Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Similar documents
Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Headspace Technology for GC and GC/MS: Features, benefits & applications

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Application Note. Abstract. Authors. Introduction

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Application. Gas Chromatography February Introduction

USP<467> residual solvents

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Determination of Volatile Substances Proof of Food Adulteration

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Study of Residual Solvents in Various Matrices by Static Headspace

Analysis of BTEX in Natural Water with SPME

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

GAS CHROMATOGRAPHY (GC)

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns Advanced Technology for Fast, Reliable Measurement of Alcohol in Blood

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Introduction to Capillary GC

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

NEW Rtx -BAC Plus 1 and Rtx -BAC Plus 2 Columns

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Application Note. Abstract. Authors. Introduction. Food Safety

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Volatile Organic Compounds in Every Day Food

Analytical Trap Comparison for USEPA Method 8260C

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Sulfotepp impurities in Chlorpyrifos EC formulations

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Evaluation of Capillary Columns for General Performance Parameters

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Author. Abstract. Introduction

METHANOLYSIS OF ACETAL

Methanol Extraction of high level soil samples by USEPA Method 8260C

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

*Correspondence to:

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

NCASI METHOD CI/SG/PULP-94.02

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Understanding Gas Chromatography

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

GC/MS Application Note

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Expectations for GC-MS Lab

Chemistry Gas Chromatography: Separation of Volatile Organics

A biphasic oxidation of alcohols to aldehydes and ketones using a simplified packed-bed microreactor

Certified Reference Material - Certificate of Analysis

Transcription:

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench Application Note Forensic Toxicology and Drug Testing Author Jared Bushey Agilent Technologies, Inc. 285 Centerville Rd. Wilmington, DE 1988 USA Abstract Preparing a set of calibration standards is useful for determining the amount of analyte in an unknown sample and for establishing the linearity of a system. Manually preparing the standards can be time consuming and susceptible to error. Automated preparation is more time efficient and can avoid potential errors introduced by manual methods. Previous work has demonstrated the use of an Agilent 7696 WorkBench for automating a serial dilution protocol [1]. This article describes the use of the Agilent 7696 WorkBench to successfully prepare a set of blood alcohol calibration standards for analysis by an Agilent 7697 Headspace Sampler. The standards will be prepared in 2-mL vials which will themselves be placed within 2-mL headspace vials. The analysis of the standards prepared by the WorkBench system is compared to the analysis of standards prepared manually. The results show that there is no sacrifice in area precision or system linearity performance from using an automated preparation.

Introduction Calibration solutions are used extensively in forensic applications such as blood alcohol (BAC) analyses. While a BAC headspace (HS) and gas chromatographic (GC) analysis may be relatively short, the preparation of the necessary calibration standards can be time consuming. By using an automated preparation method a significant time savings can be realized by allowing the user to address other tasks while the samples are being prepared. The Agilent 7696 WorkBench was designed for automated sample preparation and uses 2-mL vials commonly used in liquid auto-samplers. The 7697 headspace sampler can use 1, 2, and 22-mL headspace vials. A sample volume of ~ 1 ml is more than sufficient to generate enough usable signal for HS GC analysis. Therefore, the volumes of the calibration standards prepared by the 7696 WorkBench are adequate for headspace analysis. The challenge lies in how to match the two platforms into one analytical solution. Automating the dilution preparation is beneficial not only for reducing the amount of chemicals that need to be used, since larger volume volumetric glassware is replaced with 2-mL vials, but also for reducing the amount of human intervention and its concomitant potential for user error. The design of the Agilent 7697 headspace and 7696 WorkBench vials are such that a 2-mL 7696 vial can be completely inserted into a 7697 vial. Operation of the 7697 relies on the sampling of the gaseous headspace within each vial, and not on the sample probe directly contacting the sample solution. Placing a 2-mL 7696 vial into a 2-mL 7697 vial ensures that the 7697 probe will not impact the 2-mL vial during sampling, which could potentially damage the probe. Static headspace analysis involves the sampling of volatile components out of a nonvolatile matrix once the components have reached a thermodynamic equilibrium between the two phases. Heating the sample containing vial increases the amount of volatile analyte in the vial's headspace allowing for better detection response. To decrease the amount of heating time necessary to achieve a given analytical response, headspace vials are often agitated or shaken during heating. The precise control of both sample heating and shaking, along with pneumatic and temporal control, give the 7697 Headspace Sampler industry leading precision and linearity response. The results described in this application note show that the use of 7696 2-mL vials within 7697 2-mL vials does not compromise the area precision or linearity performance of the 7697 Headspace Sampler. In addition, the results also show that the 7696 WorkBench can be used to accurately prepare calibration standards for headspace applications. Experimental Equipment Agilent 7696 WorkBench (1-µL syringe, back tower and 5-µL syringe, front tower) Agilent 7697 Headspace Sampler, high capacity Agilent 789 Gas Chromatograph 25-mL and 1-mL volumetric flasks, ±.12 ml and ±.8 ml, respectively Gilson pipetman pipets Sample.5,.1,.5,.1, and.2% each vol/vol: methanol (MeOH), ethanol (EtOH), acetone, isopropanol (IPA), acetonitrile, ethyl acetate, methyl-ethyl ketone (MEK) were made with both manual and automated preparation from a.5% stock solution per the steps in Tables 1 and 2, respectively. The.5% stock solution was prepared by adding 1.25 ml of each of the analytes (neat) to a 25-mL volumetric flask and diluting to the mark with water. For the manual preparation, 1.5-mL aliquots of the final 1-mL solution of each concentration were manually pipeted Table 1. Vol prepared (ml) Volume of.5% stock.2 1 4 6.1 1 2 8.5 1 1 9.1 1 2. 98.5 1 1. 99 Table 2. Manual Sample Preparation Protocol Automated Sample Preparation Protocol Vol prepared (ml) Volume of.5% stock Water (ml) Water (ml).2 1.5.6.9.1 1.5.3 1.2.5 1.5.15 1.35.1 1.5.3 1.47.5 1.5.15 1.49 2

into a 2-mL headspace vial and promptly capped. In total, six headspace vials were prepared for each dilution level for a total of 36 vials. For the WorkBench preparation, the 1.5-mL dilution levels prepared for each concentration were kept in the 2-mL vials they were prepared in, but their 2-mL screw caps were removed, and each vial was placed in its own 2-mL headspace vial which was promptly capped after receiving its 2-mL vial. As with the manual prep, six headspace vials (with 2-mL 7696 vials inside) were prepared for each dilution level for a total of 36 vials, therefore the WorkBench method was repeated six times to generate the necessary number of samples. For each preparation procedure, one dilution series was analyzed sequentially (that is,.5% to.5%) before any one dilution level was repeated. Two water blanks were run between each dilution series. The headspace and gas chromatograph parameters are given in Table 3: Table 3. HS and GC Parameters Agilent 7697 Headspace parameters Temperatures Oven = 6 C, Loop = 6 C, Tr line = 1 C Times Vial equilibration = 35 min, GC cycle = 6 min, pressure equib. =.1 min, Inj =.5 min Vial Mode = flow-to-pressure, Initial pressure = 15psi, Fill flow = 5 ml/min, Loop fill mode = Custom, Loop fill rate = 2 psi/min, Final loop pressure = 1 psi, Loop equib =.5 min, Vial = 2 ml, Shaking = 1 Carrier GC control Adv func Purge flow = 1 ml/min, Purge time = 1 min Config Vial pressurization gas type = He, Sample loop volume = 1 ml, Standby flow = 2 ml/min Agilent 789 GC parameters Inlet S/Sl, 5183-4647 liner (wool removed), 2 C, 33.55 psi, 3 ml/min septum purge, split = 5:1 Columns DB-ALC2 (J&W 123-9234),.32 mm 3 m, 1.2 µm, constant flow = 12 ml/min Dual column configuration for 7696-prepared samples: 16-cm piece of 53 µm deactivated fused silica (16-2535) connected the inlet to a passive CFT splitter. The splitter was then connected to a DB-ALC2 and a DB-ALC1 column. DB ALC1 (J&W 123-9134),.32 mm 3 m, 1.8 µm, constant flow = 12 ml/min DB-ALC2 (J&W 123-9234),.32 mm 3 m, 1.2 µm, constant flow = 12 ml/min Oven 35 C, hold 4 min Detector FID, 25 C Signal FID, 5 Hz (front or back detector) 3

Results and Discussion The samples prepared by the Agilent 7696 WorkBench were analyzed with the Agilent 7697 Headspace Sampler using the dual column (DB-ALC1 and DB-ALC2 columns) setup in the Agilent 789 GC. The manually prepared samples were analyzed using a standard single column (DB-ALC2) setup in the 789 GC. Representative chromatograms acquired using the dual column system are shown in Figure 1. pa 3 25 2 15 1 5 DB-ALC2 1 2 3 4 5 6 7 Peak identification 1. Methanol 2. Ethanol 3. Acetone 4. Isopropanol 5. Acetonitrile 6. Ethyl acetate 7. Methyl-ethyl ketone To ensure the HS vials reached thermal equilibrium, the Parameter Increment feature of the Agilent 7697HS Sampler was used to determine at what Vial Equilibration time a constant area response was observed for all analytes. The Vial Equilibration time determination was done using the.5% dilution level in an experiment separate from the dilution series evaluation and the results are shown in Figure 2. Normalized area (pa*s) 1.6 1.4 1.2 1..8.6.4.2 5 1 15 2 25 3 35 4 Vial equib time at 6 C (min) A MeOH EtOH Acetone Isopropanol Acetonitrile Ethyl Acetate MEK 1.6 B pa 3 25 2 15.5 1. 1.5 2. 2.5 3. 3.5 min 3,5 DB-ALC1 6 7 Normalized area (pa*s) 1.4 1.2 1..8.6.4.2 MeOH EtOH Acetone Isopropanol Acetonitrile Ethyl Acetate MEK 1 5 1 2 4 5 1 15 2 25 3 35 4 Vial equib time at 6 C (min) Figure 1..5 1. 1.5 2. 2.5 3. 3.5 min Representative chromatograms of BAC mixture prepared by the Agilent 7696 WorkBench. The data in Figure 1 represent analytes at the.1% concentration level prepared by the 7696 WorkBench. Each chromatogram is actually six overlaid traces from the replicate injections, illustrating the system s precision. Figure 2. Peak area as a function of Agilent 7697 Headspace Sampler vial equilibration time. A. 1.5 ml of sample in a 2-mL HS vial. Average of three trials. B. 1.5 ml of sample in a 2-mL vial within a 2-mL HS vial. Average of three trials. The experimental results in Figure 2 were used to determine that a 35-min vial equilibration in the 7697 oven was sufficient for the headspace vials containing 2-mL ALS vials. At an equilibration time of 35 min, the area precision that was observed indicated that a thermal equilibrium had been reached. In addition, Figure 2 also shows that a 35-min vial equilibration time was not detrimental to the analysis of headspace vials containing only liquid sample. 4

Manual Prep Using a 35-min vial equilibration time, the manually prepared dilution series was analyzed on the 7697 Headspace Sampler using the conditions listed in Table 3. The resulting calibration curve is shown in Figure 3. Table 4 contains the regression values for the manually prepared samples. The results in Figure 3 represent the average areas for six replicate measurements of each concentration level with 1.5 ml of each sample placed directly into the HS vial. Based on the data in Figure 3, the precision for each analyte at each concentration level and the normalization of ethanol to the internal standard, isopropanol, are given in Table 5. Table 4. Data for Figure 3 1 3 5 45 4 35 3 25 2 15 1 5.5.15.25.35.45 Average area, pa*s (n = 6) Figure 3. MeOH EtOH Acetone Isopropanol Acetonitrile Ethyl Acetate MEK Agilent 7697 Headspace Sampler analysis of manually prepared BAC calibration standards. Analyte m b R 2 MeOH 26.61 4.25.99999 EtOH 49.95 19.23.99998 Acetone 196.8 178.7.99963 Isopropanol 83.13 71.332.99984 Acetonitrile 11.5 48.276.99995 Ethyl acetate 456.69 524.77.99945 MEK 373.91 451.77.99942 The bottom table of Table 5 illustrates the precision of ethanol analysis when an internal standard, in this case, isopropanol is used. The data shows that across the entire calibration range no trending is observed that is unique to any one analyte. For example, if trending was occurring in the ethanol area response then the same magnitude of trending was also occurring with the isopropanol internal standard. Table 5. Precision of BAC Analysis of Each Manually Prepared Concentration Level Area precision n = 6 Conc % MeOH EtOH Acetone Isopropanol Acetonitrile Ethyl acetate MEK.5 1.4 1.1.54.53.57 1.8.85.1 3.3 3.4 1.1 2.2 1.7 1.3.89.5 1.1.93.45.6.53.9.52.1 1.1.98.46.66.58.89.5.2 2.4 1.9.46 1.1.77.83.54.5 1.8 1.8.78 1.6 1.2 1.1.72 EtOH/IPA ratio Conc % Trial 1 Trial 2 Trial 3 Trial 4 Trial 5 Trial 6 Avg Stdev RSD %.5.56.56.56.57.57.56.56.35.62.1.55.55.56.57.56.56.56.76 1.4.5.56.56.56.56.56.56.56.19.34.1.58.58.58.58.59.58.58.2.35.2.58.59.59.59.59.59.59.51.87.5.6.6.6.6.6.6.6.19.32 5

Workbench prep DB-ALC2 Using a 35-min vial equilibration time, the Agilent 7696 prepared dilution series was analyzed on the Agilent 7697 Headspace Sampler using the conditions listed in Table 3. The calibration curve and its regression values are given in Figure 4 and Table 6 for the analysis run on the DB-ALC2 column. The points in the curves represent the average area response of the six replicate analyses made at each concentration level. The linear regression values show that linearity was maintained across the entire calibration range for all analytes. Table 7 contains the area precision for each analyte at each concentration level and the normalization of ethanol to the internal standard isopropanol for the DB-ALC2 analyses. Table 6. Linear Regression Results for Automated Sample Preparation from the DB-ALC2 Column Analyte m b R 2 MeOH 22.545-7.2643.99992 EtOH 41.948-3.8325.99999 Acetone 172.1-5.854.99997 IPA 66.71 43.59.99964 Acetonitrile 91.56-11.48.99999 Ethyl acetate 365.62-14.323 1. MEK 312.3 9.117.99991 The data illustrates a consistent analyte response for different compounds, independent of analyte concentration. The data in the bottom panel of Table 7 again shows that ethanol and isopropanol trend together with respect to area response as a function of concentration, rather than independently. This result shows that the use of the automated sample preparation procedure has not introduced any concentration biases due to erroneous sample handling. 1 3 4 Average area, pa*s (n = 6) 35 3 25 2 15 1 5 Figure 4..5.15.25.35.45 MeOH EtOH Acetone Isopropanol Acetonitrile Ethyl Acetate MEK Agilent 7697 Headspace Sampler analysis of BAC standards prepared on the Agilent 7696 WorkBench and analyzed with a DB-ALC2 column. Table 7. Precision Performance of the Agilent 7697 Headspace Sampler for the Analysis of BAC Standards Prepared on the Agilent 7696 WorkBench and Analyzed with a DB-ALC2 Column Area precision n = 6 Conc % MeOH EtOH Acetone Isopropanol Acetonitrile Ethyl acetate MEK.5 1.2 1.1 1.5 1.4 1.5 1.8 1.6.1.52.76.84.7.59 1.8 1.1.5.44.5.53.49.37 1.3.69.1.62.59.58.59.59 1..66.2 2.2 1.8.75 1.2 1.3 1.4.83.5.65.56.6.62.48 1.1.75 EtOH/IPA ratio Conc % Trial 1 Trial 2 Trial 3 Trial 4 Trial 5 Trial 6 Avg Stdev RSD %.5.57.57.57.57.57.58.57.24.43.1.57.58.58.57.58.58.58.18.31.5.57.57.57.57.57.58.57.14.25.1.57.57.57.57.57.57.57.11.2.2.62.62.62.62.62.61.62.48.77.5.63.62.63.62.63.63.63.13.2 6

Workbench Prep DB ALC1 The results from analyzing the Agilent 7696 prepared dilution series on the Agilent 7697 Headspace Sampler using the DB-ALC1 column under the conditions listed in Table 3 are shown in Figure 5. Due to the nature of the DB-ALC1 column, acetonitrile and acetone are not able to be separated and methyl ethyl ketone and ethyl acetate are not baseline resolved, as observed in Figure 1. The data points in the curves represent the average area response of the six replicate analyses made at each concentration level. The linear regression results are given in Table 8, and again they show that over the range tested, the analytical response for each analyte was linear. Table 9 contains the precision for each analyte at each concentration level and the normalization of ethanol to the internal standard isopropanol for the DB-ALC1 analyses. Table 8. Linear Regression Results for Automated Sample Preparation from the DB-ALC1 Column Analyte m b R 2 Consistent with the previous results, the data in Table 9 shows acceptable raw area precision for each of the compounds. The ethanol area normalized to the iso propanol area displays the same precision as the results from the DB-ALC2 column as well as from the analysis of the manually prepared samples. 1 3 5 Average area, pa*s (n = 6) 45 4 35 3 25 2 15 1 5.5.15.25.35.45 MeOH EtOH Isopropanol Acteone/ Acetonitrile Ethyl acetate MEK MeOH 3.515-13.55.9998 EtOH 52.77-12.55.99996 Isopropanol 86.119 -.42683 1. Acetone/Acetonitrile 316.62-27.949 1. MEK 442.2-656.6.9987 Ethyl acetate 431.5 555.17.99813 Figure 5. Agilent 7697 Heasdspace Sampler analysis of BAC standards prepared on the Agilent 7696 WorkBench and analyzed with a DB-ALC1 column. Table 9. Precision Performance of the Agilent 7697 Headspace Sampler for the Analysis of BAC Standards Prepared on the Agilent 7696 WorkBench and Analyzed with a DB-ALC1 Column Area precision n = 6 Conc % MeOH EtOH Isopropanol Acetone/Acetonitrile MEK Ethyl acetate.5 1.2 1.3 1.1 1.4 1.5 1.8.1.4.72.73.72 1. 1.8.5.53.42.46.44.69 1.3.1.52.55.55.53.64 1.1.2 2.3 1.8 1.2.83.95 1.5.5.52.54.59.54 1..96 EtOH/IPA ratio Conc % Trial 1 Trial 2 Trial 3 Trial 4 Trial 5 Trial 6 Avg Stdev RSD %.5.59.6.59.59.59.59.59.24.41.1.6.59.59.59.59.6.6.16.27.5.6.6.6.6.6.6.6.15.26.1.6.6.6.6.6.6.6.14.23.2.6.6.6.61.6.59.6.47.78.5.61.61.61.61.61.61.61.13.2 7

Summary The results in Figures 4 and 5 were acquired during the same GC runs using a passive CFT splitter and the dual column arrangement described in Table 3. The dual column configuration produced accurate and precise data and demonstrates the utility of using the Agilent 7696 WorkBench to prepare standards for a common forensics (blood alcohol analysis) application. To summarize the data presented above, four key metrics have been accumulated in Table 1. Table 1. Summary Table Comparing Results From Analyzing Samples Prepared Manually to Those Prepared Using the Automated 7696 System EtOH area precision.5.1.5.1.2.5 Manual prep (DB-ALC2) 1.1 3.4.93.98 1.9 1.8 Automated prep (DB-ALC2) 1.1.76.5.59 1.8.56 Automated prep (DB-ALC1) 1.3.72.42.55 1.8.54 IPA area precision.5.1.5.1.2.5 Manual prep (DB-ALC2).53 2.2.6.66 1.1 1.6 Automated prep (DB-ALC2) 1.42.7.49.59 1.2.62 Automated prep (DB-ALC1) 1.1.73.46.55 1.2.59 EtOH/IPA ratio precision.5.1.5.1.2.5 Manual prep (DB-ALC2).62 1.4.34.35.87.32 Automated prep (DB-ALC2).43.31.25.2.77.2 Automated prep (DB-ALC1).41.27.26.23.78.2 Linearity (R 2 ) MeOH EtOH Acetone Isopropanol Acetonitrile Ethyl acetate MEK Manual prep (DB-ALC2).99999.9998.99963.99984.99995.99945.99942 Automated prep (DB-ALC2).99992.99999.99997.99964.99999 1..99991 Automated prep (DB-ALC1).9998.99996 not resolved 1. not resolved.99813.9987 The data in Table 1 shows that for ethanol, isopropanol, and their area ratio the 7696 prepared samples provide more accurate data than the manually prepared alternative except at the lowest concentration level,.5%. However, it can be seen in Table 1 that at a concentration of.5% the 7696 WorkBench prepared samples still gave area precisions < 1.5% for both ethanol and iso propanol. The linear regression results summarized in Table 1 also illustrate that the linearity performance was unaffected by the type of sample preparation employed or the type of column configuration used. 8

Conclusion 1. Calibration standards prepared by an Agilent WorkBench demonstrate the same Agilent 7697 Headspace Sampler linearity as manually prepared standards. 2. The same peak area precision was obtained from samples prepared by WorkBench compared to ones prepared manually. 3. The 2-mL vials prepared by WorkBench can be nested within 2-mL headspace vials and successfully analyzed using the 7697 sampler. 4. The nesting of a 2-mL vial within a 2-mL headspace vial will increase the amount of vial equilibration time or agitation needed to achieve a thermodynamic equilibrium between the liquid in the 2-mL vial and the headspace in the 2-mL vial. The impact on equilibration time and agitation level will depend on the method s headspace oven temperature, characteristics of the glass vials, and the properties of the chemical analytes and solvents under test. References 1. W. Dale Snyder, Agilent 7696A Sample Prep WorkBench: How to Automate Preparation of a Sample Set by Serial Dilution for Measurement of Flame Ionization Detector Performance, Agilent Application Note, Publication number 599-685EN, 21 For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 9

www.agilent.com/chem For Forensic Use Only. Information is subject to change without notice. Agilent Technologies, Inc., 211 Printed in the USA September 28, 211 599-925EN