Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Similar documents
Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Chloramine (Mono) and Nitrogen, Free Ammonia

Quaternary Ammonium Compounds

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Method to 0.50 mg/l NH 3 N Powder Pillows

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Method to 0.50 mg/l NH 3 N Powder Pillows

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Ammonia Gas Sensing Combination Electrode

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Ammonia Gas Sensing Combination Electrode

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

PAR Method Method to 2.0 mg/l Pb TNTplus 850

DR/4000 PROCEDURE. IRON, Total

Nitrogen, Total Inorganic

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Oxygen Demand, Chemical

Trihalomethane Formation Potential (THMFP)

Oxygen Demand, Chemical

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Chlorine, Free and Total, High Range

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

DR/4000 PROCEDURE. HARDNESS, Total

Chlorine, Free and Total, High Range

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

NITRITE, Low Range (0 to mg/l NO 2- -N)

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Chlorine, Free and Total, Low Range

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

CHLORINE, TOTAL (0 to 4.00 mg/l)

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Operating Instructions for Ammonium ISE Specifications

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

TPH (Total Petroleum Hydrocarbons)

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

NITROGEN, TOTAL, HR, Test N Tube

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

NITROGEN, TOTAL, Test N Tube * Method 10071

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Mercaptoacetic Acid Method*

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

GENERAL INSTRUCTIONS GENERAL PREPARATION

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

Polychlorinated Biphenyls (PCB) in Soil

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Ion Selective Electrode Probe

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

[Agilent Technologies] [Electrochemistry Products] Quick Reference

1) A ph/mv meter or an ion meter, either line operated or portable.

INSTRUCTION MANUAL. Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN Introduction Helpful Operating Tips. 2

DOC /2013, Edition 1. User Manual

NITROGEN, AMMONIA, High Range, Test N Tube

symphony Multisensor Probes

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

DPD Test N Tube Method *

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode

User Guide. High Performance Ammonia Ion Selective Electrode

Steam-Assisted Gravity Drainage Water Analysis Handook

Transcription:

, drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific information The Instrument-specific information table displays requirements that may vary between instruments. To use this table, select an instrument then read across to find the corresponding information required to perform this test. Table 452 Instrument-specific information Meter sension 4 meters 1 Electrode 5192000 1 Only sension 4 meters can be used for this analysis. Before starting the test: Refer to the meter user manual for meter operation. Refer to electrode manual for electrode maintenance and care. When using the electrode for the first time, condition the reference electrode for eight hours in a 100 mg/l nitrate nitrogen standard. After initial conditioning (before the first use or after long-term storage), condition the electrode for 60 minutes only, unless the electrode is stored dry for longer than four weeks. If the electrode is used daily, store the electrode in a nitrate standard having a concentration similar to the measured samples. The low range nitrate test requires several conditioning steps and careful attention to temperature. Read through the procedure steps and the Sample collection, preservation and storage section before proceeding. The 2-hour conditioning step must be completed before the calibration step. Sample concentration for this test must be 3.0 mg/l NO 3 N or less. Use a test strip to determine nitrate concentration. If the electrode is conditioned to a low nitrate-nitrogen concentration (as in this procedure), putting it in a concentrated NO 3 N sample (10 mg/l or higher) will swamp the electrode with nitrate ion and the 2-hour conditioning step must be repeated. Note A: If the electrode is conditioned properly, the mv potential should not change more than 0.1 mv every five minutes in step 7. It may be necessary to press the electrolyte dispenser button again to stabilize the potential reading. If the potential drifts one direction more than 0.2 mv every 5 minutes, leave the electrodes in the 150-mL beaker until the drift has slowed. If the drift is higher than 0.1 mv per minute, the electrode should be conditioned longer in 100 ml of 0.040 mg/l standard. Do not put ISA into the standard. Prepare the electrode. Refer to Electrode assembly and half-cell preparation in this procedure. Use identical amounts of ionic strength adjustor (ISA) in the standard beaker and in sample measurements. Use liquid ISA. Temperature variation causes inaccurate measurements. Calibration and sample measurements should be made at the same temperature ± 1 C. The stable sample reading is the nitrate-nitrogen (NO 3 N) concentration expressed as elemental nitrogen (N). The results can be expressed as mg/l nitrate (NO 3 ) by multiplying the results by 4.4. Page 1 of 10

Collect the following items: Description Quantity Ammonium sulfate reference electrolyte gel cartridge 1 half-cell internal filling solution ISA solution 0.5 ml 25 ml per sample or standard Electrode Membrane Tip (replacement) 1 Pipet, TenSette, 0.1 to 1.0 ml 1 Pipet, TenSette, 1.0 to 10.0 ml 1 -Nitrogen standard solutions: 10-mg/L as NO 3 N 100-mg/L as NO 3 N Water, deionized 25 ml 25 ml varies Beaker, 150-mL, polypropylene 3 Bottle, wash, 500-mL 1 Cylinder, graduated, 25-mL, polypropylene 3 Combination Electrode, BNC 1 sension 4 (laboratory) 1 Stir Bar, 7/8 X 3/16 in. (22.2 x 4.8 cm) 1 Stirrer, electromagnetic with stand and stir bar 1 Thermometer, Digital 1 100 ml plastic volumetric flask 2 See Consumables and replacement items for reorder information. Page 2 of 10

-Nitrogen, liquid ISA method 1. Condition the electrode in 100 mg/l NO 3 N standard for one hour (refer to halfcell preparation and Electrode Assembly in this procedure). Then condition the electrode in 0.010 mg/l NO 3 N for at least 2 hours. To make 100 ml of 0.010 mg/l NO 3 N, use a TenSette Pipet to deliver 0.10 ml of 10 mg/l NO 3 N into a 100 ml volumetric flask and dilute to the mark. Mix well. 2. Use a TenSette Pipet to pipet 5.0 ml of of nitrate liquid ISA into a plastic 100-mL volumetric flask. Bring to the 100-mL mark with deionized water. The deionized water must be at room temperature. 3. Pour the solution from the flask into a 150-mL beaker. Add a large stir bar (50.8 x 7.9 mm) to the beaker, place the beaker on an electromagnetic stirrer and begin stirring at a moderate rate. 4. Remove the electrode from the 0.010 mg/l nitrate nitrogen standard and rinse it well with deionized water. Place the electrode in the 150-mL beaker, submerging the tip below the solution surface. 5. Use a TenSette Pipet to add 0.4 ml of 10 mg/l nitrate nitrogen standard solution to the solution in the beaker (this makes 100 ml of 0.04 mg/l nitrate nitrogen). Allow the electrode to condition for 30 minutes in this solution before proceeding. 6. Turn the meter on. Press SETUP, set the electrode type to BNC and scroll to Stabilizing. Press ENTER and edit the display to show 0.1 mg/ min. Accept the value and EXIT the setup menu. 7. Press ISE MV until the mv potential shows on the display. Refer to Before starting the test: Note A. After the measurement is stable, record the potential value. Measure the temperature of the standard ( C) with a lab grade thermometer. Record the temperature. 8. Press ISE MV to toggle to concentration units. Press CAL and scroll to mg/l. Press ENTER and accept the concentration units. Page 3 of 10

-Nitrogen, liquid ISA method (continued) Repeat step 10 9. Edit the display to show the concentration of the solution in the 150 ml beaker (0.040 mg/l). Refer to the Low level nitrate calibration table. Press ENTER to accept the concentration. 10. When the measurement stabilizes, pipet the corresponding additional volume from the Low level nitrate calibration table. Wait the amount of time specified for step 2 in the table to allow the membrane to respond. Enter the concentration in mg/l (0.080 mg/l). Accept the concentration. If Error 2 occurs, use 0.08 mg/l as the first standard for this calibration. Then try conditioning in 0.04 mg/l NO 3 N solution for subsequent attempts. 11. Repeat step 10, adding the additional volumes of 10 mg/l and 100 mg/l NO 3 N standard from the Low level nitrate calibration table until all seven standards have been measured. Store the calibration and return to measurement mode. 12. Remove the electrode from the last standard, rinse well with deionized water, and blot dry. Save the solution in the 150-mL beaker with the 3.84 mg/l NO 3 N for later calibration checks. Page 4 of 10

-Nitrogen, liquid ISA method (continued) 13. Accurately pipet 5 ml of nitrate liquid ISA into a Class A 100-mL volumetric flask. Bring to the mark with the sample being measured. The sample must be at the same temperature as the standard solution in the 150-mL beaker used to perform the calibration. A white precipitate will form if chloride or other ions that react with silver are present in solution. This will not hurt the electrode or interfere with the analysis. 14. Pour this 100 ml of sample into a 150-mL beaker. Add a magnetic stir bar, place the beaker on an electromagnetic stirrer. Stir at a moderate rate. 15. Put the electrode into the sample and press the electrolyte dispenser button once. 16. Wait 5 10 minutes to allow electrode to condition to the low level nitrate in solution. Read the nitrate concentration from the display after it stabilizes. This is the sample nitrate concentration expressed as elemental nitrogen (NO 3 N ). Note: Record this value. Between uses, the electrode can be stored in the sample (if not an extreme ph). See electrode manual for details. Calibration Make sure to measure millivolt potentials of all nitrate standards at the same temperature ± 0.5 C. The sample and standards must be measured at the same temperature, ± 1 C. This procedure keeps temperature error to a minimum by using a spiked additions method of calibration. Even so, care should be taken to use the 100-mg/L NO 3 N standard at room temperature. Use a water bath slightly above room temperature (25 C) to equilibrate the standard temperature and sample temperature before measuring mv potentials. Use a laboratory-grade thermometer to monitor the temperature. A one degree centigrade difference may result in as much as a 0.4 mv inaccuracy. This temperature variation will, in turn, decrease accuracy of concentration measurements. Page 5 of 10

Table 453 Low level nitrate calibration table Step Volume 10 mg/l NO 3 N Volume 100 mg/l NO 3 N Concentration standard added standard added mg/l Time 1 0.4 ml 0.040 30 min. 2 0.4 ml 0.080 10 min. 3 1.2 ml 0.20 5 min. 4 0.2 ml 0.40 5 min. 5 0.4 ml 0.80 5 min. 6 1.2 ml 1.96 5 min. 7 2.0 ml 3.84 5 min. half-cell preparation Electrode assembly 1. Condition the nitrate membrane tip by immersing the membrane tip in beaker with 100 mg/l NO 3 N solution for approximately one hour prior to use.when using th electrode for the first time, condition the reference electrode for eight hours in a 100 mg/l nitrate nitrogen standard. 2. Clean all gel from the electrode tip and glass stem. Rinse with Deionized water and dry completely with a soft paper towel. 3. Use a soft paper towel soaked in isopropyl or rubbing alcohol to wipe the glass stem. Dry the stem with a dry, soft paper towel. Make sure that alcohol does not get in the glass stem. 4. Use the provided syringe to fill the membrane to with the electrode filling solution. Fill with solution to the top of the tip. Alternatively, fill the glass stem as shown in the electrode manual. Do not fill the stem completely or the pressure will cause the membrane tip to fall off. 5. Use a gentle, twisting action to carefully slide a clean, dry nitrate membrane tip over the glass stem until the end of the stem rests midway through the Membrane Tip and some resistance is met. Leave a 1 /8-inch gap between the tip and the electrode body. 6. Shake the electrode as if shaking down mercury in a thermometer to make sure the Electrode Filling Solution contacts the end of the nitrate membrane tip. Verify that air bubbles are not present in the tip. 1. Remove the cap from the electrolyte cartridge. 2. Visually inspect the Luer tip of the electrolyte cartridge. If air is present, rotate the feed-screw counter-clockwise until gel expels the air and fills the tip. 3. Fit the cartridge outlet tube firmly onto the inlet tube of the electrode body. 4. Place the dispenser unit over the electrolyte cartridge. Screw the dispenser unit onto the electrode body until reaching the stop. Do not over tighten. 5. Dispense the electrolyte gel by pressing the pump button. Repeat this procedure until gel is visible at the reference outlet. If readings become erratic make sure that the electrolyte gel is completely purged through the reference line. 6. Rinse the electrode with deionized water. Do not scrub the electrode tip. 7. Connect the BNC connector of the electrode to the BNC connector on the meter. Page 6 of 10

Interferences For the nitrate electrode, the major interferences include perchlorate, iodide, nitrite, bromide and chloride. The addition of ISA will eliminate most of these interferences. The highest level of chloride the ISA can accommodate is 40 mg/l Cl. Concentrations greater than 40 mg/l Cl in the sample will cause false high nitrate values. For more information on selectivity coefficients without ISA, refer to the Electrode Characteristics section of the electrode manual. Sample collection, preservation and storage Accuracy check Collect samples in clean plastic or glass bottles. Start nitrate measurements promptly after sampling. If storage is necessary, store for up to 24 hours at 4 C or lower (as per the USEPA MUR in March 2007, the storage criteria changed from 4 C to 0 6 C). Do not adjust sample ph. This method is for low ionic strength use only. Adjusting the ph with acid will change the ionic strength and make this method unusable. Checking electrode response To verify electrode response at these low levels of nitrate, the millivolt potential should decrease upon each addition of 100 mg/l NO3-N. Using the Low Level -Nitrogen procedure, at least a 5.0 mv drop should be observed from step 1 to step 2 (0.040 mg/l to 0.080 mg/l NO3--N). Each additional spike should decrease the mv reading substantially from the previous change. If this is not the case, check the purity of the standard. If this is not the problem, use a new membrane. Checking calibration accuracy 1. Pipet 1 ml liquid ISA into a 100-mL volumetric flask. 2. Fill the 100-mL volumetric flask to the mark with 1.0 mg/l NO 3 N. Pour this solution in a 100-mL beaker and add a stir bar. 3. Put the beaker on an electromagnetic stirrer and measure the concentration of the solution with the calibrated electrode. The measurement should be 1.0 mg/l ± 0.1 mg/l. Note: The beaker containing the 3.0 mg/l NO 3 N standard used in the calibration may also be used as a check on the calibration. It should read close to 3.0 mg/l NO 3 N ± 0.1 mg/l. Page 7 of 10

Checking the accuracy of the sample reading To verify sample measurement accuracy, add a spike of standard NO 3 N solution with a TenSette or volumetric pipet. Use the Spike volumes of 100-mg/L standard table and the formulas in Percent recovery. Table 454 Spike volumes of 100-mg/L standard Measured sample Concentration NO 3 N Volume of 100 ppm Standard to Add C x V 0.04 to 0.3 0.1 ml 10 0.30 to 0.60 0.3 ml 30 0.6 to 0.9 0.5 ml 50 0.9 to 3.0 1.0 ml 100 Percent recovery To calculate the percent recovery: M = S 100 + ( C V) E = ------------------- M 100 + V A R = --- 100% E Where: Method performance M = calculated mass of nitrate as nitrogen present after the spike (micrograms) S = mg/l of NO 3 N in sample (before spike) C = concentration of standard used for spiking (mg/l) V = spike volume from the Spike volumes of 100-mg/L standard table (ml) E = expected concentration after spiking (mg/l) R = percent recovery (should be 95 100%) A = actual reading on meter after spike (mg/l NO 3 N) Instrument sension 4 1 Standard Precision 95% Confidence Limits of Distribution 0.1 mg/l 0.095 0.1005 mg/l 1 With a stabilization criteria of 0.1 mv/min Page 8 of 10

Summary of method ions are selectively absorbed by the ISE membrane, establishing a potential (voltage) that is proportional to the concentration of nitrate in the sample. This potential is compared to the constant potential of a reference electrode by measuring the potential of known standard. A calibration curve can be constructed to determine the concentration of nitrate in unknown samples. The solvent-polymer membrane is a nitrate ion-exchanger in an inert polyvinyl chloride (PVC) plastic matrix. The nitrate electrode has an internal silver/silver chloride element, which establishes a fixed potential when in contact with the internal filling solution. The ion selective membrane undergoes ion exchange with nitrate in the sample, creating a potential across the membrane which varies with the amount of nitrate ion in the sample. This potential will decrease by about 58 mv for every tenfold increase in nitrate concentration in the linear operating range at 25 C. Consumables and replacement items Required reagents Description Quantity/Test Unit Catalog number Ammonium Sulfate Reference Electrolyte Gel Cartridge varies 2/pkg 2597102 Electrode Membrane Tips varies 16/pkg 4613300 -Nitrogen Standard Solutions 10 mg/l as NO 3 N varies 500 ml 30749 ISA liquid varies 500 ml 2488349 half-cell filling solution varies 50 ml 4456369 -Nitrogen Standard Solution 100 mg/l NO 3 N varies 500 ml 194749 test strips varies 25/pkg 2745425 Water, deionized varies 4 L 27256 Required apparatus Description Unit Catalog number Beaker, 150-mL, polypropylene each 108044 Bottle, wash, 500-mL each 62011 Pipet, TenSette, 0.1 1.0 m each 1970001 Flask, volumetric, poly, 100 ml each 1406042 Thermometer, Digital each 2630600 Flask, volumetric, 100 ml, polypropylene each 1406041 Combination Electrode, Platinum series, BNC each 5192000 sension 4 Laboratory ph/ise Meter each 5177500 Stir Bar, 7 /8 x 3 /16 in. (22.2 x 4.8 cm) each 4531500 Stir Bar, 1 1 /8 x 3 /10 in. (28.6 x 7.9 cm) each 2095355 Select one based on available voltage: Stirrer, electromagnetic, 115 VAC, with stand and stir bar each 4530001 Stirrer, electromagnetic, 230 VAC, with stand and stir bar each 4530002 Electrode Membrane Tips (replacements) 6/pkg 4613300 Page 9 of 10

Optional reagents Description Unit Catalog number Ionic Strength Adjustor, powder 454 g 4456301 Nitrogen Standard Solutions 1 mg/l as NO 3 N 500 ml 204649 Optional apparatus Description Unit Catalog number Beaker, 100 ml, polypropylene each 108042 Pipet Tips, for TenSette Pipet 50/pkg 2185696 Water Bath, circulating each 2616300 Pipet, Mohr, 2.00 ml, glass each 2093636 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company, 2007, 2010, 2012. All rights reserved. Printed in the U.S.A. Edition 7