RESPONSE OF A NAI(TL) SCINTILLATION DETECTOR IN A WIDE TEMPERATURE INTERVAL

Similar documents
Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

Sample Spectroscopy System Hardware

1 of :32

ɣ-radiochromatography

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

Detection and measurement of gamma-radiation by gammaspectroscopy

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

Mass of the electron m 0

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

New Scintillators for the Border Monitoring Equipment

Survey of the performance of scintillation materials at low temperatures

ORTEC AN34 Experiment 10 Compton Scattering

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

Gamma and X-Ray Detection

THE COMPTON EFFECT Last Revised: January 5, 2007

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

AIR PRESSURE MEASUREMENT USING SILICON BASED ANALOG SENSOR SPD015A

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Slides by: Prof. Abeer Alharbi

LaBr 3 (Ce) detector and PMT tests at IPN Orsay

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

Gamma Spectroscopy. References: Objectives:

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

K 40 activity and Detector Efficiency

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

The Best Gamma-Ray Detector

Chemical Engineering 412

He-3 Neutron Detectors

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

"Neutron Flux Distribution"

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Gamma ray coincidence and angular correlation

The Compton Effect. Martha Buckley MIT Department of Physics, Cambridge, MA (Dated: November 26, 2002)

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

The Mössbauer Effect

AGATA preamplifier performance on large signals from a 241 Am+Be source. F. Zocca, A. Pullia, D. Bazzacco, G. Pascovici

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts

Wiper Gold Edition. Premium Wipe Test Counter. Setting the bar higher

Quality Assurance. Purity control. Polycrystalline Ingots

Figure 1. Decay Scheme for 60Co

Determination of the shielding power of different materials against gamma radiation

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Pulse height non-linearity in LaBr 3 :Ce crystal for gamma ray spectrometry and imaging

Analysis of γ spectrum

*Corresponding author,

Wiper Gold Edition. Premium Wipe Test Counter. Setting the bar higher

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER

Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals

Study of Isospin simmetry using the PARIS detector. Alice Mentana

arxiv: v3 [nucl-ex] 12 Jan 2012

Prognostication of Ozone Concentration in the Air

High Purity Germanium Detector Calibration at ISOLDE

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Natural Radiation K 40

Characterization and Monte Carlo simulations for a CLYC detector

arxiv: v2 [physics.ins-det] 8 Feb 2013

arxiv: v1 [physics.ins-det] 3 Feb 2011

ß + RADIATION. Choose your application of. chromatography controlled. chemical purity TLC PET GABI. PET GABI 2x2 Nal (Tl) 0 1V output.

Chapter Seven (Nuclear Detectors)

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

Radioactivity. Lecture 6 Detectors and Instrumentation

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

Measurements of Neutron Total and Capture Cross Sections at the TOF spectrometers of the Moscow Meson Factory

THE spectroscopic performance of large volume CdZnTe

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

Compton suppression spectrometry

Application of positrons in materials research

Radionuclide Imaging MII Detection of Nuclear Emission

Absolute activity measurement

BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING. P TheodÛrsson 1 S Ingvarsdottir G I Gudjonsson

CapeSym, Inc. Radiation Detection Solutions

Development of a 3D-Imaging Calorimeter in LaBr 3 for Gamma-Ray Space Astronomy

Fast detectors for Mössbauer spectroscopy )

DETECTORS. I. Charged Particle Detectors

Radiation Detection and Measurement

Gamma Ray Spectroscopy

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

CCD readout of GEM-based neutron detectors

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

K 40 activity and Detector Efficiency

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09

QUANTITATIVE Cs-137 DISTRIBUTIONS FROM AIRBORNE GAMMA RAY DATA

Measurement of Shielding Effectiveness of Building Blocks against 662 KeV Photons

DELAYED COINCIDENCE METHOD FOR PICOSECOND LIFETIME MEASUREMENTS

COMPTON SCATTERING OF GAMMA RAYS

Dependence Of Gamma Ray Attenuation On Concentration Of Manganese (II) Chloride Solution

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting

Positron-Electron Annihilation

Preparatory experiments for cold-neutron induced fission studies at IKI

Efficiency. Calculations for Selected Scintillators. > Detector Counting. Efficiency. > Photopeak Efficiency of Various Scintillation Materials

Energy Resolution of Scintillation Detectors New Observations

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Transcription:

RESPONSE OF A NAI(TL) SCINTILLATION DETECTOR IN A WIDE TEMPERATURE INTERVAL Ludmil Todorov Tsankov 1, Mityo Georgiev Mitev 2 1 Department of Nuclear Engineering, University of Sofia, 5, J.Bourchier Str., 1164-Sofia, Bulgaria, e-mail: ludmil@phys.uni-sofia.bg 2 Department of Electronic Technique, Technical University - Sofia, 8, Kliment Ohridski Str., 1000-Sofia, Bulgaria, e-mail: mitev@ecad.tu-sofia.bg The temperature dependence of a NaI(Tl) scintillation detector's response has been studied in the temperature interval (-30 0 C to +50 0 C). Besides the common temperature drift of the light output and that of the decay time of the scintillator, some other very-long-term factors have been observed which give rise to a remarkable hysteresis of the output signal drift. They are most probably due to the PMT temperature instability. A post-processing method is proposed to compensate their effect and the results of its application are discussed. 1. INTRODUCTION Keywords: Scintillation detectors, long-term behaviour Scintillation detectors are amongst the most widely used instruments for measurement of the gamma-radiation. They have very good efficiency, reasonable energy resolution and moderate price. A significant drawback of the scintillation spectrometers is their relatively large gain drift which smears the observed photopeaks. It is mainly due to the following factors: the temperature dependence of the light output of the scintillator; the temperature dependence of the decay constant of the scintillations in NaI(Tl); the photomultiplier's temperature drift; the temperature behaviour of the detector front-end electronics; the ageing of the photomultipliers. Fortunately, in most applications scintillation spectrometers are used either in laboratory conditions where the temperature changes are within 1-2 0 C, or for short time measurements in the environment. So, in those situations the temperature effects are not so serious. However, when used for continuous monitoring of the environmental radiation, the whole spectrometric system has to work long time in open air conditions (large temperature range, humidity etc.). Then the long-term characteristics of both the scintillator and the photomultiplier tube (PMT) are of crucial importance for the system performance. 138

The goal of the present work is a study of the long-term performance of a NaI(Tl) scintillation spectrometer in a temperature interval covering the whole operation range (-30 0 C +50 0 C) for outdoor radiation monitoring. A post-processing method is also suggested which is used to compensate the effects of gain instability. 2. EXPERIMENTAL A cylindrical 45x40mm NaI(Tl) scintillator coupled to a XP2000 PMT is used. The scintillation detector and the front-end electronics (3.5μs RC-shaping chain and a buffering preamplifier) are put inside a refrigerator camera. The temperature inside the camera is controlled automatically within ±0.3 0 C in the range (-29 0 C+55 0 C) following a preset temperature profile. Weak radioactive sources are used to produce gamma-rays ( 241 Am - 26keV, 60keV; 137 Cs - 662keV; 60 Co - 1173keV, 1332keV); the total count rate is about 4000 cps. In order to retain all modules inside the camera at the same temperature, the temperature change speed is kept rather slow (1-3 0 C/h). As a consequence, the duration of one cycle is typically one week. The spectrometer behaviour is monitored by recording 2048 channel spectra every 300s. The photopeak centroids of the gamma-lines are computed by a least-squares analysis of individual spectra. 3. EXPERIMENTAL RESULTS Fig.1 shows a typical change in the 1332keV 60 Co peak position during a temperature cycle: +20 0 C (+1 0 C/h) +55 0 C (-1 0 C/h) -25 0 C (+1 0 C/h) +20 0 C. Besides the typical temperature drift, a deep hysteresis is observed that makes the system behaviour ambigious. Multiple loops make the picture still more complex as it can be seen from Fig.2. Temperature Loop 1900 1332 kev Peak Position 1850 1800 1750 1700 1650 1600-30 -20-10 0 10 20 30 40 50 60 Temperature [deg.c] Fig.1. 139

Multiple Temp. Loops 1332keV Peak Position 1950 1900 1850 1800 1750 1700 1650 1600 1550 1500-30 -20-10 0 10 20 30 40 50 60 Temperature [deg.c] Fig.2. The observed hysteresis effect is definitly related to a gain drift of the spectrometer rather than to a spectrum shift as a whole. To establish that fact we have applied a linear energy calibration procedure of the type: Pi = a + b* Ei, i = 1, n (1) where Pi is the position of the peak centroid at energy Ei, to all recorded spectra during the cycle, and then the coefficients a and b have been plotted vs temperature. Fig.3 shows the results corresponding to the temperature loop of Fig.1. P=a+b*E intercept (a) 14 1.40 13 1.35 12 1.30 11 1.25 10 1.20 9 1.15 8 1.10-40 -30-20 -10 0 10 20 30 40 50 60 Temperature [C] Fig.3. slope (b) a b 140

It is seen that the hysteresis in the intercept ( a ) does not exceed 1 channel, while the gain hysteresis (b ) is about 0.05chn/keV (i.e. 50 channels for 1000keV which is comparible with the gamma line width at that energy). Additional experiments at a constant temperature preceeded by a temperature stress (temperature cycle: +50 0 C (-3 0 C/h) +20 0 (T=const)) (Fig.4) have shown that the observed hysteresis is due to a very long-term factor. The experimental data can be fitted as a sum of two exponents: P = b + a1. exp( ln2. t / τ1) + a2.exp( ln2. t / τ 2) (2) with the following values of the parameters: parameter value σ b 1985 0.6 a 1 28 0.4 τ 1 116 5 a 2 12.3 0.3 τ 10.3 0.3 2 1332 kev Peak Position 2030 2025 2020 2015 2010 2005 2000 1995 1990 t=20 deg.c Exper.data Fit 0 24 48 72 96 120 144 168 Time [h] 4. STABILIZATION METHOD Fig.4 Summarizing the experimental results, we have observed two effects: strong temperature drift, mainly of the gain, and a long term hysteresis of the gain. The second phenomenon excludes the possibility to use the current temperature value for gain stabilization of the spectrometer, since the present gain status depends also on its temperature history a couple of days ago. Obviously, an active feedback stabilization system has to be developed in order to compensate influence of the the observed effects on the spectrum quality. 141

At this stage of the investigation, we have decided to use a phenomenological post-processing method [1]. It is based on a least-squares comparison of successive spectral records with a reference spectrum (say, the first record) when the x-scale parameters (in the simplest linear approximation these are the zero-energy channel and the channel/energy ratio) are varied until a maximal consistency is achieved. Then the channel contents of the examined spectrum is redistributed (rebinned) according to the new values of the x-scale parameters. Spline technique is used to ensure continuity of the rebinning procedure. The method allows any analytical parametrization of the spectrum x-scale to be used; in most cases a linear or quadratic 'energy vs channel' function is relevant. Fig.5 shows the 137 Cs peak position from the Fig.1 cycle before and after the application of the x-scale adjustment algorithm. The peak position variation is now diminished 5 times and falls well within the instrumental linewidth. 662-keV peak position 1000 950 60 40 900 20 850 0 800-20 750-40 05.01 06.01 07.01 08.01 09.01 10.01 11.01 12.01 Time original rebinned Temp[C] Fig.5 5. CONCLUSION The total instability of a NaI(Tl) scintillation spectrometer is investigated in the whole temperature interval of open air environment operation. The well known temperature dependent factors (the light output and the decay time of the scintillator) determine the main profile of the spectrometer's temperature drift. Besides them, some very-long-term drift factors have been observed which give rise to a remarkable hysteresis of the output amplitude drift. Since we are interested of the response of the spectrometer as a whole, we have not yet performed a special survey to identify their origin. It seems that they arise from the PMT temperature instability. The important consequence is the fact that it is impossible to stabilize spectrometer gain on the basis of its current temperature only. At this stage, a post-processing method is proposed to 142

compensate the gain drift. Its application results in diminishing the drift 5 times. A detailed study of the system stabilization by a deep feedback is forthcoming. 6. ACKNOWLEDGEMENTS This work is supported by the Bulgarian National Science Fund under contract ВУФ-10/05. 7. REFERENCES [1] L.T.Tsankov, A Method for Comparison of Discrete Spectra, J.Phys.III France (1993) Vol.3, pp.1525-1530. 143