Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017

Similar documents
METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

MARS 6 Batch Microwave Acid Digestion System

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Determination of Orthophosphate Ion

Spectrophotometric Determination of Ferrocyanide in Effluents

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Determination of Orthophosphate Ion

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

National standard of People s Republic of China

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

Test Method: CPSC-CH-E

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Chloramine (Mono) and Nitrogen, Free Ammonia

Oxygen Demand, Chemical

Method for estimation of iodine in urine

Procedure for the Determination of Permanganate Oxidizable Carbon

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Oxygen Demand, Chemical

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Chapter 12 Tex-617, Determining Chloride in Concrete

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Sodium Chloride - Analytical Standard

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

The Fluorometric Determination of Acetylsalicylic Acid in an Aspirin Tablet

National standard of People s Republic of China

Method to 0.50 mg/l NH 3 N Powder Pillows

Cyanide, colorimetric, pyridine-pyrazolone

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Recovery of Copper Renee Y. Becker Manatee Community College

GENERAL INSTRUCTIONS GENERAL PREPARATION

Method to 0.50 mg/l NH 3 N Powder Pillows

BATTERY INDUSTRY STANDARD ANALYTICAL METHOD

NICKEL-63/59 IN WATER

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

TECHNETIUM-99 IN SOIL

Sodium Chloride - Analytical Standard

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

RAPID KJELDAHL BENCHNOTES

maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline)

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Chemical Reactions: The Copper Cycle

DR/4000 PROCEDURE. IRON, Total

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

The analysis of organic acid content of additives, premix, feed, and water.

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

Chemistry Determination of Mixed Acids

Experimental Procedure Lab 402

ISO INTERNATIONAL STANDARD. Soil quality Determination of ph. Qualité du sol Détermination du ph. Second edition

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

IT DOESN T GET ANY EASIER THAN THIS!

Distribution equilibrium

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon

UGANDA STANDARD. Fruit and vegetable products Determination of titratable acidity US ISO 750. First Edition 2009-mm-dd

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

Experiment 24. Chemical recycling of poly(ethylene) terephthalate (PET)

NITROGEN, TOTAL, HR, Test N Tube

Sodium Chloride - Analytical Standard

DR/4000 PROCEDURE NITRATE. 2. The display will show:

SOIL ORGANIC CONTENT USING UV-VIS METHOD

Titrimetric Determination of EASTMAN Color Developing Agent, CD-3, in Process ECN-2 Developer with Sulfato Cerate

Determination the elemental composition of soil samples

Experiment 20-Acid-Base Titration: Standardization of KOH and Determination of the Molarity and/or Percent Composition of an Acid Solution

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Potentiometric measurement of ph

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

Experiment 7: ACID-BASE TITRATION: STANDARDIZATION OF A SOLUTION

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

TEST REPORT. The following sample(s) was(were) identified by the customer as : PP-NON WOVEN

GRAVIMETRIC ANALYSIS OF A CHLORIDE SALT. REFERENCES: Nelson, J., Chemistry: The Central Science, 3 rd edition, Prentice-Hall, 1985

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

National Research Council Institute for Ecosystem Study Verbania Pallanza - Italy

Transcription:

Page 1 of 3 Abstract The increasing use of consumer and electronic device components worldwide has drawn increased attention to their impact on the environment. The correct disposal of these materials is of critical importance due to the potential for hexavalent chromium contamination of soil and water. Most countries around the world have adopted regulations regarding the Restriction of Hazardous Substances (RoHS). Recently the International Electrotechnical Commission (IEC) has established a new test method, IEC 62321-7-2:2017, to test for hexavalent chromium in many of these products. This new method is a replacement for parts of the corresponding clauses in IEC 62311:2008. Introduction The recent passage of IEC 62321-7-2:2017 provides determination of hexavalent chromium in polymers and electronics using colorimetric determination by a UV-Vis Spectrophotometer. Prior to analysis, a sample preparation method using microwave digestion is used. The IEC method has several post digestion steps and requires multiple reagents to be made up prior to the procedure. This application note will establish the correct microwave equipment, options, and program to be followed in order to be in compliance with the method. It will also attempt to make the method easier to follow by alerting the analyst to make up specific reagents prior to starting the procedure. Instrumentation The microwave procedure for the IEC method (62321-7-2:2017 e) does not require rigorous conditions. It only requires the samples to reach a temperature of 150 160 C and to hold for 90 minutes. A MARS 6 equipped with standard IR temperature control and the stirring option was used. The samples were prepared in the CEM 55 ml MARSXpress vessel, a simple to use 3 part vessel with vent and reseal capabilities. (See Figure 1) Alternatively, a MARS One with 55 ml MARSXpress Vessels or a MARS 6 with EasyPrep or iprep vessels can also be used for this method. Liner - Add sample and digestion reagents Pressure Plug - Place plug on top of liner Cap - Hand tighten cap Figure 1: MARSXpress 3-Part Vessel

Page 2 of 3 Procedure Note: Prior to the digestion procedure, you will need to grind or cut up samples into small pieces. Reagents Preparation You must make up these reagents prior to sample preparation. All reagents used must be reagent grade or better. Digestion Solution Dissolve 20 g of NaOH and 30 g of NaCO 3 in water in 1 L volumetric flask and dilute to mark with clean deionized water. This should be stored at 20-25 C and be made fresh monthly. Test ph prior to use and discard solution if ph is below 11.5. Phosphate Buffer Disolve 87.09 g K 2 HPO 4 and 68.04 g KH 2 PO 4 into 700 ml of clean deionized water. Transfer to a 1 L volumetric flask and make up to volume with clean deionized water. Nitric Acid at 35% Dilute 50 ml of reagent grade HNO 3 to 100 ml with clean deionized water. Store at 20 C. Diphenylcarbazide Dissolve 250 mg 1,5-Diphenylcarbazide in 50 ml of acetone. Store in a brown bottle. Prior to use, check solution for discoloration. Store up to two weeks and if solution becomes discolored discard and prepare a fresh batch. Sulfuric Acid at 10% Dilute 10 ml of distilled reagent grade or spectroscopic grade H 2 SO 4 with clean deionized water to 100 ml into a volumetric flask. Other required Reagents Toluene (Analytical Grade) MgCl 2 anhydrous (Analytical Grade) Microwave Digestion Procedure 1. Weigh approximately 0.15 grams of sample into 55 ml MARSXpress vessel and add magnetic stir bar. 2. Add 10 ml of Digestion Solution 3. Add 5 ml of Toluene (analytical grade) 4. Add 400 mg of MgCl 2 (anhydrous) analytical grade 5. Add 0.5 ml of Phosphate Buffer 6. Evenly space vessels into MARSXpress turntable and place into microwave cavity. Create Classic microwave digestion program as defined in Table 1. Table 1: Microwave Digestion Custom Program Settings Stage Ramp Temp ( C) Hold 1 15 mins 155 60 mins 2 1 sec 155 30 mins Results The MARS 6 with MARSXpress vessels can be used to successfully perform the closed vessel microwave digestion portion (62321-7-2:2017 e) of the hexavalent chromium determination sample preparation. As shown in Figure 2, the precise power control can easily obtain the necessary digestion conditions required for the duration of the run. Although the method does not call for it, we would recommend the stirring option to help extract all of the hexavalent chromium into the aqueous phase in the vessel prior to separation. Once the microwave portion is complete a post digestion procedure is required, which is outlined in the following section. Temperature ( C) 250 C 200 C 150 C 100 C 50 C 0 C 0 15 30 45 60 Time (min) 75 90 105 1,800 W 1,500 W 1,200 W 900 W 600 W 300 W 0 Power (W) Figure 2: Temperature Curve The Green power line represents the precise control of the MARS system while the Red Temperature line shows that the samples are being held in the correct temperature range of 150-160 C for 90 minutes.

Page 3 of 3 Post Digestion Preparation 1. Cool and transfer solution to a separatory funnel to separate organic phase. Discard organic phase. 2. Filter aqueous phase with a 0.45 µm membrane filter. Rinse the digestion vessel three times with water and filter rinse solution as well. If filter becomes clogged use a larger pore size filter. 3. Rinse inside of flask and filter pad with water and transfer filtrate and rinse solution into a 150 ml beaker with magnetic stir bar. 4. Add Nitric Acid (35%) dropwise while stirring and monitor ph to adjust to 7.5 +/- 0.5. 5. Check to see if sample is clear. If sample is clear: a. Add 2.5 ml of diphenylcarbazide solution to each beaker. b. Slowly add sulfuric acid (10%) to the vessel to adjust ph to 2.0 +/- 0.5. c. Transfer the contents quantitatively to a 50 ml volumetric flask and adjust sample volume to 50 ml with deionized water and invert several times. d. Let stand 5 10 minutes to develop full color. 2. Transfer appropriate portion to a 1 cm absorption cell and measure at 540 nm with colorimetric instrument. Analysis must be complete within 30 minutes of color development. 3. Correct the absorbance reading of the sample by subtracting the absorbance of a blank carried through the color development procedures. 4. From corrected absorbance determine the concentration by referring to calibration curve from IEC method. If sample is turbid or colored: Filter the sample with a 0.45 µm membrane filter. If sample is colored, filter solution with a C 18 cartridge syringe before proceeding to develop color. If sample is clear after filtration proceed to develop color. a. Add 2.5 ml of diphenylcarbazide solution to each beaker. b. Slowly add sulfuric acid (10%) to the vessel to adjust ph to 2.0 +/- 0.5. c. Transfer the contents quantitatively to a 50 ml volumetric flask and adjust sample volume to 50 ml with deionized water and invert several times. d. Remove 5 ml from the flask and record and measure with colorimetric instrument. This is the background absorption measurement. e. Perform background correction by adding 2.5 ml diphenylcarbazide solution to each sample digestion solution. f. Mix and add deionized water to adjust volume to 50 ml inverting several times. g. Let stand 5 10 minutes to develop full color. 2. Transfer appropriate portion to a 1 cm absorption cell and measure at 540 nm with colorimetric instrument. Analysis must be complete within 30 minutes of color development. 3. Correct the absorbance reading by subtracting the reading from the measurement taken above for background absorbance. 4. From corrected absorbance determine the concentration by referring to calibration curve from IEC method. Discussion The IEC method 62321-7-2:2017 is a suitable alternative to analysis by ICP; however, great care should be taken to minimize error in each step. Technicians should be well trained in proper analytical technique to minimize sample to sample variability and minimize introduction of error, which could lead to false or inaccurate results. United States (Headquarters) 800-726-3331 704-821-7015 Fax: 704-821-7894 info@cem.com France 33 (01) 69 35 57 80 Fax: 33 (01) 60 19 64 91 info.fr@cem.com Germany, Austria, Switzerland (49) 2842-9644-0 Fax: (49) 2842-9644-11 info@cem.de Ireland +353 (0) 1 885 1752 Fax: +353 (0) 1 885 1601 info.ireland@cem.com Italy Japan United Kingdom www.cem.com (39) 35-896224 Fax: (39) 35-891661 info.srl@cem.com +81-3-5793-8542 Fax: +81-3-5793-8543 info@cemjapan.co.jp (44) 1280-822873 Fax: (44) 1280-822873 info.uk@cem.com All rights reserved. This may not be reproduced or published without written permission from CEM.

Flow Chart of IEC Method 62321-7-2:2017 Reagents Preparation You must make up these reagents prior to sample preparation. All reagents used must be reagent grade or better. Digestion Solution Dissolve 20 g of NaOH and 30 g of NaCO 3 in water in 1 L volumetric flask and dilute to mark with clean deionized water. This should be stored at 20-25 C and be made fresh monthly. Test ph prior to use and discard solution if ph is below 11.5. Phosphate Buffer Disolve 87.09 g K 2 HPO 4 and 68.04 g KH 2 PO 4 into 700 ml of clean deionized water. Transfer to a 1 L volumetric flask and make up to volume with clean deionized water. Nitric Acid at 35% Dilute 50 ml of reagent grade HNO 3 to 100 ml with clean deionized water. Store at 20 C. Diphenylcarbazide Dissolve 250 mg 1,5-Diphenylcarbazide in 50 ml of acetone. Store in a brown bottle. Prior to use, check solution for discoloration. Store up to two weeks and if solution becomes discolored discard and prepare a fresh batch. Sulfuric Acid at 10% Dilute 10 ml of distilled reagent grade or spectroscopic grade H 2 SO 4 with clean deionized water to 100 ml into a volumetric flask. Other required Reagents Toluene (Analytical Grade) MgCl 2 anhydrous (Analytical Grade) Microwave Digestion Procedure Weigh approximately 0.15 grams of ground sample into 55 ml MARSXpress vessel and add magnetic stir bar. To each vessel add: Add 10 ml of Digestion Solution Add 5 ml of Toluene (analytical grade) Add 400 mg of MgCl 2 (anhydrous) analytical grade Add 0.5 ml of Phosphate Buffer Evenly space vessels into MARSXpress turntable and place into microwave cavity. Create Classic microwave digestion program as defined below. Microwave Digestion Custom Program Settings: Stage Ramp Temp ( C) Hold 1 15 mins 155 60 mins 2 1 sec 155 30 mins Post Digestion Procedure 1. Cool and transfer solution to a separatory funnel to separate organic phase. Discard organic phase. 2. Filter aqueous phase with a 0.45 µm membrane filter. Rinse the digestion vessel three times with water and filter rinse solution as well. If filter becomes clogged use a larger pore size filter. 3. Rinse inside of flask and filter pad with water and transfer filtrate and rinse solution into a 150 ml beaker with magnetic stir bar. 4. Add Nitric Acid (35%) dropwise while stirring and monitor ph to adjust to 7.5 +/- 0.5. 5. Check to see if sample is clear.

Flow Chart of IEC Method 62321-7-2:2017 Yes a. Add 2.5 ml of diphenylcarbazide solution to each beaker. b. Slowly add sulfuric acid (10%) to the vessel to adjust ph to 2.0 +/- 0.5. c. Transfer the contents quantitatively to a 50 ml volumetric flask and adjust sample volume to 50 ml with deionized water and invert several times. d. Let stand 5 10 minutes to develop full color. 2. Transfer appropriate portion to a 1 cm absorption cell and measure at 540 nm with colorimetric instrument. Analysis must be complete within 30 minutes of color development. 3. Correct the absorbance reading of the sample by subtracting the absorbance of a blank carried through the color development procedures. 4. From corrected absorbance determine the concentration by referring to calibration curve from IEC method. Is the solution clear? No, solution is turbid or colored Filter the sample with a 0.45 µm membrane filter. Is the solution clear? Yes a. Add 2.5 ml of diphenylcarbazide solution to each beaker. b. Slowly add sulfuric acid (10%) to the vessel to adjust ph to 2.0 +/- 0.5. c. Transfer the contents quantitatively to a 50 ml volumetric flask and adjust sample volume to 50 ml with deionized water and invert several times. d. Remove 5 ml from the flask and record and measure with colorimetric instrument. This is the background absorption measurement. e. Perform background correction by adding 2.5 ml diphenylcarbazide solution to each sample digestion solution. f. Mix and add deionized water to adjust volume to 50 ml inverting several times. g. Let stand 5 10 minutes to develop full color. No, solution is colored If digest is colored filter solution with a C 18 cartridge syringe before proceeding to develop color. 2. Transfer appropriate portion to a 1 cm absorption cell and measure at 540 nm with colorimetric instrument. Analysis must be complete within 30 minutes of color development. 3. Correct the absorbance reading by subtracting the reading from the measurement taken above for background absorbance. 4. From corrected absorbance determine the concentration by referring to calibration curve from IEC method.