Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Similar documents
Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Fast methods for the determination of ibuprofen in drug products

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

LC Technical Information

Automated and accurate component detection using reference mass spectra

Dionex IonPac AS28-Fast-4µm column

APPLICATION NOTE Goal. Introduction. Keywords

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of urea in ultrapure water by IC-MS/MS

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Charged Aerosol Detection 101

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Simultaneous Determination of Aromatic Amines and Pyridines in Soil

Determination of trace anions in concentrated hydrofluoric acid

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Determination of Carbonyl Compounds in Workplace Air

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Acclaim HILIC-10 Columns

Monitoring Protein PEGylation with Ion Exchange Chromatography

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Acclaim Mixed-Mode WAX-1 Columns

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Pharmaceutical Analysis of API and Counter Ions in Complex Formulations in a Single Injection

Discover what you re missing

Thermo Scientific Syncronis HPLC Columns. Remarkable separations guaranteed time after time

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

HILIC Method Development in a Few Simple Steps

Thermo Scientific. Acclaim Trinity P2. Column Product Manual. P/N: October Part of Thermo Fisher Scientific

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Rapid, sensitive, and easy UHPLC-MS/MS analysis of fungicides in fruit juices with QuEChERS

Expanded capabilities in ammonium and amine detection

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Acclaim Mixed-Mode WCX-1

for Acclaim Trinity TM P1

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

for Acclaim Mixed-Mode WCX-1

Determination of Polymer Modifier in Asphalt

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Inline sample preparation for the determination of anions in sodium hydroxide

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

ProPac WCX-10 Columns

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

for Acclaim Mixed-Mode HILIC-1 Column

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Determination of Additives in Carbonated Beverages

Transcription:

APPLICATION NOTE 21734 Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners Authors Aaron Lamb, Joanne Jones, Thermo Fisher Scientific, Runcorn, UK Keywords Environment, food safety, food and beverage analysis, UHPLC, HILIC, DAD Flow Cell LightPipe, UV, QA/QC, natural sweeteners, steviol glycosides, dulcoside A, stevioside, rebaudioside C, steviobioside, rebaudioside A, rebaudioside B,, Quaternary, Acclaim Trinity P1 Application benefits Sensitivity of steviol glycoside determination by HILIC UHPLC was found to increase by up to 40% by utilizing the newly developed Thermo Scientific LightPipe Diode Array Detector (DAD) Flow Cell technology. Compared to the Thermo Scientific UltiMate 3000 system, the %RSD of peak width at half height was found to be up to 80% lower in the case of dulcoside A using the Thermo Scientific Vanquish Flex system s improved proprietary pump technology. The unique chemistry of the Thermo Scientific Acclaim Trinity P1 columns allows for the HILIC separation of highly polar and structurally similar steviol glycosides, which are difficult to separate on traditional reversed-phase columns. Goal To demonstrate the seamless method transfer from an RS system to a system using an Acclaim Trinity P1, 3 µm column with unique chemistry and functionality for the HILIC analysis of steviol glycosides. To demonstrate the increased reproducibility and sensitivity of the system obtained by improved pump features and newly developed DAD Flow Cell LightPipe technology.

Introduction The inter- and intra-laboratory transfer of methods between different instruments is common over the lifetime of an HPLC method. The seamless transfer of methods is vital to ensure re-validation costs remain minimal. In this analysis we demonstrate the transfer of a method from an system to a UHPLC system. The UHPLC system provides the user the flexibility expected from a quaternary, low pressure mixing pump. However, the additional benefits of improved autosampler and pump technology result in unrivalled retention time precision providing the user with greater data confidence and more freedom in method development and application transfer. The Acclaim Trinity P1 column is a novel, high-efficiency, silica-based column designed for maximum flexibility in method development through unique chemistry and functionality. The inner-pore surface is modified with an organic layer that provides both reversed-phase and anion-exchange properties, while the outer-pore surface is modified with cation-exchange functionality. This method uses the column in hydrophilic interaction liquid chromatography (HILIC) mode, not only for the separation of six steviol glycosides, but with the additional advantage of volatile mobile phases that can be used for charged aerosol or mass spectrometric detection (CAD or MS) giving further flexibility to the method. Steviol glycosides are natural sweeteners widely used for beverages, which are significantly sweeter than sugar and do not have a caloric value. The chromatographic separation, however, can be challenging as they are structurally highly related (Figure 1). Due to their strong polarity, analysis by reversed-phase HPLC is particularly challenging. This method demonstrates the full resolution of six steviol glycosides using a HILIC based method. Experimental Consumables and apparatus Acclaim Trinity P1, 3 µm HPLC column, 100 mm 2.1 mm (P/N 071389) Acclaim Trinity P1, guard, 3 µm, 10 mm 2.1 mm (P/N 071391) Acclaim SST Guard cartridge holder (P/N 069580) LC/MS grade 18 MΩ water from Thermo Scientific Barnstead Smart2Pure Water Purification system (P/N 50129845) Fisher Scientific HPLC grade acetonitrile (P/N A/0626/17) Fisher Scientific Formic acid (P/N 10559570) Fisher Scientific Ammonium formate (P/N 10131850) Thermo Scientific Virtuoso 9 mm wide opening, 2 ml screw thread vial and cap kit (P/N 60180-VT400) All standards were purchased from a reputable supplier. Instrumentation Analyses were performed using an UHPLC system consisting of: LPG-3400RS Pump (P/N 50436) WPS-3000RS Autosampler (P/N 58410) TCC-3000RS Column Oven (P/N 57300) DAD-3000RS Diode Array Detector (P/N 5082.0020) Analytical Flow Cell for DAD-3000, 13 µl, 10 mm (P/N 6082.0100) Analyses were also performed using a UHPLC system consisting of: Quaternary Pump F (P/N VF-P20-A) System Base (P/N VF-S01-A) Split Sampler FT (P/N VF-A10-A) Column Compartment H (P/N VH-C10-A) Active Pre-heater (P/N 6732.0110) Diode Array Detector HL (P/N VH-D10-A) Figure 1. Chemical structure of rebaudioside A, rebaudioside B and stevioside. LightPipe Flow Cell, 2 µl 10 mm (P/N 6083.0100) Thermo Scientific Virtuoso Vial Identification System (P/N 60180-VT-100) 2

Software Thermo Scientific Chromeleon Chromatography Data System 7.2 SR4 Sample preparation Solutions of the six compounds were prepared by dissolving a known amount in mobile phase to produce 2 mg/ml or 1 mg/ml (rebaudioside C) primary solutions. A mixed spiking solution was used to assess both systems and was prepared in mobile phase at a concentration of 0.19 mg/ml per compound. Vial labelling was supported by the Virtuoso Vial Identification System. UHPLC conditions HPLC columns: Acclaim Trinity P1, 3 µm HPLC column, 100 mm 2.1 mm Acclaim Trinity P1, guard, 3 µm, 10 mm 2.1 mm Mobile phase: 10 mm ammonium formate ph 3.0 / acetonitrile (19:81 v/v) Flow rate: 0.30 ml/min Column temperature: 40 C Pre-heater temperature: 40 C Thermostatting mode: ForcedAir Injection volume 5 µl UV detection: 210 nm Mixer: 350 µl static + 50 µl capillary Results and discussion Relative separation of all six sweeteners was achieved within 10 minutes on both the system and the system using an Acclaim Trinity P1 column and the same chromatographic method (Figure 2). mau 80 70 60 50 40 30 20 10 0-10 -20 120 10 80 60 40 20 0-20 -40-60 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 Figure 2. Chromatograms showing the separation of six sweeteners on both the system (top trace) and the system (bottom trace). Min 8.0 9.0 10.0 In the current configuration, both systems are equipped with quaternary, low pressure mixing pumps with very similar dwell volumes. The average component retention time of n=6 injections for the two systems were found to be comparable and therefore no adjustment to system dwell volume was required (Figures 2 and 3). 8.00 7.00 6.00 5.00 4.00 3.00 2.00 1.00 Figure 3. Average retention times (n=6) for six sweeteners on the system and system. 3

The system and system both exhibit comparable peak widths at 50% height, demonstrating similar chromatographic efficiencies of the two. In both cases the majority of peak widths at 50% height are less than 0.2 minutes (Figure 4). 0.25 0.20 0.15 0.10 0.05 Figure 4. Average peak widths (min) at half height (n=6) for six sweeteners on the system and system. The autosampler utilizes the proprietary Thermo Scientific SmartInject technology to reduce flow inconsistencies during injection and pressure shocks to the HPLC column. The improvements adopted to the pump technology differentiate it in terms of precision. When comparing the %RSD of peak widths at 50% height for six injections it is apparent that the system is superior to the system, leading to greater confidence in analytical results (Figure 5). 0.35 0.30 0.25 0.20 0.15 0.10 0.05 The system also exhibits the new LightPipe flow cell technology designed for the Diode Array Detector, which provides the user with increased sensitivity for analytes due to the very long light path and minimum peak dispersion due to small internal volume. These features coupled with total internal reflection maximize the incident UV light intensity, thus increasing the sensitivity for low level analytes. This can be observed when comparing peak heights and peak areas for the analysis of steviol glycosides. Here the Vanquish Flex system was found to increase peak heights and areas by up to 40% compared to the system when analyzing the same standard at the same time (Figures 6 and 7). 12 10 8 6 4 2 Figure 6. Average peak height (n=6) for six sweeteners on the system and system. 18.00 16.00 14.00 12.00 1 8.00 6.00 4.00 2.00 Figure 7. Average peak area (n=6) for six sweeteners on the system and system. Figure 5. %RSD of peak widths at half height (n=6) for six sweeteners on the system and system.

Conclusions Sensitivity of steviol glycoside determination by HILIC UHPLC was found to increase by up to 40% by utilizing the newly developed DAD Flow Cell LightPipe technology. Compared to the system, the %RSD of peak width at half height was found to be up to 80% lower in the case of dulcoside A using the Vanquish Flex system s improved pump and autosampler technology. The unique chemistry of the Acclaim Trinity P1, 3 µm columns allows for the HILIC separation of highly polar and structurally similar steviol glycosides, which are difficult to separate on traditional reversed-phase columns. Find out more at thermofisher.com For Research Use Only. Not for use in diagnostic procedures. 2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN21734-EN 1017S