Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

Similar documents
A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Application Note S/SL

Virtually Particle-Free Rt -Silica BOND Columns

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Chromatography. Gas Chromatography

Introduction to Capillary GC

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Monomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Chapter 31 Gas Chromatography. Carrier Gas System

Muliple Gas#3 plus Sulfur GC Configuration

Application Note 032

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Introduction to Capillary GC

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Sulfotepp impurities in Chlorpyrifos EC formulations

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

Understanding Gas Chromatography

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

GC Instruments. GC Instruments - Sample Introduction

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

The Pulsed Flame Photometric Detector for analysis of phosphorus pesticides in wastewater and food commodities

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

STANDARD OPERATING PROCEDURES

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

1.0 Reactive Gas Chromatography Design and Setup Procedure for Catalyst Pelletization & Inlet Packing Software and Method.

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

Organochlorine Pesticides by GCxGC-ECD

Activity in the FID Detection Port: A Big Problem if Underestimated

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Standard Test Method for Determination of Trace Thiophene in Refined Benzene by Gas Chromatography 1

Chemistry Gas Chromatography: Separation of Volatile Organics

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Permanent Gas Analysis by Gas Chromatography Vacuum Ultraviolet Spectroscopy

PROCESS AND ENVIRONMENTAL APPLICATIONS OF GAS CHROMATOGRAPHY WITH THE REDUCTION GAS DETECTOR

2401 Gas (liquid) Chromatography

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Real Time On-Site Odor and VOC Emission Measurements Using a znose

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Application Note. Abstract. Authors. Introduction

Chapter 27: Gas Chromatography

Polymer plants continue to seek ways to increase production and efficiency without compromising safety.

CHAPTER 6 GAS CHROMATOGRAPHY

Highly Sensitive and Rugged GC/MS/MS Tool

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Principles of Gas- Chromatography (GC)

TO-15 Checklist Determination of VOCs in Air by GC-MS

HYPHENATED TECHNOLOGY GUIDE

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Pittcon P

Making Improvements to Today s Natural Gas Analysis Systems

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

U.S. EPA Method 8270 for multicomponent analyte determination

REAL TIME MEASUREMENTS OF ACETYLENE IN ETHYLENE MANUFACTURING PROCESS. FIELD TRIALS OF THE TDLAS BASED ANALYZER

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Maximizing Sample Throughput In Purge And Trap Analysis

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Agilent G2350A Atomic Emission Detector (AED)

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

costech instruments ECS 4010 Elemental Combustion System CHNS-O

STANDARD OPERATING PROCEDURES

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Gas Chromatography. Introduction

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Transcription:

Application Note 36720111 Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary PLOT Column Keywords Pulsed Flame Photometric Detector (PFPD) Carbonyl Sulfide (COS) SPRO-Select PLOT Column Propane-propylene matrix Introduction High grade ethylene and propylene feedstocks are used in production of a variety of fine chemicals and end products. Determination of low-level sulfur impurities, such as S and COS, in the feed is important because they can corrode equipment, poison catalysts during production, and affect polymer yield. Analysis of volatile sulfur compounds in light hydrocarbon streams is traditionally done by gas chromatography (GC) with sulfur-selective detection, and is generally very successful for most feedstocks. However, with a propane-propylene feed, most available GC columns will not separate the impurities from the C3 matrix, will irreversibly absorb S, or result in poor peak shape for the sulfur compounds. Some solutions require additional equipment for heart-cutting or even two analytical lines with two detectors. A turn-key system is now available, which is based on the Pulsed Flame Photometric Detector (PFPD) (1) and a new capillary PLOT column (2). The column provides complete baseline separation of the volatile sulfur impurities from the propane-propylene matrix, sharp, symmetric peak shape, and excellent response for S. The PFPD provides unambiguous sulfur detection with long-term stability and minimal maintenance. System Description All data were acquired using the OI Analytical SPRO-Select System shown in Figure 1. The SPRO-Select is a fully integrated GC system based on the Agilent 7890 GC that incorporates calibration, QA/QC capability, automated sample introduction, baseline separation of the C3 matrix from the target compounds, and compound detection and quantitation. Presented at the 2011 Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy Atlanta, Georgia, March 13 18, 2011

Figure 1. The OI Analytical SPRO-Select is designed to detect and quantify sulfur species in gas-phase samples. Sample Introduction An automated air-actuated 4-port sample selection valve is programmed to select either the analytical sample or the calibration standard generated by the integrated permeation oven. The sample is introduced into the analytical system using a 6-port valve and a 1-mL gas sample loop. The sample is then introduced onto the GC column through the OI split/splitless volatiles interface optimized for low-volume injections. The entire sample pathway resides inside a heated valve oven, and is treated with Sulfinert surface coating to minimize adsorption of sulfur on system surfaces. OI Volatiles Interface The OI volatiles interface is designed specifically for introducing low-concentration gas samples at varying split ratios from a gas sampling valve. The temperature of the heated inlet, the carrier gas flow for the column, and the split ratios are controlled directly through the 7890 keypad or via ChemStation. The column installs quickly and easily from inside the GC oven. Permeation Oven The permeation oven is fully integrated into the system with temperature and dilution gas flow rate controlled through the 7890 keypad or ChemStation. The oven accommodates up to five individual permeation devices and is used to generate gas-phase analytical standards at point-of-use for either automated calibration or sequenced introduction of QA/QC check standards. A pure compound is sealed inside a permeation device (tube or wafer) and diffuses across the permeable membrane at a defined rate for a given temperature. All permeation devices are supplied with a certificate stating the precise permeation rate and uncertainty range at a given reference temperature. The permeation oven is held at a constant temperature to produce a constant diffusion rate. A precise flow rate of dilution gas, controlled using the AUX EPC flow control module, is passed through the oven and across the permeation device; the result is an accurate gas standard always available at point-of-use without the need for expensive standards in cylinders. Detection The SPRO-Select system is configured with the OI Analytical Pulsed Flame Photometric Detector (PFPD) for detection and quantitation of individual sulfur compounds. The linear, equimolar response is used to generate a single response factor for sulfur, which is then used to quantify the sulfur content in both known and unknown sulfur species. When configured for sulfur detection, the PFPD produces simultaneous, mutually selective sulfur and hydrocarbon chromatograms from a single detector. The constant pulsing of the flame results in a self-cleaning detector with long term stability, virtually no coking, and minimal maintenance requirements. 2

Chromatography Baseline separation of the sulfur contaminants from the matrix is achieved using the CP8575 Agilent J&W Select Low Sulfur PLOT column. Unlike other PLOT columns, this new column completely separates the S and COS sulfur species from the propane-propylene matrix allowing for unambiguous, low-level identification and quantitation without interference from the C3 hydrocarbons. Experimental All instrument operating conditions are shown in Table 1. The instrument was calibrated for S and COS using certified wafer-type permeation devices. The permeation oven was held at a constant temperature of 40 C, and the concentration of S and COS were varied by changing the dilution gas flow rate through the permeation oven. Ten replicate injections were made for each calibration point to confirm that the gas flow rate and new concentration had reached equilibrium for each level, and the average peak area was used to generate a linear calibration curve. The calibration ranges for S and COS were 0.14 to 5.60 ppmv and 0.39 to 7.80 ppmv, respectively. A mid-point standard with S at 1.40 ppmv and COS at 3.90 ppmv was generated for the long-term stability study. Details of the permeation oven conditions and calibration standards are shown in Table 2. Table 1. Instrument Configuration and Operating Conditions Permeation Oven Automated Injection System Volatiles Interface GC Column Oven Program Sulfur Detection SPRO-Select GC System 40 C Helium dilution gas Dilution gas flow rate 5 to 200 ml/minute 4-port selection valve 6-port GSV with 1-mL, Sulfinert -coated sample loop Automated, air actuated valves All lines Sulfinert coated Valve oven temperature 110 C 200 C Split mode Split ratio 50:1 Sulfinert coated CP 8575 Agilent J&W Select Low Sulfur PLOT Column 60-m x 0.32-mm ID Helium carrier gas, 2 ml/minute 65 C for 6.5 minutes 10 C/minute to 185 C Hold for one minute Total run time 19.5 minutes Pulsed Flame Photometric Detector (PFPD) 2-mm combustor, BG-12 filter, R1924 PMT Detector base temperature 200 C /air ratio tuned for optimum sulfur emission 6-24 msec sulfur gate (linear mode) 1-3 msec hydrocarbon gate 3

Table 2. Generation of Analytical Standards Standards Generation Permeation Oven Permeation Devices S wafer device, permeation rate = 17.8 ng/minute COS wafer device, permeation rate = 87.6 ng/minute Dilution Gas Helium Reference Temperature 30 C Permeation Oven Temperature 40 C Valve Oven Temperature 110 C Volume of Sample Loop 1 ml Dilution Gas Flow Rate (ml/minute) S Concentration (ppmv) COS Concentration (ppmv) Calibration Standards 200 0.14 0.39 100 0.28 0.78 50 0.56 1.56 25 1.12 3.12 10 2.80 7.80 5 5.60 Out of range, not used Long-Term Repeatability Standards 20 1.40 3.90 Results and Discussion All analytical results are summarized in Table 3. Table 3. Analytical Results Obtained with the SPRO-Select GC System Parameter Measured S Result COS Result Calibration Linearity Long-Term Repeatability n = 2,400 Signal-To-Noise Estimated Minimum Detectable Concentration Analytical Standards Range 0.14 to 5.60 ppmv R 2 = 0.9999 Peak Response RSD = 3.7% Concentration = 0.14 ppmv S/N = 4 0.14 ppmv with 50:1 0.03 ppmv with 10:1 Range 0.39 to 7.80 R 2 = 0.9996 Peak Response RSD = 1.6% Concentration = 0.39 ppmv S/N = 16 0.10 ppmv with 50:1 0.02 ppmv with 10:1 Calibration The 6-point calibration curve for S covered a range from 0.14 to 5.60 ppmv, and had a linear correlation coefficient (R 2 ) of 0.9999. The calibration curve for COS covered a range from 0.39 to 7.80 ppmv, with an R 2 value of 0.9996. Each point on the curves represents the average of 10 replicate injections; the relative standard deviation (RSD) for each set of replicates was under 5%. Figures 2 and 3 illustrate the calibration curves and linear R 2 values for the two analytes. 4

Figure 2. Calibration of S using a certified wafer-type permeation device and the permeation oven Figure 3. Calibration of COS using a certified wafer-type permeation device and the permeation oven 5

Long-Term Stability A long-term stability study was done by making continuous sequential injections of a mid-level standard over a period of approximately two months, roughly 2,400 injections. Every 10 th injection, about 240 data points were plotted to illustrate the system s stability over time. The RSD for S and COS for all 2,400 analyses were 3.7% and 1.6%, respectively. Figure 4 is a chart of every 10 th data point collected over the two-month period, and illustrates the long-term stability of the system. In Figure 5, chromatograms from approximately every 100 th analysis (21 chromatograms) are overlaid to demonstrate the consistency in peak shape, retention time, and baseline stability. Figure 4. Chart of every 10 th data point collected over two-month test period, illustrating the long-term stability of the system 6

Figure 5. Chromatograms from approximately every 100 th analysis overlaid, demonstrating consistency in peak shape, retention time, and baseline stability During this period, the detector air and hydrogen gas tanks each has to be changed once, and on one occasion, the entire system lost power during an electrical storm. As soon as the gas or power was restored, the sequence was restarted immediately and data collection resumed without delay. No other maintenance or operator intervention was required over the extended test period. Estimated Minimum Detectable Calibration A signal-to-noise ratio was calculated for S and COS using the peak height from a representative analysis of the low-level calibration standards, acquired using a 1-mL loop and a 50:1 split ratio. The minimum detectable concentration for each analyte was then estimated based on a signal-to-noise ratio of 4:1 and a split ratio of 10:1. The estimated minimum detectable concentrations for S and COS were 0.03 ppmv and 0.02 ppmv, respectively, and are consistent with the levels reported by Agilent. These results are summarized in Table 3. Analyte Separation from C3 Matrix Three compressed gas cylinders containing unknown concentrations of S and COS in different proportions of propane and propylene were used to demonstrate baseline separation of the sulfur compounds from the C3 matrix mix. Analyte concentrations were calculated based on the original calibration curves. Composition of the individual tanks is shown in Table 4, and chromatograms illustrating baseline separation of the target compounds from the C3 matrix are shown in Figures 6, 7, and 8. 7

. Table 4. Composition of Tanks Parameter Measured S Result COS Result Calculated Concentration in Unknown Samples Tank #3002 100% Propane 0.79 ppmv 9.88 ppmv Tank #3007 1% Propylene 99% Propane ND 9.49 ppmv Trace C2 Tank #3012 77% Propylene 23% Propane 0.17 ppmv 8.23 ppmv COS 9.88 ppmv S 0.79 ppmv ~100% Propane Figure 6. Tank #3002 contained S at 0.79 ppmv and COS at 9.88 ppmv, and the balance was 100% propane. The COS was baseline separated from the propane by one full minute. 8

COS 9.49 ppmv S ND Trace C2 1% Propylene ~99% Propane Figure 7. Tank #3007 contained 9.49 ppmv COS and no detectable S. The 1% propylene and 99% propane were fully separated from the COS, as was a trace amount of C2. COS 8.23 ppmv S 0.17 ppmv 77% Propylene 23% Propane Figure 8. Tank #3012 contained 8.23 ppmv COS and a trace amount of S at 0.17 ppmv. The matrix consisted of 77% propylene, which eluted before the 23% propane and was still completely baseline separated from the sulfur compounds. 9

Summary and Conclusions The OI Analytical SPRO-Select system is a simple, turn-key solution for fast analysis of sulfur in propanepropylene feedstock using a Pulsed Flame Photometric Detector (PFPD) and a new capillary PLOT column. The new column provides complete baseline separation of the volatile sulfur impurities from the C3 propane-propylene matrix, good symmetric peak shape, and exhibits excellent response for S. Certified permeation devices were used to generate analytical standards at point-of-use for calibration and stability testing. The PFPD remained stable over the two-month test period, in spite of temporary interruptions in operation. Minimum detectable concentrations of the sulfur compounds are estimated to be approximately 20 to 30 ppbv. References 1. Amirav, A., Pulsed Flame Detector Method and Apparatus. U.S. Patent 5,153,673, Israel Patent 95617, Europe Patent 0475250, and Japan Patent 2759854. 2. CP 8575 Agilent J&W Select Low Sulfur PLOT Column. Acknowledgement Thanks to DCG Partnership in Pearland, Texas for providing the tanks with variable ratios of propane and propylene, and unknown concentrations of S and COS. Sulfinert is a registered trademark of Restek. Publication 36720111