Calibration of the BABAR CsI (Tl) calorimeter

Similar documents
Wolfgang Gradl. on behalf of the BABAR collaboration. Tau 2016, Beijing 23 rd September 2016

Results from BABAR/PEP-II - One Year of Operations

Hadronic vs e + e - colliders

Study of e + e annihilation to hadrons at low energies at BABAR

Recent results on exclusive hadronic cross sections measurements with the BABAR detector

Validation of Geant4 Physics Models Using Collision Data from the LHC

Search for Lepton Flavour Violation (LFV) in Three-Body Tau Decays at BaBar

Energy Loss of Muons in the EMC

Muon reconstruction performance in ATLAS at Run-2

CLEO c. Anders Ryd Cornell University June 7, e e cc D D. D K,D K e

Abstract. To be published in the Proceedings of the 8th International Workshop On Tau Lepton Physics September 2004, Nara, Japan

Performance of muon and tau identification at ATLAS

CALICE scintillator HCAL

Study of τ Decays to Four-Hadron Final States with Kaons

Rare Exclusive Decays. Results from BaBar on the. DPF-2002, College of William and Mary. University of California, Jeffrey Berryhill.

Absolute D Hadronic Branching Fractions at CLEO-c

Measurements of R K and R K* at LHCb

7 Particle Identification. Detectors for Particle Physics Manfred Krammer Institute of High Energy Physics, Vienna, Austria

Last Lecture 1) Silicon tracking detectors 2) Reconstructing track momenta

Usage of GEANT 4 versions: 6, 7 & 8 in BABAR

Measurement of the mass of the W boson at DØ

NA62: Ultra-Rare Kaon Decays

arxiv:hep-ex/ v1 31 Dec 2001

CsI Calorimeter for KOTO experiment

Jet Reconstruction and Energy Scale Determination in ATLAS

Studies of charmonium production in e + e - annihilation and B decays at BaBar

The BABAR Electromagnetic Calorimeter: Status and Performance Improvements

Radiative penguins at hadron machines. Kevin Stenson! University of Colorado!

New Physics search in penguin B-decays

Tesi di Laurea Specialistica

hadronic decays at BESIII Abstract hadronic decays using a double tag technique. Among measurement for twelve Λ c

Overview of recent HERMES results

LHCb Semileptonic Asymmetry

Conference Report Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

LHCb Calorimetry Impact

Dalitz Plot Analyses of B D + π π, B + π + π π + and D + s π+ π π + at BABAR

International Workshop on Heavy Quarkonium Oct. 2007, DESY Hamburg. Prospects for Panda. Charmonium Spectroscopy

Study of Baryon Form factor and Collins effect at BESIII

PoS(EPS-HEP2015)565. Study of the radiative tau decays τ γlνν with the BABAR detector. Roger Barlow

Hadronic D Decays and the D Meson Decay Constant with CLEO c

The Mystery of Vus from Tau decays. Swagato Banerjee

Nicolas Berger SLAC, Menlo Park, California, U.S.A.

D Hadronic Branching Fractions and DD Cross Section at ψ (3770) from CLEO c

Relative branching ratio measurements of charmless B ± decays to three hadrons

Prospects for X(3872) Detection at Panda

First Observation of the

Calorimetry in. in Nuclear and Particle Physics Experiments

Simulation and validation of the ATLAS Tile Calorimeter response

Measurement of Properties of Electroweak Bosons with the DØ Detector

The CMS Particle Flow Algorithm

Measurements of Leptonic B Decays from BaBar

PoS(KAON13)012. R K Measurement with NA62 at CERN SPS. Giuseppe Ruggiero. CERN

FATRAS. A Novel Fast Track Simulation Engine for the ATLAS Experiment. Sebastian Fleischmann on behalf of the ATLAS Collaboration

Physics at Hadron Colliders

The BaBar CsI(Tl) Electromagnetic Calorimeter

The LHC Experiments. TASI Lecture 2 John Conway

ISR physics at BABAR

KEKB collider and Belle detector

D0 and D+ Hadronic Decays at CLEO

Guest Lecture PHY 7361: Harnessing Cherenkov Radiation SteveSekula, 13 April 2010 (created 9 April 2010)

Particle Detectors. Summer Student Lectures 2010 Werner Riegler, CERN, History of Instrumentation History of Particle Physics

Dual readout with tiles for calorimetry.

Lepton Universality Test in Υ(1S) decays at BABAR

PoS(KAON)049. Testing the µ e universality with K ± l ± ν decays

THE PEP-II B-FACTORY AND THE BABAR DETECTOR

Interaction of particles in matter

Calorimetry in particle physics experiments

Future prospects for the measurement of direct photons at the LHC

The reaction p(e,e'p)π 0 to calibrate the Forward and the Large Angle Electromagnetic Shower Calorimeters

Searches for Exotics in Upsilon Decays in BABAR

Recent results from SND detector

Measurement of the η mass by the NA48 Experiment

Correction for PMT temperature dependence of the LHCf calorimeters

Hadronic Cross Section Measurements at BES-III Sven Schumann Johannes Gutenberg Universität Mainz

Measurement of the associated production of direct photons and jets with the Atlas experiment at LHC. Michele Cascella

Studies of the diffractive photoproduction of isolated photons at HERA

Electroweak Physics at the Tevatron

Testing the m-e universality with ±

Studies of D ( ) Production in B Decays and e + e c c Events

PoS(ICHEP2012)300. Electroweak boson production at LHCb

arxiv: v1 [hep-ex] 21 Aug 2011

Introduction. The Standard Model

ATLAS EXPERIMENT : HOW THE DATA FLOWS. (Trigger, Computing, and Data Analysis)

arxiv:nucl-ex/ v1 21 Dec 2004

Searches for New Physics in quarkonium decays at BaBar/Belle

Commissioning of the ATLAS LAr Calorimeter

CLEO Results From Υ Decays

Lepton Universality in ϒ(nS) Decays at CLEO

Dario Barberis Evaluation of GEANT4 Electromagnetic and Hadronic Physics in ATLAS

Measurement of Collins Asymmetries in the inclusive production of hadron pairs

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Missing Transverse Energy. Performance with CMS

New Hadroproduction results from the HARP/PS214 experiment at CERN PS

b Physics Prospects For The LHCb Experiment Thomas Ruf for the LHCb Collaboration Introduction Detector Status Physics Program

New Results from the DREAM project

PHENIX measurements of bottom and charm quark production

Non-collision Background Monitoring Using the Semi-Conductor Tracker of ATLAS at LHC

arxiv: v2 [hep-ex] 12 Feb 2014

Recent BaBar results on CP Violation in B decays

CDF top quark " $ )(! # % & '

Transcription:

Journal of Physics: Conference Series Calibration of the BABAR CsI (Tl) calorimeter To cite this article: Jörg Marks and the Calorimeter Group of the BARBAR Collaboration 2009 J. Phys.: Conf. Ser. 60 02005 View the article online for updates and enhancements. Related content - Radiation hardness test of un-doped CsI crystals and Silicon Photomultipliers for the Mu2e calorimeter S Baccaro, A Cemmi, M Cordelli et al. - Study of radiation hardness of pure CsI crystals for Belle-II calorimeter A. Boyarintsev, Y. Boyarintseva, A. Gektin et al. - Light yield of an undoped CsI crystal coupled directly to a photomultiplier tube at 77 Kelvin J. Liu, M. Yamashita and A.K. Soma This content was downloaded from IP address 48.25.232.83 on 8/08/208 at 08:

Calibration of the BABAR CsI(Tl) Calorimeter Jörg Marks for the Calorimeter Group of the BABAR Collaboration Physikalisches Institut, Universität Heidelberg, Philosophenweg 2, D-6920 Heidelberg, Germany E-mail: marks@physi.uni-heidelberg.de Abstract. After nine years of operation, the BABAR experiment at the e + e B factory PEP-II (Standford Linear Accelerator Center) stopped data taking in April 2008. An important part of the experiment is the electromagnetic calorimeter which consists of 6580 CsI crystals doped with thallium and read out by Si-PIN photodiodes. The light yield of the CsI crystals is changing in time due to radiation exposure. In addition to the changing light yield, passive material in front of and between the crystals as well as signal thresholds during the reconstruction influence the reconstructed energies. This requires a time-dependent calibration of the calorimeter. The calibration issues are reviewed and the calibration results obtained from various data samples are presented.. Introduction The photon energy calibration of the BABAR electromagnetic calorimeter (EMC) [] is performed in three main steps. The first two, the calibration of the electronic readout chain and the chargeto-energy conversion at the single crystal level were discussed in the previous presentation [2]. This presentation describes a third calibration step applied at the cluster level. The limited shower containment due to the total CsI crystal length and the dead material in front of the calorimeter (0.3 to 0.7 radiation length (X 0 )) and between the CsI crystals lead to an energy loss of the total cluster energy measurement. Moreover, the noise suppression and the energy deposited in crystals not associated with the clusters during the reconstruction process cause an additional reduction. Therefore, the measured cluster energy has to be calibrated to the true energy of the primary electron or photon which is performed by the cluster energy calibration. 2. Cluster energy calibration 2.. Cluster energy calibration scheme The cluster calibration requires processes which allow a calorimeter-independent energy determination of the calibration signature (mostly photons). Their predicted energy has to be compared to the measured energy in the calorimeter over a range from well below 00 MeV up to the highest energies. Since no single process covers the whole energy range, photons from π 0 decays (π 0 ) produced in multi-hadron events are used in the low-energy regime (70 MeV < E < 2 GeV). The high-energy regime (400 MeV < E < 6 GeV) is covered by photons from radiative muon events (e + e µ + µ ). The calibration function C = f(e, θ ) () c 2009 IOP Publishing Ltd

depends on the energy E and the polar angle θ of the photon. The emphasis of the calibration is to correct the measured cluster energies to the absolute energy scale and to describe the measured energy scale and their dependencies by the BABAR detector simulation based on GEANT4 [3]. Due to the poor statistical coverage of the full parameter space (E, θ ), only the projections of the calibration function can be measured C = g(e ) h(θ ). (2) In a first step, the polar-angle-dependent part h(θ ) is determined using radiative muon events. After applying the polar-angle-dependent part h(θ ), the energy-dependent part g(e ) is extracted from both the π 0 and the radiative muon sample over the full energy range. 2.2. Calibration with radiative muon data Since the initial state parameters of the process e + e µ + µ are known and the muon energies are measured in a calorimeter-independent way, the reaction is kinematically over-determined. Thus, the energy and the polar angle of the photon can be determined by a kinematic fit independent of any calorimetric measurement. The predicted photon energy E fit is compared to the photon energy measured in the EMC Erec by considering the ratio distribution Erec/E fit. The average values of the ratio distributions Erec/E fit are quantitatively evaluated by the peak position of a fit of a modified Log-normal distribution [4]. In the MC simulation, the cluster calibration function is extracted using the true photon energy E true instead of the predicted one. The asymmetric line shape of the photon response distribution Erec causes systematic shifts of the ratio Erec/E fit and therefore the data-over-mc-simulation ratio has to be used to extract the calibration function. The angular part h(θ ) of the calibration function is determined in the different areas of the θ-modules (7 crystals each[2]) θ. i The following ratio yields the θ-dependent calibration function for data: h data (θ) i Erec/E fit = data(θ) i h MC (θ ) Erec/E fit MC(θ) i. (3) The function is measured for three different ranges of the photon energy (I. 0.5.0 GeV, II..0 2.0 GeV, III. 2.0 4.0 GeV) depending on the polar angle of the crystals () and is shown for the data taking period 2006 in Figure. The full lines represent the parametrizations. In order to determine the energy-dependent calibration function g(e ) for simulated data, the θ-dependent function h(θ ) is applied and g MC (E ) is obtained using the true photon energy. (θ ).085 Parametrization µ µ data, 0.5 GeV.08 µ µ data, 2 GeV µ µ data, 2 4 GeV.075.07.065 h data (E ) data g.035.03 5 5 µ µ data (2006) π 0 data (2006) Parametrization.06.055.05 0 0 20 30 40 50 60 5 2.5 2.5 0.5 0 0.5.5 ln (E / GeV) Figure. θ-dependent calibration in three energy ranges for µµ data of the data taking period 2006. Figure 2. Energy-dependent calibration for π 0 and µµ data after applying the θ- dependent part (data taking period 2006). 2

The calibration function for data is extracted from g data (E i ) = h data (θ ) E rec/e fit data(e i ) g MC (E ) h MC (θ ) E rec/e fit MC(E i ). (4) The red symbols in Figure 2 show the results obtained for the radiative muon data on a logarithmic E scale. 2.3. Calibration with photons of π 0 decays In order to measure g(e i ) in the low-energy regime, π 0 decays with the two decay photons in the same energy range (symmetric π 0 s) are selected. The distribution of the invariant mass m of the π 0 candidates is shown as the blue distribution in Figure 3 for photons in the range 70 MeV < E, E 2 < 90 MeV. At low photon energies, peaking combinatorial background appears in the π 0 signal region. The combinatorial background contribution shown in Figure 3 (red histogram) is determined by combining photons from different events. Subtracting the background from the invariant mass distribution yields the π 0 signal distribution (Figure 4). The asymmetric line Entries / MeV 0.2 0. 0.08 Entries / MeV 0.04 0.02 0.0 0.008 0.06 0.006 0.04 π 0 Signal + Background 0.004 0.02 Combinatorial Backgr. 0.002 0 0.06 0.08 0. 0.2 0.4 0.6 0.8 0.2 2 m (GeV/c ) 0 0.06 0.08 0. 0.2 0.4 0.6 0.8 0.2 2 m (GeV/c ) Figure 3. Invariant mass of all π 0 candidates (blue) and of the combinatorial background (red) for the data taking period 2006 (70 MeV < E, E 2 < 90 MeV). Figure 4. Invariant mass of the π 0 signal after subtraction of the combinatorial background contribution for the data taking period 2006 (70 MeV < E, E 2 < 90 MeV). shape of the photon response causes the invariant mass m of the π 0 candidates to be shifted by m = m(π 0 ) m, even if the photon response is perfectly calibrated. m is a function of the photon energy resolution, the tail of the photon response function and the position resolution of the photons. To account for these effects, the calibration function g data (E ) in the low-energy regime is determined as the ratio of the average invariant mass values obtained with calibrated photon energies in the MC simulation and uncalibrated energies in the data, g data (E i ) = C MC(E ) m MC (E i ) h data (θ ) m data (E i ). (5) The average values of the invariant mass distributions m are quantitatively evaluated by the determination of the peak position by a fit of a modified Log-normal distribution [4]. The blue symbols in Figure2 show g data (E i ) on a logarithmic E scale for the π 0 sample of the data taking period 2006. The full line represents the parametrization of the energy-dependent part of the photon calibration. The π 0 mass distribution is a convolution of the two asymmetric photon response functions. 3

XIII International Conference on Calorimetry in High Energy Physics (CALOR 2008) IOP Publishing This scheme implies a perfect description of the data by the MC simulation. Any deviation causes systematic shifts in the invariant mass and therefore shifts of the calibration function. Using typical discrepancies between data and MC simulation, the systematic uncertainties are estimated to be about 0.3 % in case of the π 0 sample. In case of the radiative muon sample, the systematic uncertainties decrease from 0.5 % at the lowest energy value to 0. % at the highest. [MC] [data] / m m 5 5 π 0 data 2007 cluster calib.,, cluster calib.,, crystal calib.,, different E bins / E fit )[MC] / E fit )[data] / (E 5 5 µ µ data 2006 cluster calibration crystal calibration 2.4. Cluster calibration results The invariant mass ratio of data and simulation for the π 0 sample of the 2007 data taking period before (blue symbols) and after (red symbols) applying the calibration is shown in Figure 5. In case of the green symbols, the daughter photons of the π 0 decay originate from different energy ranges. Without cluster calibration, the deviations between data and MC simulations are as large as.5 % and (E, θ )-dependent. After applying the calibration, within the uncertainties the data are well described by the MC simulation and are independent of E and θ. The data- 0 0 20 30 40 50 60 (E 0 0 20 30 40 50 60 [MC] [data] / m m 5 5 π 0 data 2007 cluster calib.,, cluster calib.,, crystal calib.,, different E bins / E fit )[MC] / E fit )[data] / (E 5 5 µ µ data 2006 cluster calibration crystal calibration (E 2.5 2.5 0.5 0 0.5 ln (E / GeV) 0.5 0 0.5.5 ln (E / GeV) Figure 5. Invariant mass ratio of data and MC simulation for the π 0 sample of the 2007 data taking period. Figure 6. Erec/E fit data-mc-simulation ratio for the radiative muon sample of the 2006 data taking period. MC-simulation ratio of the quantity Erec/E fit for the radiative muon sample of the 2006 data taking period is shown in Figure 6. As in case of the π 0 sample, after applying the calibration, the data are well described within the estimated uncertainties from the data-mc-simulation agreement. No remaining dependencies on E and θ are observed. 2.5. Validation of the cluster energy calibration In the process Σ Λ pπ the photon energy is constrained and can be used to perform an independent test of the calibration at low photon energies (50 MeV < E < 400 GeV). The photon energy is calculated using tracking and particle identification information E calc = M 2 Σ M2 Λ 2(E Λ p Λ cos θ Λ ), (6) 4

where M X denotes the invariant mass, E the energy, p the momentum of the particles and θ Λ the opening angle between Λ and. The ratio of the predicted to the measured photon energy in the projections of E and θ is measured and is found to be consistent with within the total uncertainties of about %. In the high-energy regime, the photon energy scale can be probed by the process B 0 K 0 K + π. The photon energy in this process ranges from.5 GeV to 3.5 GeV. The difference between the reconstructed energy Erec B0 of the B 0 and the beam energy Ebeam (both in the center of mass frame) E = Erec B0 Ebeam (7) is determined with quantities measured by the tracking system, particle identification and the photon energy. The measurements yield E data = ( 8.7 ± 2.3)MeV for the data and E MC = ( 8.5 ± 0.2)MeV for the MC simulation. This translates to a deviation of the photon energy scale of ( 0.35 ± 0.09)% for data. 3. Smearing of the photon response in MC data The line shape of the photon response is not fully described by the MC simulation. As a consequence, it is difficult to model selection criteria and efficiencies in physics analyses. The main goal is the improvement of the MC simulation description by tuning the MC model to data. Because of the large amount of MC data already simulated, a simplified approach is to introduce an additional term to the simulated photon response. It reflects the difference between the measurement and the MC simulation. Data of the process e + e µ + µ allows to determine a smearing matrix (Students t distribution [5]) S ij which connects the bin i of the distribution of the normalized photon response Erec/E fit in MC simulation with the bin j of the data distribution. The smearing matrix depends on the photon energy and the calorimeter area. After applying the smearing procedure to the MC data, the measured photon line shape in data is very well modeled. As an independent test, the E distribution of B 0 K 0 K + π processes is compared in data and MC simulation. The E line shape in data is very well described by the smeared MC simulation. 4. Monitoring of the EMC with muons The signatures of minimum ionizing muons (from e + e µ + µ ) measured in the EMC provide a stable signal in time with an energy loss in a CsI crystal of about 200 MeV depending on the muon energy. Selecting muons which fully pass one crystal and normalizing the muon signal by the crystal length allows to monitor the time stability of the crystal response. The average energy deposition of each crystal is determined as the peak position of a fit of a modified Log-normal distribution [4] to the muon signal. The time stability over the full time of the BABAR operation can be seen by building the average over φ of the mean energy deposition of each crystal. The muon energy increases with decreasing θ (in forward direction). Therefore, the relativistic rise of de/dx [6] leads to an increase of the average muon response in forward direction. The time dependence is illustrated in Figure 7 for the barrel part of the calorimeter. The variation over the data taking period from 200 to 2008 is about ±0.5 %. The normalized muon response can be used to determine calibration factors for each crystal. The width of the measured cluster energy of the photons from radiative muons over the predicted photon energy Erec/E fit judges the improvement in the resolution. The crystal calibration system considers the non-homogeneous light yield over the crystal length by using low energy and high energy signals (see [2]). However, in case of a calibration with muons this effect is not included and therefore the energy resolution degrades. 5

Response to µ s (Average in φ) [MeV/cm] 6.7 6.6 6.5 6.4 6.3 6.2 Run 2 Run 3 Run 4 Run 5 Run 6 Run 7 0 5 20 25 30 35 40 45 50 Figure 7. Normalized average muon response in φ as function of θ for different data taking periods. Figure 8. Normalized response Erec/E fit across three CsI crystals before and after the lateral non-uniformity correction. 5. Lateral non-uniformity correction Between the CsI crystals, dead material can not be avoided. As a result, a drop of the response Erec/E fit at the crystal edges by about 2-3 % is observed (see Figure 8 red histogram). As in case of the cluster calibration, photons of the process e + e µ + µ are used to obtain a correction function (lateral non-uniformity correction) depending on the impact position on the crystal surface. The correction function is determined in two dimensions and in five energy bins (400 MeV < E < 8 GeV). Applying the lateral non-uniformity correction, the photon response is found to be independent of the impact position as shown in Figure 8 as green histogram. The energy resolution is measured to improve by about 0 %. 6. Association of EMC cluster and DIRC information The dead material in front of the central part of the EMC varies between 0.3 and 0.6 X 0. The largest contribution is from the quartz bars of the BABAR particle identification system, an internally reflecting Cherenkov counter (DIRC) []. The fraction of photons with an electromagnetic shower starting in front of the EMC is about 8 %. These showers have a significant energy loss outside of the calorimeter. Therefore, the mean energy is smaller and a major contribution to the low-energy tail in the line shape of the photon response is observed. The additional fluctuations degrade the energy resolution. To study the calorimeter performance and the influence of resolution effects on data selection and efficiencies, it is of advantage to tag showers which started in front of the EMC. These early showers result in Cherenkov rings in the DIRC caused by the charged particles of the shower. Associating the Cherenkov rings and the EMC cluster allows for a selection of the early showering photons. In a sample of generic B 0 B 0 MC events, the fraction of true early showers in the number of reconstructed showers is 37 %. This number increases to 60 % for the e + e µ + µ sample due to the smaller number of charged tracks. In data, 2 % of the π 0 s have at least one daughter photon with a selected early shower. The average of the invariant mass distribution of these reconstructed π 0 s is by about.5 MeV smaller than the invariant mass of π 0 s without early showering photon. 7. Reference list [] B. Aubert et al. 2002 Nuc. Instr. Meth. A 479-6 [2] A. Ruland et al. 2008 these Proceedings [3] S. Agostinelli et al. 2003 Nuc. Instr. Meth. A 506 250-303 [4] E. L. Crow and K. Shimizu 987 Stat.: Textb. and Monogr. 88 [5] M. Abramowitz and I. A. Stegun 964 Handb. of Math. Func. Dover 0-486-6272-4 [6] H. Bethe and E. Bloch 935 An. d. Phys. 337 Issue 3 325-400 6