Alternatives to Sub-2 µm UHPLC Columns

Similar documents
Changing the way you think about HPLC

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

Phenyl-Hexyl. UHPLC Columns. Alternate, complementary selectivity to C18 and C8 bonded phases

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Bringing uhplc Performance to the Separation of Peptides

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns

Separation of Structurally Similar Steroids on HALO C18 and PFP

Superficially Porous Silica Particles with Wide Pores for Biomolecular Separations

penta-hilic UHPLC COLUMNS

C18 Column. Care & Use Sheet

LC Technical Information

Breaking the speed limit: Fast, high-resolution peptide and tryptic digest separations using fused-core particles

Improve HPLC Separations by Carefully Matching Column Pore-Size to Solute-Size

Trends in Method Development for HPLC and UHPLC... the movement toward using solid-core particles

Impact of Pore Exclusion on Reversed-Phase HPLC Column Performance

penta-hilic UHPLC COLUMNS

Isomer Separation on ZirChrom -CARB

Small Molecule Selectivity Sampler

HPLC & LC-MS. Techniques to Improve Speed, Resolution and Sensitivity without Investing in Capital Equipment

High Resolution Fast LC

LC and LC/MS Column Selection Flow Chart

DEVELOPING HPLC METHODS WHEN C18 COLUMNS DON T WORK

Advantages and Applications of Revolutionary Superficially Porous Particle Columns in Liquid Chromatography

HICHROM. Chromatography Columns and Supplies. LC COLUMNS HALO and HALO-5. Catalogue 9. Hichrom Limited

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Liquid Chromatography

Analysis of Positional Isomers with Agilent Poroshell 120 PFP Columns

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

CHANGING THE WAY YOU THINK ABOUT HPLC

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Uptisphere CS Evolution Core Shell columns for fast & highly efficient identification & quantification of small molecules

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

FluoroPhenyl. Stationary Phase: Get the Power of HILIC and RP Modes in One LC Column. Pure Chromatography. Selectivity Accelerated

Ultra Performance Liquid Chromatography (UPLC): An Introduction

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

How proteins separate on reverse-phase HPLC

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Packings for HPLC. Packings for HPLC

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools

Chemistry Instrumental Analysis Lecture 28. Chem 4631

Separation Methods Based on Distributions in Discrete Stages (02/04/15)

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Thermo Scientific Syncronis HPLC Columns. Remarkable separations guaranteed time after time

Open Column Chromatography, GC, TLC, and HPLC

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

FluoroPhenyl. Stationary Phase: Get the Power of HILIC and RP Modes in One LC Column. Pure Chromatography. Selectivity Accelerated

Hypersil BDS Columns TG 01-05

Recent developments in support materials for use in high performance liquid chromatography.

Reversed Phase Solvents

Too Polar for Reversed Phase What Do You Do?

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

LIQUID CHROMATOGRAPHY

Alltech Alltima HP Introduction

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

High Temperature Ultra Fast Liquid Chromatography. Peter W. Carr, Jon Thompson, Dwight Stoll, and Adam Schallinger

High Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

Validated HPLC Methods

Chromatographic Analysis

High Performance Liquid Chromatography

Basic Principles for Purification Using Supercritical Fluid Chromatography

A Look at Agilent HPLC Columns Choices

How to Troubleshoot a Failed System Suitability Test

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

for Acclaim Mixed-Mode HILIC-1 Column

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Fast HPLC Theory and Practice of Fast HPLC

Method Development Kits

Modernizing the USP Monograph for Acetaminophen

Choosing Columns and Conditions for the Best Peak Shape

Automated LC Method Development: Fundamentals and Columns, Hardware, and Software

New 3-Micron Polysaccharide-based Chiral Columns for Fast HPLC and SFC Geoffrey B. Cox, Norbert M. Maier Chiral Technologies, Inc

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

synergi Ultra-Performance HPLC The Concept Behind Synergi: Separate the Entire Polarity Spectrum Material Characteristics

Chromatography. Gas Chromatography

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

ANALYSIS OF BASIC COMPOUNDS

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

CHEM340 Tutorial 4: Chromatography

NANOLCMS SOLUTIONS HPLC BASICS

Nexera UC Unified Chromatography

SELECTING HPLC COLUMNS: GOING UNDER THE HOOD. HPLC Column Selection Guide

Biochemistry. Biochemical Techniques HPLC

Forensic and toxicological drug screening using liquid chromatography-tandem mass spectrometry with MS/MS library based identification

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Stability of Silica-Based, Monofunctional C 18 Bonded- Phase Column Packing for HPLC at High ph

SRT -10/10C (10 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC. Sepax Technologies, Inc. Preparative Size Exclusion Chromatography

High Performance Liquid Chromatography

Part 2. Overview of HPLC Media

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003

Transcription:

Alternatives to Sub-2 µm UHPLC Columns J. J. DeStefano, B. E. Boyes, S. Schuster, W. L. Miles, and J. J. Kirkland Advanced Materials Technology, Inc. 3521 Silverside Rd., Quillen Bld., Ste. 1-K Wilmington, DE 19810 USA

Why Use Sub-2µm Particles? Smaller particles improve efficiency allowing faster separations Efficiency is directly proportional to the reduction in particle size N ~ (1/d p ) High efficiency in short columns Fast method development Short run times Improved productivity Less solvent usage Sharper peaks for more sensitivity

Are Sub-2µm SPP Needed When Separating Small Molecules? SPP shown to have unusually high efficiency 2.6 2.7 µm SPP have efficiency of sub-2µm TPP Theory predicts efficiency advantages of smaller SPP particles Sub-2µm SPP already available General consensus is Yes

Downside of Using Sub-2µm Particles Pressure goes up as the square of the reduction in particle size P ~ (1/d p ) 2 Specially designed (expensive) instruments required for optimum use 400 600 bar often insufficient for optimum flow Low-dispersion design required to minimize extra-column effects for highest efficiency Small ID tubing and flow cells significantly add to operational pressure Maintenance is expensive and often not userfriendly

Downside of Using Sub-2µm Particles Column frits with small pores (0.2 0.5µm) required to retain particles in columns More subject to plugging than 2µm frits Additional efforts needed to avoid particulate fouling (filter samples and mobile phases) Frictional heating of columns More pronounced as d p is reduced Can result in band-broadening and changes in retention 3 mm i.d. columns required to minimize frictional heating effects Columns may not exhibit expected efficiency or stability Small particles harder to pack into homogeneous beds for highest efficiency

Effect of Particle Size on H vs v Plots 8.0 7.5 4.6 µm Plate Height, micrometers 7.0 6.5 6.0 5.5 5.0 4.5 4.0 4.1 µm 2.7 µm 2.0 µm 3.5 Data fitted to Knox equation 3.0 1 2 3 4 5 6 7 Linear Mobile Phase Velocity, mm/sec The Plate Heights of columns packed with SPP particles of different sizes, as expected, get smaller as the particle size gets smaller.

Effect of Particle Size on h vs v Plots 2.6 2.4 2.2 Reduced Plate Height 2.0 1.8 1.6 1.4 2.0 µm 2.7 µm 4.1 µm 4.6 µm 1.2 1.0 1 2 3 4 5 6 7 Linear Mobile Phase Velocity, mm/sec Data fitted to Knox equation Reduced Plate Heights (h = H/d p ) get smaller as the particle size is increased, indicating less homogeneity in packed beds for the smaller particles.

For Superficially Porous Particles (SPP) (assumes h = 1.6) particle size column length theoretical plates pressure time 5 microns SPP 250 mm 31,200 200 bar 1.0

For Superficially Porous Particles (SPP) (assumes h = 1.6) particle size column length theoretical plates pressure time 5 microns SPP 250 mm 31,200 200 bar 1.0 2.7 microns SPP 150 mm 34,700 410 bar 0.6

For Superficially Porous Particles (SPP) (assumes h = 1.6) particle size column length theoretical plates pressure time 5 microns SPP 250 mm 31,200 200 bar 1.0 2.7 microns SPP 150 mm 34,700 410 bar 0.6 1.7 microns SPP 100 mm 36,700 700 bar 0.4

For Superficially Porous Particles (SPP) (assumes h = 1.6) particle size column length theoretical plates pressure time 5 microns SPP 250 mm 31,200 200 bar 1.0 2.7 microns SPP 150 mm 34,700 410 bar 0.6 1.7 microns SPP 100 mm 36,700 700 bar 0.4 1.3 microns SPP 75 mm 36,000 900 bar 0.3

What are the Alternatives to Operating at High Pressures?

Method Development Needed When Peaks of Interest are not Fully Separated 1 2 3 4 5 6 7 8 R s = 1/4 [(N) 1/2 ][(α -1)/α][k 2 /(1+k ave )]

Resolution Equation Shows that Selectivity is More Effective Parameter to Change R s 1 1 k2 N 4 1 k Resolution (R) 3.0 2.5 2.0 1.5 1.0 0.5 N k 0.0 1.00 1.05 1.10 1.15 1.20 1.25 0 5000 10000 15000 20000 25000 0 5 10 15 20 25 Resolution (Rs) 45.0 40.0 35.0 30.0 25.0 20.0 15.0 10.0 Source: Jun Mao, PhD Thesis with Professor Peter Carr, U. of Minnesota, 2001 5.0 0.0 R s vs. alpha is nearly linear between 1 < < 2 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 Alpha Rs

Most Effective Parameters to Change Selectivity The analysis condition parameters that most affect selectivity, are 1 : Column type (C18, phenyl, amide, etc.) ++ B-solvent (acetonitrile, methanol, etc.) ++ Mobile phase ph ++ Ion-pair concentration ++ %B solvent/gradient steepness + Column temperature + Buffer concentration + more effective less effective 1 adapted from Introduction to Modern Liquid Chromatography, 3 rd Edition, L. R. Snyder, J. J. Kirkland, J. W. Dolan; p. 29, 2010, John Wiley & Sons, Inc.

Absorbance Change Bonded Phase to Vary Selectivity Test Conditions: Column size: 4.6 x 50 mm A = 25 mm Ammonium Acetate B = Acetonitrile Flow rate = 1.5 ml/min. Gradient = 34 63 %B in 3.5 min. Pressure = 200 bar Temperature = 35 C UV=254 nm, 1 µl Injection Peak Identities: 1. Oxazepam 2. Lorazepam 3. Nitrazepam 4. Alprazolam 5. Clonazepam 6. Temazepam 7. Flunitrazepam 8. Diazepam 1 2 3 4 5 6 7 8 HALO C18 1 3 2 4 5 6 7 8 HALO Phenyl-Hexyl 1.5 2 2.5 3 3.5 4 Time (min.)

Change Organic Modifier to Vary Selectivity

Method Development Recommendations 1. Use short, high-efficiency 2.7 µm SPP C18 columns to initially evaluate separation - 2.7 µm gives high efficiency at moderate pressures - Short columns give short run times for rapid method development - SPP shown to have ~40% efficiency advantage over totally porous particles of same size - C18 phases are rugged and effective for RPLC 2. Increase efficiency (longer column, smaller particles) if close to adequate resolution, if not: 3. Change Selectivity of the separation - Modify mobile phase (type of organic modifier, ph, etc.) - Change C18 phase to other type bonded phase

An Alternative to Sub-2µm Particle Columns 2µm SPP Solid Core 1.2 µm 2 µm 0.4 µm Shell with 90 Å pores

Particle Size Distributions 1.7 µm totally porous 2.0 µm superficially porous

For Superficially Porous Particles (SPP) (assumes h = 1.6) particle size column length theoretical plates pressure time 5 microns SPP 250 mm 31,200 200 bar 1.0 2.7 microns SPP 150 mm 34,700 410 bar 0.6 2 microns SPP 100 mm 31,200 500 bar 0.4 1.7 microns SPP 100 mm 36,700 700 bar 0.4 1.3 microns SPP 75 mm 36,000 900 bar 0.3

Aborbance @ 254 nm HALO 2 C18 vs. solid-core sub-2-µm column Columns: 2.1 x 50 mm Instrument: Shimadzu Nexera Injection Volume: 0.2 µl Detection: 254 nm Temperature: 25 C Mobile Phase A: water Mobile Phase B: acetonitrile Ratio A/B: 15/85 Flow rate: 0.5 ml/min Peak Identities: 1. Uracil 2. Pyrene 3. Decanophenone 4. Dodecanophenone 1 2 3 4 HALO 2 µm C18 N = 15500 P = 190 bar Plates/Bar = 82 47% higher pressure at the ~ same N SPP 1.6 µm C18 N = 15800 P = 280 bar Plates/Bar = 56 0.0 0.5 1.0 1.5 2.0 2.5 Time, min

Explosives: HALO 2 C18 vs. Non-Core Sub-2- m An average of 57% taller peaks at 15% lower pressure Temperature: 42 o C Flow rate: 0.4 ml/min Mobile Phase A: Water Mobile Phase B: Methanol Ratio A/B: 72/28 Detection: PDA @ 254 nm Peak Identities: 1. HMX 2. RDX 3. 1,3,5-Trinitrobenzene 4. 1,3-Dinitrobenzene 5. Nitrobenzene 6. Tetryl 7. 2,4,6-Trinitrotoluene 8. 2-Amino-4,6-Dinitrotoluene 9. 4-amino-2,6-dinitrotoluene 10. 2,4-Dinitrotoluene 11. 2,6-Dinitrotoluene 12. 2-Nitrotoluene 13. 4-Nitrotoluene 14. 3-Nitrotoluene

An Alternative to Sub-2µm 2µm SPP 2µm SPP keeps pressure within a comfort zone and retains most of advantages of sub-2µm columns Higher efficiencies than sub-2µm TPP columns Lower pressure than sub-2µm columns (TPP or SPP) Short columns exhibit the high efficiencies wanted for fast method development Minimizes disadvantages of sub-2µm columns Greater efficiencies than sub-2µm TPP with lower pressure requirements Similar efficiencies as sub-2µm SPP with lower pressure requirements Uses 1-micron frits that are less prone to plugging Reduced frictional heating

Are Sub-2µm SPP Needed for Small Molecules? Our conclusion: not necessary Advantages of very small particles are not sufficient to overcome the disadvantages for most small molecule applications Selectivity manipulations via bonded-phase or mobile phase are more effective at improving resolution than is increasing efficiency Conclusion may be different for large molecules. Large molecules may require shorter diffusion paths of small particle size SPP for adequate mass transfer

Conclusions Sub-2 µm SPP not needed for most routine small molecule applications Larger SPP are less problematic for high throughput operation (e.g., QC Labs) Columns of 2-µm SPP appear to be a good compromise of speed and efficiency with superior advantages for small molecule applications

Acknowledgements Thanks go to Robert Moran who supplied much of the data with 2-µm particles for this presentation