A calibration facility for automatic weather stations

Similar documents
Components of uncertainty in measurements by means of AWS

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

Towards SI traceable humidity measurements with radiosondes

Comparative analysis of the influence of solar radiation screen ageing on temperature measurements by means of weather stations

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

GRUAN Station Report for Ny-Ålesund

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

Specifications for a Reference Radiosonde for the GCOS Reference. Upper-Air Network (GRUAN)

Humidity Calibration Solutions

From MeteoMet to MeteoMet2. and more. Andrea Merlone

Ny-Ålesund: Site Update & Atmosphere Flagship Program. Marion Maturilli, Sandro Dahlke, Roland Neuber, Andrea Merlone

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

Humidity Calibration Solutions

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

The Fundamentals of Moisture Calibration

Nanoholes for leak metrology

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

The AWS based operational urban network in Milano: achievements and open questions.

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba

Abstract. Introduction

473 Dew Point Hygrometer

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR

Calibration of Paroscientific Model 205 Pressure Sensor for use at Heard Island.

473 Dew Point Hygrometer

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

The calibration of Hygrometer (LectureandTraining )

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard

Application of the Multi-current Transient Hot-Wire Technique for Absolute Measurements of the Thermal Conductivity of Glycols

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

DEVELOPMENT OF MEASUREMENT STANDARD FOR DYNAMIC PRESSURE AT MIKES

Calibration of Temperature Block Calibrators

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS

Automated volume measurement for weihts using acoustic volumeter

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Introduction of RIC Tsukuba (Japan, RAII)

Introduction to Blackbody Sources

473-SHX Dew Point Mirror

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

CALIBRATION CENTER. Republic of Turkey. Ministry of Forestry and. Water Affairs. Turkish State Meteorological Service.

World Meteorological Organization

EURAMET Project 858: Hydrostatic weighing exchange of experiences. Project outline

CERTIFICATE OF CALIBRATION

CERTIFICATE OF ACCREDITATION

Edgetech Instruments Inc.

Renewal of the gage-block interferometer at INRIM

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Received: 17 December 2009 / Accepted: 11 March 2010 / Published online: 10 April 2010 Springer Science+Business Media, LLC 2010

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

for JMA/WMO Training Workshop on CALIBRATION AND MAINTENANCE OF METEOROLOGICAL INSTRUMENTS

Argon Triple-Point Device for Calibration of SPRTs

BIPM, EURAMET AND METEOMET

MEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G.

Introductions to RIC-Beijing. NAN Xuejing, CUI Xiai Meteorological Observation Center China Meteorological Administration March,2018

The AWS based operational urban network in Milano: achievements and open questions.

Chilled Mirror Hygrometry

Il nuovo Sistema Internazionale (SI) Le unità elettromagnetiche

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

He II Heat transfer through a Corrugated Tube - Test Report

Calibrating the Thermal Camera

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

Report of CoreTemp2017: Intercomparison of dual thermistor radiosonde (DTR) with RS41, RS92 and DFM09 radiosondes

DEVELOPMENT OF A FLOW-THROUGH CELL FOR ACCURATE MEASUREMENTS OF LOW ELECTROLYTIC CONDUCTIVITY

Supporting Information

Metrological requirements for a laboratory rainfall simulator. M. Colli* 1,2, M. Stagnaro 1,2, L. Lanza 1,2, P. La Barbera 1

A Case for Periodic Calibration or Verification of RTDs

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

DATA SHEET Metal Element Current Sensing Chip Resistor (Jumper) CLS Series

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators

2.1 OBSERVATIONS AND THE PARAMETERISATION OF AIR-SEA FLUXES DURING DIAMET

Final Report August 2010

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

REGIONAL INSTRUMENT CENTER (RIC) MANILA Philippines (Regional Association V)

Exploitation of ground based GPS for Climate and Numerical Weather Prediction applications COST action 716

AN EXPERIMENTAL STUDY OF THE FROST FORMATION ON A COLD SURFACE IN FREE CONVECTIVE FLOW

Development of procedures for calibration of meteorological sensors. Case study: calibration of a tipping-bucket rain gauge and data-logger set

OverviewofRIC Tsukuba (Japan,RAII )

European Project Metrology for Radioactive Waste Management

CERTIFICATE OF ACCREDITATION

Evaluation of Vaisala HMP45D Humidity Probes

Determination of Minimum Detectable Partial Pressure (MDPP) of QMS and its Uncertainty

Available online at ScienceDirect. Andrej Godina*, Bojan Acko

Climate & Earth System Science. Introduction to Meteorology & Climate. Chapter 05 SOME OBSERVING INSTRUMENTS. Instrument Enclosure.

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Standard Practices for Air Speed Calibration Testing

1 Phone , fax , 2 Phone , fax ,

Certificate of Accreditation

57:020 Mechanics of Fluids and Transfer Processes OVERVIEW OF UNCERTAINTY ANALYSIS

Measurement of Tritium in Helium

Ion Beam Sources for Neutral Beam Injectors: studies and design for components active cooling and caesium ovens

FROM SHAPE AND TEMPERATURE TO ENERGY CONVERSION ANALYSIS: A CASE STUDY

Transcription:

METEOROLOGICAL APPLICATIONS Meteorol. Appl. 22: 842 846 (2015) Published online 16 November 2015 in Wiley Online Library (wileyonlinelibrary.com) DOI: 10.1002/met.1514 A calibration facility for automatic weather stations G. Lopardo, a * S. Bellagarda, a F. Bertiglia, a A. Merlone, a G. Roggero a and N. Jandric b a Department of Thermodynamics, Istituto Nazionale di Ricerca Metrologica INRiM, Torino, Italy b Department of Thermodynamics, Institute of Metrology of Bosnia and Herzegovina, Sarajevo, Bosnia and Herzegovina ABSTRACT: Environmental monitoring by automatic weather stations (AWSs) is growing as a result of the increasing number and reliability of surface observations. In order to ensure data traceability and obtain more comprehensive data on the performance of AWSs, a new transportable calibration facility was manufactured at the Italian Institute of Metrology (INRiM) in the framework of the project MeteoMet. The facility is equipped with temperature and pressure reference sensors directly traceable to the International System of Units (SI) to obtain meteorological data with well-documented calibration uncertainty. In this calibration system, temperature and pressure can be controlled simultaneously and independently so that all combinations over the ranges are possible. The nominal ranges are: absolute pressure 50 to 110 kpa and temperature 25 to 50 C. The availability of a large range of atmospheric variability and the possibility of studying the mutual influence effects on sensors response are important characteristics of the facility. This apparatus is also designed to permit a control in humidity, in order to complete the characterization of the whole AWS pressure temperature humidity modulus. As a matter of fact, the final version of the facility will be equipped with a humidity generator for hygrometer calibrations. Finally, the calibration system is designed to be transportable, therefore allowing the calibration of AWSs located at sites that are difficult to access such as Ny-Ålesund (Svalbard) stations, where the facility was employed. The design and the technical characteristics are reported in this paper. KEY WORDS automatic weather station calibration; calibration chamber; pressure; simultaneous calibration of pressure and temperature; temperature; thermometer calibration in air; traceability for meteorological observations; uncertainty Received 12 September 2014; Revised 8 April 2015; Accepted 25 April 2015 1. Introduction The growing number of data recorded by automatic weather stations (AWSs) as replacement for conventional observation instruments requires assurance of reliability and quality for this kind of measurements. The metrological traceability to International System of Units (SI) through national standards is a milestone to address this objective. In order to provide data with good accuracy, it is important to define a standard calibration procedure to be carried out in chambers that keep measured quantities both stable during the calibration period at each measuring point and uniform in the calibration volume around the sensors (Lopardo et al., 2012). Currently, AWS sensors of pressure, temperature and humidity are usually calibrated separately (Saxholm and Heinonen, 2010). This kind of calibration procedure implies that mutual influences, for example the possible temperature-related behaviour of the pressure sensor, cannot be detected. A new transportable calibration system, Earth Dynamics Investigation Experiment 1 (EDIE1), capable of simultaneous and independent control of pressure and temperature, was developed at the Italian Institute of Metrology (INRiM). The system is composed of a test chamber, an external thermostat (PolyScience PD07R-40-A12E), pressure (Druck DPI 740) and temperature sensors (100 Ω Platinum Resistance Thermometers (Pt100)), a vacuum pump and a pressure regulator pump. The system diagram is illustrated in Figure 1. * Correspondence: G. Lopardo, Department of Thermodynamics, Istituto Nazionale di Ricerca Metrologica INRiM, Strada delle Cacce 91, Torino 10135, Italy. E-mail: g.lopardo@inrim.it This apparatus is also designed to allow the control in humidity, therefore completing the characterization of the whole AWS pressure temperature humidity modulus. In fact, the final version of the system will be equipped with a humidity generator for hygrometer calibrations in variable temperature conditions, also allowing the calibration of thermometers and barometers under different humidity values as the quantity of influence. At the moment, the development of a humidity generator is in progress and this paper reports only about pressure and temperature performances of the facility. This chamber has similar dimensions and similar auxiliary external fluid-based thermal control as other devices previously manufactured at the INRiM for thermal metrology (Merlone et al., 2007). The most important feature, with respect to other systems, is that EDIE1 allows temperature calibrations in air, instead of in a bath, in order to simulate the real operational conditions and characterize the sensor response more accurately. A characteristic of this facility is its ability to cover a wide range of atmospheric variability with one medium. The temperature is controlled between 25 and +50 C and the pressure between 50 and 110 kpa. This temperature range together with adjustable pressure conditions in the chamber gives extensive simulation scenarios for instruments and other components which are exposed to weather changes during operation. One of the biggest advantages of this apparatus is its reduced dimensions. The good compromise between the measuring inner chamber (inner diameter 220 mm and volume of about 15 l) and the external total dimensions (350 650 mm) makes it transportable for in situ calibration campaigns (Bertiglia et al., 2015), allowing at the same time the housing of the whole sensor and the datalogger. In this manner, the whole measurement 2015 Royal Meteorological Society

Calibration facility 843 PC Digital multimeter Test chamber Control valve EDIE1 calibration facility PRT Thermostat Vacuum pump Barometer Pressure regulator pump Figure 1. The system diagram of the Earth Dynamics Investigation Experiment 1 (EDIE1) calibration chamber and the acquisition system (digital multimeter and PC). (a) (b) Figure 2. (a) Test chamber without covers. The vacuum interspace hosting the steel pipe connections and the inner cylinder with thermalized copper cylinder are visible. (b) The test chamber and the external thermostat. The pressure controller and the vacuum pump are not shown in the picture. chain can be tested in working conditions to reduce further calibration uncertainty contributions due to the datalogger, Analog to Digital Converter and software elaboration. Today, the most used approach is a field verification, consisting of a travelling high-quality reference sensor, which is installed for some days in the proximity of operational sensors; then an inter-comparison trial is carried out. This test is not considered as a calibration process but only a check, allowing the identification of possible sensor malfunctions or larger deviations at one point (Lopardo et al., 2012). A robust metrological traceability is therefore missing when such a procedure is adopted. The use of this new facility can avoid these problems by also allowing metrological traceability for these AWSs installed in remote areas, where access difficulties can delay or prevent periodic calibrations (Merlone and Roggero, 2014). During June and July 2014, the EDIE1 calibration facility was shipped to the Ny-Ålesund research site, in the Norwegian polar area (Musacchio et al., 2014). The site hosts a GRUAN (Global Climate Observing System (GCOS) Reference Upper Air Network) station, so the facility was used for the calibration of those pressure and temperature measuring instruments used as ground check for the radiosonde sensors just before the launch. In this paper, the constructed apparatus and its performance characterization are described. Tests covering the temperature control in the whole range of interest are reported. 2. Test chamber The test chamber (Figure 2(a)) is formed by two concentric steel cylinders; between them, there is an interspace in which a vacuum is created as the thermal insulator. The inner cylinder is the test zone; it contains a copper cylinder on which a cooling

844 G. Lopardo et al. coil, directly connected to the external thermostat, is soldered (Figure 2(b)). The copper bottom and cover of this internal volume are to guarantee the best thermal uniformity with the walls. The thermostatic fluid flow is used for thermal control. A steel pipe, connecting the inner chamber with the outside, allows pressure control and is also used for permanent electrical wiring. The humidity control, not yet realized, will be obtained by a dedicated small saturator placed on the external part of the chamber; traceability will be guaranteed by a standard hygrometer directly traceable to Italian national primary standard. The humid gas generator will be based on a single-pass isothermal saturator, targeted to generate dew/frost point temperatures between 25 and 50 C. The reference relative humidity will be calculated by means of the dew/frost point temperature measured by a reference chilled-mirror hygrometer and the air temperature will be measured with a reference platinum resistance thermometer inside the EDIE1 chamber. The duct, through humid air fluxes, will be thermal-conditioned by long proximity with thermostatic coils in order to allow the simultaneous control of temperature and humidity. Air will leave the chamber through a short pipe near the top directly opened towards the laboratory, without any kind of pressure valve; in this case, the simultaneous control of pressure and humidity will not be possible. The temperature data were recorded by a digital multimeter, having a resolution of 8-1/2 digits (Model Keithley 2002), equipped with a scanner. The sampling time was 10 s. The contribution to standard uncertainty due to the nanovoltmeter readings was 0.004 C, as reported in Table 1 (Table 1, line 2). The repeatability is assumed to be the standard deviation of the readings along 10 min. In order to study the temperature vertical profile, three Pt100s were placed inside the test chamber. The measuring volume, where the major thermal uniformity is expected, is limited by the upper and lower thermometers and the middle thermometer is located in the centre of this area. The distance between each thermometer is 10 cm, the bottom one being placed at 17 cm from the bottom, the second at 27 cm and the top one at 37 cm. After a certain period, necessary to reach and stabilize the temperature, the readings of the three thermometers were compared averaged over 10 min of measurements. The thermal vertical gradients as a function of temperature in the two regions (between middle and upper thermometers, and lower and middle one) and in the whole measuring volume (between lower and upper thermometers) are reported in Figure 3. The upper volume (between middle and upper thermometers) turns out to be more uniform, 3. Performance characterization The pressure characterization of the test chamber was carried out through a Druck DPI 740 barometer calibrated against the INRiM secondary standard. The expanded uncertainty for pressure readings inside the chamber is 10 Pa (k = 2) in the range from 50 to 110 kpa. No pressure leaks were detected during the long-lasting tests in this range. The temperature characterization aims to determine the uniformity and stability of temperature in the test chamber, at different set points, in still air. The uniformity was tested in two different experiments: one test regarded axial uniformity and a second one dealt with vertical temperature changes along the measuring volume. The selected set points were 25, 10, 0, 10, 25, 40 and 50 C in order to cover the whole calibration range for atmospheric sensors. The measurements were performed with three Pt100 thermometers. All the Pt100s were calibrated in a bath against the INRiM secondary standards in the temperature range from 40 to 50 C; the maximum value of their calibration standard uncertainty was 0.011 C as reported in Table 1 (Table 1, line 1). Gradient ( C cm 1 ) 0.010 0.008 0.006 0.004 0.002 0.000 40 20 0 20 40 60 0.002 0.004 0.006 0.008 0.010 0.012 mid-up Low-mid low-up Temperature ( C) Figure 3. Thermal vertical gradients as a function of temperature. Gradients are calculated between the middle and upper thermometers (mid-up), between the lower and middle thermometers (low-mid) and between the lower and upper thermometers (low-up). Gradient is defined here as the ratio of the difference in temperature and the distance between the thermometers. Table 1. Summary of the sources of uncertainty for temperature measurements. The standard uncertainty is the quadratic sum of all uncertainty contributions and expanded uncertainties is the standard uncertainty multiplied for the coverage factor (k = 2) corresponding to a level of confidence of 95%. Uncertainty contribution Probability distribution Contribution to standard uncertainty ( C) PT100 thermometer Calibration uncertainty 0.02 C Normal 0.011 Fitting uncertainty 0.01 C Digital multimeter Resolution 0.005 mω Rectangular 7.5 10 6 Uncertainty 2.8 mω Rectangular 0.004 Repeatability 0.1 mω Normal 0.001 Chamber Axial uniformity 0.075 C Rectangular 0.021 Vertical uniformity 0.1 C Rectangular 0.028 Stability 0.01 C Normal 0.01 Standard uncertainty 0.038 Expanded uncertainty (k = 2) 0.076

Calibration facility 845 Gradient ( C cm 1 ) cent-rad 0.025 0.02 0.015 0.01 0.005 0 30 20 10 0 10 20 30 40 50 60 0.005 0.01 0.015 0.02 Temperature ( C) Figure 4. The axial gradient as a function of temperature. The gradient is defined as the ratio of the difference in temperature and the distance between the thermometers. Figure 5. Long-term stability test in a period of 12 h. The set point temperature was about 25 C. while the lower thermometer is more subject to external temperature perturbations. This figure also shows a different type of air temperature stratification for temperatures lower and higher than the ambient one. Regarding the axial uniformity, two Pt100s, located, respectively, at 5 and 1.5 cm from the centre of the measuring volume, were compared. Temperature differences between 0.004 and 0.075 C were detected in the calibration range. The axial gradient as a function of temperature is reported in Figure 4. As expected, the readings of the farthest thermometer from the centre are more influenced by the cooling coil: the temperature in the external area is higher than that in the central one at 50 Cand lower at 25 C. The temperature stability was determined at different temperatures, by recording the outputs of a thermometer located in the centre of the measuring volume. Figure 5 shows the record for a set point near 25 C. A temperature drift of 0.08 Cwas observed during a 12 h period. At +25 C, the maximum deviation during a 2 h period was 0.03 C. 4. Conclusions The calibration facility designed and manufactured at the Italian Institute of Metrology (INRiM) provides a suitable instrument for accurate in situ calibration of the automatic weather stations (AWSs), and in general of meteorological sensors, located in remote regions thereby allowing the metrological traceability to key observables for climate change evaluation. This calibration facility allows performance of calibrations in a wide range of atmospheric conditions (temperature: from 25 to 50 C, pressure: from 50 to 110 kpa); using a more powerful external thermostat may even extend the temperature range towards lower temperatures (from 50 to 50 C) to cover extreme atmospheric conditions such as those occurring in the Himalayas or the Polar regions. The combined pressure and temperature conditions, which are not possible were the devices tested or calibrated separately, are achievable in this apparatus. Besides, the final version of the facility, equipped with a humidity generator and a controller, will allow extension of the calibration and mutual influence analysis to humidity sensors. The test chamber characteristics allow to perform calibrations by comparison against the reference standards with limited uncertainty contributions: axial chamber uniformity contribution to standard calibration uncertainty: 0.075 C (maximum difference recorded in the calibration range) vertical chamber uniformity contribution to standard calibration uncertainty: 0.1 C (maximum difference recorded in the calibration range) chamber stability contribution to standard calibration uncertainty: 0.01 C (stability evaluated during a 1 h observation period).

846 G. Lopardo et al. Table 1 reports the contributions to the total calibration standard uncertainty for temperature measurements: Pt100 calibration uncertainty, uncertainty due to the digital multimeter and uncertainty due to chamber manufacturing characteristics. The expanded uncertainty is equal to 0.076 Cink = 2, corresponding to a confidence level of 95%. The expanded uncertainty (k = 2) for pressure is 10 Pa. The Earth Dynamics Investigation Experiment 1 (EDIE1) is the first example of transportable calibration facility for atmospheric sensors developed at the INRiM in the framework of the European Project MeteoMet (http://www.meteomet.org). It is now used for in-field calibration activities in co-operation with the weather and climate observation services and for scientific research. Its design and technical characteristics have allowed extension of the project to other specific needs of traceability. For example, a second chamber, EDIE2, dedicated to measurements in the Everest area, was developed. This chamber, working in a wider range of atmospheric variability and with smaller dimensions than those of the EDIE1, was installed on the EV-K2-CNR Everest Pyramid Observatory (Merlone and Roggero, 2014). Acknowledgements This work is being developed within the frame of the European Metrology Research Program (EMRP) joint research project ENV07 METEOMET. The EMRP is jointly funded by the EMRP participating countries within the EURAMET and the European Union. References Bertiglia F, Lopardo G, Merlone A, Roggero G, Cat Berro D, Mercalli L, et al. 2015. Traceability of ground based air temperature measurements: a case study on the meteorological observatory of Moncalieri (Italy). Int. J. Thermophys. 36: 589 601. Lopardo G, Marengo D, Meda A, Merlone A, Moro F, Pennecchi FR, et al. 2012. Traceability and online publication of weather station measurements of temperature, pressure and humidity. Int. J. Thermophys. 33: 1633. Merlone A, Iacomini L, Tiziani A, Marcarino P. 2007. A liquid bath for accurate temperature measurements. Measurement 40(4): 422 427. Merlone A, Roggero G. 2014. In situ calibration of meteorological sensor in extreme environment. Metrology for Meteorology and Climate Conference, 15 18 September 2014, Brdo, Slovenia. Musacchio C, Bellagarda S, Merlone A, Maturilli M, Graeser J, Vitale V, Viola A, Liberatori E, Sparapani R. 2014. Metrology activities in Ny-Ålesund (Svalbard). Metrology for Meteorology and Climate Conference, 15 18 September 2014, Brdo, Slovenia. Saxholm S, Heinonen M. 2010. A calibration system for PTU devices. Measurement 43: 1583 1588.