0.0 0.1 0.0 Methyl Bromide 0.0 0.5 0.0 0.0 0.0 0.0 1,2 Dichloroethane 0.0 0.0 0.0 0.0 0.0 0.0 Methyl Isobutyl Ketone 22.0 349.9 0.7 2.3 11.8 0.0 Chloroform 52.9 495.4 2.2 0.5 2.1 0.0 TOC 692.00 3914.4 232.0 36.3 132.3 7.3 Comparison of MS and FID Previously the site had used Flame Ionization Detector systems to monitor TOC levels. However the FID s response is extremely dependent on the species being measured, with halide and oxygen containing species giving a very different response from hydrocarbons [1], so that errors of 50% relative are possible. Calibration of an FID analyzer for this application is also problematic, because the component make up of the process gas is so variable. Using a FID to control bed switching therefore runs the risk of exceeding emission standards for individual or groups of components. Thermo Scientific Prima PRO s accurate multi-component analysis has allowed the site to control bed switching without exceeding emission standards. Online process MS results were also compared with laboratory results using sample collection by adsorption trapping, followed by Gas Chromatograph-Mass Spectrometer analysis. Good agreement was obtained, which served as a final consistency check. Advantages of Prima PRO Process MS At the heart of the Prima PRO is a magnetic sector analyzer which offers unrivalled precision and accuracy compared with other mass spectrometers. Thermo Fisher Scientific manufactures both quadrupole and magnetic sector mass spectrometers; over thirty years of industrial experience have shown the magnetic sector based analyzer offers the best performance for industrial online gas analysis. Key advantages of magnetic sector analyzers include improved precision, accuracy, long intervals between calibrations and resistance to contamination. Typically, analytical precision is between 2 and 10 times better than a quadrupole analyzer, depending on the gases analyzed and complexity of the mixture. Neutral gas atoms and molecules are first converted into positively charged ions in the Prima PRO ion source. This is an enclosed type for high sensitivity, minimum background interference and maximum contamination resistance. It is a high-energy (1000 ev) analyzer that offers extremely rugged performance in the presence of gases and vapors that have the potential for contaminating the internal vacuum components. Prima PRO has a proven track record of monitoring high percent level concentrations of organic compounds without experiencing drift or contamination. Ions are then accelerated through a flight tube, where they are separated by their mass to charge ratios in a magnetic field of variable strength. Since the magnetic sector mass spectrometer produces a focused ion beam at the detector, the peak shape obtained is flat-topped and uniform response is observed over a finite mass width. As the height of the peak is directly proportional to the number of ions striking the detector it is also directly proportional to the concentration of the component being measured. As long as the measurement is taken anywhere on the peak s flat top, high precision analysis will be observed.

The system has a dual Faraday/Secondary Electron Multiplier (SEM) detector, with the Faraday detector used to measure percentage level major air components (nitrogen, oxygen and argon) and the SEM detector used to measure the level components (carbon dioxide and VOCs). The relative response of the SEM detector compared to the Faraday detector is measured using the 16 O 18 O mass 34 peak present at 800 in air, which can be measured with high accuracy on both detectors. Selectivity Table 2 shows the relative sensitivities and mass spectral fragmentation patterns of the VOCs of interest, obtained after calibration. The VOC molecules generate a significant amount of spectral overlap; it is therefore extremely important that the fragment peaks are measured with high precision. There is an additional complication for measurements at mass 31, caused by the presence of a very large peak at mass 32 from the percentage levels of oxygen in the plant air. The intensity of the oxygen tail at mass 31 compared with the intensity of the oxygen molecular peak at mass 32 is 0.02%. So when the oxygen concentration is around 20%, this means the signal at mass 31 is equivalent to around 40. During calibration this interference is recorded and used to correct subsequent measurements of mass 31. With Prima PRO s magnetic sector instrument the measurement is very reproducible; on a quadrupole MS this interference level is much greater and also variable, resulting in excessive uncertainty in low level measurements of mass 31. Low level signals effectively tend to get buried in the noise from the oxygen peak. Table 2 Prima PRO VOC analysis set-up Component Relative Sensitivity Fragmentation Patterns 28 31 32 33 40 44 45 46 50 57 58 62 71 72 83 86 92.2 96 100 Nitrogen 1 100 Oxygen 0.85 100 Argon 1.57 100 Carbon dioxide 1.28 100 1.2 0.4 Methanol 0.63 100 0.3 0.2 0.1 0.2 0.1 0.0 0.0 0.0 0.1 0.1 0.1 0.0 0.0 Ethanol 1.01 100 38.2 19.4 0.1 0.1 0.1 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 Methyl chloride 0.87 7.1 1.2 0.7 100 0.1 0.1 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 Acetone 0.40 11.4 1.4 0.3 0.2 3.2 100 0.0 0.1 0.1 0.8 0.1 0.1 0.2 0.0 2-Propanol 1.61 8.7 100 2.5 0.1 0.4 2.7 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 Methyl ethyl ketone 0.36 13.3 12.5 0.5 2.3 34.5 2.3 0.0 4.5 100 0.2 0.1 0.1 0.0 0.0 Tetrahydrofuran 0.35 16.8 6.9 0.4 0.7 0.7 0.7 0.0 88.3 100 0.1 0.0 0.1 0.0 0.0 Dichloromethane 0.33 10.5 4.7 0.6 9.7 0.3 0.4 0.0 0.1 0.4 5.1 100 0.1 0.0 0.0 Toluene 1.41 2.8 9.4 6.3 5.6 0.1 0.1 3.9 0.0 0.0 0.1 1.0 100 0.0 0.0 Methyl bromide 0.45 8.3 17.6 10.1 0.6 0.3 0.2 0.1 0.0 0.1 0.2 0.1 4.9 100 0.0 1,2 Dichloroethane 0.39 8.8 1.0 0.2 2.4 0.4 0.3 100 0.0 0.1 0.3 0.0 0.0 0.5 16.4 Methyl isobutyl ketone 0.82 5.1 1.7 0.1 0.8 45.2 100 0.1 0.4 1.4 1.0 1.6 0.1 0.3 29.7 Chloroform 0.79 4.4 0.4 0.1 4.5 0.2 0.2 0.0 0.0 0.4 100 1.3 0.0 0.0 0.0

Figure 6 shows the logarithmic mass spectrum around mass 31 for air on the Prima PRO, Figure 7 shows the logarithmic spectrum for air containing around 400 ethanol. The magnetic sector analyzer s flat-topped peaks are clearly shown in these figures. Figure 6 Spectrum of air Figure 7 Spectrum of air with 400 ethanol The potential effects of cross interference can be tested by analyzing each VOC calibration gas sequentially. Typical results obtained by running each calibration gas in turn for 3 minutes are shown in Table 3. The results from this test are generally extremely accurate, with no significant interferences and no interferences for the more harmful compounds such as the chlorides and methyl bromide, which have the lowest

Table 3 Calibration gas checks for accuracy and cross-interference Calibration gas Methanol at 97.2 in N 2 Recorded mean result for calibrated VOC 96.7 Other VOCs recorded above 0.2 Ethanol at 105.5 in N 2 105.0 0.7 Methanol, 0.2 2-Propanol Methyl Chloride at 100.0 in N 2 99.5 0.3 Methanol Acetone at 98.1 in N 2 98.0 0.6 Methanol 2-Propanol at 92.2 in N 2 92.2 0.7 Methanol Methyl Ethyl Ketone at 97.1 in N 2 97.4 0.4 Methanol, 0.4 Methyl Ethyl Ketone Tetrahydrofuran at 99.4 in N 2 99.8 0.4 Methanol, 0.3 Acetone Dichloromethane at 103.5 in N 2 104.7 0.2 Methanol Toluene at 99.0 in N 2 99.7 0.3 Methanol Methyl Bromide at 99.9 in N 2 99.9 0.5 Methanol 1,2 Dichloroethane at 100.4 in N 2 99.9 0.2 Methanol Methyl Isobutyl Ketone at 104.1 in N 2 105.5 0.7 Methanol, 0.2 Methyl Ethyl Ketone Chloroform at 98.2 in N 2 99.3 0.6 2-Propanol Air 78.09% N 2, 20.94% O 2, 0.930% Ar, 389 CO 2 acceptable emission standards. For example, the emission limit for methanol is twenty times higher than that of dichloromethane. Stability The stability of the system can be checked by the daily analysis of one of the calibration gases. A typical stability check on a cylinder containing 100 of 1, 2 dichloroethane over 14 days without re-calibration is shown in Figure 8. The day-to-day and longer term accuracy is typically ± 5% at the 100 level. Over the short term (one hour), the reproducibility at 100 is typically ± 1%. Longer term fluctuations are mainly due to changes in the SEM gain. These effects could be reduced if the results were normalized relative to a SEM peak representative of air, such as the 16 O 18 O mass 34 peak; however it is more informative to see the additional analysis of the major air components nitrogen, oxygen and argon. These indicate that analysis results are correct - one of the problems with trace environmental analysis is that it is not always obvious that the results are real. Including air components in the analysis serves as a consistency check. 110 100 90 80 70 60 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Day Figure 8 Stability check analysis of 1, 2 dichloroethane over 14 days without recalibration

Communication with Plant Control System Analysis results of the inlet and outlet streams for each component can be transmitted from the Prima PRO to the plant control system by serial protocols such as Modbus or OPC, or by analog outputs. Digital outputs are available for stream identification, hardware faults, analyzer status (analyzing, standby or calibration) and component high concentration alarms. In normal operation, Prima PRO is set to run a calibration check daily at a set time. If the analysis of one of the calibration gases is outside a certain tolerance (for example 5%), the instrument is automatically re-calibrated in respect of the SEM gain. If the check analysis is outside a wider tolerance (for example 10%), the instrument performs an automatic re-calibration of sensitivities and fragmentation patterns for all components. A digital signal tells the plant control system when the Prima PRO is performing a calibration. Real-time monitoring of individual component concentrations can alert plant operation departments to limit the usage of a particular solvent or solvents. Conclusions Prima PRO provides fast, accurate multi-component VOC analysis of inlet and outlet streams TOC analysis by Prima PRO has been shown to be more accurate than by FID, because of the variation in concentration levels of different VOC species Prima PRO s multi-component analysis means that emission sources can be identified and traced to source Component alarms can be used as warnings to limit individual solvent usage, preventing over-limit emissions On a multi-bed carbon adsorption plant, bed switchover is controlled by Prima PRO s analysis of emissions, leading to a considerable reduction in steam usage and plant running costs Prima PRO generates a continuous and complete log of all emissions, providing evidence to regulating authorities that compliance with emission standards was being maintained Reference 1. Analytical Instrumentation, editors R.E. Sherman and L. Rhodes, Instrument Society of America, 627-629 (1996) USA 27 Forge Parkway Franklin, MA 02038 Ph: (800) 437-7979 Fax: (713) 272-2273 orders.process.us@thermofisher.com India C/327, TTC Industrial Area MIDC Pawane New Mumbai 400 705, India Ph: +91 22 4157 8800 india@thermofisher.com China +Units 702-715, 7th Floor Tower West, Yonghe Beijing, China 100007 Ph: +86 10 84193588 info.eid.china@thermofisher.com Europe Ion Path, Road Three, Winsford, Cheshire CW73GA UK Ph: +44 1606 548700 Fax: +44 1606 548711 sales.epm.uk@thermofisher.com Find out more at thermofisher.com/primapro For Research Use Only. Not for use in diagnostic procedures. 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. AN_CEPMS_1016