Interface (backside) & Extraction Lens

Similar documents
ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

Plasma Interface Element2

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

ICPMS Doherty Lecture 1

Mass Spectrometry in MCAL

MS Goals and Applications. MS Goals and Applications

Extrel Application Note

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry

A Multi-beamlet Injector for Heavy Ion Fusion: Experiments and Modeling

Simulation of the Plasma Meniscus with and without Space Charge using Triode Extraction System

Chemistry 311: Topic 3 - Mass Spectrometry

Introduction to the Q Trap LC/MS/MS System

Basic structure of SEM

MS Goals and Applications. MS Goals and Applications

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Harris: Quantitative Chemical Analysis, Eight Edition

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

Chemistry Instrumental Analysis Lecture 35. Chem 4631

CEE 772: Instrumental Methods in Environmental Analysis

1. The range of frequencies that a measurement is sensitive to is called the frequency

= 6 (1/ nm) So what is probability of finding electron tunneled into a barrier 3 ev high?

Secondary Ion Mass Spectrometry (SIMS)

The Plasma Phase. Chapter 1. An experiment - measure and understand transport processes in a plasma. Chapter 2. An introduction to plasma physics

Suzanne Bell Second Edition

Technology Solution. Principles of Sheath Technology and Low Maintenance Ionizers. The Operation of Corona Ionizers. The Creation of White Fuzzballs

Spectrometric Methods of Analysis. OCN 633 Fall 2013

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

Comparing Theory and Experiment for Analyte Transport in the First Vacuum Stage of the Inductively Coupled Plasma Mass Spectrometer

UNIT 15: NUCLEUS SF027 1

TANDEM MASS SPECTROSCOPY

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

The Analyzer. We will deal mainly with magnetic sector instruments here although we will talk briefly about other separation methods later.

Introduction to Elemental Analysis

ELECTROSTATIC FIELDS

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

object objective lens eyepiece lens

ENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques

ICP-MS. High Resolution ICP-MS.

VLSI Technology Dr. Nandita Dasgupta Department of Electrical Engineering Indian Institute of Technology, Madras

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

Defining quality standards for the analysis of solid samples

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

1 (a) Define magnetic flux [1]

tip conducting surface

Chapter 10: Wave Properties of Particles

Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray

1. Draw in the magnetic field inside each box that would be capable of deflecting the particle along the path shown in each diagram.

PROTEAN : Neutral Entrainment Thruster Demonstration

Ion Implanter Cyclotron Apparatus System

Review of Fundamentals - Fluid Mechanics

Secondary ion mass spectrometry (SIMS)

PRINCIPLE OF ICP- AES

Jet Aircraft Propulsion Prof. Bhaskar Roy Prof. A.M. Pradeep Department of Aerospace Engineering

Power transmission from the ITER model negative ion source on MANTIS

Rocket Propulsion Prof. K. Ramamurthi Department of Mechanical Engineering Indian Institute of Technology, Madras

Table of Content. Mechanical Removing Techniques. Ultrasonic Machining (USM) Sputtering and Focused Ion Beam Milling (FIB)

Electron beam scanning

The Cyclotron I. 1. Motion of the charges occurs in two semicircular containers, D 1. and D 2

The deposition efficiency and spatial thickness distribution of films created by Directed

Brown University PHYS 0060 Physics Department LAB B -190

Contents. Preface... xvii

Assessment of the Azimuthal Homogeneity of the Neutral Gas in a Hall Effect Thruster using Electron Beam Fluorescence

Langmuir Probes as a Diagnostic to Study Plasma Parameter Dependancies, and Ion Acoustic Wave Propogation

Secondary Ion Mass Spectroscopy (SIMS)

Secondary Ion Mass Spectrometry (SIMS) Thomas Sky

Chapter 7 Plasma Basic

The Q Machine. 60 cm 198 cm Oven. Plasma. 6 cm 30 cm. 50 cm. Axial. Probe. PUMP End Plate Magnet Coil. Filament Cathode. Radial. Hot Plate.

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Analysis of Polar Metabolites using Mass Spectrometry

Chapter test: Probing the Heart of Matter

Enhanced High Aspect Ratio Etch Performance With ANAB Technology. Keywords: High Aspect Ratio, Etch, Neutral Particles, Neutral Beam I.

Huashun Zhang. Ion Sources. With 187 Figures and 26 Tables Э SCIENCE PRESS. Springer

RESEARCH ON TWO-STAGE ENGINE SPT-MAG

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

Low Level Detection of Priority Inorganic Pollutants in Water. Anthony Neumann, MSc Darren Crichton, BSc

Chapter 7. Plasma Basics

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur

The 30-Minute Guide to ICP-MS

Particle physics experiments

Electromagnetism Notes 1 Magnetic Fields

Ratio of Charge to Mass (e/m) for the Electron

Low-energy ion beamline scattering apparatus for surface science investigations

Compressible Flow. Professor Ugur GUVEN Aerospace Engineer Spacecraft Propulsion Specialist

Mass spectrometry.

Secondary ion mass spectrometry (SIMS)

Chapter 23 Electric Potential. Copyright 2009 Pearson Education, Inc.

THE HEAVY ION BEAM PROBE

A) m B) m C) m D) m E) m. 5. Which one of the following circuits has the largest resistance?

Force Due to Magnetic Field You will use

Robert A. Meger Richard F. Fernster Martin Lampe W. M. Manheimer NOTICE

CHAPTER 5 MASS AND ENERGY ANALYSIS OF CONTROL VOLUMES

Physique des plasmas radiofréquence Pascal Chabert

ELECTROMAGNETIC WAVES

Lecture Sound Waves EM Waves. Physics Help Q&A: tutor.leiacademy.org. The Doppler Effect 11/11/2014

HIGH PERFORMANCE MASS SPECTROMETERS FOR EARTH AND PLANETARY SCIENCE

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

An advanced simulation code for Hall effect thrusters

Transcription:

Plasma Interface

Interface (backside) & Extraction Lens

Extraction Lens (-2000 volts)

ION OPTICS Tip of the sampler cone is positioned to be in the region of maximum ionization Ions no longer under control Adiabatic expansion decrease in gas density and temperature enthalpy converted to directional flow = speed increases forms a free supersonic jet between sampler and skimmer cones

ION OPTICS The position of the mach disk relative to the skimmer cone is extremely important Mach disk represents a region where the supersonic jet impacts background gas Within Mach disk, jet becomes subsonic, ions can be lost due to scattering and collisions Thus, we should sample the jet BEFORE the mach disk

ION OPTICS Position of the Mach disk (relative to sampler cone) X m = 0.67D 0 (P 0 /P 1 ) 1/2 D 0 = diameter of sampler orifice P 0 = pressure in ICP P 1 = pressure in expansion chamber

ION OPTICS turns out the ideal position of the skimmer cone is ~ 2/3 of X m

Sampler cone - plasma interactions Boundary layer and Sheath (remember - sampler cone is water cooled) Boundary layer = layer of gas at intermediate temperature between that of the cone and the plasma Sheath - electrical interaction (as opposed to the physical interaction that forms the boundary layer)

ION OPTICS Why do we care about the Boundary Layer and Sheath? the Boundary Layer is a cool region that is a significant source of oxide formation design of the interface and Z axis position critical

ION OPTICS Implications of the sheath In the case of a grounded sampler cone the sheath causes the plasma to have a positive potential sheath acts as couple between the plasma and cone due to impedance differences between the sheathplasma couple and the load coil- plasma couple, the plasma acquires an even greater positive (+) potential

ION OPTICS IF this positive potential is great enough, a discharge from the plasma into the orifice is observed called secondary discharge ( ARCing ) host of bad effects including - orifice (cone) erosion - multiply charged ions - variance in the kinetic energy of the ions

ION OPTICS The question you should be asking... how representative is the distribution of ions entering the high vacuum region of the ions in the plasma?

ION OPTICS Factors that could affect the representativeness of the sampled ion stream - cooling of ion stream - cool boundary layer reactions - ion recombination - ion collisions with background gas

ION OPTICS Evidence for a representative ion stream: Short duration of extraction (~ 3 us) Lack of substantial collisions between ions in stream Electron density does not change greatly from plasma to supersonic jet

ION OPTICS Gas flow rate through sampler ~ 10 21 atoms/sec 1% also make it through the skimmer ~ 10 19 atoms/sec majority (nearly 100%) of these atoms are from the centerline essentially all ion sampling occurs in zone of silence

Ion Focusing Purpose - deliver ions to mass analyzer focuses and directs ion beam

ION LENS (TRANSFER LENS) STACK

ION LENS (TRANSFER LENS) STACK The Transfer Lens system is used to: 1. extract the ions entering the analyzer section through the orifices of the cones with very high velocity; i.e. supersonic speed; 2. focus the divergent ion beam onto the target (Entrance Slit), 3. correct the direction of the beam to the target (Entrance Slit) 4. accelerate the ions to their full speed by applying 8KV 5. shape the ion beam into a flat shape so as to make it through the Entrance slit. The ions are subsequently travelling with the desired velocity through the focus point (Entrance Slit) and start to diverge again slightly. In order to control the ion beam during the travel path through the Magnetic Field and the Electric Field, the system controls the rotation of the beam and focuses again to the next focus point.

The proportion of ions sampled from the plasma that make it through to the detector is actually extremely small (1 part in 10 6 to 10 8 ), therefore, a high ion gain detection system is required if low detection limits are to be achieved

Ion Focusing General Assumptions 1. All ions are free particles with positive charge 2. Density of ion beam is not great enough to induce repulsion ( space charge effects) 3. Presence of ions does not change electrostatic fields 4. Vacuum conditions are adequate to give ions necessary mean free path 5. Ions originate with a constant kinetic energy

IONS IN ARGON FLOW SAMPLER SKIMMER ICP SHOCK WAVES

IONS ENTRAINED IN Ar FLOW ACCELERATED TO SAME VELOCITY AS Ar AVG. KE OF Ar = AVG. KE IN ICP = 2.5 kt gas = 0.5 m Ar v Ar 2 ALL IONS (i) ACHIEVE SAME VELOCITY v i = v Ar KE i = 0.5 m i v i 2 IONS OF DIFFERENT MASS HAVE DIFFERENT KINETIC ENERGIES

Basics of Ion Transfer Positive ion of charge (z) forms in a region of potential V initial (plasma) This positve ion has a potential energy of z*v initial (z= charge) The ion will travel through any region with a potential LESS than V initial

Ion Optics Potential energy of ion is converted to kinetic energy of ion = z(v initial -V 1 ); in a region of potential V 1 The velocity of ion (v) = [2z(V initial V 1 )/m] 1/2 while in a region of constant V 1, the ion will experience a constant speed and direction once the ion nears the exit of the lens, it enters a new field region, which changes the kinetic energy (velocity) and potentially direction

Ion Optics In an ideal world... All ions leaving the plasma would have the same kinetic energy regardless of m/z ratio This would result in uniform ion transmission through the lens stack

Ion Optics In the real world... All ions are essentially accelerated to the supersonic velocity of Ar in the supersonic jet This results in a range of kinetic energy that is a function of mass

ION ENERGY vs m/z m/z MAX. ION KE (ev)

Ion Optics Sources of initial kinetic energy: - Supersonic expansion - Plasma potential (remember the sheath!) - Amount of initial kinetic energy is related to m/z ratio of the ion

Ion Optics What does this mean? A single set of ion lens settings will not be appropriate for all elements Therefore, must COMPROMISE (unless looking at a narrow mass range)

Space charge effects The mutual repulsion of ions of like (similar) charge limits the total number of ions that can be compressed into a beam of given size

Space Charge Effects Plasma - ion flux is balanced by the electron flux - essentially neutral Supersonic jet - ion flux is balanced by the electron flux - essentially neutral Lens stack - electrons are repelled by negative potential - ion beam gains positive charge

Space Charge Effects Repulsion of like charges limits the ion density of an ion beam ICP-MS beam current: ~ 1mA Beam currents of ~ 1 ua should be enough to have substantial space charge effects

EINZEL LENS EFFECT OF SPACE CHARGE BEAM CURRENT 0 1 A

Space Charge Effects Remember our assumptions #3 - density of ion beam not great enough to induce space charge effects ASSUMPTION NOT MET -further reason for non-ideal behavior and different response across the mass range

Question: what elements are most likely to be effected by space charge effects?

Space Charge Effects Space charge effect is most strongly felt by lighter mass elements The space charge force (positive-positive repulsion) acts on all ions equally

Space Charge Effects Recall most of the ions present are Ar = mass 40 Elements lighter than mass 40 are going to undergo space charge effects even if there is no other matrix element!

Question: why are the low mass elements effected by space charge effects to a greater degree?

Space Charge Effects Heavy elements are effected less than light elements Heavy matrices cause more problems than light matrices Best case scenario = analysis of uranium in water Worst case scenario = analysis of Li in U-rich solution

ESA Detector ELEMENT SCANNING HIGH RES ICP-MS DEVICE Entrance slit Quad lenses Magnet & flight tube Extraction lenses Skimmer Sampler ICP Neb & Spray chamber

MAGNET

MAGNETIC SECTOR MASS ANALYZER + ION MOVING THRU MAGNETIC FIELD STRENGTH B v B F m v F m F m F m = MAGNETIC FORCE ALWAYS ACTS PERPENDICULAR TO DIR. OF MOTION v

EXAMPLE B = 10 3 Gauss m = 100 z = +1 m/z = 100 V = 2000 volts r m = 64 cm m z = 2 B r 2 V 2 m SCAN m/z BY SCANNING EITHER: B (vary mag. field) V (vary acc. voltage) AND / OR r m (array detector)

Electrostatic Analyzers (ESA)