Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Similar documents
AppNote 12/2008 KEYWORDS ABSTRACT

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances KEYWORDS

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to En vi ron men tal Aqueous Samples INTRODUCTION

TechNote 5/2002 KEYWORDS ABSTRACT. SBSE, Twister, Thermal Desorption

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

AppNote 1/2011 KEYWORDS ABSTRACT

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

AppNote 4/2004. KEYWORDS Multi-residue method, pesticides, vegetables, fruits, green tea, Stir bar sorptive extraction (SBSE), GC-MS.

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Application. Gas Chromatography February Introduction

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Analysis of BTEX in Natural Water with SPME

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Highly Sensitive and Rugged GC/MS/MS Tool

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Author. Abstract. Introduction

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

AppNote 2/2012. Determining Phenolic Compounds in Whisky using Direct Large Volume Injection and Stir Bar Sorptive Extraction KEYWORDS ABSTRACT

GC/MS Application Note

AppNote 15/2012. Large Volume Injection with Solvent Venting - Application to Trace Detection of Analytes in Water INTRODUCTION

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

Application. Gas Chromatography March 1998

AppNote 6/2010. part a. Alternative Procedure for Extraction and Analysis of PAHs in Seafood by QuEChERS-SBSE-GC-MS KEYWORDS ABSTRACT

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of volatiles in Food sample preparation strategies and techniques Dr Kathy Ridgway

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Application Note. Abstract. Authors. Introduction

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

ENVIRONMENTAL analysis

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

AppNote 1/2004. Use of a Mass Spectral Based Chemical Sensor to Discriminate Food and Beverage Samples: Olive Oils and Wine as Examples KEYWORDS

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Detection of Chemical Warfare Agents by Transportable GC/MS

PAL SPME Arrow The Better SPME

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

Sulfotepp impurities in Chlorpyrifos EC formulations

AppNote 8/2008. A New Purge Tool for Use with Automated Headspace Analysis KEYWORDS ABSTRACT

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

PAL SPME Arrow The Better SPME

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

AppNote 3/2014. Analysis of Aroma Compounds in Edible Oils by Direct Thermal Desorption GC/MS using Slitted Microvials KEYWORDS ABSTRACT

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Multi-residue analysis of pesticides by GC-HRMS

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Off-gassing of Rubber Particles Used for Athletic Fields using Automated Dynamic Headspace Sampling

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface

Thermal Desorption Unit TDU

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

AppNote 14/2012. Thermal Gravimetric Analysis/Mass Spectrometry Simulation using the GERSTEL Automated Pyrolyzer KEYWORDS ABSTRACT

Signal, Noise, and Detection Limits in Mass Spectrometry

A New Way to Extract Off-Flavor Compounds in the Aquatic Environment. Stir Bar Sorptive Extraction (SBSE)

Application Note 032

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

STANDARD OPERATING PROCEDURES

AppNote 10/2012. Large Volume Injection with GC/MS Multi Column Switching for Direct Determination of Ethyl Carbamate in Alcoholic Beverages KEYWORDS

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

AppNote 1/2007. Automated Dynamic Headspace Sampling using Replaceable Sorbent Traps KEYWORDS ABSTRACT

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Application Note. Abstract. Authors. Environmental

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

AppNote 8/2002 KEYWORDS ABSTRACT. Electronic Nose, Fast GC, Headspace Analysis, Chemometrics,

AppNote 12/2012. Comparison of EG-Silicone-SBSE and Derivatization-PDMS-SBSE for the Analysis of Phenolic Compounds and Off-fl avors in Water KEYWORDS

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

AppNote 2/2002. Discrimination of Soft Drinks using a Chemical Sensor and Principal Component Analysis KEYWORDS ABSTRACT

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Solid Phase Micro Extraction of Flavor Compounds in Beer

Determination of Volatile Substances Proof of Food Adulteration

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Rapid Determination Method for 12 Pyrethroid Pesticide Residues in Tea by Stir Bar Sorptive Extraction-Thermal Desorption-Gas Chromatography*

The Analysis of Organophosphate Pesticides by LC/MS Application

Transcription:

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS Application Note Food Authors Nobuo Ochiai and Kikuo Sasamoto GERSTEL K.K. 2-13-18 Nakane Meguro-ku, Tokyo, 152-0031 Japan Abstract The performance of sequential stir bar sorptive extraction (sequential SBSE) in combination with thermal desorption (TD)-GC/MS for screening of pesticide residues in water was described. Compared to conventional SBSE, sequential SBSE provides more uniform enrichment over the entire polarity and volatility range for organic pollutants at ultra trace levels in water. Sequential SBSE consists of an SBSE performed first on a 5-mL sample without modifier using one stir bar, then on the same sample after addition of 30% NaCl using a second stir bar. After extraction the two stir bars are placed in a single glass desorption liner and are thermally desorbed simultaneously. The presence of pesticide is elucidated with a retention time locked GC/MS method (RTL-GC/MS). Screening of pesticides at the ng/l level in river water samples was successfully carried out with sequential SBSE-TD-RTL-GC/MS operated in the scan mode.

Introduction The determination of pesticide residues in environmental samples, such as water, soil, and agricultural products, has been a major subject for many years. This is because of their potential risk for human health, persistence and tendency to bio-accumulate. For water samples, analytical methods usually include extraction and enrichment steps for determining pesticide residues at very low levels (<sub-µg/l). Therefore, miniaturized methods such as solid phase micro-extraction (SPME) and stir bar sorptive extraction (SBSE) were introduced because they are simple, solvent-free techniques that allow extraction and concentration in a single step. [1,2]. These sorptive extraction methods have been successfully applied to the determination of organic compounds in various sample matrices, such as water, soil, food, and biological fluid [3-5]. These methods also provide enhanced sensitivity because the extracted fraction (on a fiber or on a stir bar) can be introduced quantitatively into a GC system by thermal desorption. In addition, the enrichment factor for SBSE is higher than that of SPME because of the 50-250 times larger volume of extraction phase on the stir bar. Several authors indicated that the SBSE method allows high recovery and an extremely low limit of detection (LOD) at the sub-ng/l level, particularly for some solutes having hydrophobic characteristics [4, 5]. In 2008, Ochiai et al, developed an SBSE procedure termed sequential SBSE for exhaustive enrichment of organic pollutants in aqueous samples. This procedure is performed sequentially for one aliquot under two extraction conditions using two stir bars [6]. The technique counters the difficulty in recovering both hydrophobic and hydrophilic compounds. This new approach produced a remarkable improvement in recoveries for a set of 80 pesticides (log K o/w : 1.70-8.35) in water, 82-113% for the majority and less than 80% for only five hydrophilic compounds. This application note describes the performance of sequential SBSE-thermal desorption (TD) in combination with retention time locked (RTL)-GC/MS for the screening of pesticide residues in water. Sequential SBSE-TD-RTL-GC/MS For the first SBSE, five milliliters of water sample were transferred to 10 ml headspace vials. A stir bar (Gerstel Twister; 24 µl of PDMS) was added and the vial was sealed with a screw cap. SBSE of several samples was performed simultaneously at room temperature (24 ºC) for 60 min while stirring at 1500 rpm. After the first extraction, the stir bar was removed with forceps, dipped briefly in Milli-Q water, dried with a lint-free tissue, and placed in a glass thermal desorption liner. The glass liner was temporarily placed and stored in a sealed sample tray of the MPS 2. For the second extraction, 30 % NaCl was dissolved in the sample. Then, a second stir bar was added and the vial was capped again. The second extraction was performed under the same conditions as the first extraction. After the second extraction, the stir bar was removed with forceps, dipped briefly in Milli-Q water, dried with a lint-free tissue, and placed in the glass liner which contained the first SBSE stir bar. Finally, the glass liner was placed in the thermal desorption unit. No further sample preparation was necessary. Figure 1 shows the experimental setup of sequential SBSE. Reconditioning of stir bars was done after use by soaking in Milli-Q purified water and a mixture of methylene chloridemethanol (1:1) for 24 h each. Stir bars were then removed from the solvent and dried on a clean surface at room temperature for 1 h. Finally, the stir bars were thermally conditioned for 30 min at 300 ºC in a flow of helium. Typically, 30 extractions could be performed with the same stir bar. 1 st Extraction 2 nd Extraction 30% NaCl Experimental SBSE 1h@1500 rpm SBSE 1h@1500 rpm Instrumentation Analyses were performed with a GERSTEL TDU thermal-desorption unit equipped with a MPS 2 auto-sampler (Gerstel) and a CIS 4 programmed temperature vaporization (PTV) inlet (Gerstel) installed on an Agilent 7890 GC with an Agilent 5975 Series GC/MSD triple-axis detector (TAD). Solutes: log K O/W > 4 Solutes: log K O/W < 4 Figure 1. Experimental setup of sequential SBSE 2

The two stir bars were thermally desorbed by the thermal desorption unit (TDU) with 50 ml/min desorption flow. Desorbed compounds were cryo-focused on quartz wool packed liner in the PTV inlet for subsequent GC/MS analysis. The analytical conditions are summarized in Table 1. Table 1 Stir bar Sequential SBSE TDU PTV Column RTL GC Oven MSD Analytical conditions GERSTEL Twister; 24 µl PDMS 5 ml sample volume 60 min extraction for each SBSE Non-modifier for 1st SBSE 30 % NaCl for 2nd SBSE 1500 rpm stirring speed Splitless 50 ml/min desorption flow 40 C (0.2 min); 720 C/min; 280 C (5 min) Quartz wool packed liner Splitless 100 C (0.5 min); 12 C/s; 280 C (hold) 30 m 0.25 mm id 0.25 µm Agilent HP-5ms Chlorpyrifos methyl locked to 15.59 min. 70 C (2 min); 25 C/min; 150 C; 3 C/min; 200 C; 8 C/min; 300 C (2 min) Scan mode Scan: m/z 58-510; 3.2 scans/s Results and Discussion Figure 2 shows a comparison of the recoveries of SBSE without modifier, SBSE with 30% NaCl, and sequential SBSE for representative pesticides with various log K o/w values in natural water spiked at 500 ng/l. The recoveries for solutes with log K o/w of less than 4.0 dramatically increased with salt addition, for example, for pirimicarb (carbamate; log K o/w : 1.70), fenobucarb (carbamate; log K o/w : 2.79), and pacrobutrazol (other; log K o/w : 3.36), the recovery increased from 15% to 74%, 41% to 90%, and 31% to 95%, respectively. However, recovery for solutes with log K o/w of more than 4.0 drastically decreased, for example, for terbufos (organophosphorus; log K o/w : 4.24), pyridaben (other; log K o/w : 5.47) and permethrin 1.2 (pyrethroid; log K o/w : 7.43), the recovery decreased from 89% to 68%, 100% to 57%, and 101% to 54%, respectively. In contrast with conventional SBSE with or without salt addition, the sequential approach could eliminate the negative effect of the salt for solutes with log K o/w of more than 4.0, while maintaining increased recovery for hydrophilic solutes with salt addition, resulting in high recovery. Therefore, using sequential SBSE, results in higher sensitivity for a wide range of solutes with different polarities. Figure 3 shows the mass chromatogram (m/z 304) of diazinon (log K o/w : 3.86) in 5 ml of natural water spiked at 20 ng/l. Excellent sensitivity (peak-to-peak S/N of 40) and a well defined mass spectrum were obtained for ng/l level samples even with the scan mode. 120 100 SBSE SBSE with 30 % NaCl Seq. SBSE 900 800 Recovery (%) 80 60 40 20 0 Abundance 700 600 500 400 300 200 100 Figure 2. Comparison of the recovery of SBSE without modifier, SBSE with 30 % NaCl, sequential SBSE for representative pesticides with various log K o/w in natural water spiked at 500 ng/l. 0 Figure 3. 12.00 13.00 14.00 15.00 16.00 Retention time (min) 17.00 Mass chromatogram (m/z 304) of diazinon (log K o/w : 3.86) in 5-mL natural water spiked at 20 ng/l. 3

Figure 4 shows a Wiley library search result, which obtained a match factor of 94. RTL GC/MS method can eliminate many false positives and give more confidence in compound identification with not only mass spectral information but also locked retention times. Recently, Agilent updated the RTL pesticide library with 926 pesticides, endocrine disruptors, and related compounds [7]. For the present paper, the presence of pesticides in river water sample is elucidated automatically via the RTL screener in combination with the RTL library for 926 compounds. (a) (b) Figure 4. Wiley library search result of diazinon (log K o/w : 3.86) in 5-mL natural water spiked at 20 ng/l. (a) Measured mass spectrum (b) Wiley library spectrum. 4

Figures 5 and 6 show the screener software windows for the positive detection and identification of symetryn (log K o/w : 2.90) and pyributicarb (log K o/w : 5.34) in Ayase river water sample. The ratio of four qualifier ions are measured and compared with those listed in the database. The software window also shows the deviation of measured retention time with the RTL value. After identification of these pesticides by the RTL screener, quantification was performed in six replicate analyses with standard addition calibration method. The determined concentrations were 240 ng/l (RSD: 2.9 %, n = 6) for symetryn and 25 ng/l (RSD: 8.9 %, n = 6) for piributicarb, respectively. (a) (b) (c) Figure 5. Results screener window of the positive identification of symetryn (log K o/w : 2.90) in Ayase river water (a) Mass chromatograms (m/z 213, 170, 155 and 198) (b) Measured mass spectrum (c) Expected and measured relative ion abundance ratio and deviation of the RTL value. (a) (b) (c) Figure 6. Results screener window of the positive identification of pyributicarb (log K o/w : 5.34) in Ayase river water (a) Mass chromatograms (m/z 165, 108, 181 and 93) (b) Measured mass spectrum (c) Expected and measured relative ion abundance ratio and deviation of the RTL value. 5

Conclusion The combination of sequential SBSE, thermal desorption and RTL GC/MS operated in the scan mode can provide a very powerful system to screen wide range of pesticide residues at ultra trace levels in water. References 1. C. L. Arthur, J. Pawliszyn, Anal. Chem, 1990, 62: 2145. 2. E. Baltussen, P. Sandra, F. David, C. A. Cramers, J. Microcol. Sep, 1999, 11: 737. 3. H. Lord, J. Pawliszyn, J. Chromatogr. A, 2000, 885: 153. 4. F. David, P. Sandra, J. Chromatogr. A, 2007, 1152: 54. 5. F. M. lancas, M. E. C. Queiroz, P. Grossi, I. R. B. Olivares, J. Sep. Sci, 2009, 32: 813. 6. N. Ochiai, K. Sasamoto, H. Kanda, E. Pfannkoch, J. Chromatogr. A, 2008, 1200: 72. 7. P. Wiley, Screening for 926 pesticides and endocrin disruptors by GC/MS with deconvolution reporting software and a new pesticide library, Agilent Technologies, publication 5989-5076EN, www.agilent.com/chem www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2010 Printed in the USA January 19, 2010 5990-5217EN