Low Level Detection of Priority Inorganic Pollutants in Water. Anthony Neumann, MSc Darren Crichton, BSc

Similar documents
Low Level Detection of Priority Inorganic Pollutants in Groundwater By Darren Crichton and Anthony Neumann

IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS

Trace Analyses in Metal Matrices Using the ELAN DRC II

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages

Analytical Approaches to Arsenic

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Arsenic and Other Trace Elements in Groundwater in the Southern San Joaquin Valley of California

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Primary chimney for mercury wet-deposition collection funnel Warm air path. Glass capillary thistle tube. Mercury sample bottle

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

ALLOWAY METHOD OUTLINE

PRINCIPLE OF ICP- AES

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters

Arsenic Speciation in Chicken. Mary Carson, Ph.D. FDA/CVM/OR/ Division of Residue Chemistry

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

Determination of the Halogen Elements in the Deep UV Region of the Spectrum by ICP - OES

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Interface (backside) & Extraction Lens

EXPANDING THE CAPABILITIES OF ICP-MS. ELAN DRC-e. Uncompromised sensitivity and performance in a robust system designed for routine laboratories

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Agilent Technologies at TIAFT 2013

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

The New Agilent 7700 Series ICP-MS

Agilent 7700 Series. The Smallest, Most Powerful ICP MS Ever Made

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

A Comparison of Arsenic Speciation

A Comparison of Three Methods for Arsenic Speciation in Biological Tissues. May Nguyen Brooks Rand Labs Seattle, WA

Prof. Dr. Biljana Škrbić, Jelena Živančev

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

The 30-Minute Guide to ICP-MS

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Opening the analytical black box: Insights into interferences, corrections, and data quality

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

William O. Moellmer, Theron G. Miller, Steve Wilbur, and Emmett Soffey

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

RISK ASSESSMENT, IMPLEMENTATION AND VALIDATION OF ELEMENTAL IMPURITIES FOR ICH Q3D, USP <232><233> AND EP 5.20

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Rapid Sample Assessment and Simplified Method Development with IntelliQuant

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Proposed new USP general chapters <232> and <233> for elemental impurities: The application of ICP-MS for pharmaceutical analysis

Preparation of Food and Feed Samples for Metals Analysis

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Fergus Keenan Shona McSheehy Julian Wills

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

METHOD 6010A INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROSCOPY

APPENDIX II QUALITY CONTROL SUMMARY TABLES

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Partial Energy Level Diagrams

Analysis for Arsenic and Hexavalent Chromium:

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

METHOD 6020A INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

speciation in water Page 1

Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology

Theory, Design, and Operation of a Dynamic Reaction Cell for ICP-MS

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

Tap Water Analysis Lockwood 1 TAP WATER ANALYSIS BY ATOMIC ABSORPTION SPECTROSCOPY AND ION CHROMATOGRAPHY. Susan Lockwood

The challenges of field sensors for trace metal detection. Chris Searle Product Manager Trace2o Ltd

Transcription:

Low Level Detection of Priority Inorganic Pollutants in Water Anthony Neumann, MSc Darren Crichton, BSc

Priority List of Pollutants Consists of 16 metals and/or transitional elements. Enter the human food chain via: 1. dissolution of metal rich ores in water 2. Plant and animal uptake 3. Contamination from Industrial Processes

Arsenic 20 th most abundant element in the earth s crust. Acutely toxic and carcinogenic Inorganic forms found in water are more toxic than organic forms found in tissue Is mobile at neutral ph and in reductive or oxidizing conditions.

International guidelines for Arsenic Ranges from 7ug/L (ppb) in Australia to 50 ppb in the U.S. Current Canada guideline = 25ppb Alberta proposing a reduction to 5ppb Public Health Goal for Arsenic = 0.004ppb Problem Standard lab detection limits =0.2 to 1.0 ppb

Laboratory methods of analysis of Inorganic Pollutants Atomic Absorption One element analyzed at a time Long run times and higher costs ICP-OES Rapid analysis of several elements Not sensitive enough to meet regulatory criteria for certain elements ICP-MS Rapid analysis of all elements Low detection limit Large dynamic range

Polyatomic Interferences Are created within the plasma Results in the formation of a product that has the same mass:charge ratio as the analyte of interest e.g. 40 Ar 35 Cl + Arsenic, selenium, chromium, vanadium Can be corrected for by: 1. Mathematical corrections 2. Chemical Resolution

PE ELAN 9000 DRCII DRC gas Inlet DRC Vent Skimmer cone Torch Ion Lens DRC Quad Analyzer Quad Detector Sampler cone Entrance Aperture Exit Aperture

Chemical Resolution Ion lens system Dynamic Reaction cell Analyzing quadrupole 75 As + 40 Ar 35 Cl +

Chemical Resolution

Objective To compare the operation of an ICP-MS in Standard mode (mathematical corrections) and in Dynamic Reaction Cell mode (chemical resolution) Instrument detection limits Effect of different matrices on precision Accuracy and precision at low pollutant concentrations Effect of different matrices on accuracy

Comparison of Instrument Detection Limits for Arsenic IDLs determined in Standard mode and DRC mode at three salt concentrations: 1. 0 ppm chloride 2. 1000 ppm chloride (elevated groundwater) 3. 19000 ppm chloride (seawater)

Matrix Effect on IDLs for Arsenic Instrument detection Limit (ug/l) 8 6 4 2 0 7.898 0.077 0.349 0.533 0.028 0.035 Water 1000 ppm Cl Seawater DRC Mode Standard Mode

Matrix effect on Instrument Detection Limits (IDL) Instrument Detection Limits ug/l 0.4 0.3 0.2 0.1 0 Standard Mode 0ppm K 1000ppm K Instrument Detection Limits ug/l 0.4 0.3 0.2 0.1 0 DRC Mode 0ppm K 1000ppm K Vanadium Chromium Selenium Vanadium Chromium Selenium

Accuracy and precision of Low Level Element Detection Analysed for Vanadium, arsenic chromium, selenium Each analyte was determined with 10 replicates at 2 concentrations: 1. 0.2ug/L 2. 2.0ug/L Compared the relative standard deviation and the mean of each analyte in DRC and Standard mode

Low Level Detection of Elements % Relative Standard Deviation 60 50 40 30 20 10 0.17 0 V51 As75 Cr52 Se78/82 Se80/77 V51 As75 Cr52 Se78/82 Se80/77 0.22 0.2ug/L 0.21 0.17 0.22 0.24 0.21 0.17 0.19 0.39 1.7 2.2 2.0ug/L 2.0 2.1 1.9 2.2 1.9 2.0 DRC Standard 1.9 2.2

Detection of Arsenic in Different Matrices Four levels of arsenic (0, 2, 10, 25ug/L) were spiked into 3 salt solutions: 1. 0ppm Chloride 2. 1000ppm Chloride 3. Seawater (19000ppm Chloride)

Measured Concentration of Arsenic ug/l 30.00 25.00 20.00 15.00 10.00 5.00 0.00-5.00-10.00 * * 0 2 10 25 0 2 10 25 0 2 10 25 Actual concentration of Arsenic DRC Mode Standard Mode 0 ppm Cl - 1000 ppm Cl - 19000 ppm Cl -

Detection of Arsenic in potassium solutions Four levels of arsenic (0, 2, 10, 25ug/L) were spiked into: 1. 0 ppm potassium chloride 2. 1900ppm potassium chloride (1000 ppm of potassium) Compared DRC to Standard mode

120 % Recovery 110 100 90 DRC - no K DRC - 1000K Standard No K Standard - 1000K 80 0 2 10 25 Spiked Arsenic Concentration

Summary ICP-DRC-MS shows improved performance over standard ICP-MS: 1. DRC improves the precision of the results for low pollutant concentrations and in the presence of interfering substrates (most environmental samples) using chemical resolution 2. Accuracy of the DRC is consistent with or better than standard mode operation 3. DRC uses chemical resolution to prevent errors that may arise due to mathematical corrections in standard mode

Acknowledgements Thanks to the ICP group at Norwest: For allowing us to work on this research project while they found ways to get your analysis done on time.