Mass Spectrometry in MCAL

Similar documents
TANDEM MASS SPECTROSCOPY

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry

Lecture 15: Introduction to mass spectrometry-i

Analysis of Polar Metabolites using Mass Spectrometry

Quattro Micro - How does it work?

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

CEE 772: Instrumental Methods in Environmental Analysis

Introduction to the Q Trap LC/MS/MS System

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 34. Chem 4631

M M e M M H M M H. Ion Sources

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Translational Biomarker Core

MS Goals and Applications. MS Goals and Applications

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

Molecular Mass Spectrometry

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Harris: Quantitative Chemical Analysis, Eight Edition

Molecular Mass Spectrometry

LC-MS Based Metabolomics

Introduction to LC-MS

MS/MS .LQGVRI0606([SHULPHQWV

Mass spectrometry.

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

MS Goals and Applications. MS Goals and Applications

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

Introduction to GC/MS

Electrospray Ion Trap Mass Spectrometry. Introduction

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

Types of Analyzers: Quadrupole: mass filter -part1

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Mass Spectrometry. Electron Ionization and Chemical Ionization

Finnigan LCQ Advantage MAX

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Ion traps. Quadrupole (3D) traps. Linear traps

Extrel Application Note

Chemistry 311: Topic 3 - Mass Spectrometry

Interface (backside) & Extraction Lens

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

CHROMATOGRAPHY AND MASS SPECTROMETER

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

Mass Spectrometry. General Principles

Questions on Electric Fields

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

ICPMS Doherty Lecture 1

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

Electron Ionization Source Parameters in GC-MS

Ionization Techniques Part IV

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

BIOINF 4399B Computational Proteomics and Metabolomics

Ch. 22 Mass Spectrometry (MS)

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Thermo Finnigan LTQ. Specifications

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC)

Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

Mass spectrometry gas phase transfer and instrumentation

Technical Note. Introduction

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Basics of Mass Spectrometry

Quadrupole Mass Spectrometry Concepts. Mass spectrometers for residual gas analysis: Intermediate Level Users Guide

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

1. The range of frequencies that a measurement is sensitive to is called the frequency

2401 Gas (liquid) Chromatography

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results

Particle Position Relative Mass Relative Charge Proton Nucleus 1 +1 Neutron Nucleus 1 0 Electron Orbitals 1/ Atomic Symbol

Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4.

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Secondary Ion Mass Spectroscopy (SIMS)

Determination of Volatile Substances Proof of Food Adulteration

HPLC-MS of the Initial Perfluorinated Surfactants and Shorter Chains:

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION

Mass Spectrometry. What is Mass Spectrometry?

Introduction to Mass Spectrometry (MS)

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:

Vacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular

Space Charge Effects in Linear. Quadrupole Ion Traps

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

API Optimization, Source Design, and Other Considerations of LC/MS

Gas Chromatography (GC)

Choosing the metabolomics platform

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray

Transcription:

Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring MSMS monitoring Sample can be injected as a liquid, a gas or even a solid

Different Types of MS GC-MS - Gas Chromatography MS separates volatile compounds in gas column and ID s by mass LC-MS - Liquid Chromatography MS separates delicate compounds in HPLC column and ID s by mass MS-MS - Tandem Mass Spectrometry separates compound fragments

How Does MS Measure Mass Deflection of ions in electric or magnetic field which is related to mass of ion. Ions need to be charged. Can be either positive or negative.

GC-MS Injection port MS 320 GC 3800 Probe

GC-MS

EI Fragmentation of CH 3 OH CH 3 OH CH 3 OH + CH 3 OH CH 2 O=H + + H CH 3 OH + CH 3 + OH CH 2 O=H + CHO=H + + H

Electron Impact MS of CH 3 OH Molecular ion EI Breaks up Molecules in Predictable Ways

The GC-MS A mixture of compounds is separated by gas chromatography, then identified by mass spectrometry. =>

e vials Liquid sampling syringe Heating block with stirrer

Probe (solid samples) Sample

Probe

Triple Quodrapole From GC Collect full scan or SIM Ion volume Detector Collect ms-ms fragment ions in this quod CID in this quod ms-ms

Mass Spectrometer (Q3) Q2 Q1 Q3 ple from GC Diffusion pump Ion Source tungsten filament Detector photomultip

GC-MS The MS is equipped with a triple quadrupole analyzer and allows several types of mass detection to be performed. Full scan (direct MS) mode is used. Sim can be used MS-MS can be used

Key Components of MS Ionization gas phase ions created in source High vacum creates free path Mass Analyser- sorts ions by M/Z ratio - electric field Detector creates signal multiplication

GC Settings

Varian (Agilent) FactorFour Capillary GC Columns. Capillary column: thin fusedsilica capillary. 50 m in length and 250 µm inner diameter. The stationary phase is CPsil 8, with 5% phenyl 95% dimethyl polysiloxane coated on the inner surface.

Electron Ionization It uses a heated filament to produce electrons. The filament is usually made of rhenium or tungsten. Once the electrons are produced they are accelerated through a potential difference of around 70V this gives electrons with 70eV of energy.

Electron Impact Electrons produced by the source will then collide with the sample and remove an electron to give ions e- + M M+. + 2e- Fragmentation is a result of an excited molecular ion, which in attempting to gain stability Fragmentation is not random and occurs through real chemical reactions.

Caffeine fragmentation pattern GC-MS 70eV

Caffeine Fragmentation Pattern 109 55 67 82 194 44 137 165

Q1 Window Ion Range

Various constituents

Quantitation of Ingredients You usually quantitate the ingredient that you are interested in by: SIM which measures one of the ions in your sample or MSMS which measures one of the ions produced in the collission cell (Quad 2)

Mass 294 Q1 only Single ion monitoring SIM or SIR

Fractionating the 294 ion in Quad 2 will produce product ions which can be scanned at 100 to 300 to identify the product ions. MW 294 - molecular ion Product ions

MSMS product ions 183-185

Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS LC seperates small, volatile, polar and nonpolar material MS is detection devise (Agilent 500-MS IonTrap (IT) Mass Spectrometer Full scan monitoring SIM single ion monitoring MSMS monitoring Liquid samples are analysed

Agilent LCMS The LCMS can be used either with the LC seperating the sample on the HPLC column before introduction into MS or Stand alone MS where sample is introduced via the syringe pump.

mixer solvents LC-MS Column ES Pumps Syringe pump Sample rack (autosampler) Injection valve

From HPLC or Syringe pump Ion guid ES unit Hexapo N 2 drying gas @150-300 O C Scimmer cone

Electrospray Ionization During standard electrospray ionisation the sample is dissolved in a polar, volatile solvent and pumped through a narrow, stainless steel capillary (needle). A high voltage of 3 to 6 kv is applied to the tip of the needle, which is situated within the ionisation source of the mass spectrometer As a consequence of this strong electric field, the sample emerging from the tip is dispersed into an aerosol of highly charged droplets,

Electrospray Ionization The charged droplets diminish in size by solvent evaporation, (N 2 drying gas) Charged sample ions, free from solvent, are released from the droplets, Some pass through an orifice in the cone into an intermediate vacuum region, Then into the analyser of the mass spectrometer, which is held under high vacuum.

Electrospray Ionization Spray Needle Solvent droplets N 2 flow Cone Liquid from Syringe pump or LC Capillary To MS Atmospheric Pressure

Electrospray Ionization (ESI)

500-MS Operation

500 MS Operation Gas phase ions are generated from solution at atmospheric pressure using either Electrospray Ionization (ESI) or Atmospheric Pressure Chemical Ionization (APCI). The ions are transported through two vacuum interfaces via a metal capillary tube.

500-MS Operation Continued After exiting the capillary, the ion stream expands into a supersonic beam. The sample passes through a skimmer cone (which is at ground potential). If the potential difference between the capillary and skimmer cone is high enough, some molecules may fragment (Capillary Induced Dissociation (CID)). After passing through the skimmer cone, ions enter an 8 m Torr vacuum chamber containing ion optics. The ions enter a hexapole ion guide. The ion guide uses an oscillating potential across six cylindrical rods to confine the ions and a direct current (dc) potential to facilitate ion transport. Collisions with neutral gas in the ion guide reduce the kinetic energy of the ions to facilitate focusing. The ions pass through the ion guide exit and focus lenses, The ions then pass through a split cylinder that acts like a gate for the ion trap by either focusing the beam or deflecting it. The final end cap lens focuses the ions through the entrance hole of the trap.

Electron Multiplier

Ion Trap

Operation of Ion Trap There is a hole in the center of both the entrance and exit end cap electrodes. This hole allows the ions to enter the trap and exit to the detector. The holes in the edge of the end caps contain posts that make contact with pogo pins that carry supplemental waveform signals. An RF generator provides a high voltage, 781 khz RF, that is applied to the ring electrode. With the proper RF voltage, the ion trap electrodes create a three-dimensional, hyperbolic electric field. This field traps the ions in stable orbits. In the presence of helium damping gas, the ions are cooled towards the center of the trap. As the RF voltage increases, the ion trajectories become unstable in increasing order of mass to charge ratio. The ion trap ejects the ions and sends them to the conversion dynode and then to the electron multiplier for detection.

500-MS Detector

Ion Detection fter exiting the trap, ions accelerate toward an off axis conversion dynode that gene combination of positive ions and electrons through secondary electron emission. or the detection of positive ions, the conversion dynode is set to a large negative vo typically -15 kv). The secondary electrons are attracted to the relatively positive mul lectrons or ions emitted from the conversion dynode strike the cathode with sufficie elocity to dislodge additional electrons from the inner curving surface of the cathod ncreasingly positive potential gradient draws the ejected electrons into the electron ultiplier, further accelerating them in the process., the ejected electrons strike the c nner surface of the multiplier, resulting in the emission of more electrons. his configuration produces a cascade of electrons that accelerate toward ground otential at the exit end of the cathode. he anode collects the electrons and passes the resulting ion signal to the ion ampli he ion current is proportional to the total number of ions the ion trap ejects. ach electron or positive ion that enters the electron multiplier generates approximat 0^5 electrons.

Separation Of Analgesics Seperation is based on HPLC and mixture of Tylenol (acetaminophen, caffeine) and aspirin are separated. MS substance in mixture, can be optimized for each of RF, needle voltage, capillary voltage and CID.

Seperation Under Optimum Conditions Precursor (parent) ion produces breakdown ions (daughter ions) which can be used to quantitate the substances in a HPLC separated mixture.

Caffeine fragmentation pattern