Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Similar documents
A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Study of Residual Solvents in Various Matrices by Static Headspace

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Methanol Extraction of high level soil samples by USEPA Method 8260C

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Analytical Trap Comparison for USEPA Method 8260C

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Volatile Organic Compounds in Every Day Food

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Solid Phase Micro Extraction of Flavor Compounds in Beer

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Optimizing. Abstract: is standardd. procedures. altered to

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Determination of Volatile Organic Compounds in Air

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

USP<467> residual solvents

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Maximizing Sample Throughput In Purge And Trap Analysis

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Optimizing Standard Preparation

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

GC/MS Application Note

Analysis of Fast Moving Consumer Goods (FMCG) using GCMS with static and dynamic headspace (HS)

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

Author. Abstract. Introduction

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

STANDARD OPERATING PROCEDURES

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Detection of Chemical Warfare Agents by Transportable GC/MS

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Solid Phase Micro Extraction of Flavor Compounds in Beer

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Headspace Technology for GC and GC/MS: Features, benefits & applications

USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns

The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note

Ed George and Anaïs Viven Varian, Inc.

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

U.S. EPA Method 8270 for multicomponent analyte determination

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Application Note 032

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection)

Technical Report Abstract: Keywords: RoHS, phthalate esters, Py-GC/MS, Py-Screener 1. Introduction

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

ICBA Guidance Document to Mitigate the Potential for Benzene Formation in Beverages Adopted by the ICBA Council on 29 April 2006

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Maximizing Production While Minimizing Costs

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

AUTOMATED HEADSPACE. Headspace Principles. Graham Broadway. Headspace Gas Chromatography Background to the HS Technique

2017 EPA Method Update Rule and EPA Method 624.1

Transcription:

Application Note Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform ( 염화비닐, 아크릴로니트릴, 브로모포름 ) with Static and Dynamic Headspace GC/MS Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Introduction The Korean Method ES 04602.1b requires either static or dynamic headspace with GC/MS to monitor vinyl chloride, acrylonitrile, and bromoform. The Minimum Quantitation Limit () must be less than 5 ppb. The method requires the calibration curve to have a correlation coefficient greater than 0.98 or a Relative Standard Deviation (RSD) of the Response factors (Rf) less than 25%. The Teledyne Tekmar HT3 Automated Static and Dynamic Headspace Vial Sampler was used to meet these requirements for vinyl chloride, acrylonitrile and bromoform in drinking water by the static and dynamic headspace GC/MS method. Standards 2 ppm Internal Standard (IS) containing Fluorobenzene, chlorobenzene-d5 and 1,4-dichlorobenzene-d4 5 ppm Stock Standard containing acrylonitrile, vinyl chloride and bromoform Calibration Curve and All standards and samples were prepared similarly. Each headspace vial contained 3 g of sodium chloride, 10 ml of deionized water and 10 µl of the 2 ppm IS solution. A calibration curve and seven standards were prepared according to Table I. Table I Calibration Curve Standards and Standard Dilution Standard Level Stock Standard Sample Volume 0 0 µl 10 ml 5 and 10 µl 10 ml 10 20 µl 10 ml 25 50 µl 10 ml Instrument Conditions Table II Static (Loop) HT3 Headspace Instrument Parameters Variable Value Variable Value Constant Heat Time On Mixing Time 5.00 min G.C. Cycle Time 20.00 min Mixing Level Level 5 Valve Oven Temp 150 C Mixer Stabilization Time 0.5 min Transfer Line Temp 150 C Pressurize 11 psig Standby Flow Rate 50 ml/min Pressurize Time 1.00 min Platen/Sample Temp 60 C Pressurize Equil Time 0.20 min Platen Temp Equil Time 0.10 min Loop Fill Pressure 7 psig Sample Equil Time 30.00 min Loop Fill Time 2.00 min Mixer Off Inject Time 1.00 min Sales/Support: 800-874-2004 Main: 513-229-7000

Page 2 Table III Dynamic (Trap) HT3 Headspace Instrument Parameters Variable Value Variable Value Valve Oven Temp 180 C Sweep Flow Rate 75 ml/min Transfer Line Temp 180 C Sweep Flow Time 5.00 min Standby Flow Rate 100 ml/min Dry Purge Time 2.00 min Trap Standby Temp 30 C Dry Purge Flow 50 ml/min Trap Sweep Temp 0 C Dry Purge Temp 25 C Platen/Sample Temp 45 C Desorb Preheat 245 C Sample Preheat Time 0.00 min Desorb Temp 260 C Preheat Mixer On Desorb Time 1.00 min Preheat Mixing Level Level 10 Trap Bake Temp 265 C Preheat Mixing Time 5.00 min Trap Bake Time 5.00 min Preheat Mixer Stabilize Time 0.50 min Trap Bake Flow 150 ml/min Trap K Table IV Agilent 7890B GC with 5977A MS Parameters Variable Column Oven Program Inlet MS Scan/SIM Mode Value Agilent DB-624UI, 20 m, 0.18 mm ID, 1 µm; Constant Flow 0.9 ml/min: Average Velocity 42.02 cm/sec 35 C for 3 min; 13 C/min to 85 C, 25 C/min to 225 C, hold for 0 min Temp 200 C; Helium Carrier Gas; Septum Purge Flow 0.5 ml/min, 1 mm IP Deact. Liner Static Headspace Split Ratio - 30:1 Dynamic Headspace Split Ratio - 100:1 Source Temp 230 C; Quad Temp 150 C ; Solvent Delay 7.00 min; Atune; Transfer Line 225 C ;Scan/SIM Mode; Trace Ion Detection Static - On Dynamic - Off; Gain Factor: Static - 10 Dynamic - 5 Scan - 35.0 m/z to 270.0 m/z, Threshold 10, Sampling Rate 3 SIM Group 1-0.75 min; 62.00 m/z, 64.00 m/z, 200 msec dwell - Vinyl Chloride Group 2-2.20 min; 51.00 m/z, 52.00 m/z, 53.00 m/z, 200 msec dwell - Acrylonitrile Group 3-4.50 min; 62.00 m/z, 64.00 m/z, 200 msec dwell - Fluorobenzene Group 4-7.50 min; 117.00 m/z, 171.00 m/z, 173.00 m/z, 175.00 m/z,200 msec dwell Chlorobenzene-d5, Bromoform Group 5-9.90min; 115.00 m/z, 150.00 m/z, 152.00 m/z, 200 msec dwell - 1,4-Dichlorobenzene-d4

Page 3 Static (Loop) Headspace SIM Mass Spectrometry Results The Selected Ion Monitoring (SIM) chromatograms were evaluated using the Agilent Environmental ChemStation software. Figure 1 is the SIM chromatogram of a 5 ppb sample by the static headspace method. The SIM ions that were used for the Internal Standard calculation are shown in Figure 1 and Table V. The Response factor (Rf) of vinyl chloride and chlorobenzene were calculated versus both the fluorobenzene and the chlorobenzene-d5 Internal Standards. The Rf of bromoform was calculated versus the 1,4-dichlorobenzene-d4 internal standard. The four standards were evaluated for linearity and RSD of the Rf. The concentrations of the seven 5 ppb samples were calculated by both the average Rf value and the linear calibration curve. The was calculated by multiplying the standard deviation of the calculated amount of the seven samples times 10. The %RSD of the Rf and its calculated, and the linear correlation coefficient (r 2 ) and its calculated data are presented in Table V. Figure 1 Static headspace SIM quantitation ion chromatogram of a 5 ppb vinyl chloride, chlorobenzene and bromoform standard with fluorobenzene, chlorobenzene-d5 and 1,4-dichlorobenzene-d4 Internal Standard. Table V %RSD, Linearity and Results with Static SIM GC/MS Rf Calibration Linear Calibration Compound Quant Ion %RSD r 2 Fluorobenzene IS 96 11.8 NA NA NA Vinyl Chloride 62 2.4 2.1 0.9998 2.2 Acrylonitrile 53 2.1 4.9 0.9999 5.0 Chlorobenzene-d5 IS 117 11.8 NA NA NA Vinyl Chloride 62 11.7 2.0 0.9917 1.8 Acrylonitrile 53 9.1 3.5 0.9948 3.2 1,4-Dichlorobenzene-d4 IS 152 9.1 NA NA NA Bromoform 173 4.4 1.2 0.9988 1.3

Page 4 Dynamic (Trap) Headspace Full Scan Mass Spectrometry Results The Total Ion Chromatograms (TIC) were evaluated using the Agilent Environmental ChemStation software. Figure 2 is the TIC of a 5 ppb sample by the dynamic headspace method displaying the quantitation ion (quant ion) for each compound. The quant ions used for calculations are shown in Figure 2 and Table VI. The Response factor (Rf) of vinyl chloride and acrylonitrile were calculated versus both the fluorobenzene and the chlorobenzene-d5 internal standards. The Rf of bromoform was calculated versus the 1,4-dichlorobenzene-d4 internal standard. The four standards were evaluated for linearity and RSD of the Rf. The concentrations of the seven 5 ppb samples were calculated by both the average Rf value and the linear calibration curve. The was calculated by multiplying the standard deviation of the calculated amount of the seven samples times 10. The %RSD of the Rf and its calculated, and the linear correlation coefficient (r 2 ) and its calculated data are presented in Table VI. Figure 2 Dynamic headspace TIC quantitation ion of a 5 ppb vinyl chloride, acrylonitrile and bromoform standard with fluorobenzene, chlorobenzene-d5 and 1,4-dichlorobenzene-d4 Internal Standard. Table VI %RSD, Linearity and Results with Dynamic Full Scan GC/MS Rf Calibration Linear Calibration Compound Quant Ion %RSD r 2 Fluorobenzene IS 96 8.7 NA NA NA Vinyl Chloride 62 0.5 1.8 1.0000 2.0 Acrylonitrile 53 16.0 5.7 0.9992 5.2 Chlorobenzene-d5 IS 117 5.8 NA NA NA Vinyl Chloride 62 3.8 2.4 0.9997 2.5 Acrylonitrile 53 14.9 3.5 0.9998 4.9 1,4-Dichlorobenzene-d4 IS 152 7.7 NA NA NA Bromoform 173 2.8 3.2 0.9993 3.3

Page 5 Dynamic (Trap) Headspace SIM Mass Spectrometry Results The Selected Ion Monitoring (SIM) chromatograms were evaluated using the Agilent Environmental ChemStation software. Figure 3 is the SIM chromatogram of a 5 ppb sample by the dynamic headspace method. The SIM ions that were used for the internal standard calculation are shown in Figure 3 and Table VII. The Response factor (Rf) of vinyl chloride and acrylonitrile were calculated versus both the fluorobenzene and the chlorobenzene-d5 internal standards. The Rf of bromoform was calculated versus the 1,4-dichlorobenzene-d4 internal standard. The four standards were evaluated for linearity and RSD of the Rf. The concentrations of the seven 5 ppb samples were calculated by both the average Rf value and the linear calibration curve. The was calculated by multiplying the standard deviation of the calculated amount of the seven samples times 10. The %RSD of the Rf and its calculated, and the linear correlation coefficient (r 2 ) and its calculated data are presented in Table VII. Figure 3 Dynamic headspace SIM quantitation ion chromatogram of a 5 ppb vinyl chloride, acrylonitrile and bromoform standard with fluorobenzene, chlorobenzene-d5 and 1,4-dichlorobenzene-d4 Internal Standard. Table VII %RSD, Linearity and Results with Dynamic SIM GC/MS Rf Calibration Linear Calibration Compound Quant Ion %RSD r 2 Fluorobenzene IS 96 6.2 NA NA NA Vinyl Chloride 62 3.6 2.1 0.9996 1.8 Acrylonitrile 53 4.7 1.7 0.9998 1.7 Chlorobenzene-d5 IS 117 2.4 NA NA NA Vinyl Chloride 62 2.2 1.7 0.9999 1.6 Acrylonitrile 53 2.3 1.3 0.9997 1.3 1,4-Dichlorobenzene-d4 IS 152 4.1 NA NA NA Bromoform 173 6.5 1.8 0.9992 2.0

Conclusion The HT3 static and dynamic headspace method for the detection of vinyl chloride, acrylonitrile and bromoform surpassed the method requirements for the response factor Relative Standard Deviation, correlation coefficient and as required by the Ministry of Environment. Acrylonitrile detection is greatly improved when determining the concentration using SIM GC/MS analysis with the dynamic capability of the HT3. Page 6 References Korean: ES 04602.1b, 염화비닐, 아크릴로니트릴, 브로모포름 - 드스페이스 / 기체크로마토그래피 - 질량분석법 English: ES 04602.1b, Vinyl Chloride, Acrylonitrile, Bromoform-Headspace/Gas Chromatography/Mass Spectrometry Special thanks to Ju-Hyun, Han of Young-In Scientific Co Ltd of Seoul Korea for translation of the method and the method requirements.