Legacy Calibration of the Automatic Weather Station Model 2 of the United States Antarctic Program

Similar documents
Antarctic Automatic Weather Station Data for the calendar year 2000

Antarctic Automatic Weather Station Data for the calendar year 1998

Exercise 1: Thermocouple Characteristics

THE NEXT GENERATION POLAR CLIMATE AND WEATHER STATION

Exercise 1: Thermistor Characteristics

Antarctic Automatic Weather Stations Abbreviated Field Report for

Evaluation of Vaisala HMP45D Humidity Probes

Smartec Pressure Sensor (bridge output)

Exercise 2: The DC Ohmmeter

Remote and Autonomous Measurements of Precipitation in Antarctica

WeatherHawk Weather Station Protocol

M. Mielke et al. C5816

How Does Weather in Antarctica Impact Me?

Meteorological Service

PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL

COMPOSITE ANALYSIS OF EL NINO SOUTHERN OSCILLATION EVENTS ON ANTARCTICA

WEATHER STATION FOR SOLAR FARM MONITORING

ECNG3032 Control and Instrumentation I

1. Mark the correct statement(s)

Robotics Errors and compensation

Instrument Test Report ITR 656. Evaluation of Milos 500 AWS unit

PMF2000 Series DESCRIPTION FEATURES APPLICATIONS MAXIMUM RATINGS MASS AIR FLOW SENSOR

The Meteorological Observatory from Neumayer Gert König-Langlo, Bernd Loose Alfred-Wegener-Institut, Bremerhaven, Germany

Performance of Radar Wind Profilers, Radiosondes, and Surface Flux Stations at the Southern Great Plains (SGP) Cloud and Radiation Testbed (CART) Site

Biosensors and Instrumentation: Tutorial 2

WEATHER MULTI-SENSOR. Vaisala Weather Transmitter WXT510. Change the Way You Measure Weather

EAS 535 Laboratory Exercise Weather Station Setup and Verification

THM80X Series Feature Application. Temperature & Humidity / Flow Measuring Specialist

ASEN 2002 Experimental Laboratory 1: Temperature Measurement and an Blow Dryer Test

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE

PRESENT STATE OF METEOROLOGICAL INSTRUMENTS IN MONGOLIA

Development of Innovative Technology to Provide Low-Cost Surface Atmospheric Observations in Data-sparse Regions

resistance in the circuit. When voltage and current values are known, apply Ohm s law to determine circuit resistance. R = E/I ( )

EVALUATION OF ANTARCTIC MESOSCALE PREDICTION SYSTEM (AMPS) FORECASTS FOR DIFFERENT SYNOPTIC WEATHER PATTERNS

Neutrinos with a cosmic ray detector. Cosmic rays with a neutrino detector

Simple calibration for ceramic sensing elements using an ME651

Automatic weather station research by Japanese Antarctic Research Expedition in East Droning Maud Land, East Antarctica.

M E 345 Professor John M. Cimbala Lecture 42

Calibration of Paroscientific Model 205 Pressure Sensor for use at Heard Island.

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

UNITED STATES ANTARCTIC PROGRAM

Antarctic Automatic Weather Station Field Report: Season

Design of Intelligent Meteorological System Based on MEMS Wang Hao 1,a, Nie Meng 2,b, Huang Qing an 3,c

* * * Table (1) Table (2)

Millivolt Output Pressure Sensors

Measurements & Instrumentation. Module 3: Temperature Sensors

BMXART0814 analog input module M340-8 inputs - temperature

MODEL HMP45C TEMPERATURE AND RELATIVE HUMIDITY PROBE INSTRUCTION MANUAL

Observational Needs for Polar Atmospheric Science

Introductions to RIC-Beijing. NAN Xuejing, CUI Xiai Meteorological Observation Center China Meteorological Administration March,2018

Amplified Low Pressure Sensors

APPENDIX 1 DESCRIPTION OF HOT WIRE ANEMOMETER

COMPOSITE ANALYSIS OF EL-NIÑO SOUTHERN OSCILLATION TELECONNECTIONS IN ANTARCTICA

Digital Multi-Day Max/ Min/Current Air and Soil Temperatures Protocol

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Validation of Water Depth Measurements at the Mass Balance Site, 2009 and 2010

AUTOMOTIVE CURRENT TRANSDUCER OPEN LOOP TECHNOLOGY HAH1BVW S/08

Designing a Thermostat Worksheet

Amplified Low Pressure Sensors

MATERIAL TEST REPORT STATIC DECAY AND SURFACE RESISTIVITY TESTING OF SOCK AND SPUNBOND SAMPLES NEW PIG CORPORATION FEBRUARY 28, 2008

2.4 The ASME measurement-uncertainty formulation

Differential Pressure Sensor

A free web support in Education. Internal resistance of the battery, r = 3 Ω. Maximum current drawn from the battery = I According to Ohm s law,

HAL501...HAL506, HAL508 Hall Effect Sensor ICs MICRONAS INTERMETALL MICRONAS. Edition May 5, DS

473-SHX Dew Point Mirror

technical bulletin Installation Instructions: Wall-Mount Dewpoint Probe ( )

Fig 1. Power Tower during Operation

CERTIFICATE OF ACCREDITATION

Understanding Hot-Wire Anemometry

for JMA/WMO Training Workshop on CALIBRATION AND MAINTENANCE OF METEOROLOGICAL INSTRUMENTS

Moisture transport to Syowa Station and Dome Fuji Station, Antarctica

Exercise 1: RC Time Constants

SENSORS CLIMATRONICS CORPORATION (631) MANUAL TACMET II WEATHER STATION, EMI 2.0 INSTALLATION 1.0 INTRODUCTION

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Using a Mercury itc with thermocouples

Complex atmospheric-lithospheric observations of acoustic emission at «Karymshina» site in Kamchatka

STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING

Experiment 5: Thermocouples (tbc 1/14/2007, revised 3/16/2007, 3/22,2007, 2/23/2009, 3/13/2011)

Measurements in Mechatronic design. Transducers

OPERATION MANUAL. 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER. Model : TM-2000

Advances in Military Technology Vol. 3, No. 2, December Prediction of Ice Formation on Earth s surface. K. Dejmal 1 and V. Répal *2.

NOVALYNX CORPORATION MODEL 110-WS-16BP BAROMETRIC PRESSURE SENSOR INSTRUCTION MANUAL

A TEST OF THE PRECIPITATION AMOUNT AND INTENSITY MEASUREMENTS WITH THE OTT PLUVIO

Electronics Resistive Sensors and Bridge Circuits

3WHB10K 3-WIRE HALF BRIDGE TERMINAL INPUT MODULE INSTRUCTION MANUAL

Low Cost Weather Stations for Developing Countries (Kenya)

Lab E3: The Wheatstone Bridge

Technology Brief 9: Capacitive Sensors

INC 331 Industrial Process Measurement. Instrument Characteristics

Ambient Weather WS-228TBH 9" Brushed Aluminum Traditional Barometer with Temperature and Humidity, Radiant Blue User Manual

Community College of Allegheny County Unit 9 Page #1. Thermocouples R1 = 1K

Exercise 1: Capacitors

WM9280. Pro Family weather station with T/H sensor, pluviometer, anemometer, PC connection and Meteotime weather forecasts until 3 days

BMXART0814 analog input module M340-8 inputs - temperature

PURCHASE OF LAB EQUIPMENTS FOR CHEMISTRY [Kindly enclose a signed copy of this document along with the TENDER Document]

Amundsen-Scott South Pole Station CLIMAT Message Project

473 Dew Point Hygrometer

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba

Chapter 1 - Basic Concepts. Measurement System Components. Sensor - Transducer. Signal-conditioning. Output. Feedback-control

Intercomparison of Mobility Particle Size Spectrometers

Transcription:

Legacy Calibration of the Automatic Weather Station Model 2 of the United States Antarctic Program G. A. Weidner 2, J. E. Thom 1, and M. A. Lazzara 1 1 Antarctic Meteorological Research Center Space Science and Engineering Center 2 Deparment of Atmospheric and Oceanic Sciences University of Wisconsin-Madison Madison, WI, USA

Stanford electronics 1978 CMOS ICs A/D Circuits Argos TX er AWS2A Model of the Stanford AWS Paroscientific Gauge

AWS2B design UW SSEC

The AWS2B version The UW AWS2B as it was designed in the mid 1980 s and is in use today. It also included an acoustic depth gauge (ADG) that measures the distance to the snow surface, thereby giving a measurement of net snow accumulation. An amplifier board was designed in 1980 to enable measurement snow temperature profiles from thermocouples.

Reason for AWS

AWS Hardware check

AWS2A Measurements Pressure counts for 5.120 seconds (two counts) 1 MHz reference oscillator for 0.1 mb accuracy Temperature - Ratiometric with precision resistor Accuracy of +/- 0.5 C over temperature range Wind speed to 0.25 meters per second Wind direction to +/- 1.5 Degrees Humidity (HMP14) +/- 5% at best

AWS2B Measurements

AWS2A retirement

Weed temperature bridge

Temperature calibration A fixed calibration point for the AWS was chosen to be 0.0 C. A 1000 ohm (at 0.0C) PRT was selected as the temperature sensor. In order to set the PRT to 0C for a particular AWS, a 0.05% resister was substituted for the PRT and the AWS temperature output was observed. The output was set to 0C by setting an offset value in the AWS onboard software that compensated for the variation of the resistors in the bridge from their stated values. Typical errors from the PRT calibration table were less than 1.0 C over the temperature range from 25C to -75C.

AWS Temperature sensors

Summary for Weeds All Weed temperature probes returned in the last three years are within the stated +/- 0.5C accuracy at 0.0C All Weeds will be matched to respective AWS electronics and cycle down to -60C in an environmental chamber.

Temperature calibration in chamber

Humidity Vaisala series of humidity probes HMP 14 1979-1982 HMP 31UT 1983-1989 HMP 35A 1989-1999 HMP 45A/D 1999-2009 HMP155 2009 HMP45 RH Measuring range: 0.8 to 100% RH RH Output scale 0 to 100% RH equals 0 to 1 VDC RH Accuracy at +20 C against factory references: +/-1% RH Field calibration against references: +/-2% RH (0 to 90 %RH); +/-3 %RH (90 to 100% RH) RH Typical long-term stability:< 1% RH / year RH Temperature dependence: +/-0.05% RH/ C RH Response time (90% at +20 C): 10s with membrane filter Each model improved on performance. Recalibrated a HMP35A deployed on Ross Ice Shelf for three years within 2% of original calibration.

Humidity Probe Summary

One returned HMP31UT from Marilyn this season: Lab reference HMP45A read 37% Marilyn HMP 31UT read 40% (this is after 25 years) HMP35A summary for 2011 and 2012 for room temperature only. HMP 35A # HMP45 reference reads 42% 1: 47 2: 44 3: 45 4: BAD 5: 44 6: 43 7: BAD 8: 43 9: 45 Yet to be done is a calibration at over the temperature range of 20C down to -60 C. Note that below -40C these probes act more like a temperature probe

Wind Speed and Direction Wind Sensors for the AWS2 stations all required a voltage output. Wind sensors typically failed at various rates depending on their location. All CPU board A/D circuits were identical so that wind sensors could be swapped without regard to AWS unit.

WS calibration for Byrd

Paroscientific 215A

The only pressure sensor used for all AWS2 stations has been the Paroscientific model 215A pressure transducer. The 215A coupled with the AWS2 provided an initial accuracy of +/-0.1 hpa. There have been four generations (my classification) of 215As, with each generation achieving better accuracy and lower long term drift. The long term drift results in an error in pressure with a negative bias due to out gassing in the reference vacuum chamber and water vapor migration through sealants in early models (later generations solved this problem). The output signal is a frequency (nominal 40 KHz) that is a function of both the air pressure and the temperature. Early models required a separate temperature measurement which the AWS2 provided with a Weed PRT laid against the outer case of the 215A. Later 215A models incorporated an onboard temperature sensor at the point of the pressure sensor. However, all AWS2B units continued to use the Weed PRT temperature to compute the pressure from the observed output frequency of the 215A. The early 215A models (the original 20 or so gauges purchased for the AWS2A units) that have been returned for calibration have a long term drift of between 2 and 6 hpa over 30 years. Second and third generation models have shown a drift of around 2 to 3 hpa over 15 to 20 years. The latest generation of 215A have less than 2 hpa of drift observed over 10 years.

Early pressure checks