Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Similar documents
columns IonPac AS17-C Anion Exchange Column

Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment and Ion Chromatography

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Inline sample preparation for the determination of anions in sodium hydroxide

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Dionex IonPac AS28-Fast-4µm column

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Thermo Scientific. Anion-Exchange Column

columns IonPac AS18 Anion-Exchange Columns

columns IonPac AS22 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Determination of trace anions in concentrated hydrofluoric acid

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Determination of trace anions in organic solvents using matrix elimination and preconcentration

Determination of urea in ultrapure water by IC-MS/MS

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

Determination of Trace Cations in Power Plant Waters Containing Morpholine

The Determination of Trace Anions in Concentrated Phosphoric Acid

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

The determination of trace anions in concentrated phosphoric acid

Determination of silicate in high-purity water using ion chromatography and online sample preparation

anion-exchange column

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of trace anions in high-nitrate matrices by ion chromatography

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Direct Determination of Cyanate in a Urea Solution and a Urea-Containing Protein Buffer

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

IonPac Trace Cation Concentrator

Determination of Trace Anions in Organic Solvents

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Determination of Trace Anions in Concentrated Glycolic Acid

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Determination of trace anions in concentrated glycolic acid

Determination of Trace Anions in Concentrated Hydrofluoric Acid

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

columns IonPac SCS 1 Silica Cation Separator

Overcoming Challenging Matrices in Ion Chromatography

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm)

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Sample Preparation for Ion Chromatography

for IonPac AG11-HC IonPac AS11-HC

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Determination of Methylamine in Drug Products

Utility of the Charge Detector in Ion Chromatography Applications

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts By Ion Chromatography

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Determination of Inorganic Anions in Wastewater by Ion Chromatography

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

columns I o n P a c AS11 and AS11-H C Anion-E x c h a n g e C o l u m n s

Determination of Adsorbable Organic Halogen in Wastewater

Application Note 176. Brian DeBorba and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Organic Acids in Beer Samples Using a High-Pressure Ion Chromatography System

Determination of Common Inorganic Anions in Environmental Waters Using a Hydroxide Selective Column

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

suppressors Eluent Suppressors for Ion Chromatography

Thermo Scientific Dionex IonPac AS11 and AS11-HC Anion-Exchange Column

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

IonPac SCS 1 IonPac SCG 1

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

DETERMINATION OF PERCHLORATE IN DRINKING WATER USING TWO-DIMENSIONAL ION CHROMATOGRAPHY WITH SUPPRESSED CONDUCTIVITY DETECTION. Version 1.

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent

Thermo Scientific Dionex eluent suppressors for ion chromatography

Analysis of various clays with Metrohm Combustion IC

Expanded capabilities in ammonium and amine detection

ProPac WCX-10 Columns

suppressors SRS -ULTRA Self-Regenerating Suppressor

Determination of adsorbable organic halogen in wastewater using a combustion ion chromatography system

California Environmental Protection Agency Air Resources Board

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY

The power to increase productivity. Thermo Scientific Dionex Reagent-Free Ion Chromatography (RFIC) System Capabilities

ION CHROMATOGRAPHY SYSTEM S 150

Advances in Sample Preparation for Accelerated Solvent Extraction

Transcription:

Application Note 9 Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography Introduction The computer, semiconductor, and food industries need analytical methods to determine trace anions in concentrated bases. Anions at very low concentrations can cause corrosion, deposition defects, and electronic shorts in computer devices. In the food industry, sodium hydroxide and other caustic solutions used to remove the outer layers of fruit and vegetables can affect taste. Other anions are toxic (bromate, oxalate, nitrate, and nitrite) or carcinogenic (nitrite and bromate) when ingested at high levels. Ion chromatography (IC) with suppressed conductivity detection is the preferred method for determining ionic species, especially at µg/l and ng/l concentrations. Direct injections of concentrated base samples result in high concentrations of the base anion that overload the column, resulting in poor chromatography and quantification. Therefore, concentrated bases must be diluted prior to anion analysis. However, this dilution sacrifices anion determinations at µg/l and ng/l concentrations. The introduction of AutoNeutralization by Dionex in 99, embodied in the SP0 AutoNeutralization module, eliminated the need to dilute the concentrated base and thus allowed µg/l and ng/l anion determinations. This technique used multiple cycles through the neutralizer to completely neutralize the strong acid or base in the sample. The second embodiment of AutoNeutralization replaced the SP0 with two, six-port valves and an external pump installed on the ICS-00 chromatography system. This method, reported in the previous version of this application note, effectively neutralized strong bases with a series of neutralizing cycles. This same technique was used to neutralize acids. In the latest version of AutoNeutralization, the sample was pumped into the ASRN II neutralizer by an external pump and trapped for a set period. This technique was coined Park and AutoNeutralize. Here we discuss neutralization with Park and AutoNeutralize using the ICS-000 ion chromatography system with dual pumps, detectors, and valves. The system configurations, plumbing, valve operations, and program are described in this application note. The application includes the Carbonate Removal Device (CRD) to improve the determination of sulfate and other anions eluting near carbonate. The CRD removes carbonate from the sample prior to detection. The CRD and carbonate removal process are discussed thoroughly in Technical Note. This application note replaces the previous version of AN 9 and updates the method to use an IonPac AS9 column set and an IonPac UTAC-XLP concentrator column. The IonPac AS9 is a high capacity, hydroxide-selective column (0 µeq/column) with selectivity for oxyhalides and inorganic anions. The AS9 contains a novel, hyper-branched anion-exchange condensation polymer attached to the 7. µm super ma- Application Note 9

croporous resin. The IonPac UTAC-XLP ( mm) concentrator column was selected for this application to ensure low sulfate blanks. The IonPac UTAC-XLP contains a latex agglomerated super macroporous resin and has a low void volume (< µl). It was designed for high purity water analysis. It is a latex agglomerated super macroporous resin. The IonPac UTAC-XLP column is similar to the older concentrator column, but to ensure minimal sulfate backgrounds, the resin is repeatedly carboxylated instead of sulfonated. Additionally, the new mm configuration has the added benefit of very low backpressure. This method successfully determined trace anions from low µg/l to low mg/l concentrations in 0% and 0% sodium hydroxide, 0% potassium hydroxide, and concentrated ammonium hydroxide (9% as NH, 9.7% as NH OH). Experimental Equipment Dionex ICS-000 Reagent-Free Ion Chromatography system with: DC Detector/Chromatography module with Single Temperature zone, AM Automation Manager (P/N 07), 0-port valve (P/N 09), one injection valve, a standard bore temperature stabilizer, and two CD Conductivity Detector and Integrated Cells (P/N 07) DP Dual Pump module with Degas module, Gradient Mixer IonPac GM- (P/N 09), and IonPac ATC-HC trap column (9 7 mm, P/N 090) EG Eluent Generator module with EluGen II potassium hydroxide (Dionex P/N 08900), and Continuously Regenerating Anion Trap Column (CR-ATC, P/N 0077) Dionex AS Autosampler with 0 ml sample tray and vial kit (P/N 008), and 00 or 000 µl sample syringe Chromeleon Chromatography Management system, Chromeleon.7 Tubing: Black PEEK (0.-mm or 0.0-in i.d., P/N 00 for ft) for liquid line connections for both systems and backpressure loops for the ASRS ULTRA II suppressor Yellow PEEK (0.7-mm or 0.00-in i.d., P/N 00 for ft) for system backpressure loops Green PEEK (0.7-mm or 0.0-in i.d., P/N 0077 per inch) for eluent waste lines Low pressure Teflon tubing (E.I. du Pont de Nemours,.-mm or 0.0-in i.d., P/N 07) for the CRD, external water regenerant lines for the ASRN II neutralizer, and ASRS ULTRA II suppressor regenerant and degas waste lines 00 µl sample loop (P/N 9) Suppressor External Regenerant Installation Kit (P/N 0808) for external water with one additional -L bottle -ml polystyrene sterile flasks for trace anion standards and samples (Corning P/N: 08 or VWR P/N: 98-99) Reagents and Standards Note: Use only ACS reagent grade chemicals for all reagents and standards. Deionized water, type reagent-grade, 8. MΩ-cm resistivity or better Oxalic acid, dihydrate, crystalline (Fisher, P/N A9-00) Sodium bromate (Sigma-Aldrich, P/N 87) Sodium bromide (Sigma-Aldrich, P/N,00-) Sodium chlorate (Sigma-Aldrich ReagentPlus, P/N,-7) Sodium chloride, crystalline (JT Baker Ultrapure Bioreagent, P/N JT-) Sodium chlorite (Fisher, P/N LC0) Sodium fluoride (Fisher, P/N S99-00) Sodium hydroxide solution, 0% (w/w) (Fisher Chemicals, P/N SS-00) Sodium nitrate, crystalline (Fisher, P/N S-00) Sodium nitrite (JT Baker, P/N JT780-) Sodium phosphate, monobasic (EMD Biosciences Inc., P/N 8008-0) Sodium sulfate, granular (EM Science, P/N EM-SX070-) Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

s Sodium hydroxide, 0% and 0% (w/w) prepared from 0% (w/w) sodium hydroxide solution (Fisher Chemicals, P/N SS-00) Potassium hydroxide, 0% (w/w) from % (w/w) potassium hydroxide solution (JT Baker, P/N -0) Ammonium hydroxide, concentrated, 9% (w/w) as NH (9.7% as NH OH), (Sigma, P/N 8) Conditions Preparation: System Trap Column: IonPac ATC-HC, 9 7 mm Flow Rate: 0. ml/min Carrier: Degassed deionized water purified inline with a trap column Inj. Volume: 00 µl (full-loop, calibrated value of 0 µl, which is the value used for all calculations in this application note) Neutralization: ASRN II Anion Self-Regenerating Neutralizer (P/N 0), external water mode, 9 ma Run Time: 9. min for Park and AutoNeutralize and Concentrate Typical Backpressure: ~800 psi Analytical: System Columns: IonPac AG9, AS9 ( 0 mm, P/N 0887), ( 0 mm, P/N 088) Flow Rate:.0 ml/min Eluent (EGC-KOH): 0 to mm Potassium hydroxide from 0 to min, mm from to min Column Temp.: 0 ºC Concentrator: IonPac UTAC-XLP, mm (P/N 09) CRD: -mm CRD (P/N 098) installed between suppressor and the detector Detection: Suppressed Conductivity, ASRS ULTRA II Anion Self-Regenerating Suppressor, recycle mode, ma Typical Background: < µs Typical Backpressure: ~00 psi Typical Noise: < ns Run Time: min Preparation of Solutions and Reagents Preparation It is essential to use high quality, Type water, 8. MΩ-cm. It should contain as little dissolved carbon dioxide as possible. Standard Preparation To prepare individual stock solutions of 000 mg/l fluoride, chloride, chlorite, bromate, nitrite, bromide, nitrate, sulfate, oxalate, and phosphate, weigh the amount of reagent grade compound (Table ) in a ml HDPE sample bottle and dilute with deionized water to 00.00 g total weight. Shake each stock solution to fully dissolve the reagent. Table. Amount of Compound Used to Prepare 00 ml of 000 mg/l Individual Stock Solutions Anion Compound Mass (g) Fluoride Sodium fluoride (NaF) 0. Chlorite Sodium chlorite (NaClO ) 0. Bromate Sodium bromate (NaBrO ) 0.80 Chloride Sodium chloride (NaCl) 0. Nitrite Sodium nitrite (NaNO ) 0.0 Chlorate Sodium chlorate (NaClO ) 0.8 Bromide Sodium bromide (NaBr) 0.9 Nitrate Sodium nitrate (NaNO ) 0.7 Sulfate Sodium sulfate (Na SO ) 0.8 Oxalate Oxalic acid dihydrate (HOOCCOOH H O) 0. Phosphate Sodium phosphate, monobasic (NaH PO ) 0. Application Note 9

Table. Amount (µl) of 0 mg/l Intermediate Standards Used to Prepare 00 ml Working Standards of Fluoride, Chlorite, Bromide, Chlorate, Bromate, Oxalate, Phosphate, Nitrite, and Nitrate a 0 mg/l Intermediate standards Working Standards 7 8 9 Fluoride, chlorite, bromide, and chlorate 0 00 00 00 800 00 000 Bromate, oxalate, and phosphate 00 00 00 800 00 00 000 Nitrite 00 000 00 000 000 Nitrate 0 00 00 00 800 00 000 000 000 a Combined working standards of 00, 00, 000, 00, 000, 00, 000, and 00 µg/l chloride and sulfate were prepared from 000 mg/l stock standards. Intermediate and Working Standards To determine anions in sodium and potassium hydroxide reagent samples, prepare separate 0.0 mg/l intermediate standards from fluoride, chlorite, bromate, nitrite, chlorate, bromide, nitrate, oxalate, and phosphate stock solutions in water. Pipette 000 µl each of the 000 mg/l individual stock solutions into separate 0 ml polypropylene bottles. Dilute with deionized water to 00.00 g total weight. Prepare the Corning -ml polystyrene sterile flasks for the µg/l standards two days or more prior to the standard preparation. Rinse each flask five times with deionized water, fill it to the top with deionized water, and let it soak overnight. Repeat this daily until flasks are needed for the µg/l standards. (See Application Update for additional precautions needed for determination of µg/l anion concentrations. 7 ) To prepare working standards (, 0, 0, and 0 µg/l fluoride, chlorite, bromide, chlorate, and nitrate, and 0, 0, 0, and 80 µg/l bromate, oxalate, and phosphate), pipette the amounts in Table into four -ml polystyrene sterile flasks. Dilute these working standards with deionized water to 00.00 g total weight and mix thoroughly. Prepare working standards 7 (80, 0, and 00 µg/l fluoride, chlorite, bromide, chlorate and nitrate, 0, 00, and 0 µg/l nitrite, and 0, 0, and 00 µg/l bromate, oxalate, and phosphate) and working standards 8 9 (00 and 00 µg/l nitrite and nitrate) in a similar manner. Prepare 00, 00, 000, 00, 000, 00, 000, and 00 µg/l working standards of chloride and sulfate, similarly with 0, 0, 00, 0, 00, 0, 00, and 0 µl of 000 mg/l stock standards of chloride and sulfate. These standards can be prepared in 0 ml polypropylene bottles. Prepare low µg/l standards daily, higher µg/l standards weekly, and mg/l standards monthly. Preparation To prepare samples of 0% (w/w) and 0% (w/w) (. M and 7. M) of sodium hydroxide solution from 0% (w/w) (. M) sodium hydroxide and degassed deionized water, first clean a -ml transfer pipette by rinsing it three times with deionized water. Shake out the excess liquid. Also clean a -ml polystyrene flask as described in the standard preparation section. To prepare the 0% (w/w) and 0% (w/w) sodium hydroxide sample, transfer 0.0 g and 0.0 g, respectively, of 0% sodium hydroxide with the pre-rinsed transfer pipette into a pre-cleaned -ml polystyrene flask containing 0.0 g of degassed deionized water. Dilute with degassed deionized water to 0.0 g total weight. Cap the bottle and gently shake bottle until the solution is thoroughly mixed. Prepare the 0% (w/w) potassium hydroxide solution in a similar manner with.7 g of % (w/w) potassium hydroxide in deionized water to 0.0 g total weight. The concentrated (9% as NH, 9.7% ammonium hydroxide) ammonium hydroxide solution was analyzed directly without dilution. The Autoneutralization pretreatment method is recommended for concentrated bases up to 0%. Concentrated bases above 0% have higher viscosities that can cause longer loading times and result in lower reproducibilities. Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Spike Recovery s in Water and Reagents To prepare spike recovery samples in sodium hydroxide and potassium hydroxide solutions, the 000 mg/l stock standards should be spiked directly into the water diluent. For example, to prepare spike recovery samples in 0% sodium hydroxide, pipette 0 µl of 000 mg/l stock standards of bromate, chlorate, bromide, nitrate, and oxalate into a -ml precleaned polystyrene flask containing 0.00 g of degassed deionized water. Also, pipette 0 µl of 000 mg/l stock standard of phosphate, and 0 µl of 000 mg/l stock standards of chloride and sulfate. Using a pre-rinsed -ml transfer pipette, fill the pipette with 0% sodium hydroxide and dispense it into the flask. Fill the pipette a second time and dispense until 0.00 g of 0% sodium hydroxide are added to the flask. Add degassed deionized water to a total weight of 00.00 g. Cap the flask and gently shake the flask until it is thoroughly mixed. The final concentrations are 0 µg/l bromate, chlorate, bromide, nitrate, and oxalate, 00 µg/l phosphate, and 00 µg/l chloride and sulfate. Prepare the spike recovery samples for 0% sodium hydroxide and 0% potassium hydroxide and in a similar manner. To prepare 00 µg/l chloride, nitrite, and sulfate, and 0 µg/l bromide, and 90 µg/l of nitrate and oxalate spiked into concentrated ammonium hydroxide, pipette 0, 8, and 8 µl respectively, of 000 mg/l stock standards into a pre-cleaned -ml polystyrene flask. Add concentrated ammonium hydroxide to a total weight of 00.00 g. Anion Self-Regenerating Neutralizer ASRN II Operation The AutoNeutralization pretreatment is based on the neutralization of the strong base to low conductivity water through the high capacity, electrolytic Anion Self-Regenerating Neutralizer (ASRN II) device. The ASRN II device neutralizes the strong base by exchanging the cation of the base with hydronium that is electrolytically generated. The neutralized sample is predominantly water with trace levels of anions. The neutralization process is thoroughly discussed in the ASRN II product manual. 8 The Park and AutoNeutralization pretreatment includes two steps. First, the sample slug is routed through the eluent channel of the ASRN II and the flow of the carrier stream is diverted to place the sample slug within the eluent channel. This is called Park. Second, the sample slug is neutralized completely, then routed to a concentrator column for pre-concentration and analysis. System Setup The setup for the individual modules, components, and system are thoroughly described in the Operator s and Installation manuals 9,0 for the ICS-000 Chromatography system, and the Chromeleon Help menus. The sample loading and neutralization by the Anion Self-Regenerating Neutralizer ASRN II occurs on System. The remaining trace anions in the neutralized sample are concentrated, injected, separated, and detected by suppressed conductivity on System. System, Preparation preparation for the Park and AutoNeutralization application uses the AS Autosampler, Pump with IonPac ATC-HC (9 mm) trap column, Automation Manager (AM) with a 0-port, high pressure valve (AM-HP), Conductivity Detector CD, and the ASRN II neutralizer. To set up System, connect black PEEK (0.-mm or 0.00-in I.D.) tubing from Pump to the IonPac ATC-HC trap column and from the trap column to the inside of the DC. Add a 0-cm ( in) length of yellow (0.07-mm or 0.00-in i.d.) PEEK tubing to make an approximately 800 psi backpressure loop. After priming Pump, verify that the System pressure is between 800 and 900 psi. Remove or add yellow PEEK backpressure tubing to adjust the system pressure to 800 900 psi. Flush the ATC-HC trap column with deionized water for 0 min. Open the upper chamber door of the ICS-000 DC module and locate the Automation Manager and the AM-HP 0-port high pressure valve, (top left corner of the Automation Manager). Plumb System with black PEEK tubing according to Figure. Add black PEEK tubing from the yellow backpressure tubing to Port on AM-HP. Install the 00-µL loop (~9 cm of orange PEEK tubing) in Ports and. Connect a. to 7 cm (. to.8 in) piece of black PEEK tubing from Port to Port 7 to connect the sample load side of the valve (Ports ) to the neutralization side of the valve (Ports 0). Connect the AS Autosampler pink injection line tubing into Port. Connect one end of green (0.7-mm or 0.0-in i.d.) PEEK tubing into Port and direct the other end into the waste container. Connect to 7 cm (.8 to. in) of black PEEK tubing from Port 0 to the inlet of CD Conductivity Detector. Application Note 9

Set-up for the ASRN II Neutralizer Hydrate the ASRN II neutralizer for 0 min, according to the QuickStart procedure in the neutralizer product manual. Install the neutralizer in the suppressor slot for System. Connect a.0 to.0 cm (8. to 9.0 in) length of black PEEK tubing from Port 8 of the AM- HP valve to Eluent In on ASRN II neutralizer and a to cm (7. to 8. in) length of tubing from Eluent Out to Port 9. Install the external water regenerant on the neutralizer using the Suppressor External Regenerant Installation Kit. Connect Teflon tubing (.-mm or 0.0-in i.d.) to the Regen In port of the ASRN II neutralizer and from Regen Out to waste. Adjust the flow rate to - ml/min by adjusting the pressure on the external water regenerant system. To determine the flow rate, collect the water from the neutralizer into a -ml graduated cylinder for min. The total volume should be to ml. It is critical that the external water regenerant is flowing through the ASRN II neutralizer before the pump and the neutralizer are turned on. When the pump is turned off, the regenerant should also be turned off. It is also critical that the backpressure to the ASRN II neutralizer is less than 00 psi. Measure the backpressure on the neutralizer channel by disconnecting the line out of the eluent out port of the ASRN II (B). Replace the line and measure the system pressure with the line in (A). The difference A minus B should be less than 00 psi. To reduce background sulfate contamination, install a temporary waste line from the CD detector outlet and flush System with deionized water overnight or until the background conductivity is < µs. Open the DC chamber and verify by sight that the external water is flowing through neutralizer before turning on the pump. To turn on the neutralizer, select the Detector Compartment tab (System ) in the Chromeleon Panel (Default Panel Tabset), select SRS as the suppressor type, and enter 9 ma for the current. After the background conductivity is < µs, remove the temporary waste line on the CD outlet and install 8 to 0 cm (7. to 7.9 in) of black PEEK tubing from the CD outlet to the Port (S) in the Inj. Valve located in lower chamber of the DC. System is now connected to System. AS Autosampler Setup On the AS Autosampler module, enter the sample loop volume of 00 µl or the calibrated volume and press the Enter key (Plumbing Parameters under Module Setup Menu). Also verify that the Syringe Volume and the Prep Syringe Volumes are correct. For this application, use a 00 µl or 000 µl sample syringe. For samples with extreme concentration differences, adjust the cut volume to µl (Module Setup Menu under System Parameters). The Autosampler should also be in Normal mode (Use the _ key to select Normal for the Mode. Press Enter.) Set the Flush and Prime volumes on the Chromeleon Panel. Enter 00 µl (or times the sample loop) for the Flush volume and enter a large volume for the Prime volume (000 µl). The injection volume, entered on the Sequence, is the same as sample loop volume. The volume removed from the vial is four times the injection volume. More information can be found in the AS Autosampler operator s manual. System, Analytical System analyzes the neutralized sample and includes Pump, eluent generation (EG with degas and CR-ATC), injection (Inj. Valve ), and concentration (IonPac Trace Anion Concentrator column, UTAC-XLP, mm), separation (IonPac AS9, 0 mm column and guard, 0 mm), and suppressed conductivity detection (ASRS ULTRA II suppressor with CRD). The neutralized sample is moved from the ASRN neutralizer onto the conductivity detector in System (CD ) through Inj. Valve in the DC, and onto the concentrator column. Then the concentrated neutralized sample is moved from the concentrator column into the IonPac AS9 column set. To set up System, install the gradient mixer (GM-), EGC II KOH cartridge, IonPac CR-ATC, ASRS ULTRA II suppressor, CRD, columns, and backpressure loops for the suppressor and the eluent generator. Connect black PEEK (0.-mm or 0.00-in I.D.) tubing from Pump to the gradient mixer (GM-) and from the gradient mixer into the EG module. Install and condition the EluGen II KOH cartridge, hydrate and flush the CR-ATC, and install a backpressure loop Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

according to the start up instructions in the ICS-000 operator s manual. Install the CR-ATC after the cartridge and before the degas module and the Port P of Inj. Valve. Install a backpressure loop after the CR-ATC and before Port P of Inj. Valve (Figure ). Install the IonPac UTAC-XLP concentrator column in Ports L and L according to the product manual instructions. Position the concentrator column closer to Port than Port and install it with a minimal amount of black PEEK tubing from the column outlet to Port L. Connect the inlet of the concentrator column to Port L. The sample is loaded and injected from Port L. Install one end of the black PEEK tubing from the temperature stabilizer into Port C. Install the other end into the guard and analytical columns following the instructions in their product manuals. Connect green PEEK tubing into Port W and direct the other end to waste. Green line Inj. Valve UTAC-XLP Analytical Temp Stabilizer EGC, CR-ATC Pump A S R N External Water Figure. Valve configuration. Green Line 0 0 cm Black PEEK 7 8 0 cm 9 Black PEEK CD AM-HP Valve cm Black PEEK Trap ATC-HC Pump AS Autosampler 00 µl Loop 8 tighten the fittings on the CRD. Set the Pump flow rate to.0 ml/min and prime the pump. Adjust the system pressure to about 00 psi by adding or removing yellow PEEK tubing to the system backpressure loop. Configuration Using the Chromeleon Server Configuration program, create one timebase and add the ICS-000 modules and AS Autosampler. 0 Under DC Properties, assign the ASRS ULTRA II suppressor to System and ASRN II neutralizer to System. Rename the ASRN II neutralizer as Neutralizer, for convenience and clarity. This application uses two high-pressure valves (AM-HP and Inj. Valve ). Verify that the boxes in front of AM-HP and Inj. Valve are checked, and that no other valves are checked. Assign control of the AM-HP valve to the AS Autosampler, and control of the Inj. Valve to the DC. Verify that the EGC II cartridge is listed on System and that it is linked to Pump. Under the AS Autosampler properties, verify that Overlap box is checked. Overlap reduces the total sample analysis time. Creating the Program Use Chromeleon Wizard, the eluent conditions (Table ), and valve commands (Table ) to create a program for Park and AutoNeutralization. The eluent conditions are programmed to wash the column before returning to the initial conditions for the next run. The eluent concentration remains at mm potassium hydroxide from the end of the last program to the beginning of the 0 mm potassium hydroxide equilibrium Before installing the backpressure loop on the cell, measure the backpressure directly from the pump to the cell. Install the backpressure loop on the cell out position of the conductivity cell and measure the backpressure. The difference of these two measurements is the backpressure due to the cell and backpressure loop. It must be between 0 and 0 psi to prevent suppressor damage. Follow the QuickStart instructions for the ASRS ULTRA II suppressor and the -mm CRD to hydrate and install them. 7,8 Install the suppressor in recycle mode in the System suppressor slot between the columns and the CD. Install the CRD on the suppressor. Tighten the fitting to finger-tight. Do not over- Table. Chromeleon Entries for Eluent Conditions Time (min) Potassium Hydroxide Description Concentration (mm) 0 Continue column wash.9 0 Start. min equilibration.0 0 9.0 0 Start gradient.0 End gradient Equilibration (.9 9. min) and initial conditions (9. 9 min). Start column wash.0 Column wash Application Note 9 7

Table. Chromeleon Wizard Valve Commands to Table Control. Chromeleon Inj. Valve Entries and the for AM-HP Eluent Valves Conditions Time (min) Valve Position Action 0. AM-HP A 8. Inj. Valve Load 8. AM-HP B 9. Inj. Valve Inject The sample is isolated onto ASRN II and neutralized. Inj. Valve (System ) is turned to Load. The neutralized sample is released from the neutralizer. Start concentrating the sample onto the UTAC-XLP concentrator column. Inject neutralized, concentrated sample into analytical system. Analytical Column Inj. Valve UTAC-XLP Eluent In Pump A S R N External Water CD 8 0 AM-HP Valve 7 9 Trap ATC-HC Figure. AM-HP valve (A) in sample load position. In Loop Water Carrier Pump 9 (.9 min). During this period, the sample is loaded and partially neutralized. To enter the suppressor type for the ASRN II neutralizer, select SRS-MPIC. Set the current at the maximum level (9 ma). The ASRS ULTRA II, -mm is listed under suppressor types. Select it and set the current for ma. Enter the valve commands for Inj. Valve and the AM-HP valves in the Relay & State Devices Option section. In the program review mode, verify the valve timing and the gradient timing. Remove any Duration=number on the valve commands. Save the program. autosampler loads the sample onto the sample loop, the water carrier by-passes the ASRN II neutralizer and sample loop. At time zero, the sample is injected as AM-HP valve switches to state B (Figure ) and the water carrier flushes the sample from the sample loop. The valve is switched to state A (Figure ) at 0. min as the sample is carried and trapped in the neutralizer or parked in the neutralizer where the base is neutralized and the anions are converted to their acid forms. At 8. min, Inj. Valve (Figure ) is switched to Load. At 8. min, the AM-HP valve is switched to state B, which releases the neutralized sample to the concentrator on System Results and Discussion This application uses both systems on the ICS-000 Ion Chromatography system. Before discussing the experiments, it is helpful to review the flow path of the AM-HP valve. The valve has two states: A and B. In sample load (Figure ), the valve is in state A. The Analytical Column Inj. Valve UTAC-XLP A S R N 8 0 AM-HP Valve 7 9 In Loop Eluent In Pump External Water CD Trap ATC-HC Water Carrier Pump 0 Figure. AM-HP valve (B) in sample inject position. 8 Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Analytical Column Inj. Valve UTAC-XLP Eluent In Pump External Water A S R N 8 0 CD AM-HP Valve 7 9 Trap ATC-HC In Loop Water Carrier Pump. After concentrating the anions, the concentrated and neutralized sample is injected (Inj. Valve to Inject at 9. min) to the analytical system (Figure ). The present study determined the retention times of the eleven anions using 0 µg/l standards of fluoride, chlorite, bromate, nitrite, chlorate, bromide, nitrate, oxalate, and phosphate and 00 µg/l standards of chloride and sulfate. Trace anions in water (Figure 7), 9.9% (w/w) and 9.9% (w/w) sodium hydroxide (Figures 8 9) were determined with this method. The chloride, sulfate, and oxalate concentrations show proportional increases with Figure. is Parked and Neutralized. Analytical Column Inj. Valve UTAC- XLP Eluent In Pump A S R N External Water Figure. concentrate. 8 0 CD AM-HP Valve 7 9 Trap ATC-HC In Loop Water Carrier Pump Prep. Carrier: Degassed deionized water Temperature: 0 ºC Carrier Flow Rate: 0. ml/min Inj. Volume: 0 µl loop Neutralization: Park and AutoNeutralize for 8 min in ASRN II, external water, 9 ma Analytical Column: IonPac AG9, AS9, -mm Eluent (EluGen KOH): 0 to mm potassium hydroxide from 0 to min, mm potassium hydroxide from to min Column Temp.: 0 ºC AS Temperature: 0 ºC Flow Rate:.0 ml/min Concentration: IonPac UTAC-XLP Detection: Suppressed conductivity, recycle, ma 0.7 Peaks:. Fluoride 0.8 µg/l. Acetate. Formate. Chlorite.8. Bromate 7.. Chloride 89. 7. Nitrite 9. 8. Unknown 9. Chlorate 8. 0. Bromide 0.. Nitrate 8.7. Unknown. Unknown. Carbonate. Sulfate 8.. Oxalate. 7. Unknown 8. Phosphate. 9. Unknown 0. Unknown Analytical Column Inj. Valve UTAC- XLP Eluent In Pump External Water Figure. Inject and analyze. A S R N 8 0 CD AM-HP Valve 7 9 Trap ATC-HC In Loop Water Carrier Pump µs 7 8 9 0 0. 9..0 0.0.0 0.0.0 0.0.0 Minutes 7 8 9 0 Figure 7. Park and AutoNeutralization of anion standard in water (0 µg/l fluoride, chlorite, bromate, chlorate, bromide, and nitrate, µg/l oxalate, 0 µg/l phosphate, and 00 µg/l chloride and sulfate). Application Note 9 9

Prep. Carrier: Degassed deionized water Temperature: 0 ºC Carrier Flow Rate: 0. ml/min Inj. Volume: 0 µl loop Neutralization: Park and AutoNeutralize for 8 min in ASRN II, external water, 9 ma Analytical Column: IonPac AG9, AS9, -mm Eluent (EluGen KOH): 0 to mm potassium hydroxide from 0 to min, mm potassium hydroxide from to min Column Temp.: 0 ºC AS Temperature: 0 ºC Flow Rate:.0 ml/min Concentration: IonPac UTAC-XLP Detection: Suppressed conductivity, recycle, ma.00 Peaks:. Fluoride.8 µg/l. Acetate. Formate. Chlorite.. Bromate 9.7. Chloride 7. 7. Nitrite.8 8. Chlorate.9 9. Bromide. 0. Nitrate 9.9. Unknown. Carbonate. Sulfate 7.. Oxalate 8.8. Unknown. Unknown 7. Phosphate. 8. Unknown 9. Unknown Prep. Carrier: Degassed deionized water Temperature: 0 ºC Carrier Flow Rate: 0. ml/min Inj. Volume: 0 µl loop Neutralization: Park and AutoNeutralize for 8 min in ASRN II, external water, 9 ma Analytical Column: IonPac AG9, AS9, -mm Eluent (EluGen KOH): 0 to mm potassium hydroxide from 0 to min, mm potassium hydroxide from to min Column Temp.: 0 ºC AS Temperature: 0 ºC Flow Rate:.0 ml/min Concentration: IonPac UTAC-XLP Detection: Suppressed conductivity, recycle, ma.00 Peaks:. Fluoride. µg/l. Acetate. Formate. Chlorite.0. Bromate 9.. Chloride 89. 7. Nitrite. 8. Chlorate.9 9. Bromide.9 0. Nitrate 7.. Unknown. Carbonate. Sulfate. Oxalate 8.8. Unknown. Unknown 7. Unknown 8. Phosphate 09. 9. Unknown 0. Unknown µs 0 7 8 9 0. 9..0 0.0.0 0.0.0 0.0.0 Minutes 7 8 9 Figure 8. 9.9% (. M) sodium hydroxide spiked with 0 µg/l of fluoride, chlorite, bromate, chlorate, bromide, nitrate, and oxalate, 9 µg/l of phosphate, and 80 µg/l of chloride and sulfate. µs 0 7 8 9 8 7 0 0. 9..0 0.0.0 0.0.0 0.0.0 Minutes Figure 9. 9.9% (7. M) sodium hydroxide spiked with 0 µg/l of fluoride, chlorite, bromate, nitrite, chlorate, bromide, and nitrate, 8 µg/l of oxalate and phosphate, and 0 µg/l of chloride and sulfate. 9 concentration. This confirms that the timing is correct and that the ASRN II neutralizer has sufficient neutralizing capacity. In this application the recorded retention times include 9. min of sample preparation. To determine reproducibility based on the true retention, we subtracted the 9. min sample preparation time and manually recalculated the RSDs. These retention time reproducibilities of the 0 anions in water, 9.9% and 9.9% sodium hydroxide (Table ) were good, <% RSDs. The peak area reproducibilities of the same anions (Table ) in the same solutions ranged from 0. to.% RSDs. 0 Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Table. Retention Time Reproducibilities of Ten Anions in Water, 9.9% (. M), and 9.9% (7. M) Sodium Hydroxide Spiked into water 9.9% Sodium Hydroxide 9.9% Sodium Hydroxide with added anions with added anions Amount Retention RSD Amount Retention RSD Amount Retention RSD Present (µg/l) Time (min) a (%) Present (µg/l) Time (min) a (%) Present (µg/l) Time (min) a (%) Fluoride..7 0. 0.9.8 0. 7.. 0.8 Chlorite 0. 7.0 0. 0. 7.0 0..8 7.0 0. Bromate. 7. 0. 9.7 7. 0.. 7. 0. Chloride 9.9 8. 0. 89. 8. 0. 87 8. 0. Chlorate 0..07 0...08 0. 7..07 0. Bromide 8.8.0 0. 7..0 0. 9..0 0. Nitrate..9 0...70 0...9 0. Sulfate.9 9.8 <0.. 9.8 <0. 9.8 0. Oxalate 79.0.9 <0. 9..9 <0..0.9 <0. Phosphate 99.. <0. 0.. <0... <0. n=7 a The sample preparation times (9. min) were subtracted from the retention times and the RSDs were re-calculated for the actual retention times. Spiked into water 9.9% Sodium Hydroxide 9.9% Sodium Hydroxide with added anions with added anions Amount Peak Area RSD Amount Peak Area RSD Amount Peak Area RSD Present (µg/l) (µs-min) (%) Present (µg/l) (µs-min) (%) Present (µg/l) (µs-min) (%) Fluoride. 0.0 0.8 0.9 0.00. 7. 0.0 0.8 Chlorite 0. 0.00. 0. 0.00..8 0.00. Bromate. 0.00. 9.7 0.00.9. 0.007.7 Chloride 9.9 0.8 0.9 89. 0.08 0.8 87.07 0. Chlorate 0. 0.0 0.. 0.00. 7. 0.0 0. Bromide 8.8 0.0. 7. 0.00. 9. 0.0. Nitrate. 0.0.8. 0.0.. 0.0. Sulfate.9 0.9.. 0. 0.. 0. Oxalate 79.0 0.. 9. 0.0.0.0 0. 0. Phosphate 99. 0.0. 0. 0.009.0. 0.07 0. n=7 Table. Peak Area Reproducibilities of Ten Anions in Water, 9.9% (. M), and 9.9% (7. M) Sodium Hydroxide Application Note 9

Method Qualification The Park and AutoNeutralization method was qualified in water, and 9.9% (w/w) and 9.9% (w/w) sodium hydroxide solutions before determination of trace anions in other concentrated-base samples. The linearity of eleven anions over a four-fold or more concentration range, typical noise, method detection limits (MDLs), reproducibilities, and spike recoveries were determined in water solutions. Spike recovery and -day ruggedness were determined in 9.9% and 9.9% sodium hydroxide solutions. The linearity of peak response for fluoride, chlorite, bromide, and chlorate was measured by determining each anion in five replicates each of seven standards (, 0, 0, 0, 80, 0, and 00 µg/l). The linearities of peak response for bromate, oxalate, and phosphate were determined from 0 00 µg/l (0, 0, 0, 80, 0, 0, and 00 µg/l). The calibration results showed good linearities over these concentration ranges, r >0.999, (Table 7). Chlorite was the exception and did not show linearity in this range (not shown). Table 7. Method Detection Limit (MDL) and Linearity MDL a (µg/l) RSD Linearity b (r ) Fluoride 0..9 0.999 Bromate.9. 0.9990 Chloride.7 c 9. 0.9999 Nitrite N.A. 0.999 Chlorate.70 8. 0.9999 Bromide.07.8 0.999 Nitrate 0. 9.0 0.9989 Sulfate. c.9 0.9997 Oxalate.7 c.7 0.999 Phosphate.0. 0.9999 n=7 a MDL is defined as three times the noise. The linearity of peak response was measured at higher concentrations for nitrite, nitrate, chloride, and sulfate. The linearities of peak response for chloride and sulfate were measured with triplicate injections of eight standards from 00 to 00 µg/l (00, 00, 000, 00, 000, 00, 000, and 00 µg/l). The linearity of peak response for nitrate was measured with triplicate injections of nine standards from to 00 µg/l (, 0, 0, 0, 80, 0, 00, 00, and 00 µg/l). Nitrite was measured in a similar way. The response was not linear at the lower concentrations (<00 µg/l) but it exhibited good linearity (r > 0.999) from 00 to 00 µg/l. Nitrate, chloride, and sulfate had good linearity for the full concentration ranges. The noise was determined over five, 90-min runs, when water was injected, by measuring the noise in -min intervals from to 90 min. The noise value determined by this experiment was. ± 0. ns (n=). The method detection limit (MDL) was defined as three times the noise level,.8 ±. ns. For this application, the MDL concentrations for the eight analytes (except chloride, sulfate, and oxalate) ranged from 0. to. µg/l. The MDL concentrations for chloride, sulfate, and oxalate were set to their baseline concentrations found in the blank,.7,., and. µg/l, respectively. Spike recovery was determined for chloride, bromate, chlorate, bromide, nitrate, sulfate, oxalate, and phosphate in 9.9% sodium hydroxide, and 9.9% sodium hydroxide solutions. In the first spike recovery experiments, 7. µg/l bromate, chlorate, bromide, nitrate, and oxalate, 8. µg/l chloride and sulfate, and 9. µg/l phosphate were spiked into 9.9% sodium hydroxide. The recovery was good, 8. to 09.% (Table 8). Similar spike recoveries (97..%) were found when spiking 9.9% sodium hydroxide with. µg/l bromate, chlorate, bromide, and nitrate, µg/l chloride and sulfate, and 8. µg/l oxalate and phosphate. b The linearity of fluoride, chlorate, and bromide was determined from to 00 µg/l. The linearity of bromate, phosphate and oxalate were determined from 0 to 00 µg/l. The linearity of chloride and sulfate was determined from 00 to 00 µg/l, the linearity of nitrite and nitrate from 00 to 00 µg/l and to 00 µg/l, respectively. c The MDLs of chloride, sulfate, and oxalate are the concentrations in the blank. Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Table 8. Recovery of Eight Anions Spiked in 9.9% (. M), and 9.9% (7. M) Sodium Hydroxide Amount in Amount Amount RSD Amount in Amount Amount RSD 9.9% NaOH Spiked into Recovered from 9.9% NaOH Spiked into Recovered from (µg/l) 9.9% NaOH 9.9% NaOH (µg/l) 9.9% NaOH 9.9% NaOH (µg/l) (%) (µg/l) (%) Bromate N.D. 7. 99.. N.D.. 08..7 Chloride 78. 8. 9.0 0. 90. 0.0. Chlorate.7 7. 0. 0.7... 0. Bromide N.D. 7. 09...... Nitrate 8. 7. 90...7. 0.7.8 Sulfate. 8. 88.7 0. 00 97. 0.7 Oxalate. 7. 8. 0.9 8. 8. 97.8. Phosphate N.D. 9. 0.7. N.D. 8.. 0. n= N.D. not detected. We determined the reproducibility and ruggedness of the Park and AutoNeutralization method over 0 injections, about 8 h. During this study we measured four separately prepared solutions each of 9.9% (w/w) and 9.9% (w/w) sodium hydroxide, each spiked with 00 µg/l of chloride and sulfate and 0 µg/l of oxalate. Deionized water injections were inserted between the sample groups. The results (Figures 0 ) show that retention times and peak areas were stable over 8 h of the experiment. The retention time reproducibilities for chloride, sulfate, and oxalate were 8. ± 0.0, 9.87 ± 0.0, and. ± 0.0 min, respectively. Notice that the sample preparation times were subtracted for reproducibility determinations. The peak areas were reproducible and stable (0.7 ± 0.0, 0. ± 0.0, and 0.0 ± 0.00 µs-min, for chloride, sulfate, and oxalate, respectively). Retention Time (min) 0 0 Oxalate Sulfate Chloride 0 0 0 0 80 00 0 0 0 Number of Injections 7 Figure 0. Retention time stability of chloride, sulfate, and oxalate over days. 00 µg/l of chloride and sulfate, and 0 µg/l of oxalate spiked into 0% sodium hydroxide. Peak Area (µs-min) 0. 0. 0. Chloride Sulfate Oxalate 0 0 0 0 0 80 00 0 0 0 Number of Injections 8 Figure. Peak area stability of chloride, sulfate, and oxalate over days, 00 µg/l of chloride and sulfate, and 0 µg/l of oxalate spiked into 0% sodium hydroxide. Application Note 9

Chloride, sulfate, and oxalate also had stable retention times in 9.9% sodium hydroxide: 8.8 ± 0.0, 9.8 ± 0.0, and.0 ± 0.0 min, respectively (Figures ). The peak areas for chloride, sulfate, and oxalate were approximately three times those for the 9.9% sodium hydroxide solutions:.0 ± 0.0,.8 ± 0.0, and 0.07 ± 0.00 µs-min, respectively. These experiments showed that the ASRN II neutralizer neutralized the 0% sodium hydroxide solutions in 8 min. Retention Time (min) 0 0 Oxalate Sulfate Determination of Trace Anions in 9.9% Potassium Hydroxide and Concentrated (9.7%) Ammonium Hydroxide We applied the Park and AutoNeutralization method to determine trace anions in 9.9% potassium hydroxide and 9.7% (concentrated) ammonium hydroxide. Spike recovery of selected anions and background sulfate concentrations were determined for both solutions. The 9.9% potassium hydroxide solution had low concentrations of chlorate (. ± 0.9 µg/l), moderate concentrations of oxalate (08.9 ±. µg/l), and high concentrations of sulfate (7. ±.8 µg/l) and chloride (08 ± µg/l) (Figure ). Spike recoveries of 9. µg/l bromate, chlorate, bromide, and nitrate, and 79. µg/l oxalate, 99. µg/l phosphate, 08 µg/l chloride, and 9 µg/l sulfate were good, at 9. 09.% (Table 9). Peak Area (µs-min)..0..0 Chloride 0 0 0 0 80 00 0 0 0 Number of Injections 9 Figure. Retention time stability of chloride, sulfate, and oxalate over days, 00 µg/l of chloride and sulfate and 0 µg/l of oxalate spiked into 0% sodium hydroxide. Sulfate Chloride 0. Oxalate 0 0 0 0 0 80 00 0 0 0 Number of Injections 0 Prep. Carrier: Degassed deionized water Temperature: 0 ºC Carrier Flow Rate: 0. ml/min Inj. Volume: 0 µl loop Neutralization: Park and AutoNeutralize for 8 min in ASRN II, external water, 9 ma Analytical Column: IonPac AG9, AS9, -mm Eluent (EluGen KOH): 0 to mm potassium hydroxide from 0 to min, mm potassium hydroxide from to min Column Temp.: 0 ºC AS Temperature: 0 ºC Flow Rate:.0 ml/min Concentration: IonPac UTAC-XLP Detection: Suppressed conductivity, recycle, ma.00 7 Peaks:. Fluoride. µg/l. Butyrate. Acetate. Formate. Chlorite.. Bromate 0. 7. Chloride 8 8. Nitrite 7.8 9. Chlorate. 0. Bromide 9.8. Nitrate.. Unknown. Carbonate. Sulfate 70.0. Unknown. Oxalate 0.9 7. Unknown 8. Phosphate 9. 9. Unknown 0. Unknown Figure. Peak area stability of chloride, sulfate, and oxalate over days, 00 µg/l of chloride and sulfate, and 0 µg/l of oxalate spiked into 0% sodium hydroxide. µs 8 8 9 0 7 9 0 0. 9..0 0.0.0 0.0.0 0.0.0 Minutes Figure. 9.9% potassium hydroxide spiked with 0 µg/l of fluoride, chlorite, bromate, nitrite, chlorate, bromide, and nitrate, 80 µg/l of oxalate, 00 µg/l of phosphate, 0 µg/l of chloride, and 00 µg/l of sulfate. Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Table 9. Recovery of Eight Anions Spiked in 9.9% (. M) Potassium Hydroxide Amount in RSD Amount Amount RSD 9.9% KOH Added Recovered (µg/l) (µg/l) (%) Bromate N.D. 9. 0..7 Chloride 08 0.8 08 9.7. Chlorate. 0.8 9. 9.7. Bromide N.D. 9. 00.. Nitrate N.D. 9. 9.7. Sulfate 7.. 9 9.7. Oxalate 08.9. 79. 9..0 Phosphate N.D. 99. 09.. n= N.D. not detected. Prep. Carrier: Degassed deionized water Temperature: 0 ºC Carrier Flow Rate: 0. ml/min Inj. Volume: 0 µl loop Neutralization: Park and AutoNeutralize for 8 min in ASRN II, external water, 9 ma Analytical Column: IonPac AG9, AS9, -mm Eluent (EluGen KOH): 0 to mm potassium hydroxide from 0 to min, mm potassium hydroxide from to min Column Temp.: 0 ºC AS Temperature: 0 ºC Flow Rate:.0 ml/min Concentration: IonPac UTAC-XLP Detection: Suppressed conductivity, recycle, ma.0 Peaks:. Fluoride 8. µg/l. Butyrate. Acetate. Formate. Chloride 09.. Nitrite 9. 7. Bromide 8.0 8. Nitrate 8. 9. Unknown 0. Carbonate. Sulfate 7.0. Oxalate.9. Unknown. Unknown. Phosphate.. Unknown 7. Unknown 8. Unknown The concentrated (9.7%,. M) ammonium hydroxide solution (Figure ) had low concentrations of chlorite, oxalate, and bromide (< µg/l), moderate concentrations of chloride (09. ±. µg/l) and nitrate (8. ±. µg/l), and high concentrations of nitrite (9. ±. µg/l), and sulfate (7.0 ±.9 µg/l). Spike recoveries of.8 µg/l bromide, 8.0 µg/l nitrate, 00.8 8.8 µg/l of chloride, nitrite, and sulfate, and 8.0 µg/l oxalate were good, at 9. to 07.0% (Table 0). µs 7 8 0. 9..0 0.0.0 0.0.0 0.0.0 Minutes Figure. Concentrated (9.7% as NH OH,. M) ammonium hydroxide. 9 0 7 8 Table 0. Recovery of Six Anions Spiked in 9.7% (. M) Ammonium Hydroxide Amount in RSD Amount Amount RSD Concentrated Added Recovered NH OH (µg/l) (%) (µg/l) Chloride 09..9 00.8 9. 0. Nitrite 9..0 8.8 07.0.8 Bromide N.D..8 00.. Nitrate 8.. 8.0 9.9. Sulfate 7.0 0.9 8.8 99. 0. Oxalate.9.9 9.0 00.9 0. n= N.D. not detected. Conclusion Using the Automation Manager, the dual pump, dual detectors, and eluent generation of an ICS-000, this application note demonstrates that strong bases can be neutralized to allow determination of µg/l anion concentrations. The neutralization occurs automatically within the temperature-controlled environment of the ICS-000 DC using the ASRN II neutralizer. The setup described here is an easy, automated, and reproducible method to determine µg/l anion concentrations in strong bases. Although not discussed in this application note, trace concentrations of cations can also be determined in concentrated acids using the same techniques and the corresponding cation components. Application Note 9

Precautions Always handle concentrated bases with extreme caution. Consult the Material Safety Data Sheets (MSDS) for protective clothing, storage, disposal, and health effects. Use only low pressure concentrator columns, such as the IonPac UTAC-XLP, with the ASRN II neutralizer. Pressures >00 psi can cause irreversible damage. Always verify that the external water regenerant is flowing through the neutralizer before applying current and turn it off when the pump is off. In this application, baseline sulfate levels should be determined for your samples because sulfate from the neutralizer can be extracted by the sample during the neutralization process. Fluoride and chlorite can not be determined quantitatively with this method. Nitrite was not stable at low µg/l concentrations and therefore quantitative determinations are not recommended below 00 µg/l. Always use high-quality deionized water (>8 M-ohm) for preparing reagents and samples. Do not over-tighten the fittings on the Carbonate Removal Device (CRD): tighten to finger-tight. References. Preparing Vegetables, Home of Australia s Favorite Food Brands.htm, Simplot Australia Pty Ltd, 00.. Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment/Ion Chromatography. Application Note 9; LPN 0. Dionex Corporation, Sunnyvale, CA, 99.. Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment/Ion Chromatography. Application Note 9; LPN -0. Dionex Corporation, Sunnyvale, CA, 00.. Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment / Ion Chromatography. Application Note 9; LPN. Dionex Corporation, Sunnyvale, CA, 00.. Srinivasan, K.; Lin, R.; Avdalovic, N.; Pohl, C.; A New Neutralizer Method for Analysis of Trace Anions in Bases and Trace Cations in Acids; LPN 9. Poster presented at the 7th Annual International Ion Chromatography Symposium (IICS), Trier, Germany, September 0-, 00.. Reducing Carbonate Interference in Anion Determinations with the Carbonate Removal Device (CRD). Technical Note ; LPN 7. Dionex Corporation, Sunnyvale, CA, 00. 7. Determination of Trace Anions in High Purity Waters by High-Volume Direct Injection with the EG0. Application Update ; LPN 90. Dionex Corporation, Sunnyvale, CA, 00. 8. Product Manual for Anion Self-Regenerating Neutralizer ASRN II and Cation Self- Regenerating Neutralizer CSRN II. LPN 09. Dionex Corporation, Sunnyvale, CA, 998. 9. Operator s Manual for ICS-000 Ion Chromatography System; LPN 00. Dionex Corporation, Sunnyvale, CA, 00. 0. Installation Instructions for ICS-000 Ion Chromatography System; LPN 00. Dionex Corporation, Sunnyvale, CA, 00.. Product Manual for IonPac Anion Trap Column ATC-HC; LPN 097. Dionex Corporation, Sunnyvale, CA, 00.. Product Manual for Anion Self-Regenerating Neutralizer ASRN II and Cation Self- Regenerating Neutralizer CSRN II: QuickStart; LPN 09. Dionex Corporation, Sunnyvale, CA, 998.. Operator s Manual for AS Autosampler; LPN 00. Dionex Corporation, Sunnyvale, CA, 00.. Operator s Manual for ICS-000 Ion Chromatography System; LPN 00. Installing a New EGC Cartridge, Installing a New CR-ATC, and Installing a Backpressure Coil, Sections 9.. 9.., 9.. 9.., and 9., P. 0 7, P. 9 and P.. Dionex Corporation, Sunnyvale, CA, 00.. Product Manual for IonPac UTAC; LPN 009. Dionex Corporation, Sunnyvale, CA, 00.. Product Manual for IonPac AG9 Guard and AS9 Analytical Columns; LPN 000. Dionex Corporation, Sunnyvale, CA, 00. 7. Product Manual for Anion Self-Regenerating Suppressor ULTRA II and Cation Self- Regenerating Suppressor ULTRA II; LPN 09. Dionex Corporation, Sunnyvale, CA, 00. 8. Product Manual for Carbonate Removal Device; LPN 008. Dionex Corporation, Sunnyvale, CA, 00. Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Suppliers Fisher Scientific International Inc., Liberty Lane, Hampton, NH 08 USA, -800-7-7000, www.fisherscientific.com VWR International, Inc., Goshen Corporate Park West, 0 Goshen Parkway, West Chester, PA 980 USA, -800-9-000, www.vwrsp.com Sigma-Aldrich, Inc., P.O. Box 9, Dallas, TX 79- USA, -800--00, www.sigmaaldrich.com Teflon is a registered trademark of E.I. du Pont de Nemours. PEEK is a trademark of Victrex PLC. Reagent-Plus is a registered trademark of Sigma-Aldrich Biotechnology LP. All other trademarks or registered trademarks are property of Dionex Corporation. Passion. Power. Productivity. Dionex Corporation 8 Titan Way P.O. Box 0 Sunnyvale, CA 9088-0 (08) 77-0700 North America U.S. (87) 9-700 Canada (90) 8-90 South America Brazil () 7 0 Europe Austria () Benelux () 0 8 978 () 9 Denmark () 90 90 France () 9 0 0 0 Germany (9) 99 0 Ireland () 00 Italy (9) 0 7 Switzerland () 0 99 United Kingdom () 7 97 Asia Pacific Australia () 90 China (8) 8 8 India (9) 77 Japan (8) 88 Korea (8) 80 Singapore () 89 90 www.dionex.com Application Note 9 LPN 88 090/ 007 Dionex Corporation