FEASIBILITY OF IN SITU TXRF

Similar documents
Laboratory and Synchrotron Radiation total. reflection X-ray fluorescence: New Perspectives in. Detection Limits and Data Analysis

CHARACTERIZATION OF SILICON WAFER SURFACES WITH SR-TXRF

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

X-ray Spectroscopy. Interaction of X-rays with matter XANES and EXAFS XANES analysis Pre-edge analysis EXAFS analysis

Inelastic soft x-ray scattering, fluorescence and elastic radiation

X- ray Photoelectron Spectroscopy and its application in phase- switching device study

Beamline practice at BL01B1 (XAFS) In-situ XAFS measurement of catalyst samples

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

Introduction to XAFS. Grant Bunker Associate Professor, Physics Illinois Institute of Technology. Revised 4/11/97

Supporting Information. Chlorine in PbCl 2 -Derived Hybrid-Perovskite Solar Absorbers

Lecture 23 X-Ray & UV Techniques

Portable type TXRF analyzer: Ourstex 200TX

CHEM*3440. X-Ray Energies. Bremsstrahlung Radiation. X-ray Line Spectra. Chemical Instrumentation. X-Ray Spectroscopy. Topic 13

Interaction of particles with matter - 2. Silvia Masciocchi, GSI and University of Heidelberg SS2017, Heidelberg May 3, 2017

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

How Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source

NEXAFS data normalization procedure: A case study at the C K edge *

Auger Electron Spectroscopy

Ma5: Auger- and Electron Energy Loss Spectroscopy

X-Ray Emission and Absorption

Auger Electron Spectroscopy (AES)

Structural Characterization of Giant Magnetoresistance Multilayers with New Grazing Incidence X-ray Fluorescence

Interaction X-rays - Matter

Small-Angle X-ray Scattering (SAXS)/X-ray Absorption Near Edge Spectroscopy (XANES).

ANALYSIS OF SYNCHROTRON RADIATION USING SYNRAD3D AND PLANS TO CREATE A PHOTOEMISSION MODEL

X-ray Fluorescence Imaging Following Synchrotron Beam Excitation

QUESTIONS AND ANSWERS

Electron Spectroscopy

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Energy Spectroscopy. Ex.: Fe/MgO

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE

Advanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7

Wolter Imaging On Z. Chris Bourdon, Manager Z Imaging and Spectroscopy Julia Vogel, LLNL; Ming Wu, SNL ICF Diagnostics Workshop, October 5 th 2015

Study of Distributed Ion-Pumps in CESR 1

Basics of Synchrotron Radiation Beamlines and Detectors. Basics of synchrotron radiation X-ray optics as they apply to EXAFS experiments Detectors

A DIVISION OF ULVAC-PHI. Quantera II. Scanning XPS Microprobe

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Secondary ion mass spectrometry (SIMS)

Energy Spectroscopy. Excitation by means of a probe

Development of a Compact XAFS Measurement Chamber under Atmospheric Pressure in the Soft X-ray Region

Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis

MS482 Materials Characterization ( 재료분석 ) Lecture Note 12: Summary. Byungha Shin Dept. of MSE, KAIST

EUV lithography and Source Technology

EXAFS. Extended X-ray Absorption Fine Structure

Why do I chose this topic? By: Medani Sangroula 12/6/2013 1

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5

Spin-resolved photoelectron spectroscopy

Methods of surface analysis

Secondary Ion Mass Spectrometry (SIMS)

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

Probing Matter: Diffraction, Spectroscopy and Photoemission

Applications of Synchrotron Radiation. Presentation at JASS02 Seminar; Jordan, Oct , 2002 Herman Winick, SSRL/SLAC, Stanford University

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

The Use of Synchrotron Radiation in Modern Research

X-Ray Photoelectron Spectroscopy (XPS)

Effective testing for wafer reject minimization by terahertz analysis and sub-surface imaging

Chapter 6. Fiber Optic Thermometer. Ho Suk Ryou

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

Applications of Synchrotron Radiation in Chemistry Zafar Hussain Ibupoto

Development of 2-Dimentional Imaging XAFS System at BL-4

Recent results of Synchrotron radiation induced Total Reflection X-ray Fluorescence Analysis at HASYLAB, Beamline L

X-ray Absorption Spectroscopy

Supplementary Information. Atomic Layer Deposition of Platinum Catalysts on Nanowire Surfaces for Photoelectrochemical Water Reduction

Appearance Potential Spectroscopy

Photoemission Spectroscopy

X-Ray Emission Spectroscopy

Auger Electron Spectroscopy (AES) Prof. Paul K. Chu

LOW-TEMPERATURE Si (111) HOMOEPITAXY AND DOPING MEDIATED BY A MONOLAYER OF Pb

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

LAB 01 X-RAY EMISSION & ABSORPTION

Chapter 4 Scintillation Detectors

X-Ray Photoelectron Spectroscopy (XPS)

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

PHOTOELECTRON SPECTROSCOPY (PES)

X-ray Absorption at the Near-edge and Its Applications

An Introduction to Diffraction and Scattering. School of Chemistry The University of Sydney

Photon Interaction. Spectroscopy

SAXS and SANS facilities and experimental practice. Clement Blanchet

PHI. Scanning XPS Microprobe

Biophysics Collaborative Access Team Basic Techniques for EXAFS Revision date 6/25/94 G.Bunker. Optimizing X-ray Filters

Damage to Molecular Solids Irradiated by X-ray Laser Beam

Removal of Cu Impurities on a Si Substrate by Using (H 2 O 2 +HF) and (UV/O 3 +HF)

USPAS course on Recirculated and Energy Recovered Linacs Ivan Bazarov, Cornell University Geoff Krafft, JLAB. ERL as a X-ray Light Source

Surface Sensitivity & Surface Specificity

Secondary ion mass spectrometry (SIMS)

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE

SYNCHROTRON RADIATION INDUCED X-RAY EMISSION - SRIXE W.M. KWIATEK

Lecture 22 Ion Beam Techniques

In-Situ FTIR Spectroscopy and Metrology of a Tungsten CVD Process

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

X-RAY MICRO-TOMOGRAPHY OF PORE-SCALE FLOW AND TRANSPORT. University of California Davis. Dorthe Wildenschild & Annette Mortensen

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

Introduction. Motivation

Studying Metal to Insulator Transitions in Solids using Synchrotron Radiation-based Spectroscopies.

Lecture 5. X-ray Photoemission Spectroscopy (XPS)

X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES)

PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy

Transcription:

FEASIBILITY OF IN SITU TXRF A. ngh 1, P. Goldenzweig 2, K. Baur 1, S. Brennan 1, and P. Pianetta 1 1. Stanford Synchrotron Radiation Laboratory, Stanford, CA 94309, US 2. Binghamton University, New York, NY 13905, US Abstract Synchrotron radiation based total external reflection x-ray fluorescence spectroscopy (TXRF) is now being routinely used at the Stanford Synchrotron Radiation Laboratory (SSRL) to carry out industrially relevant measurements. For transition elements on wafer surfaces, detection limits of 8x10 7 atoms/cm 2 that correspond to 1 impurity atom among 30 million substrate atoms have proven more than adequate for the studies undertaken to date. New studies, which aim to develop a better understanding of the kinetics of metal deposition during wafer cleaning processes, are being pursued. By using a sample cell with a water delivery system that has been custom designed, silicon samples can be analyzed directly beneath a water layer (in situ TXRF). Initial results measuring silicon samples with known concentrations are presented. Due to the increased scatter and background from the water layer along with the attenuation of fluorescence signals, minimum detection limits for the in situ TXRF setup increased to 8x10 10 atoms/cm 2, in close agreement to predicted values extracted from modeling calculations. Introduction For over 20 years, the device density of silicon chips has been increasing, leading to more stringent requirements for surface cleanliness. According to National Semiconductor Technology Roadmap developed by SEMATECH, transition metal impurities and light metallic elements (Z<14) must now be reduced to concentrations lower than 3x10 8 atoms/cm 2 and 5x10 9 atoms/cm 2 respectively, in order to achieve acceptable yields with current processing technologies [1]. At beamline 6-2 at the Stanford Synchrotron Radiation Laboratory (SSRL), these requirements have been met by applying synchrotron radiation based total reflection x-ray

fluorescence (SR-TXRF) to detect trace impurities on silicon wafer surfaces in a nondestructive manner [2]. Minimum detection limits (MDL) of 8x10 7 atoms/cm 2 (with a 1000 second spectrum) have been routinely achieved for transition metals, while an MDL of 8x10 9 atoms/cm 2 has been reached for aluminum using a bending magnet source [2,3]. Having exceeded the needs of the semiconductor industry for the minimum detection limit of transition metals, new experiments of studying silicon samples directly beneath a liquid layer have begun (in situ TXRF). rrently, the semiconductor industry uses a variety of cleaning procedures, but relatively little research has been done to understand the kinetics of these reactions. Some work has been done with ex situ techniques [6], but these studies often suffer from environmental interference and cannot resolve the kinetics that a real-time experiment can give. By observing samples directly beneath a liquid layer, it is believed that these cleaning mechanisms may be elucidated without environmental interference. With this in mind, a sample cell was designed with extreme cleanliness in mind and preliminary studies were conducted at beamline 6-2. Extending TXRF to samples beneath a water layer poses numerous challenges. First, the applied water layer attenuates fluorescence signals that are emitted from the surface of the sample. Moreover, it increases the dominant sources of background in TXRF: photoelectron bremsstrahlung and inelastic scattering. To minimize these effects it is imperative that the water layer above the sample be kept as thin as possible. Finally, since the metal contamination on silicon surfaces during silicon wafer processing usually happens on the timescales of minutes, the measuring time needs to be shorter than what is used in ex situ TXRF (usually 1000 seconds). Theoretical models simulating a 20 µm thick water layer and a 10-second measuring time predict a detection limit of 8x10 11 atoms/cm 2 (8x10 11 atoms/cm 2 for a 1000 second spectrum). Experimental Setup The current setup for the ex situ TXRF station uses an x-ray beam produced by a 54 pole wiggler on beamline 6-2. The beam is focused with a quartz mirror and tuned to an energy of 11.2 kev by using a double multilayer monochromator and the size of the beam is defined with slits to be

.25 mm x 3 mm (h x v). The sample is placed vertically with the incident beam at a glancing angle less than the critical angle for total external reflection (~0.16 deg). In an effort to reduce the observed elastic scatter, the detector is positioned along the polarization vector of the x-ray beam. The TXRF setup is shown in Figure 1. Figure 1 Schematic representation of the experimental station used at SSRL. The beam is focused onto the wafer using the focusing mirror and monochromatized by the multilayers. The detector is aligned along the polarization vector of the synchrotron radiation to minimize scatter. Using an in situ sample cell in place of the wafer chamber, it is possible to measure various samples beneath a water layer using a similar setup shown in Figure 1. The in situ cell used in these experiments is comprised of a machined Teflon chuck, upon which a clean PFA Teflon petri dish is placed. The sample is placed on the backside of the petri dish and an aluminized Mylar film covers it. A seal is created by using an o-ring in conjunction with a clamp. Teflon tubing is fed through the chuck and holes in the petri dish in order to provide a supply of water. A syringe pump is used with the inlet tubing, while the outlet tubing drains the water to a reservoir. By varying the flow rate of the syringe pump the thickness of the water layer can be controlled. A detailed figure of the cell design is shown in Figure 2.

Results Solutions spiked with 20 to 200 ppb were used to intentionally contaminate silicon samples that were initially measured in vacuum using ex situ TXRF for comparison. With the surface concentrations established, these samples were then measured using the in situ setup without Mylar or water. Shown below in figure 3 is a representative spectrum comparing the ex situ vacuum setup to the in situ air setup normalized such that the integrated and non-attenuated Kα fluorescence signature is the same for both spectra. As can be seen, the addition of an air volume results in additional elastic and inelastic scatter in the high energy part of spectrum as well as an additional Ar peak at 2958 ev. The silicon fluorescence signal from the substrate is also attenuated heavily due to the presence of the air and an additional Teflon filter that was used with the in situ setup. Finally, it should be noted that there was slight unintentional contamination when measuring the sample with the in situ setup resulting in a small but noticeable Fe peak. Figure 2 In situ cell design. The liquid is filled using a syringe pump, which delivers a constant flow rate of water. Figure 3 Two spectra comparing a sample in vacuum (ex situ setup) to a sample in air (in situ setup). Both spectra were normalized to the integrated Kα signal since this concentration was assumed to remain constant. The presence of air increases the scatter and background, attenuates the signal, and gives rise to an argon peak.

Next, the Mylar and water were applied to the sample with the in situ setup and the resulting spectrum overlaid with the ex situ vacuum measurement, shown in figure 4, was measured. As can be seen, the addition of Mylar and water resulted in Fe and Zn contamination and attenuation of the signal. Moreover, the elastic and inelastic scatter contribution continued to increase, all of which degrade our detection limits. Lastly, it should be noted that while this spectrum shows complete attenuation of the signal, other samples had a sufficiently high concentration to be visible beneath the water layer thereby allowing for quantification. Quantification Using the Lambert-Beer law as shown below, it is possible to calculate the water layer thickness by measuring the attenuation of the signal due to the presence of the water layer. I = I o * exp[ µ * x] where Io is the incident flux, µ is the linear absorption coefficient, and x is the thickness of the sample. For a flow rate of 2 ml/min, a water layer thickness of 10 µm is calculated. nce the water thickness is known, it is possible to calculate the concentration by normalizing the signal to a known fluorescence signal present on the sample. The known signal used is the silicon peak, as its surface concentration is well known to be 3x10 15 atoms/cm 2. The fluorescence signal from an impurity Z is given by [5]: S = N * X * σ ( E ) * ω * κ * T ( E ) d Ω z ph z z z z z * where N ph is the incident photon flux (photons/s), σ z is the photo-ionization cross-section, ω z is the fluorescence yield, κ z is the relative transition probability of Kα to Kβ, T is the transmission factor, Ε z is the fluorescence x-ray energy, and dω is solid angle subtended by detector. Normalizing the signal to the silicon signal removes the dependence of constant terms such as incident flux and solid angle subtended and gives: S S = X X * σ ( E) * ω * κ * T ( E * σ ( E) * ω * κ * T ( E ) From this equation, concentrations for several spectra can be calculated and are in close agreement to values measured with the ex situ SR-TXRF station with no water layer present. )

From these calculations, the MDL for the in situ TXRF setup with a 10 µm thick water layer is 7x10 10 atoms/cm 2 (for a thousand second spectrum), well within the range for kinetic studies. To insure that impurities from the ultra pure water (UPW) are negligible compared to the samples being studied, clean samples were measured with the ex situ TXRF setup before and after dipping into a bath of UPW. The results are shown in figure 5, and as can be seen, trace impurities of, Fe, and Zn were present in concentrations of 5.4x10 8, 1.4x10 9, and 4.7x10 8 atoms/cm 2, corresponding to a 10 ppt level concentration in the water itself [4]. nce the solutions used to contaminate samples measured with the in situ experiments were all above 20 ppb, it can safely concluded that contamination from the water to the sample is insignificant. Figure 4 Two spectra comparing in situ versus ex situ TXRF. Due to unintentional contamination from the mylar layer, Ca, Fe, and Zn peaks are present. Moreover, the peak is completely attenuated due to the presence of the water layer. Figure 5 The cleanliness of the ultra pure water was tested by measuring silicon samples before and after dipping them into a bath of ultra pure water. As can be seen, trace amounts of Fe,, and Zn were deposited onto the silicon surface after dipping. Conclusion From these early experiments it is apparent that further cleanliness is required as Zn and Fe contamination from the environment interferes with the in situ spectra. Especially detrimental is the Zn signal due to its proximity to the signal being measured. The development of a cleanroom environment for sample preparation along with clean chemicals will aid in this endeavor.

After developing a clean environment, it will also be possible to use spiked solutions to observe the deposition of other transition metals as a function of time. Models showing the enhanced deposition of in the presence of Fe ions have been proposed by Homma et al [6] and in situ studies could confirm this result. Finally, by extending this technique, it will be possible to conduct in situ x-ray absorption near edge spectroscopy (XANES) to determine the chemical state of the deposited on the silicon. Homma et al [4] have proposed that the presence of dissolved oxygen disturbs the reductive deposition of while enhancing hydroxide formation. nce the chemical shifts of the various oxides are well characterized, measuring silicon samples with spiked solutions using XANES will provide useful results. References [1] SEMATECH Roadmap Meteorology, http://notes.sematech.org/1997pub.htm [2] P. Pianetta, K. Baur, A. ngh, S. Brennan, J. Kerner, D. Werho, and J. Wang, Application of Synchrotron Radiation to TXRF Analysis of Metal Contamination on licon Wafer Surfaces, Thin Solid Films, v.373, 222 (1999). [3] K. Baur, J. Kerner, S. Brennan, A. ngh, and P. Pianetta, Aluminum Impurities in licon: Investigation of X-ray Raman Scattering in Total Reflection X-ray Fluorescence Spectroscopy, Journal of Applied Physics, v.88, 4642 (2000). [4] T. Homma, J. Tsukano and T. Osaka, Deposition Mechanism of Trace Metals on licon Wafer Surface in Ultra Pure Water, xth International Symposium on High Purity licon,v. 00-17, 670-676 (2000). [5] N. Takaura, Synchrotron Radiation Total Reflection X-Ray Fluorescence of licon Wafer Surfaces, Department of Materials Science and Engineering, Thesis, Stanford University (1997). [6] T. Homma, J. Tsukano and T. Osaka, Induced Codeposition of Trace Metals on H- (111) Surface of Buffered Fluoride Solutions, Third Electrochemical Technology Applications in Electronics Proceedings, v. 99-34, 95-101 (1999). [7] J. Stohr, NEXAFS Spectroscopy, Springer Series in Surface Science, v. 25 (1992).