Three-Fold Increase in Productivity for Pesticide Residue Analysis in Baby Food Using Fast Triple Quadrupole GC-MS/MS

Similar documents
High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Routine-grade robustness with the Advanced Electron Ionization source: confirmation of contaminants in food products at trace levels by GC-MS/MS

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

QuEChERS Pesticide Analysis for Fresh Produce*

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Validation of GC-MS/MS for Detection and Confirmation of Low-Level Dioxins

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Technical Report. Residual Pesticides Analysis of Botanical Ingredients Using Gas Chromatography Triple Quadrupole Mass Spectrometry.

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Multi-residue analysis of pesticides by GC-HRMS

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Multi-residue Method for Agricultural Chemicals by GC-MS (Animal and Fishery Products)

Implementing GC-HRAM MS for More Efficient and Effective Routine Pesticide Residues Analysis

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Highly Sensitive and Rugged GC/MS/MS Tool

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Analysis of Pesticides and Contaminants using HS-SPME-GC/MS and DRS Detection. Blair Surridge, Tyler Dyer, Rene Bilideau, Paloma Bertolin Presoti

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Rugged GC/MS/MS Pesticide Residue Analysis Fulfilling the USDA Pesticide Data Program (PDP) Requirements

Fast GC-MS/MS for High Throughput Pesticides Analysis

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

of mass spectrometry

Simultaneous determination of 74 kinds of pesticides in Chinese herbal medicine listed in Chinese pharmacopoeia (2015) by GPC-GC-MS/MS

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry

GC/Q-TOF MS Surveillance of Pesticides in Food

Automated and accurate component detection using reference mass spectra

High Throughput Multi-Residue Screening of Drinking Water using the SPE-DEX~ and Pegasus GC-TOF-MS

[AMI] Multi-residue method, by LC-MSMS, for the quantification of amitraz and its metabolites in fruit, vegetables, cereals and cereal products

Expect more productivity

Thermo Scientific TSQ 8000 Evo Triple Quadrupole GC-MS/MS. Expect more productivity. with the evolution of ultimate performance GC-MS/MS

Safeguarding Food from Pesticides by UHPLC After Extraction with the QuEChERS Method

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Screening for Hundreds of Pesticide Residues Using a GC/Q-TOF with an Exact Mass Pesticide Database in Food

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

The key to your laboratory s success

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Analysis of pesticide residues using various GC/MS systems. Radim Štěpán Czech Agriculture and Food Inspection Authority Prague, Czech Republic

Certificate of pesticide multi-residue analysis by MSMS detection

Date ACCURATE-MASS DATABASE FOR SCREENING OF PESTICIDE RESIDUES IN FRUITS AND VEGETABLES BY GAS CHROMATOGRAPHY TIME-OF-FLIGHT MASS SPECTROMETRY

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Introduction to Fourier Transform Infrared Spectroscopy

Improved GC/MS Analysis of Tomato Pesticides with Agilent Deactivated Silica Tubing

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Validation Report 18

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Key Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia

HILIC Method Development in a Few Simple Steps

Yokogawa Analytical Systems Inc., Nakacho, Musashino-shi, Tokyo , Japan.

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Trim Time Not Columns Multiresidue Pesticides Analysis with an Intuvo GC/TQ

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Fast, Effective XPS Point Analysis of Metal Components

Reduce Cost of Pesticide Residue Analysis

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Multiresidue Analysis of 100 Pesticides in Food Samples by LC/Triple Quadrupole Mass Spectrometry Application

Validation Report 8 Determination of pesticide residues in rice and wheat by GC-MS/MS and LC-MS/MS (QuEChERS method) Appendix 2

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Analysis of Alkylphenols Using GC-MS/MS and Automated SRM Development

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

Transcription:

Three-Fold Increase in Productivity for Pesticide Residue Analysis in Baby Food Using Fast Triple Quadrupole GC-MS/MS Cristian Cojocariu,1 Michael T. Hetmanski,2 Paul Silcock,1 and Richard J. Fussell2 1Thermo Fisher Scientific, Runcorn, UK 2Food and Environment Research Agency (FERA), York, UK Key Words Pesticide Analysis, Baby Food, GC-MS/MS, TraceFinder, Food Safety Goal To assess the performance and productivity of the Thermo Scientific TSQ 8000 Evo GC-MS/MS for pesticide residues analysis. Introduction Pesticides include more than 1000 different substances used to control or eradicate pests. Strict regulatory controls are in place to ensure that these chemicals are used safely and effectively without harmful effects to humans, wildlife, and the environment. Maximum residue levels (MRLs) of pesticides in food and feed have been set by many international bodies including the EU. 1 Detection, quantification, and correct identification of pesticide residues at trace levels requires sensitive, selective, and robust analytical instrumentation. With ever-increasing pressure to analyze a greater number of samples of perishable commodities with shorter turnaround times, high throughput laboratories seek continuous improvements in analytical productivity. In recent times, substantial productivity gains have been achieved using the QuEChERS (quick, easy, cheap, effective, robust and safe) sample extraction approach in combination with gas or liquid chromatography (GC or LC) mass spectrometry (MS). Here, we report the possibility of further productivity gains using advanced, rapid GC-MS/MS technology in combination with new software developments to reduce the time needed to acquire and process the data. Acetonitrile is commonly used as the extraction solvent for QuEChERS. Direct analysis of pesticide residues in acetonitrile is preferred to avoid the need for solvent exchange, which is time consuming and, hence, costly. However, the polar nature of acetonitrile results in poor focusing of chromatographic peaks and the high expansion coefficient limits the injection volume that can be used. In this study, a fast, easy, and robust workflow was used to analyze pesticide residues in baby food. Accurate and sensitive detection, quantification, and identification of pesticides in baby foods is of particular importance because babies are more vulnerable to adverse health effects from these chemicals. This work shows that laboratory productivity can be accelerated by direct injection of low sample volumes of QuEChERS acetonitrile extracts, in combination with fast temperature ramps to shorten GC run times. This is made possible using the innovative EvoCell collision chamber technology combined with the efficient selected reaction monitoring (SRM) scheduling of timed-srm software in the TSQ 8000 Evo triple quadrupole GC-MS/MS. A thorough assessment of the robustness of this fast GC analysis using acetonitrile was conducted following the SANCO guidelines. 2 Always whats next.

2 Instrument and Method Setup A TSQ 8000 Evo triple quadrupole GC-MS/MS instrument coupled with a Thermo Scientific TRACE 1310 GC was used. Sample introduction was performed a Thermo Scientific TriPlus RSH autosampler, and chromatographic separation using a Thermo Scientific TraceGOLD TG-5SilMS 15 m 0.25 mm I.D. 0.25 µm film capillary column (P/N: 26096-1300). Additional details of instrument parameters are displayed in tables below. GC and Injector Conditions TRACE 1310 GC Injection Volume (μl): 1.0 Liner: SSL single taper (P/N: 453A2342) Inlet ( C): 240 Inlet Module and Mode: Splitless Carrier Gas, (ml/min): He, 1.2 Oven Temperature Program Temperature 1 ( C): 60 Hold Time (min): 1 Temperature 2 ( C): 180 Rate ( C/min) 50 Temperature 3 ( C): 320 Rate ( C/min) 35 Hold Time (min): 4 Mass Spectrometer Conditions TSQ 8000 Evo Mass Spectrometer Transfer Line ( C): 280 Ionization Type: EI Ion Source ( C): 320 Electron Energy (ev): 70 Acquisition Mode: t-srm Q2 Gas Pressure(argon)(psi): 60 Q1 Peak Width (Da): 0.7 Q3 Peak Width (Da): 0.7 The TSQ 8000 Evo triple quadrupole mass spectrometer was operated in MS/MS mode using electron ionization (EI+). For each pesticide, two SRM transitions were chosen one for quantification and one for identification purposes. A total of 264 SRM transitions were acquired with dwell times varying from 1 ms to 52 ms, depending on the number of SRM transitions monitored simultaneously. Chromatographic data was acquired data using timed-selected reaction monitoring (t-srm) with a minimum of 12 points/peak. Sample Preparation Baby food samples were extracted using the citrate buffered QuEChERS protocol. The homogenized sample was extracted (10 g) with acetonitrile (10 ml) followed by the addition of MgSO4 (4 g), NaCl (1.0 g), disodium hydrogen citrate sesquihydrate (0.5 g), and trisodium citrate dihydrate (1.0 g). Dispersive solid phase extraction [MgSO4 (150 mg), C18 (50 mg), PSA (50 mg) and carbon (7.5 mg) per ml of extract] was used for sample clean-up. Final extracts (1 g/ml in acetonitrile) were spiked with a mixture of 132 pesticides at concentrations corresponding to 0.5 100 ng/g (ppb) and 1.0 200 ng/g (ppb) for some analytes. Data Processing Data were acquired and processed using the Thermo Scientific TraceFinder version 3.2 software, a single software package that integrates instrument control, method development functionality, and quantitationfocused workflows. For each compound, one SRM transition was used for quantitation and the second one for positive identification of the pesticide. Results and Discussion This study describes the methodology used for multiresidue pesticides analysis in baby food using fast GC for increasing laboratory productivity. The results described below were obtained with acetonitrile as the final extract solvent from the QuEChERS extraction and low-volume, hot splitless injection. The performance of the TSQ 8000 Evo GC-MS/MS system was evaluated by assessing the chromatography, sensitivity, linearity, and reproducibility of the target pesticides analyzed in the extracts of baby food samples.

3 Three-Fold Increase in Sample Throughput Typically, a GC analysis of 132 target pesticides has a run time of around 42 minutes in order to obtain a sufficient number of scans per chromatographic peak (Figure 1), especially in time windows containing many co-eluting peaks. At least 10 12 scans across a chromatographic peak are needed in order to accurately integrate the peaks of interest. Previously, fast scan speeds compromised instrument sensitivity, especially when several SRM transitions were monitored simultaneously. Using the fast GC conditions described above, the GC run time was decreased to 6.32 100 ~11 min with no compromise in the number of data points acquired for each chromatographic peak (Figures 2 and 3). This advance is possible because the fast EvoCell technology allows fast clearance of ions from the collision cell and hence faster data acquisition, without adversely affecting instrument sensitivity. Fast data acquisition enables more information to be collected in a shorter time, ultimately resulting in faster GC runs. Using this fast methodology, sample productivity is improved by approximately three-fold, as around three times as many injections of sample/standard extracts can be carried out in an overnight sequence. 6.59 Full Scan @ 100 200 ng/g TG-5 SILMS, 30 m 0.25 mm 0.25 µm GC Run Time: 37 min Relative Abundance 7.00 11.13 10.12 11.99 25.20 9.50 14.96 23.13 29.61 17.38 7.31 8.89 5.77 12.20 15.57 15.84 19.13 8.20 12.87 17.95 26.47 13.77 23.58 20.62 25.36 28.71 5.09 21.70 19.25 23.83 24.12 27.77 5 35.78 34.81 32.92 34.51 11 31.00 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 Minutes Figure 1. Total ion chromatogram (TIC, full scan) for a typical GC-MS chromatographic run of 132 pesticides at 100 200 ng/g with a total run time of approximately 40 minutes. The first (dichlorvos, RT = 5.77 min) and the last (deltamethrin, RT = 29.61 min) eluting pesticides are highlighted. 8.00 100 SRM @ 100 200 ng/g TG-5 SILMS, 15 m 0.25 mm 0.25 µm GC Run Time: 11 min Relative Abundance 4.69 6.29 7.05 7.21 6.52 6.60 7.63 7.57 7.41 7.65 7.81 8.11 8.23 8.49 0 4.73 6.35 6.78 6.80 8.34 8.69 6.90 7.83 8.74 6.12 6.16 5.07 5.61 5.72 6.00 7.86 8.91 5.08 8.58 4.33 4.85 5.21 8.95 4.61 5.57 4.99 4.48 5.49 5.53 4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8 6.0 6.2 6.4 6.6 6.8 7.0 7.2 7.4 7.6 7.8 8.0 8.2 8.4 8.6 8.8 9.0 9.2 Minutes Figure 2. SRM chromatogram for a fast GC-MS chromatographic run of 132 pesticides at 100 200 ng/g with a total run time of 11 minutes. The first (dichlorvos, RT = 4.33 min) and the last (deltamethrin, RT = 9.15 min) eluting pesticides are highlighted. 9.15

4 100 Relative Abundance 10 100 6.88 6.89 6.88 6.89 6.88 6.89 6.87 6.89 6.90 6.87 6.90 6.87 6.90 6.90 6.88 6.89 6.88 6.89 Sensitivity Almost all pesticides (97%) were detected at a concentration of 0.5 or 1.0 ng/g (ng/ml) and calibration curves were linear over the range 0. 5 100 ng/g (or 1.0 200 ng/g). Examples of chromatography at this low concentration and calibration curves are shown in Figure 4. At the lowest calibration concentration of 5 10 ng/g (0.5 1 default MRL), all compounds were comfortably detected with all the ion ratios for compound identification within 15% of the average ion ratio values derived from the calibration curve across all concentrations. 10 6.87 6.88 6.89 6.87 6.89 6.90 6.90 6.87 6.88 6.89 6.90 6.91 Minutes Figure 3. SRM chromatogram for parathion ethyl eluting at RT = 6.89 min showing 13 scans/peak (peak width 1.8 sec, dwell time of 1.7 ms). Estimation of Instrument Detection Limit (IDL) and Peak Area Repeatability The IDL of the target pesticides was determined empirically by repeatedly injecting (n=20) the 5 ng/g (and 10 ng/g) matrix-matched standard and taking into account the Student s-t critical values for the corresponding degrees of freedom (99% confidence). Figure 4. Examples of chromatography (0.5 pg on column) and linearity (no internal standard correction) for trifluralin, pendimethalin, and folpet.

5 The results of this experiment showed an average %RSD for the peak area reproducibility of 7.3 % and IDL values varying from 0.2 ng/g for dimethamid to 3.7 ng/g for captan (Table 3). By using internal standard correction to compensate for the injection errors both %RSD for peak area repeatability values can be improved even further. Table 1. Peak area reproducibility (% RSD, n=20) at 5 or 10 pg absolute amount on column and calculated instrument IDL99 (in ng/g). No Compound RT (min) pg on Column % RSD IDL 1 Acetochlor 6.53 5 4.5 0.6 2 Aclonifen 7.64 10 8.3 2.1 3 Aldrin 6.90 5 5.9 0.7 4 Azinphos-ethyl 8.34 10 10.0 2.5 5 Benalaxyl 7.74 5 4.4 0.6 6 BHC, Alpha 5.96 5 5.3 0.7 7 BHC, Beta 6.11 5 8.7 1.1 8 BHC, gamma 6.18 5 6.6 0.8 9 Bifenox 8.09 10 10.0 2.6 10 Bifenthrin 8.00 5 4.3 0.5 11 Biphenyl 4.85 5 10.0 1.3 12 Bromophos-ethyl 7.21 5 6.8 0.9 13 Bromopropylate 8.04 5 5.7 0.7 14 Bupirimate 7.43 5 3.0 0.4 15 Buprofezin 7.45 5 4.9 0.6 16 Cadusafos 5.88 5 5.1 0.6 17 Captan 7.16 10 14.0 3.7 18 Carbetamide 6.90 10 9.9 2.5 19 Chlorbufam 6.09 10 7.6 1.9 20 Chlordane alpha-cis 7.25 5 6.4 0.8 21 Chlordane gamma-trans 7.32 5 5.7 0.7 22 Chlorothalonil 6.29 5 5.7 0.7 23 Chlorpropham 5.77 10 4.6 1.2 24 Chlorpyrifos-ethyl 6.85 5 4.3 0.5 25 Chlorpyrifos-methyl 6.55 5 4.2 0.5 26 Chlozolinate 7.07 5 7.9 1.0 27 Clomazone 6.12 5 4.1 0.5 28 Coumaphos 8.48 5 14.0 1.8 29 Cyanazine 6.85 5 17.0 2.2 30 Cycloate 5.72 5 8.8 1.1 31 Cyfluthrin peaks I-IV 8.59 10 11.0 2.8 32 Cyhalothrin-S 8.23 10 6.8 1.7 33 Cypermethrin peaks I-IV 8.68 10 10.0 2.6 34 DDD p,p 7.63 5 5.6 0.7 35 DDE p, p 7.42 5 4.0 0.5 36 DDT o,p 7.65 5 8.2 1.0 37 DDT p,p 7.80 5 12.0 1.6 38 Deltamethrin 9.15 10 9.8 2.5 39 Diazinon 6.21 5 4.9 0.6 40 Dichlobenil 4.69 10 8.3 2.1 41 Dichlofluanid 6.80 5 3.7 0.5 42 Dichloran 6.04 5 13.0 1.7 43 Dichlorvos 4.33 5 12.0 1.5 44 Dicrotophos 5.79 5 6.8 0.9 45 Dieldrin 7.47 5 8.0 1.0 46 Diflufenican 7.86 5 5.0 0.6 47 Dimethenamid 6.52 5 1.9 0.2 48 Diphenylamine 5.68 10 6.8 1.7 49 Endosulfan I 7.36 5 5.6 0.7 50 Endosulfan II 7.62 5 4.3 0.5 51 Endosulfan sulfate 7.81 5 4.1 0.5 52 Endrin 7.58 5 13.0 1.6 53 EPN 8.03 5 6.0 0.8 54 EPTC 4.73 10 10.0 2.6 55 Ethion 7.61 5 4.2 0.5 56 Ethofumesate 6.74 5 8.6 1.1 57 Ethoprop (Ethoprophos) 5.70 5 7.1 0.9 58 Etoxazole 8.04 10 7.5 1.9 59 Etridiazole 5.07 10 1.0 2.6 60 Etrimfos 6.33 5 4.4 0.6 61 Fenazaquin 8.11 5 4.5 0.6 62 Fenitrothion 6.74 5 4.1 0.5 63 Fenpropathrin 8.05 10 8.9 2.3 64 Fenvalerate I 8.91 5 7.4 0.9 65 Fenvalerate II 8.98 5 7.6 1.0 66 Flucythrinate I 8.69 10 8.9 2.3 67 Flucythrinate II 8.74 5 7.5 1.0 No Compound RT (min) pg on Column % RSD 68 Flurochloridone 6.92 5 4.9 0.6 69 Flutolanil 7.33 5 10.0 1.2 70 Fluvalinate 8.93 5 11.0 1.4 71 Folpet 7.19 5 9.2 1.2 72 Furalaxyl 7.13 5 3.6 0.5 73 Heptachlor 6.67 5 3.7 0.5 74 Heptachlor epoxide-cis 7.12 5 6.3 0.8 75 Heptachlor epoxide-trans 7.14 5 9.4 1.2 76 Hexachlorobenzene 6.00 5 9.6 1.2 77 Hexazinone 7.83 5 6.9 0.9 78 Iprodione 7.99 10 12.0 3.1 79 Malaoxon 6.55 5 4.1 0.5 80 Mephosfolan 7.11 5 9.2 1.2 81 Metazachlor 7.06 10 6.4 1.6 82 Methacrifos 5.21 5 9.6 1.2 83 Methidathion 7.21 5 4.7 0.6 84 Methoxychlor 8.05 5 8.6 1.1 85 Metribuzin 6.54 10 3.8 1.0 86 Napropamide 7.34 10 6.2 1.6 87 Nitrofen 7.54 5 7.0 0.9 88 Nitrothal-isopropyl 6.92 5 4.3 0.5 89 Oxadiazon 7.39 5 4.7 0.6 90 Oxychlordane 7.12 5 5.9 0.7 91 Oxyfluorfen 7.42 10 7.7 2.0 92 Paraoxon-methyl 6.30 5 10.0 1.3 93 Parathion (ethyl) 6.88 5 6.5 0.8 94 Parathion-methyl 6.59 5 4.4 0.6 95 Pendimethalin 7.04 5 6.7 0.9 96 Pentachloroaniline 6.48 5 5.4 0.7 97 Pentanochlor (Solan) 6.78 5 5.0 0.6 98 Permethrin I 8.43 5 10.0 1.3 99 Permethrin II 8.46 5 9.7 1.2 100 Phosalone 8.18 5 8.5 1.1 101 Phosmet 8.02 5 6.8 0.9 102 Pirimiphos methyl 6.72 5 4.0 0.5 103 Pirimiphos-ethyl 6.95 5 4.0 0.5 104 Procymidone 7.15 5 5.7 0.7 105 Propachlor 5.61 5 6.9 0.9 106 Propanil 6.52 10 13.0 3.2 107 Propargite 7.86 5 10.0 1.3 108 Propetamphos 6.16 5 4.3 0.5 109 Propham 5.08 5 5.7 0.7 110 Prosulfocarb 6.71 10 4.2 1.1 111 Prothiofos 7.37 5 7.7 1.0 112 Pyrazophos 8.29 5 9.0 1.1 113 Pyridaben 8.49 5 10.0 1.3 114 Pyridaphenthion 7.97 5 16.0 2.0 115 Pyrifenox-E 7.11 5 11.0 1.4 116 Quinalphos 7.13 5 6.9 0.9 117 Quinomethionate 7.25 5 8.8 1.1 118 Quintozene 6.16 5 6.3 0.8 119 Resmethrin 7.89 5 9.6 1.2 120 Spirodiclofen 8.42 5 7.9 1.0 121 Tecnazene 5.57 5 9.6 1.2 122 Tefluthrin 6.30 5 4.5 0.6 123 Terbuthylazine 6.18 10 5.1 1.3 124 Terbutryn 6.73 5 5.6 0.7 125 Tetrachlorvinphos 7.24 5 8.7 1.1 126 Tetradifon 8.16 5 8.0 1.0 127 Tetramethrin 8.01 5 6.2 0.8 128 Tolclofos-methyl 6.60 5 4.6 0.6 129 Tolylfluanid 7.09 5 5.3 0.7 130 Triallate 6.36 10 5.9 1.5 131 Trifluralin 5.78 5 7.5 1.0 132 Vinclozolin 6.57 5 7.1 0.9 IDL

6 Figure 5. Coefficient of determination (R2) and residuals values (%RSD) calculated for a linear range of 0.5 100 ng/g (or 1.0 200 ng/g). Dashed lines represent the 10% and 20% RSD residual limits. Linearity of Response Linearity of the GC-MS/MS system was evaluated across a concentration range of 0.5 100 ng/g (or 1 200 ng/g for some analytes) using matrix-matched standards. In all cases the coefficient of determination (R 2 ) was higher than 0.99 with an average value of R 2 = 0.997. Moreover, individual residual values were <20% with an average value of 10% (Figure 5). Comprehensive Analysis of Additional Pesticides Targeted screening and quantification of a given number of pesticides is important, but there is increasing interest in screening samples for compounds other than those in a target list. To answer the question, What else is in my sample?, samples have to be screened for unexpected or new pesticides or for metabolic/transformation products that could be present in the samples in addition to the targeted compounds. The capability of fast analytical instrumentation enables simultaneous acquisition of full scan and SRM/SIM data.

7 Figure 6. Comprehensive analysis of baby food contaminants using simultaneous full scan/srm data acquisition. Compound at RT = 5.89 min identified as metolachlor (using NIST) in the full scan acquisition window. Using the TSQ 8000 Evo GC-MS/MS system, the baby food samples were screened for additional compounds. Data was acquired in full scan and SRM modes simultaneously. An example of a full scan/srm chromatogram is shown in Figure 6. The extracted mass spectrum of the peak eluting at RT = 5.89 min was submitted to NIST mass spectral library and identified as metolachlor (a compound not in the spiking solution or the target list) with a probability of 95%. This result shows the advantage of using such simultaneous data acquisition, which is possible only using fast instrumentation such as the TSQ 8000 Evo GC-MS/MS.

8 Conclusion The results of this work show that laboratory productivity can be tripled using the Thermo Scientific TSQ 8000 Evo triple quadrupole GC-MS system. Acceleration of sample analysis is made possible by: direct analysis of acetonitrile extracts with no need for an additional solvent exchange step. shorter GC run times using fast data acquisition with the EvoCell fast collision cell technology. comprehensive detection of target pesticides and nontargeted pesticides using simultaneous full scan and SRM data acquisition. Additional pesticides were identified by searching the full scan data against the NIST library. Excellent sensitivity was achieved. All pesticides were detected and identified at a concentration of 5 10 ng/g with IDL values from 0.2 3.7 ng/g. References 1. Commission Regulation (EU) No 396/2005 on maximum residue levels of pesticides in or on food and feed of plant and animal origin and amending Council Directive 91/414/EC, 16.3.2005, p. 1 16. 2. SANCO/12571/2013 (2014), Guidance document on analytical quality control and validation procedures for pesticide residues analysis in food and feed, 19.11.2013 rev. 0. 3. EN 15662 Version 2.2, Date: 2008-04, Foods of plant origin Determination of pesticide residues using GC-MS and/or LC-MS/MS following acetonitrile extraction/partitioning and cleanup by dispersive SPE QuEChERS-method. These results demonstrate that fast GC data acquisition using the TSQ 8000 Evo GC-MS/MS system delivers excellent peak area reproducibility and compound linearity. www.thermofisher.com 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 AN10432-EN 0716S Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 45 453 9100 Korea +82 2 3420 8600 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Spain +34 914 845 965 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752