Luminescence Spectroscopy Excitation is very rapid (10-15 s). Vibrational relaxation is a non-radiational process. It involves vibrational levels of

Similar documents
Introduction to Chromatographic Separations

Chapter 15 Molecular Luminescence Spectrometry

Analytical Chemistry

Introduction to Chromatography

Chromatography. Gas Chromatography

HPLC Background Chem 250 F 2008 Page 1 of 24

A Fluorometric Analysis of Quinine in Tonic Water

Molecular Luminescence Spectroscopy

Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis

LABORATORY OF ELEMENTARY BIOPHYSICS

Chromatographic Analysis

Separation Methods Based on Distributions in Discrete Stages (02/04/15)

Molecular Luminescence. Absorption Instrumentation. UV absorption spectrum. lg ε. A = εbc. monochromator. light source. Rotating mirror (beam chopper)

Remember - Ions are more soluble in water than in organic solvents. - Neutrals are more soluble in organic solvents than in water.

Singlet. Fluorescence Spectroscopy * LUMO

What is Chromatography?

Ch 313 FINAL EXAM OUTLINE Spring 2010

Chemistry Instrumental Analysis Lecture 26. Chem 4631

What the Einstein Relations Tell Us

Information given in these slides are, either in part or all, recollection from the followings:

Instrumental Analysis II Course Code: CH3109. Chromatographic &Thermal Methods of Analysis Part 1: General Introduction. Prof. Tarek A.

Chromatography and other Separation Methods

Luminescence transitions. Fluorescence spectroscopy

Chromatographic Separation

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

Chem 454 instrumental Analysis Exam 1 February 6 th, 2008

3) In CE separation is based on what two properties of the solutes? (3 pts)

Liquid Chromatography

CHEM 429 / 529 Chemical Separation Techniques

Chapter 26. An Introduction to Chromatographic Separations. Chromatography

Open Column Chromatography, GC, TLC, and HPLC

Chapter 26: An Introduction to Chromatographic Separations

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

PHOTOCHEMISTRY NOTES - 1 -

Reflection = EM strikes a boundary between two media differing in η and bounces back

High Performance Liquid Chromatography

Separations---Chromatography and Electrophoresis

Fluorescence (Notes 16)

CHAPTER 1. Introduction, Chromatography Theory, and Instrument Calibration

Chemistry Instrumental Analysis Lecture 3. Chem 4631

High Pressure/Performance Liquid Chromatography (HPLC)

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

Chapter 17: Fundamentals of Spectrophotometry

1. Photoreduction of Benzophenone in 2-Propanol

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Analytical Technologies in Biotechnology Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

CHEM Outline (Part 15) - Luminescence 2013

CHAPTER 13 Molecular Spectroscopy 2: Electronic Transitions

高等食品分析 (Advanced Food Analysis) I. SPECTROSCOPIC METHODS *Instrumental methods: 1. Spectroscopic methods (spectroscopy): a) Electromagnetic radiation

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography

Chapter 6 Photoluminescence Spectroscopy

Chromatography. Mrs. D. MEENA MPharm PA & QA KTPC

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Chapter 17: Fundamentals of Spectrophotometry

Cork Institute of Technology. Summer 2005 Instrumental Analysis (Time: 3 Hours) Section A

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Excited State Processes

Chromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?

Introduction to Gas Chromatography

Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. CHAPTER 26

2. a) R N and L N so R L or L R 2.

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Electronic Spectra of Complexes

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

1. Cyclic voltammetry involves the measurement of a diffusion controlled at an electrode in which the is controlled. (4 points)

Volatile organic compounds (VOCs):

LAB #3: FLUROESCENCE SPECTROSCOPY AND ELECTRON TRANSFER (This lab is adapted from the U of MN Phsyical Chemistry lab manual)

Instrumental Chemical Analysis

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

Instrumental Analysis

Lecture 3: Light absorbance

ERT320 BIOSEPARATION ENGINEERING CHROMATOGRAPHY

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

2] The plate height in chromatography is best described as 2

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Lecture- 08 Emission and absorption spectra

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid

Chapter content. Reference

Fluorescence 2009 update

Chapter 23 Introduction to Analytical Separations

Fluorescence Workshop UMN Physics June 8-10, 2006 Quantum Yield and Polarization (1) Joachim Mueller

Chromatography & instrumentation in Organic Chemistry

An Introduction to Chromatographic Separations

I. Proteomics by Mass Spectrometry 1. What is an internal standard and what does it accomplish analytically?

Determination of Caffeine by HPLC

Chromatography Outline

Chem Homework Set Answers

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

single-molecule fluorescence resonance energy transfer

A very brief history of the study of light

Spectroscopy Problem Set February 22, 2018

LC Technical Information

Biochemistry. Biochemical Techniques HPLC

Direct, Reliable, and Correct Oxygen Measurement

two slits and 5 slits

Fluorescence Excitation and Emission Fundamentals

Transcription:

Luminescence Spectroscopy Excitation is very rapid (10-15 s). Vibrational relaxation is a non-radiational process. It involves vibrational levels of the same electronic state. The excess of vibrational energy is released by the excited molecule in the form of thermal or vibrational motion to the solvent molecules. It takes between 10-11 and 10-10 s. Internal conversion is a non-radiational process. It is the crossover between states of the same multiplicity. It occurs when the high vibrational levels of the lower electronic state overlap with the low vibrational levels of the high electronic state. It can occur between excited states (S 2 and S 1 ) or between excited and ground states (S 1 and S 0 ). Between excited states is rapid (10-12 s). Fluorescence is a radiational transition between S 1 and S 0. It occurs from the ground vibrational level of S 1 to various vibrational levels in S 0. It requires 10-10 to 10-6 s to occur. External conversion is a nonradiative process in which excited states transfer their excess energy to other species, such as solvent or solute molecules. Intersystem crossing is a non-radiative process between electronic states of different multiplicities. It requires a change in electronic spin and, therefore, it has a much lower probability to occur than spin-allowed transitions. Because the time scale is similar to the one for fluorescence (10-8 - 10-7 s), it competes with fluorescence for the deactivation of the S 1 state Phosphorescence is a radiational deactivation process between electronic states of different multiplicity, i.e. T 1 to S 0. It usually takes between 10-4 to 10s to occur because the process is spin forbidden. 139

Intensity Intensity of fluorescence (or phosphorescence) emission as a function of fluorophor (or phosphor) concentration The power of fluorescence (or phosphorescence) radiation (F or P) is proportional to the radiant power of the excitation beam that is absorbed by the system: F = K (P 0 P) (1) The transmittance of the sample is given by Beer s law: P/P 0 = 10 -ebc (2) Re-arranging Eq. 1: F = K P 0 (1 P / P 0 ) Substituting Eq. 2 above: F = K (1 10 -ebc ) (3) The exponential term in Eq. 3 can be expanded as a Maclaurin series to: F = K.P 0.[2.303ebc (2.303ebc) 2 + (2.303ebc) 3 - ] (4) 2! 3! For diluted solutions, i.e. 2.303ebc < 0.05, all of the subsequent terms in the brackets become negligible with respect to the first, so: F = K.P 0.2.303ebc or F = 2.303.P 0.K.ebc (5) A plot of fluorescence (or phosphorescence) intensity as a function of concentration should be linear up to a certain concentration. Three are the main reasons for lack of linearity at high concentrations: a) 2.303ebc > 0.05 b) self-quenching c) self-absorption P 0 P F Detector LDR Concentration 140

Fluorescence quantum yield The slope of the calibration curve is equal to 2.303.P 0.K.ebc. K is also known as the fluorescence quantum yield (f F ). The fluorescence quantum yield is the ratio between the number of photons emitted as fluorescence and the number of photons absorbed: f F = # of fluorescence photons # of absorbed photons Fluorescence quantum yields may vary between zero and unity: 0 < f F < 1 The higher the quantum yield, the stronger the fluorescence emission. In terms of rate constants, the fluorescence quantum yield is expressed as follows: f F = kf kf + ki + kec + kic + kpd +kd kf kic kec ki Consider a dilute solution of a fluorescent species A whose concentration is [A] (in mol.l -1 ). A very short pulse of light at time 0 will bring a certain number of molecules A to the S 1 excited state by absorption of photons: A + hν A* The excited molecules then return to S 0, either radiatively or non-radiatively, or undergo intersystem crossing. As in classical kinetics, the rate of disappearance of excited molecules is expressed by the following differential equation: -d[a*] / dt = (kf + knr) [A*] knr = ki + kec + kic + kpd + kd are the rate constants of competing processes that ultimately reduce the intensity of fluorescence. k units = s -1. 141

Fluorescence lifetime Integration of equation: -d[a*] / dt = (kf + knr) [A*] yields the time evolution of the concentration of excited molecules [A*]. Let [A*] 0 be the concentration of excited molecules at time 0 resulting from the pulse light excitation. Integration leads to: [A*] = [A*] 0 exp(-t / t) where t is the lifetime of the excited state S 1. The fluorescence lifetime is then given by: t = 1 / kf + knr Typical fluorescence lifetimes are in the ns range. Typical phosphorescence lifetimes are in the ms to s range. The fluorescence (or phosphorescence lifetime) is the time needed for the concentration of excited molecules to decrease to 1/e of its original value. The fluorescence lifetime correlates to the fluorescence quantum yield as follows: f f = kf. t 142

Excitation and emission spectra Excitation spectra appear in the same wavelength region as absorption spectra. However, it is important to keep in mind that absorption spectra are not the same as excitation spectra. Fluorescence and phosphorescence spectra appear at longer wavelength regions than excitation spectra. Phosphorescence spectra appear at longer wavelength regions than fluorescence spectra. In some cases, it is possible to observe vibrational transitions in room-temperature fluorescence spectra. There are several parameters (many wavelengths and two lifetimes) for compound identification. 143

Fluorescence and Structure General rule: Most fluorescent compounds are aromatic. An increase in the extent of the p-electron system (i.e. the degree of conjugation) leads to a shift of the absorption and fluorescence spectra to longer wavelengths and an increase in the fluorescence quantum yield. Example: Aromatic hydrocarbon Naphthalene Anthracene Naphthacene Fluorescence ultraviolet blue green Pentacene red The lowest-lying transitions of aromatic hydrocarbons are of the p p* type, which are characterized by high molar absorption coefficients and relatively high quantum fluorescence quantum yields. No fluorescence Fluorescence 144

Substituted Aromatic Hydrocarbons The effect of substituents on the fluorescence characteristics of aromatic hydrocarbons varies with the type of substituent. Heavy atoms: the presence of heavy atoms (e.g. Br, I, etc.) results in fluorescence quenching (internal heavy atom effect) because of the increase probability of intersystem crossing (ISC). Electron- donating : -OH, -OR, -NH 2, -NHR, -NR 2 This type of substituent generally induces and increase in the molar absorption coefficient and a shift in both absorption and fluorescence spectra. Electron-withdrawing substituents: carbonyl and nitro-compounds Carbonyl groups: There is no general rule. Their effect depends on the position of the substituent group in the aromatic ring. Nitro groups: No detectable fluorescence. 145

Rigidity Rigidity usually enhances fluorescence emission ph of solution Additional resonance forms lead to a more stable first excited state 146

Quenching: nonradiative energy transfer from an excited species to other molecules. Quenching results in deactivation of S 1 without the emission of radiation causing a decrease in fluorescence intensity. Types of quenching: dynamic, static and others. Dynamic quenching: or collisional quenching requires contact between the excited species and the quenching agent. Dynamic quenching is a diffusion controlled process. As such, its rate depends on the temperature and viscosity of the sample. High temperatures and low viscosity promote dynamic quenching. For dynamic quenching with a single quencher, the Stern-Volmer expression is valid: F 0 /F = 1 + K q [Q] Where F 0 and F are the fluorescence intensities in the absence and the presence of quencher, respectively. [Q] is the concentration of quencher (mols.l -1 ) and K q is the Stern-Volmer quenching constant. The Stern-Volmer constant is defined as: K q = kq / kf + ki + kic where kq is the rate constant for the quenching process (diffusion controlled). Static Quenching: the quencher and the fluorophor in the ground state form a complex. The complex is no fluorescent (dark complex). The Stern-Volmer equation is still valid but Kq in this case is the equilibrium constant for complex formation: A + Q <=> AQ. In static quenching, the fluorescence lifetime of the fluorophor is not affected. In dynamic quenching, the fluorescence lifetime of the fluorophor is affected. So, lifetime can be used to distinguish between dynamic and static quenching. Quenching Fluorescence quenching of quinine sulfate as a function of chloride concentration Oxygen sensors O 2 is paramagnetic, i.e. its natural electronic configuration is the triplet state. When O 2 interacts with the fluorophor, it promotes conversion and deactivation of excited fluorophor molecules. Upon interaction, O 2 goes into the singlet state (diamagnetic). 147

Instrumentation Typical spectrofluorometer (or spectrofluuorimeter) configuration Recording excitation and emission spectra Recording synchronous fluorescence spectra 148

Spectrofluorimeter capable to correct for source wavelength dependence Spectrofluorimeter with the ability to record total luminescence 149

Measuring phosphorescence Using a spectrofluorimeter with a continuous excitation source: be careful with fluorescence and second order emission! Rotating-can phosphoroscope: good-old approach to discriminate against fluorescence. A. True representation B. Approximate representation t e = exposure time; t d = shutter delay time; t t = shutter transit time; t C = time for one cycle of excitation and observation; t E = t e + t t ; t D = t d + t t Spectrofluorimeters with a pulsed source: employ a gated PMT for fluorescence discrimination. Excitation source on => PMT off = fluorescence decays Excitation source off => PMT on phosphorescence measurement 150

Introduction to Chromatographic Separations Analysis of complex samples usually involves previous separation prior to compound determination. Two main separation methods instrumentation are available: based on Chromatography Electrophoresis Chromatography is based on the interaction of chemical species with a mobile phase (MP) and a stationary phase (SP). The MP and the SP are immiscible. The sample is transported by the MP. The interaction of species with the MP and the SP separates chemical species in zones or bands. The relative chemical affinity of chemical species with the MP and the SP dictates the time the species remain in the SP. MP + sample SP Detector Two general types of chromatographic techniques exist: Planar: flat SP, MP moves through capillary action or gravity Column: tube of SP, MP moves through gravity or pressure. 151

Classification of Chromatographic Methods Chromatographic methods can be classified on the type of MP and SP and the kinds of equilibrium involved in the transfer of solutes between phases: Column 1: Type of MP Column 2: Type of MP and SP Column 3: SP Column 4: Type of equilibrium Concentration profiles of solute bands A and B at two different times in their migration down the column. 152

Migration Rates of Solutes Two-component chromatogram illustrating two methods for improving separation: (a) Original chromatogram with overlapping peaks; (b) improvement brought about an increase in band separation; (c) improvement brought about by a decrease in the widths. Distribution constant or partition ratio or partition coefficient: It describes the partition equilibrium of an analyte between the SP and the MP. If K = K c and SP = S and MP = M K c = c S / C M = n S / V S n M / V M Where V S and V M are the volumes of the two phases and n S and n M are the moles of A in SP and MP, respectively. 153

Retention Time Retention time is a measured quantity. From the figure: t M = time it takes a non-retained species (MP) to travel through the column = dead or void time. t R = retention time of analyte. The analyte has been retained because it spends a time t S in the SP. The retention time is then given by: t R = t S + t M The average migration rate (cm/s) of the solute through the column is: Where L is the length of the column. The average linear velocity of the MP molecules is: 154

Relationship Between Retention Time and Distribution Constant The Rate of Solute Migration: The Retention Factor Retention factor = capacity factor = k A = k A. However: k A K C or k A K A Where V S and V M are the volumes of SP and MP in the column. Knowing that: => An equation can be derived to obtain K C of A (K A ) as a function of experimental parameters. => 155

Band Broadening and Column Efficiency The shape of an analyte zone eluting from a chromatographic column follows a Gaussian profile. Some molecules travel faster than the average. The time it takes them to reach the detector is: t R Dt. Some molecules travel slower than the average molecule. The time it takes them to reach the detector is: t R + Dt The selectivity factor can be measured from the chromatogram. Selectivity factors are always greater than unity, so B should be always the compound with higher affinity by the SP. Selectivity factors are useful parameters to calculate the resolving power of a column. So, the Gaussian provides an average retention time (most frequent time) and a time interval for the total elution of an analyte from the column. The magnitude of Dt depends on the width of the peak. 156

Considering that: v_ = L / t R => L = v_ x or L ± DL = v_ x [t R ± Dt] The equation above correlates chromatographic peaks with Gaussian profiles to the length of the column. Both Dt and DL correspond to the standard deviation of a Gaussian peak: Dt DL s If s = Dt, s units are in minutes. If s = DL, s units are in cm. The width of a peak is a measure of the efficiency of a column. The narrower the peak, the more efficient is the column. => The efficiencies of chromatographic columns can be compared in terms of number of theoretical plates. 157

The Plate Theory The plate theory supposes that the chromatographic column contains a large number of separate layers, called theoretical plates. Separate equilibrations of the sample between the stationary and mobile phase occur in these "plates". The analyte moves down the column by transfer of equilibrated mobile phase from one plate to the next. It is important to remember that theoretical plates do not really exist. They are a figment of the imagination that helps us to understand the processes at work in the column. As previously mentioned, theoretical plates also serve as a figure of merit to measuring column efficiency, either by stating the number of theoretical plates in a column (N) or by stating the plate height (H); i.e. the Height Equivalent to a Theoretical Plate. The number of theoretical plates is given by: N = L / H If the length of the column is L, then the HETP is: H = s 2 / L Note: For columns with the same length (same L): The smaller the H, the narrower the peak. The smaller the H, the larger the number of plates. => column efficiency is favored by small H and large N. It is always possible to using a longer column to improve separation efficiency. 158

Experimental Evaluation of H and N Calling s in time units (minutes or seconds) as t: => s / t = cm / s Considering that: v_ = L / t R = cm /s We can write: L / t R = s / t or t = s = s L / t R v_ If the chromatographic peak is Gaussian, approximately 96% of its area is included within ± 2s. This area corresponds to the area between the two tangents on the two sides of the chromatographic peak. The width of the peak at its base (W) is then equal to: W = 4t in time units or W = 4s in length units. Substituting t = W / 4 in the equation above we obtain: s = LW / 4t R Substituting s = W / 4 in the same equation we obtain: t = Wt R / 4L Substituting s = LW / 4t R in the HEPT equation: H = LW 2 / 16t R 2 Substitution of this equation in N gives: N = 16 (t R / W) 2 These two equations allow one to estimate H and N from experimental parameters. If one considers the peak of the width at the half maximum (W 1/2 ): N = 5.54 (t R / W 1/2 ) 2 In comparing columns, N and H should be obtained with the same compound! 159

Kinetic Variables Affecting Column Efficiency The plate theory provides two figures of merit (N and H) for comparing column efficiency but it does not explain band broadening. Table 26-2 provides the variables that affect band broadening in a chromatographic column and, therefore, affect column efficiency. The effect of these variables in column efficiency is best explained by the theory of band broadening. This theory is best represented by the van Deemter equation: H = A + B / u+ C S. u + C M. U where H is in cm and u is the velocity of the mobile phase in cm.s -1. The other terms are explained in Table 26-3. Table 26-3: f(k) and f (k) are functions of k l and g are constants that depend on the quality of the packing. B is the coefficient of longitudinal diffusion. C S and C M are coefficients of mass transfer in stationary and mobile phase, respectively. Before we try to understand the meaning of the van Deemter equation, a better understanding of chromatographic columns is needed. 160

Some Characteristics of Gas Chromatography (GC) Columns Two general types => Open Tubular Columns (OTC) or Capillary Columns => Packed Columns OTC: => Wall Coated Open Tubular (WCOT) Columns => Support Coated Open Tubular (SCOT) Columns => The most common inner diameters for capillary tubes are 0.32 and 0.25mm. Packed Columns: => Glass tubes with 2 to 4mm inner diameter. => Packed with a uniform, finely divided packing material of solid support, coated with a thin layer (005 to 1mm) of liquid stationary phase. Solid Support Material in OTC and Packed Columns Diatomaceous Earth: skeletons of species of single-celled plants that once inhabited ancient lakes and seas. 161

Some Characteristics of HPLC Columns Only packed columns are used in HPLC. Current packing consists of porous micro-particles with diameters ranging from 3 to 10mm. The particles are composed of silica, alumina, or an ion-exchange resin. Silica particles are the most common. Thin organic films are chemically or physically bonded to the silica particles. The chemical nature of the thin organic film determines the type of chromatography. SP for normal-phase chromatography Typical microparticle SP for partition chromatography 162

Another Look at the van Deemter Equation H = A + B/u + C S u + C M u The multi-path term A or Eddy-Diffusion: => This term accounts for the multitude of pathways by which a molecule (or ion) can find its way through a packed column. A = 2ld p where: l is a geometrical factor that depends on the shape of the particle: 1 l 2. d p is the diameter of the particle. Using packing with spherical particles of small diameters should reduce eddy -diffusion. Injection Detector 163

The Longitudinal Diffusion Term B/u: Longitudinal diffusion is the migration of solute from the concentrated center of the band to the more diluted regions on either side of the analyte zone. B/u = 2gD M /u where: g is a constant that depends on the nature of the packing and it varies from 0.6 g 0.8. D M is the diffusion coefficient in the mobile phase. D M a T / m, where T is the temperature and m is the viscosity of the mobile phase. As u infinite, B/u zero. So, the contribution of longitudinal diffusion in the total plate height is only significant at low MP flow rates. Its contribution is potentially more significant in GC than HPLC because of the relatively high column temperatures and low MP viscosity (gas). Initial band Diffusion of the band with time The initial part of the curve is predominantly due to the B/u term. Because the term B/u in GC is larger than HPLC, the overall H in GC is about 10x the overall H in HPLC. 164

The Stationary-Phase Mass Transfer Term C S u: C S = mass transfer coefficient in the SP. C S a d f 2 / D S where d f is the thickness of the SP film and D S is the diffusion coefficient in the SP. => Thin-film SP and low viscosity SP (large D S ) provide low mass transfer coefficients in the SP and improve column efficiency. The Mobile-Phase Mass Transfer Term C M u: C M = mass transfer coefficient in the MP. C M a d p 2 / D M where d p is the diameter of packing particles and D M is the diffusion coefficient in the MP. The contribution of mass transfer in the SP and MP on the overall late height depends on the flow velocity of the mobile phase. Both phenomena play a predominant role at high MP flow rates. Liquid SP coated on solid support Cartoon with example of SP mass-transfer in a liquid SP: different degrees of penetration of analyte molecules in the liquid layer of SP lead to band-broadening Porous in silica particle Cartoon with example of MP masstransfer: stagnant pools of MP retained in the porous of silica particles lead to bandbroadening. 165

Summary of Methods for Reducing Band Broadening Packed Columns: Most important parameter that affects band broadening is the particle diameter. If the SP is liquid, the thickness of the SP is the most important parameter. Capillary Columns: No packing, so there is no Eddy-diffusion term. Most important parameter that affects band broadening is the diameter of the capillary. Gaseous MP (GC): The rate of longitudinal diffusion can be reduced by lowering the temperature and thus the diffusion coefficient. The effect of temperature is mainly noted at low flow rate velocities where the term B/u is significant. Temperature has little effect on HPLC. Effect of particle diameter in GC Effect of particle diameter in HPLC Note: m = mm 166

Optimization of Column Performance Optimization experiments are aimed at either reducing zone broadening or altering relative migration rates of components. The time it takes for chromatographic analysis is also an important parameter that should be optimized without compromising chromatographic resolution. Column Resolution: It is a quantitative measure of the ability of the column to separate two analytes. It can be obtained from the chromatogram with the equation: In terms of retention factors k A and k B for the two solutes, the selectivity factor and the number of theoretical plates of the column: From the last equation we can obtain the number of theoretical plates needed to achieve a given resolution: 167

For compounds with similar capacity factors, i.e. k A k B : The time it takes to achieve a separation can be predicted with the formula: Where k = k A + k B / 2 168

The General Elution Problem The general elution problem occurs in the separation of mixtures containing compounds with widely different distribution constants. The best solution to the general elution problem is to optimize eluting conditions for each compound during the chromatographic run. In HPLC, this is best accomplished by changing the composition of the mobile phase (Gradient Elution Chromatography). In GC, this is best accomplished by changing the temperature of the column during the chromatographic run. 169

Qualitative and Quantitative Analysis in GC and HPLC Both are done with the help of standards. Qualitative analysis, i.e. compound identification is done via retention time. The retention time from the pure standard is compared to the retention time of the analyte in the sample. Note: retention times are experimental parameters and as such are prone to standard deviation. Quantitative analysis is done via the calibration curve method (or external standard method) or the internal standard method. Calibration curve or external standard method: the procedure is the same as usual. The calibration curve can be built plotting the peak height or the peak area versus standard concentration. The same volume of sample was injected in each case, but Sample B has a much smaller peak. Since the t R at the apex of both peaks is 2.85 minutes, this indicates that they are both the same compound, (in this example, acrylamide (ID)). The Area under the peak ( Peak Area Count ) indicates the concentration of the compound. This area value is calculated by the Computer Data Station. Notice the area under the Sample A peak is much larger. In this example, Sample A has 10 times the area of Sample B. Therefore, Sample A has 10 times the concentration, (10 picograms) as much acrylamide as Sample B, (1 picogram). Note, there is another peak, (not identified), that comes out at 1.8 min. in both samples. Since the area counts for both samples are about the same, it has the same concentration in both samples. 170

Internal Standards An internal standard is a known amount of a compound that is added to the unknown. The signal from analyte is compared to the signal from the standard to find out how much analyte is present. This method compensates for instrumental response that varies slightly from run to run and deteriorates reproducibility considerably. This is the case of mobile phase flow rate variations in chromatographic analysis. Internal standards are also desirable in cases where the possibility of loosing sample during analysis exists. This is the case of sample separation in the chromatographic column. The internal standard should be chosen according to the analyte. Their chemical behavior with regards to the SP and MP should be similar. How to use an internal standard?: a mixture with the same known amount of standard and analyte is prepared to measure the relative response of the detector for the two species. The factor (F) is obtained from the relative response of the detector. Once the relative response of the detector has been found, the analyte concentration is calculated according to the formula: Area of analyte signal Concentration of analyte = F x Area of standard signal Concentration of standard A known amount of standard is added to the unknown X. The relative response is measured to obtain the detector s response factor F. 171

Example of Internal Standards In a preliminary experiment, a solution containing 0.0837M X and 0.066M S gave peak areas of A X = 423 and A S = 347. Note that areas are measured in arbitrary units by the instrument s computer. To analyze the unknown, 10.0mL of 0.146M S were added to 10.0mL of unknown, and the mixture was diluted to 25.0mL in a volumetric flask. This mixture gave a chromatogram with peak areas A X = 553 and A S = 582. Find the concentration of X in the unknown. First use the standard mixture to find the response factor: A X / [x] = F x {A S / [S]} Standard mixture: 423 / 0.0837 = F {347 / 0.0666} => F = 0.9700 In the mixture of unknown plus standard, the concentration of S is: [S] = (0.146M)(10.0mL / 25.0mL) = 0.0584M where: 10.0mL / 25.0mL is the dilution factor Using the known response factor and S concentration of the diluted sample in the equation above: 553 / [X] = 0.9700 (582 / 0.0584) =>[X] = 0.05721M 172

Liquid Chromatography Size exclusion or gel: polystyrene-divinylbenzene Silica with various porous sizes 173

Instrumentation 174

Pumping Systems Pumping systems that allow to change the MP composition during the chromatographic run provide better separation of compounds with wide range of k factors. 175

Detectors UV-VIS absorption cell for HPLC 176