High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Similar documents
The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TOF LC/MS Combined with Advanced MassHunter Data Processing Software

Analysis of E.coli Tryptic Digest and Intact Protein Using the Agilent 1290 Infinity 2D-LC Solution with Diode Array Detection and Q-TOF LC/MS

Method Development in Comprehensive 2D-LC Finding the Most Orthogonal Separation Systems for RPLC RPLC Using Column and Solvent Screening

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Accurate Mass Analysis of Hydraulic Fracturing Waters

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Application Note. Edgar Naegele. Abstract

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Routine MS Detection for USP Chromatographic Methods

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Application Note. Authors. Abstract. Food Testing & Agriculture. Stephen E. Reichenbach University of Nebraska Lincoln, NE, USA

The Agilent 1290 Infinity II Multi Method Solution

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Measuring Drug-to-Antibody Ratio (DAR) for Antibody-Drug Conjugates (ADCs) with UHPLC/Q-TOF

with diode array detection

Agilent LC & LC-MS Cannabis Testing

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Application Note. Edgar Naegele. Abstract

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Modernizing the USP Monograph for Acetaminophen

Analysis of E.coli Tryptic Digest and Intact Protein Using the Agilent 1290 Infinity 2D-LC Solution with Diode Array Detection and Q-TOF LC/MS

Separation of Large and Small Peptides by Supercritical Fluid Chromatography and Detection by Mass Spectrometry

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

Agilent All Ions MS/MS

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

The new Agilent 1290 Infinity Quaternary LC

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract

Varian Galaxie Chromatography Data System for Preparative HPLC

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

UPLC Intact MASS Analysis Application Kit

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Agilent ESI and APCI sources: for polar to non-polar compounds

Screening and Quantitation of 240 Pesticides in Difficult Food Matrixes Using the Agilent 6545 Q-TOF LC/MS System

Fast methods for the determination of ibuprofen in drug products

MassHunter TOF/QTOF Users Meeting

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Shodex TM ODP2 HP series columns

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Analysis of PEGylated Proteins with Agilent AdvanceBio SEC Columns

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

The Theory of HPLC. Quantitative and Qualitative HPLC

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Quantitation of Cystine and Identification of Related Metabolites in White Blood Cells Using a High Resolution Accurate Mass LC/MS Approach

Transcription:

High-Resolution Sampling D-LC for Pharmaceutical Impurity Analysis Detection of Impurities Hidden Under the API Peak at Relevant Levels Application Note Small Molecule Pharmaceuticals and Generics Authors Susanne Stephan and Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany Abstract The analysis of impurities in low concentrations relative to an active pharmaceutical ingredient (API) is important for quality control of drug substances. When impurities are structurally similar to the API and the concentration difference is high, chromatographic separation and detection can be challenging. The Agilent 9 Infinity II D-LC Solution offers the possibility to switch easily between comprehensive (LCxLC), multiple heart-cutting (MHC), or high resolution sampling D-LC (HiRes D-LC). In this Application Note, HiRes D-LC is used to achieve the separation of two closely eluting compounds, one of which is present at very low concentration, and is hidden under the other more highly concentrated compound peak. Chlorodifluorobenzoic acids and deamidated insulin were analyzed as standard substances and a real sample, respectively.

Introduction Purity analysis of drug substances is of high importance regarding quality and patient safety. According to ICH Guideline Q3A(R), an impurity in a new drug substance should be reported above a threshold of. % relative to the API. Separation and detection of impurities at these levels can be challenging, especially when the impurity is structurally related to the active pharmaceutical ingredient (API). This Application Note demonstrates the use of high-resolution (HiRes) sampling D-LC for the separation and detection of impurities, exemplified for different isomers of chlorodifluorobenzoic acid at relative concentrations of % and. %. Precision and accuracy could further be improved for this application using the Agilent 9 Infinity II High Dynamic Range Diode Array Detection (HDR-DAD) Impurity Analyzer Solution as the D detector. Deamidated insulin is analyzed as a real sample to show the detection of a structurally similar impurity with HiRes sampling D-LC/MS. Here, D-LC has the additional benefit that the D separation acts as a desalting step, avoiding ion suppression caused by high salt amounts from the D buffer in the MS ion source 3. The D Chromatogram Creator for Agilent MassHunter is used for viewing and analyzing the D-LC/MS data in MassHunter. Experimental Equipment The Agilent 9 Infinity II D-LC Solution was composed of the following modules: Two Agilent 9 Infinity II High Speed Pumps (G7A) Agilent 9 Infinity II Multisampler (G77B) with Infinity II Sample Cooler (Option #) Two Agilent 9 Infinity II Multicolumn Thermostats (G7B) Three Agilent 9 Infinity II Diode Array Detectors (G77B) with 3.7 mm, -mm, and mm Max Light Cartridge Cells (G 8) Agilent 9 Infinity Valve Drive (G7A) with -position/-port-duo valve (D-LC valve head, G3A) Two Agilent 9 Infinity Valve Drives (G7A) with multiple heart-cutting valves (G-) equipped with -µl loops Cut no. 3 For the analysis of deamidated insulin, the Agilent 9 Infinity II D-LC Solution was coupled to an Agilent Q-TOF equipped with a Dual Agilent Jet Stream ESI source. To avoid high amounts of salt from the D buffer entering the ion source, a timetable was used to switch the MS diverter valve, directing the flow to waste for the first minute of each D analysis, as recommended in a previous Application Note. Software Agilent OpenLAB CDS ChemStation Edition Rev. C..7 SR [] with Agilent D-LC Software, Product Version A..3 [], and Agilent HDR-DAD ChemStation AddOn, Product Version A.. []. Agilent MassHunter Workstation Software LC/MS Data Acquisition, Version B.8., Build 8..8. Agilent MassHunter Workstation Software Qualitative Analysis, Version B.7., Build 7..7. Agilent D Chromatogram Creator for MassHunter, Rev... In HiRes sampling D-LC, target compounds can be determined by collecting several small fractions over a selected time range, covering the entire area of a peak from the D chromatogram, as shown in Figure. Each cut is parked in a sampling loop, and all cuts are analyzed in the second dimension consecutively. In this mode, it is ensured that all of the selected compound is transferred to, and analyzed in the second dimension. This way, selected coeluting compounds can be subjected to the HiRes sampling process for analysis in the second dimension. This enables a reliable quantification, as was shown in a Technical Overview. LC LC Cut Cut Cut 3 Cut Cut Cut Figure. Illustration of high-resolution sampling D-LC.

Chemicals All solvents were LC grade. Acetonitrile and sodium sulfate were purchased from Merck, Darmstadt, Germany. Formic acid, phosphoric acid, sodium phosphate dibasic, and ammonium phosphate monobasic were from Sigma-Aldrich, Steinheim, Germany. Fresh ultrapure water was obtained from a Milli-Q Integral system equipped with a.-µm membrane point-of use cartridge (Millipak, EMD Millipore, Billerica, MA, USA). Standards of -chloro, difluorobenzoic acid, 3-chloro-,-difluorobenzoic acid, -chloro-,-difluorobenzoic acid, and insulin from bovine pancreas were purchased from Sigma-Aldrich, Steinheim, Germany. Sample Stock solutions of chlorodifluorobenzoic acids were prepared at a concentration of mg/ml in / water/acetonitrile. Mixtures of the three compounds with different concentration ratios were prepared from these stock solutions. A mg/ml amount of insulin was dissolved in mm sodium phosphate buffer at ph.. To trigger deamidation, the insulin solution was adjusted to ph 9, and kept at room temperature for hours. High-resolution sampling D-LC method for the analysis of chlorodifluorobenzoic acids Columns First dimension Second dimension D Pump Solvent A Solvent B Flow rate D Detector: DAD with -mm Max-Light cartridge cell Agilent ZORBAX Eclipse Plus C8 RRHD,. mm,.8 µm (p/n 9978-9) Agilent ZORBAX Eclipse Plus PAH,. mm,.8 µm (p/n 997-98) Water +. % phosphoric acid Acetonitrile. ml/min Gradient minutes 3 %B minutes 3 %B minutes 8 %B minutes 8 %B minutes 3 %B D Pump Solvent A Solvent B Flow rate Water +. % phosphoric acid Acetonitrile. ml/min Gradient minutes %B minutes %B minutes %B D Gradient stop time. minutes D Cycle time. minutes Stop time High-resolution sampling minutes D Detector: HDR-DAD solution Time based. minutes 7.83 minutes Sampling time 7 seconds 9 seconds Number of cuts 9 Multicolumn thermostat First dimension 3 C Second dimension 3 C Multisampler Injection volume µl µl Needle wash seconds in methanol/water / D Diode array detector Wavelength Data rate D Detector Wavelength Data rate nm/ nm, reference 39 nm/ nm Hz DAD ( mm cell) nm/ nm, reference 39 nm/ nm Hz HDR-DAD 3

High-resolution sampling D-LC/MS method for the analysis of deamidated insulin (modified from) 3 Parameter Columns First dimension Second dimension D Pump Value Agilent Poroshell EC-C8. mm,.7 µm (p/n 9377-9) Agilent ZORBAX Bonus RP. mm,.8 µm (p/n 8778-9) Solvent A mm Na SO + mm NH H PO, ph. Solvent B Acetonitrile/water 8/ Flow rate.3 ml/min Gradient minutes - 9 %B 8 minutes - 3 %B 9 minutes - 9 %B 3 minutes - 9 %B 3 minutes - 9 %B D Pump Solvent A Solvent B Flow rate Gradient Water +.% formic acid Acetonitrile/water 8/ +.% formic acid. ml/min minutes -.% B. minutes - 3% B. minutes - 3% B.7 minutes - 9% B.8 minutes - 9% B.9 minutes -.% B D Gradient stop time 3. minutes D Cycle time. minutes Stop time High-resolution sampling Time based Sampling time 38 minutes. minutes s Number of cuts Parameter Value Multicolumn thermostat First dimension C Second dimension C Multisampler Injection volume µl Needle wash seconds in / methanol/water D Diode array detector Wavelength 9 nm/ nm, reference 39 nm/ nm Data rate Hz D Diode array detector Wavelength 9 nm/ nm, reference 39 nm/ nm Data rate Hz MS parameters Mode positive Gas temperature C Gas flow 3 L/min Nebulizer 3 psig Sheath gas temperature 37 C Sheath gas flow L/min VCap, V Nozzle voltage 3 V Fragmentor 7 V Skimmer V Oct RF Vpp 7 V Mass range 3, m/z Acquisition rate spectra/s MS diverter valve timetable Time segment no. Start time (min) MS. Waste 3 7. MS. Waste. MS. Waste 7 7. MS 8 3. Waste 9 3. MS 3. Waste 37. MS. Waste Diverter valve position Time segment no. Start time (min) 3. MS. Waste 7. MS. Waste 7. MS 8. Waste 9 7. MS. Waste. MS. Waste 3 7. MS Diverter valve position

Method setup for high-resolution sampling D-LC High-resolution sampling D-LC was performed with the Agilent 9 Infinity II D-LC Solution. The valve configuration (Figure ) consisted of a -position/ port-duo valve connected to two multiple heart cutting valves, holding sampling loops. With this setup, up to consecutive cuts can be sampled and stored until analysis. For HiRes sampling D-LC, a maximum loop filling of 8 % is recommended to prevent loss of sample. Figure 3 shows the method setup used for the D pump for the analysis of chlorodifluorobenzoic acids. First, a D-LC separation of the sample was run, and the chromatogram was loaded as a reference signal in the preview window. HiRes sampling was set up time-based, according to the peak of interest, with nine cuts for the chlorodifluorobenzoic acids covering the entire peak width. Under the given D conditions, a sampling time of 7 seconds equals a loop filling of 8 %. The higher injection volume applied with the Agilent 9 Infinity II HDR-DAD Impurity Analyzer Solution as the D detector required cuts and a higher sampling time of 9 seconds to cover the entire D peak width. This resulted in a loop filling of 7 %. For HiRes sampling of deamidated insulin, the method was set up following the same procedure. Ten cuts were taken with a sampling time of seconds, which equals a loop filling of %. 3 Waste IN 3 ¹D-Column OUT 3 7 8 3 IN 3 ²D-Pump ²D-Column Figure. Valve configuration of the Agilent 9 Infinity II D-LC Solution, holding sampling loops. OUT

Figure 3. Method setup for the D pump. Results and Discussion Analysis of chlorodifluorobenzoic acids Mixtures of the three isomers -chloro,-difluorobenzoic acid (Compound ), -chloro-,difluorobenzoic acid (Compound ) and 3-chloro-,-difluorobenzoic acid (Compound 3, structures in Figure ) in different concentration ratios were analyzed using HiRes sampling D-LC, first with the Agilent 9 Infinity II DAD with a -mm Max-Light cartridge cell as the D detector. OH O OH O Cl OH O F F Cl Cl F F F Compound -Chloro-,-difluorobenzoic acid Compound -Chloro-,-difluorobenzoic acid F Compound 3 3-Chloro-,-difluorobenzoic acid Figure. Chemical structures of the three analyzed chlorodifluorobenzoic acids.

Figure shows the D chromatograms of two mixtures containing all three compounds at the same low concentration level (blue), and in relative concentration levels of % of Compound 3 and. % of Compounds and, respectively (red). In the blue chromatogram, three separated peaks can be observed, whereas in the red chromatogram, Compound (. % relative concentration) is hidden underneath Compound 3 ( % relative concentration), and could not be detected as a separate peak. In D-LC analysis of these compounds, HiRes sampling was used to sample the entire width of peak 3, followed by consecutive analysis of nine cuts on the D column. Figure shows the sampling scheme of the D peak and the resulting D chromatograms of Cuts. In Cut, no peak was detected. Cut 3 shows two separated peaks of Compounds and 3, which overlapped in the D. Cut contains mainly Compound 3 and a very small amount of Compound. In Cut, only Compound 3 was detected. mau 3 7 Time (min) 8 9 Figure. D chromatograms of two mixtures in relative concentrations of % Compound 3,. % Compounds and (red curve), and. % of all three compounds (blue curve). Cut Cut 3 Cut Cut Figure. D chromatograms with cuts 9 sampled over the entire peak width of Compound 3 (top). D Chromatograms are shown for the analysis of Cuts, 3,, and. 7

To determine the usability of this method according to the reporting threshold given in the ICH guideline, different mixtures of the three chlorodifluorobenzoic acids were prepared, containing relative concentrations of % of Compound 3 and. %,. %, or. % of Compounds and. Peak areas of Compound 3 (3-chloro-,-difluorobenzoic acid) and Compound (-chloro-,difluorobenzoic acid) were calculated as the sum of D peaks in all cuts using the D-LC Viewer in Chemstation, as shown in Figure 7. Each mixture was analyzed in six replicates. Table shows the average peak ares, as well as the relative standard deviations (RSDs) calculated for the three different mixtures. The peaks of the lower-concentrated Compound have average signal-to noise ratios (S/N) of 3.,.9, and. for Mixes,, and 3, respectively, which is in the region of or below the limit of quantification (LOQ). This explains the relatively high RSD values of up to 7.9 % for the integrated peaks of Compound. Compound could be detected in a relative area ratio of.3 %,.7 %, and. % compared to Compound 3. Considering a response factor of.7, which was determined from D runs of both compounds, accuracy values of 7. 8. % can be calculated, which is already a good result in light of the high concentration difference between the main compound and the impurity. This demonstrates that separation and detection of impurities in APIs at relevant levels according to the reporting threshold (. %) given in ICH guideline Q3A(R) is possible using high resolution sampling D-LC for an impurity coeluting with the main peak in the first dimension. Figure 7. Results of the high-resolution sampling D-LC analysis of chlorodifluorobenzoic acids with peak areas of Compounds (-chloro-,-difluorobenzoic acid) and 3 (3-chloro-,-difluorobenzoic acid), as shown in the D-LC Viewer of Chemstation. Table. Average peak areas and relative standard deviations calculated for three different mixtures from six consecutive runs with the Agilent 9 Infinity II DAD with a -mm Max-Light cartridge cell as the D detector. The area ratio of Compound to Compound 3 is given in %. For calculation of accuracy, a response factor of.7, calculated from D peaks, was considered. 3-Chloro-,difluorobenzoic acid (Compound 3) -Chloro-,difluorobenzoic acid (Compound ) Area RSD (%) Area RSD (%) S/N Area ratio (%) Accuracy (%) Mix (.:.: %),8.. 3.39 7.9 3..3 7. Mix (.:.: %),.7.39 7.7 3..9.7 8.3 Mix 3 (.:.: %),8.79.3.7.7.. 8. 8

For further improvement of precision and accuracy, the Agilent 9 Infinity II HDR DAD Impurity Analyzer solution was used as the D detector. This solution combines the signals from two diode array detectors with Max-Light cartridge cells of different path length to increase the linear dynamic UV range. This enabled a higher injection volume without exceeding the linear UV range for the high-concentrated Compound 3. Average S/N ratios were all above the LOQ, with.7,., and 3. for Mixes 3. The area precision of Compound could significantly be improved, with RSD values below.8 % (Table 3). Accuracy was in the range of 9. 8. %, which is excellent for quantification. Analysis of deamidated insulin Bovine insulin was deamidated under basic conditions to generate a real sample of a pharmaceutically relevant substance containing impurities. The sample was analyzed with HiRes sampling D-LC/MS. The D Chromatogram Creator for MassHunter combines measurement data from Agilent OpenLAB CDS ChemStation Edition and MassHunter. The software creates D chromatograms, one chromatogram per cut in the first dimension, from the acquired full D Table 3. Average peak areas and relative standard deviations calculated for three different mixtures from six consecutive runs with the Agilent 9 Infinity II HDR-DAD Impurity Analyzer Solution as the D detector. The area ratio of Compound to Compound 3 is given in %. For calculation of accuracy, a response factor of.7, calculated from D peaks, was considered. Mix (.:.: %) Mix (.:.: %) Mix 3 (.:.: %) 3-Chloro-,- difluorobenzoic acid (Compound 3) -Chloro-,- difluorobenzoic acid (Compound ) Area RSD (%) Area RSD (%) S/N MS data. These chromatograms can be displayed in MassHunter similarly to what is offered by the D-LC viewer in OpenLAB CDS ChemStation Edition. The D UV chromatogram, enriched by a signal indicating cuts, can be displayed in MassHunter, as shown in Figure 8A. Figure 8B shows the full D-MS, and Figure 8C the full D-UV Area ratio (%) Accuracy (%) 7.7.9 3.3.8.7. 8. 9.3.8....9 99.9 898.7.3.37.7 3.. 9. Response units Response units Response units 3. A..8.....8... -. -. -. 8...8... 3.... -. -. 3789 3. 3. 3. 3.8.....8.....8.....8 7. 7. 7. 7. 7.8 8. 8. 8. 8. 8.8 9. 9. Acquisition time (min) B C 3 F 9 8 7 3 7 9 3 7 9 3 7 9 3 33 3 37 39 3 7 9 3 7 9 3 7 9 Acquisition time (min) Figure 8. Signals displayed in Agilent MassHunter for the analysis of deamidated insulin. A) D-UV signal with cuts taken in HiRes sampling. B) Full D-MS signal. C) Full D-UV signal. 9

signal in MassHunter. For each of the cuts, one individual chromatogram was created using the D Chromatogram Creator for MassHunter. These total ion chromatograms (TICs) are stacked for all cuts in Figure 9. During the first minute of each D separation, the flow was directed to waste by switching the MS diverter valve. Buffer salts contained in the fractions were excluded from the MS ion source in this manner. In each TIC in Figure 9, a step is visible at minute run time when the diverter valve is switched back, and the flow is directed to the MS. The main insulin peak can be observed with a retention time of. minutes at different intensities in Cuts. The D gradient includes a short section with 9 %B for flushing the column, leading to a rapid increase and decrease of the baseline at approximately 3. minutes. Figure shows a closer view of the TIC and the corresponding D-UV signal for Cut 7, which was taken from the first dimension in the tailing of the main insulin peak. In the D chromatogram, two peaks can be observed in both the MS and the UV signal. Figure shows that the mass spectra of both were extracted. The isotopic pattern of peak can be referred to as the [M+H] + ion of insulin. The mass spectrum of peak shows an isotopic pattern corresponding to a single deamidation product of insulin (neutral mass shift of.98 Da). The same compound could be detected in Cut (data not shown). Counts Cut - at. minutes 7 Cut - at.7 minutes 7 Cut -3 at.77 minutes 7 Cut - at.8 minutes 7 Cut - at.9 minutes 7 Diverter valve switch to MS Cut - at.97 minutes Insulin peak 7 Cut -7 at. minutes Cut -8 at. minutes Cut -9 at.7 minutes 7 Cut - at. minutes..3..7.9..3..7.9..3..7.9 3. 3.3 3. 3.7 3.9..3..7 Acquisition time (min) Figure 9. Individual D chromatograms of all cuts generated with the D Chromatogram Creator for Agilent MassHunter.

Response 9 8 7 A Cut 7 MS (TIC).7.39 Response - - - -8 B Cut 7 DAD (9 nm)..33...3....7.8.9....3....7.8.9 3. Acquisition time (min) Figure. D chromatogram of Cut 7, generated from the MS (TIC) and the UV signal (DAD). Counts 3 8 7 3 A Peak in Cut 7,.93,7.3,7.33,7.,7.7,7.99,8.,8.37,8.8 Counts 3..8.....8... B Peak in Cut 7,7.8,7.3,7.7,7.,7.9,8.3,8.33,8.,8.7,8.9,.8,7.,7.,7.,7.,7.8,8.,8.,8.,8.,8.8,9. Mass-to-charge (m/z) Figure. Mass spectra of peaks and observed in the D chromatogram of Cut 7.

Conclusions This Application Note demonstrates the detection of low-level impurities contained in a sample of a structurally related main compound using high resolution sampling D-LC. Using a mixture of chlorodifluorobenzoic acids, it is shown that compounds at relative concentrations of. %,. %,. %, and % according to the ICH Guideline about impurities in new drug substances, which cannot be detected separately in D analysis, are transferred to and separated on the D column by HiRes sampling D-LC. The compound -chloro-,-difluorobenzoic acid was detected with accuracy values of 7. 8. %. Area precision was.7 7.9 %. The Agilent 9 Infinity II HDR-DAD Impurity Analyzer Solution as the D detector allowed a higher injection volume, resulting in improved accuracy of 9. 8. % and area precision of..8 %. Deamidated insulin was used as a real sample for analysis with HiRes sampling D-LC/MS. Data were evaluated using the D Chromatogram Creator for Agilent MassHunter. A single deamidation product of insulin was detected and identified through its isotopic pattern in two cuts, which were taken from the first dimension in the tailing of the main insulin peak. References. Venkatramani, C. J.; et al. Assessing Stability-Indicating Methods for Coelution by Two-Dimensional Liquid Chromatography with Mass Spectrometric Detection. J. Sep. Sci., 37, 3 3.. International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use Impurities in New Drug Substances Q3A(R),. 3. Petersson, P.; Haselmann, K.; Buckenmaier, S. Multiple Heart-Cutting Two Dimensional Liquid Chromatography Mass Spectrometry: Towards Real Time Determination of Related Impurities of Bio-Pharmaceuticals in Salt Based Separation Methods. J. Chromatogr. A, 8, 9.. Stephan, S. High-Resolution Sampling D-LC with the Agilent 9 Infinity II D-LC Solution. Agilent Technologies Technical Overview, publication number 99-737EN,.. Schneider, S. D-LC/MS Characterization of Charge Variants Using Ion Exchange and Reversed Phase Chromatography Multiple Heart-Cutting D-LC Analysis of Innovator versus Biosimilar Monoclonal Antibodies. Agilent Technologies Application Note, publication number 99-73EN,. www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 7 Published in the USA, June, 7 99-88EN