On the use of Deming and quadratic calibrations in analytical chemistry

Similar documents
Quality Control Procedures for Graphite Furnace AAS using Avanta Software

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

VAM Project Development and Harmonisation of Measurement Uncertainty Principles

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer

Examples of Method Validation Studies Conducted in Different Economies

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation.

Direct Determination of Cadmium and Lead in honey by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Test Method: CPSC-CH-E

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

Elemental Scientific. Spike Recovery. Elemental Scientific

Standard Operating Procedures. Determination of Anions and/or Cations Extracted from Nylon Filters by Ion Chromatography (IC)

Comparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Analysis of domestic sludge using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

Technical Procedure for General Laboratory Equipment

Elemental Scientific. Stability. Elemental Scientific

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Measurement Uncertainty, March 2009, F. Cordeiro 1

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

ALLOWAY METHOD OUTLINE

Measurement uncertainty and legal limits in analytical measurements

The stimation of Uncertainty by the Utilization of Validation and Quality Control Data; Lead in Gasoline by AAS

ANALYSIS CERTIFICATE OF

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Elemental Scientific. Spike Recovery. Elemental Scientific

Chapter 4: Verification of compendial methods

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

V. LAB REPORT. PART I. ICP-AES (section IVA)

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.)

United States EPA Method 415.3

DRAFT EAST AFRICAN STANDARD

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Example method validation and MU approach for Cadmium in seafood test method

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Technical Report. Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis

Hydra-C Application Note: 1076

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

Analytical Performance & Method. Validation

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Application Note PB 401. Improve HPLC Sample Preparation in an Analytical Laboratory A New Automated Sample Preparation Process

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

What is measurement uncertainty?

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Analysis of mineral elements in wines using ICP-AES (inductively coupled plasma / atomic emission spectrometry) (Resolution Oeno 478/2013)

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Accreditation of radiochemical analyses, from NAMAS to ISO 17025:2005 and beyond

á233ñ ELEMENTAL IMPURITIES PROCEDURES

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

Analytische Qualitätssicherung Baden-Württemberg

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

110 Lowell Rd Hudson, NH USA Tel:

METHOD 8000A GAS CHROMATOGRAPHY

Annex 2 Formaldehyde

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Traceability is easy. All results are traceable. To what is the issue!

Analytische Qualitätssicherung Baden-Württemberg

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE

2. QUALITY CONTROL IN MASS SPECTROMETRY

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Measurement uncertainty revisited Alternative approaches to uncertainty evaluation

Hexavalent Chromium by Flow Injection Analysis (FIA)

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

Automation of the radiochemical procedures for the sequential separation of radionuclides

SECTION D.2 AMMONIA NITROGEN

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Orthophosphate Ion

Product Descriptions and Offerings

Cadmium and lead content

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

Chemistry 119: Experiment 4. Titrimetric Determination of Acetic Acid Samples And Preparation of Youden Plot

Nordic Innovation Centre Holbergsgate 1, 0166 OSLO Telephone Fax

Transcription:

On the use of Deming and quadratic calibrations in analytical chemistry Henry Foden Rand Water October 09-2013 PO Box 3526, Vereeniging, 1930, Republic of South Africa e-mail: hfoden@randwater.co.za Phone: 016 430 8849 Fax: 016 455 2055

Introduction The objective of this presentation is to introduce the audience to instrumental calibration techniques which are less familiar than the traditional linear calibration techniques normally used in analytical chemistry. These include 1) The Deming regression. 2) The quadratic regression. Purpose The calibration (regression) techniques already mentioned carry advantages associated with their use which are not available from those associated with linear regression methods.

Challenges Modern high volume production laboratories face many challenges. These include technical, quality control (QC), and production (sample load) issues. Consequently, analysts are faced with ever increasing demands for faster analytical turnaround times, increasingly strict quality control measures, lower detection limits, greater numbers of samples for analysis and more analytes per sample than was previously the case.

Given that any laboratory has only a given quantity of resources in terms of available finances, laboratory staff, analytical instrumentation and ancillary equipment; it becomes obvious that methodologies for reducing the overall workload which are acceptable from a technical and QC point of view, become targets for investigation. On the purely technical side of chemical analysis, stricter demands on QC practices by governmental and other statutory bodies occasionally require innovative solutions to satisfy their requirements.

Objectives The purpose of this presentation is to give an overview of two of the potential solutions to the issues mentioned above, both of which use regression methods which differ from the classical linear calibration techniques commonly used in analytical chemistry.

1 - The Deming regression This takes the point of view that unlike a least linear curves approximation to a linear calibration, where all uncertainty is attributed to that which is contained in the ordinate (i.e. the Y axis of a graph normally containing the raw instrument response), that a portion of the uncertainty is also present in the abscissa (i.e. the X axis of the graph normally containing the concentration of standards used for instrument calibration). This relationship has found realistic application in these laboratories for the calibration of microsyringes used for sample and reagent dispensing in discrete analysers and graphite furnace atomic absorption spectrometers for the definitive calibration of microsyringes, such as those having capacities of 100 µl or less.

The calibration of micropipettes Many modern analytical instruments utilise microsyringes for dispensing known volumes of sample as well as for the autodilution of a high level standard to produce calibration standards. Examples include discrete analysers, graphite furnace atomic absorption spectrometers and chromatographic instruments. Using traditional gravimetric methods for the determination of the accuracy of microsyringes is sometimes impractical due to the small volumes of liquid involved as well as the possibility that they are built into the instrument and are not readily removable by an analyst. In addition, the syringes have to normally have to be removed from the instrument and may not have a convenient hypodermic syringe adaptor.

Practical determination - 1 In order to determine the accuracy of the microsyringe of a graphite furnace atomic absorption spectrometer (GFAA) autosampler, the technique outlined below was performed utilising the element selenium as the test analyte. The instrument was a Thermofisher ice 3000 series GFAA. The instrument was calibrated from 0-50 µg/l as normal using the autodiluter to produce the standards from the master standard, which had a concentration of 50 µg/l. A second set of standards of the same concentrations as the calibration standards was then analysed. These standards however were manually prepared and then ran as analytical unknowns. They were prepared from the same stock solution as the autodiluter calibration standard to minimise the number of experimental variables.

Practical determination - 2 The absorbance data for the two sets of calibration data were then analysed using the Deming model II linear regression. As both data sets were obtained via the same method, they were assumed to possess equal uncertainties. The commercially available statistical package GraphPad Prism 4 was used to perform the Deming regression. In table 1 below, the output absorbance of the autopipetted and the manually prepared standards as a function of concentration are shown. The output data from Graphpad was then pasted into EXCEL, and t and F tests performed.

Table 1. Absorption values for manual and autosampler prepared standards Se concentration µg/l Manual standard absorption values Microsyringe standard absorption values 0 0.008 0.009 10 0.143 0.144 20 0.296 0.313 30 0.437 0.453 40 0.59 0.635 50 0.748 0.735

Table 2 - t-test: Two-Sample Assuming Equal Variances Microsyringe Manual Mean 0.381500001 0.370333337 t Stat 0.363699994 P(T<=t) one-tail 0.358233107 t Critical one-tail 1.652508101

Table 3 - F-Test: Microsyringe Manual Mean 0.381500001 0.370333337 F 1.0236831 P(F<=f) one-tail 0.453532459 F Critical one-tail 1.391719552

Statistical summary of results for microsyringe calibration As no statistically significant differences are shown between the results of the t and of the F tests for the two different calibration methods, it is therefore proven that the microsyringe is accurate. It should be noted that the Deming regression allows for the input of the standard deviation of replicate samples, thus also making it a powerful tool for the comparison of different methods for the analysis of the same analyte. In this case, this was unnecessary as the analytical method (and specific instrument) was the same for both sets of data obtained. Both sets of data were obtained from the same analytical run.

2 - The Quadratic calibration It has long been the practice in analytical chemistry and chemometrics courses for instrumental methods of analysis, to present topics related to the determination of the linear range of an analytical method together with the required statistical background. Such topics would naturally include correlation methods e.g. the method of least linear squares, and, in more advanced courses, the methods attached to the determination of the uncertainty of regression, and suitable analytical spacing of the calibration standards. Although seldom explicitly stated as such, these didactic methods leave an implicit assumption unstated, namely that other methods of calibration are unacceptable.

The Quadratic calibration Whilst objections have been raised to the practice of quadratic calibration, it has been conclusively shown in these laboratories that when used in conjunction with well-defined ISO practices and statistical procedures, and despite algebraic difficulties which may arise during the course of the determination of more difficult analytical parameters e.g. the uncertainty of calibration regression of the method, that these may be easily overcome on a routine basis.

Main references used in this section of the presentation South African National Standard SANS 8466-2:2002, Water quality- Calibration and evaluation of analytical methods and estimation of performance characteristics, Part 1: Statistical evaluation of the linear function. pp 5. South African National Standard SANS 8466-2:2002, Water quality- Calibration and evaluation of analytical methods and estimation of performance characteristics, Part 2: Calibration strategy for non linear second-order calibration functions. pp 7.

Quadratic regressions - 1 The calibration of an instrument may be viewed as the analysis of a series of standards of known concentration to obtain an instrument output for each of the standards. The derivation of a defined mathematical formula relating the magnitude of the instrument output to the standard concentration, normally utilising a mathematical regression technique, results in an equation by which the instrument output for unknown samples can then be utilised to calculate the analyte concentration. Given that a linear calibration methodology has been the traditional and mathematically simpler approach, it is reasonable to examine the underlying reasons for the selection of a quadratic approach.

Quadratic regressions - 2 Accredited laboratories which are primarily involved in high sample throughput routine analytical work, with tight deadlines on turnaround time require analytical procedures which are as simple and fast as possible, whilst bearing in mind the technical and quality control requirements of the specific analysis. As a natural corollary of this requirement, analytical procedures with simple sample preparation procedures would naturally be preferred over those with more complex preparation procedures.

Quadratic regressions - 3 In some cases where the linear dynamic range of a method is high, covering many orders of magnitude, e.g. ICP-OES or ICP-MS, quadratic calibrations may be unnecessary, however other methods e.g. those which are colorimetrically or atomic absorption based which have relatively low linear dynamic ranges can benefit greatly from the use of quadratic calibrations.

Quadratic regressions - 3 In some cases where the linear dynamic range of a method is high, covering many orders of magnitude, e.g. ICP-OES or ICP-MS, quadratic calibrations may be unnecessary, however other methods e.g. those which are colorimetrically or atomic absorption based which have relatively low linear dynamic ranges can benefit greatly from the use of quadratic calibrations.

Quadratic regressions - 3 In some cases where the linear dynamic range of a method is high, covering many orders of magnitude, e.g. ICP-OES or ICP-MS, quadratic calibrations may be unnecessary, however other methods e.g. those which are colorimetrically or atomic absorption based which have relatively low linear dynamic ranges can benefit greatly from the use of quadratic calibrations.

Quadratic regressions - 3 In some cases where the linear dynamic range of a method is high, covering many orders of magnitude, e.g. ICP-OES or ICP-MS, quadratic calibrations may be unnecessary, however other methods e.g. those which are colorimetrically or atomic absorption based which have relatively low linear dynamic ranges can benefit greatly from the use of quadratic calibrations.

Fundamentals of linear a quadratic equations These are shown below in equations one and two Equation 1 Linear equation Equation 2 Quadratic equation By way of example, figuires 1 and 2 show the linear and quadratic calibrations for the analysis of potassium in sewage samples by flame atomic absorption utilising a high range calibration methodology. Table 5 contains the raw data upon which the two calibrations are based.

Table 4 Raw data for Potassium calibrations Calibration 1 Nominal concentration mg/l Absorbance K Blank 0 0.0027 Standard 1 20 0.3116 Standard 2 40 0.5847 Standard 3 60 0.8073 Standard 4 80 0.9694 Standard 5 100 1.0826

Linear calibration for high range K analysis

Quadratic calibration for high range K analysis

Table 5 Raw data for Potassium calibrations If we sequentially remove the calibration points from the high concentration end of the calibration curve, we obtain the information in table 6. Calibration points Top calibration standard mg/l R Squared R 1 2 3 4 5 6 N/A N/A 40 60 80 100 N/A N/A N/A N/A 0.9987 0.9994 0.9948 0.9973 0.9858 0.9929 0.9715 0.9856

As can be readily seen in the two previous graphs of the linear and quadratic approaches to the regression analysis presented in the data, the quadratic approach leads to substantially higher correlation coefficients than the linear approach.

Accuracy of quadratic calibrations Possibly the most critical question that can be asked of any calibration technique is how accurate is it? In order to answer this question, test samples spreading across the analytical range consisting of a certified reference standard (CRM), a spiked sewerage sample, together with mid and high range QC verification standards were analysed.

Accuracy of quadratic calibrations Table 6 Analysis of CRM RTC QCI-046 Lot 015740/015741 PARAMETER Mean analysed value of CRMs mg/l 8.316 Accuracy of CRMs. % 102.80 Bias of CRMs. % +2.80 Manufacturers certified value mg/l 8.09 Manufacturers specified standard deviation mg/l 0.572 Calculated Z Score 0.395 Manufacturers acceptance limits mg/l 6.37 9.80 K

Accuracy of quadratic calibrations To establish the accuracy at higher analytical levels (± 30 mg/l), samples of sewerage were analysed with and without spike addition. The spiking was performed as follows; A 100 ml volumetric flask was rinsed out with sample and then approximately half filled with the sample. 2 ml of 1000 mg/l K standard was added with a piston pipette. The flasks were then made up to the mark with the sample, mixed and used for spike recovery tests. The data obtained is in table eight below.

Accuracy of quadratic calibrations Table seven CONCENTRATION OF UNSPIKED SAMPLE mg/l 10.591 MEAN VALUE OF SPIKED SAMPLE mg/l 31.210 SIGMA 0.607 R.S.D. % 1.944 MAXIMUM mg/l 32.032 MINIMUM mg/l 30.211 SPIKE RECOVERY % 102.73

Table eight accuracy of a 50 mg/l K standard Accuracy of Q.C. verifications % 102.231 R.S.D. OF Q.C. verifications % 1.552 Bias of Q.C. verifications % + 2.231

Table nine accuracy of an 80 mg/l K standard standard MEAN mg/l 78.612 SIGMA 2.237 R.S.D. % 2.846 MAXIMUM mg/l 81.033 MINIMUM mg/l 74.731 Accuracy % 98.27

Potential pitfalls to avoid in quadratic calibration

Potential pitfalls to avoid in quadratic calibration Samples with concentrations higher than the analytical rollover point will be incorrectly interpreted by the instrument software as have a lower value than is actually the case, therefore caution has to be observed using the quadratic method, and the samples analysed should preferably have been historically well characterised to ensure that they fall within the method calibration range. In cases of doubt, it is preferable to dilute the sample prior to analysis.

CONCLUSIONS DEMING REGRESSIONS The Deming technique is both a mathematically and analytically elegant method for the determination of the accuracy of microsyringes for instrumental analysis as it is used without removal of the syringe from its position in the instrument. As all test samples are analysed with the syringe in position, leading to analytically realistic results as the output obtained from the instrument are directly used as the input for statistical analysis. Quadratic calibration, although somewhat more complex for the determination of uncertainty budgets, where the uncertainty of calibration regression is required, are still relatively easy to perform, and result in the extension of analytical range. This leads to material savings in terms of resource requirements due to the smaller amount of dilutions

CONCLUSIONS QUADRATIC REGRESSIONS Quadratic calibration, although somewhat more complex for the determination of uncertainty budgets, where the uncertainty of calibration regression is required, are still relatively easy to perform, and result in the extension of analytical range. This leads to material savings in terms of resource requirements due to the smaller amount of dilutions required when samples which exceed the calibration range of a linearly calibrated method are used.

THANK YOU Are their any questions?