Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Similar documents
Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit

USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns

U.S. EPA Method 8270 for multicomponent analyte determination

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph

Understanding Gas Chromatography

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Methanol Extraction of high level soil samples by USEPA Method 8260C

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Solid Phase Micro Extraction of Flavor Compounds in Beer

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Author. Abstract. Introduction

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

NCFS Sample Preparation and Instrumental Parameters

Pittcon P

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Analytical Trap Comparison for USEPA Method 8260C

STANDARD OPERATING PROCEDURES

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Ultra-Inert chemistry for Trace Level Analysis

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles

Maximizing Sample Throughput In Purge And Trap Analysis

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a Combination of New Software and Low-Bleed Columns

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

GC-CI-MS analysis of TMS derivatives

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Selection of a Capillary

The Role of the Carbon Cartridge in the One-Pass System

Headspace Technology for GC and GC/MS: Features, benefits & applications

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Optimizing. Abstract: is standardd. procedures. altered to

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

DETERMINATION OF PHENOLS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) Revision 1.

Identifying Disinfection Byproducts in Treated Water

STANDARD OPERATING PROCEDURES

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Agilent G3212 GC-APCI Source

Volatile Organic Compounds in Every Day Food

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

ENVIRONMENTAL analysis

Selection of a Capillary GC Column

Highly Sensitive and Rugged GC/MS/MS Tool

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

GAS CHROMATOGRAPHY (GC)

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Transcription:

Application Note #1820230 Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS GC-MS Introduction USEPA Method 8270 [1] for semivolatiles is is used by by laboratories to measure a mixture of of acids, bases, and and neutrals in a wide variety of matrices. Sample extracts can can range from clean water to sludge coming from Superfund clean-up sites. The complexity of of these extracts demand a robust instrument that is is easy to to operate and maintain. Adding to method complexity is is the uncertainty in in both cost and supply of helium, forcing some laboratories to to consider hydrogen as carrier gas. Hydrogen is is not not an an inert inert gas; it is reactive and can be an explosion hazard if if allowed to build up in either the GC oven or or manifold of of a a mass spectrometer. The SCION Bruker Helium-free SCION Helium-free Package will Package will ensure safe routine operation, with little or or no no performance change for EPA Method 8270. Scion s Bruker s unique unique axial axial ion ion source design provides excellent robust operation and and minimizes unwanted protonation and spectral distortions. In addition, the Gas Chromatograph with Scion Bruker Split/ Split/ Splitless (SSL) injector and inert pathway prevent compound degradation and reactions with hydrogen. Author Ed George, Bruker Daltonics Inc, Fremont, USA Keywords Environmental Testing GC/MS Instrumentation and Software SCION Select SQ SCION 436-GC 8270 PEAK Precision 500 Hydrogen Generator SW-846 Semi-volatiles Hydrogen Carrier Gas CP-8400 Autosampler Bruker MSWS MSWS 8 Software 8 Software

Table 1: Gas Chromatograph and Mass Spectrometer Conditions. Gas Chromatograph Bruker SCION SCION 436 436 Inlet Bruker SSL (Split/Splitless) SSL (Split/Splitless) Mode Pulsed Split Injection, 0.5 µl Injection Temp 290 C Pressure Pulse Split Flow Gas Type Gas Supply 40 psi for 0.3 min 70 ml/min Hydrogen PEAK Precision 500 TRACE Hydrogen Generator Inlet Liner SGE 4 mm ID Liner w/ goose-neck, p/n 092017 Oven Oven Ramp C/min Temp C Hold (min) Initial 45 3.0 Ramp 1 25 100 1.0 Ramp 2 10 310 1.0 Total Run Time: 28 min Column SGE BP-5 MS, 20 m x 0.18mm x 0.18 µm, p/n 054301 Constant Flow 1.0 ml/min Hydrogen Carrier Mass Spectrometer Solvent Delay Scan Range Dwell Time SCION Select SQ with Helium-free Package 2 minutes 45-500 da 250 ms Ion Source Temp 330 C Transfer line Temp 300 C Experimental The recommended instrument and operating parameters are listed are listed in Table in Table 1. 1. When using hydrogen carrier for for the first time, more hydrocarbon background is is observed but but will will clean up over time. This initial background can can be be reduced significantly by increasing the ion source temperature to to 350 C, with hydrogen column flow at at 44 ml/min and and filament filament turned ON turned for 4 hours. ON for 4 hours. Calibration standards were obtained from Restek Calibration standards were obtained from Restek Corporation, Bellefonte PA, MegaMix p/n 31850, containing 76 target Corporation, Bellefonte PA, MegaMix p/n 31850, containing 76 target compounds. The calibration for most analytes compounds. The calibration for most analytes ranged from 1 ranged from 1 to 200 ppm. Internal and surrogate standards to 200 ppm. Internal and surrogate standards were added at were added at a concentration of 40 ppm. Solvent: a concentration of 40 ppm. Solvent: Dichloromethane. Dichloromethane. A pulsed split injection was used to to minimize contact and residence residence time of compounds time of compounds inside the inside inlet. the This inlet. is critical This in is and critical hydrogen in hydrogen carrier gas, carrier due to gas, its low due viscosity to its low and viscosity tendency and to tendency react with to dichloromethane react with dichloromethane and form HCI. and Figure form 1 illustrates HCI. Figure the pulsed 1 illustrates split injection the pulsed technique split injection technique with the Scion Bruker SSL. SSL. The single goose-neck 4 mm open inlet liner is commonly The single goose-neck 4 mm open inlet liner is commonly used with Method 8270. It does not contain glass wool. used with Method 8270. It does not contain glass wool. Glass wool could contribute to compound degradation, Glass wool could contribute to compound degradation, especially when using reactive hydrogen as carrier especially when using reactive hydrogen as carrier gas. gas. Pulsed Split Injection High pressure (40 psi) during injection Split flow out Figure 1: During the pulsed split injection, the inlet is set to 40 psi with split flow at 70:1. Under this condition, the column flow is near 5 ml/min, allowing rapid transport of analytes to column.

Results In order for hydrogen carrier gas to be considered for use, Method 8270 specifications in terms of tuning, resolution, calibration, peak shape (Gaussian Factor), and system performance checks (SPCCs) must be met. In addition, the solvent specified in the method is dichloromethane (DCM), therefore essential to minimize degradation in the inlet. The GC/MS system must also produce mass spectra that match NIST library and demonstrate robust operation in heavy matrices. Tune with Hydrogen Carrier Gas 75% 50% 25% DFTPP 8270A Report 15.045-15.058 min, Scans: 3236-3240, Merged 100% 197.9 1.092e+7 77.0 5.740e+6 74.9 970409.744 109.9 2.998e+6 186.0 1.580e+6 255.0 5.648e+6 275.0 2.335e+6 441.9 7.521e+6 443.0 1.280e+6 The SCION with Helium-free Package can be auto-tuned normally and used with the tune-to-target feature for passing DFTPP tune. Figure 2 illustrates a DFTPP spectrum and report generated with the EnivroPro software. The viscosity of hydrogen changes at a much slower rate vs. temperature than helium, and is an advantage for maintaining good solubility in the stationary phase during the GC oven ramp. Also, at a flow of 1.0 ml/min using the 20 meter 0.18 mm ID column, an optimal linear velocity for peak separation is maintained. This assures that the peak resolution criteria in the method is met, as illustrated in Figure 3 for Benzo(b) and Benzo(k)fluoranthene. The peak Gaussian Factor or peak Tailing Factor for Pentachlorphenol demonstrates the inert behavior of the system. 0% 100 200 300 400 500 m/z Lab File ID 0326_S2_H2_BFB2.xms Injection Date: 3/26/2013 Injection Time: 14:24 m/z Acceptance Criterion Value Pass/Fail 51 30-60% of m/z 198 46.95 PASS 68 less than 2% of m/z 69 0.02 PASS 69 present 40.10 PASS 70 less than 2% of m/z 69 1.44 PASS 127 40 to 60% of m/z 198 40.97 PASS 197 <1% of m/z 198 0.33 PASS 198 base peak 100.00 PASS 199 5-9% of m/z 198 6.93 PASS 275 10-30% of m/z of m/z 198 21.40 PASS 365 >1% of m/z 198 2.34 PASS 441 present and <m/z 443 87.90 PASS 442 >40% of m/z 198 68.94 PASS 443 17-23% of m/z 442 17.02 PASS Figure 2: Auto-tune and tune-to-target work without modification using hydrogen carrier. DFTPP report easily generated from EnviroPro Software Package.

EnviroPro Software Reports Calibration with Hydrogen Carrier Gas Figure 3: SPCC Peak Gaussian report for Pentachlorophenol (left) and Peak Resolution Test for Benzo(b) and (k) Fluoranthene generated using EnviroPro Software. The SPCCs and initial calibration criteria are also met with hydrogen carrier. For the SPCCs the relative response factors (RRFs) are measured and compared to the criteria given in Table 2. Table 2: Relative Response Factors with Hydrogen Carrier Gas. Compound Method 8270 Min. RRF N-nitroso-di-n-propylamine 0.05 0.715 Hexachlorocyclopentadiene 0.05 0.178 2,4-Dinitrophenol 0.05 0.070 4-Nitrophenol 0.05 0.134 RRF on SCION Select SQ An example calibration curve of Hexachlorocyclopentadiene is shown in Figure 4, with % RSD and corr. coefficient (r 2 ) for this compound of 3.9% and 0.9998, respectively. Figure 4: Calibration curve for Hexachlorocyclopentadiene with % RSD and corr. coefficient (r2) of 3.9% and 0.9998, respectively. The overall average % RSD of all 76 compounds was 8.5%. Calibration for the critical quality control (QC) check and SPCC compounds are listed in Table 3, along with some %RSDs shown for very active compounds in Figure 5. Table 3: Quality control check compounds for Method 8270 pass using hydrogen as carrier gas. QC Check Compounds Corr. Coeff. Avg. RRF % RSD CCC SPCC Phenol 0.9980 0.722 11.9 PASS Benzene, 1,4-dichloro- 0.9985 0.980 7.6 PASS 1-Propanamine, N-nitroso- N-propyl- 0.9991 0.719 8.5 PASS Phenol, 2-nitro- 0.9968 0.116 13.3 PASS Phenol, 2,4-dichloro- 0.9981 0.114 8.4 PASS 1,3-Butadiene, 1,1,2,3,4,4- hexachloro 0.9991 0.315 7.0 PASS Phenol, 4-chloro-3-methyl- 0.9958 0.147 9.8 PASS Hexachlorocyclopentadiene 0.9998 0.178 3.9 PASS Phenol, 2,4,6-trichloro- 0.9997 0.142 8.0 PASS Acenaphthene 0.9997 1.16 4.9 PASS Phenol, 2,4-dinitro- 0.9993 0.070 14.5 PASS Phenol, 4-nitro- 0.9940 0.134 6.8 PASS Phenol, pentachloro- 0.9994 0.082 14.2 PASS Fluoranthene 0.9998 1.42 6.3 PASS Di-n-octyl phthalate 0.9998 0.199 4.6 PASS Benzo[a]pyrene 0.9993 1.57 14.8 PASS The SPCC compounds must have average RF > 0.05 for initial calibration The CCC compounds must have % RSD of RRFs for the initial calibration < 30% The overall average of all compound % RSDs must be < 15% PASS PASS PASS

Calibration Response Factors for Active Compounds 10 9 8 7 6 5 4 3 2 1 0 Hexachlorocyclopentadiene Dimethyl phthalate % RSDs 1-200 ppm CAL Range Toluene-2,6-dinitro- Toluene 2,4-dinitro- Diethyl Phthalate Dibutyl phthalate Benzyl butyl phthalate Bis(2-ethylhexyl) phthalate Di-n-octyl phthalate Figure 5: Active compounds show calibration response factors less than 10% indicating minimal reactivity with liner and ion source with hydrogen carrier and dichloromethane solvent. Spectral quality may be problematic using hydrogen because unwanted protonation or other reactions in the ion source are possible. The Scion Bruker axial ion ion source design with Helium-free Package minimizes these reactions, as evidenced by good quality library matches to NIST. During this study, all 76 compounds were detected using automated AMDIS deconvolution/library search against NIST 11. Figure 6 illustrates example library match hits at 900+ for 1,2,4-Trichlorobenzene and Dibutyl phthalate. Excellent Library Matches with Hydrogen Carrier Gas % RSD Method 8270 performance was tested using repeated injections of a contaminated sludge extract. A total of fifty injections were made, with an injection of the continuing calibration check (CCC) after every 10 extracts. The calculated concentration of the surrogates (target 40 ppm) and the % differences in response of the compounds were compared to the initial injection and then plotted. Figure 7 shows a TIC of the sludge extract. Figures 8 illustrates an example surrogate (2,4,6-Tribromophenol) and Figure 9 shows percent differences observed for the CCCs. As can be seen from the plots, excellent robust operation of the SCION Select SQ is maintained with hydrogen carrier. Sludge Extract Injections Figure 7: A Method 8270 sludge extract (right) was injected 50 times. A CCC was injected between every 10 extract injections for monitoring the % difference of surrogate and CCC compounds. Surrogate Standard Recovery 45.0 SS-2,4,6-Tribromophenol 40.0 Concnetration, ppm 35.0 30.0 25.0 20.0 15.0 10.0 5.0 0.0-5.0-10.0 1 2 3 4 5 6 7 8 9 10 Observed Conc 36.9 38.7 37.1 37.2 36.3 39.7 36.1 37.7 36.5 38.4 % Difference -3.8 0.9-3.1-2.9-5.2 3.7-5.7-1.6-4.7 0.2 Figure 8: The target concentration = 40 ppm, with % difference measured relative to initial CCC before extract injections. Figure 6: NIST library matches for 1,2,4-Trichlorobenzene and Dibutyl phthalate.

Bruker Daltonics is continually improving its products and reserves the right to change specifications without notice. BDAL 04-2013 Continuing Calibration Check Stability Average % Difference for CCCs Phenol, pentachloro- Phenol, 4-chloro-3-methyl- Phenol, 2-nitro- Phenol, 2,4-dichloro- Phenol, 2,4,6-trichloro- Phenol Fluoranthene Diphenylamine Di-n-octyl phthalate Benzo[a]pyrene Benzene, 1,4-dichloro- Acenaphthene 1,3-Butadiene, 1,1,2,3,4,4-hexachloro- References [1] USEPA Method 8270D, Revision 4, February 2007 Acknowledgements PEAK Scientific for use of the Precision 500 TRACE Hydrogen Generator -100-80 -60-40 -20 0 20 40 60 80 100 Average % Difference Figure 9: Target concentration = 20 ppm. The plot shows the average % difference for the base neutral and acid compounds in the CCC mixture over 50 matrix injections. Conclusion Method 8270 is is a challenging method covering a a wide variety of compound classes and matrix types. As As the the cost and and scarcity of helium rises, laboratories have begun to to convert methods to hydrogen carrier gas, due to to the ability to to safely produce it it on demand with generators. Since hydrogen is is aa reactive gas and can develop explosive build-up in in certain situations, a GC/MS solution must be be safe and and also also produce data that pass both the qualitative and quantitative aspects of the method. The Bruker SCION Select SQ has demonstrated excellent The SCION Select SQ has demonstrated excellent performance for Method 8270. The axial ion source, as well performance for Method 8270. The axial ion source, as well as the pulsed-split injection technique with the Bruker SSL, as the pulsed-split injection technique with the Scion SSL, was shown to produce excellent library-searchable mass was shown to produce excellent library-searchable mass spectra, passing quality control criteria, and robust operation spectra, passing quality control criteria, and robust operation in heavy matrices. in heavy matrices. For research use only. Not for use in diagnostic procedures. Scion Instruments Fremont, CA USA Phone +1 (510) 683-4300 Fax +1 (510) 490-6586 sales@scioninst.com www.scioninstruments.com